Integrated Microfluidic Capillary Electrophoresis-Electrospray Ionization Devices with Online MS Detection for the Separation and Characterization of Intact Monoclonal Antibody Variants
The DOI matches its Crossref record; 46 of the 46 references it cites were found in a registry or an index.
Abstract
Here, we demonstrate an integrated microfluidic capillary electrophoresis-electrospray ionization (CE-ESI) device for the separation of intact monoclonal antibody charge variants with online mass spectrometric (MS) identification. The need for dynamic coating and zwitterionic background electrolyte (BGE) additives has been eliminated by utilizing surface chemistry within the device channels to control analyte adsorption and electroosmotic flow (EOF) while maintaining separation efficiency. The effectiveness of this strategy was illustrated with the separation of charge variants of Infliximab. Three major species corresponding to C-terminal lysine variants were separated with an average resolution of 0.80 and identified by mass difference. In addition to the lysine variants, masses were determined for minor acidic and basic species. The separation of these variants prior to MS analysis facilitated the identification of glycosylation patterns for each of the variants. The general applicability of this method was demonstrated by analyzing two additional monoclonal antibody species: an IgG2 antibody and an IgG1 antibody conjugate. The IgG2 proved to have similar modifications to Infliximab with lower relative abundances of the lysine variants. Analysis of the IgG1 drug conjugate further exemplified the advantages of MS detection; differences in the extent of antibody conjugation were detectable despite limited CE resolution. The CE-ESI-MS methodology described here is a rapid and generic strategy for the separation of intact mAb charge variants and facilitates the identification of variants through MS detection.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“The use of microchip CE-MS has also been demonstrated for the separation of intact monoclonal antibody variants where fast analysis was achieved (~ 4 min)137.”
From Recent trends of capillary electrophoresis‐mass spectrometry in proteomics research (2019) · cites this paper as [137]
Citations
Cites
The Impact of Glycosylation on the Biological Function and Structure of Human Immunoglobulins
Arnold, J. N., Wormald, M. R., Sim, R. B., Rudd, P. M., & Dwek, R. A. (2007) · Annual Review of Immunology · doi:10.1146/annurev.immunol.25.022106.141702 · Cited by 1,307
Jacobson, S. C., Hergenroder, R., Koutny, L. B., Warmack, R. J., & Ramsey, J. M. (1994) · Analytical Chemistry · doi:10.1021/ac00079a028 · Cited by 669
Self-citationHeterogeneity of Monoclonal Antibodies
Liu, H., Gaza-Bulseco, G., Faldu, D., Chumsae, C., & Sun, J. (2008) · Journal of Pharmaceutical Sciences · doi:10.1002/jps.21180 · Cited by 481
High-Speed Separations on a Microchip
Jacobson, S. C., Hergenroder, R., Koutny, L. B., & Ramsey, J. M. (1994) · Analytical Chemistry · doi:10.1021/ac00079a029 · Cited by 450
Self-citationMass spectrometry for structural characterization of therapeutic antibodies
Zhang, Z., Pan, H., & Chen, X. (2009) · Mass Spectrometry Reviews · doi:10.1002/mas.20190 · Cited by 316
Non-covalent capillary coatings for protein separations in capillary electrophoresis
Lucy, C. A., MacDonald, A. M., & Gulcev, M. D. (2008) · Journal of Chromatography A · doi:10.1016/j.chroma.2007.10.114 · Cited by 299
Disentangling electrospray spectra with maximum entropy
Ferrige, A. G., Seddon, M. J., Green, B. N., Jarvis, S. A., Skilling, J., & Staunton, J. (1992) · Rapid Communications in Mass Spectrometry · doi:10.1002/rcm.1290061115 · Cited by 261
Mellors, J. S., Jorabchi, K., Smith, L. M., & Ramsey, J. M. (2010) · Analytical Chemistry · doi:10.1021/ac902218y · Cited by 235
