Comparison of the Numerical and Clinical Accuracy of Four Continuous Glucose Monitors
Open accessAbstract
OBJECTIVE—The purpose of this study was to compare the numerical and clinical accuracy of four continuous glucose monitors (CGMs): Guardian, DexCom, Navigator, and Glucoday. RESEARCH DESIGN AND METHODS—Accuracy data for the four CGMs were collected in two studies: Study 1 enrolled 14 adults with type 1 diabetes at the University of Virginia (UVA), Charlottesville, Virginia; study 2 enrolled 20 adults with type 1 diabetes at the Profil Institute for Metabolic Research, Neuss, Germany. All participants underwent hyperinsulinemic clamps including 1.5–2 h of maintained euglycemia at 5.6 mmol/l followed by descent into hypoglycemia, sustained hypoglycemia at 2.5 mmol/l for 30 min, and recovery. Reference blood glucose sampling was performed every 5 min. The UVA study tested Guardian, DexCom, and Navigator simultaneously; the Profil study tested Glucoday. RESULTS—Regarding numerical accuracy, during euglycemia, the mean absolute relative differences (MARDs) of Guardian, DexCom, Navigator, and Glucoday were 15.2, 21.2, 15.3, and 15.6%, respectively. During hypoglycemia, the MARDs were 16.1, 21.5, 10.3, and 17.5%, respectively. Regarding clinical accuracy, continuous glucose–error grid analysis (CG-EGA) revealed 98.9, 98.3, 98.6, and 95.5% zones A + B hits in euglycemia. During hypoglycemia, zones A + B hits were 84.4, 97.0, and 96.2% for Guardian, Navigator, and Glucoday, respectively. Because of frequent loss of sensitivity, there were insufficient hypoglycemic DexCom data to perform CG-EGA. CONCLUSIONS—The numerical accuracy of Guardian, Navigator, and Glucoday was comparable, with an advantage to the Navigator in hypoglycemia; the numerical errors of the DexCom were ∼30% larger. The clinical accuracy of the four sensors was similar in euglycemia and was higher for the Navigator and Glucoday in hypoglycemia.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Evaluating the Accuracy of Continuous Glucose-Monitoring Sensors
Kovatchev, B. P., Gonder-Frederick, L. A., Cox, D. J., & Clarke, W. L. (2004) · Diabetes Care · doi:10.2337/diacare.27.8.1922 · Cited by 312
Self-citationContinuous Glucose Sensors: Continuing Questions about Clinical Accuracy
Clarke, W. L., & Kovatchev, B. (2007) · Journal of Diabetes Science and Technology · doi:10.1177/193229680700100510 · Cited by 71
Self-citation
Cited by
Electrochemical Glucose Sensors and Their Applications in Diabetes Management
Heller, A., & Feldman, B. (2008) · Chemical Reviews · doi:10.1021/cr068069y · Cited by 1,707
Continuous glucose monitoring and intensive treatment of type 1 diabetes
(2008) · New England Journal of Medicine · doi:10.1056/nejmoa0805017 · Cited by 1,559
Electrochemical Non-enzymatic Glucose Sensors: A Perspective and an Evaluation
Toghill, K. E., & Compton, R. G. (2010) · International Journal of Electrochemical Science · doi:10.1016/s1452-3981(23)15359-4 · Cited by 926
Artificial Pancreas: Past, Present, Future
Cobelli, C., Renard, E., & Kovatchev, B. (2011) · Diabetes · doi:10.2337/db11-0654 · Cited by 600
Self-citationDiabetes: Models, Signals, and Control
Cobelli, C., Dalla Man, C., Sparacino, G., Magni, L., De Nicolao, G., & Kovatchev, B. P. (2009) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2009.2036073 · Cited by 511
Self-citationHovorka, R., Allen, J. M., Elleri, D., Chassin, L. J., Harris, J., Xing, D., Kollman, C., Hovorka, T., Larsen, A. M. F., Nodale, M., et al. (2010) · The Lancet · doi:10.1016/s0140-6736(09)61998-x · Cited by 450
