“Turn it Off!”: Diabetes Device Alarm Fatigue Considerations for the Present and the Future
Open accessThe DOI matches its Crossref record; 10 of the 10 references it cites were found in a registry or an index.
Abstract
Safe and widespread use of diabetes technology is constrained by alarm fatigue: When someone receives so many alarms that he or she becomes less likely to respond appropriately. Alarm fatigue and related usability issues deserve consideration at every stage of alarm system design, especially as new technologies expand the potential number and complexity of alarms. The guiding principle should be patient wellbeing, while taking into consideration the regulatory and liability issues that sometimes contribute to building excessive alarms. With examples from diabetes devices, we illustrate two complementary frameworks for alarm design: A “patient safety first” perspective and a focus on human factors. We also describe opportunities and challenges that will come with new technologies such as remote monitoring, adaptive alarms, and ever-closer integration of glucose sensing with insulin delivery.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Drew, B. J., Califf, R. M., Funk, M., Kaufman, E. S., Krucoff, M. W., Laks, M. M., Macfarlane, P. W., Sommargren, C., Swiryn, S., & Van Hare, G. F. (2004) · Circulation · doi:10.1161/01.cir.0000145144.56673.59 · Cited by 609
Tanenberg, R. J., Newton, C. A., & Drake, A. J. (2010) · Endocrine Practice · doi:10.4158/ep09260.cr · Cited by 225
Prevention of Nocturnal Hypoglycemia Using Predictive Alarm Algorithms and Insulin Pump Suspension
Buckingham, B., Chase, H. P., Dassau, E., Cobry, E., Clinton, P., Gage, V., Caswell, K., Wilkinson, J., Cameron, F., Lee, H., et al. (2010) · Diabetes Care · doi:10.2337/dc09-2303 · Cited by 169
Phantom vibrations among undergraduates: Prevalence and associated psychological characteristics
Drouin, M., Kaiser, D. H., & Miller, D. A. (2012) · Computers in Human Behavior · doi:10.1016/j.chb.2012.03.013 · Cited by 104
Atzema, C., Schull, M. J., Borgundvaag, B., Slaughter, G. R. D., & Lee, C. K. (2006) · The American Journal of Emergency Medicine · doi:10.1016/j.ajem.2005.05.015 · Cited by 98
Nocturnal hypoglycemia in type 1 diabetes: an assessment of preventive bedtime treatments
Raju, B., Arbelaez, A. M., Breckenridge, S. M., & Cryer, P. E. (2006) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2005-2798 · Cited by 93
Realistic Expectations and Practical Use of Continuous Glucose Monitoring for the Endocrinologist
Hirsch, I. B. (2009) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2008-2625 · Cited by 93
Ahmet, A., Dagenais, S., Barrowman, N. J., Collins, C. J., & Lawson, M. L. (2011) · The Journal of Pediatrics · doi:10.1016/j.jpeds.2011.01.064 · Cited by 42
Ward, W. K., Castle, J. R., & Youssef, J. E. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500608 · Cited by 36
van Bon, A. C., Dragt, D., & DeVries, J. H. (2012) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2011.0259 · Cited by 25
Cited by
The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System
Bailey, T., Bode, B. W., Christiansen, M. P., Klaff, L. J., & Alva, S. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0378 · Cited by 685
“The System displays trends and alerts on the Reader but does not have real-time alarms, which may provide a good option for individuals who complain of alarm fatigue.(11,13) The objective of this article is to describe the performance and usability of the System by comparing its scanned sensor results with capillary B…”
BackgroundMethodClosed-Loop Artificial Pancreas Systems: Engineering the Algorithms
Doyle, F. J., Huyett, L. M., Lee, J. B., Zisser, H. C., & Dassau, E. (2014) · Diabetes Care · doi:10.2337/dc13-2108 · Cited by 394
Self-citationContinuous glucose monitoring: A review of the technology and clinical use
Klonoff, D. C., Ahn, D., & Drincic, A. (2017) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2017.08.005 · Cited by 320
Assessing Sensor Accuracy for Non-Adjunct Use of Continuous Glucose Monitoring
Kovatchev, B. P., Patek, S. D., Ortiz, E. A., & Breton, M. D. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0272 · Cited by 234
Moser, O., Riddell, M. C., Eckstein, M. L., Adolfsson, P., Rabasa-Lhoret, R., van den Boom, L., Gillard, P., Nørgaard, K., Oliver, N. S., Zaharieva, D. P., et al. (2020) · Diabetologia · doi:10.1007/s00125-020-05263-9 · Cited by 214
Fokkert, M. J., van Dijk, P. R., Edens, M. A., Abbes, S., de Jong, D., Slingerland, R. J., & Bilo, H. J. G. (2017) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2016-000320 · Cited by 209
“FSL-CGM could be considered as CGM on demand or a viable alternative to frequent finger prick readings, and might also benefit individuals who have ceased sensor use due to alarm fatigue.(28) Owing to its 14-day longevity, costs are lower compared with available CGM systems,(27) 29 30 bringing the FSL-CGM within easie…”
MethodLawton, J., Blackburn, M., Allen, J., Campbell, F., Elleri, D., Leelarathna, L., Rankin, D., Tauschmann, M., Thabit, H., & Hovorka, R. (2018) · BMC Endocrine Disorders · doi:10.1186/s12902-018-0239-1 · Cited by 174
