Home blood pressure monitoring: methodology, clinical relevance and practical application: a 2021 position paper by the Working Group on Blood Pressure Monitoring and Cardiovascular Variability of the European Society of Hypertension
Open accessThe DOI matches its Crossref record, with a small difference in title or year; 281 of the 294 references it cites were found in a registry or an index; 30 of the 413 sources its claims cite were checked for support.
Abstract
The present paper provides an update of previous recommendations on Home Blood Pressure Monitoring from the European Society of Hypertension (ESH) Working Group on Blood Pressure Monitoring and Cardiovascular Variability sequentially published in years 2000, 2008 and 2010. This update has taken into account new evidence in this field, including a recent statement by the American Heart association, as well as technological developments, which have occurred over the past 20 years. The present document has been developed by the same ESH Working Group with inputs from an international team of experts, and has been endorsed by the ESH.
Overview
CoChat Summary
TL;DR
This 2021 ESH position paper updates prior home blood pressure monitoring (HBPM) recommendations using evidence published through January 2021 and provides practical guidance for clinical and research use. It concludes that HBPM is a reliable, widely accepted out-of-office BP method with strong prognostic value, but it should be used with validated upper-arm devices, standardized schedules, and awareness of important limitations and unresolved questions.
Key findings
- HBPM has better reproducibility than office BP (OBP) and similar reproducibility to ABPM; the paper notes that HBPM’s better reproducibility vs OBP is partly due to a higher number of readings and more standardized conditions.
- For diagnosis, HBPM and ABPM are complementary rather than interchangeable: HBPM has higher specificity and negative predictive value but lower sensitivity and positive predictive value than ABPM for white-coat and masked hypertension.
- Prognostically, home SBP/DBP correlated with left ventricular mass index in a meta-analysis of 14 studies with pooled correlation coefficients of 0.46/0.28; these were similar to ABPM (0.37/0.26, 9 studies) and stronger than OBP (0.23/0.19, 10 studies).
- In outcome prediction, HBP was associated with higher odds ratios than OBP in a meta-analysis: for total mortality 1.14/1.10 vs 1.07/1.02 per 10 mmHg SBP/DBP increase, cardiovascular mortality 1.29/1.17 vs 1.15/1.07, and cardiovascular events 1.14/1.13 vs 1.10/1.07.
- The paper recommends a diagnostic HBPM threshold of ≥135/85 mmHg and suggests treatment targets of systolic HBP between 125 and 130 mmHg and diastolic HBP ≤80 mmHg for most individuals, while noting these targets still need randomized trial confirmation.
Methods
- Position paper/update by the ESH Working Group on Blood Pressure Monitoring and Cardiovascular Variability, developed with input from international experts and endorsed by ESH.
- Based on evidence from papers published up to January 2021; a draft was prepared by a writing committee and circulated among all authors for review and approval.
- Synthesizes evidence from prior position papers, meta-analyses, cohort studies, trials, and validation studies to issue recommendations on methodology, device selection, clinical application, and research use.
Limitations
- The paper explicitly states several areas still need more evidence: diagnostic agreement between ABPM and HBPM, outcome-based thresholds/targets for HBPM, especially in special populations, and whether HBPM-guided management improves outcomes over OBP-guided management.
- It notes that evidence is not convincing that either HBP or ABP is superior for predicting outcomes, because of the lack of head-to-head comparisons.
- Evident from the text: many recommendations are based on heterogeneous observational studies and some trial data rather than definitive randomized intervention trials, and several special populations (children, pregnancy, CKD, large arms, atrial fibrillation, nocturnal HBPM, cuffless devices) remain incompletely validated.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
18 Valid · 15 Not valid · 217 Not checked · 60 Own results and interpretations
Every source the paper cites for this claim was checked and supports it.
