Stimulation and Artifact-Suppression Techniques for In Vitro High-Density Microelectrode Array Systems
Open accessAbstract
We present novel voltage stimulation buffers with controlled output current, along with recording circuits featuring adjustable high-pass cut-off filtering to perform efficient stimulation while actively suppressing stimulation artifacts in high-density microelectrode arrays. Owing to the dense packing and close proximity of the electrodes in such systems, a stimulation through one electrode can cause large electrical artifacts on neighboring electrodes that easily saturate the corresponding recording amplifiers. To suppress such artifacts, the high-pass corner frequencies of all available 2048 recording channels can be raised from several Hz to several kHz by applying a "soft-reset" or pole-shifting technique. With the implemented artifact suppression technique, the saturation time of the recording circuits, connected to electrodes in immediate vicinity to the stimulation site, could be reduced to less than 150 μs. For the stimulation buffer, we developed a circuit, which can operate in two modes: either control of only the stimulation voltage or control of current and voltage during stimulation. The voltage-only controlled mode employs a local common-mode feedback operational transconductance amplifier with a near rail-to-rail input/output range, suitable for driving high-capacitive loads. The current/voltage controlled mode is based on a positive current conveyor generating adjustable output currents, whereas its upper and lower output voltages are limited by two feedback loops. The current/voltage controlled circuit can generate stimulation pulses up to 30 μA with less than ±0.1% linearity error in the low-current mode and up to 300 μA with less than ±0.2% linearity error in the high-current mode.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Pole shifting helps to suppress stimulus artifacts by adjusting the amplifier's frequency response with limited switching artifacts that are typically caused by alternative front-end recovery methods.
“This method helps to suppress stimulus artifacts by adjusting the amplifier's frequency response with limited switching artifacts that are typically caused by alternative front-end recovery methods [190], such as amplifier charge reset and active electrode discharge.”
From Active Neural Interface Circuits and Systems for Selective Control of Peripheral Nerves: A Review (2024) · cites this paper as [190]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Neural Stimulation and Recording Electrodes
Cogan, S. F. (2008) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.10.061807.160518 · Cited by 2,243 · see citing works
Electrical stimulation of excitable tissue: design of efficacious and safe protocols
Merrill, D. R., Bikson, M., & Jefferys, J. G. R. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2004.10.020 · Cited by 2,078 · see citing works
Revealing neuronal function through microelectrode array recordings
Obien, M. E. J., Deligkaris, K., Bullmann, T., Bakkum, D. J., & Frey, U. (2015) · Frontiers in Neuroscience · doi:10.3389/fnins.2014.00423 · Cited by 705 · see citing works
ILLINGWORTH, J. A. (1987) · Biochemical Education · doi:10.1016/0307-4412(87)90120-8 · Cited by 368 · see citing works
Berdondini, L., Imfeld, K., Maccione, A., Tedesco, M., Neukom, S., Koudelka‐Hep, M., & Sergio Martinoia, A. (2009) · Lab on a Chip · doi:10.1039/b907394a · Cited by 320 · see citing works
Role of electrode design on the volume of tissue activated during deep brain stimulation
Butson, C. R., & McIntyre, C. C. (2005) · Journal of Neural Engineering · doi:10.1088/1741-2560/3/1/001 · Cited by 312 · see citing works
Effective parameters for stimulation of dissociated cultures using multi-electrode arrays
Wagenaar, D. A., Pine, J., & Potter, S. M. (2004) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2004.03.005 · Cited by 300 · see citing works
CMOS nanoelectrode array for all-electrical intracellular electrophysiological imaging
Abbott, J., Ye, T., Qin, L., Jorgolli, M., Gertner, R. S., Ham, D., & Park, H. (2017) · Nature Nanotechnology · doi:10.1038/nnano.2017.3 · Cited by 267 · see citing works
Switch-Matrix-Based High-Density Microelectrode Array in CMOS Technology
Frey, U., Šedivý, J., Heer, F., Pedron, R., Ballini, M., Müller, J., Bakkum, D. J., Hafizovic, S., Faraci, F. D., Greve, F., et al. (2010) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2009.2035196 · Cited by 251 · see citing works
Self-citationBallini, M., Müller, J., Livi, P., Chen, Y., Frey, U., Stettler, A., Shadmani, A., Viswam, V., Jones, I. L., Jäckel, D., et al. (2014) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2014.2359219 · Cited by 236 · see citing works
Self-citationDesign and Development of Medical Electronic Instrumentation
Prutchi, D., & Norris, M. (2004) · Wiley eBooks · doi:10.1002/0471681849 · Cited by 216 · see citing works
Noorsal, E., Sooksood, K., Xu, H., Hornig, R., Becker, J., & Ortmanns, M. (2012) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2011.2164667 · Cited by 212 · see citing works
Dragas, J., Viswam, V., Shadmani, A., Chen, Y., Bounik, R., Stettler, A., Radivojević, M., Geissler, S. A., Obien, M. E. J., Müller, J., et al. (2017) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2017.2686580 · Cited by 207 · see citing works
Self-citationAn active approach for charge balancing in functional electrical stimulation.
