Wireless Power Transfer Techniques for Implantable Medical Devices: A Review
Open accessAbstract
Wireless power transfer (WPT) systems have become increasingly suitable solutions for the electrical powering of advanced multifunctional micro-electronic devices such as those found in current biomedical implants. The design and implementation of high power transfer efficiency WPT systems are, however, challenging. The size of the WPT system, the separation distance between the outside environment and location of the implanted medical device inside the body, the operating frequency and tissue safety due to power dissipation are key parameters to consider in the design of WPT systems. This article provides a systematic review of the wide range of WPT systems that have been investigated over the last two decades to improve overall system performance. The various strategies implemented to transfer wireless power in implantable medical devices (IMDs) were reviewed, which includes capacitive coupling, inductive coupling, magnetic resonance coupling and, more recently, acoustic and optical powering methods. The strengths and limitations of all these techniques are benchmarked against each other and particular emphasis is placed on comparing the implanted receiver size, the WPT distance, power transfer efficiency and tissue safety presented by the resulting systems. Necessary improvements and trends of each WPT techniques are also indicated per specific IMD.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“A value closer to 1 not only maximizes energy transfer but also minimizes potential thermal damage to neural tissues—a concern for BMIs requiring high computational power for real-time neural decoding (Haerinia and Shadid, 2020; Khan et al., 2020).”
From Comparative analysis of energy transfer mechanisms for neural implants (2024) · cites this paper as Khan et al., 2020
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“This system utilizes a closed current loop and relies on the displacement current, IDisp., between the conductor plates to establish wireless power transfer across tissue layers (Khan et al., 2020).”
From Comparative analysis of energy transfer mechanisms for neural implants (2024) · cites this paper as Khan et al., 2020
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Wireless Power Transfer via Strongly Coupled Magnetic Resonances
Kurs, A., Karalis, A., Moffatt, R. A., Joannopoulos, J. D., Fisher, P. H., & Marin Soljacic, A. (2007) · Science · doi:10.1126/science.1143254 · Cited by 5,584 · cited 6×
“where QL (=2πfLRX/RL) [15,87,88] is the loaded Q-factor calculated by using the inductance of the RX coil, LRX, and the load resistance, RL, and the imensi nless figure-of-merit, FOM, defined as [87]: = , (26)”
BackgroundMethodInfluentialGabriel, S., Lau, R.-W.-M., & Gabriel, C. (1996) · Physics in Medicine and Biology · doi:10.1088/0031-9155/41/11/002 · Cited by 4,487 · cited 2×
“The optimized PTE, η, can be approximated using the dielectric tissue dispersion of the Cole-Cole relaxation model [55,57]: η = RL RL + RT (1− |τ|2) (1)”
MethodGabriel, S., Lau, R. W., & Gabriel, C. (1996) · Physics in Medicine and Biology · doi:10.1088/0031-9155/41/11/003 · Cited by 4,175 · cited 2×
“ro and ∞ are the relative permittivity values of the tissue at dc and infinite frequencies and can be found in [159].”
BackgroundPhysical Limitations of Omni‐Directional Antennas
Chu, L. (1948) · doi:10.1063/1.1715038 · Cited by 2,267
A piezoelectric vibration based generator for wireless electronics
Roundy, S., & Wright, P. (2004) · Smart Materials and Structures · doi:10.1088/0964-1726/13/5/018 · Cited by 2,017 · cited 1×
“Research on mid-field [37], acoustic piezoelectric vibration-based [38] and optical [39] WPT techniques in the 2000s have also introduced the possibility of miniaturized WPT RX coils for deep implants.”
BackgroundSimple accurate expressions for planar spiral inductances
Mohan, S. S., Hershenson, M. D. M., Boyd, S., & Lee, T. H. (1999) · IEEE Journal of Solid-State Circuits · doi:10.1109/4.792620 · Cited by 1,627 · cited 2×
“The coefficients Ci are layoutdependent as shown in Table 1 [73].”
