Recent developments in time-of-flight PET
Open accessThe DOI matches its Crossref record; 64 of the 64 references it cites were found in a registry or an index.
Abstract
While the first time-of-flight (TOF)-positron emission tomography (PET) systems were already built in the early 1980s, limited clinical studies were acquired on these scanners. PET was still a research tool, and the available TOF-PET systems were experimental. Due to a combination of low stopping power and limited spatial resolution (caused by limited light output of the scintillators), these systems could not compete with bismuth germanate (BGO)-based PET scanners. Developments on TOF system were limited for about a decade but started again around 2000. The combination of fast photomultipliers, scintillators with high density, modern electronics, and faster computing power for image reconstruction have made it possible to introduce this principle in clinical TOF-PET systems. This paper reviews recent developments in system design, image reconstruction, corrections, and the potential in new applications for TOF-PET. After explaining the basic principles of time-of-flight, the difficulties in detector technology and electronics to obtain a good and stable timing resolution are shortly explained. The available clinical systems and prototypes under development are described in detail. The development of this type of PET scanner also requires modified image reconstruction with accurate modeling and correction methods. The additional dimension introduced by the time difference motivates a shift from sinogram- to listmode-based reconstruction. This reconstruction is however rather slow and therefore rebinning techniques specific for TOF data have been proposed. The main motivation for TOF-PET remains the large potential for image quality improvement and more accurate quantification for a given number of counts. The gain is related to the ratio of object size and spatial extent of the TOF kernel and is therefore particularly relevant for heavy patients, where image quality degrades significantly due to increased attenuation (low counts) and high scatter fractions. The original calculations for the gain were based on analytical methods. Recent publications for iterative reconstruction have shown that it is difficult to quantify TOF gain into one factor. The gain depends on the measured distribution, the location within the object, and the count rate. In a clinical situation, the gain can be used to either increase the standardized uptake value (SUV) or reduce the image acquisition time or administered dose. The localized nature of the TOF kernel makes it possible to utilize local tomography reconstruction or to separate emission from transmission data. The introduction of TOF also improves the joint estimation of transmission and emission images from emission data only. TOF is also interesting for new applications of PET-like isotopes with low branching ratio for positron fraction. The local nature also reduces the need for fine angular sampling, which makes TOF interesting for limited angle situations like breast PET and online dose imaging in proton or hadron therapy. The aim of this review is to introduce the reader in an educational way into the topic of TOF-PET and to give an overview of the benefits and new opportunities in using this additional information.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Performance Measurements of the Siemens mMR Integrated Whole-Body PET/MR Scanner
Delso, G., Fürst, S., Jakoby, B., Ladebeck, R., Ganter, C., Nekolla, S. G., Schwaiger, M., & Ziegler, S. I. (2011) · Journal of Nuclear Medicine · doi:10.2967/jnumed.111.092726 · Cited by 874
Surti, S., Kuhn, A., Werner, M. E., Perkins, A. E., Kolthammer, J. A., & Karp, J. S. (2007) · PubMed · Cited by 639
Self-citationBenefit of Time-of-Flight in PET: Experimental and Clinical Results
Karp, J. S., Surti, S., Daube-Witherspoon, M. E., & Muehllehner, G. (2008) · Journal of Nuclear Medicine · doi:10.2967/jnumed.107.044834 · Cited by 584
Self-citationPhysical and clinical performance of the mCT time-of-flight PET/CT scanner
Jakoby, B. W., Bercier, Y., Conti, M., Casey, M. E., Bendriem, B., & Townsend, D. W. (2011) · Physics in Medicine and Biology · doi:10.1088/0031-9155/56/8/004 · Cited by 492
Design and performance evaluation of a whole-body Ingenuity TF PET–MRI system
Zaidi, H., Ojha, N., Morich, M., Griesmer, J., Hu, Z., Maniawski, P., Ratib, O., Izquierdo-Garcia, D., Fayad, Z. A., & Shao, L. (2011) · Physics in Medicine and Biology · doi:10.1088/0031-9155/56/10/013 · Cited by 402
Physical Performance of the new hybrid PET/CT Discovery-690
Bettinardi, V., Presotto, L., Rapisarda, E., Picchio, M., Gianolli, L., & Gilardi, M. C. (2011) · Medical Physics · doi:10.1118/1.3635220 · Cited by 380
Nuyts, J., Dupont, P., Stroobants, S., Benninck, R., Mortelmans, L., & Suetens, P. (1999) · IEEE Transactions on Medical Imaging · doi:10.1109/42.774167 · Cited by 362
List-mode likelihood: EM algorithm and image quality estimation demonstrated on 2-D PET
Parra, L., & Barrett, H. H. (1998) · IEEE Transactions on Medical Imaging · doi:10.1109/42.700734 · Cited by 357
Time of flight in PET revisited
Moses, W. W. (2003) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2003.817319 · Cited by 336
Focus on time-of-flight PET: the benefits of improved time resolution
Conti, M. (2011) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-010-1711-y · Cited by 335 · cited 1×
“The main motivation for TOF-PET has always been the potential image quality improvement or reduction in image acquisition time of TOF-PET [9, 15, 16].”
