Implementation of Continuous Glucose Monitoring in the Hospital: Emergent Considerations for Remote Glucose Monitoring During the COVID-19 Pandemic
Open accessThe DOI matches its Crossref record; 31 of the 31 references it cites were found in a registry or an index.
Abstract
Continuous glucose monitoring (CGM) has become a widely used tool in the ambulatory setting for monitoring glucose levels, as well as detecting uncontrolled hyperglycemia, hypoglycemia, and glycemic variability. The accuracy of some CGM systems has recently improved to the point of manufacture with factory calibration and Food and Drug Administration clearance for nonadjunctive use to dose insulin. In this commentary, we analyze the answers to six questions about what is needed to bring CGM into the hospital as a reliable, safe, and effective tool. The evidence to date indicates that CGM offers promise as an effective tool for monitoring hospitalized patients. During the current coronavirus disease 2019 crisis, we hope to provide guidance to healthcare professionals, who are seeking to reduce exposure to SARS-Cov-2, as well as preserve invaluable personal protective equipment. In this commentary, we address who, what, where, when, why, and how CGM can be adopted for inpatient use.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures

Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“During the coronavirus disease 2019 (COVID-19) pandemic, many institutions used CGM in ICU and non-ICU settings, with the aim of minimizing exposure time and saving personal protective equipment, under an FDA policy of enforcement discretion (49,50).”
From 16. Diabetes Care in the Hospital: Standards of Care in Diabetes—2026 (2026) · cites this paper as [50]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Data on the safety and efficacy of real-time CGM use in the hospital, particularly with implementation of remote monitoring (e.g., a glucose telemetry system), has grown (47,50-52).”
From 16. Diabetes Care in the Hospital: Standards of Care in Diabetes—2026 (2026) · cites this paper as [50]
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“This approach has been taken to reduce the use of personal protective equipment and more closely monitor patients so that health care personnel do not have to go into a patient room solely to measure a glucose level (283-286).”
From 7. Diabetes Technology: Standards of Care in Diabetes—2025 (2025) · cites this paper as [286]
The cited source says what the claim attributes to it.
In the citing paper
“Studies have been published assessing the effectiveness of this approach, which may ultimately lead to the approved use of CGM for monitoring hospitalized individuals (277,286-295).”
From 7. Diabetes Technology: Standards of Care in Diabetes—2025 (2025) · cites this paper as [286]
The check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“The transition of persons from the hospital to ambulatory or subacute settings is a high-risk period for medication errors, but also an opportunity to improve glycemic control, avoiding what is often termed clinical inertia.102,1285,1286,1380”
From American Association of Clinical Endocrinology Clinical Practice Guideline: Developing a Diabetes Mellitus Comprehensive Care Plan-2022 Update. (2022) · cites this paper as [1380]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Sourcing Personal Protective Equipment During the COVID-19 Pandemic
Livingston, E., Desai, A., & Berkwits, M. (2020) · JAMA · doi:10.1001/jama.2020.5317 · Cited by 637
Wearable Devices for Ambulatory Cardiac Monitoring
Sana, F., Isselbacher, E. M., Singh, J. P., Heist, E. K., Pathik, B., & Armoundas, A. A. (2020) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2020.01.046 · Cited by 362
Umpierrez, G., Cardona, S., Pasquel, F., Jacobs, S., Peng, L., Unigwe, M., Newton, C. A., Smiley-Byrd, D., Vellanki, P., Halkos, M., et al. (2015) · Diabetes Care · doi:10.2337/dc15-0303 · Cited by 281
Self-citation15. Diabetes Care in the Hospital: Standards of Medical Care in Diabetes—2020
American Diabetes Association (2020) · Diabetes Care · doi:10.2337/dc20-s015 · Cited by 265 · cited 1×
“13 The current recommended target glucose range in critically ill patients is 140-180 mg/dL. 17 It is difficult to achieve this range with multiple daily SC insulin injections, particularly in patients receiving vasoactive agents, high doses of steroids, or medical nutrition therapy (MNT).”
Khanna, A. K., Bergese, S. D., Jungquist, C. R., Morimatsu, H., Uezono, S., Lee, S., Ti, L. K., Urman, R. D., McIntyre, R., Tornero, C., et al. (2020) · Anesthesia & Analgesia · doi:10.1213/ane.0000000000004788 · Cited by 209
Downey, C. L., Chapman, S., Randell, R., Brown, J. M., & Jayne, D. G. (2018) · International Journal of Nursing Studies · doi:10.1016/j.ijnurstu.2018.04.013 · Cited by 185
Feldman, B., Brazg, R., Schwartz, S., & Weinstein, R. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503322526978 · Cited by 184 · cited 1×
“22 Ascorbic acid may falsely elevate sensor glucose readings, while aspirin may slightly lower glucose readings with FreeStyle Libre.”