BackgroundEfficient wireless non-radiative mid-range energy transfer
Karalis, A., Joannopoulos, J., & Soljačić, M. (2006) · Annals of Physics · doi:10.1016/j.aop.2007.04.017 · Cited by 1,498 · cited 3×
“where QL (=2πfLRX/RL) [15,87,88] is the loaded Q-factor calculated by using the inductance of the RX coil, LRX, and the load resistance, RL, and the imensi nless figure-of-merit, FOM, defined as [87]: = , (26)”
BackgroundMethodDesign considerations for a contactless electric vehicle battery charger
Wang, C.-S., Stielau, O., & Covic, G. (2005) · IEEE transactions on industrial electronics (1982. Print) · doi:10.1109/tie.2005.855672 · Cited by 1,474
A Critical Review of Recent Progress in Mid-Range Wireless Power Transfer
Hui, S., Zhong, W., & Lee, C. (2014) · IEEE transactions on power electronics · doi:10.1109/tpel.2013.2249670 · Cited by 1,373 · cited 1×
“Furthermore, the authors in [42,43] described the theory behind mid-field WPT and its recent improvements.”
BackgroundCharacterization of Polydimethylsiloxane (PDMS) Properties for Biomedical Micro/Nanosystems
Mata, Á., Fleischman, A. J., & Roy, S. (2005) · Biomedical Microdevices · doi:10.1007/s10544-005-6070-2 · Cited by 1,321
Wang, C.-S., Covic, G., & Stielau, O. (2004) · IEEE transactions on industrial electronics (1982. Print) · doi:10.1109/tie.2003.822038 · Cited by 1,144
Wireless Charging Technologies: Fundamentals, Standards, and Network Applications
Lu, X., Wang, P., Niyato, D. T., Kim, D. I., & Han, Z. (2015) · IEEE Communications Surveys & Tutorials · doi:10.1109/comst.2015.2499783 · Cited by 1,059 · cited 1×
“In [41], a generalized and fundamental overview of WPT was presented.”
BackgroundRamRakhyani, A. K., Mirabbasi, S., & Chiao, M. (2010) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2010.2072782 · Cited by 950 · cited 7×
“In [67,140,141] and [137,142], L is calculated using Equations (6) and (8), and Mij is determined using (9) and (11).”
BackgroundMethodInfluentialRobust wireless power transfer using a nonlinear parity–time-symmetric circuit
Assawaworrarit, S., Yu, X., & Fan, S. (2017) · Nature · doi:10.1038/nature22404 · Cited by 834 · cited 1×
“Some attempts have been reported to address the self-tuning of the NRIC WPT resonators [125,126].”
BackgroundWheeler, H. A. (1942) · Proceedings of the IRE · doi:10.1109/jrproc.1942.232015 · Cited by 759 · cited 1×
“The commonly used formula for AC resistance of a WWC is given as [75,81,82]:”
MethodOptimal Design of ICPT Systems Applied to Electric Vehicle Battery Charge
Sallán, J., Villa, J., Llombart, A., & Sanz, J. (2009) · IEEE transactions on industrial electronics (1982. Print) · doi:10.1109/tie.2009.2015359 · Cited by 731
Jow, U.-M., & Ghovanloo, M. (2007) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2007.913130 · Cited by 628 · cited 7×
“Its performance is commonly limited by the skin effect and proximity effect resistances [36,69,72], large separation distance between TX and RX, translational and angular misalignment of the coils and tissue losses.”
BackgroundMethodInfluentialStability and control of inductively coupled power transfer systems
Boys, J., Covic, G., & Green, A. W. (2000) · IEE Proceedings - Electric Power Applications · doi:10.1049/ip-epa:20000017 · Cited by 588 · cited 1×
“Used on the 1970s for high-frequency heating equipment [21,22], WPT systems were applied for short-range applications, such as inductive power transfer systems in the 1990s [21–25] and wireless charging systems for portable equipment in the 2000s [26–30].”