BackgroundThe digital silicon photomultiplier — Principle of operation and intrinsic detector performance
Frach, T., Prescher, G., Degenhardt, C., de Gruyter, R., Schmitz, A., & Ballizany, R. (2009) · 2009 IEEE Nuclear Science Symposium Conference Record (NSS/MIC) · doi:10.1109/nssmic.2009.5402143 · Cited by 317
Update on time-of-flight PET imaging
Surti, S. (2015) · Journal of Nuclear Medicine · doi:10.2967/jnumed.114.145029 · Cited by 303 · cited 1×
“A more extensive description of introduction of TOF gain into NEC can be found in [86].”
BackgroundMuehllehner, G., & Karp, J. S. (2006) · Physics in Medicine and Biology · doi:10.1088/0031-9155/51/13/r08 · Cited by 288
Self-citationTime-of-flight positron emission tomography: status relative to conventional PET.
Budinger, T. F. (1983) · PubMed · Cited by 283
Impact of Time-of-Flight on PET Tumor Detection
Kadrmas, D. J., Casey, M. E., Conti, M., Jakoby, B. W., Lois, C., & Townsend, D. W. (2009) · Journal of Nuclear Medicine · doi:10.2967/jnumed.109.063016 · Cited by 277 · cited 2×
“A more recent study [91] used a more realistic anthropomorphic phantom.”
BackgroundMethodAn Assessment of the Impact of Incorporating Time-of-Flight Information into Clinical PET/CT Imaging
Lois, C., Jakoby, B. W., Long, M. J., Hubner, K. F., Barker, D. W., Casey, M. E., Conti, M., Panin, V. Y., Kadrmas, D. J., & Townsend, D. W. (2010) · Journal of Nuclear Medicine · doi:10.2967/jnumed.109.068098 · Cited by 265
Simultaneous Reconstruction of Activity and Attenuation in Time-of-Flight PET
Rezaei, A., Defrise, M., Bal, G., Michel, C., Conti, M., Watson, C., & Nuyts, J. (2012) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2012.2212719 · Cited by 261
PET-MRI: a review of challenges and solutions in the development of integrated multimodality imaging
Vandenberghe, S., & Marsden, P. K. (2015) · Physics in Medicine and Biology · doi:10.1088/0031-9155/60/4/r115 · Cited by 258
Self-citationProspects for time-of-flight PET using LSO scintillator
Moses, W. W., & Derenzo, S. E. (1999) · IEEE Transactions on Nuclear Science · doi:10.1109/23.775565 · Cited by 240
Time-of-flight PET data determine the attenuation sinogram up to a constant
Defrise, M., Rezaei, A., & Nuyts, J. (2012) · Physics in Medicine and Biology · doi:10.1088/0031-9155/57/4/885 · Cited by 234
Improvement in Lesion Detection with Whole-Body Oncologic Time-of-Flight PET
El Fakhri, G., Surti, S., Trott, C. M., Scheuermann, J., & Karp, J. S. (2011) · Journal of Nuclear Medicine · doi:10.2967/jnumed.110.080382 · Cited by 198
Self-citationSimultaneous reconstruction of activity and attenuation in Time-of-Flight PET
Rezaei, A., Nuyts, J., Defrise, M., Bal, G., Michel, C., Conti, M., & Watson, C. (2011) · 2011 IEEE Nuclear Science Symposium Conference Record · doi:10.1109/nssmic.2011.6153883 · Cited by 185
Investigation of time-of-flight benefit for fully 3-DPET
Surti, S., Karp, S., Popescu, L. M., Daube-Witherspoon, E., & Werner, M. (2006) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2006.871419 · Cited by 166
Why is TOF PET reconstruction a more robust method in the presence of inconsistent data?
