Pulse transit time measured from the ECG: an unreliable marker of beat-to-beat blood pressure
Abstract
The arterial pulse-wave transit time can be measured between the ECG R-wave and the finger pulse ( rPTT), and has been shown previously to have a linear correlation with blood pressure (BP). We hypothesized that the relationship between rPTT, preejection period (PEP; the R-wave/mechanical cardiac delay), and BP would vary with different vasoactive drugs. Twelve healthy men (mean age 22 yr) were studied. Beat-to-beat measurements were made of rPTT (using ECG and photoplethysmograph finger probe), intra-arterial radial pressure, PEP (using cardiac bioimpedance), and transit time minus PEP ( pPTT). Four drugs (glyceryl trinitrate, angiotensin II, norepinephrine, salbutamol) were administered intravenously over 15 min, with stepped dosage increase every 5 min and a 25-min saline washout between agents. All subjects in all conditions had a negative linear correlation ( R 2 = 0.39) between rPTT and systolic BP (SBP), generally constant between different drugs, apart from four subjects who had a positive rPTT/SBP correlation with salbutamol. The 95% limits of agreement between measured and rPTT-predicted SBP were ±17.0 mmHg. Beat-to-beat variability of rPTT showed better coherence with SBP variability than it did with heart rate variability ( P < 0.001). PEP accounted for a substantial and variable proportion of rPTT (12–35%). Diastolic (DBP) and mean arterial BP (MAP) correlated poorly with rPTT ( R 2 = 0.02 and 0.08, respectively) but better with pPTT ( rPTT corrected for PEP, R 2 = 0.41 and 0.45, respectively). The 95% limits of agreement between measured and pPTT-predicted DBP were ± 17.3 mmHg. In conclusion, the negative correlation between rPTT and SBP is generally constant, even with marked hemodynamic perturbations. However, the relationship is not reliable enough for rPTT to be used as a surrogate marker of SBP, although it may be useful in assessing BP variability. DBP and MAP cannot be predicted from rPTT without correction for PEP. The significant contribution of PEP to rPTT means that rPTT should not be used as a marker of purely vascular function.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“The results for HR and SpO2 are within the regulatory guidelines set by the US Food and Drug Administration (FDA), which require errors less than ± 10 % or ± 5 bpm for HR and less than 3.5 % for Arms for reflectance mode SpO226,27.”
From Skin-interfaced biosensors for advanced wireless physiological monitoring in neonatal and pediatric intensive-care units (2020) · cites this paper as [27]
Citations
Cites
McDonald's blood flow in arteries; theoretical, experimental and clinical principles
(1992) · General Pharmacology The Vascular System · doi:10.1016/0306-3623(92)90079-y · Cited by 2,395
Parati, G., Saul, J. P., Di Rienzo, M., & Mancia, G. (1995) · Hypertension · doi:10.1161/01.hyp.25.6.1276 · Cited by 849
Dave, K. A., & White, S. W. (2002) · World Highways/Routes du Monde · Cited by 767
The revision of the Declaration of Helsinki: past, present and future
Carlson, R. V., Boyd, K., & Webb, D. J. (2004) · British Journal of Clinical Pharmacology · doi:10.1111/j.1365-2125.2004.02103.x · Cited by 723
Self-citationPhotoelectric measurement of blood pressure, volume and flow in the finger
Peňáz, J. (1973) · Medical Entomology and Zoology · Cited by 528
Brien, E. O., Petrie, J. C., Little, W. C., de Swiet, M., Padfield, P. L., Altma, D. G., Bland, M., Coats, A. J. S., & Atkins, N. (1993) · Journal of Hypertension · doi:10.1097/00004872-199306000-00013 · Cited by 518
Chen, W., Kobayashi, T., Ichikawa, S., Takeuchi, Y., & Togawa, T. (2000) · Medical & Biological Engineering & Computing · doi:10.1007/bf02345755 · Cited by 425
Lantelme, P., Mestre, C., Lièvre, M. M., Gressard, A., & Milon, H. (2002) · Hypertension · doi:10.1161/01.hyp.0000019132.41066.95 · Cited by 353
Arterial baroreflex function and cardiovascular variability: interactions and implications
Lanfranchi, P. A., & Somers, V. K. (2002) · American Journal of Physiology-Regulatory, Integrative and Comparative Physiology · doi:10.1152/ajpregu.00051.2002 · Cited by 340
Determination of pulse wave velocities with computerized algorithms
Chiu, Y. C., Arand, P. W., Shroff, S. G., Feldman, T. E., & Carroll, J. D. (1991) · American Heart Journal · doi:10.1016/0002-8703(91)90153-9 · Cited by 313
Flagged first, then by citations
1 of the 41 references couldn’t be matched to a known paper.
