Clinical Accuracy of a Continuous Glucose Monitoring System With an Advanced Algorithm
Open accessAbstract
Background: We assessed the performance of a modified Dexcom G4 Platinum system with an advanced algorithm, in comparison with frequent venous samples measured on a laboratory reference (YSI) during a clinic session and in comparison to self-monitored blood glucose (SMBG) during home use. Methods: Fifty-one subjects with diabetes were enrolled in a prospective multicenter study. Subjects wore 1 sensor for 7-day use and participated in one 12-hour in-clinic session on day 1, 4, or 7 to collect YSI reference venous glucose every 15 minutes and capillary SMBG test every 30 minutes. Carbohydrate consumption and insulin dosing and timing were manipulated to obtain data in low and high glucose ranges. Results: In comparison with the laboratory reference method (n = 2,263) the system provided a mean and median absolute relative differences (ARD) of 9.0% and 7.0%, respectively. The mean absolute difference for CGM was 6.4 mg/dL when the YSIs were within hypoglycemia ranges (≤ 70 mg/dL). The percentage in the clinically accurate Clarke error grid A zone was 92.4% and in the benign error B zone was 7.1%. Majority of the sensors (73%) had an aggregated MARD in reference to YSI ≤ 10%. The MARD of CGM-SMBG for home use was 11.3%. Conclusions: The study showed that the point and rate accuracy, clinical accuracy, reliability, and consistency over the duration of wear and across glycemic ranges were superior to current commercial real-time CGM systems. The performance of this CGM is reaching that of a self-monitoring blood glucose meter in real use environment.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“continuous glucose monitoring (CGM) and intermittently viewed CGM devices for at-home patient use with minimally invasive devices have become available, which have similar accuracy to capillary blood glucose monitors.91”
From Type 1 diabetes (2018) · cites this paper as [91]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Errors in CGM also increase the difficulty of predicting BG values (approximately 9% of the mean absolute relative difference for the best sensors70).”
From Artificial intelligence in diabetes management: Advancements, opportunities, and challenges (2023) · cites this paper as [70]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“CGM utilizes subcutaneous sensors for continuous monitoring of glucose levels with a high degree of accuracy. 17 , 18”
From Insulin pump therapy is associated with higher rates of mild diabetic ketoacidosis compared to injection therapy: A 2‐year Swedish national survey of children and adolescents with type 1 diabetes (2022) · cites this paper as [17]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Evaluating clinical accuracy of systems for self-monitoring of blood glucose
Clarke, W. L., Cox, D., Gonder-Frederick, L. A., Carter, W., & Pohl, S. L. (1987) · Diabetes Care · doi:10.2337/diacare.10.5.622 · Cited by 1,435
Freckmann, G., Schmid, C., Baumstark, A., Pleus, S., Link, M., & Haug, C. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600510 · Cited by 237
“Turn it Off!”: Diabetes Device Alarm Fatigue Considerations for the Present and the Future
Shivers, J. P., Mackowiak, L., Anhalt, H., & Zisser, H. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700324 · Cited by 159
Christiansen, M. P., Bailey, T. S., Watkins, E., Liljenquist, D. R., Price, D. A., Nakamura, K., Boock, R. J., & Peyser, T. A. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0077 · Cited by 140
