Cardiovascular magnetic resonance artefacts
Open accessThe DOI matches its Crossref record; 56 of the 56 references it cites were found in a registry or an index.
Abstract
The multitude of applications offered by CMR make it an increasing popular modality to study the heart and the surrounding vessels. Nevertheless the anatomical complexity of the chest, together with cardiac and respiratory motion, and the fast flowing blood, present many challenges which can possibly translate into imaging artefacts. The literature is wide in terms of papers describing specific MR artefacts in great technical detail. In this review we attempt to summarise, in a language accessible to a clinical readership, some of the most common artefacts found in CMR applications. It begins with an introduction of the most common pulse sequences, and imaging techniques, followed by a brief section on typical cardiovascular applications. This leads to the main section on common CMR artefacts with examples, a short description of the mechanisms behind them, and possible solutions.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
On the use of windows for harmonic analysis with the discrete Fourier transform
Harris, F. J. (1978) · Proceedings of the IEEE · doi:10.1109/proc.1978.10837 · Cited by 7,326 · cited 1×
“Hamming and Hann filters are commonly used in image and signal processing to reduce Gibbs ringing artefacts, although at the penalty of reducing spatial-resolution [62].”
BackgroundSENSE: Sensitivity encoding for fast MRI
Pruessmann, K. P., Weiger, M., Scheidegger, M. B., & Boesiger, P. (1999) · Magnetic Resonance in Medicine · doi:10.1002/(sici)1522-2594(199911)42:5<952::aid-mrm16>3.0.co;2-s · Cited by 6,251 · cited 4×
“In SENSE the incomplete k-space data is first converted into the image-space and then unwrapped using the multiple coil information (Figure 9), while with GRAPPA the missing k-space lines are generated for each coil using the acquired data of all the coils, i.e. undersampling effects are addressed before Fourier trans…”
MethodInfluentialGeneralized autocalibrating partially parallel acquisitions (GRAPPA)
Griswold, M. A., Jakob, P. M., Heidemann, R. M., Nittka, M., Jellus, V., Wang, J., Kiefer, B., & Haase, A. (2002) · Magnetic Resonance in Medicine · doi:10.1002/mrm.10171 · Cited by 5,303 · cited 4×
“In SENSE the incomplete k-space data is first converted into the image-space and then unwrapped using the multiple coil information (Figure 9), while with GRAPPA the missing k-space lines are generated for each coil using the acquired data of all the coils, i.e. undersampling effects are addressed before Fourier trans…”
MethodInfluentialImage Formation by Induced Local Interactions: Examples Employing Nuclear Magnetic Resonance
LAUTERBUR, P. C. (1973) · Nature · doi:10.1038/242190a0 · Cited by 3,611 · cited 1×
“Radial sampling was the first k-space sampling trajectory to be used in MRI with a backprojection image reconstruction [11], analogous to Computed Tomography.”
MethodAn Improved MR Imaging Technique for the Visualization of Myocardial Infarction
Simonetti, O. P., Kim, R. J., Fieno, D. S., Hillenbrand, H. B., Wu, E., Bundy, J. M., Finn, J. P., & Judd, R. M. (2001) · Radiology · doi:10.1148/radiology.218.1.r01ja50215 · Cited by 1,358 · cited 1×
“A segmented GRE sequence with an inversion pulse preparation is normally used [37].”
MethodT1, T2 relaxation and magnetization transfer in tissue at 3T
Stanisz, G. J., Odrobina, E. E., Pun, J., Escaravage, M., Graham, S. J., Bronskill, M. J., & Henkelman, R. M. (2005) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20605 · Cited by 1,266 · cited 1×
“5T, T1 relaxation times are noticeably longer at 3T [97].”
BackgroundMR imaging of motion with spatial modulation of magnetization.
Axel, L., & Dougherty, L. (1989) · Radiology · doi:10.1148/radiology.171.3.2717762 · Cited by 1,150
Motion correction with PROPELLER MRI: Application to head motion and free-breathing cardiac imaging
Pipe, J. G. (1999) · Magnetic Resonance in Medicine · doi:10.1002/(sici)1522-2594(199911)42:5<963::aid-mrm17>3.0.co;2-l · Cited by 928 · cited 1×
“The explanation of these methods falls outside the scope of this article, but the interested reader is referred to the respective bibliographical references [12,13].”
