A comparison of microelectrodes for a visual cortical prosthesis using finite element analysis
Open accessThe DOI matches its Crossref record; 17 of the 17 references it cites were found in a registry or an index.
Abstract
Altering the geometry of microelectrodes for use in a cortical neural prosthesis modifies the electric field generated in tissue, thereby affecting electrode efficacy and tissue damage. Commonly, electrodes with an active region located at the tip ("conical" electrodes) are used for stimulation of cortex but there is argument to believe this geometry may not be the best. Here we use finite element analysis to compare the electric fields generated by three types of electrodes, a conical electrode with exposed active tip, an annular electrode with active area located up away from the tip, and a striped annular electrode where the active annular region has bands of insulation interrupting the full active region. The results indicate that the current density on the surface of the conical electrodes can be up to 10 times greater than the current density on the annular electrodes of the same height, which may increase the propensity for tissue damage. However choosing the most efficient electrode geometry in order to reduce power consumption is dependent on the distance of the electrode to the target neurons. If neurons are located within 10 μm of the electrode, then a small conical electrode would be more power efficient. On the other hand if the target neuron is greater than 500 μm away-as happens normally when insertion of an array of electrodes into cortex results in a "kill zone" around each electrode due to insertion damage and inflammatory responses-then a large annular electrode would be more efficient.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Neural Stimulation and Recording Electrodes
Cogan, S. F. (2008) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.10.061807.160518 · Cited by 2,243
Electrical stimulation of excitable tissue: design of efficacious and safe protocols
Merrill, D. R., Bikson, M., & Jefferys, J. G. R. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2004.10.020 · Cited by 2,078 · cited 2×
“…may contribute to tissue damage Frontiers in Neuroengineering www.frontiersin.org September 2012 | Volume 5 | Article 23 | 1 NEUROENGINEERING (McCreery et al., 1990; Shannon, 1992), corrosion of the electrode (Merrill et al., 2005) as may affect the electrode’s efficacy (Butson and McIntyre, 2006).”
BackgroundResponse of brain tissue to chronically implanted neural electrodes
Polikov, V. S., Tresco, P. A., & Reichert, W. M. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2005.08.015 · Cited by 1,896 · cited 4×
“…greatly reduces the density of neurons within the near field (McCreery et al., 2010) and that neurons will be further separated from the electrode upon formation of a glial scar as part of the inflammatory response to a foreign body in the cortex (Xindong et al., 1999; Polikov et al., 2005).”
BackgroundInfluentialThe sensations produced by electrical stimulation of the visual cortex
Brindley, G. S., & Lewin, W. S. (1968) · The Journal of Physiology · doi:10.1113/jphysiol.1968.sp008519 · Cited by 1,170 · cited 6×
“The second approach is to stimulate the visual cortex located at the back of the brain (Brindley and Lewin, 1968; Dobelle and Mladejovsky, 1974; Troyk et al., 2005).”
BackgroundInfluentialCampbell, P. K., Jones, K. E., Huber, R. J., Horch, K. W., & Normann, R. A. (1991) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.83588 · Cited by 881 · cited 2×
“ORIGINAL RESEARCH ARTICLE published: 28 September 2012 doi: 10.3389/fneng.2012.00023”
BackgroundMethodInfluentialCharge density and charge per phase as cofactors in neural injury induced by electrical stimulation
McCreery, D. B., Agnew, W. F., Yuen, T. G. H., & Bullara, L. (1990) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.102812 · Cited by 708 · cited 4×
“This may contribute to tissue damage Frontiers in Neuroengineering www.frontiersin.org September 2012 | Volume 5 | Article 23 | 1 NEUROENGINEERING (McCreery et al., 1990; Shannon, 1992), corrosion of the electrode (Merrill et al., 2005) as may affect the electrode’s efficacy (Butson and McIntyre,…”
BackgroundInfluentialInterim results from the international trial of Second Sight's visual prosthesis
Humayun, M. S., Dorn, J. D., da Cruz, L., Dagnelie, G., Sahel, J.-A., Stanga, P. E., Cideciyan, A. V., Duncan, J. L., Eliott, D., Filley, E., et al. (2012) · Ophthalmology · doi:10.1016/j.ophtha.2011.09.028 · Cited by 700 · cited 2×
“In recent developments, stimulating the retina has elicited promising results with groups such as Second Sight successfully developing a prosthesis which is now undergoing clinical trials in the United States and Europe (Humayun et al., 2012).”
