Very Low-Noise ENG Amplifier System Using CMOS Technology
Abstract
In this paper, we describe the design and testing of a system for recording electroneurographic signals (ENG) from a multielectrode nerve cuff (MEC). This device, which is an extension of the conventional nerve signal recording cuff, enables ENG to be classified by action potential velocity. In addition to electrical measurements, we provide preliminary in vitro data obtained from frogs that demonstrate the validity of the technique for the first time. Since typical ENG signals are extremely small, on the order of 1 1 microV, very low-noise, high-gain amplifiers are required. The ten-channel system we describe was realized in a 0.8 microm CMOS technology and detailed measured results are presented. The overall gain is 10 000 and the total input-referred root mean square (rms) noise in a bandwidth 1 Hz-5 kHZ is 291 nV. The active area is 12 mm(2) and the power consumption is 24 mW from +/-2.5 V power supplies.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The choice of front-end amplifier gain is influenced by the required application, whether CNS or PNS recording, and by the system's power supply.
“The choice of gain is influenced by the required application (CNS or PNS recording) and the system’s power supply [107].”
From Implantable neurotechnologies: a review of integrated circuit neural amplifiers (2016) · cites this paper as [107]
Cites
(1988) · Integration · doi:10.1016/0167-9260(88)90044-2 · Cited by 3,213 · see citing works
A theory of the effects of fibre size in medullated nerve
Rushton, W. A. H. (1951) · The Journal of Physiology · doi:10.1113/jphysiol.1951.sp004655 · Cited by 1,075 · see citing works
A micropower low-noise monolithic instrumentation amplifier for medical purposes
Steyaert, M. S. J., & Sansen, W. M. C. (1987) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.1987.1052869 · Cited by 577 · see citing works
VLSI design techniques for analog and digital circuits
(1990) · Choice Reviews Online · doi:10.5860/choice.27-2739 · Cited by 382 · see citing works
VLSI design techniques for analog and digital circuits
(1992) · Microelectronics Journal · doi:10.1016/0026-2692(92)90033-w · Cited by 350 · see citing works
Micromachined, Polyimide-Based Devices for Flexible Neural Interfaces
Stieglitz, T., Beutel, H., Schuettler, M., & Meyer, J.-U. (2000) · Biomedical Microdevices · doi:10.1023/a:1009955222114 · Cited by 343 · see citing works
Self-citationMOS transistors operated in the lateral bipolar mode and their application in CMOS technology
Vittoz, E. A. (1983) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.1983.1051939 · Cited by 247 · see citing works
An implantable CMOS circuit interface for multiplexed microelectrode recording arrays
Ji, J., & Wise, K. D. (1992) · IEEE Journal of Solid-State Circuits · doi:10.1109/4.121568 · Cited by 173 · see citing works
Cutaneous whole nerve recordings used for correction of footdrop in hemiplegic man
Haugland, M. K., & Sinkjaer, T. (1995) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.481970 · Cited by 168 · see citing works
Waters, R. L., McNeal, D. R., Faloon, W., & Clifford, B. (1985) · The Journal of Bone & Joint Surgery · doi:10.2106/00004623-198567050-00016 · Cited by 156 · see citing works
Skin contact force information in sensory nerve signals recorded by implanted cuff electrodes
Haugland, M. K., Hoffer, J. A., & Sinkjær, T. (1994) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.296346 · Cited by 122 · see citing works
A CMOS IC for portable EEG acquisition systems
Martins, R. E., Selberherr, S., & Vaz, F. (1998) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/19.746581 · Cited by 122 · see citing works
A CMOS IC for portable EEG acquisition systems
Martins, R. E., Selberherr, S., & Vaz, F. (2002) · doi:10.1109/imtc.1998.676985 · Cited by 121 · see citing works
Biopotentials as command and feedback signals in functional electrical stimulation systems.