MethodFast Spiral Coronary Artery Imaging
Meyer, C. H., Hu, B. S., Nishimura, D. G., & Macovski, A. (1992) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910280204 · Cited by 866 · cited 1×
“Sequences with non-Cartesian forms of k-space coverage such as radial and spiral have also been used to image the heart [82-88], artefacts discussed earlier can have very different characteristics when compared to the more conventional Cartesian methods.”
Methodk‐t BLAST and k‐t SENSE: Dynamic MRI with high frame rate exploiting spatiotemporal correlations
Tsao, J., Boesiger, P., & Pruessmann, K. P. (2003) · Magnetic Resonance in Medicine · doi:10.1002/mrm.10611 · Cited by 794 · cited 1×
“However as techniques such as parallel imaging and other image acceleration methods improve they can be used to increase spatial-resolution without increasing imaging time [63-65].”
MethodLevitt, M. H. (1986) · Progress in Nuclear Magnetic Resonance Spectroscopy · doi:10.1016/0079-6565(86)80005-x · Cited by 776
McRobbie, D. W., Moore, E. A., Graves, M. J., & Prince, M. R. (2017) · Cambridge University Press eBooks · doi:10.1017/9781107706958 · Cited by 752
A flow velocity zeugmatographic interlace for NMR imaging in humans
Moran, P. R. (1982) · Magnetic Resonance Imaging · doi:10.1016/0730-725x(82)90170-9 · Cited by 745 · cited 1×
“Gradient-echo sequences can also be used to measure blood flow velocity using a technique known as phase contrast velocity mapping [31-34].”
MethodPrinciples and applications of balanced SSFP techniques
Scheffler, K., & Lehnhardt, S. (2003) · European Radiology · doi:10.1007/s00330-003-1957-x · Cited by 714
Blood Flow Imaging by Cine Magnetic Resonance
Nayler, G. L., Firmin, D. N., & Longmore, D. B. (1986) · Journal of Computer Assisted Tomography · doi:10.1097/00004728-198609000-00001 · Cited by 634 · cited 1×
“Gradient-echo sequences can also be used to measure blood flow velocity using a technique known as phase contrast velocity mapping [31-34].”
MethodSelf-citationOtazo, R., Kim, D., Axel, L., & Sodickson, D. K. (2010) · Magnetic Resonance in Medicine · doi:10.1002/mrm.22463 · Cited by 571
Projection Reconstruction Techniques for Reduction of Motion Effects in MRI
Glover, G. H., & Pauly, J. M. (1992) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910280209 · Cited by 563 · cited 1×
“Radial sequences tend to oversample the centre of k-space and this tends to reduce the visibility of breathing artefacts [89].”
Background"Black blood" T2-weighted inversion-recovery MR imaging of the heart.
Simonetti, O. P., Finn, J. P., White, R. D., Laub, G., & Henry, D. A. (1996) · Radiology · doi:10.1148/radiology.199.1.8633172 · Cited by 560 · cited 1×
“For cardiovascular morphology, a double-inversion RF pulse preparation is commonly added to a turbo spinecho sequence, removing the blood signal and providing a good contrast between the myocardium and blood, and is commonly referred as black or dark-blood preparation [23].”
Concomitant gradient terms in phase contrast MR: Analysis and correction
Bernstein, M. A., Zhou, X. J., Polzin, J. A., King, K. F., Ganin, A., Pelc, N. J., & Glover, G. H. (1998) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910390218 · Cited by 547 · cited 1×
“Some distortion of the magnetic field gradients is unavoidable from fundamental electromagnetism, and this is known in MRI as Maxwell or concomitant gradients [77].”
BackgroundMeasurement of Flow with NMR Imaging Using a Gradient Pulse and Phase Difference Technique
Bryant, D. J., Payne, J. A., Firmin, D. N., & Longmore, D. B. (1984) · Journal of Computer Assisted Tomography · doi:10.1097/00004728-198408000-00002 · Cited by 545 · cited 1×
“Gradient-echo sequences can also be used to measure blood flow velocity using a technique known as phase contrast velocity mapping [31-34].”
MethodSelf-citationMadore, B., Glover, G. H., & Pelc, N. J. (1999) · Magnetic Resonance in Medicine · doi:10.1002/(sici)1522-2594(199911)42:5<813::aid-mrm1>3.0.co;2-s · Cited by 433 · cited 2×
“Combining UNFOLD with parallel imaging techniques improves coil sensitivity profile estimation.”