BackgroundDesign and Optimization of a 3-Coil Inductive Link for Efficient Wireless Power Transmission
Kiani, M., Jow, U.-M., & Ghovanloo, M. (2011) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2011.2158431 · Cited by 560 · cited 14×
“The PTE of four-coil-based NRMRC WPT system using LCT is given as [137,142,144]: Sensors 2020, 20, 3487 22 of 58 η4−coil = ( k212Q1Q2 )( k223Q2Q3 )( k234Q3Q4 ) ( 1 + k212Q1Q2 )( 1 + k234Q3Q4 ) + k223Q2Q3 ( 1 + k223Q2Q3+k234Q3Q4 ) (32) The PTE of four-coil-based NRMRC WPT system using RLT can be written as [67,75,82,14…”
BackgroundMethodInfluentialWireless Power Transfer Strategies for Implantable Bioelectronics
Agarwal, K., Jegadeesan, R., Guo, Y., & Thakor, N. V. (2017) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2017.2683520 · Cited by 549 · cited 3×
“In [35], different WPT techniques and their design issues are reviewed.”
BackgroundInfluentialWireless power transfer to deep-tissue microimplants
Ho, J. S., Yeh, A. J., Neofytou, E., Kim, S., Tanabe, Y., Patlolla, B., Beygui, R. E., & Poon, A. S. Y. (2014) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1403002111 · Cited by 541 · cited 4×
“NRRMF 2014/[156] Cardiac Diameter: 2 mm 50 1600 0.”
BackgroundMethodInfluentialHui, S., & Ho, W. (2004) · IEEE transactions on power electronics · doi:10.1109/tpel.2005.846550 · Cited by 522
Collin, R., & Rothschild, S. (1964) · doi:10.1109/tap.1964.1138151 · Cited by 508
A contactless electrical energy transmission system for portable-telephone battery chargers
Jang, Y., & Jovanovic, M. (2000) · INTELEC. Twenty-Second International Telecommunications Energy Conference (Cat. No.00CH37131) · doi:10.1109/intlec.2000.884329 · Cited by 501
Wireless power transfer and applications to sensor networks
Xie, L.-G., Shi, Y., Hou, Y. T., & Lou, A. (2013) · IEEE wireless communications · doi:10.1109/mwc.2013.6590061 · Cited by 498 · cited 1×
“This technology is extensively used in a wide range of applications ranging from sophisticated low-power biomedical implants [3–6], to high-power electric vehicles [7–10] to white goods such as electric toothbrushes and mobile phones [11].”
BackgroundA re-examination of the fundamental limits on the radiation Q of electrically small antennas
Mclean, J. (1996) · doi:10.1109/8.496253 · Cited by 483
Optimal Frequency for Wireless Power Transmission Into Dispersive Tissue
Poon, A. S. Y., O'Driscoll, S., & Meng, T. H. (2010) · IEEE Transactions on Antennas and Propagation · doi:10.1109/tap.2010.2044310 · Cited by 469 · cited 3×
“However, reports of proper link modeling and characterization of the effects of ultrasound wave inside the human body are scarce compared to the EM WPT techniques [157,208].”
BackgroundMethodGeometric approach for coupling enhancement of magnetically coupled coils
Zierhofer, C. M., & Hochmair, E. S. (1996) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.503178 · Cited by 466 · cited 8×
“For a multi-loop, multi-layer WWC, the self-inductance is the combination of the self-inductances of all the loops and layers and the mutual inductances among different loops and layers [77,78].”