Sinkjær, T., Haugland, M., Inmann, A., Hansen, M., & Nielsen, K. D. (2003) · Medical Engineering and Physics · doi:10.1016/s1350-4533(02)00178-9 · Cited by 118 · see citing works
Haugland, M. K., Lickel, A., Haase, J., & Sinkjær, T. (1999) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.769412 · Cited by 112 · see citing works
Haugland, M. K., & Hoffer, J. A. (1994) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.296345 · Cited by 80 · see citing works
Multiple-electrode nerve cuffs for low-velocity and velocity-selective neural recording
Taylor, J. T., Donaldson, N., & Winter, J. (2004) · Medical & Biological Engineering & Computing · doi:10.1007/bf02347545 · Cited by 79 · see citing works
Self-citationDesign of a low-noise preamplifier for nerve cuff electrode recording
Rieger, R., Taylor, J., Demosthenous, A., Donaldson, N., & Langlois, P. J. (2003) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2003.814437 · Cited by 78 · see citing works
Self-citationPerformance of alternative amplifier configurations for tripolar nerve cuff recorded ENG
Pflaum, C., Riso, R. R., & Wießpeiner, G. (2002) · Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.1996.657000 · Cited by 63 · see citing works
Strange, K. D., & Hoffer, J. A. (1999) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.771189 · Cited by 39 · see citing works
Feasibility of inducing micturition through chronic stimulation of sacral roots
Schmidt, R. A., Bruschini, H., & Tanagho, E. A. (1978) · Urology · doi:10.1016/0090-4295(78)90309-6 · Cited by 37 · see citing works
Rieger, R., Taylor, J. T., Comi, E., Donaldson, N., Russold, M., Mahony, C. M. O., McLaughlin, J. A., McAdams, E. T., Demosthenous, A., & Jarvis, J. C. (2004) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2004.02.002 · Cited by 32 · see citing works
Self-citationPerelman, Y., & Ginosar, R. (2005) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/cne.2005.1419637 · Cited by 11 · see citing works
II. The Comparative Physiological Characteristics of Nerve Fibers
Erlanger, J., & Gasser, H. S. (1937) · University of Pennsylvania Press eBooks · doi:10.9783/9781512819441-006 · Cited by 4 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Recent Advances in Neural Recording Microsystems
Gosselin, B. (2011) · Sensors · doi:10.3390/s110504572 · Cited by 138 · see citing works
Implantable neurotechnologies: a review of integrated circuit neural amplifiers
Ng, K. A., Greenwald, E., Xu, Y. P., & Thakor, N. V. (2016) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-015-1431-3 · Cited by 83 · see citing works
Fibre-selective recording from the peripheral nerves of frogs using a multi-electrode cuff
Schuettler, M., Donaldson, N., Seetohul, V., & Taylor, J. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/3/036016 · Cited by 50 · see citing works
Self-citationYoshida, K., Kurstjens, G. A. M., & Hennings, K. (2009) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2009.08.005 · Cited by 43 · see citing works
Peripheral Neural Activity Recording and Stimulation System
Loi, D., Carboni, C., Angius, G., Angotzi, G. N., Barbaro, M., Raffo, L., Raspopovic, S., & Navarro, X. (2011) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2011.2123097 · Cited by 42 · see citing works
Intraneural active probe for bidirectional peripheral nerve interface
Ballini, M., Bae, J., Marrocco, N., Verplancke, R., Schaubroeck, D., Cuypers, D., Cauwe, M., O'Callaghan, J., Fahmy, A., Maghari, N., et al. (2017) · 2017 Symposium on VLSI Circuits · doi:10.23919/vlsic.2017.8008544 · Cited by 16 · see citing works
- Authors
- Rieger, R., Schuettler, M., Pal, D., Clarke, C., Langlois, P., Taylor, J., & Donaldson, N.
- Affiliations
- National Sun Yat-sen University (TW)
- Birla Institute of Technology, Mesra (IN)
- University of Bath (GB)
- University College London (GB)
- Venue
- IEEE Transactions on Neural Systems and Rehabilitation Engineering
- Published
- 2006-12
- Volume
- 14
- Issue
- 4
- Pages
- 427-437
- ISSN
- 1534-4320 (ISSN-L)
- 1558-0210
- OpenAlex
- W2151735029
- Cited by
- 57
- References
- 24
- Citations by year
- 2024: 2 · 2020: 1 · 2019: 2 · 2017: 3 · 2016: 3 · 2015: 1 · 2014: 5 · 2013: 4 · 2012: 5
- FWCI
- 4.60
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- IEEE Transactions on Neural Systems and Rehabilitation Engineering (publishedVersion)
- PubMed (publishedVersion)
- UCL Discovery (University College London) (submittedVersion)
- Funders
- Engineering and Physical Sciences Research Council (EP/D03146X/1, GR/S93783/01)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Analog and Mixed-Signal Circuit Design
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Radio Frequency Integrated Circuit Design
- Keywords
- electroneurography, low noise amplifier, CMOS integrated circuit, high-gain amplifier
- MeSH terms
- Action Potentials, Amplifiers, Electronic, Diagnostic Techniques, Neurological, Electrodes, Implanted, Equipment Design, Microelectrodes, Peripheral Nerves, Semiconductors, Sensitivity and Specificity, Reproducibility of Results, Computer-Aided Design, Equipment Failure Analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review