Self-citationMicrofluidic Devices for Electrokinetically Driven Parallel and Serial Mixing
Jacobson, S. C., McKnight, T. E., & Ramsey, J. M. (1999) · Analytical Chemistry · doi:10.1021/ac990576a · Cited by 233
Self-citationFormation of Pyroglutamic Acid from N-Terminal Glutamic Acid in Immunoglobulin Gamma Antibodies
Chelius, D., Jing, K., Lueras, A., Rehder, D. S., Dillon, T. M., Vizel, A., Rajan, R. S., Li, T., Treuheit, M. J., & Bondarenko, P. V. (2006) · Analytical Chemistry · doi:10.1021/ac051827k · Cited by 200
Research works printed but not linked: 3
- Haselberg R., 2013, Anal. Biochem., 85, 2289
- Berger S. J., 2008, Efficiently Comparing Batches of an Intact Monoclonal Antibody Using the Biopharmalynx Software Package
- Shion H., 2013, Structural Comparison of Infliximab and a Biosimilar via Subunit Analysis Using the Waters Biopharmaceutical Platform with UNIFI
Cited by
Fekete, S., Guillarme, D., Sandra, P., & Sandra, K. (2016) · Analytical Chemistry · doi:10.1021/acs.analchem.5b04561 · Cited by 258
Analytical Techniques for Metabolomic Studies: A Review
Segers, K., Declerck, S., Mangelings, D., Heyden, Y. V., & Eeckhaut, A. V. (2019) · Bioanalysis · doi:10.4155/bio-2019-0014 · Cited by 225
Sonker, M., Sahore, V., & Woolley, A. T. (2017) · Analytica Chimica Acta · doi:10.1016/j.aca.2017.07.043 · Cited by 171
“[53] reported an integrated μCE-electrospray ionization device for separation of monoclonal antibodies (mAbs).”
BackgroundCapillary Electrophoresis Separations of Glycans
Lu, G., Crihfield, C. L., Gattu, S., Veltri, L. M., & Holland, L. A. (2018) · Chemical Reviews · doi:10.1021/acs.chemrev.7b00669 · Cited by 174
Biochemical analysis on microfluidic chips
Wu, J., He, Z., Chen, Q., & Lin, J. (2016) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2016.03.013 · Cited by 162
“to separate intact monoclonal antibody charge variants with online MS identification [108].”
BackgroundRecent advances in protein analysis by capillary and microchip electrophoresis
Dawod, M., Arvin, N. E., & Kennedy, R. T. (2017) · The Analyst · doi:10.1039/c7an00198c · Cited by 146
Wang, P., & Kricka, L. J. (2018) · Clinical Chemistry · doi:10.1373/clinchem.2018.287052 · Cited by 131
Advances in capillary electrophoresis for the life sciences
Toraño, J. S., Ramautar, R., & de Jong, G. J. (2019) · Journal of Chromatography B · doi:10.1016/j.jchromb.2019.04.020 · Cited by 141
“introduced an MCE-MS device based on a surface coating procedure with chemical deposition of an aminopropylsilane layer and modification with polyethylene glycol [25,26].”
MethodInfluentialDai, J., Lamp, J., Xia, Q., & Zhang, Y. (2018) · Analytical Chemistry · doi:10.1021/acs.analchem.7b04608 · Cited by 123
Khatri, K., Klein, J. A., Haserick, J. R., Leon, D. R., Costello, C. E., McComb, M. E., & Zaia, J. (2017) · Analytical Chemistry · doi:10.1021/acs.analchem.7b00875 · Cited by 119
Opportunities and challenges of real-time release testing in biopharmaceutical manufacturing
Jiang, M., Severson, K. A., Love, J. C., Madden, H., Swann, P., Zang, L., & Braatz, R. D. (2017) · Biotechnology and Bioengineering · doi:10.1002/bit.26383 · Cited by 118
Redman, E. A., Mellors, J. S., Starkey, J. A., & Ramsey, J. M. (2016) · Analytical Chemistry · doi:10.1021/acs.analchem.5b03866 · Cited by 112
Self-citationRecent trends of capillary electrophoresis‐mass spectrometry in proteomics research
Gomes, F. P., & Yates, J. R. (2019) · Mass Spectrometry Reviews · doi:10.1002/mas.21599 · Cited by 112
Gahoual, R., Beck, A., Leize‐Wagner, E., & François, Y. (2016) · Journal of Chromatography B · doi:10.1016/j.jchromb.2016.05.028 · Cited by 116
“The CZE-ESI-MS method developed by Han and coworkers was used to perform a detailed middle-up characterization of a therapeutic IgG1 using different sample treatment.”