Sooksood, K., Stieglitz, T., & Ortmanns, M. (2010) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2010.2040277 · Cited by 170 · see citing works
Sit, J.-J., & Sarpeshkar, R. (2007) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2007.911631 · Cited by 155 · see citing works
Hierlemann, A. R., Frey, U., Hafizovic, S., & Heer, F. (2010) · Proceedings of the IEEE · doi:10.1109/jproc.2010.2066532 · Cited by 152 · see citing works
Self-citationGhovanloo, M., & Najafi, K. (2005) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2004.839797 · Cited by 137 · see citing works
Lempka, S. F., Johnson, M. D., Miocinovic, S., Vitek, J. L., & McIntyre, C. C. (2010) · Clinical Neurophysiology · doi:10.1016/j.clinph.2010.04.026 · Cited by 135 · see citing works
Super Class-AB OTAs With Adaptive Biasing and Dynamic Output Current Scaling
Galán, J., López‐Martín, A. J., Carvajal, R. G., Ramrez-Angulo, J., & Rubia-Marcos, C. (2007) · IEEE Transactions on Circuits and Systems I Fundamental Theory and Applications · doi:10.1109/tcsi.2006.887639 · Cited by 118 · see citing works
Stimulus-artifact elimination in a multi-electrode system.
Brown, E. A., Ross, J. D., Blum, R. A., Nam, Y., Wheeler, B. C., & DeWeerth, S. P. (2008) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2008.918285 · Cited by 95 · see citing works
Joucla, S., & Yvert, B. (2011) · Journal of Physiology-Paris · doi:10.1016/j.jphysparis.2011.10.003 · Cited by 88 · see citing works
Single-chip microelectronic system to interface with living cells
Heer, F., Hafizovic, S., Ugniwenko, T., Frey, U., Franks, W., Perriard, E. R., Perriard, J., Blau, A. W., Ziegler, C., & Hierlemann, A. R. (2006) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2006.10.003 · Cited by 83 · see citing works
Self-citationAn integrated system for simultaneous, multichannel neuronal stimulation and recording.
Blum, R. A., Ross, J. D., Brown, E. A., & DeWeerth, S. P. (2007) · IEEE Transactions on Circuits and Systems I Regular Papers · doi:10.1109/tcsi.2007.906071 · Cited by 74 · see citing works
Activation of ganglion cells and axon bundles using epiretinal electrical stimulation
Grosberg, L. E., Ganesan, K., Goetz, G. A., Madugula, S. S., Bhaskhar, N., Fan, V., Li, P., Hottowy, P., Dabrowski, W., Sher, A., et al. (2017) · Journal of Neurophysiology · doi:10.1152/jn.00750.2016 · Cited by 70 · see citing works
Hottowy, P., Skoczeń, A., Gunning, D. E., Kachiguine, S., Mathieson, K., Sher, A., Wiącek, P., Litke, A. M., & Dąbrowski, W. (2012) · Journal of Neural Engineering · doi:10.1088/1741-2560/9/6/066005 · Cited by 67 · see citing works
Bertotti, G., Velychko, D., Dodel, N., Keil, S., Wolansky, D., Tillak, B., Schreiter, M., Grall, A., Jesinger, P., Rohler, S., et al. (2014) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas.2014.6981723 · Cited by 64 · see citing works
Radivojević, M., Jäckel, D., Altermatt, M., Müller, J., Viswam, V., Hierlemann, A. R., & Bakkum, D. J. (2016) · Scientific Reports · doi:10.1038/srep31332 · Cited by 57 · see citing works
Self-citationElectrical stimulus artifact cancellation and neural spike detection on large multi-electrode arrays
Mena, G. E., Grosberg, L. E., Madugula, S. S., Hottowy, P., Litke, A. M., Cunningham, J. P., Chichilnisky, E. J., & Paninski, L. (2017) · PLoS Computational Biology · doi:10.1371/journal.pcbi.1005842 · Cited by 51 · see citing works
Extreme Low-Power Mixed Signal IC Design: Subthreshold Source-Coupled Circuits
Tajalli, A., & Leblebici, Y. (2010) · Springer eBooks · doi:10.1007/978-1-4419-6478-6 · Cited by 49 · see citing works
Spectral cancellation of microstimulation artifact for simultaneous neural recording in situ
