Establishment of national diagnostic dose reference levels (DRLs) for routine computed tomography examinations in Jordan
Open accessThe DOI matches its Crossref record; 45 of the 45 references it cites were found in a registry or an index.
Abstract
Background: Dose reference levels (DRLs) are used as indicators as well as guidance for dose optimization and to ensure justification of appropriate dose for a given clinical indication. The main aims of this study were to establish local DRLs for each CT imaging protocol as a reference point to evaluate the radiation dose indices and to compare our DRLs with those established in other countries and against the internationally reported guidelines. Materials and methods: 2000 CT dose reports of different adult imaging protocols from January 2021 until April 2022 were collected retrospectively at different hospitals in Jordan. Data were collected from CT scans that were performed using different types and models of CT scanners and included four adult non-enhanced, helical CT imaging protocols; Head, Chest, Abdomen-Pelvis, and Chest-Abdomen-Pelvis. Results: The average doses of CTDI vol , DLP, and effective dose were (65.11 mGy, 1232.71 mGy·cm, 2.83 mSv) for the head scan, (16.6 mGy, 586.6 mGy·cm, 8.21 mSv) for the chest scan, (17.91 mGy, 929.9 mGy·cm, 13.9 mSv) for the abdomen-pelvis scan, and (19.3 mGy, 1152 mGy·cm, 17.25 mSv) for the chest-abdomen-pelvis scan. In comparison with results from different international studies, DLP values measured in the present study were lower for the chest-abdomen-pelvis and abdomen-pelvis CT scans, and higher for the head CT and chest CT scans. Conclusions: It is very important that each country establishes its own DRLs and compares them with those reported by other countries, especially the developed ones. It is also important that these levels are regularly updated.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures



Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“A study conducted in 2023 by Radaideh et al assessed more than 2,000 CT scans from various hospitals in Jordan over the years 2021 and 2022.”
From Computed Tomography Diagnostic Reference Levels Worldwide and in Jordan: Implications for Patient Dose Reduction (2026) · cites this paper as [10]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Despite global progress in Diagnostic Reference Levels (DRLs), Jordan has failed to establish official national DRLs for routine CT imaging, resulting in a significant deficiency in dose optimization and quality assurance protocols.10”
From Computed Tomography Diagnostic Reference Levels Worldwide and in Jordan: Implications for Patient Dose Reduction (2026) · cites this paper as [10]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“The suggested diagnostic reference levels (DRLs) for adult head computed tomography (CT) are 65.1 mGy CTDIvol and 945 mGy·cm DLP, significantly exceeding the reference limits established in the UK.”
From Computed Tomography Diagnostic Reference Levels Worldwide and in Jordan: Implications for Patient Dose Reduction (2026) · cites this paper as [10]
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“The suggested DLP for chest-abdomen-pelvis CT was relatively lower.”
From Computed Tomography Diagnostic Reference Levels Worldwide and in Jordan: Implications for Patient Dose Reduction (2026) · cites this paper as [10]
The cited source says what the claim attributes to it.
In the citing paper
“The values were derived from the 75th percentile of the gathered data, in accordance with ICRP recommendations.10”
From Computed Tomography Diagnostic Reference Levels Worldwide and in Jordan: Implications for Patient Dose Reduction (2026) · cites this paper as [10]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Risk of cancer from diagnostic X-rays: estimates for the UK and 14 other countries
de González, A. B., & Darby, S. (2004) · The Lancet · doi:10.1016/s0140-6736(04)15433-0 · Cited by 1,718
ICRP Publication 135: Diagnostic Reference Levels in Medical Imaging
Vañó, E., Miller, D. L., Martin, C. J., Rehani, M. M., Kang, K., Rosenstein, M., Ortiz-López, P., Mattsson, S., Padovani, R., & Rogers, A. (2017) · Annals of the ICRP · doi:10.1177/0146645317717209 · Cited by 1,043
AAPM/RSNA Physics Tutorial for Residents: Topics in CT: Radiation Dose in CT
McNitt-Gray, M. F. (2002) · RadioGraphics · doi:10.1148/rg.226025128 · Cited by 612 · cited 1×
“48 In comparison with results from different international studies, DLP values measured in the present study were lower for the chest abdomen-pelvis and abdomen-pelvis CT scans and higher for the head CT and chest CT scans.”
