Foreign Body Reaction to Implantable Biosensors: Effects of Tissue Trauma and Implant Size
Open accessThe DOI matches its Crossref record; 25 of the 25 references it cites were found in a registry or an index.
Abstract
Background: Implantable biosensors for continuous glucose monitoring can greatly improve diabetes management. However, their applications are still associated with some challenges and one of these is the gradual functionality loss postimplantation as a consequence of the foreign body response (FBR). Sensor miniaturization in combination with drug-eluting biocompatible coatings is a promising strategy to enhance in vivo performance. However, limited study has been performed to understand the effect of initial trauma and implant size on foreign body reaction as well as in vivo performance of implantable glucose sensors. Methods: Different initial trauma was induced by implanting composite coated dummy sensors into rats using various sized needles and 3 different-sized dummy sensors were implanted to examine the size effect. Histological evaluation was performed to relate the inflammatory cell counts and foreign body capsule thickness with the implantation needle size and sensor size respectively. The effect of biocompatible coating on the performance of implantable glucose sensors was determined using both coated amperometric glucose sensors and microdialysis probes. Results: The results revealed that the degree of acute inflammation was mainly controlled by the extent of the initial trauma: the greater the trauma, the greater the acute inflammatory response. Implant size did not affect the acute inflammatory phase. However, the extent of chronic inflammation and fibrous encapsulation were affected by sensor size: the smaller the size the less the extent of chronic inflammation and fibrous encapsulation. Glucose sensors implanted using 14 gauge needles showed significantly lower initial in vivo response compared to those implanted using 16 gauge needles. This was not observed for sensors with dexamethasone-eluting biocompatible coatings since inflammation was suppressed. Conclusions: The results of the current study indicate that the extent of the inflammatory response post–sensor implantation varies as a function of the initial tissue trauma as well as the sensor size. Accordingly, miniaturization of implantable biosensors together with the utilization of a drug-eluting biocompatible composite coating may be a promising strategy to achieve long-term reliable continuous glucose monitoring.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“It is therefore unsurprising that the degree of implantation trauma is linked to the degree of FBR (Wang et al., 2015).”
From Foreign Body Reaction to Implanted Biomaterials and Its Impact in Nerve Neuroprosthetics (2021) · cites this paper as Wang et al., 2015
The cited source says what the claim attributes to it.
In the citing paper
“FBR is fundamentally tied to tissue trauma (Wang et al., 2015; Klopfleisch and Jung, 2017).”
From Foreign Body Reaction to Implanted Biomaterials and Its Impact in Nerve Neuroprosthetics (2021) · cites this paper as Wang et al., 2015
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Theoharides, T. C., Alysandratos, K.-D., Angelidou, A., Delivanis, D.-A., Sismanopoulos, N., Zhang, B., Asadi, S., Vasiadi, M., Weng, Z., Miniati, A., et al. (2012) · Biochimica et Biophysica Acta (BBA) - Molecular Basis of Disease · doi:10.1016/j.bbadis.2010.12.014 · Cited by 680
Effect of acidic pH on PLGA microsphere degradation and release
Zolnik, B. S., & Burgess, D. J. (2007) · Journal of Controlled Release · doi:10.1016/j.jconrel.2007.05.034 · Cited by 560
Self-citationA Tale of Two Compartments: Interstitial Versus Blood Glucose Monitoring
Cengiz, E., & Tamborlane, W. V. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2009.0002 · Cited by 390 · cited 1×
“The glucose concentration in the interstitial fluid is generally considered to be equal to the blood glucose level, although a short lag phase is reported.(27) Tissue injuries caused during the implantation process induce acute inflammation with associated edema.”