Galindo, R. J., Migdal, A. L., Davis, G. M., Urrutia, M. A., Albury, B., Zambrano, C., Vellanki, P., Pasquel, F. J., Fayfman, M., Peng, L., et al. (2020) · Diabetes Care · doi:10.2337/dc19-2073 · Cited by 181
Self-citationBergenstal, R. M., Ahmann, A. J., Bailey, T., Beck, R. W., Bissen, J., Buckingham, B., Deeb, L., Dolin, R. H., Garg, S. K., Goland, R., et al. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700234 · Cited by 166
Self-citationConsensus Statement on Inpatient Use of Continuous Glucose Monitoring
Wallia, A., Umpierrez, G. E., Rushakoff, R. J., Klonoff, D. C., Rubin, D. J., Hill Golden, S., Cook, C. B., Thompson, B., & Panel, T. D. C. G. M. I. T. H. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817706151 · Cited by 147 · cited 1×
“13 The current recommended target glucose range in critically ill patients is 140-180 mg/dL. 17 It is difficult to achieve this range with multiple daily SC insulin injections, particularly in patients receiving vasoactive agents, high doses of steroids, or medical nutrition therapy (MNT).”
Self-citation
Research works printed but not linked: 3
- Nacoti M Ciocca A Giupponi A, et al. At the epicenter of the Covid-19 pandemic and humanitarian crises in Italy: changing perspectives on preparation and mitigation. NEJM Catal. 2020; 1(2).
- Dexcom. Press release. https://www.dexcom.com/hospitalfacts. Accessed April 12, 2020.
- Abbott. Press release. https://abbott.mediaroom.com/2020-04-08-Abbotts-FreeStyle-R-Libre-14-Day-System-Now-Available-in-U-S-for-Hospitalized-Patients-with-Diabetes-During-COVID-19-Pandemic. Accessed April 12, 2020.
Cited by
Electrochemical glucose sensors in diabetes management: an updated review (2010–2020)
Teymourian, H., Barfidokht, A., & Wang, J. (2020) · Chemical Society Reviews · doi:10.1039/d0cs00304b · Cited by 881
Blonde, L., Umpierrez, G. E., Reddy, S. S., McGill, J. B., Berga, S. L., Bush, M., Chandrasekaran, S., DeFronzo, R. A., Einhorn, D., Galindo, R. J., et al. (2022) · Endocrine Practice · doi:10.1016/j.eprac.2022.08.002 · Cited by 507
Self-citation16. Diabetes Care in the Hospital: Standards of Medical Care in Diabetes—2022
American Diabetes Association Professional Practice Committee (2022) · Diabetes Care · doi:10.2337/dc22-s016 · Cited by 440
7. Diabetes Technology: Standards of Medical Care in Diabetes—2022
American Diabetes Association Professional Practice Committee (2022) · Diabetes Care · doi:10.2337/dc22-s007 · Cited by 358
“Studies are underway to assess the effectiveness of this approach, which may ultimately lead to the routine use of CGM for monitoring hospitalized patients (223,224).”
Background7. Diabetes Technology: Standards of Care in Diabetes—2025
American Diabetes Association Professional Practice Committee, ElSayed, N. A., McCoy, R. G., Aleppo, G., Balapattabi, K., Beverly, E. A., Briggs Early, K., Bruemmer, D., Echouffo-Tcheugui, J. B., Ekhlaspour, L., et al. (2025) · Diabetes Care · doi:10.2337/dc25-s007 · Cited by 348
Self-citationManagement of diabetes and hyperglycaemia in the hospital
Pasquel, F. J., Lansang, M. C., Dhatariya, K., & Umpierrez, G. E. (2021) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(20)30381-8 · Cited by 329
Self-citationKorytkowski, M. T., Muniyappa, R., Antinori-Lent, K., Donihi, A. C., Drincic, A. T., Hirsch, I. B., Luger, A., McDonnell, M. E., Murad, M. H., Nielsen, C., et al. (2022) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/clinem/dgac278 · Cited by 315
“Overall, the panel determined that the feasibility of introducing CGM for noncritically ill patients at high risk for hypoglycemia will vary by institution, and if implemented, a protocol for guiding the process is necessary for success (Table 2) (23, 24).”
BackgroundSelf-citation7. Diabetes Technology: Standards of Care in Diabetes — 2023
ElSayed, N. A., Aleppo, G., Aroda, V. R., Bannuru, R. R., Brown, F. M., Bruemmer, D., Collins, B. S., Hilliard, M. E., Isaacs, D., Johnson, E. L., et al. (2023) · Diabetes Care · doi:10.2337/dc23-s007 · Cited by 301
Self-citationGrunberger, G., Sherr, J., Allende, M., Blevins, T., Bode, B., Handelsman, Y., Hellman, R., Lajara, R., Roberts, V. L., Rodbard, D., et al. (2021) · Endocrine Practice · doi:10.1016/j.eprac.2021.04.008 · Cited by 288
7. Diabetes Technology: Standards of Care in Diabetes—2024
American Diabetes Association Professional Practice Committee, ElSayed, N. A., Aleppo, G., Bannuru, R. R., Bruemmer, D., Collins, B. S., Ekhlaspour, L., Hilliard, M. E., Johnson, E. L., Khunti, K., et al. (2024) · Diabetes Care · doi:10.2337/dc24-s007 · Cited by 281
Self-citation16. Diabetes Care in the Hospital: Standards of Care in Diabetes—2023
ElSayed, N. A., Aleppo, G., Aroda, V. R., Bannuru, R. R., Brown, F. M., Bruemmer, D., Collins, B. S., Hilliard, M. E., Isaacs, D., Johnson, E. L., et al. (2023) · Diabetes Care · doi:10.2337/dc23-s016 · Cited by 221
Self-citation7. Diabetes Technology: Standards of Medical Care in Diabetes—2021
American Diabetes Association (2021) · Diabetes Care · doi:10.2337/dc21-s007 · Cited by 202
“Studies are underway to assess the effectiveness of this approach, whichmay ultimately lead to the routine use of CGM for monitoring hospitalized patients (204,205).”