The Measurement, Reporting, and Management of Radiation Dose in CT
McCollough, C., Cody, D., Edyvean, S., Geise, R., Gould, B., Keat, N., Huda, W., Judy, P., Kalender, W., McNitt-Gray, M., et al. (2008) · doi:10.37206/97 · Cited by 569
U.S. Diagnostic Reference Levels and Achievable Doses for 10 Adult CT Examinations
Kanal, K. M., Butler, P. F., Sengupta, D., Bhargavan-Chatfield, M., Coombs, L. P., & Morin, R. L. (2017) · Radiology · doi:10.1148/radiol.2017161911 · Cited by 395
Rehani, M. M., Ciraj-Bjelac, O., Vañó, E., Miller, D. L., Walsh, S., Giordano, B. D., & Persliden, J. (2010) · Annals of the ICRP · doi:10.1016/j.icrp.2012.03.001 · Cited by 344
McCollough, C. H., Chen, G. H., Kalender, W., Leng, S., Samei, E., Taguchi, K., Wang, G., Yu, L., & Pettigrew, R. I. (2012) · Radiology · doi:10.1148/radiol.12112265 · Cited by 276
Establishment of CT diagnostic reference levels in Ireland
Foley, S. J., McEntee, M. F., & Rainford, L. A. (2012) · The British Journal of Radiology · doi:10.1259/bjr/15839549 · Cited by 186 · cited 3×
“…quartile for CTDIvol and the effective dose values ( Table 6 ) for the chest-abdomen-pelvis imaging protocol (19.3 mGy and 17.25 mSv, respectively) was lower than the 3 rd quartile measured in some previous studies 20,26 and higher than the 3 rd quartile measured in other studies 24,31,34,36,37 .”
InfluentialSchauer, D. A., & Linton, O. W. (2009) · Radiology · doi:10.1148/radiol.2532090494 · Cited by 171
Valentin, J. (2001) · Annals of the ICRP · doi:10.1016/s0146-6453(02)00007-6 · Cited by 149
Treier, R., Aroua, A., Verdun, F. R., Samara, E., Stuessi, A., & Trueb, P. R. (2010) · Radiation Protection Dosimetry · doi:10.1093/rpd/ncq279 · Cited by 143 · cited 4×
“…for CTDI vol and the effective dose values ( Table 3 ) for the head imaging protocol (65.1 mGy, 2.83 mSv, respectively) were lower than the 3 rd quartile measured in some previous studies 19,34,35,37,39 , and higher than the 3 rd quartile measured in other studies 10,20,26,28,32,36,38,40,41,50,52 .”
InfluentialRadiation Doses in Consecutive CT Examinations from Five University of California Medical Centers
Smith-Bindman, R., Moghadassi, M., Wilson, N., Nelson, T. R., Boone, J. M., Cagnon, C. H., Gould, R., Hall, D. J., Krishnam, M., Lamba, R., et al. (2015) · Radiology · doi:10.1148/radiol.2015142728 · Cited by 127
Vassileva, J., & Rehani, M. (2015) · American Journal of Roentgenology · doi:10.2214/ajr.14.12794 · Cited by 125
U.S. Diagnostic Reference Levels and Achievable Doses for 10 Pediatric CT Examinations
Kanal, K. M., Butler, P. F., Chatfield, M. B., Wells, J., Samei, E., Simanowith, M., Golden, D., Gress, D. A., Burleson, J., Sensakovic, W. F., et al. (2022) · Radiology · doi:10.1148/radiol.2021211241 · Cited by 101
Establishing national diagnostic reference levels (DRLs) for computed tomography in Egypt
Salama, D. H., Vassileva, J., Mahdaly, G., Shawki, M., Salama, A., Gilley, D., & Rehani, M. M. (2017) · Physica Medica · doi:10.1016/j.ejmp.2017.05.050 · Cited by 98 · cited 6×
“20 The data collected included information related to the scanner, routine CT protocols, patient weight, examination data, and dose indices (CTDI and DLP).”
InfluentialRoch, P., & Aubert, B. (2013) · Radiation Protection Dosimetry · doi:10.1093/rpd/ncs152 · Cited by 89 · cited 4×
“…quartile for CTDIvol and the effective dose values ( Table 6 ) for the chest-abdomen-pelvis imaging protocol (19.3 mGy and 17.25 mSv, respectively) was lower than the 3 rd quartile measured in some previous studies 20,26 and higher than the 3 rd quartile measured in other studies 24,31,34,36,37 .”