Self-citationA Comparative Effectiveness Analysis of Three Continuous Glucose Monitors
Damiano, E. R., McKeon, K., El-Khatib, F. H., Zheng, H., Nathan, D. M., & Russell, S. J. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814532203 · Cited by 139
Kropff, J., Bruttomesso, D., Doll, W., Farret, A., Galasso, S., Luijf, Y. M., Mader, J. K., Place, J., Boscari, F., Pieber, T. R., et al. (2014) · Diabetes Obesity and Metabolism · doi:10.1111/dom.12378 · Cited by 132
Šoupal, J., Škrha, J., Fajmon, M., Horová, E., Mráz, M., Škrha, J., & Prázný, M. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0205 · Cited by 122
Mazze, R. S., Strock, E. S., Borgman, S. K., Wesley, D., Stout, P. J., & Racchini, J. R. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0041 · Cited by 116
Accuracy and Acceptability of the 6-Day Enlite Continuous Subcutaneous Glucose Sensor
Bailey, T. S., Ahmann, A., Brazg, R., Christiansen, M. P., Garg, S. K., Watkins, E., Welsh, J. B., & Lee, S. W. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0222 · Cited by 106
Self-citationTack, C. J., Pohlmeier, H., Behnke, T., Schmid, V., Grenningloh, M., Forst, T. M., & Pfützner, A. (2011) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2011.0170 · Cited by 91
Research works printed but not linked: 5
- Christiansen M, 2013, Diabetes Technol Ther, 5, 1
- Kropff J, 2014, Diabetes Obes Metab
- Brauker JH, 2007, Review of Endocrinology
- Pettus J, 2014, Diabetes Technol Ther, 16, 198
- Hessler DM, Paper presented at: 71st American Diabetes Association meeting
Cited by
Sorted by citations
DiMeglio, L. A., Evans-Molina, C., & Oram, R. A. (2018) · The Lancet · doi:10.1016/s0140-6736(18)31320-5 · Cited by 1,561
Beck, R. W., Riddlesworth, T., Ruedy, K., Ahmann, A., Bergenstal, R., Haller, S., Kollman, C., Kruger, D., McGill, J. B., Polonsky, W., et al. (2017) · JAMA · doi:10.1001/jama.2016.19975 · Cited by 1,185
Electrochemical glucose sensors in diabetes management: an updated review (2010–2020)
Teymourian, H., Barfidokht, A., & Wang, J. (2020) · Chemical Society Reviews · doi:10.1039/d0cs00304b · Cited by 915
Artificial intelligence for diabetes management and decision support: Literature review
Contreras, I., & Vehi, J. (2018) · Journal of Medical Internet Research · doi:10.2196/10775 · Cited by 560
DiMeglio, L. A., Acerini, C. L., Codner, E., Craig, M. E., Hofer, S. E., Pillay, K., & Maahs, D. M. (2018) · Pediatric Diabetes · doi:10.1111/pedi.12737 · Cited by 604
Cappon, G., Vettoretti, M., Sparacino, G., & Facchinetti, A. (2019) · Diabetes & Metabolism Journal · doi:10.4093/dmj.2019.0121 · Cited by 477
Polonsky, W. H., Hessler, D., Ruedy, K. J., & Beck, R. W. (2017) · Diabetes Care · doi:10.2337/dc17-0133 · Cited by 346
Aleppo, G., Ruedy, K. J., Riddlesworth, T. D., Kruger, D. F., Peters, A. L., Hirsch, I., Bergenstal, R. M., Toschi, E., Ahmann, A. J., Shah, V. N., et al. (2017) · Diabetes Care · doi:10.2337/dc16-2482 · Cited by 329
Continuous glucose monitoring: A review of the technology and clinical use
Klonoff, D. C., Ahn, D., & Drincic, A. (2017) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2017.08.005 · Cited by 320
Artificial intelligence in diabetes management: Advancements, opportunities, and challenges
Guan, Z., Li, H., Liu, R., Cai, C., Liu, Y., Li, J., Wang, X., Huang, S., Wu, L., Liu, D., et al. (2023) · Cell Reports Medicine · doi:10.1016/j.xcrm.2023.101213 · Cited by 287
“Errors in CGM also increase the difficulty of predicting BG values (approximately 9% of the mean absolute relative difference for the best sensors 70 ).”