Gnadt, J., Echols, S. D., Yildirim, A., Zhang, H., & Paul, K. (2003) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2003.816077 · Cited by 45 · see citing works
Viswam, V., Dragas, J., Shadmani, A., Chen, Y., Stettler, A., Müller, J., & Hierlemann, A. R. (2016) · Digest of technical papers/Digest of technical papers - IEEE International Solid-State Circuits Conference · doi:10.1109/isscc.2016.7418073 · Cited by 45 · see citing works
Self-citationA Matching Technique for Biphasic Stimulation Pulse
Lee, E. K. F., & Lam, A. (2007) · IEEE International Symposium on Circuits and Systems proceedings · doi:10.1109/iscas.2007.378031 · Cited by 41 · see citing works
Eversmann, B., Lambacher, A., Gerling, T., Kunze, A., Fromherz, P., & Thewes, R. (2011) · European Solid-State Circuits Conference (ESSCIRC) · doi:10.1109/esscirc.2011.6044902 · Cited by 40 · see citing works
Charge Balancing in Functional Electrical Stimulators: A Comparative Study
Ortmanns, M. (2007) · IEEE International Symposium on Circuits and Systems proceedings · doi:10.1109/iscas.2007.378802 · Cited by 39 · see citing works
Compact Voltage and Current Stimulation Buffer for High-Density Microelectrode Arrays
Livi, P., Heer, F., Frey, U., Bakkum, D. J., & Hierlemann, A. R. (2010) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2010.2080676 · Cited by 29 · see citing works
Self-citationStimulus-resistant neural recording amplifier
DeMichele, G. A., & Troyk, P. R. (2004) · doi:10.1109/iembs.2003.1280857 · Cited by 28 · see citing works
Recent advances in charge balancing for functional electrical stimulation
Sooksood, K., Stieglitz, T., & Ortmanns, M. (2009) · 2009 Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2009.5333181 · Cited by 20 · see citing works
2048 action potential recording channels with 2.4 μVrms noise and stimulation artifact suppression
Viswam, V., Chen, Y., Shadmani, A., Dragas, J., Bounik, R., Radivojević, M., Müller, J., & Hierlemann, A. R. (2016) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas.2016.7833750 · Cited by 17 · see citing works
Self-citationPark, J. H., Chi, T., Su, A. M., Zhu, C., Sung, J. H., Cho, H. C., Styczynski, M. P., & Wang, H. (2016) · Symposium on VLSI Circuits · doi:10.1109/vlsic.2016.7573469 · Cited by 16 · see citing works
Design of Efficient and Safe Neural Stimulators
van Dongen, M., & Serdijn, W. A. (2016) · Analog circuits and signal processing/Analog circuits and signal processing series · doi:10.1007/978-3-319-28131-5 · Cited by 15 · see citing works
A Programmable High-Voltage Compliance Neural Stimulator for Deep Brain Stimulation in Vivo
Gong, C. A., Lai, H., Huang, S.-H., Lo, Y., Lee, N. M., Chen, P., Tu, P., Yang, C., Lin, J. C., & Chen, Y. (2015) · Sensors · doi:10.3390/s150612700 · Cited by 11 · see citing works
Self-citationCompact voltage and current stimulation buffer for high-density microelectrode arrays
Livi, P., Heer, F., Frey, U., Bakkum, D. J., & Hierlemann, A. R. (2010) · Digest of technical papers/Digest of technical papers - IEEE International Solid-State Circuits Conference · doi:10.1109/isscc.2010.5433935 · Cited by 10 · see citing works
Self-citationDesign of efficient and safe neural stimulators: a multidisciplinary approach
van Dongen, M. (2015) · CERN Document Server (European Organization for Nuclear Research) · doi:10.4233/uuid:9c215b15-552d-497b-9e1a-83a77cde0a10 · Cited by 9 · see citing works
Chun, H., Tran, N., Yang, Y., Kavehei, O., Bai, S., & Skafidas, E. S. (2012) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2012.6346595 · Cited by 7 · see citing works
Jung, H., Kim, J., & Nam, Y. (2018) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2018.04.014 · Cited by 3 · see citing works
Flagged first, then by citations
2 of the 47 references couldn’t be matched to a known paper.