Cited by
Sorted by citations
Photoplethysmography and its application in clinical physiological measurement
Allen, J. (2007) · Physiological Measurement · doi:10.1088/0967-3334/28/3/r01 · Cited by 3,967
Knowing your own heart: Distinguishing interoceptive accuracy from interoceptive awareness
Garfinkel, S. N., Seth, A. K., Barrett, A. B., Suzuki, K., & Critchley, H. D. (2015) · Biological Psychology · doi:10.1016/j.biopsycho.2014.11.004 · Cited by 1,624
Toward ubiquitous blood pressure monitoring via pulse transit time: Theory and practice
Mukkamala, R., Hahn, J.-O., Inan, O. T., Mestha, L. K., Kim, C.-S., Toreyin, H., & Kyal, S. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2441951 · Cited by 935
“1), healthy human [98], and inpatient [134], respectively.”
BackgroundMethodInfluentialUnobtrusive Sensing and Wearable Devices for Health Informatics
Zheng, Y.-L., Ding, X., Poon, C. C. Y., Lo, B. P. L., Zhang, H., Zhou, X.-L., Yang, G., Zhao, N., & Zhang, Y. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2309951 · Cited by 725
Cuffless Blood Pressure Estimation Algorithms for Continuous Health-Care Monitoring
Kachuee, M., Kiani, M. M., Mohammadzade, H., & Shabany, M. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2016.2580904 · Cited by 621
Discrepancies between dimensions of interoception in autism: implications for emotion and anxiety
Garfinkel, S. N., Tiley, C., O'Keeffe, S., Harrison, N. A., Seth, A. K., & Critchley, H. D. (2016) · Biological Psychology · doi:10.1016/j.biopsycho.2015.12.003 · Cited by 512
Chung, H. U., Rwei, A. Y., Hourlier-Fargette, A., Xu, S., Lee, K., Dunne, E. C., Xie, Z., Liu, C., Carlini, A., Kim, D. H., et al. (2020) · Nature Medicine · doi:10.1038/s41591-020-0792-9 · Cited by 482
Gesche, H., Grosskurth, D., Küchler, G., & Patzak, A. R. (2011) · European Journal of Applied Physiology · doi:10.1007/s00421-011-1983-3 · Cited by 472
Utility of the Photoplethysmogram in Circulatory Monitoring
Reisner, A. T., Shaltis, P., McCombie, D. B., Asada, H. H., Warner, D. S., & Warner, M. A. A. (2008) · Anesthesiology · doi:10.1097/aln.0b013e31816c89e1 · Cited by 365
Interoceptive dimensions across cardiac and respiratory axes
Garfinkel, S. N., Manassei, M. F., Hamilton-Fletcher, G., In den Bosch, Y., Critchley, H. D., & Engels, M. (2016) · Philosophical Transactions of the Royal Society B: Biological Sciences · doi:10.1098/rstb.2016.0014 · Cited by 326
Optical blood pressure estimation with photoplethysmography and FFT-based neural networks
Xing, X., & Sun, M. (2016) · Biomedical Optics Express · doi:10.1364/boe.7.003007 · Cited by 263
Cuff-Less and Continuous Blood Pressure Monitoring: A Methodological Review
Sharma, M., Barbosa, K., Ho, V., Griggs, D., Ghirmai, T., Krishnan, S., Hsiai, T., Chiao, J.-C., & Cao, H. (2017) · Technologies · doi:10.3390/technologies5020021 · Cited by 226
Adaptive threshold method for the peak detection of photoplethysmographic waveform
Shin, H. S., Lee, C., & Lee, M. (2009) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2009.10.006 · Cited by 211
Non-constrained Blood Pressure Monitoring Using ECG and PPG for Personal Healthcare
Yoon, Y.-Z., Cho, J. H., & Yoon, G. (2008) · Journal of Medical Systems · doi:10.1007/s10916-008-9186-0 · Cited by 199
Oscillometric Blood Pressure Estimation: Past, Present, and Future
Forouzanfar, M., Dajani, H. R., Groza, V. Z., Bolic, M., Rajan, S., & Batkin, I. (2015) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2015.2434215 · Cited by 196
Chen, Z., Zheng, T., Cai, C., & Luo, J. (2021) · ACM Annual International Conference on Mobile Computing and Networking (MobiCom) · doi:10.1145/3447993.3483251 · Cited by 195
