Recent advances in cortisol sensing technologies for point-of-care application
Abstract
Everyday lifestyle related issues are the main cause of psychological stress, which contributes to health disparities experienced by individuals. Prolonged exposure to stress leads to the activation of signaling pathways from the brain that leads to release of cortisol from the adrenal cortex. Various biomarkers have been affected by psychological stress, but cortisol "a steroid hormone" is known as a potential biomarker for its estimation. Cortisol can also be used as a target analyte marker to determine the effect of exposure such as organophosphates on central nervous system, which alters the endocrine system, leading to imbalance in cortisol secretion. Cortisol secretion of individuals depends on day-night cycle and field environment hence its detection at point-of-care (POC) is deemed essential to provide personalized healthcare. Chromatographic techniques have been traditionally used to detect cortisol. The issues relating to assay formation, system complexity, and multistep extraction/purification limits its application in the field. In order to overcome these issues and to make portable and effective miniaturized platform, various immunoassays sensing strategies are being explored. However, electrochemical immunosensing of cortisol is considered as a recent advancement towards POC application. Highly sensitive, label-free and selective cortisol immunosensor based on microelectrodes are being integrated with the microfluidic system for automated diurnal cortisol monitoring useful for personalized healthcare. Although the reported sensing devices for cortisol detection may have a great scope to improve portability, electronic designing, performance of the integrated sensor, data safety and lifetime for point-of-care applications, This review is an attempt to describe the various cortisol sensing platforms and their potential to be integrated into a wearable system for online and continuous monitoring of cortisol rhythm at POC as a function of one's environment.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“In addition, since sweat is not sampled but sensed directly on the skin, the skin topography, skin temperature[ 10 ] and pH[ 11 ] may interfere considerably with the measurement results.”
From MicroSweat: A wearable microfluidic patch for noninvasive and reliable sweat collection enables human stress monitoring (2023) · cites this paper as [11]
In the citing paper
“The vents located on top of the storage chambers are <300 µm in diameter to prevent sweat evaporation (based on the Langmuir's free evaporation equation in vacuum[ 11 , 18 ] (see Section S4, Supporting Information) and help preserve the molecular contents of the sweat in the stored fibers.”
From MicroSweat: A wearable microfluidic patch for noninvasive and reliable sweat collection enables human stress monitoring (2023) · cites this paper as [11]
Citations
Cites
Stress and the brain: from adaptation to disease
de Kloet, E. R., Joëls, M., & Holsboer, F. (2005) · Nature Reviews Neuroscience · doi:10.1038/nrn1683 · Cited by 4,621
Biosensors: sense and sensibility
Turner, A. P. F. (2013) · Chemical Society Reviews · doi:10.1039/c3cs35528d · Cited by 1,756
Microneedles for transdermal drug delivery
Prausnitz, M. R. (2003) · Advanced Drug Delivery Reviews · doi:10.1016/j.addr.2003.10.023 · Cited by 1,518
Electrochemical biosensors: Towards point-of-care cancer diagnostics
Wang, J. (2005) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2005.10.027 · Cited by 1,458
Point-of-Care Diagnostics for Global Health
Yager, P., Domingo, G. J., & Gerdes, J. (2008) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.10.061807.160524 · Cited by 1,076
Russell, E., Koren, G., Rieder, M., & Van Uum, S. (2012) · Psychoneuroendocrinology · doi:10.1016/j.psyneuen.2011.09.009 · Cited by 1,063
Shankaran, D. R., Gobi, K. V., & Miura, N. (2006) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2006.09.014 · Cited by 603
Kirschbaum, C., Tietze, A., Skoluda, N., & Dettenborn, L. (2009) · Psychoneuroendocrinology · doi:10.1016/j.psyneuen.2008.08.024 · Cited by 584
Measurement of cortisol in human Hair as a biomarker of systemic exposure
Koren, G., Grace Walsh, R. N., Tokmakejian, S., & Uum, S. H. M. V. (2007) · Clinical and investigative medicine · doi:10.25011/cim.v30i5.2894 · Cited by 578
What is stress, and how does it affect reproduction?