In the paper
“but in conjunction with co-interventions (including systematic medication titration by doctors, pharmacists, or patients; education; or lifestyle counselling), it leads to clinically significant BP reduction”
§ Home blood pressure monitoring: methodology, clinical relevance and practical application: a 2021 position paper by the Working Group on Blood Pressure Monitoring and Cardiovascular Variability of the European Society of Hypertension > HBPM: ADVANTAGES AND LIMITATIONS > Improving adherence to treatment and BP control (Box 6)
What the sources say
- Supportedcites [57] Tucker KL Sheppard JP Stevens R Bosworth HB Bove A Bray EP. Self-monitoring of blood pressure in hypertension: a systematic review and individual patient data meta-analysis. PLoS Med 2017; 14: e1002389. 28926573 PMC5604965 source
The Abstract Conclusions state that self-monitoring combined with co-interventions—including medication titration, education, or lifestyle counselling—leads to clinically significant BP reduction. The Results (Intensity of co-intervention) report a 6.1 mmHg reduction in clinic systolic BP with the most intensive support, compared with no clear reduction from self-monitoring alone. (LLM, conf=0.95)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
Every source the paper cites for this claim was checked and supports it.
In the paper
“which persists for at least 12 months [57].”
§ Home blood pressure monitoring: methodology, clinical relevance and practical application: a 2021 position paper by the Working Group on Blood Pressure Monitoring and Cardiovascular Variability of the European Society of Hypertension > HBPM: ADVANTAGES AND LIMITATIONS > Improving adherence to treatment and BP control (Box 6)
What the sources say
- Supportedcites [57] Tucker KL Sheppard JP Stevens R Bosworth HB Bove A Bray EP. Self-monitoring of blood pressure in hypertension: a systematic review and individual patient data meta-analysis. PLoS Med 2017; 14: e1002389. 28926573 PMC5604965 source
The Abstract’s Conclusions state that self-monitoring with co-interventions produces a blood-pressure reduction persisting for at least 12 months. The Results (Clinic BP) also report lower systolic and diastolic clinic blood pressure at 12 months. (LLM, conf=0.95)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“the clinical application and the accuracy of HBPM may be limited by certain conditions related either to the individual [63–66], to the procedure itself [4,67], to the oscillometric technique for BP measurement or to cuff-related issues [4,65,68–74].”
§ Home blood pressure monitoring: methodology, clinical relevance and practical application: a 2021 position paper by the Working Group on Blood Pressure Monitoring and Cardiovascular Variability of the European Society of Hypertension > HBPM: ADVANTAGES AND LIMITATIONS > Possible difficulties with HBPM use in clinical practice (Box 7)
What the sources say
- Not checked yetcites [63] Padwal RS Majumdar SR. Comparability of two commonly used automated office blood pressure devices in the severely obese. Blood Press Monit 2016; 21: 313– 315. 27100401 source
- Partly supportedcites [64] Azaki A Diab R Harb A Asmar R Chahine MN. Questionable accuracy of home blood pressure measurements in the obese population - validation of the Microlife WatchBP O3((R)) and Omron RS6((R)) devices according to the European Society of Hypertension-International Protocol. Vasc Health Risk Manag 2017; 13: 61– 69. 28280348 PMC5338962 sourceExperimental — accuracy not yet measured
The Introduction states that BP measurement 'is problematic in obese for several reasons and present a number of bias such as the arm circumference and cuff-related bias, the subcutaneous adiposity thickness, compression of the artery, and modification of the BP signal, as well as the specific hemodynamic parameters impacting the BP signal,' and that unsuitable/inappropriate cuff size is a common error leading to overestimation, with only the standard adult size available in home BP practice — so the individual-, oscillometric/signal- and cuff-related sources of inaccuracy are covered, and the Discussion (miscuffing as 'a serious source of error'; wrist characteristics and 'the BP signal itself' as accuracy issues) reinforces them. However, our copy offers no passage attributing limits of HBPM to 'the procedure itself' (measurement/user procedural errors), and the paper frames these biases specifically for the obese/large-arm population rather than as a general four-category taxonomy of HBPM clinical application and accuracy. The cuff-related part of the claim is supported; the procedure-related part is not. (LLM, conf=0.72)