Huynh, T. H., Jafari, R., & Chung, W. (2018) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2018.2865751 · Cited by 181
Cuffless Blood Pressure Measurement
Mukkamala, R., Stergiou, G. S., & Avolio, A. P. (2022) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-110220-014644 · Cited by 180
Ding, X., Zhao, N., Yang, G., Pettigrew, R. I., Lo, B. P. L., Miao, F., Li, Y., Liu, J., & Zhang, Y. (2016) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2016.2620995 · Cited by 184
An Armband Wearable Device for Overnight and Cuff-Less Blood Pressure Measurement
Zheng, Y.-L., Yan, B. P., Zhang, Y., & Poon, C. C. Y. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2318779 · Cited by 176
A Novel Continuous Blood Pressure Estimation Approach Based on Data Mining Techniques
Miao, F., Fu, N., Zhang, Y., Ding, X., Hong, X., He, Q., & Li, Y. (2017) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2017.2691715 · Cited by 176
Cattivelli, F. S., & Garudadri, H. (2009) · Wearable and Implantable Body Sensor Networks · doi:10.1109/bsn.2009.35 · Cited by 172
Non-Invasive Blood Pressure Estimation from ECG Using Machine Learning Techniques
Simjanoska, M., Gjoreski, M., Gams, M., & Bogdanova, A. M. (2018) · Sensors · doi:10.3390/s18041160 · Cited by 173
Local Pulse Wave Velocity: Theory, Methods, Advancements, and Clinical Applications
Nabeel, P. M., Kiran, V. R., Joseph, J., Abhidev, V. V., & Sivaprakasam, M. (2019) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2019.2931587 · Cited by 158
Kim, C., Carek, A. M., Mukkamala, R., Inan, O. T., & Hahn, J. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2440291 · Cited by 155
El-Hajj, C., & Kyriacou, P. A. (2020) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2020.102301 · Cited by 156
Shin, H., & Min, S. D. (2017) · BioMedical Engineering OnLine · doi:10.1186/s12938-016-0302-y · Cited by 145
A survey on signals and systems in ambulatory blood pressure monitoring using pulse transit time
Buxi, D., Redouté, J., & Yuce, M. R. (2015) · Physiological Measurement · doi:10.1088/0967-3334/36/3/r1 · Cited by 140
Multi-Sensor Fusion Approach for Cuff-Less Blood Pressure Measurement
Miao, F., Liu, Z.-D., Liu, J., Wen, B., He, Q., & Li, Y. (2019) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2019.2901724 · Cited by 137
Monitoring vital signs over multiplexed radio by near-field coherent sensing
Hui, X., & Kan, E. C.-C. (2017) · Nature Electronics · doi:10.1038/s41928-017-0001-0 · Cited by 140
Ding, X., & Zhang, Y. (2019) · Biomedical Engineering Letters · doi:10.1007/s13534-019-00096-x · Cited by 142
Pulse arrival time is not an adequate surrogate for pulse transit time as a marker of blood pressure
Zhang, G., Gao, M., Xu, D. Z., Olivier, N. B., & Mukkamala, R. (2011) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00980.2011 · Cited by 133
Zhang, Q., Zhou, D., & Zeng, X. (2017) · BioMedical Engineering OnLine · doi:10.1186/s12938-017-0317-z · Cited by 132
Heathers, J. A. J. (2013) · International Journal of Psychophysiology · doi:10.1016/j.ijpsycho.2013.05.017 · Cited by 122
Development of Nanosensors Based Intelligent Packaging Systems: Food Quality and Medicine
Chelliah, R., Wei, S., Daliri, E. B., Rubab, M., Elahi, F., Yeon, S., Jo, K. H., Yan, P., Liu, S., & Oh, D. H. (2021) · Nanomaterials · doi:10.3390/nano11061515 · Cited by 122
Zhang, Q., Zeng, X., Hu, W. W., & Zhou, D. (2017) · IEEE Access · doi:10.1109/access.2017.2707472 · Cited by 113
PPG Sensor Contact Pressure Should Be Taken Into Account for Cuff-Less Blood Pressure Measurement