Background16. Diabetes Care in the Hospital: Standards of Care in Diabetes—2025
American Diabetes Association Professional Practice Committee, ElSayed, N. A., McCoy, R. G., Aleppo, G., Balapattabi, K., Beverly, E. A., Briggs Early, K., Bruemmer, D., Echouffo-Tcheugui, J. B., Ekhlaspour, L., et al. (2025) · Diabetes Care · doi:10.2337/dc25-s016 · Cited by 195
Self-citationContinuous glucose monitors and automated insulin dosing systems in the hospital consensus guideline
Galindo, R. J., Umpierrez, G. E., Rushakoff, R. J., Basu, A., Lohnes, S., Nichols, J. H., Spanakis, E. K., Espinoza, J., Palermo, N. E., Awadjie, D. G., et al. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820954163 · Cited by 170
“122,123,126 They can be divided into the following categories: (1) patientrelated, (2) hospital-related, (3) device-related, (4) medication-related, and (5) surgical procedure-related.”
BackgroundMethodInfluentialSelf-citation16. Diabetes Care in the Hospital: Standards of Care in Diabetes—2024
American Diabetes Association Professional Practice Committee, ElSayed, N. A., Aleppo, G., Bannuru, R. R., Bruemmer, D., Collins, B. S., Ekhlaspour, L., Galindo, R. J., Hilliard, M. E., Johnson, E. L., et al. (2024) · Diabetes Care · doi:10.2337/dc24-s016 · Cited by 167
Self-citationDavis, G. M., Spanakis, E. K., Migdal, A. L., Singh, L. G., Albury, B., Urrutia, M. A., Zamudio-Coronado, K. W., Scott, W. H., Doerfler, R., Lizama, S., et al. (2021) · Diabetes Care · doi:10.2337/dc20-2856 · Cited by 160
“During the coronavirus disease 2019 (COVID-19) pandemic, critical shortages of personal protective equipment led many health care professionals to explore the use of CGM to replace capillary POC BG testing in order to minimize infectious exposures without sacrificing glycemic control and monitoring (4,5).”
BackgroundSelf-citationContinuous glucose monitoring in the intensive care unit during the COVID-19 pandemic
Agarwal, S., Mathew, J., Davis, G. M., Shephardson, A., Levine, A., Louard, R., Urrutia, A., Perez-Guzman, C., Umpierrez, G. E., Peng, L., et al. (2021) · Diabetes Care · doi:10.2337/dc20-2219 · Cited by 138
“Mean number of hours/potential POC tests on CII was 72.125 h/;72 tests, compared with mean number of actual POC tests on CGM of 28, representing an estimated 60% reduction in POC tests during rtCGM use.”
MethodSelf-citationContinuous glucose monitoring: The achievement of 100 years of innovation in diabetes technology
Galindo, R. J., & Aleppo, G. (2020) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2020.108502 · Cited by 133
Self-citationPhillip, M., Bergenstal, R. M., Close, K. L., Danne, T., Garg, S. K., Heinemann, L., Hirsch, I. B., Kovatchev, B. P., Laffel, L. M., Mohan, V., et al. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2020.0375 · Cited by 135
COVID-19 in people living with diabetes: An international consensus
Caballero, A. E., Ceriello, A., Misra, A., Aschner, P., McDonnell, M. E., Hassanein, M., Ji, L., Mbanya, J. C., & Fonseca, V. A. (2020) · Journal of Diabetes and its Complications · doi:10.1016/j.jdiacomp.2020.107671 · Cited by 132
McCall, A. L., Lieb, D. C., Gianchandani, R., MacMaster, H., Maynard, G. A., Murad, M. H., Seaquist, E., Wolfsdorf, J. I., Wright, R. F., & Wiercioch, W. (2023) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/clinem/dgac596 · Cited by 123
Davis, G. M., Faulds, E., Walker, T., Vigliotti, D., Rabinovich, M., Hester, J., Peng, L., McLean, B., Hannon, P., Poindexter, N., et al. (2021) · Diabetes Care · doi:10.2337/dc20-2085 · Cited by 94
“pandemic to preserve personal protective equipment (PPE) and reduce exposures while maintaining glucose control (4,5).”
BackgroundSelf-citationContinuous glucose monitoring in the operating room and cardiac intensive care unit
Perez-Guzman, M. C., Duggan, E., Gibanica, S., Cardona, S., Corujo-Rodriguez, A., Faloye, A., Halkos, M., Umpierrez, G. E., Peng, L., Davis, G. M., et al. (2021) · Diabetes Care · doi:10.2337/dc20-2386 · Cited by 90
“real-timeCGMwithperiodicpoint-of-care (POC) validation has been suggested (1).”