Dobson, H., & Smith, R. F. (2000) · Animal Reproduction Science · doi:10.1016/s0378-4320(00)00080-4 · Cited by 549
Flagged first, then by citations
1 of the 115 references couldn’t be matched to a known paper.
Research works printed but not linked: 1
- Mitchell, 2009, Rapid ultrasensitive measurement of salivary cortisol using nano-linker chemistry coupled with surface plasmon resonance detection, Analyst (Lond.), 134, 380, 10.1039/B817083P · DOI
Cited by
Sorted by citations
Parlak, O., Keene, S. T., Marais, A., Curto, V. F., & Salleo, A. (2018) · Science Advances · doi:10.1126/sciadv.aar2904 · Cited by 603
Wearable sensors for monitoring the physiological and biochemical profile of the athlete
Seshadri, D. R., Li, R. T., Voos, J. E., Rowbottom, J. R., Alfes, C. M., Zorman, C. A., & Drummond, C. K. (2019) · npj Digital Medicine · doi:10.1038/s41746-019-0150-9 · Cited by 537
Ramesh, M., Janani, R., Deepa, C., & Rajeshkumar, L. (2022) · Biosensors · doi:10.3390/bios13010040 · Cited by 486
Jain, S., Nehra, M., Kumar, R., Dilbaghi, N., Hu, Y., Kumar, S., Kaushik, A. K., & Li, C.-Z. (2021) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2021.113074 · Cited by 379
Self-citationA Wearable Cellulose Acetate-Coated Mouthguard Biosensor for In Vivo Salivary Glucose Measurement
Arakawa, T., Tomoto, K., Nitta, H., Toma, K., Takeuchi, S., Sekita, T., Minakuchi, S., & Mitsubayashi, K. (2020) · Analytical Chemistry · doi:10.1021/acs.analchem.0c01201 · Cited by 328
Smart, soft contact lens for wireless immunosensing of cortisol
Ku, M., Kim, J., Won, J., Kang, W., Park, Y., Park, J., Lee, J., Cheon, J., Lee, H. H., & Park, J. (2020) · Science Advances · doi:10.1126/sciadv.abb2891 · Cited by 278
Aptamer-based lateral flow assay for point of care cortisol detection in sweat
Dalirirad, S., & Steckl, A. J. (2018) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2018.11.161 · Cited by 266
Zhang, Y., Guo, H., Kim, S. B., Wu, Y., Ostojich, D., Park, S. H., Wang, X. S., Weng, Z., Li, R., Bandodkar, A. J., et al. (2019) · Lab on a Chip · doi:10.1039/c9lc00103d · Cited by 252
Recent advances and perspectives in sweat based wearable electrochemical sensors
Mohan, A. M. V., Rajendran, V., Mishra, R. K., & Mathiyarasu, J. (2020) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2020.116024 · Cited by 239
Nano-enabled biosensing systems for intelligent healthcare: towards COVID-19 management
Mujawar, M. A., Gohel, H., Bhardwaj, S. K., Srinivasan, S. S., Hickman, N., & Kaushik, A. K. (2020) · Materials Today Chemistry · doi:10.1016/j.mtchem.2020.100306 · Cited by 229
Self-citationKaushik, A. K., Dhau, J. S., Gohel, H., Mishra, Y. K., Kateb, B., Kim, N., & Goswami, D. Y. (2020) · ACS Applied Bio Materials · doi:10.1021/acsabm.0c01004 · Cited by 229
Self-citationCorrie, S. R., Coffey, J. W., Islam, J., Markey, K. A., & Kendall, M. A. F. (2015) · The Analyst · doi:10.1039/c5an00464k · Cited by 220
Flexible Substrate-Based Devices for Point-of-Care Diagnostics
Wang, S., Chinnasamy, T., Lifson, M. A., İnci, F., & Demirci, U. (2016) · Trends in biotechnology · doi:10.1016/j.tibtech.2016.05.009 · Cited by 216