BackgroundReddy, M., Jugnee, N., El Laboudi, A., Spanudakis, E., Anantharaja, S., & Oliver, N. (2018) · Diabetic Medicine · doi:10.1111/dme.13561 · Cited by 245
Wearable Continuous Glucose Monitoring Sensors: A Revolution in Diabetes Treatment
Cappon, G., Acciaroli, G., Vettoretti, M., Facchinetti, A., & Sparacino, G. (2017) · Electronics · doi:10.3390/electronics6030065 · Cited by 247
Outcome measures for artificial pancreas clinical trials: a consensus report
Maahs, D. M., Buckingham, B. A., Castle, J. R., Cinar, A., Damiano, E. R., Dassau, E., DeVries, J. H., Doyle, F. J., Griffen, S. C., Haidar, A., et al. (2016) · Diabetes Care · doi:10.2337/dc15-2716 · Cited by 211
Wadwa, R. P., Laffel, L. M., Shah, V. N., & Garg, S. K. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0150 · Cited by 203
ISPAD Clinical Practice Consensus Guidelines 2018: diabetes technologies
Sherr, J. L., Tauschmann, M., Battelino, T., de Bock, M., Forlenza, G., Roman, R., Hood, K. K., & Maahs, D. M. (2018) · Pediatric Diabetes · doi:10.1111/pedi.12731 · Cited by 197
Continuous glucose monitoring sensors: past, present and future algorithmic challenges
Facchinetti, A. (2016) · Sensors · doi:10.3390/s16122093 · Cited by 176
Kropff, J., Choudhary, P., Neupane, S., Barnard-Kelly, K. D., Bain, S. C., Kapitza, C., Forst, T. M., Link, M., DEHENNIS, A. D., & DeVries, J. H. (2016) · Diabetes Care · doi:10.2337/dc16-1525 · Cited by 171
Heinemann, L., Schoemaker, M., Schmelzeisen-Redecker, G., Hinzmann, R., Kassab, A., Freckmann, G., Reiterer, F., & Del Re, L. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819855670 · Cited by 170
Technology in the management of type 1 diabetes mellitus — current status and future prospects
Tauschmann, M., & Hovorka, R. (2018) · Nature Reviews Endocrinology · doi:10.1038/s41574-018-0044-y · Cited by 165
Christiansen, M. P., Klaff, L. J., Brazg, R., Chang, A. R., Levy, C. J., Lam, D. W.-H., Denham, D. S., Atiee, G. J., Bode, B. W., Walters, S. J., et al. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0142 · Cited by 146
Self-citationA review of optical methods for continuous glucose monitoring
Jernelv, I. L., Mileńko, K., Fuglerud, S. S., Hjelme, D. R., Ellingsen, R., & Aksnes, A. (2018) · Applied Spectroscopy Reviews · doi:10.1080/05704928.2018.1486324 · Cited by 137
Aberer, F., Hajnsek, M., Rumpler, M., Zenz, S., Baumann, P. M., Elsayed, H., Puffing, A., Treiber, G. M., Pieber, T. R., Sourij, H., et al. (2017) · Diabetes Obesity and Metabolism · doi:10.1111/dom.12907 · Cited by 132
Bailey, T. S., Grunberger, G., Bode, B. W., Handelsman, Y., Hirsch, I. B., Jovanovič, L., Roberts, V. L., Rodbard, D., Tamborlane, W. V., & Walsh, J. S. A. (2016) · Endocrine Practice · doi:10.4158/ep151124.cs · Cited by 130
Self-citationPettus, J., & Edelman, S. V. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816663747 · Cited by 122
Vettoretti, M., Facchinetti, A., Sparacino, G., & Cobelli, C. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2017.2746340 · Cited by 110
HbA1c and Glucose Management Indicator Discordance: A Real-World Analysis
Perlman, J. E., Gooley, T. A., MCNULTY, B. K., Meyers, J., & Hirsch, I. B. (2020) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2020.0501 · Cited by 101
Continuous glucose monitoring: Current use in diabetes management and possible future applications
Vettoretti, M., Cappon, G., Acciaroli, G., Facchinetti, A., & Sparacino, G. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818774078 · Cited by 96
Basu, A., Veettil, S., Dyer, R. B., Peyser, T. A., & Basu, R. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0410 · Cited by 94
Wearable physiological systems and technologies for metabolic monitoring