- Not checked yetcites [65] Stergiou GS Kollias A Destounis A Tzamouranis D. Automated blood pressure measurement in atrial fibrillation: a systematic review and meta-analysis. J Hypertens 2012; 30: 2074– 2082. 22914573 source
- Partly supportedcites [66] Stergiou GS Dolan E Kollias A Poulter NR Shennan A Staessen JA. Blood pressure measurement in special populations and circumstances. J Clin Hypertens (Greenwich) 2018; 20: 1122– 1127. 30003705 PMC8030798 sourceExperimental — accuracy not yet measured
Our copy substantiates three of the four limiting categories, but for automated (oscillometric) BP monitors generally rather than for HBPM as such: the Introduction (Section 1) lists young children, pregnancy/preeclampsia, arm circumference >42 cm and atrial fibrillation as special populations in which an otherwise validated monitor 'may not be accurate'; the Pregnancy section (A. Shennan, Section 3) states that preeclampsia's reduced intravascular volume and increased peripheral resistance 'can affect oscillometric signals' and that oscillometric devices 'tend to under-record the BP in preeclampsia'; and the children/AF sections (Stergiou & Kollias) give the cuff rule (bladder length 80%-90% of arm circumference, width 40%-50%) and note that in AF 'the evidence regarding ... the usefulness of automated home BP measurement and ABPM is scarce'. Home monitors appear only as a subset within these discussions (e.g. 'six evaluated home monitors (one failed)' in children; the office/ambulatory/home meta-analysis in AF), so the claim's HBPM-specific framing is broader than the source. What our copy does not address for HBPM is the 'procedure itself' category: its procedural content is clinic/low-resource measurement advice (rest interval, two readings 1 minute apart, triplicate auscultatory readings in AF), not limits on home BP monitoring procedure. (LLM, conf=0.72)
- Supportedcites [4] Parati G Stergiou GS Asmar R Bilo G de Leeuw P Imai Y. ESH Working Group on Blood Pressure Monitoring. European Society of Hypertension guidelines for blood pressure monitoring at home: a summary report of the Second International Consensus Conference on Home Blood Pressure Monitoring. J Hypertens 2008; 26: 1505– 1526. 18622223 source
The ‘Measurement conditions and procedures’ section says individual activity and measurement circumstances can affect readings and lead to erroneous diagnosis. ‘Device calibration’ describes inaccurate oscillometric readings in some individuals, while ‘Cuff and bladder’ explains how incorrect cuff size distorts blood-pressure estimates. Together, these passages support the stated limitations of HBPM. (LLM, conf=0.93)
- Could not verifycites [67] Edmondson D Arndt J Alcantara C Chaplin W Schwartz JE. Self-esteem and the acute effect of anxiety on ambulatory blood pressure. Psychosom Med 2015; 77: 833– 841. 26230481 PMC4565744 sourceExperimental — accuracy not yet measured
Our copy of Edmondson et al. (2015) is a multilevel analysis of trait self-esteem and momentary anxiety's acute effect on awake ambulatory systolic BP in the Masked Hypertension Study; it makes no statement about home blood pressure monitoring (HBPM) accuracy or its clinical application. The only cuff-related content is a Method (Materials, 'Ambulatory blood pressure monitoring (ABPM)') procedural detail that 'a BP cuff was fitted to the participant's non-dominant arm with cuff size determined by upper arm circumference,' and the Limitations (Discussion) section addresses sample homogeneity, exclusion of CVD/medicated participants, and unmeasured fitness/diet — not oscillometric or individual/procedure-related measurement limitations. Searches for HBPM accuracy, oscillometric technique, cuff-related issues, and ABPM advantages/limitations returned no passage on this point. (LLM, conf=0.70)
- Not checked yetcites [68] Parati G Mendis S Abegunde D Asmar R Mieke S Murray A. Recommendations for blood pressure measuring devices for office/clinic use in low resource settings. Blood Press Monit 2005; 10: 3– 10. 15687867 source
- Could not verifycites [69] Stergiou GS O’Brien E Myers M Palatini P Parati G Kollias A. STRIDE BP Scientific Advisory Board. STRIDE BP international initiative for accurate blood pressure measurement: systematic review of published validation studies of blood pressure measuring devices. J Clin Hypertens (Greenwich) 2019; 21: 1616– 1622. 31654494 PMC8030378 sourceExperimental — accuracy not yet measured