Chandrasekhar, A., Yavarimanesh, M., Natarajan, K., Hahn, J., & Mukkamala, R. (2020) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2020.2976989 · Cited by 123
Hart, N., McGowan, J. F., Minati, L., & Critchley, H. (2012) · Journal of Personality Disorders · doi:10.1521/pedi_2012_26_049 · Cited by 119
Aguirre, N. A., Grall‐Maës, E., Cymberknop, L. J., & Armentano, R. L. (2021) · Sensors · doi:10.3390/s21062167 · Cited by 110
An Ear-Worn Vital Signs Monitor
He, D. D., Winokur, E. S., & Sodini, C. G. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2459061 · Cited by 109
Liu, Q., Yan, B. P., Yu, C., Zhang, Y., & Poon, C. C. Y. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2286998 · Cited by 109
Conventional pulse transit times as markers of blood pressure changes in humans
Block, R., Yavarimanesh, M., Natarajan, K., Carek, A. M., Mousavi, A. S., Chandrasekhar, A., Kim, C., Zhu, J., Schifitto, G., Mestha, L. K., et al. (2020) · Scientific Reports · doi:10.1038/s41598-020-73143-8 · Cited by 107
Holz, C., & Wang, E. J. (2017) · Proceedings of the ACM on Interactive Mobile Wearable and Ubiquitous Technologies · doi:10.1145/3132024 · Cited by 106
Palser, E. R., Fotopoulou, A., Pellicano, E., & Kilner, J. M. (2018) · Biological Psychology · doi:10.1016/j.biopsycho.2018.05.003 · Cited by 104
Kim, C., Carek, A. M., Inan, O. T., Mukkamala, R., & Hahn, J. (2018) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2018.2797239 · Cited by 110
Electrocardiogram-Assisted Blood Pressure Estimation
Ahmad, S. S., Chen, S., Soueidan, K., Batkin, I., Bolić, M., Dajani, H. R., & Groza, V. F. (2012) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2011.2180019 · Cited by 104
Carek, A. M., Conant, J., Joshi, A. N., Kang, H., & Inan, O. T. (2017) · Proceedings of the ACM on Interactive Mobile Wearable and Ubiquitous Technologies · doi:10.1145/3130905 · Cited by 103
Cuffless blood pressure estimation using only a smartphone
Matsumura, K., Rolfe, P. J., Toda, S., & Yamakoshi, T. (2018) · Scientific Reports · doi:10.1038/s41598-018-25681-5 · Cited by 102
A Non-Invasive Continuous Blood Pressure Estimation Approach Based on Machine Learning
Chen, S., Ji, Z., Wu, H., & Xu, Y. (2019) · Sensors · doi:10.3390/s19112585 · Cited by 100
MallorquÃ-Bagué, N., Garfinkel, S. N., Engels, M., Eccles, J. A., Pailhez, G., Bulbena, A., & Critchley, H. D. (2014) · Frontiers in Psychology · doi:10.3389/fpsyg.2014.01162 · Cited by 97
Chun, K., Seo, S., & Han, C. (2022) · Advanced Materials · doi:10.1002/adma.202110082 · Cited by 97
Pulse arrival time as a surrogate of blood pressure
Finnegan, E., Davidson, S. M., Harford, M., Jorge, J., Watkinson, P., Young, D., Tarassenko, L., & Villarroel, M. (2021) · Scientific Reports · doi:10.1038/s41598-021-01358-4 · Cited by 105
Cuff-Less Blood Pressure Estimation Using Pulse Waveform Analysis and Pulse Arrival Time
Yoon, Y.-Z., Kang, J. M., Kwon, Y., Park, S., Noh, S., Kim, Y., Park, J., & Hwang, S. W. (2017) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2017.2714674 · Cited by 90
Ewing, D. L., Manassei, M., Gould van Praag, C., Philippides, A. O., Critchley, H. D., & Garfinkel, S. N. (2017) · Biological Psychology · doi:10.1016/j.biopsycho.2017.05.011 · Cited by 86
Ibrahim, B., & Jafari, R. (2022) · Scientific Reports · doi:10.1038/s41598-021-03612-1 · Cited by 86
Barvik, D., Černý, M., Penhaker, M., & Noury, N. (2021) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2021.3109643 · Cited by 85
Bi-Modal Arterial Compliance Probe for Calibration-Free Cuffless Blood Pressure Estimation