BackgroundInfluentialRecent advances in CE‐MS coupling: Instrumentation, methodology, and applications
Týčová, A., Ledvina, V., & Klepárník, K. (2017) · ELECTROPHORESIS · doi:10.1002/elps.201600366 · Cited by 99
Füssl, F., Trappe, A., Carillo, S., Jakes, C., & Bones, J. (2020) · Analytical Chemistry · doi:10.1021/acs.analchem.0c00185 · Cited by 95
Sorensen, M., Harmes, D. C., Stoll, D. R., Staples, G. O., Fekete, S., Guillarme, D., & Beck, A. (2016) · mAbs · doi:10.1080/19420862.2016.1203497 · Cited by 94
Stoll, D. R., Danforth, J., Zhang, K., & Beck, A. (2016) · Journal of Chromatography B · doi:10.1016/j.jchromb.2016.05.029 · Cited by 89
Application of microfluidic technology in food processing
He, S., Joseph, N., Feng, S., Jellicoe, M., & Raston, C. L. (2020) · Food & Function · doi:10.1039/d0fo01278e · Cited by 84
Biacchi, M., Gahoual, R., Said, N., Beck, A., Leize-Wagner, E., & François, Y.-N. (2015) · Analytical Chemistry · doi:10.1021/acs.analchem.5b00928 · Cited by 79
Towards structure-focused glycoproteomics
Chernykh, A., Kawahara, R., & Thaysen-Andersen, M. (2021) · Biochemical Society Transactions · doi:10.1042/bst20200222 · Cited by 80
Jooß, K., Hühner, J., Kiessig, S., Moritz, B., & Neusüß, C. (2017) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-017-0542-0 · Cited by 79
“During manufacture, purification, and storage of these complex proteins, different variants besides the main form can occur [4].”
BackgroundMethodLechner, A., Giorgetti, J., Gahoual, R., Beck, A., Leize‐Wagner, E., & François, Y. (2019) · Journal of Chromatography B · doi:10.1016/j.jchromb.2019.05.014 · Cited by 82
“The same year, Ramsey’s group developed and employed a CE-nanoESI-MS mi598 crochip for the analysis of intact infliximab [127].”
MethodRecent trends in capillary electrophoresis for complex samples analysis: A review
Ramos‐Payán, M., Ocaña-González, J. A., Fernández-Torres, R. M., Llobera, A., & Bello‐López, M. Á. (2017) · Electrophoresis · doi:10.1002/elps.201700269 · Cited by 76
Microfluidics for Protein Biophysics
Charmet, J., Arosio, P., & Knowles, T. P. J. (2017) · Journal of Molecular Biology · doi:10.1016/j.jmb.2017.12.015 · Cited by 74
Carillo, S., Jakes, C., & Bones, J. (2020) · Journal of Pharmaceutical and Biomedical Analysis · doi:10.1016/j.jpba.2020.113218 · Cited by 74
Wang, L., Bo, T., Zhang, Z., Wang, G., Tong, W., & Chen, D. D. Y. (2018) · Analytical Chemistry · doi:10.1021/acs.analchem.8b02175 · Cited by 68
Microfluidic-Mass Spectrometry Interfaces for Translational Proteomics
Pedde, R. D., Li, H., Borchers, C. H., & Akbari, M. (2017) · Trends in biotechnology · doi:10.1016/j.tibtech.2017.06.006 · Cited by 68
Haselberg, R., De Vijlder, T., Heukers, R., Smit, M. J., Romijn, E. P., Somsen, G. W., & Domínguez-Vega, E. (2018) · Analytica Chimica Acta · doi:10.1016/j.aca.2018.08.024 · Cited by 64
Applications of capillary electrophoresis for biopharmaceutical product characterization
Kumar, R., Guttman, A., & Rathore, A. S. (2022) · ELECTROPHORESIS · doi:10.1002/elps.202100182 · Cited by 58