Sanghavi, B. J., Moore, J. A., Chávez, J. L., Hagen, J. A., Kelley‐Loughnane, N., Chou, C., & Swami, N. S. (2015) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2015.11.044 · Cited by 188
Nanomolar Binding of Steroids to Cucurbit[n]urils: Selectivity and Applications
Lazar, A. I., Biedermann, F., Mustafina, K. R., Assaf, K. I., Hennig, A., & Nau, W. M. (2016) · Journal of the American Chemical Society · doi:10.1021/jacs.6b07655 · Cited by 186
Chemical analysis in saliva and the search for salivary biomarkers – a tutorial review
Ngamchuea, K., Chaisiwamongkhol, K., Batchelor‐McAuley, C., & Compton, R. G. (2017) · The Analyst · doi:10.1039/c7an01571b · Cited by 172
Nano-biosensors to detect beta-amyloid for Alzheimer's disease management
Kaushik, A. K., Jayant, R. D., Tiwari, S., Vashist, A., & Nair, M. G. (2016) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2016.01.065 · Cited by 167
Self-citationA low-cost miniaturized potentiostat for point-of-care diagnosis
Cruz, A. F. D., Norena, N., Kaushik, A. K., & Bhansali, S. (2014) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2014.06.053 · Cited by 165
Self-citationApplication of 2D Non-Graphene Materials and 2D Oxide Nanostructures for Biosensing Technology
Shavanova, K. E., Bakakina, Y. S., Burkova, I., Shtepliuk, I., Viter, R., Úbelis, A., Beni, V., Starodub, N. F., Yakimova, R., & Khranovskyy, V. (2016) · Sensors · doi:10.3390/s16020223 · Cited by 170
Carbon fiber based electrochemical sensor for sweat cortisol measurement
Sekar, M. K., Pandiaraj, M., Bhansali, S., Ponpandian, N., & Viswanathan, C. (2019) · Scientific Reports · doi:10.1038/s41598-018-37243-w · Cited by 167
Self-citationExtended gate field-effect-transistor for sensing cortisol stress hormone
Sheibani, S., Capua, L., Kamaei, S., Akbari, S. S. A., Zhang, J., Guerin, H., & Ionescu, A. M. (2021) · Communications Materials · doi:10.1038/s43246-020-00114-x · Cited by 158
Wang, K., Yap, L. W., Gong, S. P., Wang, R., Wang, S. J., & Cheng, W. (2021) · Advanced Functional Materials · doi:10.1002/adfm.202008347 · Cited by 160
Vabbina, P. K., Kaushik, A. K., Pokhrel, N., Bhansali, S., & Pala, N. (2014) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2014.07.026 · Cited by 155
Self-citationLee, H.-B., Meeseepong, M., Trung, T. Q., Kim, B.-Y., & Lee, N.-E. (2020) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2020.112133 · Cited by 153
Aptamer-Based Lateral Flow Biosensor for Rapid Detection of Salivary Cortisol
Dalirirad, S., Han, D., & Steckl, A. J. (2020) · ACS Omega · doi:10.1021/acsomega.0c03223 · Cited by 154
Smart Contact Lenses as Wearable Ophthalmic Devices for Disease Monitoring and Health Management
Seo, H., Chung, W. G., Kwon, Y. W., Kim, S., Hong, Y., Park, W., Kim, E., Lee, J., Lee, S., Kim, M., et al. (2023) · Chemical Reviews · doi:10.1021/acs.chemrev.3c00290 · Cited by 157
Cheng, C., Li, X., Xu, G., Lu, Y., Low, S. S., Liu, G., Zhu, L., Li, C., & Liu, Q. (2020) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2020.112782 · Cited by 153