InfluentialAdult patient radiation doses from non-cardiac CT examinations: a review of published results
Pantos, I., Thalassinou, S., Argentos, S., Kelekis, N. L., Panayiotakis, G., & Efstathopoulos, E. P. (2011) · The British Journal of Radiology · doi:10.1259/bjr/69070614 · Cited by 85 · cited 4×
“Furthermore, Greece recorded lower updated dose levels where the recent CTDI vol and effective dose levels were lower (16 mGy, 11.4 mSv, respectively) 36 than the old ones (35 mGy, 11.7 mSv, respectively) 23 .”
InfluentialUpdated Australian diagnostic reference levels for adult CT
Lee, K. L., Beveridge, T., Sanagou, M., & Thomas, P. (2020) · Journal of Medical Radiation Sciences · doi:10.1002/jmrs.372 · Cited by 75 · cited 3×
“…for CTDI vol and the effective dose values ( Table 3 ) for the head imaging protocol (65.1 mGy, 2.83 mSv, respectively) were lower than the 3 rd quartile measured in some previous studies 19,34,35,37,39 , and higher than the 3 rd quartile measured in other studies 10,20,26,28,32,36,38,40,41,50,52 .”
InfluentialRadiation doses from head, neck, chest and abdominal CT examinations: an institutional dose report
Atlı, E., Uyanık, S. A., Öğüşlü, U., Cenkeri, H. C., Yılmaz, B., & Gümüş, B. (2021) · Diagnostic and Interventional Radiology · doi:10.5152/dir.2020.19560 · Cited by 72 · cited 4×
“…for CTDI vol and the effective dose values ( Table 3 ) for the head imaging protocol (65.1 mGy, 2.83 mSv, respectively) were lower than the 3 rd quartile measured in some previous studies 19,34,35,37,39 , and higher than the 3 rd quartile measured in other studies 10,20,26,28,32,36,38,40,41,50,52 .”
InfluentialDose limits to the lens of the eye: International Basic Safety Standards and related guidance
Boal, T., & Pinák, M. (2015) · Annals of the ICRP · doi:10.1177/0146645314562321 · Cited by 69
Updated effective doses in radiology
Palop, J. V., Vilar, J., Hernández‐Aguado, I., González-Álvarez, I., & Lumbreras, B. (2016) · Journal of Radiological Protection · doi:10.1088/0952-4746/36/4/975 · Cited by 65 · cited 1×
“Some other studies, on the other hand, are now outdated and compared their results with old studies 41,43 , suffer from missing some important information 43-45 , have a limited number of patients and sites 46,47 , or are almost limited to a specific imaging protocol 21 .”
Recommendations of the International Commission on Radiological Protection: Revised December 1954
(1959) · Annals of the ICRP/ICRP Publication · doi:10.1016/s0074-2740(28)80014-6 · Cited by 65
The establishment of computed tomography diagnostic reference levels in Portugal
Santos, J., Foley, S., Paulo, G., McEntee, M. F., & Rainford, L. (2014) · Radiation Protection Dosimetry · doi:10.1093/rpd/nct226 · Cited by 64 · cited 1×
“Many countries in the Middle East or Western countries had estimated their dose reports, which were considered regional references 4,5,10,20,21,23-25 .”
Brady, S. L., Mirro, A. E., Moore, B. M., & Kaufman, R. A. (2015) · American Journal of Roentgenology · doi:10.2214/ajr.14.13317 · Cited by 53 · cited 2×
“…for CTDI vol and the effective dose values ( Table 4 ) for the chest imaging protocol (16.6 mGy, 8.21 mSv, respectively) were lower than the 3 rd quartile measured in some previous studies 20,23,35,50 , and higher than the 3 rd quartile measured in other studies 10,18,26,28-29,32,34,37-40,52 .”
InfluentialMuhammad, N. A., Abdul Karim, M. K., Abu Hassan, H., Ahmad Kamarudin, M., Ding Wong, J. H., & Ng, K. H. (2020) · Diagnostics · doi:10.3390/diagnostics10080591 · Cited by 53 · cited 2×
“…for CTDI vol and the effective dose values ( Table 3 ) for the head imaging protocol (65.1 mGy, 2.83 mSv, respectively) were lower than the 3 rd quartile measured in some previous studies 19,34,35,37,39 , and higher than the 3 rd quartile measured in other studies 10,20,26,28,32,36,38,40,41,50,52 .”
