Peripheral Neural Activity Recording and Stimulation System
Open accessAbstract
This paper presents a portable, embedded, microcontroller-based system for bidirectional communication (recording and stimulation) between an electrode, implanted in the peripheral nervous system, and a host computer. The device is able to record and digitize spontaneous and/or evoked neural activities and store them in data files on a PC. In addition, the system has the capability of providing electrical stimulation of peripheral nerves, injecting biphasic current pulses with programmable duration, intensity, and frequency. The recording system provides a highly selective band-pass filter from 800 Hz to 3 kHz, with a gain of 56 dB. The amplification range can be further extended to 96 dB with a variable gain amplifier. The proposed acquisition/stimulation circuitry has been successfully tested through in vivo measurements, implanting a tf-LIFE electrode in the sciatic nerve of a rat. Once implanted, the device showed an input referred noise of 0.83 μVrms, was capable of recording signals below 10 μ V, and generated muscle responses to injected stimuli. The results demonstrate the capability of processing and transmitting neural signals with very low distortion and with a power consumption lower than 1 W. A graphic, user-friendly interface has been developed to facilitate the configuration of the entire system, providing the possibility to activate stimulation and monitor recordings in real time.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Amplifier noise is typically designed to be lower than electrode noise and is maintained below 5 μVrms for CNS-recording amplifiers and below 1 μVrms for PNS-recording amplifiers.
“The noise introduced by the amplifier is typically designed to be lower than that introduced by the electrodes and, for a majority of neural applications, is maintained below 5 µVrms for amplifiers used to record CNS signals and less than 1 µVrms for PNS-recording amplifiers [76, 105].”
From Implantable neurotechnologies: a review of integrated circuit neural amplifiers (2016) · cites this paper as [76]
CNS-based neuroprostheses, or brain-machine interfaces, decode neural electrical activity to control robotic limbs or to stimulate muscles for restoring limb function.
“CNS-based neuroprostheses, or brain–machine interfaces (BMIs), are systems whereby neural electrical activities are decoded and used to control robotic limbs [76, 89, 105] or to stimulate muscles for restoring limb function [35].”
From Implantable neurotechnologies: a review of integrated circuit neural amplifiers (2016) · cites this paper as [76]
Cites
A low-power low-noise cmos for amplifier neural recording applications
Harrison, R. R., & Charles, C. (2003) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2003.811979 · Cited by 1,695 · see citing works
Navarro, X., Krueger, T. B., Lago, N., Micera, S., Stieglitz, T., & Dario, P. (2005) · Journal of the Peripheral Nervous System · doi:10.1111/j.1085-9489.2005.10303.x · Cited by 845 · cited 2× · see citing works
“Moreover, tf-LIFEs have 8 independent channels and, therefore, allow simultaneous recordings and/or simulations [ 2 ].”
BackgroundSelf-citationDirect neural sensory feedback and control of a prosthetic arm
Dhillon, G. S., & Horch, K. W. (2005) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2005.856072 · Cited by 573 · see citing works
An Energy-Efficient Micropower Neural Recording Amplifier
Wattanapanitch, W., Fee, M., & Sarpeshkar, R. (2007) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2007.907868 · Cited by 441 · see citing works
Double nerve intraneural interface implant on a human amputee for robotic hand control
Rossini, P. M., Micera, S., Benvenuto, A., Carpaneto, J., Cavallo, G., Citi, L., Cipriani, C., Denaro, L., Denaro, V., Di Pino, G., et al. (2010) · Clinical Neurophysiology · doi:10.1016/j.clinph.2010.01.001 · Cited by 396 · see citing works
Dhillon, G. S., Lawrence, S. M., Hutchinson, D. T., & Horch, K. W. (2004) · The Journal of Hand Surgery · doi:10.1016/j.jhsa.2004.02.006 · Cited by 354 · cited 1× · see citing works
“2123097 from small clusters of motor and sensory neurons at a subfascicular level [3], [4].”