BackgroundMethodInfluentialDesign and analysis of an adaptive transcutaneous power telemetry for biomedical implants
Wang, G.-X., Liu, W.-T., Sivaprakasam, M., & Kendir, G. (2005) · IEEE Transactions on Circuits and Systems Part 1: Regular Papers · doi:10.1109/tcsi.2005.852923 · Cited by 443
The Circuit Theory Behind Coupled-Mode Magnetic Resonance-Based Wireless Power Transmission
Kiani, M., & Ghovanloo, M. (2012) · IEEE Transactions on Circuits and Systems Part 1: Regular Papers · doi:10.1109/tcsi.2011.2180446 · Cited by 423 · cited 15×
“The PTE of four-coil-based NRMRC WPT system using LCT is given as [137,142,144]: Sensors 2020, 20, 3487 22 of 58 η4−coil = ( k212Q1Q2 )( k223Q2Q3 )( k234Q3Q4 ) ( 1 + k212Q1Q2 )( 1 + k234Q3Q4 ) + k223Q2Q3 ( 1 + k223Q2Q3+k234Q3Q4 ) (32) The PTE of four-coil-based NRMRC WPT system using RLT can be written as [67,75,82,14…”
BackgroundMethodInfluentialDesign of a contactless battery charger for cellular phone
Kim, C.-G., Seo, D., You, J., Park, J.-H., & Cho, B. (2000) · APEC 2000. Fifteenth Annual IEEE Applied Power Electronics Conference and Exposition (Cat. No.00CH37058) · doi:10.1109/apec.2000.822591 · Cited by 397
Design of radio-frequency powered coils for implant instruments
Ko, W. H., Liang, S. P., & Fung, C. D. F. (1977) · Medical & Biological Engineering & Computing · doi:10.1007/bf02457921 · Cited by 374
da Cruz, L., Coley, B. F., Dorn, J., Merlini, F., Filley, E., Christopher, P., Chen, F. K., Wuyyuru, V., Sahel, J., Stanga, P., et al. (2013) · British Journal of Ophthalmology · doi:10.1136/bjophthalmol-2012-301525 · Cited by 359 · cited 2×
“A commercial retinal prosthesis system [114,115], ARGUS II, developed by Second Sight Medical Products, is presented in Figure 18a,b.”
MethodWireless Power Transmission: From Far Field to Near Field
Garnica, J., Chinga, R., & Lin, J.-S. (2013) · Proceedings of the IEEE · doi:10.1109/jproc.2013.2251411 · Cited by 351 · cited 1×
“Wireless power transfer (WPT) can be defined as a technology capable of transmitting energy across a medium, from a power source to an electrical load, without the use of electrical wires connecting this power source to the load [1,2].”
BackgroundA Study of Loosely Coupled Coils for Wireless Power Transfer
Chen, C.-J., Chu, T.-H., Lin, C.-L., & Jou, Z.-C. (2010) · IEEE Transactions on Circuits and Systems - II - Express Briefs · doi:10.1109/tcsii.2010.2048403 · Cited by 351 · cited 1×
“The driver and load coils can be used for impedance matching [139].”
BackgroundMaximum Achievable Efficiency in Near-Field Coupled Power-Transfer Systems
Zargham, M., & Gulak, P. (2012) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2011.2174794 · Cited by 336
Modeling of Mutual Coupling Between Planar Inductors in Wireless Power Applications
Raju, S., Wu, R.-X., Chan, M., & Yue, C. (2014) · IEEE transactions on power electronics · doi:10.1109/tpel.2013.2253334 · Cited by 331 · cited 3×
“Another approach to calculate Mij stems from the approximation of the Neumann’s equation [80]:”
BackgroundMethod10 kHz inductively coupled power transfer-concept and control
Green, A. W., & Boys, J. (1994) · International Conference on Power Electronics and Variable Speed Drives (PEVD) · doi:10.1049/cp:19941049 · Cited by 325
Liu, C.-R., Guo, Y.-X., & Xiao, S. (2014) · IEEE Transactions on Antennas and Propagation · doi:10.1109/tap.2014.2307341 · Cited by 292
Metamaterials-based enhanced energy harvesting: A review
Chen, Z.-S., Guo, B., Yang, Y.-M., & Cheng, C.-C. (2014) · Physica B Condensed Matter · doi:10.1016/j.physb.2013.12.040 · Cited by 291 · cited 1×
“The possibility of using metamaterials for WPT was reviewed in [45,46].”
BackgroundCapacitive power transfer for contactless charging
Kline, M. H., Izyumin, I. I., Boser, B. E., & Sanders, S. R. (2011) · Conference proceedings/Conference proceedings - IEEE Applied Power Electronics Conference and Exposition · doi:10.1109/apec.2011.5744775 · Cited by 286 · cited 1×
“Furthermore, non-radiative techniques include inductive coupling, which includes witricity [14], magnetic resonance coupling [15] and capacitive coupling [16].”