BackgroundSelf-citationLongo, R. R., Elias, H., Khan, M., & Seley, J. J. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211008446 · Cited by 89
Diabetes and COVID-19; A Bidirectional Interplay
Kazakou, P., Lambadiari, V., Ikonomidis, I., Kountouri, A., Panagopoulos, G. N., Athanasopoulos, S., Korompoki, E., Kalomenidis, I., Dimopoulos, M. A., & Mitrakou, A. (2022) · Frontiers in Endocrinology · doi:10.3389/fendo.2022.780663 · Cited by 98
16. Diabetes Care in the Hospital: Standards of Care in Diabetes—2026
American Diabetes Association Professional Practice Committee for Diabetes*, Bajaj, M., McCoy, R. G., Balapattabi, K., Bannuru, R. R., Bellini, N. J., Bennett, A. K., Beverly, E. A., Briggs Early, K., ChallaSivaKanaka, S., et al. (2026) · Diabetes Care · doi:10.2337/dc26-s016 · Cited by 80
A Consensus Statement for Continuous Glucose Monitoring Metrics for Inpatient Clinical Trials
Spanakis, E. K., Cook, C. B., Kulasa, K. M., Aloi, J., Bally, L., Davis, G. M., Dungan, K. M., Galindo, R. J., Mendez, C. E., Pasquel, F. J., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231191104 · Cited by 93
“1,94 •• Research: Traditionally, 70% or more of the data from a 14-day report is considered acceptable for estimating HbA1c.”
BackgroundSelf-citationKanokpaka, P., Chang, Y., Chang, C.-C., Rinawati, M., Wang, P.-C., Chang, L., & Yeh, M. (2023) · Nano Energy · doi:10.1016/j.nanoen.2023.108513 · Cited by 78
Continuous Glucose Monitoring in the Hospital
Perez‐Guzman, M. C., Shang, T., Zhang, J. Y., Jornsay, D. L., & Klonoff, D. C. (2021) · Endocrinology and Metabolism · doi:10.3803/enm.2021.201 · Cited by 77
“Table 1 presents a list of currently available (in the United States) factory-calibrated subcutaneous CGMs that have product labels available and their known interferences from chemical substances [7-14].”
BackgroundSelf-citationShaw, J. L. V., Bannuru, R. R., Beach, L. A., ElSayed, N. A., Freckmann, G., Füzéry, A. K., Fung, A. W. S., Gilbert, J. D., Yun, H., Korpi-Steiner, N. L., et al. (2024) · Diabetes Care · doi:10.2337/dci24-0073 · Cited by 78
Self-citationPatel, P. R., Green, M. M., Tram, J. T., Wang, E., Murphy, M., Abd-Elsayed, A. A., & Chakravarthy, K. V. (2024) · Journal of Pain Research · doi:10.2147/jpr.s494238 · Cited by 67
“27,47 Effective integration of CGM in hospitals requires careful selection of devices based on their functionality and understanding potential interferences that could affect molecular data accuracy.”
Accuracy and Potential Interferences of Continuous Glucose Monitoring Sensors in the Hospital
Bellido, V., Freckman, G., Pérez, A., & Galindo, R. J. (2023) · Endocrine Practice · doi:10.1016/j.eprac.2023.06.007 · Cited by 57
Self-citationYeh, T., Yeung, M., & Mendelsohn Curanaj, F. A. (2021) · Current Diabetes Reports · doi:10.1007/s11892-021-01375-7 · Cited by 46
Espinoza, J., Xu, N. Y., Nguyen, K. T., & Klonoff, D. C. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211058148 · Cited by 46
Self-citationUpdate on the management of diabetes in long-term care facilities
Idrees, T., Castro-Revoredo, I. A., Migdal, A. L., Moreno, E. M., & Umpierrez, G. E. (2022) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2021-002705 · Cited by 44
Self-citationEmerging Diabetes Technologies: Continuous Glucose Monitors/Artificial Pancreases
ALMURASHI, A. M., Rodriguez, E., & Garg, S. K. (2023) · Journal of the Indian Institute of Science · doi:10.1007/s41745-022-00348-3 · Cited by 67
Hagerf, B. V., Protuš, M., Németová, L., Mráz, M., Kieslichová, E., Uchytilová, E., Indrova, V., Lelito, J., Girman, P., Haluzík, M., et al. (2024) · Diabetes Care · doi:10.2337/dc23-1663 · Cited by 49
Self-citationIdentifying the components of spirituality affecting the resilience of nurses
Meybodi, A. R., & Mohammadi, M. (2020) · Journal of Nursing Management · doi:10.1111/jonm.13235 · Cited by 36
Galindo, R. J., Parkin, C. G., Aleppo, G., Carlson, A. L., Kruger, D. F., Levy, C. J., Umpierrez, G. E., & McGill, J. B. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0107 · Cited by 39
Self-citationType 2 Diabetes Mellitus and COVID-19: A Narrative Review
Rey‐Reñones, C., Martínez‐Torres, S., Martín‐Luján, F., Pericas, C., Redondo, A., Vilaplana-Carnerero, C., Domínguez, Á., & Grau, M. (2022) · Biomedicines · doi:10.3390/biomedicines10092089 · Cited by 42
“Thus, insulin treatment results in an improvement in glycemic control in patients hospitalized in the ICU and offers the advantage of remote monitoring [57,58].”