Schwenzer, A., Kruse, L., Jooß, K., & Neusüß, C. (2023) · PROTEOMICS · doi:10.1002/pmic.202300135 · Cited by 63
Analysis of proteins and peptides by electromigration methods in microchips
Štěpánová, S., & Kašička, V. (2016) · Journal of Separation Science · doi:10.1002/jssc.201600962 · Cited by 61
Giorgetti, J., Beck, A., Leize‐Wagner, E., & François, Y. (2020) · Journal of Pharmaceutical and Biomedical Analysis · doi:10.1016/j.jpba.2020.113107 · Cited by 61
“On-line capillary electrophoresis-mass spectrometry coupling via an electrospray ionization interface (CE-ESI-MS) has already proved its ability to separate and characterize micro-variants of mAbs [11-13].”
BackgroundCharacterization of cetuximab Fc/2 dimers by off-line CZE-MS
François, Y., Biacchi, M., Saïd, N., Renard, C., Beck, A., Gahoual, R., & Leize‐Wagner, E. (2016) · Analytica Chimica Acta · doi:10.1016/j.aca.2015.12.033 · Cited by 55
“In 2015, Redman et al published a remarkable paper on characterization of intact mAb variants using microfluidic CZE-ESI devices [22].”
Dadouch, M., Ladner, Y., & Perrin, C. (2021) · Separations · doi:10.3390/separations8010004 · Cited by 53
Yang, Y., Wang, G., Song, T., Lebrilla, C. B., & Heck, A. J. R. (2017) · mAbs · doi:10.1080/19420862.2017.1290033 · Cited by 52
Camperi, J., Pichon, V., & Delaunay, N. (2019) · Journal of Pharmaceutical and Biomedical Analysis · doi:10.1016/j.jpba.2019.112921 · Cited by 51
Shen, X., Liang, Z.-J., Xu, T., Yang, Z., Wang, Q., Chen, D., Pham, L., Du, W., & Sun, L. (2021) · International Journal of Mass Spectrometry · doi:10.1016/j.ijms.2021.116541 · Cited by 43
“Le-Minh et al. investigated the conformational changes of Infliximab under stressed conditions using native CZE-MS [26].”
InfluentialMicrofluidic Approaches for the Characterization of Therapeutic Proteins
Kopp, M. R. G., & Arosio, P. (2018) · Journal of Pharmaceutical Sciences · doi:10.1016/j.xphs.2018.01.001 · Cited by 45
Redman, E. A., Ramos‐Payán, M., Mellors, J. S., & Ramsey, J. M. (2016) · Analytical Chemistry · doi:10.1021/acs.analchem.6b00622 · Cited by 41
Self-citationLe-Minh, V., Tran, N. T., Makky, A., Rosilio, V., Taverna, M., & Smadja, C. (2019) · Journal of Chromatography A · doi:10.1016/j.chroma.2019.05.050 · Cited by 42
Xu, T., Han, L., Thompson, A. M. G., & Sun, L. (2022) · Analytical Methods · doi:10.1039/d1ay01556g · Cited by 41
Lotter, C., Heiland, J. J., Thürmann, S., Mauritz, L., & Belder, D. (2016) · Analytical Chemistry · doi:10.1021/acs.analchem.5b04583 · Cited by 39
Ma, F., Raoufi, F., Bailly, M. A., Fayadat‐Dilman, L., & Tomazela, D. M. (2020) · mAbs · doi:10.1080/19420862.2020.1763762 · Cited by 40
“It has been characterized by a variety of analytical techniques, including traditional capillary electrophoresis (CE) separation, CE-MS, and other SCX-MS method.(17,29,30) The analysis of infliximab by SCXMS showed three main peaks with baseline resolution in addition to other low abundant peaks, indicating the presen…”