Conti, M. (2011) · Physics in Medicine and Biology · doi:10.1088/0031-9155/56/1/010 · Cited by 164
Extension of Single Scatter Simulation to Scatter Correction of Time-of-Flight PET
Watson, C. (2007) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2007.901227 · Cited by 164
Lewellen, T. K. (1998) · Seminars in Nuclear Medicine · doi:10.1016/s0001-2998(98)80031-7 · Cited by 160
Design considerations for a limited angle, dedicated breast, TOF PET scanner
Surti, S., & Karp, J. S. (2008) · Physics in Medicine and Biology · doi:10.1088/0031-9155/53/11/010 · Cited by 129
Self-citationOptimization of a fully 3D single scatter simulation algorithm for 3D PET
Accorsi, R., Adam, L.-E., Werner, M. E., & Karp, J. S. (2004) · Physics in Medicine and Biology · doi:10.1088/0031-9155/49/12/008 · Cited by 123
Self-citationThe imaging performance of a LaBr3-based PET scanner
Daube-Witherspoon, M. E., Surti, S., Perkins, A., Kyba, C. C. M., Wiener, R., Werner, M. E., Kulp, R., & Karp, J. S. (2010) · Physics in Medicine and Biology · doi:10.1088/0031-9155/55/1/004 · Cited by 122
Self-citationPhoton Time-of-Flight-Assisted Positron Emission Tomography
Ter-Pogossian, M. M., Mullani, N. A., Ficke, D. C., Markham, J., & Snyder, D. L. (1981) · Journal of Computer Assisted Tomography · doi:10.1097/00004728-198104000-00014 · Cited by 116
A single-ring OpenPET enabling PET imaging during radiotherapy
Tashima, H., Yamaya, T., Yoshida, E., Kinouchi, S., Watanabe, M., & Tanaka, E. (2012) · Physics in Medicine and Biology · doi:10.1088/0031-9155/57/14/4705 · Cited by 111
Surti, S., Scheuermann, J., El Fakhri, G., Daube-Witherspoon, M. E., Lim, R., Abi-Hatem, N., Moussallem, E., Benard, F., Mankoff, D., & Karp, J. S. (2011) · Journal of Nuclear Medicine · doi:10.2967/jnumed.110.086678 · Cited by 104 · cited 2×
“The same data were also presented [87] to experienced human observers, and similar findings were reported.”
BackgroundResultSelf-citationIterative image reconstruction using geometrically ordered subsets with list-mode data
Popescu, L. M., Matej, S., & Lewitt, R. M. (2005) · IEEE Symposium Conference Record Nuclear Science 2004. · doi:10.1109/nssmic.2004.1466649 · Cited by 103
Performance evaluation of the Ingenuity TF PET/CT scanner with a focus on high count-rate conditions
Kolthammer, J. A., Su, K.-H., Grover, A., Narayanan, M., Jordan, D. W., & Muzic, R. F. (2014) · Physics in Medicine and Biology · doi:10.1088/0031-9155/59/14/3843 · Cited by 101
Systematic and Distributed Time-of-Flight List Mode PET Reconstruction
Wang, W., Hu, Z., Gualtieri, E. E., Parma, M. J., Walsh, E. S., Sebok, D., Hsieh, Y.-L., Tung, C.-H., Song, X., Griesmer, J. J., et al. (2006) · 2006 IEEE Nuclear Science Symposium Conference Record · doi:10.1109/nssmic.2006.354229 · Cited by 95
Self-citationEffect of randoms on signal-to-noise ratio in TOF PET
Conti, M. (2006) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2006.875066 · Cited by 92
Direct time-of-flight for quantitative, real-time in-beam PET: a concept and feasibility study
Crespo, P., Shakirin, G., Fiedler, F., Enghardt, W., & Wagner, A. (2007) · Physics in Medicine and Biology · doi:10.1088/0031-9155/52/23/002 · Cited by 91
Impact of Time-of-Flight PET on Quantification Errors in MR Imaging–Based Attenuation Correction
Mehranian, A., & Zaidi, H. (2015) · Journal of Nuclear Medicine · doi:10.2967/jnumed.114.148817 · Cited by 90
Implementation and Evaluation of a 3D PET Single Scatter Simulation with TOF Modeling
Werner, M. E., Surti, S., & Karp, J. S. (2006) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nssmic.2006.354238 · Cited by 88