Our copy of the STRIDE BP systematic review is confined to the methodology and results of its peer-review of 419 published validation studies (Abstract; sections 1-3 and 5); it contains no statement about conditions that limit the clinical application or accuracy of HBPM in individual patients. Searches for 'oscillometric technique', 'cuff-related issues', 'individual/patient factors' and 'limitations of home blood pressure monitoring' returned nothing on that topic. The nearest passages concern validation-study quality rather than clinical use — e.g. Table 1 in section 2 lists rejection criteria such as 'Violation in the procedure or sequence of test and reference device BP measurements', 'Validation in pregnancy without data in women with hypertension and preeclampsia' and 'Minor cuff distribution criteria', and section 3.3 reports that 338 (81%) validations used upper-arm cuff devices, 72 (17%) wrist and 9 (2%) finger devices — which neither states nor contradicts the claim as worded. (LLM, conf=0.68)
- Partly supportedcites [70] Stergiou GS Palatini P Asmar R Ioannidis JP Kollias A Lacy P. European Society of Hypertension Working Group on Blood Pressure Monitoring. Recommendations and Practical Guidance for performing and reporting validation studies according to the Universal Standard for the validation of blood pressure measuring devices by the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO). J Hypertens 2019; 37: 459– 466. 30702492 sourceExperimental — accuracy not yet measured
Our copy of Stergiou et al. (2019) supports the substance only in a different setting and at a different level: the 'Validation studies in general and special populations' section states there is 'theoretical and clinical evidence of different accuracy of BP monitors' in defined individual conditions (age <3 years, pregnancy including preeclampsia, arm circumference >42 cm, atrial fibrillation, for which no agreed reference/validation procedure exists), and Table 3 excludes 'presence of any arrhythmia'; the 'Reference BP measurement' and 'Validation study data reports' sections tie cuff bladder length (75-100% of mid-arm circumference) and width (37-50%) and cuff-size subgroup reporting to accuracy. But no passage addresses home blood pressure monitoring (HBPM) or its clinical application as such — the paper is a validation-standard guidance document whose worked example is a professional office oscillometric device (Test device/DISCUSSION sections) — so the claim's HBPM-specific framing is not stated by this source. (LLM, conf=0.65)
- Not checked yetcites [71] Palatini P Frick GN. Cuff and bladder: overlooked components of BP measurement devices in the modern era? Am J Hypertens 2012; 25: 136– 138. 22262253 source
hedgedstates a possible cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
The sources with a verdict support this claim; the rest weren’t checked yet, have no full text we could read, or their verdicts aren’t rated yet.
In the paper
“However, there is some inconsistency in the recommendations by scientific societies on the issue of miscuffing [1,4,90,98–100].”
§ Home blood pressure monitoring: methodology, clinical relevance and practical application: a 2021 position paper by the Working Group on Blood Pressure Monitoring and Cardiovascular Variability of the European Society of Hypertension > TECHNOLOGY OF HBP MONITORS > Cuff-based devices
What the sources say
- Not checked — no full text of the sourcecites [1] Williams B Mancia G Spiering W Agabiti Rosei E Azizi M Burnier M. List of authors/Task Force members. 2018 Practice guidelines for the management of arterial hypertension of the European Society of Hypertension and the European Society of Cardiology: ESH/ESC Task Force for the Management of Arterial Hypertension. J Hypertens 2018; 36: 2284– 2309. 30379783 source
- Supportedcites [4] Parati G Stergiou GS Asmar R Bilo G de Leeuw P Imai Y. ESH Working Group on Blood Pressure Monitoring. European Society of Hypertension guidelines for blood pressure monitoring at home: a summary report of the Second International Consensus Conference on Home Blood Pressure Monitoring. J Hypertens 2008; 26: 1505– 1526. 18622223 source
Box 4 in the Device selection and validation section (page not given) lists different recommended cuff-bladder dimensions from the British Hypertension Society and the American Heart Association. The Cuff and bladder section explains that an undersized or oversized bladder can distort blood-pressure readings, connecting those differing recommendations to miscuffing. (LLM, conf=0.86)
- Not checked yetcites [90] Stergiou GS Alpert B Mieke S Asmar R Atkins N Eckert S. A universal standard for the validation of blood pressure measuring devices: Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Collaboration Statement. J Hypertens 2018; 36: 472– 478. 29384983 PMC5796427 source
- Not checked yetcites [98] ISO 81060-2:2018/Amd 1:2020. Non-invasive sphygmomanometers -- Part 2: Clinical investigation of intermittent automated measurement type -- Amendment 1. Available at https://www.iso.org/standard/75432.html source
Checked before — its earlier verdict is held back until a new check runs.