Nabeel, P. M., Joseph, J., Srinivasa, K., Sivaprakasam, M., & Chenniappan, M. (2018) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2018.2866332 · Cited by 83
Finnegan, E., Davidson, S. M., Harford, M., Watkinson, P., Tarassenko, L., & Villarroel, M. (2023) · Scientific Reports · doi:10.1038/s41598-022-27170-2 · Cited by 87
Pulse transit time as an indirect marker for variations in cardiovascular related reactivity
Foo, J. Y. A., & Lim, C. S. (2006) · Technology and Health Care · doi:10.3233/thc-2006-14205 · Cited by 83
Zheng, Y., Poon, C. C. Y., Yan, B. P., & Lau, J. Y. W. (2016) · Journal of Medical Systems · doi:10.1007/s10916-016-0558-6 · Cited by 82
Maqsood, S., Xu, S., Springer, M., & Mohawesh, R. (2021) · IEEE Access · doi:10.1109/access.2021.3117969 · Cited by 84
Pulse Transit Time and Blood Pressure During Cardiopulmonary Exercise Tests
Wibmer, T., DOERING, K., Kropf-Sanchen, C., RÜDIGER, S., Blanta, I., Stoiber, K., Rottbauer, W., & SCHUMANN, C. (2014) · Physiological Research · doi:10.33549/physiolres.932581 · Cited by 81
Learning Your Heart Actions From Pulse: ECG Waveform Reconstruction From PPG
Zhu, Q., Tian, X., Wong, C.-W., & Wu, M. H. (2021) · IEEE Internet of Things Journal · doi:10.1109/jiot.2021.3097946 · Cited by 85
Lin, W.-H., Wang, H., Samuel, O. W., Liu, G., Huang, Z., & Li, G. (2018) · Physiological Measurement · doi:10.1088/1361-6579/aaa454 · Cited by 78
Wearable Millimeter-Wave Device for Contactless Measurement of Arterial Pulses
Johnson, J. E., Shay, O., Kim, C., & Liao, C. (2019) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2019.2948581 · Cited by 78
Feasibility of cuff-free measurement of systolic and diastolic arterial blood pressure
Masè, M., Mattei, W., Cucino, R., Faes, L., & Nollo, G. (2011) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2010.11.019 · Cited by 78
Systolic blood pressure estimation using PPG and ECG during physical exercise
Sun, S., Bezemer, R., Long, X., Muehlsteff, J., & Aarts, R. M. (2016) · Physiological Measurement · doi:10.1088/0967-3334/37/12/2154 · Cited by 78
Towards Using Photo-Plethysmogram Amplitude to Measure Blood Pressure During Sleep
Chua, E. C.-P., Redmond, S. J., McDarby, G., & Heneghan, C. (2010) · Annals of Biomedical Engineering · doi:10.1007/s10439-009-9882-z · Cited by 77
Validating cuffless continuous blood pressure monitoring devices
Hu, J., Martin, G., Iyengar, S., Kovell, L. C., Plante, T. B., van Helmond, N., Dart, R. A., Brady, T. M., Turkson‐Ocran, R., & Juraschek, S. P. (2023) · Cardiovascular Digital Health Journal · doi:10.1016/j.cvdhj.2023.01.001 · Cited by 74
McCarthy, B. M., Vaughan, C. J., O’Flynn, B., Mathewson, A., & Mathúna, C. Ó. (2013) · Journal of Human Hypertension · doi:10.1038/jhh.2013.41 · Cited by 72
Rajala, S., Lindholm, H., & Taipalus, T. (2018) · Physiological Measurement · doi:10.1088/1361-6579/aac7ac · Cited by 70
Advances in Non-Invasive Blood Pressure Measurement Techniques
Panula, T., Sirkiä, J., Wong, D., & Kaisti, M. (2022) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2022.3141877 · Cited by 70
Wong, M. Y. M., Pickwell‐MacPherson, E., Zhang, Y., & Cheng, J. C. Y. (2010) · European Journal of Applied Physiology · doi:10.1007/s00421-010-1626-0 · Cited by 69
Identifying human body states by using a flexible integrated sensor
Jin, Y., Chen, G., Lao, K., Li, S., Lu, Y. L., Gan, Y., Li, Z., Hu, J., Huang, J., Wen, J., et al. (2020) · npj Flexible Electronics · doi:10.1038/s41528-020-00090-9 · Cited by 67
Preschool Children with Obstructive Sleep Apnea: The Beginnings of Elevated Blood Pressure?