Current Trends of Nanobiosensors for Point-of-Care Diagnostics
Noah, N. M., & Ndangili, P. M. (2019) · Journal of Analytical Methods in Chemistry · doi:10.1155/2019/2179718 · Cited by 149
Towards detection and diagnosis of Ebola virus disease at point-of-care
Kaushik, A. K., Tiwari, S., Jayant, R. D., Marty, A. M., & Nair, M. (2015) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2015.08.040 · Cited by 142
Self-citationFlexible nanoporous tunable electrical double layer biosensors for sweat diagnostics
Munje, R. D., Muthukumar, S., Panneer Selvam, A., & Prasad, S. (2015) · Scientific Reports · doi:10.1038/srep14586 · Cited by 138
Microfluidic paper-based wearable electrochemical biosensor for reliable cortisol detection in sweat
Fiore, L., Mazzaracchio, V., Serani, A., Fabiani, G., Fabiani, L., Volpe, G., Moscone, D., Bianco, G. M., Occhiuzzi, C., Marrocco, G., et al. (2022) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2022.133258 · Cited by 139
High-Affinity Nucleic-Acid-Based Receptors for Steroids
Yang, K.-A., Chun, H.-S., Zhang, Y., Pecic, S., Nakatsuka, N., Andrews, A. M., Worgall, T. S., & Stojanović, M. N. (2017) · ACS Chemical Biology · doi:10.1021/acschembio.7b00634 · Cited by 130
He, H., Fu, Y., Zhao, T., Gao, X.-C., Xing, L., Zhang, Y., & Xue, X. (2017) · Nano Energy · doi:10.1016/j.nanoen.2017.07.033 · Cited by 129
Sweat and saliva cortisol response to stress and nutrition factors
Pearlmutter, P., Derose, G., Samson, C., Linehan, N., Cen, Y., Begdache, L., Won, D., & Koh, A. (2020) · Scientific Reports · doi:10.1038/s41598-020-75871-3 · Cited by 128
Wearable and Implantable Cortisol‐Sensing Electronics for Stress Monitoring
Ok, J., Park, S., Jung, Y. H., & Kim, T. (2023) · Advanced Materials · doi:10.1002/adma.202211595 · Cited by 131
Review—Towards Wearable Sensor Platforms for the Electrochemical Detection of Cortisol
Sekar, M. K., Sriramprabha, R., Sekhar, P. K., Bhansali, S., Ponpandian, N., Manickam, P., & Viswanathan, C. (2020) · Journal of The Electrochemical Society · doi:10.1149/1945-7111/ab7e24 · Cited by 119
Self-citationElectrochemical sensors for cortisol detections: Almost there
Zea, M., Bellagambi, F. G., Halima, H. B., Zine, N., Jaffrezic-Renault, N. J., Villa, R., Gabriel, G., & Errachid, A. (2020) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2020.116058 · Cited by 122
Jia, M., Chew, W. M., Feinstein, Y., Skeath, P., & Sternberg, E. M. (2016) · The Analyst · doi:10.1039/c5an02387d · Cited by 113
Xiao, W., Huang, C., Xu, F., Yan, J., Bian, H., Fu, Q., Xie, K., Wang, L., & Tang, Y. (2018) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2018.03.110 · Cited by 113
System Architectures in Wearable Electrochemical Sensors
Steinberg, M. D., Kassal, P., & Steinberg, I. M. (2016) · Electroanalysis · doi:10.1002/elan.201600094 · Cited by 110
Nardo, F. D., Cavalera, S., Baggiani, C., Giovannoli, C., & Anfossi, L. (2019) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.9b11559 · Cited by 109
Ganguly, A., Lin, K., Muthukumar, S., & Prasad, S. (2020) · ACS Sensors · doi:10.1021/acssensors.0c01754 · Cited by 110