“This is an issue which could be explored further in psychological work, to help determine whether some individuals, especially those with sub-optimal selfmanagement behaviours and/or a high HbA1c, would benefit from psychosocial support prior to initiation of CGM to help maximise the benefits from alarms while prevent…”
Backgroundde Wit, M., Gajewska, K. A., Goethals, E. R., McDarby, V., Zhao, X., Hapunda, G., Delamater, A. M., & DiMeglio, L. A. (2022) · Pediatric Diabetes · doi:10.1111/pedi.13428 · Cited by 185
“191 However, according to a systematic review on this topic, there were no differences in body image between those with T1D using and not using technology (insulin pumps, CGM).192 Frequent exposure to device alarms (in particular false or unnecessary ones) for a device user or a caregiver can result in poor or interru…”
BackgroundClinical accuracy of a continuous glucose monitoring system with an advanced algorithm
Bailey, T. S., Chang, A., & Christiansen, M. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814559746 · Cited by 168
Visser, M. M., Charleer, S., Fieuws, S., De Block, C., Hilbrands, R., Van Huffel, L., Maes, T., Vanhaverbeke, G., Dirinck, E., Myngheer, N., et al. (2021) · The Lancet · doi:10.1016/s0140-6736(21)00789-3 · Cited by 165
Alva, S., Bailey, T., Brazg, R., Budiman, E. S., Castorino, K., Christiansen, M. P., Forlenza, G., Kipnes, M., Liljenquist, D. R., & Liu, H. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820958754 · Cited by 164
Wearable physiological systems and technologies for metabolic monitoring
Gao, W., Brooks, G. A., & Klonoff, D. C. (2018) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00407.2017 · Cited by 91
The Growing Role of Technology in the Care of Older Adults With Diabetes
Huang, E. S., Sinclair, A., Conlin, P. R., Cukierman-Yaffe, T., Hirsch, I. B., Huisingh-Scheetz, M., Kahkoska, A. R., Laffel, L., Lee, A. K., Lee, S., et al. (2023) · Diabetes Care · doi:10.2337/dci23-0021 · Cited by 73
Reference Guide for Integrating Continuous Glucose Monitoring Into Clinical Practice
Kruger, D. F., Edelman, S. V., Hinnen, D. A., & Parkin, C. G. (2019) · The Diabetes Educator · doi:10.1177/0145721718818066 · Cited by 55
Benefits and limitations of continuous glucose monitoring in type 1 diabetes
Stone, J. Y., & Bailey, T. S. (2020) · Expert Review of Endocrinology & Metabolism · doi:10.1080/17446651.2020.1706482 · Cited by 30
“Alerts and alarms can also be disruptive to sleep, and while this would be considered a good thing when the alarms are appropriate, frequent exposure to false or unnecessary alerts or alarms can lead to users becoming less likely to respond to those alarms in what has come to be known as ‘alarm fatigue’ [48].”
BackgroundBeyond A1C: exploring continuous glucose monitoring metrics in managing diabetes
Friedman, J. G., Coyne, K., Aleppo, G., & Szmuilowicz, E. D. (2023) · Endocrine Connections · doi:10.1530/ec-23-0085 · Cited by 22
Digital Solutions to Diagnose and Manage Postbariatric Hypoglycemia
Schönenberger, K. A., Cossu, L., Prendin, F., Cappon, G., Wu, J., Fuchs, K. L., Mayer, S., Herzig, D., Facchinetti, A., & Bally, L. (2022) · Frontiers in Nutrition · doi:10.3389/fnut.2022.855223 · Cited by 16
Mendez, C., Kaykayoglu, C. A., Bähler, T., Künzler, J., Lizoain, A., Rothenbühler, M., Schmidt, M. H., Laimer, M., & Witthauer, L. (2026) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251319800 · Cited by 10
Giza, S., Kotanidou, E. P., Tsinopoulou, V. R., Poukoulidou, T., Ntouma, S., & Galli-Tsinopoulou, A. (2025) · Children · doi:10.3390/children12121668 · Cited by 7
“The high frequency of alarms increases the potential of fatigue [9,16].”
InfluentialFear of hypoglycemia and sleep in children with type 1 diabetes and their parents
Schierloh, U., Aguayo, G. A., Fichelle, M., De Melo Dias, C., Schritz, A., Vaillant, M., Barnard-Kelly, K., Cohen, O., Gies, I., & de Beaufort, C. (2024) · Frontiers in Endocrinology · doi:10.3389/fendo.2024.1419502 · Cited by 7
Insulin Pump Alarms During Adverse Events: A Qualitative Descriptive Study
Estock, J. L., Codario, R. A., Zupa, M. F., Keddem, S., & Rodriguez, K. L. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231209999 · Cited by 4
“19,20 Second, there is a need to optimize insulin pump “alerts” as the frequency of “alerts” cited in insulin pump-associated AEs suggests that some “alerts” may require a more immediate patient response to prevent AEs.”
Knopp, M. I., Castleman, A. M., Schwarz, A., Belger-Wallace, J., Falciglia, M., Zuzek, A., & Mendonca, E. (2025) · JMIR Diabetes · doi:10.2196/73446 · Cited by 2
“Alarm burden [50] continues to hinder optimal use and leads to discontinuation for some participants.”
Metadata
- Authors
- Shivers, J. P., Mackowiak, L., Anhalt, H., & Zisser, H.
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2013-05
- Volume
- 7
- Issue
- 3
- Pages
- 789-794
- ISSN
- 1932-2968 (ISSN-L)
- OpenAlex
- W2144678779
- Cited by
- 158
- References
- 10
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.