Gao, W., Brooks, G. A., & Klonoff, D. C. (2018) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00407.2017 · Cited by 91
Freckmann, G., Eichenlaub, M., Waldenmaier, D., Pleus, S., Wehrstedt, S., Haug, C., Witthauer, L., Jendle, J., Hinzmann, R., Thomas, A., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231190941 · Cited by 86
Laffel, L. M. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0380 · Cited by 86
Sadhu, A. R., Serrano, I. A., Xu, J., Nisar, T., Lucier, J., Pandya, A. R., & Patham, B. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820964264 · Cited by 84
Rate-of-Change Dependence of the Performance of Two CGM Systems During Induced Glucose Swings
Pleus, S., Schoemaker, M., Morgenstern, K., Schmelzeisen‐Redeker, G., Haug, C., Link, M., Zschornack, E., & Freckmann, G. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815578716 · Cited by 84
Asarani, N. A. M., Reynolds, A. N., Styles, S. E., de Bock, M. I., & Wheeler, B. J. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819870849 · Cited by 81
Contreras, I., Oviedo, S., Vettoretti, M., VISENTIN, R., & Vehı́, J. (2017) · PLoS ONE · doi:10.1371/journal.pone.0187754 · Cited by 80
Long-term implantable glucose biosensors
Bobrowski, T., & Schuhmann, W. W. (2018) · Current Opinion in Electrochemistry · doi:10.1016/j.coelec.2018.05.004 · Cited by 77
Welsh, J. B., Gao, P., Derdzinski, M., Puhr, S., Johnson, T. K., Walker, T. C., & Graham, C. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0374 · Cited by 72
Optical sensors for continuous glucose monitoring
Yu, Z., Jiang, N., Kazarian, S. G., Taşoğlu, S., & Yetisen, A. K. (2021) · Progress in Biomedical Engineering · doi:10.1088/2516-1091/abe6f8 · Cited by 66
Wood, A. J., O'Neal, D. N., Furler, J., & Ekinci, E. I. (2018) · Internal Medicine Journal · doi:10.1111/imj.13770 · Cited by 66
Basics and use of continuous glucose monitoring (CGM) in diabetes therapy
Freckmann, G. (2020) · Journal of Laboratory Medicine · doi:10.1515/labmed-2019-0189 · Cited by 72
Freckmann, G., Link, M., Pleus, S., Westhoff, A., Kamecke, U., & Haug, C. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0105 · Cited by 64
Standardization process of continuous glucose monitoring: Traceability and performance
Freckmann, G., Nichols, J. H., Hinzmann, R., Klonoff, D. C., Ju, Y., Diem, P., Makris, K., & Slingerland, R. J. (2021) · Clinica Chimica Acta · doi:10.1016/j.cca.2020.12.025 · Cited by 62
Cappon, G., Vettoretti, M., Marturano, F., Facchinetti, A., & Sparacino, G. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818759558 · Cited by 62
Clinical Implications of Accuracy Measurements of Continuous Glucose Sensors
Bailey, T. S. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0050 · Cited by 60
Self-citationModeling Transient Disconnections and Compression Artifacts of Continuous Glucose Sensors
Facchinetti, A., Favero, S. D., Sparacino, G., & Cobelli, C. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0250 · Cited by 60
Nonadjunctive Use of Continuous Glucose Monitoring for Diabetes Treatment Decisions
Castle, J. R., & Jacobs, P. G. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816631569 · Cited by 58
Eichenlaub, M., Pleus, S., Rothenbühler, M., Bailey, T. S., Bally, L., Brazg, R., Bruttomesso, D., Diem, P., Boija, E. E., Fokkert, M. J., et al. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0465 · Cited by 51
Self-citationDeveloping Insulin Delivery Devices with Glucose Responsiveness
Wang, Z., Wang, J., Kahkoska, A. R., Buse, J. B., & Gu, Z. (2020) · Trends in Pharmacological Sciences · doi:10.1016/j.tips.2020.11.002 · Cited by 49
Avari, P., Reddy, M., & Oliver, N. S. (2019) · Diabetic Medicine · doi:10.1111/dme.13942 · Cited by 46
Forlenza, G. P., Argento, N. B., & Laffel, L. M. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0034 · Cited by 47
