Poor prognosis for existing monitors in the intensive care unit
The DOI matches its Crossref record; 8 of the 8 references it cites were found in a registry or an index.
Abstract
OBJECTIVE: To identify areas requiring the most urgent improvement in the intensive care unit (ICU); and to accurately determine the positive predictive value of routine critical care patient monitoring alarms, as well as the common causes for false-positive alarms. DESIGN: Prospective, observational study. SETTING: A multidisciplinary ICU in a university-affiliated children's hospital (excluding children with primary heart disease). INTERVENTIONS: The occurrence rate, cause, and appropriateness of all alarms from tracked monitors were recorded by a trained observer and validated by the bedside nurse over a 10-wk period for a single bedspace at a time. MEASUREMENTS AND MAIN RESULTS: After 298 monitored hrs, 86% of a total 2,942 alarms were found to be false-positive alarms, while an additional 6% were classified as clinically irrelevant true alarms. Only 8% of all alarms tracked during the study period were determined to be true alarms with clinical significance. Alarms were also classified according to whether they were clearly associated with a "patient intervention" (18%), were clearly not associated with a patient intervention (74%), or had unclear association to interventions (8%). While 11% of "nonpatient intervention" alarms were clinically significant true alarms, only 2% of "patient intervention" alarms were so. Positive predictive values for the various devices ranged from < 1% for the pulse oximeter's heart rate signal to 74% for the arterial catheter's mean systemic blood pressure signal during periods free from patient interventions. The pulse oximeter caused false-positive alarms most frequently, with common reasons being bad data format/bad connection and poor contact. CONCLUSION: Efforts to develop intelligent monitoring systems have more potential to deliver significantly improved patient care by initially targeting especially weak areas in ICU monitoring, such as pulse oximetry reliability.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Lawless (1994) · doi:10.1097/00003246-199406000-00017 · Cited by 406
Audibility and identification of auditory alarms in the operating room and intensive care unit
MOMTAHAN, K., HÉTU, R., & TANSLEY, B. (1993) · Ergonomics · doi:10.1080/00140139308967986 · Cited by 168
Are there too many alarms in the intensive care unit? An overview of the problems
Meredith, C., & Edworthy, J. R. (1995) · Journal of Advanced Nursing · doi:10.1046/j.1365-2648.1995.21010015.x · Cited by 152
Survey of alarms in an intensive therapy unit*
O'CARROLL, T. M. (1986) · Anaesthesia · doi:10.1111/j.1365-2044.1986.tb12844.x · Cited by 117
The median filter as a preprocessor for a patient monitor limit alarm system in intensive care
Mäkivirta, A., Koski, E. M. J., Kari, A., & Sukuvaara, T. (1991) · Computer Methods and Programs in Biomedicine · doi:10.1016/0169-2607(91)90039-v · Cited by 99
Frequency and reliability of alarms in the monitoring of cardiac postoperative patients
Koski, E. M. J., Mäkivirta, A., Sukuvaara, T., & Kari, A. (1990) · International journal of clinical monitoring and computing · doi:10.1007/bf01724206 · Cited by 76
Alarms and their limits in monitoring
Beneken, J. E. W., & van der Aa, J. J. (1989) · Journal of Clinical Monitoring and Computing · doi:10.1007/bf01627457 · Cited by 45
Koski, E. M. J., Sukuvaara, T., Mäkivirta, A., & Kari, A. (1994) · International journal of clinical monitoring and computing · doi:10.1007/bf01259556 · Cited by 33
Research works printed but not linked: 5
- Sara, 1986, Anaesth Intensive Care, 14, 448, 10.1177/0310057X8601400421 · doi:10.1177/0310057x8601400421