BackgroundThe 128-Channel Fully Differential Digital Integrated Neural Recording and Stimulation Interface
Shahrokhi, F., Abdelhalim, K., Serletis, D., Carlen, P. L., & Genov, R. (2010) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2010.2041350 · Cited by 263 · see citing works
Switch-Matrix-Based High-Density Microelectrode Array in CMOS Technology
Frey, U., Šedivý, J., Heer, F., Pedron, R., Ballini, M., Müller, J., Bakkum, D. J., Hafizovic, S., Faraci, F. D., Greve, F., et al. (2010) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2009.2035196 · Cited by 251 · see citing works
Wireless Neural Recording With Single Low-Power Integrated Circuit
Harrison, R. R., Kier, R. J., Chestek, C. A., Gilja, V., Nuyujukian, P., Ryu, S. I., Greger, B. E., Solzbacher, F., & Shenoy, K. V. (2009) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2009.2023298 · Cited by 209 · see citing works
Integrated circuit amplifiers for multi-electrode intracortical recording
Jochum, T., Denison, T., & Wolf, P. D. (2009) · Journal of Neural Engineering · doi:10.1088/1741-2560/6/1/012001 · Cited by 167 · see citing works
A Versatile Integrated Circuit for the Acquisition of Biopotentials
Harrison, R. R. (2007) · 2007 IEEE Custom Integrated Circuits Conference · doi:10.1109/cicc.2007.4405694 · Cited by 156 · see citing works
Implantable biomedical microsystems for neural prostheses
Stieglitz, T., Schuetter, M., & Koch, K. P. (2005) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/memb.2005.1511501 · Cited by 140 · see citing works
Lee, J., Rhew, H.-G., Kipke, D. R., & Flynn, M. P. (2010) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2010.2052403 · Cited by 140 · see citing works
An intrafascicular electrode for recording of action potentials in peripheral nerves
Malagodi, M. S., Horch, K. W., & Schoenberg, A. A. (1989) · Annals of Biomedical Engineering · doi:10.1007/bf02368058 · Cited by 107 · see citing works
Efficiency Analysis of Waveform Shape for Electrical Excitation of Nerve Fibers
Wongsarnpigoon, A., Woock, J. P., & Grill, W. M. (2010) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2010.2047610 · Cited by 104 · see citing works
An implantable wireless bidirectional communication microstimulator for neuromuscular stimulation
Lee, S.-Y., & Lee, S.-C. (2005) · IEEE Transactions on Circuits and Systems I Fundamental Theory and Applications · doi:10.1109/tcsi.2005.857770 · Cited by 104 · see citing works
Micera, S., Citi, L., Rigosa, J., Carpaneto, J., Raspopović, S., Pino, G. D., Rossini, L., Yoshida, K., Denaro, L., Dario, P., et al. (2010) · Proceedings of the IEEE · doi:10.1109/jproc.2009.2038726 · Cited by 100 · see citing works
A system for neural recording and closed-loop intracortical microstimulation in awake rodents.
Venkatraman, S., Elkabany, K., Long, J. D., Yao, Y., & Carmena, J. M. (2009) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2008.2005944 · Cited by 96 · see citing works
Design of a low-noise preamplifier for nerve cuff electrode recording
Rieger, R., Taylor, J., Demosthenous, A., Donaldson, N., & Langlois, P. J. (2003) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2003.814437 · Cited by 78 · see citing works
Characterization of signals and noise rejection with bipolar longitudinal intrafascicular electrodes
Yoshida, K., & Stein, R. B. (1999) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.740885 · Cited by 76 · see citing works
Jia, X., Koenig, M. A., Zhang, X., Zhang, J., Chen, T., & Chen, Z. (2007) · The Journal of Hand Surgery · doi:10.1016/j.jhsa.2007.02.021 · Cited by 70 · see citing works
Ye, X., Wang, P., Liu, J., et al. (2008) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2008.07.002 · Cited by 63 · see citing works
Information contained in sensory nerve recordings made with intrafascicular electrodes
Goodall, E. V., Lefurge, T. M., & Horch, K. W. (1991) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.83604 · Cited by 63 · see citing works
A low power multichannel analog front end for portable neural signal recordings
Obeid, I., Nicolelis, M. A. L., & Wolf, P. D. (2003) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2003.09.024 · Cited by 58 · see citing works
Very Low-Noise ENG Amplifier System Using CMOS Technology
Rieger, R., Schuettler, M., Pal, D., Clarke, C., Langlois, P., Taylor, J., & Donaldson, N. (2006) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2006.886731 · Cited by 57 · see citing works
A combined wireless neural stimulating and recording system for study of pain processing
Ativanichayaphong, T., He, J., Hagains, C. E., et al. (2008) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2007.12.014 · Cited by 50 · see citing works
Acute Performance of the Thin-Film Longitudinal Intra-Fascicular Electrode
Yoshida, K., Hennings, K., & Kammer, S. (2006) · IEEE/RAS-EMBS International Conference on Biomedical Robotics and Biomechatronics (BioRob) · doi:10.1109/biorob.2006.1639102 · Cited by 30 · see citing works
Design strategies for multi-channel low-noise recording systems