Nisbet, L. C., Yiallourou, S. R., Biggs, S. N., Nixon, G. M., Davey, M. J., Trinder, J. A., Walter, L. M., & Horne, R. S. C. (2013) · SLEEP · doi:10.5665/sleep.2890 · Cited by 67
Lee, J., Yang, S.-M., Lee, S., & Kim, H. C. (2019) · Journal of Clinical Medicine · doi:10.3390/jcm8111773 · Cited by 72
Continuous Blood Pressure Monitoring using ECG and Finger Photoplethysmogram
Chua, C. P., & Heneghan, C. (2006) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2006.259612 · Cited by 65
Evaluation of blood pressure changes using vascular transit time
Foo, J. Y. A., Lim, C. S., & Wang, P. (2006) · Physiological Measurement · doi:10.1088/0967-3334/27/8/003 · Cited by 62
Yang, C., & Tavassolian, N. (2017) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2017.2696703 · Cited by 62
Chan, G. S. H., Middleton, P. M., Celler, B. G., Wang, L., & Lovell, N. H. (2007) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-007-9086-8 · Cited by 62
Bentham, M., Stansby, G. P., & Allen, J. (2018) · Diseases · doi:10.3390/diseases6030081 · Cited by 60
Parametric estimation of pulse arrival time: a robust approach to pulse wave velocity
Solà, J., Vetter, R., Renevey, P., Chételat, O., Sartori, C. S., & Rimoldi, S. F. (2009) · Physiological Measurement · doi:10.1088/0967-3334/30/7/006 · Cited by 58
Continuous blood pressure measurement using pulse transit time
Hennig, A., & Patzak, A. R. (2013) · Somnologie - Schlafforschung und Schlafmedizin · doi:10.1007/s11818-013-0617-x · Cited by 58
Dimensions of interoception predict premonitory urges and tic severity in Tourette syndrome
Rae, C. L., Larsson, D. E. O., Garfinkel, S. N., & Critchley, H. D. (2019) · Psychiatry Research · doi:10.1016/j.psychres.2018.12.036 · Cited by 57
Lin, W.-H., Chen, F., Geng, Y., Ji, N., Fang, P., & Li, G. (2020) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2020.102198 · Cited by 59
Sugita, N., Yoshizawa, M., Abe, M., Tanaka, A., Homma, N., & Yambe, T. (2018) · Journal of Medical and Biological Engineering · doi:10.1007/s40846-018-0388-8 · Cited by 55
Challenges and Opportunities in Cardiovascular Health Informatics
Zhang, Y., Zheng, Y.-L., Lin, W.-H., Zhang, H., & Zhou, X.-L. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2244892 · Cited by 54
Kim, S., Song, J., Park, J.-H., Kim, J., Park, Y., & Hwang, G. (2012) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e318270a6d9 · Cited by 54
Huttunen, J. M. J., Kärkkäinen, L., & Lindholm, H. J. (2019) · PLoS Computational Biology · doi:10.1371/journal.pcbi.1007259 · Cited by 52
Arterial blood pressure feature estimation using photoplethysmography
zadi, A. S., Alex, R. M., Zhang, R., Watenpaugh, D. E., & Behbehani, K. (2018) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2018.09.013 · Cited by 51
Multimodal Photoplethysmography-Based Approaches for Improved Detection of Hypertension
Welykholowa, K., Hosanee, M., Chan, G., Cooper, R. E., Kyriacou, P. A., Zheng, D., Allen, J., Abbott, D., Menon, C., Lovell, N. H., et al. (2020) · Journal of Clinical Medicine · doi:10.3390/jcm9041203 · Cited by 51
