Potential errors in pulse oximetry III: Effects of interference, dyes, dyshaemoglobins and other pigments*
The DOI matches its Crossref record; 26 of the 26 references it cites were found in a registry or an index.
Abstract
Summary: Electrosurgery, patient motion and some types of lighting can cause errors in saturation readout; it is recommended that probes should be shielded from ambient lighting. Intravenous dyes can introduce gross but transient errors, which may also be present in in vitro measurements. Carboxyhaemoglobin causes overestimation of fractional saturation by an amount less than, but possibly close to, the percent of carboxyhaemoglobin present. Methaemoglobin causes the pulse oximeter readout to tend towards 85%. Fetal haemoglobin and bilirubin introduce no significant error, although they may interfere with in vitro measurements. Skin pigmentation can result in a slight decrease in accuracy. Nail polish may cause up to 6% underestimation of saturation; it is recommended that probes should be mounted sideways on fingers with nail polish or long nails. Adhesive tape or a vinyl glove across the probe has no demonstrable effect on accuracy. A blood sample should be analysed by a multiwavelength in vitro oximeter when an erroneous pulse oximeter reading is suspected, although errors may be introduced in the in vitro reading by fetal haemoglobin, bilirubin and intravenous dyes.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Tremper, K. K., & Barker, S. J. (1989) · Anesthesiology · doi:10.1097/00000542-198901000-00019 · Cited by 435
The Effect of Carbon Monoxide Inhalation on Pulse Oximetry and Transcutaneous Po 2
Barker, S. J., & Tremper, K. K. (1987) · Anesthesiology · doi:10.1097/00000542-198705000-00014 · Cited by 281
Principles of Pulse Oximetry: Theoretical and Practical Considerations
Alexander, C. M., Teller, L. E., & Gross, J. B. (1989) · Anesthesia & Analgesia · doi:10.1213/00000539-198903000-00035 · Cited by 230
Potential errors in pulse oximetry: II. Effects of changes in saturation and signal quality*
WEBB, R. K., RALSTON, A. C., & RUNCIMAN, W. B. (1991) · Anaesthesia · doi:10.1111/j.1365-2044.1991.tb09411.x · Cited by 213
Self-citationEffects of Intravenously Administered Dyes on Pulse Oximetry Readings
SCHELLER, M. S., UNGER, R. J., & KELNER, M. J. (1986) · Anesthesiology · doi:10.1097/00000542-198611000-00023 · Cited by 194
Principles of pulse oximetry: theoretical and practical considerations.
Alexander, C. M., Teller, L. E., & Gross, J. B. (1989) · PubMed · Cited by 175
The Effect of Nail Polish on Pulse Oximetry
Cot??, C. J., Goldstein, E. A., Fuchsman, W. H., & Hoaglin, D. C. (1988) · Anesthesia & Analgesia · doi:10.1213/00000539-198807000-00013 · Cited by 172
Potential errors in pulse oximetry: I. Pulse oximeter evaluation*
RALSTON, A. C., WEBB, R. K., & RUNCIMAN, W. B. (1991) · Anaesthesia · doi:10.1111/j.1365-2044.1991.tb09410.x · Cited by 166
Self-citationRies, A. L., Prewitt, L. M., & Johnson, J. J. (1989) · Chest · doi:10.1378/chest.96.2.287 · Cited by 145
Taylor, M. B., & WHITWAM, J. G. (1986) · Anaesthesia · doi:10.1111/j.1365-2044.1986.tb12922.x · Cited by 123
Cited by
Methemoglobinemia: Etiology, Pharmacology, and Clinical Management
Wright, R. O., Lewander, W. J., & Woolf, A. D. (1999) · Annals of Emergency Medicine · doi:10.1016/s0196-0644(99)70167-8 · Cited by 856
Effects of skin pigmentation on pulse oximeter accuracy at low saturation
Bickler, P. E., Feiner, J. R., & Severinghaus, J. W. (2005) · Anesthesiology · doi:10.1097/00000542-200504000-00004 · Cited by 427
“%) in black patients.4–7 Model simulations of errors due to various pigments were reviewed by Ralston et al.8 Cote et al.9 reported that nail polish and ink on skin surface can cause errors, a finding confirmed anecdotally by others from fingerprinting ink,10 henna,11 and meconium.12 Intravenously injected dyes cause…”
McFadden, E. R. (2003) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.200208-902so · Cited by 317
The Scoring of Respiratory Events in Sleep: Reliability and Validity
Redline, S., Budhiraja, R., Kapur, V., Marcus, C. L., Mateika, J. H., Mehra, R., Parthasarthy, S., Somers, V. K., Strohl, K. P., Gozal, D., et al. (2007) · Journal of Clinical Sleep Medicine · doi:10.5664/jcsm.26818 · Cited by 305
Meta-analysis of arterial oxygen saturation monitoring by pulse oximetry in adults
Jensen, L. A., Onyskiw, J. E., & Prasad, N. G. N. (1998) · Heart & Lung · doi:10.1016/s0147-9563(98)90086-3 · Cited by 243
Neonatal and Pediatric Pulse Oximetry
Salyer, J. W. (2003) · Respiratory Care · doi:10.4187/respcare.03480386 · Cited by 183
A review of the effect of skin pigmentation on pulse oximeter accuracy
Al-Halawani, R., Charlton, P. H., Qassem, M., & Kyriacou, P. A. (2023) · Physiological Measurement · doi:10.1088/1361-6579/acd51a · Cited by 119
Monitoring in the Intensive Care
Kipnis, E., Ramsingh, D., Bhargava, M., Dincer, E., Cannesson, M., Broccard, A., Vallet, B., Bendjelid, K., & Thibault, R. (2012) · Critical Care Research and Practice · doi:10.1155/2012/473507 · Cited by 87
“False readings are also seen in severe anemia (Hb < 5 g/dL) [128], in presence of excessive skin pigmentation [115, 129], nail polish [119], or dyes.”
Yan, L., Hu, S., Alzahrani, A., Alharbi, S., & Blanos, P. (2017) · Biosensors · doi:10.3390/bios7020022 · Cited by 40
“Ink on skin and nail polish could cause errors during measurements [5].”
BackgroundRacial effects on Masimo pulse oximetry: a laboratory study
Barker, S. J., & Wilson, W. C. (2023) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-022-00927-w · Cited by 36
Haxha, S., Nwibor, C., Ali, M., Sakel, M., Saunders, K., Dyo, V., & Nabakooza, S. (2024) · Sensors · doi:10.3390/s24113301 · Cited by 12
“Carboxyhaemoglobin develops on inhalation or exposure to carbon monoxide [36,37].”
Background
Metadata
- Authors
- RALSTON, A. C., WEBB, R. K., & RUNCIMAN, W. B.
- Venue
- Anaesthesia
- Publisher
- Wiley
- Published
- 1991-04
- Volume
- 46
- Issue
- 4
- Pages
- 291-295
- ISSN
- 0003-2409 (ISSN-L)
- 1365-2044
- OpenAlex
- W2002790965
- Cited by
- 192
- References
- 26
- FWCI
- 6.38
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›