BackgroundRemote patient monitoring after cardiac surgery: The utility of a novel telemedicine system
Atılgan, K., Onuk, B. E., Coşkun, P. K., Yeşil, F. G., Aslan, C., Çolak, A., Çelebi, S., & Bozbaş, H. (2021) · Journal of Cardiac Surgery · doi:10.1111/jocs.15962 · Cited by 40
“25,26 The limitation of the study is that the outcomes of using remote monitoring could not be compared with a control group due to ethical considerations, and thus, all patients accepting to participate in the study received remote monitoring.”
BackgroundZelada, H., Perez‐Guzman, M. C., Cherñavvsky, D. R., & Galindo, R. J. (2023) · Endocrine Connections · doi:10.1530/ec-23-0180 · Cited by 37
“As recommended by experts and guidelines, engagement, training and education of nursing personnel is required for safe implementation of CGM in the hospital (11, 16).”
BackgroundMethodInfluentialSelf-citationChagas, G. C. L., Teixeira, L., Clemente, M. R. C., Chagas, R. C. L., Pestana, D. V. S., Sombra, L. R. S., Lima, B. B., Galindo, R. J., & Abreu, M. (2025) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2024.111986 · Cited by 45
Self-citationContinuous Glucose Monitoring in the Intensive Care Unit
ARROYO, L. I. G., Faulds, E. R., Perez‐Guzman, M. C., Davis, G. M., Dungan, K. M., & Pasquel, F. J. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231169522 · Cited by 36
Self-citationGómez, A. M., Henao, D. C., Múñoz, Ó., Aschner, P. J., Yepes, C. A., Jojoa, R. I. J., Kerguelen, A., Parra, D., Jaramillo, P., & Umpierrez, G. E. (2021) · Diabetes & Metabolic Syndrome Clinical Research & Reviews · doi:10.1016/j.dsx.2021.02.008 · Cited by 31
“Nephropathy 4 [6,7] Diabetic foot 1 [1,7] Length of hospital stay, days, median (IQR) 11 [8e16] Labs at admission n (%) HbA1c > 7% 33 (58,9) Lymphocytes < 1000 28 (46,7) RCP > 20 15 (27,8) LDH > 400 22 (38,6)”
Self-citationGujral, U. P., Johnson, L. C. M., Nielsen, J., Vellanki, P., HAW, J. S., Davis, G. M., Weber, M. B., & Pasquel, F. J. (2020) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2020-001520 · Cited by 33
Self-citationBOEDER, S. C., Kobayashi, E., Ramesh, G., Serences, B., Kulasa, K. M., & Majithia, A. R. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221113865 · Cited by 31
Baker, M., Musselman, M. E., Rogers, R., & Hellman, R. (2022) · American Journal of Health-System Pharmacy · doi:10.1093/ajhp/zxab456 · Cited by 27
Zhang, J. Y., Shang, T., Ahn, D. T., Chen, K. Y., Coté, G. L., Espinoza, J., Mendez, C. E., Spanakis, E. K., Thompson, B., Wallia, A., et al. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820978399 · Cited by 33
Self-citationTu, L. M. T. P. P. L. M., Thu, V. T. A., Hoang, T. X., Selvam, S. P., Pham, H. L., Kim, J. Y., & Cho, S. (2021) · Advanced Materials Technologies · doi:10.1002/admt.202100020 · Cited by 27
Kruger, D. F., & Anderson, J. E. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0193 · Cited by 30
Reliability of continuous glucose monitoring system in the inpatient setting
Murray-Bachmann, R., Leung, T. M., Myers, A. K., Murthi, S., Sarbanes, M., Ziskovich, K., LESSER, M. L., & Poretsky, L. (2021) · Journal of Clinical & Translational Endocrinology · doi:10.1016/j.jcte.2021.100262 · Cited by 25
“[10] Galindo RJ, Aleppo G, Klonoff DC, Spanakis EK, Agarwal S, Vellanki P, et al.”
BackgroundBellido, V., & Pérez, A. (2021) · Journal of Clinical Medicine · doi:10.3390/jcm10225341 · Cited by 31
“Potential candidates include patients with moderate-severe hyperglycemia requiring treatment with multiple doses of insulin, patients with high glycemic variability or risk of hypoglycemia, and patients with hyperglycemia that is difficult to manage, such as corticosteroid-induced hyperglycemia or hyperglycemia induce…”
BackgroundGarg, S., & Norman, G. J. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2020.0656 · Cited by 24
“This decision allowed hospitals to use CGM for remote monitoring of patients to reduce personal protective equipment use and limit health care workers’ exposure to the virus.(53) Several recent publications report on inpatient hospital use of CGM during the COVID-19 pandemic.”