Self-citationFast reconstruction of 3D time-of-flight PET data by axial rebinning and transverse mashing
Vandenberghe, S., Daube-Witherspoon, M. E., Lewitt, R. M., & Karp, J. S. (2006) · Physics in Medicine and Biology · doi:10.1088/0031-9155/51/6/017 · Cited by 87
Self-citationMatej, S., Surti, S., Jayanthi, S., Daube-Witherspoon, M. E., Lewitt, R. M., & Karp, J. S. (2009) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2008.2012034 · Cited by 85
Self-citationYamamoto, M., Ficke, D. C., & Ter‐Pogossian, M. M. (1982) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.1982.4307571 · Cited by 79
Factors influencing timing resolution in a commercial LSO PET camera
Moses, W. W., & Ullisch, M. G. (2006) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2005.862980 · Cited by 74
MR-based attenuation correction for a whole-body sequential PET/MR system
Hu, Z., Ojha, N., Renisch, S., Schulz, V., Torres, I., Buhl, A., Pal, D., Muswick, G. J., Penatzer, J., Guo, T., et al. (2009) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nssmic.2009.5401802 · Cited by 69
Feasibility of time-of-flight reconstruction in positron emission tomography.
Mullani, N. A., Markham, J., & Ter‐Pogossian, M. M. (1980) · PubMed · Cited by 68
Characterization of the Vereos Digital Photon Counting PET System
Miller, M. A., Zhang, J., Binzel, K., Griesmer, J. J., Laurence, T., Narayanan, M. V., Natarajamani, D., Wang, S., & Knopp, M. V. (2015) · Journal of Nuclear Medicine · Cited by 68
Initial PET performance evaluation of a preclinical insert for PET/MRI with digital SiPM technology
Schug, D., Lerche, C., Weissler, B., Gebhardt, P., Goldschmidt, B., Wehner, J., Dueppenbecker, P. M., Salomon, A., Hallen, P., Kiessling, F. M., et al. (2016) · Physics in Medicine and Biology · doi:10.1088/0031-9155/61/7/2851 · Cited by 68
Studies on the Cherenkov Effect for Improved Time Resolution of TOF-PET
Brünner, S., Gruber, L., Marton, J., Suzuki, K., & Hirtl, A. (2013) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2013.2281667 · Cited by 66
Daube-Witherspoon, M. E., Surti, S., Perkins, A. E., & Karp, J. S. (2014) · Journal of Nuclear Medicine · doi:10.2967/jnumed.113.127035 · Cited by 62
Self-citationMollet, P., Keereman, V., Clementel, E., & Vandenberghe, S. (2012) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2012.2198831 · Cited by 56
Self-citationMollet, P., Keereman, V., Bini, J., Izquierdo‐Garcia, D., Fayad, Z. A., & Vandenberghe, S. (2014) · Journal of Nuclear Medicine · doi:10.2967/jnumed.113.125989 · Cited by 55
Self-citationPanin, V., Aykac, M., & Casey, M. E. (2013) · Physics in Medicine and Biology · doi:10.1088/0031-9155/58/11/3649 · Cited by 49
TOF PET offset calibration from clinical data
Werner, M. E., & Karp, J. S. (2013) · Physics in Medicine and Biology · doi:10.1088/0031-9155/58/12/4031 · Cited by 43
Self-citationDesign Optimization of a Time-Of-Flight, Breast PET Scanner
Lee, E., Werner, M. E., Karp, J. S., & Surti, S. (2013) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2013.2257849 · Cited by 42
Self-citationDavison, H., ter Voert, E. E. G. W., de Galiza Barbosa, F., Veit-Haibach, P., & Delso, G. (2015) · Investigative Radiology · doi:10.1097/rli.0000000000000146 · Cited by 41
Correction Technique for Cascade Gammas in I-124 Imaging on a Fully-3D, Time-of-Flight PET Scanner
Surti, S., Scheuermann, R. M., & Karp, J. S. (2009) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2008.2011805 · Cited by 35
Self-citationKarp, J. S., & Muehllehner, G. (1988) · Springer eBooks · doi:10.1007/978-1-4612-3866-9_4 · Cited by 32