- Finley, 1991, Can J Anaesth, 38, 958, 10.1007/BF03008611 · doi:10.1007/bf03008611
- Crew, 1991, Intensive Care Med, 17, 359, 10.1007/BF01716197 · doi:10.1007/bf01716197
- Koski, 1992, Int J Clin Monit Comput, 8, 289, 10.1007/BF01739130 · doi:10.1007/bf01739130
- Fukui, 1989, J Clin Monit, 5, 211, 10.1007/BF01627458 · doi:10.1007/bf01627458
Cited by
Monitor Alarm Fatigue : An Integrative Review
Cvach, M. (2012) · Biomedical Instrumentation & Technology · doi:10.2345/0899-8205-46.4.268 · Cited by 632
Sendelbach, S., & Funk, M. (2013) · AACN Advanced Critical Care · doi:10.4037/nci.0b013e3182a903f9 · Cited by 497
Monitor Alarm Fatigue: Standardizing Use of Physiological Monitors and Decreasing Nuisance Alarms
Graham, K. C., & Cvach, M. (2010) · American Journal of Critical Care · doi:10.4037/ajcc2010651 · Cited by 459
Machine Learning and Decision Support in Critical Care
Johnson, A. E. W., Ghassemi, M. M., Nemati, S., Niehaus, K. E., Clifton, D., & Clifford, G. D. (2016) · Proceedings of the IEEE · doi:10.1109/jproc.2015.2501978 · Cited by 432
Li, Q., Mark, R. G., & Clifford, G. D. (2008) · Physiological Measurement · doi:10.1088/0967-3334/29/1/002 · Cited by 426
Jubran, A. (2004) · Intensive Care Medicine · doi:10.1007/s00134-004-2399-x · Cited by 401
Orphanidou, C., Bonnici, T., Charlton, P., Clifton, D., Vallance, D., & Tarassenko, L. (2014) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2014.2338351 · Cited by 368
Steinberg, J. S., Varma, N., Cygankiewicz, I., Aziz, P., Balsam, P., Baranchuk, A., Cantillon, D. J., Dilaveris, P., Dubner, S. J., El-Sherif, N., et al. (2017) · Heart Rhythm · doi:10.1016/j.hrthm.2017.03.038 · Cited by 303
ECG Signal Quality During Arrhythmia and Its Application to False Alarm Reduction
Behar, J., Oster, J., Li, Q., & Clifford, G. D. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2240452 · Cited by 293
Masimo Signal Extraction Pulse Oximetry
Goldman, J. M., Petterson, M. T., Kopotic, R. J., & Barker, S. J. (2000) · Journal of Clinical Monitoring and Computing · doi:10.1023/a:1011493521730 · Cited by 267
Alarm Algorithms in Critical Care Monitoring
Imhoff, M., & Kuhls, S. (2006) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000204385.01983.61 · Cited by 266
Secondary Analysis of Electronic Health Records
Data, M. C. (2016) · OAPEN (The OAPEN Foundation) · doi:10.1007/978-3-319-43742-2 · Cited by 246
Improving Alarm Performance in the Medical Intensive Care Unit Using Delays and Clinical Context
Görges, M., Markewitz, B. A., & Westenskow, D. R. (2009) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e31819bdfbb · Cited by 211
Intensive care unit alarms—How many do we need?*
Siebig, S., Kuhls, S., Imhoff, M., Gather, U., Schölmerich, J., & Wrede, C. E. (2010) · Critical Care Medicine · doi:10.1097/ccm.0b013e3181cb0888 · Cited by 206
“However, the number of induced alarms was higher in our study compared with others (5, 7, 8).”
ResultReducing false alarm rates for critical arrhythmias using the arterial blood pressure waveform
Aboukhalil, A., Nielsen, L., Saeed, M., Mark, R. G., & Clifford, G. D. (2008) · Journal of Biomedical Informatics · doi:10.1016/j.jbi.2008.03.003 · Cited by 204
“Multiple expert reviews of 5,386 critical ECG arrhythmia alarms from a total of 447 adult patient records in the MIMIC II database were made using the associated 41,301 hours of simultaneous ECG and arterial blood pressure (ABP) waveforms.”
MethodClifton, L., Clifton, D. A., Pimentel, M. A. F., Watkinson, P. J., & Tarassenko, L. (2014) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2013.2293059 · Cited by 199
“Such simplistic schemes result in large numbers of false alerts, which make these devices largely unusable in clinical practice [4]–[6].”