Larson, A. C., White, R. D., Laub, G., McVeigh, E. R., Li, D., & Simonetti, O. P. (2004) · Magnetic Resonance in Medicine · doi:10.1002/mrm.10664 · Cited by 419 · cited 1×
“Similarly self-gating has been described using radial sampling with a 1DFT projection reconstruction to monitor the heart motion through the cardiac cycle and then to use this for cardiac gating [91].”
MethodMeasurement of right and left ventricular volumes in healthy individuals with cine MR imaging.
Sechtem, U., Pflugfelder, P. W., Gould, R. G., Cassidy, M. M., & Higgins, C. B. (1987) · Radiology · doi:10.1148/radiology.163.3.3575717 · Cited by 407 · cited 1×
“bSSFP and GRE sequences are commonly used to assess cardiac function [25-28].”
BackgroundImage formation by induced local interactions. Examples employing nuclear magnetic resonance. 1973.
Pc, L. (1989) · PubMed · Cited by 391
Barkhausen, J., Ruehm, S. G., Goyen, M., Buck, T., Laub, G., & Debatin, J. F. (2001) · Radiology · doi:10.1148/radiology.219.1.r01ap12264 · Cited by 380 · cited 1×
“bSSFP and GRE sequences are commonly used to assess cardiac function [25-28].”
BackgroundImproved myocardial tagging contrast
Fischer, S., McKinnon, G. C., Maier, S. E., & Boesiger, P. (1993) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910300207 · Cited by 371 · cited 2×
“The contrast of the tag lines can be improved by summing two sets of images with complementary tagging pulses, in a technique called CSPAMM [10]. k-space sampling The data collected by the scanner needs to be converted to the final image.”
MethodFirst-pass cardiac perfusion: evaluation with ultrafast MR imaging.
Atkinson, D. J., Burstein, D., & Edelman, R. R. (1990) · Radiology · doi:10.1148/radiology.174.3.2305058 · Cited by 359
Direct Cardiac NMR Imaging of Heart Wall and Blood Flow Velocity
van Dijk, P. (1984) · Journal of Computer Assisted Tomography · doi:10.1097/00004728-198406000-00012 · Cited by 343
Parallel magnetic resonance imaging
Larkman, D. J., & Nunes, R. G. (2007) · Physics in Medicine and Biology · doi:10.1088/0031-9155/52/7/r01 · Cited by 327
Stuber, M., Botnar, R. M., Danias, P. G., Sodickson, D. K., Kissinger, K. V., Van Cauteren, M., De Becker, J., & Manning, W. J. (1999) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(99)00223-5 · Cited by 320 · cited 1×
“A segmented 3D ECG-triggered acquisition is typically used to image the coronary arteries (Figure 16) [40-42].”
MethodSpoiling of transverse magnetization in steady‐state sequences
Zur, Y., Wood, M. L., & Neuringer, L. J. (1991) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910210210 · Cited by 306 · cited 1×
“For the GRE sequence RF spoiling is often used to further reduce the impact of the remaining transverse magnetisation [6].”
MethodImage reconstruction for echo planar imaging with nonequidistant k‐space sampling
Bruder, H., Fischer, H., Reinfelder, H. ‐., & Schmitt, F. (1992) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910230211 · Cited by 269 · cited 1×
“This artefact is commonly known as Nyquist ghost or N/2 artefact and it is created by a mixture of gradient imperfections [93], eddy currents, concomitant fields [94], receiver filter asymmetry [95] or susceptibility.”
BackgroundHalving MR imaging time by conjugation: demonstration at 3.5 kG.
Feinberg, D. A., Hale, J. D., Watts, J. C., Kaufman, L., & Mark, A. (1986) · Radiology · doi:10.1148/radiology.161.2.3763926 · Cited by 259
Plein, S., Bloomer, T. N., Ridgway, J. P., Jones, T. R., Bainbridge, G. J., & Sivananthan, M. U. (2001) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.1178 · Cited by 257 · cited 1×
“bSSFP and GRE sequences are commonly used to assess cardiac function [25-28].”