InfluentialAustralian diagnostic reference levels for multi detector computed tomography
Hayton, A., Wallace, A. B., Marks, P., Edmonds, K., Tingey, D., & Johnston, P. N. (2013) · Australasian Physical & Engineering Sciences in Medicine · doi:10.1007/s13246-013-0180-6 · Cited by 48 · cited 2×
“…quartile for CTDIvol and the effective dose values ( Table 6 ) for the chest-abdomen-pelvis imaging protocol (19.3 mGy and 17.25 mSv, respectively) was lower than the 3 rd quartile measured in some previous studies 20,26 and higher than the 3 rd quartile measured in other studies 24,31,34,36,37 .”
InfluentialPatient doses from CT examinations in Turkey
Atac, G. K., Parmaksiz, A., Inal, T., Bulur, E., Bulgurlu, F., Oncu, T., & Gundogdu, S. (2015) · Diagnostic and Interventional Radiology · doi:10.5152/dir.2015.14306 · Cited by 44
EXPOSURE OF THE SWISS POPULATION BY RADIODIAGNOSTICS: 2013 REVIEW
Coultre, R. L., Bize, J., Champendal, M., Wittwer, D., Ryckx, N., Aroua, A., Trueb, P. R., & Verdun, F. R. (2015) · Radiation Protection Dosimetry · doi:10.1093/rpd/ncv462 · Cited by 38
Diagnostic reference levels and patient doses in computed tomography examinations in Greece
Simantirakis, G., Hourdakis, C. J., Economides, S., Kaissas, I., Kalathaki, M., Koukorava, C., Manousaridis, G., Pafilis, C., Tritakis, P., Vogiatzi, S., et al. (2014) · Radiation Protection Dosimetry · doi:10.1093/rpd/ncu182 · Cited by 37 · cited 4×
“…quartile for CTDIvol and the effective dose values ( Table 6 ) for the chest-abdomen-pelvis imaging protocol (19.3 mGy and 17.25 mSv, respectively) was lower than the 3 rd quartile measured in some previous studies 20,26 and higher than the 3 rd quartile measured in other studies 24,31,34,36,37 .”
InfluentialAbuzaid, M., Elshami, W., Tekin, H. O., Ghonim, H., Shawki, M., & Salama, D. H. (2020) · Insights into Imaging · doi:10.1186/s13244-020-00891-6 · Cited by 37 · cited 3×
“…for CTDI vol and the effective dose values ( Table 3 ) for the head imaging protocol (65.1 mGy, 2.83 mSv, respectively) were lower than the 3 rd quartile measured in some previous studies 19,34,35,37,39 , and higher than the 3 rd quartile measured in other studies 10,20,26,28,32,36,38,40,41,50,52 .”
InfluentialRadiation doses from computed tomography practice in Johor Bahru, Malaysia
Karim, M. K. A., Hashim, S., Bradley, D. A., Bakar, K. A., Haron, M. R., & Kayun, Z. (2015) · Radiation Physics and Chemistry · doi:10.1016/j.radphyschem.2015.12.020 · Cited by 33 · cited 1×
“…vol and the effective dose values ( Table 5 ) for the abdomen-pelvis imaging protocol (17.9 mGy, 13.9 mSv, respectively) were lower than the 3 rd quartile measured in some previous studies 20,23,34,35,41 , and higher than the 3rd quartile measured in other studies 10,18,26,28-29,31,32,36-40,50,51 .”
InfluentialLiang, C. R., Chen, P. X. H., Kapur, J., Ong, M. K. L., Quek, S. T., & Kapur, S. C. (2017) · Journal of Medical Radiation Sciences · doi:10.1002/jmrs.210 · Cited by 33 · cited 1×
“Many other studies focused on how to reduce the CT radiation doses during diagnostic imaging without affecting the quality of the diagnostic images 42 but with limited information from other neighboring countries.”
Radiation exposure during paediatric CT in Sudan: CT dose, organ and effective doses
Suliman, I. I., Khamis, H. M., Ombada, T. H., Alzimami, K. S., Alkhorayef, M. A., & Sulieman, A. (2014) · Radiation Protection Dosimetry · doi:10.1093/rpd/ncu321 · Cited by 31 · cited 2×
“Some other studies, on the other hand, are now outdated and compared their results with old studies 41,43 , suffer from missing some important information 43-45 , have a limited number of patients and sites 46,47 , or are almost limited to a specific imaging protocol 21 .”