- Could not verifycites [99] Palatini P Asmar R Bilo G Parati G. Stergiou G Parati G Mancia G. Cuff design for home blood pressure monitors. Updates in hypertension and cardiovascular protection; home blood pressure monitoring 2019. 13– 22. sourceExperimental — accuracy not yet measured
The Discussion (page unavailable) mentions different standards for validating cuffless blood-pressure devices and says cuff-based devices are not error-free. It does not address miscuffing or compare scientific societies’ recommendations on cuff sizing; searches for those terms found no relevant passage in our copy. (LLM, conf=0.76)
- Not checked — no full text of the sourcecites [100] O’Brien E Petrie J Littler WA De Swiet M Padfield PD Dillon MJ. Blood pressure measurement: recommendations of the British hypertension society 3rd edn. London: BMJ Publishing Group; 1997. source
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“and to achieve more satisfactory hypertension control rates [57,112–115]”
§ Home blood pressure monitoring: methodology, clinical relevance and practical application: a 2021 position paper by the Working Group on Blood Pressure Monitoring and Cardiovascular Variability of the European Society of Hypertension > TECHNOLOGY OF HBP MONITORS > Telemonitoring of home BP (HBPT) values
What the sources say
- Partly supportedcites [57] Tucker KL Sheppard JP Stevens R Bosworth HB Bove A Bray EP. Self-monitoring of blood pressure in hypertension: a systematic review and individual patient data meta-analysis. PLoS Med 2017; 14: e1002389. 28926573 PMC5604965 sourceExperimental — accuracy not yet measured
The Results (Clinic BP control; Intensity of co-intervention) report improved hypertension control with self-monitoring interventions. However, the Conclusions state that self-monitoring alone was not associated with better control; the improvement depended on accompanying co-interventions, a qualification missing from the claim. (LLM, conf=0.85)
- Not checked — no full text of the sourcecites [112] Omboni S Gazzola T Carabelli G Parati G. Clinical usefulness and cost effectiveness of home blood pressure telemonitoring: meta-analysis of randomized controlled studies. J Hypertens 2013; 31: 455– 467. 23299557 source
- Not checked yetcites [113] Duan Y Xie Z Dong F Wu Z Lin Z Sun N Xu J. Effectiveness of home blood pressure telemonitoring: a systematic review and meta-analysis of randomised controlled studies. J Hum Hypertens 2017; 31: 427– 437. 28332506 source
- Not checked yetcites [114] Margolis KL Asche SE Bergdall AR Dehmer SP Groen SE Kadrmas HM. Effect of home blood pressure telemonitoring and pharmacist management on blood pressure control: a cluster randomized clinical trial. JAMA 2013; 310: 46– 56. 23821088 PMC4311883 source
- Not checked yetcites [115] Margolis KL Asche SE Dehmer SP Bergdall AR Green BB Sperl-Hillen JM. Long-term outcomes of the effects of home blood pressure telemonitoring and pharmacist management on blood pressure among adults with uncontrolled hypertension: follow-up of a cluster randomized clinical trial. JAMA Netw Open 2018; 1: e181617. 30646139 PMC6324502 source
hedgedstates a possible cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
188 more claims · 5 with their cited sources checked
Create a free account to see all 310 claims and to verify the 326 references still open.