BackgroundCardiovascular magnetic resonance physics for clinicians: part II
Biglands, J. D., Radjenovic, A., & Ridgway, J. P. (2012) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-14-66 · Cited by 241 · cited 1×
“Also, alternatives to these sections are available elsewhere such as the excellent physics articles from Ridgway and Biglands in JCMR 2010 and 2012 [4,5].”
BackgroundGatehouse, P. D., Rolf, M. P., Graves, M. J., Hofman, M. B., Totman, J., Werner, B., Quest, R. A., Liu, Y., von Spiczak, J., Dieringer, M., et al. (2010) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-12-5 · Cited by 234 · cited 1×
“generally available on scanners, but is available in post image-processing s sensitivity range (VENC)) can significantly affect measurements of shunt flow and cardiac output, and that these may occur easily on some systems [78].”
BackgroundSelf-citationCardiovascular magnetic resonance physics for clinicians: part I
Ridgway, J. P. (2010) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-12-71 · Cited by 218 · cited 1×
“Also, alternatives to these sections are available elsewhere such as the excellent physics articles from Ridgway and Biglands in JCMR 2010 and 2012 [4,5].”
BackgroundThiele, H., Nagel, E., Paetsch, I., Schnackenburg, B., Bornstedt, A., Kouwenhoven, M., Wahl, A., Schuler, G., & Fleck, E. (2001) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.1195 · Cited by 218 · cited 1×
“bSSFP and GRE sequences are commonly used to assess cardiac function [25-28].”
BackgroundIn vivo measurement of T*2 and field inhomogeneity maps in the human heart at 1.5 T
Reeder, S. B., Faranesh, A. Z., Boxerman, J. L., & McVeigh, E. R. (1998) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910390617 · Cited by 189 · cited 1×
“Localised field inhomogeneities due to susceptibility differences in the locality of the cardiac veins [68]; and at tissue-air interfaces such as the heart-lung interface [69], or adjacent to air pockets [70] (in the bowel, stomach, and colon for example), introduce banding type artefacts on bSSFP images that can conf…”
BackgroundField‐of‐view limitations in parallel imaging
Griswold, M. A., Kannengießer, S., Heidemann, R. M., Wang, J., & Jakob, P. M. (2004) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20249 · Cited by 160 · cited 1×
“can be seen mimicking gadolinium enhancement on image a (arrow), but sensitivity maps may also be aliased, and therefore erroneous image reconstruction will result in incomplete unwrapping of the aliased pixels [80] (some scanner manufacturers acquire sensitivity maps over a much larger FOV than the image acquisition…”
BackgroundMRI artifacts and correction strategies
Smith, T., & Nayak, K. S. (2010) · Imaging in Medicine · doi:10.2217/iim.10.33 · Cited by 145 · cited 1×
“This article, that is designed for a clinical readership, follows previous publications on the subject [1-3].”
BackgroundQuantification of left ventricular function with magnetic resonance images acquired in real time
Setser, R. M., Fischer, S., & Lorenz, C. H. (2000) · Journal of Magnetic Resonance Imaging · doi:10.1002/1522-2586(200009)12:3<430::aid-jmri8>3.0.co;2-v · Cited by 122 · cited 1×
“More recently real-time imaging strategies have been developed where the requirement of ECG triggering and or breath-holding are relaxed [55-59].”
BackgroundKellman, P., Hernando, D., & Arai, A. E. (2010) · Current Cardiovascular Imaging Reports · doi:10.1007/s12410-010-9012-1 · Cited by 103
Hofman, M. B. M., van Rossum, A. C., Sprenger, M., & Westerhof, N. (1996) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910350411 · Cited by 97 · cited 1×
“(Adapted and reprinted, with permission, from reference [50]).”
BackgroundObservation of blood flow E.M.F. in externally applied strong magnetic field by surface electrodes
Togawa, T., Okai, O., & Oshima, M. (1967) · Medical & Biological Engineering & Computing · doi:10.1007/bf02474505 · Cited by 93 · cited 1×
“triggering due to patients with arrhythmias or weak ECG signal; or due to magneto-hydrodynamic (distortion of the electrical field due to flowing ions in the blood) [54] and/ or gradient pulse distortion, can cause effects varying from minor inconsistencies to conspicuous artefacts.”
