Validation of the Omron M5-I, R5-I and HEM-907 automated blood pressure monitors in elderly individuals according to the International Protocol of the European Society of Hypertension

Omboni, Stefano, Riva, Isabella, Giglio, Alessia, Caldara, Gianluca, Groppelli, Antonella, Parati, Gianfranco · 2007 · Blood Pressure Monitoring
Cited by 129

The DOI matches its Crossref record; 8 of the 8 references it cites were found in a registry or an index.

Abstract

OBJECTIVE: This study aimed at verifying the accuracy of three automated electronic oscillometric blood pressure measuring devices, namely Omron M5-I (home use upper arm monitor), R5-I (home use wrist monitor) and HEM-907 (professional use upper arm monitor) according to the European Society of Hypertension International Protocol in elderly individuals. METHODS: Sequential measurements of systolic and diastolic blood pressure were obtained in 33 participants (aged >or=75 years) using the mercury sphygmomanometer (two observers) and each of the tested devices (one supervisor). A standard adult cuff was always employed during the study because all participants had an arm circumference compatible with such a cuff. According to the European Society of Hypertension validation protocol 99 couples (three pairs per patient) of test device and reference blood pressure measurements were obtained during phase 1 (15 participants studied) and phase 2 (a further 18 participants) for each electronic monitor. RESULTS: All devices successfully passed the validation study with a mean (+/-SD) device-observer difference for systolic and diastolic blood pressure of 0.2+/-3.6/0.2+/-3.9 mmHg (Omron M5-I), -1.5+/-6.2/-0.7+/-3.7 mmHg (Omron R5-I), and 0.1+/-5.1/-1.9+/-4.2 mmHg (Omron HEM-907). SD of the mean difference was lower and thus the precision was better for diastolic than for systolic blood pressure, and for the Omron M5-I than for the other two devices. CONCLUSIONS: According to the results of the validation study based on the European Society of Hypertension International Protocol the Omron M5-I, R5-I, and HEM-907 may be recommended for clinical use in elderly individuals, without atrial fibrillation or frequent ectopic beats.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/mbp.0b013e32813fa386