Exposure of the Swiss population to computed tomography
Aroua, A., Samara, E. T., Bochud, F. O., Meuli, R. A., & Verdun, F. R. (2013) · BMC Medical Imaging · doi:10.1186/1471-2342-13-22 · Cited by 30 · cited 1×
“…vol and the effective dose values ( Table 5 ) for the abdomen-pelvis imaging protocol (17.9 mGy, 13.9 mSv, respectively) were lower than the 3 rd quartile measured in some previous studies 20,23,34,35,41 , and higher than the 3rd quartile measured in other studies 10,18,26,28-29,31,32,36-40,50,51 .”
InfluentialHéliou, R., Normandeau, L., & Beaudoin, G. (2011) · Radiation Protection Dosimetry · doi:10.1093/rpd/ncr024 · Cited by 30
CT dose survey in adults: what sample size for what precision?
Taylor, S. E., Muylem, A. V., Howarth, N., Gevenois, P. A., & Tack, D. (2016) · European Radiology · doi:10.1007/s00330-016-4333-3 · Cited by 30 · cited 2×
“49 However, we have considered the results of the larger sample size.”
Is routine pelvic surveillance imaging necessary in patients with Wilms tumor?
Kaste, S. C., Brady, S. L., Yee, B., McPherson, V. J., Kaufman, R. A., Billups, C. A., Daw, N. C., & Pappo, A. S. (2012) · Cancer · doi:10.1002/cncr.27687 · Cited by 22
Survey of patient dose in computed tomography in Syria 2009
Kharita, M. H., & Khazzam, S. (2010) · Radiation Protection Dosimetry · doi:10.1093/rpd/ncq155 · Cited by 20 · cited 3×
“Some other studies, on the other hand, are now outdated and compared their results with old studies 41,43 , suffer from missing some important information 43-45 , have a limited number of patients and sites 46,47 , or are almost limited to a specific imaging protocol 21 .”
InfluentialHasan, N., Rizk, C., & Babikir, E. (2022) · Radiation Physics and Chemistry · doi:10.1016/j.radphyschem.2022.110147 · Cited by 17 · cited 1×
“Some other studies, on the other hand, are now outdated and compared their results with old studies 41,43 , suffer from missing some important information 43-45 , have a limited number of patients and sites 46,47 , or are almost limited to a specific imaging protocol 21 .”
Wardlaw, G. M., & Martel, N. (2016) · Medical Physics · doi:10.1118/1.4961767 · Cited by 16 · cited 4×
“…quartile for CTDIvol and the effective dose values ( Table 6 ) for the chest-abdomen-pelvis imaging protocol (19.3 mGy and 17.25 mSv, respectively) was lower than the 3 rd quartile measured in some previous studies 20,26 and higher than the 3 rd quartile measured in other studies 24,31,34,36,37 .”
InfluentialKumamaru, K. K., Kogure, Y., Suzuki, M., Hori, M., Nakanishi, A., Kamagata, K., Hagiwara, A., Andica, C., Ri, K., Houshido, N., et al. (2016) · Japanese Journal of Radiology · doi:10.1007/s11604-016-0545-3 · Cited by 13 · cited 3×
“…for CTDI vol and the effective dose values ( Table 3 ) for the head imaging protocol (65.1 mGy, 2.83 mSv, respectively) were lower than the 3 rd quartile measured in some previous studies 19,34,35,37,39 , and higher than the 3 rd quartile measured in other studies 10,20,26,28,32,36,38,40,41,50,52 .”
InfluentialKhoramian, D., Sistani, S., & Hejazi, P. (2019) · Polish Journal of Medical Physics And Engineering · doi:10.2478/pjmpe-2019-0008 · Cited by 13 · cited 1×
“…for CTDI vol and the effective dose values ( Table 3 ) for the head imaging protocol (65.1 mGy, 2.83 mSv, respectively) were lower than the 3 rd quartile measured in some previous studies 19,34,35,37,39 , and higher than the 3 rd quartile measured in other studies 10,20,26,28,32,36,38,40,41,50,52 .”
InfluentialRazali, M. A. S. M., Ahmad, M. Z., Roslee, M. A. A. M., & Osman, N. D. (2019) · Journal of Physics Conference Series · doi:10.1088/1742-6596/1248/1/012067 · Cited by 9 · cited 1×
“…quartile for CTDIvol and the effective dose values ( Table 6 ) for the chest-abdomen-pelvis imaging protocol (19.3 mGy and 17.25 mSv, respectively) was lower than the 3 rd quartile measured in some previous studies 20,26 and higher than the 3 rd quartile measured in other studies 24,31,34,36,37 .”