Recent advances in electrochemiluminescence devices for point-of-care testing
Gao, W., Saqib, M., Qi, L., Zhang, W. Y., & Xu, G. (2017) · Current Opinion in Electrochemistry · doi:10.1016/j.coelec.2017.03.003 · Cited by 103
Biomolecule sensors based on organic electrochemical transistors
Lu, Z., Xu, K., Xiao, K. J., Xu, Q., Wang, L., Li, P., Zhou, J., Zhao, D., Bai, L., Cheng, Y., et al. (2025) · npj Flexible Electronics · doi:10.1038/s41528-025-00383-x · Cited by 103
Jo, S., Lee, W., Park, J., Kim, W., Kim, W., Lee, G., Lee, H., Hong, J., & Park, J. (2019) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2019.127424 · Cited by 98
Printed microfluidic sweat sensing platform for cortisol and glucose detection
Naik, A. R., Zhou, Y., Dey, A. A., Arellano, D. L. G., Okoroanyanwu, U., Secor, E. B., Hersam, M. C., Morse, J. D., Rothstein, J. P., Carter, K. R., et al. (2021) · Lab on a Chip · doi:10.1039/d1lc00633a · Cited by 101
A Portable 3D Microfluidic Origami Biosensor for Cortisol Detection in Human Sweat
Weng, X., Fu, Z., Zhang, C., Jiang, W. R., & Jiang, H. M. (2022) · Analytical Chemistry · doi:10.1021/acs.analchem.1c04508 · Cited by 98
Sensitive electrochemical detection of Salmonella with chitosan–gold nanoparticles composite film
Xiang, C., Li, R., Adhikari, B., She, Z., Li, Y., & Kraatz, H. (2015) · Talanta · doi:10.1016/j.talanta.2015.03.033 · Cited by 94
Usha, S. P., Shrivastav, A. M., & Gupta, B. D. (2016) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2016.08.040 · Cited by 94
Current methods for stress marker detection in saliva
Giacomello, G., Scholten, A., & Parr, M. K. (2020) · Journal of Pharmaceutical and Biomedical Analysis · doi:10.1016/j.jpba.2020.113604 · Cited by 99
Inkjet Printing: A Viable Technology for Biosensor Fabrication
Hussain, A., Abbas, N., & Ali, A. (2022) · Chemosensors · doi:10.3390/chemosensors10030103 · Cited by 94
Comparison of Colorimetric Analyses to Determine Cortisol in Human Sweat
Tu, E., Pearlmutter, P., Tiangco, M., Derose, G., Begdache, L., & Koh, A. (2020) · ACS Omega · doi:10.1021/acsomega.0c00498 · Cited by 90
Luppa, P. B., Bietenbeck, A., Beaudoin, C. A., & Giannetti, A. E. (2016) · Biotechnology Advances · doi:10.1016/j.biotechadv.2016.01.003 · Cited by 90
Özgöçer, T., Yıldız, S., & Uçar, C. (2016) · Journal of Immunoassay and Immunochemistry · doi:10.1080/15321819.2016.1230130 · Cited by 88
Electrochemical Biosensors for Early Stage Zika Diagnostics
Kaushik, A. K., Tiwari, S., Jayant, R. D., Vashist, A., Nikkhah-Moshaie, R., El‐Hage, N., & Nair, M. G. (2016) · Trends in biotechnology · doi:10.1016/j.tibtech.2016.10.001 · Cited by 88
Self-citationInjectable Biomedical Devices for Sensing and Stimulating Internal Body Organs
Jung, Y. H., Kim, J. U., Lee, J. S., Shin, J. H., Jung, W., Ok, J., & Kim, T. (2020) · Advanced Materials · doi:10.1002/adma.201907478 · Cited by 88
Cortisol detection methods for stress monitoring in connected health
Iqbal, T., Elahi, A., Wijns, W. C., & Shahzad, A. (2023) · Health Sciences Review · doi:10.1016/j.hsr.2023.100079 · Cited by 94
Miniaturized high-performance liquid chromatography instrumentation