Self-citationFEASIBILITY OF TIME-OF-FLIGHT RECONSTRUCTION IN POSITRON EMISSION TOMOGRAPHY
Mullani, N. A., Markham, J., Ter-Pogossian, M. M., & Sank, V. J. (1981) · Journal of Computer Assisted Tomography · doi:10.1097/00004728-198106000-00052 · Cited by 31
Timing and Energy Resolution of New Near-UV SiPMs Coupled to ${\rm LaBr}_3$ :Ce for TOF-PET
Schmall, J. P., Wiener, R. I., Surti, S., Ferri, A., Gola, A. G., Tarolli, A., Piemonte, C., & Karp, J. S. (2014) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2014.2346579 · Cited by 24
Self-citationSon, J., Kim, K. Y., Yoon, H. S., Won, J. Y., Ko, G. B., Lee, M. S., & Lee, J. S. (2017) · Medical Physics · doi:10.1002/mp.12440 · Cited by 24
Towards very high resolution RPC-PET for small animals
Martins, P., Blanco, A., Crespo, P., Marques, M. F. F., Ferreira‐Marques, R., Gordo, P. M., Kajetanowicz, M., Korcyl, G., LOPES, L., Michel, J., et al. (2014) · Journal of Instrumentation · doi:10.1088/1748-0221/9/10/c10012 · Cited by 22
A new modular and scalable detector for a Time-of-Flight PET scanner
Burr, K. C., Wang, G.-C. J., Du, H., Mann, G. J., Balakrishnan, K., Wang, J., Li, X., Rollet, C., Kundro, E., Buhin, M., et al. (2012) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nssmic.2012.6551645 · Cited by 19
Optimization of time-of-flight reconstruction on Philips GEMINI TF
Vandenberghe, S., van Elmbt, L., Guerchaft, M., Clementel, E., Verhaeghe, J. A. J., Bol, A., Lemahieu, I. A., & Lonneux, M. (2009) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-009-1164-3 · Cited by 16
Self-citationKrishnamoorthy, S., Werner, M. E., LeGeyt, B. C., Kaul, M., Karp, J. S., & Surti, S. (2013) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nssmic.2013.6829091 · Cited by 5
Self-citation
Cited by
State of the art in total body PET
Vandenberghe, S., Moskal, P., & Karp, J. S. (2020) · EJNMMI Physics · doi:10.1186/s40658-020-00290-2 · Cited by 390
Self-citationRoadmap toward the 10 ps time-of-flight PET challenge
Lecoq, P., Morel, C., Prior, J. O., Visvikis, D., Gundacker, S., Auffray, E., Križan, P., Turtos, R. M., Thers, D., Charbon, E., et al. (2020) · Physics in Medicine & Biology · doi:10.1088/1361-6560/ab9500 · Cited by 258
Self-citationQuantification, improvement, and harmonization of small lesion detection with state-of-the-art PET
van der Vos, C. S., Koopman, D., Rijnsdorp, S., Arends, A. J., Boellaard, R., van Dalen, J. A., Lubberink, M., Willemsen, A. T. M., & Visser, E. P. (2017) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-017-3727-z · Cited by 206
“For such a radionuclide, TOF results in a better SNR for the same number of counts [5, 87].”
BackgroundMethodInfluentialPerformance Evaluation of the Vereos PET/CT System According to the NEMA NU2-2012 Standard
Rausch, I., Ruiz, A., Valverde-Pascual, I., Cal-González, J., Beyer, T., & Carrio, I. (2019) · Journal of Nuclear Medicine · doi:10.2967/jnumed.118.215541 · Cited by 168
“With these new detector technologies, the performance of PET systems can be substantially improved (16).”
BackgroundPerformance evaluation of the next generation solid-state digital photon counting PET/CT system
Zhang, J., Maniawski, P., & Knopp, M. V. (2018) · EJNMMI Research · doi:10.1186/s13550-018-0448-7 · Cited by 123
Metadata
- Authors
- Vandenberghe, S., Mikhaylova, E., D’Hoe, E., Mollet, P., & Karp, J. S.
- Venue
- EJNMMI Physics
- Published
- 2016-02-16
- Volume
- 3
- Issue
- 1
- Pages
- 3-3
- ISSN
- 2197-7364 (ISSN-L)
- OpenAlex
- W2281211957
- Cited by
- 368
- References
- 64
- FWCI
- 31.27
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.