Li, A., Riddell, M. C., Potashner, D., Brown, R. E., & Aronson, R. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0387 · Cited by 44
Comparisons of Fifth-, Sixth-, and Seventh-Generation Continuous Glucose Monitoring Systems
Welsh, J. B., Psavko, S., Zhang, X., Gao, P., & Balo, A. K. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221099879 · Cited by 43
A Review of Safety and Hazards Associated With the Artificial Pancreas
Ramkissoon, C. M., Aufderheide, B., Bequette, B. W., & Vehı́, J. (2017) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2017.2749038 · Cited by 41
Applications and pitfalls of hemoglobin A1C and alternative methods of glycemic monitoring
Lundholm, M. D., Emanuele, M. A., Ashraf, A., & Nadeem, S. (2020) · Journal of Diabetes and its Complications · doi:10.1016/j.jdiacomp.2020.107585 · Cited by 42
Avari, P., Uduku, C., George, D., Herrero, P., Reddy, M., & Oliver, N. (2020) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0276 · Cited by 40
Lupoli, R., Lembo, E., Ciciola, P., Schiavo, L., Pilone, V., & Capaldo, B. (2020) · Nutrition Metabolism and Cardiovascular Diseases · doi:10.1016/j.numecd.2020.06.029 · Cited by 41
Lindner, N., Kuwabara, A., & Holt, T. (2021) · Systematic Reviews · doi:10.1186/s13643-021-01644-2 · Cited by 42
Peyser, T. A., Nakamura, K., Price, D. A., Bohnett, L. C., Hirsch, I. B., & Balo, A. K. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0415 · Cited by 37
Thabit, H., Leelarathna, L., Wilinska, M. E., Elleri, D., Allen, J. M., Lubina-Solomon, A., Walkinshaw, E., Stadler, M., Choudhary, P., Mader, J. K., et al. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0062 · Cited by 37
Acciaroli, G., Vettoretti, M., Facchinetti, A., Sparacino, G., & Cobelli, C. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2017.2706974 · Cited by 36
Performance and Usability of Three Systems for Continuous Glucose Monitoring in Direct Comparison
Freckmann, G., Link, M., Kamecke, U., Haug, C., Baumgartner, B., & Weitgasser, R. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819826965 · Cited by 34
Mariani, H., Layden, B. T., & Aleppo, G. (2017) · Clinical Diabetes · doi:10.2337/cd16-0008 · Cited by 33
Faccioli, S., Favero, S. D., VISENTIN, R., Bonfanti, R., Iafusco, D., Rabbone, I., Marigliano, M., Schiaffini, R., Bruttomesso, D., Cobelli, C., et al. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817706377 · Cited by 32
Unblinded CGM Should Replace Blinded CGM in the Clinical Management of Diabetes
Ahn, D. T., Pettus, J., & Edelman, S. V. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816632241 · Cited by 30
FDA Approval of Nonadjunctive Use of Continuous Glucose Monitors for Insulin Dosing
Shapiro, A. R. (2017) · JAMA · doi:10.1001/jama.2017.12075 · Cited by 26
Shah, V. N., Snell‐Bergeon, J. K., Demmitt, J. K., Joshee, P., Garcetti, R., Pyle, L. L., & Polsky, S. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0093 · Cited by 25
Tissue Response to Subcutaneous Infusion Catheter
Zhang, E., & Cao, Z. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819837972 · Cited by 24
Link, M., Kamecke, U., Waldenmaier, D., Pleus, S., Garcia, A., Haug, C., & Freckmann, G. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819895022 · Cited by 23
Performance of a new real‐time continuous glucose monitoring system: A multicenter pilot study
Zhou, J., Zhang, S., Li, L., Wang, Y., Lu, W., Sheng, C., Li, Y., Bao, Y., & Jia, W. (2018) · Journal of Diabetes Investigation · doi:10.1111/jdi.12699 · Cited by 22
Acciaroli, G., Vettoretti, M., Facchinetti, A., Sparacino, G., & Cobelli, C. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0088 · Cited by 22
Nonadjunctive Use of Continuous Glucose Monitors for Insulin Dosing: Is It Safe?