Rieger, R., & Taylor, J. T. (2008) · Analog Integrated Circuits and Signal Processing · doi:10.1007/s10470-008-9230-5 · Cited by 28 · see citing works
Jaber, F. T., Labeed, F., & Hughes, M. (2009) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2009.06.013 · Cited by 27 · see citing works
Micro-multi-probe electrode array to measure neural signals
Chen, C., Yao, D., Tseng, S.-H., Lu, S.-W., Chiao, C., & Yeh, S. (2008) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2008.09.027 · Cited by 25 · see citing works
Study of microelectronics for detecting and stimulating of central neural signals
Wang, Z., Lü, X., Li, W., et al. (2005) · doi:10.1109/icnic.2005.1499875 · Cited by 22 · see citing works
Fully Integrated and Low Power CMOS Amplifier for Neural Signal Recording
Wang, Y., Wang, Z., Lü, X., & Wang, H. (2005) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2005.1615663 · Cited by 12 · see citing works
An Autonomous, Broadband, Multi-channel Neural Recording System for Freely Behaving Primates
Linderman, M. D., Gilja, V., Santhanam, G., et al. (2006) · doi:10.1109/iembs.2006.260813 · Cited by 10 · see citing works
A novel system for recording from single neurons in unrestrained animals
Sherk, H., & Wilkinson, E. J. (2008) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2008.06.010 · Cited by 7 · see citing works
Nam, Y., Brown, E. A., Ross, J. D., Blum, R. A., Wheeler, B. C., & DeWeerth, S. P. (2008) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2008.11.017 · Cited by 6 · see citing works
An Embedded Wireless Neural Stimulation and Recording System
Fernando, N. X., Macklin, D. N., Hsu, M. Y., et al. (2007) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/cne.2007.369678 · Cited by 6 · see citing works
Integrated electronics for peripheral nerve recording and signal processing
Limnuson, K., Tyler, D. J., & Mohseni, P. (2009) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2009.5334242 · Cited by 6 · see citing works
Chen, Y., Lee, Y., Yeh, S., & Chen, H. (2009) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner.2009.5109231 · Cited by 5 · see citing works
A PCB SYSTEM IMPLEMENTATION FOR NEURAL SIGNALS RECORDING AND PNS STIMULATION
Loi, D., Carboni, C., Angius, G., & Barbaro, M. (2010) · UNICA IRIS Institutional Research Information System (University of Cagliari) · doi:10.2316/journal.216.2010.7.680-067 · Cited by 2 · see citing works
Self-citationAN ELECTRONIC INTERFACE FOR NEURAL ACTIVITY RECORDING AND STIMULATION
Angius, G., Carboni, C., Loi, D., et al. (2010) · International Conference on Biomedical Electronics and Devices · doi:10.5220/0002749702110214 · Cited by 1 · see citing works
Self-citationA PCB System Implementation for Neural Signals Recording and PNS Stimulation
Loi, D., Carboni, C., Angius, G., et al. (2010) · Control and Intelligent Systems · doi:10.2316/j.2010.216.680-0067 · Cited by 1 · see citing works
Self-citation
Flagged first, then by citations
Cited by
Sorted by citations
Implantable neurotechnologies: a review of integrated circuit neural amplifiers
Ng, K. A., Greenwald, E., Xu, Y. P., & Thakor, N. V. (2016) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-015-1431-3 · Cited by 83 · see citing works
Motor-commands decoding using peripheral nerve signals: a review
Hong, K.-S., Aziz, N., & Ghafoor, U. (2018) · Journal of Neural Engineering · doi:10.1088/1741-2552/aab383 · Cited by 63 · see citing works
- Authors
- Loi, D., Carboni, C., Angius, G., Angotzi, G. N., Barbaro, M., Raffo, L., Raspopovic, S., & Navarro, X.
- Affiliations
- University of Cagliari (IT)
- Italian Institute of Technology (IT)
- Universitat Autònoma de Barcelona (ES)
- Venue
- IEEE Transactions on Biomedical Circuits and Systems
- Published
- 2011-08
- Volume
- 5
- Issue
- 4
- Pages
- 368-379
- ISSN
- 1932-4545 (ISSN-L)
- 1940-9990
- OpenAlex
- W2098006249
- Cited by
- 42
- References
- 41
- Citations by year
- 2026: 1 · 2024: 3 · 2023: 2 · 2022: 1 · 2020: 2 · 2019: 2 · 2018: 5 · 2017: 4 · 2016: 3 · 2015: 4 · 2014: 3 · 2013: 5 · 2012: 6
- FWCI
- 1.98
- Type
- journal article
- Language
- en
- Open access
- green
- zenodo.org
- Locations
- IEEE Transactions on Biomedical Circuits and Systems (publishedVersion)
- PubMed (publishedVersion)
- UNICA IRIS Institutional Research Information System (University of Cagliari) (submittedVersion)
- CINECA IRIS Institutional Research Information System (Sant'Anna School of Advanced Studies) (submittedVersion)
- Zenodo (CERN European Organization for Nuclear Research) (submittedVersion, other-oa)
- Funders
- Scuola Superiore Sant'Anna
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Engineering › Biomedical Engineering › Analog and Mixed-Signal Circuit Design
- Life Sciences › Neuroscience › Neurology › Vagus Nerve Stimulation Research
- Keywords
- peripheral nerve recording, peripheral nerve stimulation, bidirectional neural interface, implanted electrodes, microcontroller-based system, bandpass filter, variable gain amplifier, biphasic current pulses, sciatic nerve, rats, input-referred noise, low power consumption, real-time monitoring interface
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review