BackgroundPaine, C. W., Goel, V. V., Ely, E., Stave, C. D., Stemler, S., Zander, M., & Bonafide, C. P. (2016) · Journal of Hospital Medicine · doi:10.1002/jhm.2520 · Cited by 183
Neonatal and Pediatric Pulse Oximetry
Salyer, J. W. (2003) · Respiratory Care · doi:10.4187/respcare.03480386 · Cited by 183
“Pediatr Emerg Care 1992;8(5):262– 264.”
Bonafide, C. P., Lin, R., Zander, M., Graham, C. S., Paine, C. W., Rock, W., Rich, A., Roberts, K. E., Fortino, M., Nadkarni, V. M., et al. (2015) · Journal of Hospital Medicine · doi:10.1002/jhm.2331 · Cited by 180
“The nonactionable alarm proportions we found were similar to other pediatric studies, reporting greater than 90% nonactionable alarms.(1,2) One other study has reported a relationship between alarm exposure and response time.”
ResultA Review of Current and Emerging Approaches to Address Failure-to-Rescue
Taenzer, A. H., Pyke, J. B., McGrath, S. P., & Warner, D. S. (2011) · Anesthesiology · doi:10.1097/aln.0b013e318219d633 · Cited by 177
Cho, O. M., Kim, H., Lee, Y. W., & Cho, I. (2016) · Healthcare Informatics Research · doi:10.4258/hir.2016.22.1.46 · Cited by 169
“In contrast, another research [26] reported that the rate of false alarms was 86%, and the largest number of false alarms was generated by a device measuring oxygen saturation.”
BackgroundValidation of an electronic surveillance system for acute lung injury
Herasevich, V., Yilmaz, M., Khan, H., Hubmayr, R. D., & Gajic, O. (2009) · Intensive Care Medicine · doi:10.1007/s00134-009-1460-1 · Cited by 164
“The study by Tsien and Fackler showed that 86% of alarms triggered in an ICU are false, another 6% true but clinically irrelevant, and only 8% are of true clinically importance [23].”
BackgroundTemporal abstraction in intelligent clinical data analysis: A survey
Stacey, M., & McGregor, C. (2007) · Artificial Intelligence in Medicine · doi:10.1016/j.artmed.2006.08.002 · Cited by 146
“The deficiencies of the threshold alarming model are widely reported and recognised [2—4], and highlight the need for a more intelligent alarming model that includes a method for detecting and correlating temporal trends and patterns rather than simple instantaneous threshold values.”
BackgroundMethodInfluentialSurveillance of Medical Device–Related Hazards and Adverse Events in Hospitalized Patients
Samore, M. H. (2004) · JAMA · doi:10.1001/jama.291.3.325 · Cited by 142
A trend-based alarm system to improve patient monitoring in intensive care units
Charbonnier, S., & Gentil, S. (2007) · Control Engineering Practice · doi:10.1016/j.conengprac.2006.12.005 · Cited by 64
Stop the Noise: A Quality Improvement Project to Decrease Electrocardiographic Nuisance Alarms
Sendelbach, S., Wahl, S., Anthony, A., & Shotts, P. (2015) · Critical Care Nurse · doi:10.4037/ccn2015858 · Cited by 64
Reduction of false arrhythmia alarms using signal selection and machine learning
Eerikäinen, L. M., Vanschoren, J., Rooijakkers, M. J., Vullings, R., & Aarts, R. M. (2016) · Physiological Measurement · doi:10.1088/0967-3334/37/8/1204 · Cited by 48
“The number of false alarms in intensive care units (ICUs) have been reported to be 40 – 86% (Lawless 1994, Tsien and Fackler 1997, Siebig et al 2010).”
BackgroundGambarotta, N., Aletti, F., Baselli, G., & Ferrario, M. (2016) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-016-1453-5 · Cited by 45
Charbonnier, S., & Gentil, S. (2010) · International Journal of Adaptive Control and Signal Processing · doi:10.1002/acs.1123 · Cited by 26
Metadata
- Authors
- Tsien, C. L., & Fackler, J. C.
- Venue
- Critical Care Medicine
- Published
- 1997-04
- Volume
- 25
- Issue
- 4
- Pages
- 614-619
- ISSN
- 0090-3493 (ISSN-L)
- OpenAlex
- W2002971490
- Cited by
- 309
- References
- 8
- FWCI
- 8.17
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.