Cites

8 of 8 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 8 references

Research works printed but not linked: 24

  1. Hajjar, 2003, Trends in prevalence, awareness, treatment, and control of hypertension in the United States, 1988–2000, JAMA, 290, 199, 10.1001/jama.290.2.199 · DOI · Paper page
  2. Lewington, 2002, Age-specific relevance of usual blood pressure to vascular mortality: a meta-analysis of individual data for one million adults in 61 prospective studies, Lancet, 360, 1903, 10.1016/S0140-6736(02)11911-8 · DOI · Paper page
  3. Campbell, 1994, Pitfalls to avoid in the measurement of blood pressure in the elderly, Can J Public Health, 85, S26
  4. Owens, 1998, Ambulatory blood pressure in the hypertensive population: patterns and prevalence of hypertensive subforms, J Hypertens, 16, 1735, 10.1097/00004872-199816120-00005 · DOI · Paper page
  5. O'Brien, 2002, Has conventional sphygmomanometry ended with the banning of mercury?, Blood Press Monit, 7, 37, 10.1097/00126097-200202000-00007 · doi:10.1097/00126097-200202000-00007
  6. Ohkubo, 1998, Home blood pressure measurement has a stronger predictive power for mortality than does screening blood pressure measurement: a population-based observation in Ohasama, Japan, J Hypertens, 16, 971, 10.1097/00004872-199816070-00010 · DOI · Paper page
  7. Stryker, 2004, Accuracy of home blood pressure readings: monitors and operators, Blood Press Monit, 9, 143, 10.1097/01.mbp.0000130224.18777.ce · DOI · Paper page
  8. White, 1993, National Standard for measurement of resting and ambulatory blood pressures with automated sphygmomanometers, Hypertension, 21, 504, 10.1161/01.HYP.21.4.504 · DOI · Paper page
  9. O'Brien, 1993, Short report: an outline of the revised British Hypertension Society protocol for the evaluation of blood pressure measuring devices, J Hypertens, 11, 677, 10.1097/00004872-199306000-00013 · DOI · Paper page
  10. O'Brien, 2002, Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults, Blood Press Monit, 7, 3, 10.1097/00126097-200202000-00002 · DOI · Paper page
  11. O'Brien, 2001, Blood pressure measuring devices: recommendations of the European Society of Hypertension, BMJ, 322, 531, 10.1136/bmj.322.7285.531 · DOI · Paper page
  12. Ohkubo, 2004, Prediction of ischaemic and haemorrhagic stroke by self-measured blood pressure at home: the Ohasama study, Blood Press Monit, 9, 315, 10.1097/00126097-200412000-00009 · DOI · Paper page
  13. Tsuji, 1997, Proposal of reference values for home blood pressure measurement: prognostic criteria based on a prospective observation of the general population in Ohasama, Japan, Am J Hypertens, 10, 409
  14. Merrick, 1997, Factors influencing the accuracy of home blood pressure measurement, South Med J, 90, 1110, 10.1097/00007611-199711000-00011 · doi:10.1097/00007611-199711000-00011
  15. Campbell, 1999, Is usual measurement of blood pressure meaningful?, Blood Press Monit, 4, 71, 10.1097/00126097-199902000-00003 · DOI
  16. Altunkan, 2006, Validation of the Omron 637IT wrist blood pressure measuring device with a position sensor according to the International Protocol in the elderly, Blood Press Monit, 11, 97, 10.1097/01.mbp.0000200486.95281.5c · DOI
  17. Bortolotto, 1999, Validation of two devices for self-measurement of blood pressure by elderly patients according to the revised British Hypertension Society protocol: the Omron HEM-722C and HEM-735C, Blood Press Monit, 4, 21, 10.1097/00126097-199904010-00004 · DOI · Paper page
  18. El Assaad, 2003, Evaluation of two devices for self-measurement of blood pressure according to the international protocol: the Omron M5-I and the Omron 705IT, Blood Press Monit, 8, 127, 10.1097/00126097-200306000-00006 · DOI · Paper page
  19. White, 2001, Evaluation of the overall efficacy of the Omron office digital blood pressure HEM-907 monitor in adults, Blood Press Monit, 6, 107, 10.1097/00126097-200104000-00007 · DOI · Paper page
  20. El Assaad, 2002, Validation of the Omron HEM-907 device for blood pressure measurement, Blood Press Monit, 7, 237, 10.1097/00126097-200208000-00006 · DOI · Paper page
  21. Go, 2005, The epidemiology of atrial fibrillation in elderly persons: the tip of the iceberg, Am J Geriatr Cardiol, 14, 56, 10.1111/j.1076-7460.2005.02278.x · DOI · Paper page
  22. Ryder, 1999, Epidemiology and significance of atrial fibrillation, Am J Cardiol, 84, 131R, 10.1016/S0002-9149(99)00713-4 · DOI · Paper page
  23. Sudlow, 1998, Prevalence of atrial fibrillation and eligibility for anticoagulants in the community, Lancet, 352, 1167, 10.1016/S0140-6736(98)01401-9 · DOI · Paper page
  24. Parati, 2002, Self blood pressure monitoring at home by wrist devices: a reliable approach?, J Hypertens, 20, 573, 10.1097/00004872-200204000-00005 · DOI · Paper page

Cited by

See all 129

Metadata

Authors
Omboni, S., Riva, I., Giglio, A., Caldara, G., Groppelli, A., & Parati, G.
Venue
Blood Pressure Monitoring
Published
2007-08
Volume
12
Issue
4
Pages
233-242
ISSN
1359-5237 (ISSN-L)
OpenAlex
W1992031528
Cited by
129
References
8
FWCI
4.97
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Validation of the Omron M5-I, R5-I and HEM-907 automated blood pressure monitors in elderly individuals according to the International Protocol of the European Society of Hypertension. https://app.cochat.ai/papers/10.1097/mbp.0b013e32813fa386
The paper
Omboni, S., Riva, I., Giglio, A., Caldara, G., Groppelli, A., & Parati, G. (2007). Validation of the Omron M5-I, R5-I and HEM-907 automated blood pressure monitors in elderly individuals according to the International Protocol of the European Society of Hypertension. Blood Pressure Monitoring. https://doi.org/10.1097/mbp.0b013e32813fa386

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.