Research works printed but not linked: 1
- [40]Tajalli A Leblebici Y Extreme Low-Power Mixed Signal IC Design Springer New York 2010
Cited by
Sorted by citations
Wang, Y., Luo, H., Chen, Y., Jiao, Z., Sun, Q., Dong, L., Chen, X., Wang, X., & Zhang, H. (2021) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2021.3102261 · Cited by 63 · see citing works
Self-citationDidier, C. M., Kundu, A., DeRoo, D., & Rajaraman, S. (2020) · Journal of Micromechanics and Microengineering · doi:10.1088/1361-6439/ab8e91 · Cited by 52 · see citing works
Advances in large-scale electrophysiology with high-density microelectrode arrays
Schröter, M. S., Cardes, F., Bui, C.-V. H., Dodi, L. D., Gänswein, T., Bartram, J., Sadiraj, L., Hornauer, P., Kumar, S. S., Pascual-Garcia, M., et al. (2025) · Lab on a Chip · doi:10.1039/d5lc00058k · Cited by 43 · see citing works
Self-citationWang, H., Bi, Z.-Y., Fan, W.-J., Zhou, Y., Zhou, Y., Li, F., Wang, K., Lü, X., & Wang, Z. (2021) · IEEE Access · doi:10.1109/access.2021.3077644 · Cited by 16 · see citing works
A Bootstrapped Pseudo Resistor by Reusing OTA-C Filter for Neural Signal Processing
Zhou, Z., Liao, H., Zhu, L., Liu, Y., Zhu, L., Wang, K., & Wang, Z. (2021) · IEEE Transactions on Circuits & Systems II Express Briefs · doi:10.1109/tcsii.2021.3102249 · Cited by 14 · see citing works
Active Neural Interface Circuits and Systems for Selective Control of Peripheral Nerves: A Review
Habibollahi, M., Jiang, D., Lancashire, H. T., & Demosthenous, A. (2024) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2024.3430038 · Cited by 11 · see citing works
“Alteration of the poles is often achieved using a variable pseudo-resistor in the amplifier’s feedback path, which determines its high-pass characteristics [190].”
Anti-artifacts techniques for neural recording front-ends in closed-loop brain-machine interface ICs
Chen, W., Liu, X., Wan, P., Chen, Z., & Chen, Y. (2024) · Frontiers in Neuroscience · doi:10.3389/fnins.2024.1393206 · Cited by 11 · see citing works
Self-citationCarlino, M. F., & Gielen, G. E. (2023) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2023.3344889 · Cited by 10 · see citing works
Erbslöh, A., Viga, R., Seidl, K., & Kokozinski, R. (2021) · IEEE Sensors Journal · doi:10.1109/jsen.2021.3133716 · Cited by 10 · see citing works
Narayanan, R., Khaleghi, A., Veletić, M., & Balasingham, I. (2024) · PLoS ONE · doi:10.1371/journal.pone.0297114 · Cited by 8 · see citing works
A closed-loop neural recording analog front-end with low noise and automatic gain control amplifiers
Mohtashamnia, M., & Yavari, M. (2025) · AEU - International Journal of Electronics and Communications · doi:10.1016/j.aeue.2025.155976 · Cited by 8 · see citing works
Gencturk, E., Ülgen, K. Ö., & Mutlu, S. (2020) · Biomicrofluidics · doi:10.1063/5.0008462 · Cited by 8 · see citing works
Lee, T., & Je, M. (2020) · IEEE Transactions on Circuits & Systems II Express Briefs · doi:10.1109/tcsii.2020.3046451 · Cited by 7 · see citing works
An Active Microchannel Neural Interface with Artifact Reduction
Habibollahi, M., Hanzaee, F. F., Jiang, D., Lancashire, H., & Demosthenous, A. (2021) · 2021 IEEE International Symposium on Circuits and Systems (ISCAS) · doi:10.1109/iscas51556.2021.9401318 · Cited by 5 · see citing works
Programmable Electrochemical Stimulation on a Large-Scale CMOS Microelectrode Array
Joshi, P., Hu, K., Larkin, J., & Rosenstein, J. K. (2022) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas54905.2022.9948674 · Cited by 4 · see citing works