- Probably not a substantial claim. 27
- It may state two separate claims as one. 100
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“out-of-office BP evaluation using 24 h ambulatory or home BP monitoring is recommended by all major medical associations for obtaining further insights into the BP profile of an individual and how it relates to their daily activities.”
From Blood pressure and its variability: classic and novel measurement techniques (2022) · cites this paper as [44]
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“HBPM is currently recommended as the best method for long-term follow-up of patients who are being treated for hypertension6,12,44.”
From Blood pressure and its variability: classic and novel measurement techniques (2022) · cites this paper as [44]
The cited source says what the claim attributes to it.
In the citing paper
“Home BP monitoring, when performed according to recommendations, can provide complementary information surrounding BP control to that of clinic and ambulatory BP [54,55].”
From The Identification and Management of High Blood Pressure Using Exercise Blood Pressure: Current Evidence and Practical Guidance (2022) · cites this paper as [54]
In the citing paper
“Home BP was measured using the validated semiautomatic Omron M10-IT oscillometric device and following the guidelines of the European Society of Hypertension [23,24].”
From A low dose of daily licorice intake affects renin, aldosterone, and home blood pressure in a randomized crossover trial (2024) · cites this paper as [23]
Citations
Cites
(2013) · European Heart Journal · doi:10.1093/eurheartj/eht151 · Cited by 13,655
Self-citationWhelton, P. K., Carey, R. M., Aronow, W. S., Casey, D. E., Collins, K. J., Dennison Himmelfarb, C., DePalma, S. M., Gidding, S., Jamerson, K. A., Jones, D. W., et al. (2018) · Hypertension · doi:10.1161/hyp.0000000000000065 · Cited by 7,391
Self-citationMancia, G., Fagard, R., Narkiewicz, K., Redón, J., Zanchetti, A., Böhm, M., Christiaens, T., Cifkova, R., De Backer, G., Dominiczak, A., et al. (2013) · Journal of Hypertension · doi:10.1097/01.hjh.0000431740.32696.cc · Cited by 5,655
Self-citationWhelton, P. K., Carey, R. M., Aronow, W. S., Casey, D. E., Collins, K. J., Dennison Himmelfarb, C., DePalma, S. M., Gidding, S., Jamerson, K. A., Jones, D. W., et al. (2018) · Hypertension · doi:10.1161/hyp.0000000000000066 · Cited by 4,347
Self-citationFlynn, J. T., Kaelber, D. C., Baker-Smith, C. M., Blowey, D., Carroll, A. E., Daniels, S. R., de Ferranti, S. D., Dionne, J. M., Falkner, B., Flinn, S. K., et al. (2017) · Pediatrics · doi:10.1542/peds.2017-1904 · Cited by 3,651
A Controlled Trial of Renal Denervation for Resistant Hypertension
Bhatt, D. L., Kandzari, D. E., O'Neill, W. W., D'Agostino, R., Flack, J. M., Katzen, B. T., Leon, M. B., Liu, M., Mauri, L., Negoita, M., et al. (2014) · New England Journal of Medicine · doi:10.1056/nejmoa1402670 · Cited by 2,186
O'Brien, E., Asmar, R., Beilin, L., Imai, Y., Mallion, J.-M., Mancia, G., Mengden, T., Myers, M., Padfield, P., Palatini, P., et al. (2003) · Journal of Hypertension · doi:10.1097/00004872-200305000-00001 · Cited by 1,655
Self-citationEuropean Society of hypertension position paper on ambulatory blood pressure monitoring
O’Brien, E., Parati, G., Stergiou, G., Asmar, R., Beilin, L., Bilo, G., Clement, D., de la Sierra, A., de Leeuw, P., Dolan, E., et al. (2013) · Journal of Hypertension · doi:10.1097/hjh.0b013e328363e964 · Cited by 1,419
Self-citationLurbe, E., Agabiti-Rosei, E., Cruickshank, J. K., Dominiczak, A., Erdine, S., Hirth, A., Invitti, C., Litwin, M., Mancia, G., Pall, D., et al. (2016) · Journal of Hypertension · doi:10.1097/hjh.0000000000001039 · Cited by 1,379
Self-citation
Cited by