MethodFüssl, F., Criscuolo, A., Cook, K., Scheffler, K., & Bones, J. (2019) · Journal of Proteome Research · doi:10.1021/acs.jproteome.9b00375 · Cited by 39
Dada, O. O., Zhao, Y., Jaya, N., & Salas‐Solano, O. (2017) · Analytical Chemistry · doi:10.1021/acs.analchem.7b03643 · Cited by 40
Intact NIST monoclonal antibody characterization—Proteoforms, glycoforms—Using CE-MS and CE-LIF
Chen, C., Feng, H., Guo, R., Li, P., Laserna, A. K. C., Ji, Y., Ng, B. H., Li, S. F. Y., Khan, S. H., Paulus, A., et al. (2018) · Cogent Chemistry · doi:10.1080/23312009.2018.1480455 · Cited by 37
Cheng, M., Shu, H., Ye, P., Feng, X., Yan, G., Zhang, L., Yao, J. H., Bao, H., & Lu, H. (2021) · Analytical Chemistry · doi:10.1021/acs.analchem.1c00064 · Cited by 37
Black, W. A., Stocks, B. B., Mellors, J. S., Engen, J. R., & Ramsey, J. M. (2015) · Analytical Chemistry · doi:10.1021/acs.analchem.5b01179 · Cited by 35
Self-citationGoyon, A., François, Y., Colas, O., Beck, A., Veuthey, J., & Guillarme, D. (2018) · Electrophoresis · doi:10.1002/elps.201800131 · Cited by 35
Aich, U., Lakbub, J. C., & Liu, A. (2016) · Electrophoresis · doi:10.1002/elps.201500551 · Cited by 36
Recent advances of capillary electrophoresis-mass spectrometry instrumentation and methodology
Jiang, Y., He, M.-Y., Zhang, W.-J., Luo, P., Guo, D., Fang, X., & Xu, W. (2017) · Chinese Chemical Letters · doi:10.1016/j.cclet.2017.05.008 · Cited by 30
Nys, G., & Fillet, M. (2018) · Journal of Pharmaceutical and Biomedical Analysis · doi:10.1016/j.jpba.2018.07.011 · Cited by 33
Rapid and comprehensive monoclonal antibody Characterization using microfluidic CE-MS
Cao, L., Fabry, D., & Lan, K. (2021) · Journal of Pharmaceutical and Biomedical Analysis · doi:10.1016/j.jpba.2021.114251 · Cited by 29
Wu, Z., Wang, H., Wu, J., Huang, Y., Zhao, X., Nguyen, J. B., Rosconi, M. P., Pyles, E. A., Qiu, H., & Li, N. (2022) · Journal of Pharmaceutical and Biomedical Analysis · doi:10.1016/j.jpba.2022.115147 · Cited by 29
Xu, T., Zhang, F., Chen, D., Sun, L., Tomazela, D. M., & Fayadat‐Dilman, L. (2023) · mAbs · doi:10.1080/19420862.2023.2229102 · Cited by 30
Kerr, R. A., Keire, D. A., & Ye, H. (2019) · mAbs · doi:10.1080/19420862.2019.1599632 · Cited by 30
Microfluidic antibody affinity profiling of alloantibody-HLA interactions in human serum
Schneider, M. M., Scheidt, T., Priddey, A. J., Xu, C. K., Hu, M., Meisl, G., Devenish, S. R. A., Dobson, C. M., Kosmoliaptsis, V., & Knowles, T. P. J. (2023) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2023.115196 · Cited by 28
Identification of adeno-associated virus capsid proteins using ZipChip CE/MS
Zhang, Y., Wang, Y., Sosic, Z., Zang, L., Bergelson, S., & Zhang, W. (2018) · Analytical Biochemistry · doi:10.1016/j.ab.2018.06.006 · Cited by 27
Svejdal, R. R., Sticker, D., Sønderby, C., Kutter, J. P., & Rand, K. D. (2020) · Analytica Chimica Acta · doi:10.1016/j.aca.2020.09.062 · Cited by 24