Tarotin, I., Mastitskaya, S., Ravagli, E., PERKINS, J. D., Holder, D. S., & Aristovich, K. Y. (2022) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac669a · Cited by 4 · see citing works
Virtual white matter: a novel system for cross-dish neural interaction and modulation
Khantan, M., Lim, J., Napoli, A., Obeid, I., & Serruya, M. D. (2025) · Journal of Neural Engineering · doi:10.1088/1741-2552/add49c · Cited by 4 · see citing works
Sato, Y., Yamamoto, H., Ishikawa, Y., Sumi, T., Sono, Y., Sato, S., Katori, Y., & Hirano‐Iwata, A. (2024) · Japanese Journal of Applied Physics · doi:10.35848/1347-4065/ad7ec1 · Cited by 3 · see citing works
A Novel Stimulus Artifact Suppression System With Fast Template Subtraction
Liu, Y., Chang, Q., Li, X., & Liu, X. (2025) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2025.3591110 · Cited by 2 · see citing works
A Charge Balanced Neural Stimulator Silicon Chip for Human-Machine Interface
Liu, X., Li, J., Mao, W., Chen, Z., Chen, Z., Wan, P., & Yu, H. (2021) · Frontiers in Electronics · doi:10.3389/felec.2021.773812 · Cited by 2 · see citing works
Erbslöh, A., Viga, R., Seidl, K., & Kokozinski, R. (2020) · Proceedings of IEEE Sensors/Proceedings of IEEE Sensors ... · doi:10.1109/sensors47125.2020.9278607 · Cited by 1 · see citing works
Lin, Y., Huang, N.-H., Chou, Y., Chen, B.-C., Lin, L.-J., Hsu, W., & Lee, C.-F. (2025) · Otology & Neurotology · doi:10.1097/mao.0000000000004406
Seseri, N., Nigrisoli, D., Faraci, F., D’Angelo, A., Freddi, R., Chandran, S., Barbieri, R., Corti, S. P., Ottoboni, L., & Russo, S. (2025) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2025.09.01.673532
Artifact Cancellation in Neurostimulators: A Review
Das, V. R., Karşilayan, A. İ., Silva-Martínez, J., Genzer, D., & Stotts, L. J. (2025) · IEEE International Midwest Symposium on Circuits and Systems (MWSCAS) · doi:10.1109/mwscas53549.2025.11244388
- Authors
- Shadmani, A., Viswam, V., Chen, Y., Bounik, R., Dragas, J., Radivojevic, M., Geissler, S., Sitnikov, S., Muller, J., & Hierlemann, A.
- Affiliations
- ETH Zurich (CH)
- ORCID
- Venue
- IEEE Transactions on Biomedical Engineering
- Published
- 2019-09
- Volume
- 66
- Issue
- 9
- Pages
- 2481-2490
- ISSN
- 0018-9294 (ISSN-L)
- 1558-2531
- PMC
- PMC6711758
- OpenAlex
- W2907874759
- Cited by
- 24
- References
- 47
- Citations by year
- 2025: 7 · 2024: 4 · 2023: 1 · 2022: 2 · 2021: 6 · 2020: 4
- FWCI
- 1.06
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- Locations
- IEEE Transactions on Biomedical Engineering (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- ERC Advanced Grants (NeuroCMOS / AdG 267351, neuroXscales / AdG 694829)
- FP7-MTN (EngCaBra / 264417)
- Schweizerischer Nationalfonds zur Furderung der Wissenschaftlichen Forschung (205321_157092/1, MHV-Grant 171267)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Analog and Mixed-Signal Circuit Design
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Advanced MEMS and NEMS Technologies
- Keywords
- high-density microelectrode arrays, electrical stimulation, stimulation artifacts, artifact removal, high-pass filtering, pole shifting, operational transconductance amplifier, current conveyor
- MeSH terms
- Superconductivity, Animals, Electric Stimulation, Electrophysiology, Equipment Design, Microelectrodes, Neurons, Signal Processing, Computer-Assisted, Artifacts, Rats, Wistar, Rats
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review