Mancia, G., Kreutz, R., Brunström, M., Burnier, M., Grassi, G., Januszewicz, A., Muiesan, M. L., Tsioufis, K., Agabiti-Rosei, E., Algharably, E. A. E., et al. (2023) · Journal of Hypertension · doi:10.1097/hjh.0000000000003480 · Cited by 3,473
Self-citation2024 ESC Guidelines for the management of elevated blood pressure and hypertension
McEvoy, J. W., McCarthy, C. P., Bruno, R. M., Brouwers, S., Canavan, M. D., Ceconi, C., Christodorescu, R. M., Daskalopoulou, S. S., Ferro, C. J., Gerdts, E., et al. (2024) · European Heart Journal · doi:10.1093/eurheartj/ehae178 · Cited by 2,298
Self-citationBlood pressure and its variability: classic and novel measurement techniques
Schutte, A. E., Kollias, A., & Stergiou, G. S. (2022) · Nature Reviews Cardiology · doi:10.1038/s41569-022-00690-0 · Cited by 335
Self-citationParati, G., Bilo, G., Kollias, A., Pengo, M., Ochoa, J. E., Castiglioni, P., Stergiou, G. S., Mancia, G., Asayama, K., Asmar, R., et al. (2023) · Journal of Hypertension · doi:10.1097/hjh.0000000000003363 · Cited by 220
Self-citationHypertension in children and adolescents
SIMONE, G. D., Mancusi, C., Hanssen, H., Genovesi, S., Lurbe, E., Parati, G., Sendžikaitė, S., Valerio, G., Bonito, P. D., Salvo, G. D., et al. (2022) · European Heart Journal · doi:10.1093/eurheartj/ehac328 · Cited by 184
Self-citation
Metadata
- Authors
- Parati, G., Stergiou, G. S., Bilo, G., Kollias, A., Pengo, M., Ochoa, J. E., Agarwal, R., Asayama, K., Asmar, R., Burnier, M., De La Sierra, A., Giannattasio, C., Gosse, P., Head, G., Hoshide, S., Imai, Y., Kario, K., Li, Y., Manios, E., Mant, J., McManus, R. J., Mengden, T., Mihailidou, A. S., Muntner, P., Myers, M., Niiranen, T., Ntineri, A., O’Brien, E., Octavio, J. A., Ohkubo, T., Omboni, S., Padfield, P., Palatini, P., Pellegrini, D., Postel-Vinay, N., Ramirez, A. J., Sharman, J. E., Shennan, A., Silva, E., Topouchian, J., Torlasco, C., Wang, J. G., Weber, M. A., Whelton, P. K., White, W. B., & Mancia, G.
- Affiliations
- IRCCS Istituto Auxologico Italiano (IT)
- National and Kapodistrian University of Athens (GR)
- Hypertension Center STRIDE-7, National and Kapodistrian University of Athens, School of Medicine, Third Department of Medicine, Sotiria Hospital, Athens, Greece
- Richard L. Roudebush VA Medical Center (US)
- Tohoku Institute for Management of Blood Pressure (JP)
- Foundation for laboratory medicine (CH)
- Centre Hospitalier Universitaire Vaudois (CH)
- Hypertension Unit, Department of Internal Medicine, Hospital Mútua Terrassa, University of Barcelona, Barcelona, Spain
- Azienda Socio Sanitaria Territoriale Grande Ospedale Metropolitano Niguarda (IT)
- Hôpital Saint-André (FR)
- Baker Heart and Diabetes Institute (AU)
- Jichi Medical University (JP)
- Shanghai Jiao Tong University (CN)
- University of Cambridge (GB)
- University of Oxford (GB)
- Kerckhoff Klinik (DE)
- Royal North Shore Hospital (AU)
- University of Alabama at Birmingham (US)
- Sunnybrook Health Science Centre (CA)
- University of Turku (FI)
- The Conway Institute, University College Dublin, Dublin, Ireland
- Experimental Cardiology, Department of Tropical Medicine Institute, Universidad Central de Venezuela, Venezuela
- Sechenov University (RU)
- University of Edinburgh (GB)
- University of Padua (IT)
- Ospedale Papa Giovanni XXIII (IT)
- Hypertension Unit, European Georges Pompidou Hospital, Paris, France
- Arterial Hypertension and Metabolic Unit, University Hospital, Fundacion Favaloro, Argentina
- University of Tasmania (AU)
- King's College London (GB)
- Research Institute of Cardiovascular Diseases of the University of Zulia, Venezuelan Foundation of Arterial Hypertension. Maracaibo, Venezuela
- Assistance Publique – Hôpitaux de Paris (FR)
- State University of New York (US)
- Tulane University (US)
- University of Connecticut (US)
- University of Milano-Bicocca (IT)
- Corresponding author
- Parati, Gianfranco
- ORCID
- Parati, Gianfranco
- Stergiou, George S.