Dadouch, M., Ladner, Y., Bich, C., Larroque, M., Larroque, C., Morel, J., Bonnet, P., & Perrin, C. (2019) · The Analyst · doi:10.1039/c9an01906e · Cited by 26
Coupling Microchip Electrospray Ionization Devices with High Pressure Mass Spectrometry
Gilliland, W. M., Mellors, J. S., & Ramsey, J. M. (2017) · Analytical Chemistry · doi:10.1021/acs.analchem.7b03484 · Cited by 24
Self-citationMorbioli, G. G., Mazzu-Nascimento, T., Aquino, A., Cervantes, C. A. C., & Carrilho, E. (2016) · Analytica Chimica Acta · doi:10.1016/j.aca.2016.06.019 · Cited by 25
Hartanto, H., Wu, M., Lam, M.-L., & Chen, T. (2020) · Biomicrofluidics · doi:10.1063/5.0031521 · Cited by 22
Dykstra, A. B., Flick, T. G., Lee, B., Blue, L. E., & Angell, N. (2021) · Journal of the American Society for Mass Spectrometry · doi:10.1021/jasms.0c00424 · Cited by 22
Dada, O. O., Zhao, Y., Jaya, N., & Salas-Solano, O. (2017) · Analytical Chemistry · doi:10.1021/acs.analchem.7b03405 · Cited by 20
Development of a Microchip CE-HPMS Platform for Cell Growth Monitoring
Gilliland, W. M., & Ramsey, J. M. (2018) · Analytical Chemistry · doi:10.1021/acs.analchem.8b03708 · Cited by 20
Self-citationKristoff, C. J., Li, C., Li, P., & Holland, L. (2020) · Analytical Chemistry · doi:10.1021/acs.analchem.9b03994 · Cited by 21
da Silva, M. R., Zaborowska, I., Carillo, S., & Bones, J. (2021) · Journal of Chromatography A · doi:10.1016/j.chroma.2021.462336 · Cited by 20
Mechanism of droplet-formation in a supersonic microfluidic spray device
Kartanas, T., Toprakcioglu, Z., Hakala, T. A., Levin, A., Herling, T. W., Daly, R., Charmet, J., & Knowles, T. P. J. (2020) · Applied Physics Letters · doi:10.1063/1.5145109 · Cited by 18
Henley, W. H., He, Y., Mellors, J. S., Batz, N. G., Ramsey, J. M., & Jorgenson, J. W. (2017) · Journal of Chromatography A · doi:10.1016/j.chroma.2017.07.059 · Cited by 18
“The aminopropyl silane surface was further modified with polyethylene glycol using a modification of the previously described method [28, 29].”
MethodSelf-citationMicrofluidic devices for protein analysis using intact and top‐down mass spectrometry
Shao, X., Huang, Y., & Wang, G. (2022) · Open Research (University of Surrey) · doi:10.1002/viw.20220032 · Cited by 15
Cheng, J., Wang, L., Rive, C. M., Holt, R. A., Morin, G. B., & Chen, D. D. Y. (2020) · Journal of Proteome Research · doi:10.1021/acs.jproteome.0c00223 · Cited by 15
Schairer, J., Höchsmann, A., Bauer, B., Höfer, V., Römer, J., & Neusüß, C. (2025) · mAbs · doi:10.1080/19420862.2025.2537116 · Cited by 14
“131 Giorgetti et al. also show some separation of variants using a 3% HAc BGE, but the deconvoluted masses for acidic, main, and basic proteoforms were similar and the author thus concluded that no specific proteoforms can be assigned.”
Höchsmann, A., Dhellemmes, L., Leclercq, L., Cottet, H., & Neusüß, C. (2024) · Electrophoresis · doi:10.1002/elps.202400084 · Cited by 14
“When using microchip electrophoresis (ME)-MS, it was possible to separate the main form and the two lysine variants from each other when using an acidic BGE [17–19].”