Shapiro, A. R. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816688303 · Cited by 22
Chaugule, S., & Graham, C. (2017) · Journal of Medical Economics · doi:10.1080/13696998.2017.1360312 · Cited by 22
Schiavon, M., Man, C. D., & Cobelli, C. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0248 · Cited by 22
Hypoglycemia Reduction and Accuracy of Continuous Glucose Monitoring
Kovatchev, B. P. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0144 · Cited by 22
Wersäll, J. H., Adolfsson, P., Forsander, G., & Hanas, R. (2022) · Pediatric Diabetes · doi:10.1111/pedi.13377 · Cited by 21
Longato, E., Acciaroli, G., Facchinetti, A., Hakaste, L., Tuomilehto, J., Maran, A., & Sparacino, G. (2018) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2018.03.007 · Cited by 19
A Model of Self-Monitoring Blood Glucose Measurement Error
Vettoretti, M., Facchinetti, A., Sparacino, G., & Cobelli, C. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817698498 · Cited by 18
Joseph, J. I., EISLER, G. J., Diaz, D., Khalf, A., Loeum, C., & Torjman, M. C. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0106 · Cited by 20
Ghosh, J., Choudhury, S. R., Singh, K., & Koner, S. (2024) · International Journal of Advancement in Life Sciences Research · doi:10.31632/ijalsr.2024.v07i02.004 · Cited by 18
Sharifi, A., Varsavsky, A., Ulloa, J., Horsburgh, J. C., McAuley, S. A., Krishnamurthy, B., Jenkins, A. J., Colman, P. G., Ward, G. M., MacIsaac, R. J., et al. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815612096 · Cited by 16
Camerlingo, N., Vettoretti, M., Sparacino, G., Facchinetti, A., Mader, J. K., Choudhary, P., & Favero, S. D. (2021) · Diabetes Obesity and Metabolism · doi:10.1111/dom.14483 · Cited by 16
Methodology for Safe and Secure AI in Diabetes Management
Foppen, R. J. G., Gioia, V. L., Gupta, S., Johnson, C. L., Giantsidis, J., & Papademetris, M. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968241304434 · Cited by 16
Hochfellner, D. A., Simic, A., Taucher, M. T., Sailer, L. S., Kopanz, J., Pöttler, T., & Mader, J. K. (2022) · Biosensors · doi:10.3390/bios12020106 · Cited by 15
Vettoretti, M., Facchinetti, A., Sparacino, G., & Cobelli, C. (2015) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2015.7318868 · Cited by 15
Pettus, J., Gill, J., Paranjape, S., Stewart, J. A., Malla, S., Edelman, S. V., Bergenstal, R. M., & Bode, B. W. (2019) · Diabetes Obesity and Metabolism · doi:10.1111/dom.13751 · Cited by 15
Price, D. A., & Walker, T. C. (2016) · touchREVIEWS in Endocrinology · doi:10.17925/ee.2016.12.01.24 · Cited by 17
Graf, A., McAuley, S. A., Sims, C. M., Ulloa, J., Jenkins, A. J., Voskanyan, G. R., & O'Neal, D. N. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816682762 · Cited by 14
Data Reliability and Quality in Body Area Networks for Diabetes Monitoring
Huzooree, G., Khedo, K. K., & Joonas, N. (2018) · EAI/Springer Innovations in Communication and Computing · doi:10.1007/978-3-030-00865-9_4 · Cited by 13
Electrochemical Glucose Biosensors for Diabetes Care
Ocvirk, G., Buck, H., & DuVall, S. H. (2016) · Bioanalytical reviews · doi:10.1007/11663_2016_3 · Cited by 13
Pleus, S., Link, M., Hinzmann, R., Öter, Ş., Eichenlaub, M., Baumstark, A., Haug, C., & Freckmann, G. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0489 · Cited by 12