BackgroundUǧurbil, K., Garwood, M., & Bendall, M. R. (1987) · Journal of Magnetic Resonance (1969) · doi:10.1016/0022-2364(87)90186-7 · Cited by 92 · cited 1×
“Optimal saturation recovery preparation is generally achieved by the use of special RF pulse designs (composite [7], B1 Independent Rotation (BIR) [8]).”
BackgroundKaji, S., Yang, P. C., Kerr, A. B., Tang, W. H. W., Meyer, C. H., Macovski, A., Pauly, J. M., Nishimura, D. G., & Hu, B. S. (2001) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(01)01399-7 · Cited by 85
Adaptive technique for three-dimensional MR imaging of moving structures.
Korin, H. W., Felmlee, J. P., Ehman, R. L., & Riederer, S. J. (1990) · Radiology · doi:10.1148/radiology.177.1.2399320 · Cited by 78
Assessment of Myocardial Viability Using Delayed Enhancement Magnetic Resonance Imaging at 3.0 Tesla
Klumpp, B. D., Fenchel, M. C., Hoevelborn, T., Helber, U., Scheule, A. M., Claussen, C. D., & Miller, S. (2006) · Investigative Radiology · doi:10.1097/01.rli.0000233321.82194.09 · Cited by 78 · cited 1×
“The same changes in relaxation time can also be used to reduce the dose of Gd contrast agent in first-pass perfusion and LGE studies decreasing possible susceptibility problems [98,99].”
Free-Breathing Black-Blood Coronary MR Angiography: Initial Results
Stuber, M., Botnar, R. M., Kissinger, K. V., & Manning, W. J. (2001) · Radiology · doi:10.1148/radiology.219.1.r01mr07278 · Cited by 77 · cited 1×
“frequency selective fat suppression prior to image acquisition [44].”
BackgroundNayak, K. S., Lee, H., Hargreaves, B. A., & Hu, B. S. (2007) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21296 · Cited by 67
Hori, Y., Yamada, N., Higashi, M., Hirai, N., & Nakatani, S. (2003) · Journal of Cardiovascular Magnetic Resonance · doi:10.1081/jcmr-120022260 · Cited by 67 · cited 1×
“More recently real-time imaging strategies have been developed where the requirement of ECG triggering and or breath-holding are relaxed [55-59].”
BackgroundPlein, S., Smith, W. H., Ridgway, J. P., Kassner, A., Beacock, D. J., Bloomer, T. N., & Sivananthan, M. U. (2001) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.1146 · Cited by 52 · cited 1×
“More recently real-time imaging strategies have been developed where the requirement of ECG triggering and or breath-holding are relaxed [55-59].”
BackgroundMyocardial wall tagging with undersampled projection reconstruction†
Peters, D. C., Epstein, F. H., & McVeigh, E. R. (2001) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1076 · Cited by 38 · cited 1×
“Sequences with non-Cartesian forms of k-space coverage such as radial and spiral have also been used to image the heart [82-88], artefacts discussed earlier can have very different characteristics when compared to the more conventional Cartesian methods.”
MethodKung, G. L., Ajijola, O. A., Ramírez, R. J., Gahm, J. K., Zhou, W., Wisniewski, N., Mahajan, A., Garfinkel, A., Shivkumar, K., & Ennis, D. B. (2012) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-14-s1-p66 · Cited by 38
Ferreira, P., Gatehouse, P. D., Bucciarelli‐Ducci, C., Wage, R., & Firmin, D. N. (2008) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21723 · Cited by 21 · cited 1×
“5 T [75], although when combined with other sources of field”
BackgroundSelf-citation
Cited by
Goldstein, S. A., Evangelista, A., Abbara, S., Arai, A., Asch, F. M., Badano, L. P., Bolen, M. A., Connolly, H. M., Cuéllar-Calàbria, H., Czerny, M., et al. (2015) · Journal of the American Society of Echocardiography · doi:10.1016/j.echo.2014.11.015 · Cited by 641
Metadata
- Authors
- Ferreira, P. F., Gatehouse, P. D., Mohiaddin, R. H., & Firmin, D. N.
- Venue
- Journal of Cardiovascular Magnetic Resonance
- Publisher
- Elsevier BV
- Published
- 2013-01
- Volume
- 15
- Issue
- 1
- Pages
- 41
- ISSN
- 1097-6647 (ISSN-L)
- OpenAlex
- W2051693142
- Cited by
- 191
- References
- 56
- FWCI
- 7.84
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.