BackgroundEmerging synergy between nanotechnology and implantable biosensors: A review
Vaddiraju, S., Tomazos, I., Burgess, D. J., Jain, F. C., & Papadimitrakopoulos, F. (2010) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2009.12.001 · Cited by 368
Self-citationWard, W. K. (2008) · Journal of Diabetes Science and Technology · doi:10.1177/193229680800200504 · Cited by 325
Evaluation of in vivo–in vitro release of dexamethasone from PLGA microspheres
Zolnik, B. S., & Burgess, D. J. (2008) · Journal of Controlled Release · doi:10.1016/j.jconrel.2008.01.004 · Cited by 286
Self-citationHickey, T., Kreutzer, D., Burgess, D. J., & Moussy, F. (2002) · Biomaterials · doi:10.1016/s0142-9612(01)00291-5 · Cited by 268
Self-citationElevated temperature accelerated release testing of PLGA microspheres
Zolnik, B. S., Leary, P. E., & Burgess, D. J. (2006) · Journal of Controlled Release · doi:10.1016/j.jconrel.2006.02.015 · Cited by 233
Self-citationConcurrent delivery of dexamethasone and VEGF for localized inflammation control and angiogenesis
Patil, S. D., Papadmitrakopoulos, F., & Burgess, D. J. (2007) · Journal of Controlled Release · doi:10.1016/j.jconrel.2006.10.013 · Cited by 230 · cited 1×
“In addition, coatings that codeliver growth factors and dexamethasone to control FBR and promote blood vessel growth simultaneously have also been reported.(10,12) Accordingly, miniaturization of the implantable biosensor in combination with the utilization of a drug-eluting biocompatible coatings would be promising t…”
BackgroundSelf-citationImplantable enzyme amperometric biosensors
Kotanen, C. N., Moussy, F. G., Carrara, S., & Guiseppi‐Elie, A. (2012) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2012.03.016 · Cited by 185
Norton, L. W., Koschwanez, H. E., Wisniewski, N. A., Klitzman, B., & Reichert, W. M. (2007) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.31088 · Cited by 128 · cited 1×
“In addition, coatings that codeliver growth factors and dexamethasone to control FBR and promote blood vessel growth simultaneously have also been reported.(10,12) Accordingly, miniaturization of the implantable biosensor in combination with the utilization of a drug-eluting biocompatible coatings would be promising t…”
BackgroundHelton, K. L., Ratner, B. D., & Wisniewski, N. A. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500317 · Cited by 125
Polymeric “smart” coatings to prevent foreign body response to implantable biosensors
Wang, Y., Papadimitrakopoulos, F., & Burgess, D. J. (2013) · Journal of Controlled Release · doi:10.1016/j.jconrel.2012.12.028 · Cited by 114
Self-citationPLGA/PVA hydrogel composites for long-term inflammation control following s.c. implantation
Bhardwaj, U., Sura, R., Papadimitrakopoulos, F., & Burgess, D. J. (2009) · International Journal of Pharmaceutics · doi:10.1016/j.ijpharm.2009.09.046 · Cited by 102
Self-citationCalibration of a subcutaneous amperometric glucose sensor implanted for 7 days in diabetic patients
Choleau, C., Klein, J. C., Reach, G., Aussedat, B., Demaria-Pesce, V., Wilson, G. S., Gifford, R., & Ward, W. K. (2002) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(01)00304-9 · Cited by 76 · cited 2×
“I2/G2 (S = (I1/G1+I2/G2)/2), wherein I1 and I2 represent the peak amperometric sensor response for the 2 glycemic events, while G1 and G2 represent the maximum glucose concentrations of the 2 glycemic peaks determined via a Bayer Contour® Blood Glucose Monitoring System.(22,23) As shown in Figure 6A, the initial sensi…”
MethodInfluentialIntegrated implantable device for long-term glucose monitoring
Wilkins, E. S., Atanasov, P., & Muggenburg, B. A. (1995) · Biosensors and Bioelectronics · doi:10.1016/0956-5663(95)96894-5 · Cited by 68
Calibration of a subcutaneous amperometric glucose sensor
Choleau, C., Klein, J. C., Reach, G., Aussedat, B., Demaria-Pesce, V. H., Wilson, G. S., Gifford, R., & Ward, W. K. (2002) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(01)00306-2 · Cited by 66
Effect of ethanol as a processing co-solvent on the PLGA microsphere characteristics
Rawat, A., & Burgess, D. J. (2010) · International Journal of Pharmaceutics · doi:10.1016/j.ijpharm.2010.05.013 · Cited by 56
Self-citationAvula, M., Rao, A. N., McGill, L. D., Grainger, D. W., & Solzbacher, F. (2013) · Biomaterials · doi:10.1016/j.biomaterials.2013.08.090 · Cited by 48
Design and Fabrication of a High-Performance Electrochemical Glucose Sensor
Vaddiraju, S., Legassey, A., Wang, Y., Qiang, L., Burgess, D. J., Jain, F., & Papadimitrakopoulos, F. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500504 · Cited by 46
Self-citationWARD, W. K., & TROUPE, J. E. (1999) · ASAIO Journal · doi:10.1097/00002480-199911000-00009 · Cited by 43
A miniaturized transcutaneous system for continuous glucose monitoring
Croce, R. A., Vaddiraju, S., Kondo, J., Wang, Y., Zuo, L., Zhu, K., ISLAM, S. K., Burgess, D. J., Papadimitrakopoulos, F., & Jain, F. (2012) · Biomedical Microdevices · doi:10.1007/s10544-012-9708-x · Cited by 43
Self-citationThe role of H2O2 outer diffusion on the performance of implantable glucose sensors
Vaddiraju, S., Burgess, D. J., Jain, F., & Papadimitrakopoulos, F. (2008) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2008.08.015 · Cited by 42
Self-citationSanders, J. E., Cassisi, D. V., Neumann, T., Golledge, S. L., Zachariah, S. G., Ratner, B. D., & Bale, S. D. (2003) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.10525 · Cited by 35 · cited 1×
“The influence of polymer fiber diameter and surface charge on the fibrous capsule and vessel density in the surrounding tissue, has been investigated for single-fiber implants, with diameters in the micrometer range.(14) However, there are no reports describing a detailed understanding of the effect of the initial tis…”
BackgroundSINGARAYAR, S., KISTLER, P. M., DE WINTER, C., & MOND, H. (2005) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2005.40055.x · Cited by 21
Research works printed but not linked: 4
- 5. David JAS, Cunningham D. In Vivo Glucose Sensing . Hoboken, NJ: John Wiley; 2009.