Ding, X., Zhang, Y., & Tsang, H. K. (2016) · Physiological Measurement · doi:10.1088/0967-3334/37/2/227 · Cited by 50
ECG Reconstruction via PPG: A Pilot Study
Zhu, Q., Tian, X., Wong, C.-W., & Wu, M. (2019) · IEEE-EMBS International Conference on Biomedical and Health Informatics (BHI ...) · doi:10.1109/bhi.2019.8834612 · Cited by 50
Beutel, F., Hoof, C. A. V., Rottenberg, X., Reesink, K. D., & Hermeling, E. (2021) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2021.3055154 · Cited by 48
Muehlsteff, J., Aubert, X. L., & Morren, G. (2008) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2008.4649501 · Cited by 48
Beat-to-Beat Continuous Blood Pressure Estimation Using Bidirectional Long Short-Term Memory Network
Lee, D., Kwon, H., Son, D., Eom, H., Park, C., Lim, Y., Seo, C., & Park, K. (2020) · Sensors · doi:10.3390/s21010096 · Cited by 49
Stojanova, A., Koceski, S., & Koceska, N. (2019) · Journal of Medical Systems · doi:10.1007/s10916-018-1138-8 · Cited by 46
Multi-level information fusion for learning a blood pressure predictive model using sensor data
Simjanoska, M., Kochev, S., Tanevski, J., Bogdanova, A. M., Papa, G., & Eftimov, T. (2019) · Information Fusion · doi:10.1016/j.inffus.2019.12.008 · Cited by 46
A Blood Pressure Monitoring Method for Stroke Management
Ma, H. T. (2014) · BioMed Research International · doi:10.1155/2014/571623 · Cited by 46
An implantable optical blood pressure sensor based on pulse transit time
Fiala, J., Bingger, P., Ruh, D., Foerster, K., Heilmann, C., Beyersdorf, F., Zappe, H. P., & Seifert, A. (2012) · Biomedical Microdevices · doi:10.1007/s10544-012-9689-9 · Cited by 45
Metadata
- Authors
- Payne, R. A., Symeonides, C. N., Webb, D. J., & Maxwell, S. R. J.
- Affiliations
- Western General Hospital (GB)
- University of Edinburgh (GB)
- Venue
- Journal of Applied Physiology
- Publisher
- American Physiological Society
- Published
- 2006-01
- Volume
- 100
- Issue
- 1
- Pages
- 136-141
- ISSN
- 0161-7567 (ISSN-L)
- 8750-7587
- 1522-1601
- OpenAlex
- W2059208589
- Cited by
- 418
- References
- 41
- Citations by year
- 2026: 12 · 2025: 17 · 2024: 15 · 2023: 18 · 2022: 16 · 2021: 25 · 2020: 39 · 2019: 32 · 2018: 40 · 2017: 28 · 2016: 31 · 2015: 18 · 2014: 19 · 2013: 13 · 2012: 16
- FWCI
- 8.20
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Journal of Applied Physiology (publishedVersion)
- PubMed (publishedVersion)
- Explore Bristol Research (submittedVersion)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Keywords
- pulse transit time, beat-to-beat blood pressure, pre-ejection period, systolic blood pressure, diastolic blood pressure, mean arterial pressure, vasoactive drugs, nitroglycerin, angiotensin II, norepinephrine, salbutamol, photoplethysmography, blood pressure variability, heart rate variability
- MeSH terms
- Adolescent, Adult, Blood Pressure, Blood Pressure Determination, Diagnosis, Computer-Assisted, Electrocardiography, Heart Rate, Humans, Male, Manometry, Pulsatile Flow, Sensitivity and Specificity, Reproducibility of Results, Photoplethysmography
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review