Johnson, R. T., To, N. H., Stobaugh, J. F., & Lunte, C. E. (2017) · Chromatographia · doi:10.1007/s10337-017-3326-y · Cited by 14
“Iterations of this device have since been used to detect peptides, intact proteins, and cell lysate [16, 17].”
MethodMoini, M., & Rollman, C. M. (2015) · Journal of the American Society for Mass Spectrometry · doi:10.1007/s13361-015-1314-8 · Cited by 13
“been introduced in both capillary and chip formats [18–21].”
Deyanova, E. G., Huang, R. Y. ‐., Madia, P. A., Nandi, P., Gudmundsson, O. S., & Chen, G. (2020) · Electrophoresis · doi:10.1002/elps.202000132 · Cited by 12
Nagy, C., Andrási, M., Hamidli, N., Gyémánt, G., & Gáspár, A. (2021) · Journal of Chromatography Open · doi:10.1016/j.jcoa.2021.100024 · Cited by 12
Pickens, C. A., Isenberg, S. L., Cuthbert, C. D., & Pétritis, K. (2021) · Clinical Chemistry · doi:10.1093/clinchem/hvab171 · Cited by 11
Schilling, M. R., Feng, P., Sosic, Z., & Traviglia, S. L. (2020) · Bioengineered · doi:10.1080/21655979.2020.1842651 · Cited by 11
“In addition, capillary electrophoresis coupled to mass spectrometry (CE-ESI-MS) is an emerging technology for the characterization of mAb glycosylation [6,7].”
Li, W., Chaihu, L., Jiang, J., Wu, B., Zheng, X., Dai, R., Tian, Y., Huang, Y., Wang, G., & Men, Y. (2022) · Analytical Chemistry · doi:10.1021/acs.analchem.2c00077 · Cited by 11
Wang, Z., Li, M., Xu, S., Sun, L., & Li, L. (2024) · Analytica Chimica Acta · doi:10.1016/j.aca.2024.342905 · Cited by 10
Füssl, F., Carillo, S., Millán‐Martín, S., Jakes, C., Bora, K., Liberatori, S., Graham, J. W. A., & Bones, J. (2023) · Journal of Pharmaceutical and Biomedical Analysis · doi:10.1016/j.jpba.2023.115494 · Cited by 10
Biopharmaceutical Applications of Capillary Electromigration Methods
Gahoual, R., Giorgetti, J., Beck, A., Leize‐Wagner, E., & François, Y. (2018) · Elsevier eBooks · doi:10.1016/b978-0-12-809375-7.00021-6 · Cited by 10
Cageling, R., Carillo, S., Boumeester, A. J., Lubbers-Geuijen, K., Bones, J., Jooß, K., & Somsen, G. W. (2024) · Journal of Pharmaceutical and Biomedical Analysis · doi:10.1016/j.jpba.2024.116301 · Cited by 8
Cotham, V. C., Liu, A. P., Wang, S., & Li, N. (2023) · Journal of Pharmaceutical and Biomedical Analysis · doi:10.1016/j.jpba.2023.115337 · Cited by 7
Atia, M. A., Šmejkal, P., Gupta, V., Haddad, P. R., & Breadmore, M. C. (2023) · Analytica Chimica Acta · doi:10.1016/j.aca.2023.341847 · Cited by 8
Shah, A., Desai, R., Cui, W., Harrahy, J., & Ivanov, A. R. (2023) · The Analyst · doi:10.1039/d2an01724e · Cited by 7
Rollo, D., Kulkarni, A., Yu, K., & Fabris, D. (2024) · Journal of the American Society for Mass Spectrometry · doi:10.1021/jasms.3c00411 · Cited by 7
Chen, Y., Xia, L., Xiao, X., & Li, G. (2022) · Journal of Separation Science · doi:10.1002/jssc.202101007 · Cited by 7
Progress in Top-Down LC-MS Analysis of Antibodies: Review
You, J., & Park, H.-M. (2023) · Biotechnology and Bioprocess Engineering · doi:10.1007/s12257-023-0011-x · Cited by 7
Biophysical Analysis of Therapeutic Antibodies in the Early Development Pipeline
