Recent developments in time-of-flight PET
Open accessThe DOI matches its Crossref record; 64 of the 64 references it cites were found in a registry or an index.
Abstract
While the first time-of-flight (TOF)-positron emission tomography (PET) systems were already built in the early 1980s, limited clinical studies were acquired on these scanners. PET was still a research tool, and the available TOF-PET systems were experimental. Due to a combination of low stopping power and limited spatial resolution (caused by limited light output of the scintillators), these systems could not compete with bismuth germanate (BGO)-based PET scanners. Developments on TOF system were limited for about a decade but started again around 2000. The combination of fast photomultipliers, scintillators with high density, modern electronics, and faster computing power for image reconstruction have made it possible to introduce this principle in clinical TOF-PET systems. This paper reviews recent developments in system design, image reconstruction, corrections, and the potential in new applications for TOF-PET. After explaining the basic principles of time-of-flight, the difficulties in detector technology and electronics to obtain a good and stable timing resolution are shortly explained. The available clinical systems and prototypes under development are described in detail. The development of this type of PET scanner also requires modified image reconstruction with accurate modeling and correction methods. The additional dimension introduced by the time difference motivates a shift from sinogram- to listmode-based reconstruction. This reconstruction is however rather slow and therefore rebinning techniques specific for TOF data have been proposed. The main motivation for TOF-PET remains the large potential for image quality improvement and more accurate quantification for a given number of counts. The gain is related to the ratio of object size and spatial extent of the TOF kernel and is therefore particularly relevant for heavy patients, where image quality degrades significantly due to increased attenuation (low counts) and high scatter fractions. The original calculations for the gain were based on analytical methods. Recent publications for iterative reconstruction have shown that it is difficult to quantify TOF gain into one factor. The gain depends on the measured distribution, the location within the object, and the count rate. In a clinical situation, the gain can be used to either increase the standardized uptake value (SUV) or reduce the image acquisition time or administered dose. The localized nature of the TOF kernel makes it possible to utilize local tomography reconstruction or to separate emission from transmission data. The introduction of TOF also improves the joint estimation of transmission and emission images from emission data only. TOF is also interesting for new applications of PET-like isotopes with low branching ratio for positron fraction. The local nature also reduces the need for fine angular sampling, which makes TOF interesting for limited angle situations like breast PET and online dose imaging in proton or hadron therapy. The aim of this review is to introduce the reader in an educational way into the topic of TOF-PET and to give an overview of the benefits and new opportunities in using this additional information.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Performance Measurements of the Siemens mMR Integrated Whole-Body PET/MR Scanner
Delso, G., Fürst, S., Jakoby, B., Ladebeck, R., Ganter, C., Nekolla, S. G., Schwaiger, M., & Ziegler, S. I. (2011) · Journal of Nuclear Medicine · doi:10.2967/jnumed.111.092726 · Cited by 874
Surti, S., Kuhn, A., Werner, M. E., Perkins, A. E., Kolthammer, J. A., & Karp, J. S. (2007) · PubMed · Cited by 639
Self-citationBenefit of Time-of-Flight in PET: Experimental and Clinical Results
Karp, J. S., Surti, S., Daube-Witherspoon, M. E., & Muehllehner, G. (2008) · Journal of Nuclear Medicine · doi:10.2967/jnumed.107.044834 · Cited by 584
Self-citationPhysical and clinical performance of the mCT time-of-flight PET/CT scanner
Jakoby, B. W., Bercier, Y., Conti, M., Casey, M. E., Bendriem, B., & Townsend, D. W. (2011) · Physics in Medicine and Biology · doi:10.1088/0031-9155/56/8/004 · Cited by 492
Design and performance evaluation of a whole-body Ingenuity TF PET–MRI system
Zaidi, H., Ojha, N., Morich, M., Griesmer, J., Hu, Z., Maniawski, P., Ratib, O., Izquierdo-Garcia, D., Fayad, Z. A., & Shao, L. (2011) · Physics in Medicine and Biology · doi:10.1088/0031-9155/56/10/013 · Cited by 402
Physical Performance of the new hybrid PET/CT Discovery-690
Bettinardi, V., Presotto, L., Rapisarda, E., Picchio, M., Gianolli, L., & Gilardi, M. C. (2011) · Medical Physics · doi:10.1118/1.3635220 · Cited by 380
Nuyts, J., Dupont, P., Stroobants, S., Benninck, R., Mortelmans, L., & Suetens, P. (1999) · IEEE Transactions on Medical Imaging · doi:10.1109/42.774167 · Cited by 362
List-mode likelihood: EM algorithm and image quality estimation demonstrated on 2-D PET
Parra, L., & Barrett, H. H. (1998) · IEEE Transactions on Medical Imaging · doi:10.1109/42.700734 · Cited by 357
Time of flight in PET revisited
Moses, W. W. (2003) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2003.817319 · Cited by 336
Focus on time-of-flight PET: the benefits of improved time resolution
Conti, M. (2011) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-010-1711-y · Cited by 335 · cited 1×
“The main motivation for TOF-PET has always been the potential image quality improvement or reduction in image acquisition time of TOF-PET [9, 15, 16].”