MethodOprea, A. D., Kalra, S. K., Duggan, E. W., Russell, L., Urman, R. D., Abdelmalak, B., Patel, P., Pfeifer, K. J., Grant, P., Charitou, M. M., et al. (2024) · Journal of Clinical Anesthesia · doi:10.1016/j.jclinane.2024.111627 · Cited by 22
Self-citationOlsen, M. T., Klarskov, C. K., Pedersen‐Bjergaard, U., Hansen, K. B., & Kristensen, P. L. (2024) · BMC Endocrine Disorders · doi:10.1186/s12902-024-01595-4 · Cited by 22
Wang, Y., Li, S., Lu, J., Feng, K., Huang, X., Hu, F., Sun, M., Zou, Y., Li, Y., Huang, W., et al. (2024) · Diabetologia · doi:10.1007/s00125-024-06136-1 · Cited by 23
“Furthermore, the use of CGM in hospitals has been accelerated due to the COVID-19 pandemic [9–11].”
Efficacy of Hybrid Closed-Loop Insulin Delivery System in a Hospital Setting: A Case Series
Atif, Z., Halstrom, A., Peragallo‐Dittko, V., & Klek, S. P. (2021) · AACE Clinical Case Reports · doi:10.1016/j.aace.2020.12.013 · Cited by 23
Blissett, R., Blissett, D. B., Levrat‐Guillen, F., Deshmukh, H., Wilmot, E. G., Ryder, R. E. J., Walton, C., & Sathyapalan, T. (2022) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2021-002580 · Cited by 24
Galindo, R. J., Aleppo, G., Parkin, C. G., Baidal, D. A., Carlson, A. L., Cengiz, E., Forlenza, G. P., Kruger, D. F., Levy, C. J., McGill, J. B., et al. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221144052 · Cited by 19
Self-citationDumitrascu, A. G., Perry, M. F., Boone, R. J., Guzman, M. P., Chirilă, R. M., McNally, A. W., Colibaseanu, D., Meek, S., Ball, C. T., White, L. J., et al. (2022) · Endocrine Practice · doi:10.1016/j.eprac.2022.12.012 · Cited by 15
“The median 155 duration of hospitalization was 11 (4-61) days.”
BackgroundContinuous glucose monitoring (CGM) in a non-Icu hospital setting: The patient's journey
Irace, C., Coluzzi, S., Cianni, G. D., Forte, E., Landi, F., Rizzo, M. R., Sesti, G., Succurro, E., & Consoli, A. (2023) · Nutrition Metabolism and Cardiovascular Diseases · doi:10.1016/j.numecd.2023.06.021 · Cited by 19
How use of continuous glucose monitoring can address therapeutic inertia in primary care
Martens, T. W., & Parkin, C. G. (2022) · Postgraduate Medicine · doi:10.1080/00325481.2022.2080419 · Cited by 16
Longo, R. R., & Joshi, R. (2022) · Diabetes Spectrum · doi:10.2337/dsi22-0009 · Cited by 20
Tian, T., Aaron, R. E., Seley, J. J., Longo, R. R., Nayberg, I., Umpierrez, G. E., Levy, C. J., & Klonoff, D. C. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231200847 · Cited by 18
Self-citationHughes, M. S., Pasquel, F. J., Davis, G. M., Lal, R., Buckingham, B. A., Boughton, C. K., Brown, S. A., & Bally, L. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2024.0343 · Cited by 14
“8,85,88 While these solutions do not integrate with EHR or current inpatient telemetry systems, a sophisticated work fl ow using cloud platforms for remote 24/7 centralized telemetry monitoring of CGM for hypoglycemia has been implemented at Scripps Health and is being tested in a large RCT…”
Self-citationRajpal, A., Kassem, L. S., & Aron, D. C. (2021) · Expert Review of Endocrinology & Metabolism · doi:10.1080/17446651.2021.1927708 · Cited by 18
“also be used for non-critically ill patients [41,48,49], it is important to prioritize this resource to critically ill patients or patients on general wards requiring complex insulin regimens or experiencing hypoglycemic events or in patients with”
BackgroundMethodCardio‐renal‐metabolic disease in primary care setting
Ibrahim, M. A., Ba-Essa, E. M., Baker, J. C., Cahn, A., Ceriello, A., Cosentino, F., Davies, M. J., Eckel, R. H., Gaal, L. V., Gæde, P., et al. (2023) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.3755 · Cited by 14
Self-citationReview Article – Diabetes Technology in the Hospital: An Update
Visser, M. M., Vangoitsenhoven, R., Gillard, P., & Mathieu, C. (2024) · Current Diabetes Reports · doi:10.1007/s11892-024-01545-3 · Cited by 17
“During the COVID-19 pandemic, CGM demonstrated the potential to avoid bedside POCT and reduce the burden of diabetes care on healthcare providers, while reducing the risk of viral exposure [35].”
Yu, T. S., Song, S., Yea, J., & Jang, K. (2024) · Advanced Sensor Research · doi:10.1002/adsr.202400048 · Cited by 16
Klarskov, C. K., & Kristensen, P. L. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/1932296821993169 · Cited by 12
Continuous Glucose Monitoring in Pregnancy
Horgan, R., Diab, Y. H., Bartal, M. F., Sibai, B. M., & Saade, G. R. (2023) · Obstetrics and Gynecology · doi:10.1097/aog.0000000000005374 · Cited by 15
Ang, L., Lin, Y. K., Schroeder, L. F., Huang, Y., DeGeorge, C. A., Arnold, P., Akanbi, F., Knotts, S., DuBois, E., Desbrough, N., et al. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968241241005 · Cited by 11
One Size Fits All Versus Individualized Medicine in Type 1 Diabetes Management
Midyett, L. K. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0109 · Cited by 18
Shang, J., Yuan, Z., Zhang, Z., Zhou, Q., Zou, Y., & Wang, W. (2025) · Journal of Medical Internet Research · doi:10.2196/67012 · Cited by 12
“It is now widely used, medically prescribed, or commercially available for both patients with and without diabetes [2,3].”