- 7. Wang Y, Papadimitrakopoulos F, Burgess DJ. Polymeric 'smart' coatings to prevent foreign body response to implantable biosensors. J Control Release . 2013.
- 14. performance and the foreign body response-part I: theoretical framework. J Diabetes Sci Technol . 2011;5:632-646.
- 24. Villarreal G, Zagorski J, Wahl SM. Inflammation: Acute . Hoboken, NJ: John Wiley; 2001.
Cited by
Electrochemical glucose sensors in diabetes management: an updated review (2010–2020)
Teymourian, H., Barfidokht, A., & Wang, J. (2020) · Chemical Society Reviews · doi:10.1039/d0cs00304b · Cited by 881
Foreign Body Reaction to Implanted Biomaterials and Its Impact in Nerve Neuroprosthetics
Carnicer-Lombarte, A., Chen, S.-T., Malliaras, G. G., & Barone, D. G. (2021) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2021.622524 · Cited by 454
“It is therefore unsurprising that the degree of implantation trauma is linked to the degree of FBR (Wang et al., 2015).”
BackgroundPiech, D. K., Johnson, B. C., Shen, K., Ghanbari, M. M., Li, K. Y., Neely, R. M., Kay, J. E., Carmena, J. M., Maharbiz, M. M., & Muller, R. (2020) · Nature Biomedical Engineering · doi:10.1038/s41551-020-0518-9 · Cited by 442
Mayer, M., & Baeumner, A. J. (2019) · Chemical Reviews · doi:10.1021/acs.chemrev.8b00719 · Cited by 300
Domesticating the foreign body response: recent advances and applications
Veiseh, O., & Vegas, A. J. (2019) · Advanced Drug Delivery Reviews · doi:10.1016/j.addr.2019.08.010 · Cited by 245
Metadata
- Authors
- Wang, Y., Vaddiraju, S., Gu, B., Papadimitrakopoulos, F., & Burgess, D. J.
- Affiliations
- University of Connecticut (US)
- Biorasis (United States) (US)
- Corresponding author
- Burgess, Diane J.
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2015-08-25
- Volume
- 9
- Issue
- 5
- Pages
- 966-977
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC4667354
- OpenAlex
- W1830091241
- Cited by
- 107
- References
- 25
- Citations by year
- 2026: 12 · 2025: 16 · 2024: 13 · 2023: 10 · 2022: 8 · 2021: 10 · 2020: 7 · 2019: 8 · 2018: 15 · 2017: 4 · 2016: 4
- FWCI
- 3.69
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- Europe PMC (PubMed Central) (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Heart, Lung, and Blood Institute (R21-HL090458-01, R21HL090458)
- National Institute of Biomedical Imaging and Bioengineering (R43EB011886)
- Topics
- Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Engineering › Biomedical Engineering › Wireless Body Area Networks
- Keywords
- foreign body reaction, implantable glucose sensor, continuous glucose monitoring, diabetes management, implant size, chronic inflammation, fibrous encapsulation, sensor miniaturization, amperometric glucose sensor, microdialysis probe
- MeSH terms
- Animals, Foreign-Body Reaction, Male, Biosensing Techniques, Rats, Sprague-Dawley, Prostheses and Implants, Rats
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.