Lynch, K. B., Chen, A., & Liu, S. (2017) · Talanta · doi:10.1016/j.talanta.2017.09.016 · Cited by 85
Karuppaiah, G., Lee, M., Bhansali, S., & Manickam, P. (2023) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2023.115600 · Cited by 87
Self-citationSchlotz, W. (2018) · Psychoneuroendocrinology · doi:10.1016/j.psyneuen.2018.11.038 · Cited by 85
Jang, H.-J., Lee, T., Song, J., Russell, L., Li, H., Dailey, J., Searson, P. C., & Katz, H. E. (2018) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.7b18855 · Cited by 80
Lee, M. A., Wang, S., Jin, X., Bakh, N. A., Nguyen, F. T., Dong, J. I., Silmore, K. S., Gong, X., Pham, C., Jones, K. K., et al. (2020) · Advanced Healthcare Materials · doi:10.1002/adhm.202000429 · Cited by 80
Electrochemical Sensing of Cortisol: A Recent Update
Singh, A., Kaushik, A. K., Kumar, R., Nair, M. G., & Bhansali, S. (2014) · Applied Biochemistry and Biotechnology · doi:10.1007/s12010-014-0894-2 · Cited by 80
Self-citationSingh, N. K., Chung, S., Sveiven, M., & Hall, D. A. (2021) · ACS Omega · doi:10.1021/acsomega.1c03552 · Cited by 80
Jayant, R. D., Yndart, A., Sagar, V., Bhansali, S., Nair, M. G., Kaushik, A. K., & Atluri, V. S. R. (2015) · International Journal of Nanomedicine · doi:10.2147/ijn.s75514 · Cited by 78
Self-citationMitigation of stress: new treatment alternatives
Subhani, A. R., Kamel, N., Saad, M. N. M., Nandagopal, D., Kang, K., & Malik, A. S. (2017) · Cognitive Neurodynamics · doi:10.1007/s11571-017-9460-2 · Cited by 75
Prakash, J., Parveen, A., Mishra, Y. K., & Kaushik, A. K. (2020) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2020.112562 · Cited by 73
Self-citationTian, G., Zhou, Z., Li, M., Li, X., Xu, T., & Zhang, X. (2023) · Analytical Chemistry · doi:10.1021/acs.analchem.3c02392 · Cited by 74
Disposable electrochemical immunosensor for cortisol determination in human saliva
Kämäräinen, S., Mäki, M., Tolonen, T., Palleschi, G., Virtanen, V., Micheli, L., & Sesay, A. M. (2018) · Talanta · doi:10.1016/j.talanta.2018.05.039 · Cited by 72
Unobtrusive and Multimodal Wearable Sensing to Quantify Anxiety
Zheng, Y., Wong, T. C. H., Leung, B. H. K., & Poon, C. C. Y. (2016) · IEEE Sensors Journal · doi:10.1109/jsen.2016.2539383 · Cited by 71
Point of Care Sensing Devices: Better Care for Everyone
Kaushik, A. K., & Mujawar, M. A. (2018) · Sensors · doi:10.3390/s18124303 · Cited by 70
Self-citationNeužil, P., Campos, C. D. M., Wong, C. C., Soon, J. B. W., Reboud, J., & Manz, A. (2014) · Lab on a Chip · doi:10.1039/c4lc00310a · Cited by 68
Salivary Cortisol Determination on Smartphone-Based Differential Pulse Voltammetry System
Liu, J., Xu, N., Men, H., Li, S., Lu, Y., Low, S. S., Li, X., Zhu, L., Cheng, C., Xu, G., et al. (2020) · Sensors · doi:10.3390/s20051422 · Cited by 68
Kim, K. S., Ryeol, L. S., Kim, S., Lee, J. Y., Chung, C., Choe, W., & Yoo, P. J. (2016) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2016.09.135 · Cited by 66
Yadav, A. K., Verma, D., Kumar, A., Kumar, P., & Solanki, P. R. (2021) · Materials Today Chemistry · doi:10.1016/j.mtchem.2021.100443 · Cited by 65