Ullal, J., Spanbauer, C., Baran, J. D., Bais, R., Chiang, H.-H., Chao, J. H., Khakpour, D., Panfil, P., Pasquel, F. J., Gligorijevic, N., et al. (2025) · Diabetes Technology & Therapeutics · doi:10.1177/15209156251368933 · Cited by 8
“5 CGM may reduce the need for point-of-care testing or other invasive glucose testing by venipuncture or arterial draws while meeting the need for frequent glucose measurements and monitoring of glucose fl uctuations, with continuous tracking of glycemic variability in critically ill patients.”
Self-citationIflaifel, M., Lim, R. H. M., Crowley, C. M., Ryan, K., & Greco, F. (2021) · BMJ Open · doi:10.1136/bmjopen-2020-041848 · Cited by 9
Insurance Coverage for Emerging Diabetes Technologies: Payers' Perspective
Nolan, T., & Frazzitta, M. A. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0262 · Cited by 6
Moscardó, V., García, A., Bondía, J., Diaz, J., Ramos‐Prol, A., & Rossetti, P. (2022) · Sensors · doi:10.3390/s22093101 · Cited by 7
“Nevertheless, one of the recognized limitations of CGM implementation in the hospital setting is the unknown accuracy under conditions of metabolic acidosis [7] although factors other than pH (such as hypovolemia with peripheral tissue hypoperfusion) may well affect glucose-sensor readings in these cases.”
BackgroundHypoglycemia symptoms are reduced in hospitalized patients with diabetes
Criner, K. E., Kim, H. N., Ali, H., Kumar, S., Kanter, J. E., Wang, L., & Korytkowski, M. T. (2021) · Journal of Diabetes and its Complications · doi:10.1016/j.jdiacomp.2021.107976 · Cited by 8
Consumer Wearables for Patient Monitoring in Otolaryngology: A State of the Art Review
Somani, S. N., Yu, K. M., Chiu, A. G., Sykes, K. J., & Villwock, J. A. (2021) · Otolaryngology · doi:10.1177/01945998211061681 · Cited by 7
Demidowich, A. P., Stanback, C., & Zilbermint, M. F. (2024) · Annals of the New York Academy of Sciences · doi:10.1111/nyas.15190 · Cited by 14
Wang, R., Kyi, M., Krishnamoorthi, B., Tjahyadi, J., Connell, A., Chiang, C. Y., Renouf, D., Barmanray, R., & Fourlanos, S. (2025) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2024.0604 · Cited by 11
“15 Establishing CGM device accuracy in hospital is crucial to ensure this technology can be safely applied to the inpatient setting.”
Philis‐Tsimikas, A., Diego, E. R. N. S., Vincent, L., Lohnes, S., & Singleton, C. L. (2024) · Current Diabetes Reports · doi:10.1007/s11892-024-01549-z · Cited by 6
“Several groups adopted and reported on the use of rtCGM during the COVID-19 pandemic when the FDA provided an allowance for use in the hospital setting [49–52].”
Design of Breath Acetone Detection for Diabetes Detection and Classification using Electronic Nose
Evangelista, G. G., Paglinawan, A. C., & Paglinawan, C. C. (2024) · International Conference on Computer and Automation Engineering · doi:10.1109/iccae59995.2024.10569837 · Cited by 5
“However, in other cases, a negative correlation was observed [19].”
BackgroundEndocrinology in the Time of COVID-19: A Rapid Evolution of Knowledge and Care
Rizvi, A. A., Stoian, A. P., Lessan, N. G., & Rizzo, M. (2021) · Medicina · doi:10.3390/medicina57080805 · Cited by 5
“The management of diabetes in the hospital setting received a boost with no objection from the FDA to the use of continuous glucose monitoring techniques; this step was initially taken to enhance the protection of health professionals caring for patients with known or suspected COVID-19 virus [20].”
BackgroundMüssig, K., Gallwitz, B., Haak, T., Kellerer, M., & Siegel, E. G. (2025) · Diabetologie und Stoffwechsel · doi:10.1055/a-2592-9700 · Cited by 7
Role of hyperglycemia in cystic fibrosis pulmonary exacerbations
Merjaneh, L., Sidhaye, A. R., Vu, P. T., Heltshe, S. L., Goss, C. H., Flume, P. A., Kelly, A. T., & Rosenfeld, M. (2023) · Journal of Cystic Fibrosis · doi:10.1016/j.jcf.2023.06.011 · Cited by 4
Kanjee, Z., Brown, F. M., Taxin, Z. H., & Smetana, G. W. (2024) · Annals of Internal Medicine · doi:10.7326/annals-24-01100 · Cited by 5
Non-invasive blood glucose level monitoring using antennas: a comprehensive review report
Upadhyay, N., Kaur, A., Pattanayak, A., & Singh, A. (2025) · Discover Electronics · doi:10.1007/s44291-025-00068-9 · Cited by 5
“Hence, developing a true non-invasive technique for reliable BGL monitoring is the essential need of the hour [17].”