Willis, L. F., Kapur, N., Radford, S. E., & Brockwell, D. J. (2024) · Biologics · doi:10.2147/btt.s486345 · Cited by 8
Madren, S. M., & Yi, L. (2022) · Electrophoresis · doi:10.1002/elps.202200129 · Cited by 6
Wu, Z., Wang, H., Zhao, X., Gong, C., Sidnam, S., Cantero‐Tubilla, B., Nedjic-Dugic, B., Li, M., Wu, J., Su, Y., et al. (2024) · Journal of Pharmaceutical Sciences · doi:10.1016/j.xphs.2024.05.022 · Cited by 5
Tools for the analysis and characterization of therapeutic protein species
Schlüter, H., Fuh, M. M., & Steffen, P. (2016) · Biosimilars · doi:10.2147/bs.s68103 · Cited by 5
Kaya, S. I., Çetinkaya, A., Çağlayan, M. G., & Özkan, S. A. (2021) · Electrophoresis · doi:10.1002/elps.202100193 · Cited by 5
Dai, J., Xia, Q., & Ji, C. (2022) · Methods in molecular biology · doi:10.1007/978-1-0716-2325-1_6 · Cited by 5
Zhu, G., Rodriguez, D. A. A., Gould, N., Ivanov, A. R., Jooß, K., & Sun, L. (2026) · Mass Spectrometry Reviews · doi:10.1002/mas.70028 · Cited by 5
Separation techniques for intact antibody analysis by mass spectrometry
Schneck, N. A., Khatri, K., Maust, M. D., Jennings, M. E., Peltier, J., & Kellie, J. F. (2022) · Journal of Liquid Chromatography & Related Technologies · doi:10.1080/10826076.2023.2199328 · Cited by 4
“The strategy was further applied to an antibody drug conjugate (ADC) and demonstrated the possibility to distinguish non-conjugated species from mono- and di-conjugated species.([57]) The same research group used the same microchip CE-nanoESI-MS interface for the characterization of lysine-linked ADCs.”
Background
Metadata
- Authors
- Redman, E. A., Batz, N. G., Mellors, J. S., & Ramsey, J. M.
- Affiliations
- University of North Carolina at Chapel Hill (US)
- Corresponding author
- Ramsey, J. Michael
- Venue
- Analytical Chemistry
- Publisher
- American Chemical Society (ACS)
- Published
- 2015-01-29
- Volume
- 87
- Issue
- 4
- Pages
- 2264-2272
- ISSN
- 0003-2700 (ISSN-L)
- 1520-6882
- OpenAlex
- W2315996892
- Cited by
- 136
- References
- 46
- Citations by year
- 2026: 6 · 2025: 5 · 2024: 8 · 2023: 11 · 2022: 11 · 2021: 15 · 2020: 14 · 2019: 13 · 2018: 13 · 2017: 19 · 2016: 17 · 2015: 4
- FWCI
- 8.79
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Analytical Chemistry (publishedVersion)
- PubMed (publishedVersion)
- Funders
- Center for Biomedical Microdevices, University of North Carolina, Chapel Hill
- National Institute of General Medical Sciences (5R01GM066018-11)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Microfluidic and Capillary Electrophoresis Applications
- Physical Sciences › Chemistry › Spectroscopy › Mass Spectrometry Techniques and Applications
- Life Sciences › Immunology and Microbiology › Immunology › Biosimilars and Bioanalytical Methods
- Keywords
- microfluidic capillary electrophoresis, infliximab, glycosylation patterns, electroosmotic flow control, surface chemistry, IgG2 antibodies, antibody conjugates, on-line mass spectrometry
- MeSH terms
- Immunoglobulin G, Electrophoresis, Capillary, Spectrometry, Mass, Electrospray Ionization, Microfluidic Analytical Techniques
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.