Chronic and acute stress monitoring by electrophysiological signals from adrenal gland
Sunwoo, S. H., Lee, J. S., Bae, S., Shin, Y. J., Kim, C. S., Joo, S. Y., Choi, H. S., Suh, M., Kim, S. W., Choi, Y. J., et al. (2019) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1806392115 · Cited by 64
Çay, M. (2017) · Northern Clinics of Istanbul · doi:10.14744/nci.2017.42103 · Cited by 72
SERS-based immunoassay for monitoring cortisol-related disorders
Villa, J. E. L., Garcı́a, I., de Aberasturi, D. J., Pavlov, V. V., Sotomayor, M. D. P. T., & Liz‐Marzán, L. M. (2020) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2020.112418 · Cited by 63
Cortisol Immunosensors: A Literature Review
Karachaliou, C., Koukouvinos, G., Goustouridis, D., Raptis, I., Kakabakos, S. E., Petrou, P. S., & Livaniou, E. (2023) · Biosensors · doi:10.3390/bios13020285 · Cited by 63
Electrochemical Immunosensing of Saliva Cortisol
Pasha, S. K., Kaushik, A. K., Vasudev, A., Snipes, S. A., & Bhansali, S. (2013) · Journal of The Electrochemical Society · doi:10.1149/2.017402jes · Cited by 63
Self-citationKhan, M. S., Dighe, K., Wang, Z., Srivastava, I., Schwartz‐Duval, A. S., Misra, S. K., & Pan, D. (2018) · The Analyst · doi:10.1039/c8an02085j · Cited by 63
Singh, A., Ahmed, A., Sharma, A., & Arya, S. K. (2022) · Biosensors · doi:10.3390/bios12100910 · Cited by 65
Self-citationPruna, R., Palacio, F., Baraket, A., Zine, N., Streklas, A., Bausells, J., Errachid, A., & López, M. (2017) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2017.09.049 · Cited by 62
Grand Challenges in Bio-Nanotechnology to Manage the COVID-19 Pandemic
Paliwal, P., Sargolzaei, S., Bhardwaj, S. K., Bhardwaj, V., Dixit, C. K., & Kaushik, A. K. (2020) · Frontiers in Nanotechnology · doi:10.3389/fnano.2020.571284 · Cited by 61
Self-citationChen, Y., Fu, Q., Li, D., Xie, J., Ke, D., Song, Q., Tang, Y., & Wang, H. (2017) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-017-0605-2 · Cited by 62
Portable and wearable real-time stress monitoring: A critical review
Parlak, O. (2021) · Sensors and Actuators Reports · doi:10.1016/j.snr.2021.100036 · Cited by 60
Shirtcliff, E. A., Buck, R. L., Laughlin, M. J., Hart, T., Cole, C. R., & Slowey, P. D. (2015) · Clinical Therapeutics · doi:10.1016/j.clinthera.2015.02.014 · Cited by 61
Pandit, P., Crewther, B. T., Cook, C. J., Punyadeera, C., & Pandey, A. K. (2024) · Materials Advances · doi:10.1039/d3ma00937h · Cited by 58
Pusomjit, P., Teengam, P., Thepsuparungsikul, N., Sanongkiet, S., & Chailapakul, O. (2021) · Microchimica Acta · doi:10.1007/s00604-020-04692-y · Cited by 56
A Wearable Molecularly Imprinted Electrochemical Sensor for Cortisol Stable Monitoring in Sweat
Chen, Y., He, Z., Wu, Y., Bai, X., Li, Y., Yang, W., Liu, Y., & Li, R. (2025) · Biosensors · doi:10.3390/bios15030194 · Cited by 56
Pandey, A. K., Sharma, A. K., & Marques, C. (2020) · Materials · doi:10.3390/ma13071623 · Cited by 56
Shajari, S., Salahandish, R., Zare, A., Hassani, M., Moossavi, S., Munro, E., Rashid, R., Rosenegger, D., Bains, J. S., & Sanati Nezhad, A. (2023) · Advanced Science · doi:10.1002/advs.202204171 · Cited by 55