- Bilo, Grzegorz
- Pengo, Martino
- Ochoa, Juan Eugenio
- Agarwal, Rajiv
- Asayama, Kei
- Asmar, Roland
- Burnier, Michel
- Giannattasio, Cristina
- Gosse, Philippe
- Head, Geoffrey
- Hoshide, Satoshi
- Imai, Yutaka
- Kario, Kazuomi
- Li, Yan
- Manios, Efstathios
- Mant, Jonathan
- McManus, Richard J.
- Mihailidou, Anastasia S.
- Muntner, Paul
- Myers, Martin
- Niiranen, Teemu
- Ntineri, Angeliki
- Ohkubo, Takayoshi
- Omboni, Stefano
- Palatini, Paolo
- Pellegrini, Dario
- Sharman, James E.
- Shennan, Andrew
- Topouchian, Jirar
- Torlasco, Camilla
- Wang, Ji Guang
- Weber, Michael A.
- Whelton, Paul K.
- White, William B.
- Mancia, Giuseppe
- Venue
- Journal of Hypertension
- Published
- 2021-07-26
- Volume
- 39
- Issue
- 9
- Pages
- 1742-1767
- ISSN
- 0263-6352 (ISSN-L)
- 1473-5598
- PMC
- PMC9904446
- OpenAlex
- W3181754241
- Semantic Scholar
- 29e2249cfcdb891204befa8440135890efa4dccc
- Cited by
- 178
- References
- 294
- Citations by year
- 2026: 28 · 2025: 39 · 2024: 27 · 2023: 40 · 2022: 39 · 2021: 4
- FWCI
- 18.42
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Locations
- Journal of Hypertension (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- SERVAL (Université de Lausanne) (publishedVersion, cc-by)
- IRIS (submittedVersion, cc-by)
- Lirias (submittedVersion, other-oa)
- UTAS Research Repository (submittedVersion)
- BOA (University of Milano-Bicocca) (submittedVersion)
- Figshare (submittedVersion, cc-by)
- Institutional Repositories DataBase (IRDB) (submittedVersion)
- Research Portal (King's College London) (submittedVersion, other-oa)
- IUScholarWorks (Indiana University) (submittedVersion, cc-by)
- STM:n Hallinnonalan avoin julkaisuarkisto (Julkari) (submittedVersion)
- WOS
- Funders
- Pfizer
- AstraZeneca
- Sanofi
- Regeneron Pharmaceuticals
- ReCor Medical
- Omron Healthcare
- National Institute for Health and Care Research (NIHR200287, NIHR200201)
- Servier
- Akebia Therapeutics
- Relypsa
- Daiichi Sankyo Europe
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Keywords
- home blood pressure monitoring, hypertension diagnosis, ambulatory blood pressure monitoring, cardiovascular variability, clinical guidelines, European Society of Hypertension
- MeSH terms
- Blood Pressure, Blood Pressure Determination, Humans, Hypertension, Societies, Medical, Blood Pressure Monitoring, Ambulatory
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol and qwen3.8-omni-flash).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.