Biocompatible Materials for Continuous Glucose Monitoring Devices
Nichols, S. P., Koh, A., Storm, W. L., Shin, J. H., & Schoenfisch, M. H. (2013) · Chemical Reviews · doi:10.1021/cr300387j · Cited by 347
Closed-loop insulin delivery from bench to clinical practice
Hovorka, R. (2011) · Nature Reviews Endocrinology · doi:10.1038/nrendo.2011.32 · Cited by 342
Function of an Implanted Tissue Glucose Sensor for More than 1 Year in Animals
Gough, D. A., Kumosa, L. S., Routh, T. L., Lin, J. T., & Lucisano, J. Y. (2010) · Science Translational Medicine · doi:10.1126/scitranslmed.3001148 · Cited by 179
Heinemann, L., Schoemaker, M., Schmelzeisen-Redecker, G., Hinzmann, R., Kassab, A., Freckmann, G., Reiterer, F., & Del Re, L. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819855670 · Cited by 170
Self-citationModeling the glucose sensor error
Facchinetti, A., Del Favero, S., Sparacino, G., Castle, J. R., Ward, W. K., & Cobelli, C. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2284023 · Cited by 130
Amperometric Glucose Sensors: Sources of Error and Potential Benefit of Redundancy
Castle, J. R., & Ward, W. K. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400127 · Cited by 95
Realistic Expectations and Practical Use of Continuous Glucose Monitoring for the Endocrinologist
Hirsch, I. B. (2009) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2008-2625 · Cited by 93
Freckmann, G., Eichenlaub, M., Waldenmaier, D., Pleus, S., Wehrstedt, S., Haug, C., Witthauer, L., Jendle, J., Hinzmann, R., Thomas, A., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231190941 · Cited by 86
Facchinetti, A., Del Favero, S., Sparacino, G., & Cobelli, C. (2015) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-014-1226-y · Cited by 80
Clinical Implications of Accuracy Measurements of Continuous Glucose Sensors
Bailey, T. S. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0050 · Cited by 60
Cosson, E., Hamo-Tchatchouang, E., Dufaitre-Patouraux, L., Attali, J.-R., Pariès, J., & Schaepelynck-Bélicar, P. (2009) · Diabetes & Metabolism · doi:10.1016/j.diabet.2009.02.006 · Cited by 57
Robust Fault Detection System for Insulin Pump Therapy Using Continuous Glucose Monitoring
Herrero, P., Calm, R., Vehí, J., Armengol, J., Georgiou, P., Oliver, N., & Tomazou, C. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600518 · Cited by 52
Evolution of Devices in Diabetes Management
Penfornis, A., Personeni, E., & Borot, S. (2011) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2011.0058 · Cited by 49
Limits to the Evaluation of the Accuracy of Continuous Glucose Monitoring Systems by Clinical Trials
Schrangl, P., Reiterer, F., Heinemann, L., Freckmann, G., & Del Re, L. (2018) · Biosensors · doi:10.3390/bios8020050 · Cited by 46
Self-citationHigh performance boronic acid-containing hydrogel for biocompatible continuous glucose monitoring
Dou, Q., Hu, D., Gao, H., Zhang, Y., Yetisen, A. K., Butt, H., Wang, J., Nie, G., & Dai, Q. (2017) · RSC Advances · doi:10.1039/c7ra06965k · Cited by 31
Clinical impact of blood glucose monitoring accuracy: an in-silico study
Campos-Náñez, E., Fortwaengler, K., & Breton, M. D. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817710474 · Cited by 22
Metadata
- Authors
- Kovatchev, B., Anderson, S., Heinemann, L., & Clarke, W.
- Venue
- Diabetes Care
- Publisher
- American Diabetes Association
- Published
- 2008-06-01
- Volume
- 31
- Issue
- 6
- Pages
- 1160-1164
- ISSN
- 0149-5992 (ISSN-L)
- 1935-5548
- OpenAlex
- W2158137085
- Cited by
- 236
- References
- 2
- FWCI
- 17.87
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.