Balakrishnan, S. R., Hashim, U., Gopinath, S. C. B., POOPALAN, P., Ramayya, H. R., Omar, M. I., Haarindraprasad, R., & Perumal, V. (2015) · PLoS ONE · doi:10.1371/journal.pone.0137891 · Cited by 55
Hu, X., Wei, T., Wang, J., Liu, Z.-E., Li, X., Zhang, B., Li, Z., Li, L., & Yuan, Q. (2014) · Analytical Chemistry · doi:10.1021/ac5032308 · Cited by 55
Microfluidic immunosensor for rapid and highly-sensitive salivary cortisol quantification
Pinto, V. C., Sousa, P. J., Catarino, S. O., Correia‐Neves, M., & Minas, G. M. H. (2016) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2016.11.067 · Cited by 57
Rapid micromotor-based naked-eye immunoassay
de Ávila, B. E., Zhao, M., Campuzano, S., Ricci, F., Pingarrón, J. M., Mascini, M., & Wang, J. (2017) · Talanta · doi:10.1016/j.talanta.2017.02.068 · Cited by 54
Manickam, P., Fernandez, T. F., Umasankar, Y., Muneeswaran, G., Arizaleta, F., Urizar, G. G., & Bhansali, S. (2018) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2018.07.133 · Cited by 54
Self-citationHalicka, K., Meloni, F., Czok, M. J., Spychalska, K., Baluta, S., Malecha, K., Pilo, M. I., & Cabaj, J. (2022) · ACS Omega · doi:10.1021/acsomega.2c04134 · Cited by 54
Gevaerd, A., Watanabe, E. Y., Belli, C., Junior, L. H. M., & Bergamini, M. F. (2021) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2021.129532 · Cited by 53
Cardiac troponin biosensors: where are we now?
Upasham, S., Tanak, A. S., & Prasad, S. (2018) · Advanced Health Care Technologies · doi:10.2147/ahct.s138543 · Cited by 52
Metadata
- Authors
- Kaushik, A., Vasudev, A., Arya, S. K., Pasha, S. K., & Bhansali, S.
- Affiliations
- Florida International University (US)
- Agency for Science, Technology and Research (SG)
- Corresponding author
- Kaushik, Ajeet
- Bhansali, Shekhar
- Venue
- Biosensors and Bioelectronics
- Publisher
- Elsevier BV
- Published
- 2014-03
- Volume
- 53
- Pages
- 499-512
- ISSN
- 0956-5663 (ISSN-L)
- OpenAlex
- W2013008702
- Cited by
- 316
- References
- 115
- Citations by year
- 2026: 13 · 2025: 23 · 2024: 30 · 2023: 31 · 2022: 35 · 2021: 33 · 2020: 36 · 2019: 21 · 2018: 24 · 2017: 18 · 2016: 31 · 2015: 12 · 2014: 8
- FWCI
- 1.20
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Biosensors and Bioelectronics (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Science Foundation
- Forensic Sciences Foundation
- National Institute of Mental Health (R44MH085474)
- Topics
- Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Molecular Biology › Advanced biosensing and bioanalysis techniques
- Physical Sciences › Chemistry › Electrochemistry › Electrochemical Analysis and Applications
- Keywords
- cortisol detection, point-of-care testing, electrochemical immunosensor, microelectrodes, microfluidics, wearable sensors, diurnal cortisol rhythm, organophosphate exposure, personalized healthcare, label-free detection
- MeSH terms
- Humans, Hydrocortisone, Immunoassay, Biosensing Techniques, Point-of-Care Systems, Microfluidics, Precision Medicine
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review