InfluentialSaeed, M. K., Alzoubi, A. S., & Alqahtani, J. M. (2014) · Australasian Physical & Engineering Sciences in Medicine · doi:10.1007/s13246-014-0256-y · Cited by 8 · cited 1×
“Some other studies, on the other hand, are now outdated and compared their results with old studies 41,43 , suffer from missing some important information 43-45 , have a limited number of patients and sites 46,47 , or are almost limited to a specific imaging protocol 21 .”
Larsson, C. (2009) · Journal of Environmental Radioactivity · doi:10.1016/j.jenvrad.2009.07.003 · Cited by 7
Research works printed but not linked: 4
- 2. Rothenberg LN, Pentlow KS. CT dosimetry and radiation safety. In: Goldman LW, Fowlkes JB, eds. Medical CT and Ultrasound: Current Technology and Applications. Madison, Wis: Advanced Medical Publishing; 1995:519-553
- 4. IAEA Safety Standards Series No. GSR Part 3. Radiation protection and safety of radiation sources: International basic safety standards. General safety requirements Part 3 (Spanish Edition). 2016.
- 10. Shrimpton PC, Hillier MC, Lewis MA, Dunn M. NRPB-W67 doses from computed tomography (CT) examinations in the UK-2003 review. National Radiation Protection Board.2005.
- 45. Vilar-Palop J, Vilar J, Hernández-Aguado I, et al. Updated effective doses in radiology. Journal of Radiological Protection. 2016;36(4):975-990. https://doi.org/10.1088/0952-4746/36/4/975 · DOI
Cited by
A closer look at utilized radiation doses during chest CT for COVID-19 patients
Khallouqi, A., Sekkat, H., Halimi, A., & Rhazouani, O. E. (2024) · Radiation Physics and Chemistry · doi:10.1016/j.radphyschem.2024.112079 · Cited by 19
An Analysis of Computed Tomography Diagnostic Reference Levels in India Compared to Other Countries
Malik, M. M. U. D., Alqahtani, M. M., Hadadi, I., AlQhtani, A. G. M., & Alqarni, A. (2024) · Diagnostics · doi:10.3390/diagnostics14151585 · Cited by 17
Diagnostic Reference Levels of Radiographic and CT Examinations in Jordan: A Systematic Review
Alzyoud, K., Al-Murshedi, S., & England, A. (2024) · Health Physics · doi:10.1097/hp.0000000000001778 · Cited by 6
Botwe, B. O., Anim-Sampong, S., & Saizi, R. (2025) · International Journal of Biomedical Imaging · doi:10.1155/ijbi/6464273 · Cited by 4
Hamd, Z. Y., Almohammed, H. I., Manssor, E., Bakry, A. H. A., & Gareeballah, A. (2025) · Tomography · doi:10.3390/tomography11060065 · Cited by 3
Metadata
- Authors
- Radaideh, K., Al-Radaideh, A., Ramli, R. M., Saleh, A., & Alshayeb, R.
- Affiliations
- Isra University (JO)
- Department of Medical Imaging , The Hashemite University , Jordan
- Universiti Sains Malaysia (MY)
- Corresponding author
- Radaideh, Khaldoon
- Venue
- Polish Journal of Medical Physics and Engineering
- Publisher
- Walter de Gruyter GmbH
- Published
- 2023-02-01
- Volume
- 29
- Issue
- 1
- Pages
- 26-34
- ISSN
- 1425-4689 (ISSN-L)
- 1898-0309
- OpenAlex
- W4318766078
- Cited by
- 16
- References
- 44
- Citations by year
- 2026: 3 · 2025: 6 · 2024: 6 · 2023: 1
- FWCI
- 2.66
- Type
- journal article
- Language
- en
- Open access
- diamond
- doi.org
- Licence
- Locations
- Polish Journal of Medical Physics And Engineering (publishedVersion, cc-by-nc-nd)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiation Dose and Imaging
- Physical Sciences › Physics and Astronomy › Radiation › Advanced Radiotherapy Techniques
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Digital Radiography and Breast Imaging
- Keywords
- dose reference levels, computed tomography, radiation dose, CTDIvol, dose-length product, effective dose, head CT, chest CT, abdomen-pelvis CT, dose optimization, Jordan
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.