Intraperitoneal Glucose Sensing is Sometimes Surprisingly Rapid
Fougner, A. L., Kölle, K., Skjærvold, N. K., Elvemo, N.-A. L., Hjelme, D. R., Ellingsen, R., Carlsen, S. M., & Stavdahl, Ø. (2016) · Modeling Identification and Control A Norwegian Research Bulletin · doi:10.4173/mic.2016.2.4 · Cited by 12
Muhs, A., Bobrowski, T., Lielpētere, A., & Schuhmann, W. W. (2022) · Angewandte Chemie International Edition · doi:10.1002/anie.202211559 · Cited by 12
Predicting Insulin Treatment Scenarios with the Net Effect Method: Domain of Validity
Vettoretti, M., Facchinetti, A., Sparacino, G., & Cobelli, C. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0148 · Cited by 12
Jia, Z., Huang, L., Liu, H., Huang, Y., Li, W., Pi, X., & Zheng, X. (2019) · Applied Biochemistry and Biotechnology · doi:10.1007/s12010-019-03142-7 · Cited by 11
Camerlingo, N., Vettoretti, M., Sparacino, G., Facchinetti, A., Mader, J. K., Choudhary, P., & Favero, S. D. (2021) · Diabetic Medicine · doi:10.1111/dme.14758 · Cited by 10
Continuous Glucose Monitoring Use in Clinical Trials for On-Market Diabetes Drugs
Fox, B. Q., Benjamin, P. F., Aqeel, A., Fitts, E., Flynn, S., Levine, B., Maslak, E., Milner, R. L., Ose, B. M., Poeschla, M., et al. (2021) · Clinical Diabetes · doi:10.2337/cd20-0049 · Cited by 12
McAuley, S. A., Dang, T. T., Horsburgh, J. C., Bansal, A., Ward, G. M., Aroyan, S., Jenkins, A. J., MacIsaac, R. J., Shah, R. V., & O'Neal, D. N. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816629982 · Cited by 10
Continuous Glucose Monitoring for the Internist
Lee, G. S., & Lupsa, B. C. (2021) · Medical Clinics of North America · doi:10.1016/j.mcna.2021.06.004 · Cited by 9
Schiavon, M., Galderisi, A., Basu, A., Kudva, Y. C., Cengiz, E., & Dalla Man, C. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2022.0397 · Cited by 8
Retrofitting Real-Life Dexcom G5 Data
Favero, S. D., Facchinetti, A., Sparacino, G., & Cobelli, C. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0413 · Cited by 8
Metadata
- Authors
- Bailey, T. S., Chang, A., & Christiansen, M.
- Affiliations
- AMCR Institute (US)
- John Muir Health (US)
- Diablo Clinical Research (US)
- Corresponding author
- Bailey, Timothy S.
- ORCID
- Bailey, Timothy S.
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2014-11-03
- Volume
- 9
- Issue
- 2
- Pages
- 209-214
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC4604574
- OpenAlex
- W2165138941
- Cited by
- 168
- References
- 24
- Citations by year
- 2026: 7 · 2025: 7 · 2024: 9 · 2023: 7 · 2022: 7 · 2021: 14 · 2020: 15 · 2019: 23 · 2018: 21 · 2017: 30 · 2016: 18 · 2015: 10
- FWCI
- 15.54
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- Europe PMC (PubMed Central) (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Funders
- Dexcom
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Treatment and Management
- Keywords
- continuous glucose monitoring, diabetes mellitus, clinical accuracy, hypoglycemia, self-monitoring of blood glucose, Clarke error grid, mean absolute relative difference, sensor reliability, insulin dose
- MeSH terms
- Adult, Aged, Aged, 80 and over, Algorithms, Blood Glucose, Diabetes Mellitus, Female, Humans, Male, Middle Aged, Blood Glucose Self-Monitoring, Young Adult
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review