BackgroundAnalysis of Models for External Stimulation of Axons
Rattay, F. (1986) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1986.325670 · Cited by 681
Schmidt, E. M., Bak, M. J., Hambrecht, F. T., Kufta, C. V., O'Rourke, D. K., & Vallabhanath, P. (1996) · Brain · doi:10.1093/brain/119.2.507 · Cited by 672 · cited 8×
“…via electrical stimulation of the visual cortex has been demonstrated in many studies (Brindley and Lewin, 1968; Dobelle and Mladejovsky, 1974; Schmidt et al., 1996) and it is now well established that stimulating the surface of the visual cortex can induce spots of light termed phosphenes…”
BackgroundMethodInfluentialA model of safe levels for electrical stimulation
Shannon, R. V. (1992) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.126616 · Cited by 526 · cited 3×
“…may contribute to tissue damage Frontiers in Neuroengineering www.frontiersin.org September 2012 | Volume 5 | Article 23 | 1 NEUROENGINEERING (McCreery et al., 1990; Shannon, 1992), corrosion of the electrode (Merrill et al., 2005) as may affect the electrode’s efficacy (Butson and McIntyre, 2006).”
BackgroundMethodElectrical stimulation of neural tissue to evoke behavioral responses
Tehovnik, E. J. (1996) · Journal of Neuroscience Methods · doi:10.1016/0165-0270(95)00131-x · Cited by 478
Dobelle, W. H., & Mladejovsky, M. G. (1974) · The Journal of Physiology · doi:10.1113/jphysiol.1974.sp010766 · Cited by 468
Visual sensations produced by intracortical microstimulation of the human occipital cortex
Bak, M., Girvin, J. P., Hambrecht, F. T., Kufta, C. V., Loeb, G. E., & Schmidt, E. M. (1990) · Medical & Biological Engineering & Computing · doi:10.1007/bf02442682 · Cited by 299 · cited 2×
“The most common penetrating electrode design used for stimulation of the visual cortex has the active area located in the tip (Bak et al., 1990; Campbell et al., 1991; Schmidt et al., 1996; Tehovnik, 1996; Cogan et al., 2004).”
BackgroundMethodOver-pulsing degrades activated iridium oxide films used for intracortical neural stimulation
Cogan, S. F., Guzelian, A. A., Agnew, W. F., Yuen, T. G. H., & McCreery, D. B. (2004) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2004.02.019 · Cited by 210
Maynard, E. M. (2001) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.3.1.145 · Cited by 126
Agnew, W. F., Yuen, T. G. H., & McCreery, D. B. (1983) · Experimental Neurology · doi:10.1016/0014-4886(83)90221-2 · Cited by 96 · cited 2×
“Many studies have shown that increased charge density at the tip of the electrode is associated with increased histological damage (Pollen, 1977; Yuen et al., 1981; Agnew et al., 1983; McCreery et al., 1990).”
BackgroundCurrent Density Profiles of Surface Mounted and Recessed Electrodes for Neural Prostheses
Rubinstein, J. T., Spelman, F. A., Soma, M., & Suesserman, M. F. (1987) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1987.326007 · Cited by 89
Cited by
Lewis, P. M., Ackland, H. M., Lowery, A. J., & Rosenfeld, J. V. (2015) · Brain Research · doi:10.1016/j.brainres.2014.11.020 · Cited by 228
Self-citationBrunton, E. K., Winther-Jensen, B., Wang, C., Yan, E. B., Hagh Gooie, S., Lowery, A. J., & Rajan, R. (2015) · Frontiers in Neuroscience · doi:10.3389/fnins.2015.00265 · Cited by 46
Self-citationIntroducing the Monash vision group's cortical prosthesis
Lowery, A. J. (2013) · 2013 IEEE International Conference on Image Processing · doi:10.1109/icip.2013.6738316 · Cited by 45
Self-citationNeurophysiological considerations for visual implants
Meikle, S. J., & Wong, Y. T. (2022) · Brain Structure and Function · doi:10.1007/s00429-021-02417-2 · Cited by 15
“It should also be noted that the injected charge is dependent on the current density at the surface of the electrode, meaning different electrode shapes and sizes will dictate the amount of depolarized tissue or spread of cortical activity when the same amount of current is used during stimulation (Bosking et al. 2017…”
Background
Metadata
- Authors
- Brunton, E., Lowery, A. J., & Rajan, R.
- Venue
- Frontiers in Neuroengineering
- Publisher
- Frontiers Media SA
- Published
- 2012
- Volume
- 5
- Pages
- 23-23
- ISSN
- 1662-6443 (ISSN-L)
- OpenAlex
- W2148324796
- Cited by
- 19
- References
- 17
- FWCI
- 1.32
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.