Background
Cited by
State of the art in total body PET
Vandenberghe, S., Moskal, P., & Karp, J. S. (2020) · EJNMMI Physics · doi:10.1186/s40658-020-00290-2 · Cited by 390
Self-citationRoadmap toward the 10 ps time-of-flight PET challenge
Lecoq, P., Morel, C., Prior, J. O., Visvikis, D., Gundacker, S., Auffray, E., Križan, P., Turtos, R. M., Thers, D., Charbon, E., et al. (2020) · Physics in Medicine & Biology · doi:10.1088/1361-6560/ab9500 · Cited by 258
Self-citationQuantification, improvement, and harmonization of small lesion detection with state-of-the-art PET
van der Vos, C. S., Koopman, D., Rijnsdorp, S., Arends, A. J., Boellaard, R., van Dalen, J. A., Lubberink, M., Willemsen, A. T. M., & Visser, E. P. (2017) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-017-3727-z · Cited by 206
“For such a radionuclide, TOF results in a better SNR for the same number of counts [5, 87].”
BackgroundMethodInfluentialPerformance Evaluation of the Vereos PET/CT System According to the NEMA NU2-2012 Standard
Rausch, I., Ruiz, A., Valverde-Pascual, I., Cal-González, J., Beyer, T., & Carrio, I. (2019) · Journal of Nuclear Medicine · doi:10.2967/jnumed.118.215541 · Cited by 168
“With these new detector technologies, the performance of PET systems can be substantially improved (16).”
BackgroundPerformance evaluation of the next generation solid-state digital photon counting PET/CT system
Zhang, J., Maniawski, P., & Knopp, M. V. (2018) · EJNMMI Research · doi:10.1186/s13550-018-0448-7 · Cited by 123
Nuclear medicine and molecular imaging advances in the 21st century
Vaz, S. C., Oliveira, F., Herrmann, K., & Veit-Haibach, P. (2020) · The British Journal of Radiology · doi:10.1259/bjr.20200095 · Cited by 86
Reduction of 18 F-FDG Dose in Clinical PET/MR Imaging by Using Silicon Photomultiplier Detectors
Sekine, T., Delso, G., Zeimpekis, K. G., de Galiza Barbosa, F., ter Voert, E. E. G. W., Huellner, M., & Veit-Haibach, P. (2018) · Radiology · doi:10.1148/radiol.2017162305 · Cited by 79
“For example, images of the liver usually have a high degree of noise, and this certainly in our results and has already been reported (21).”
ResultSchug, D., Nadig, V., Weissler, B., Gebhardt, P., & Schulz, V. (2019) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2018.2884564 · Cited by 56
“[coincidence resolving time (CRT)] with a timing resolution in the order of a few hundred picoseconds [1], [2].”
BackgroundCaribé, P. R. R. V., Koole, M., D’Asseler, Y., Van Den Broeck, B., & Vandenberghe, S. (2019) · EJNMMI Physics · doi:10.1186/s40658-019-0264-9 · Cited by 51
Self-citationSalvadori, J., Imbert, L., Perrin, M., Karcher, G., Lamiral, Z., Marie, P.-Y., & Verger, A. (2019) · EJNMMI Research · doi:10.1186/s13550-019-0526-5 · Cited by 44
Pilz, J., Hehenwarter, L., Zimmermann, G., Rendl, G., Schweighofer-Zwink, G., Beheshti, M., & Pirich, C. (2021) · EJNMMI Research · doi:10.1186/s13550-021-00784-9 · Cited by 6
Metadata
- Authors
- Vandenberghe, S., Mikhaylova, E., D’Hoe, E., Mollet, P., & Karp, J. S.
- Venue
- EJNMMI Physics
- Published
- 2016-02-16
- Volume
- 3
- Issue
- 1
- Pages
- 3-3
- ISSN
- 2197-7364 (ISSN-L)
- OpenAlex
- W2281211957
- Cited by
- 368
- References
- 64
- FWCI
- 31.27
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.