Shimizu, N., Hayashi, A., Ito, S., Suzuki, A., Fujishima, R., Matoba, K., Wada, T., Takano, K., Katagiri, M., & Shichiri, M. (2021) · Endocrine Journal · doi:10.1507/endocrj.ej21-0540 · Cited by 4
“Libre system has the advantage of being cheaper than previously reported remote realtime CGMs [5-11].”
BackgroundResultSiemens, R. L. (2022) · Clinical Diabetes · doi:10.2337/cd21-0088 · Cited by 6
Guideline Development for Medical Device Technology: Issues for Consideration
Battelino, T., Brosius, F. C., Ceriello, A., Cosentino, F., Green, J., Kellerer, M., Koob, S., Kosiborod, M., Lalic, N. M., Marx, N., et al. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221093355 · Cited by 5
Survey of Diabetes Technology in the Pediatric Inpatient Setting
Hamilton, A. B., Rajagopal, N., Ngo, C., Hendrix-Dicken, A. D., Mercer, S., & Condren, M. (2025) · The Journal of Pediatric Pharmacology and Therapeutics · doi:10.5863/1551-6776-30.1.123 · Cited by 3
Müssig, K., Gallwitz, B., Haak, T., Kellerer, M., & Siegel, E. G. (2024) · Diabetologie und Stoffwechsel · doi:10.1055/a-2251-2772 · Cited by 3
Müssig, K., Gallwitz, B., Haak, T., Kellerer, M., & Siegel, E. G. (2025) · Diabetes aktuell · doi:10.1055/a-2500-1766 · Cited by 5
Nugroho, B. S. N., Rizal, M. F., Bangkara, B. M. A. S. A., Ganiem, L. M., & Purwoko, R. Y. (2022) · Science Midwifery · doi:10.35335/midwifery.v10i5.930 · Cited by 4
“The subjects of two recent commentaries published in Nature Medicine are the promise of digital tools during a pandemic and some ethical and societal issues that need to be addressed to make their promise for precision public health a reality in the long run (Galindo et al., 2020).”
BackgroundMüssig, K., Gallwitz, B., Haak, T., Kellerer, M., & Siegel, E. G. (2024) · Diabetologie und Stoffwechsel · doi:10.1055/a-2312-1160 · Cited by 3
Wang, R., Krishnamoorthi, B., Connell, A., Chiang, C. Y., Renouf, D., Barmanray, R., Kyi, M., & Fourlanos, S. (2025) · Internal Medicine Journal · doi:10.1111/imj.70095 · Cited by 2
Setford, S. J. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251377019 · Cited by 2
“Published literature is sparse regarding the clinical accuracy of first-generation CGM devices under hypoxic conditions, 51 which may prevail in some critically ill patients or maybe due to compression artifacts at the CGM wear site due to body tissue pressurization and subsequent gradual oxygen…”
Metadata
- Authors
- Galindo, R. J., Aleppo, G., Klonoff, D. C., Spanakis, E. K., Agarwal, S., Vellanki, P., Olson, D. E., Umpierrez, G. E., Davis, G. M., & Pasquel, F. J.
- Affiliations
- Emory University (US)
- Northwestern University (US)
- San Mateo Medical Center (US)
- University of Maryland, Baltimore (US)
- Albert Einstein College of Medicine (US)
- Corresponding author
- Pasquel, Francisco J.
- ORCID
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2020-06-14
- Volume
- 14
- Issue
- 4
- Pages
- 822-832
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC7673156
- OpenAlex
- W3034681791
- Cited by
- 155
- References
- 31
- Citations by year
- 2026: 15 · 2025: 24 · 2024: 26 · 2023: 23 · 2022: 20 · 2021: 36 · 2020: 11
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Locations
- Journal of Diabetes Science and Technology (publishedVersion, cc-by)
- PubMed (publishedVersion)
- PubMed Central (publishedVersion)
- Funders
- National Institute of Diabetes and Digestive and Kidney Diseases (P30-DK-111024)
- national institutes of health (1K23DK123384-01, P30DK11102, K23DK115896, P30DK111022, 1K23DK113241, 1K23GM128221-01A1)
- VA MERIT (#1I01CX001825)
- NIH/NATS (UL1 TR002378, 1P30DK111024-01)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Hyperglycemia and glycemic control in critically ill and hospitalized patients
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Treatment and Management
- Keywords
- continuous glucose monitoring, hospital implementation, remote glucose monitoring, COVID-19 pandemic, personal protective equipment, hospitalized patients, glycemic variability, hypoglycemia, hyperglycemia, insulin dose, inpatient diabetes management
- MeSH terms
- Betacoronavirus, COVID-19, SARS-CoV-2, Blood Glucose, Calibration, Communicable Disease Control, Diabetes Mellitus, Hospitalization, Hospitals, Humans, Hyperglycemia, Inpatients, Insulin Infusion Systems, Pneumonia, Viral, United States, United States Food and Drug Administration, Blood Glucose Self-Monitoring, Coronavirus Infections, Monitoring, Ambulatory, Diabetes Complications, Electronic Health Records, Pandemics
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.