MethodPower harvesting and telemetry in CMOS for implanted devices
Sauer, C. M., Stanaćević, M., Cauwenberghs, G., & Thakor, N. (2004) · IEEE Transactions on Circuits and Systems Part 1: Regular Papers · doi:10.1109/tcsi.2005.858183 · Cited by 281
Huang, F.-J., Lee, C.-M., Chang, C.-L., Chen, L.-K., Yo, T.-C., & Luo, C.-H. (2011) · IEEE Transactions on Antennas and Propagation · doi:10.1109/tap.2011.2152317 · Cited by 265
Designing Efficient Inductive Power Links for Implantable Devices
Harrison, R. (2007) · 2007 IEEE International Symposium on Circuits and Systems · doi:10.1109/iscas.2007.378508 · Cited by 263
Magnetic MIMO: how to charge your phone in your pocket
Jadidian, J., & Katabi, D. (2014) · ACM/IEEE International Conference on Mobile Computing and Networking · doi:10.1145/2639108.2639130 · Cited by 252
Review of Near-Field Wireless Power and Communication for Biomedical Applications
Kim, H., Hirayama, H., Kim, S., Han, K. J., Zhang, R., & Choi, J.-W. (2017) · IEEE Access · doi:10.1109/access.2017.2757267 · Cited by 252 · cited 2×
“Recently, in [12,51], the theoretical aspects of non-radiative near-field WPT were examined in details.”
BackgroundExperimental study of large rectenna array for microwave energy transmission
Shinohara, N., & Matsumoto, H. (1998) · doi:10.1109/22.661713 · Cited by 250
Modeling and optimization of printed spiral coils in air, saline, and muscle tissue environments
Jow, U.-M., & Ghovanloo, M. (2009) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2009.2025366 · Cited by 241 · cited 5×
“Its performance is commonly limited by the skin effect and proximity effect resistances [36,69,72], large separation distance between TX and RX, translational and angular misalignment of the coils and tissue losses.”
BackgroundInfluentialUltrasonic transcutaneous energy transfer for powering implanted devices.
Ozeri, S., & Shmilovitz, D. (2010) · Ultrasonics · doi:10.1016/j.ultras.2009.11.004 · Cited by 241
Choi, B., Nho, J., Cha, H., Ahn, T., & Choi, S.-C. (2004) · IEEE transactions on industrial electronics (1982. Print) · doi:10.1109/tie.2003.822039 · Cited by 239
A mm-Sized Wireless Implantable Device for Electrical Stimulation of Peripheral Nerves
Charthad, J., Chang, T. C., Liu, Z., Sawaby, A., Weber, M. J., Baker, S., Gore, F., Felt, S. A., & Arbabian, A. (2018) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2018.2799623 · Cited by 230
A Critical Review of Wireless Power Transfer via Strongly Coupled Magnetic Resonances
Wei, X., Wang, Z., & Dai, H. (2014) · Energies · doi:10.3390/en7074316 · Cited by 229 · cited 7×
“In [75], a comprehensive analysis of the method concluded that three coils could increase power transfer efficiency compared to four coils and simultaneously keep the transfer efficiency constant over changes in the transferred power [63].”
BackgroundMethodInfluentialElectromagnetic Energy Harvesting and Wireless Power Transmission: A Unified Approach
Costanzo, A., Dionigi, M., Masotti, D., Mongiardo, M., Monti, G., Tarricone, L., & Sorrentino, R. (2014) · Proceedings of the IEEE · doi:10.1109/jproc.2014.2355261 · Cited by 229 · cited 1×
“Wireless power transfer (WPT) can be defined as a technology capable of transmitting energy across a medium, from a power source to an electrical load, without the use of electrical wires connecting this power source to the load [1,2].”
BackgroundFully Wireless Implantable Cardiovascular Pressure Monitor Integrated with a Medical Stent
Chow, E. Y., Chlebowski, A. L., Chakraborty, S., Chappell, W. J., & Irazoqui, P. P. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2010.2041058 · Cited by 223
Midfield Wireless Powering for Implantable Systems
Ho, J. S., Kim, S., & Poon, A. S. Y. (2013) · Proceedings of the IEEE · doi:10.1109/jproc.2013.2251851 · Cited by 223 · cited 2×
“However, the level of miniaturization required for NRIC and NRMRC implantable systems affects the PTE performance at these low frequencies [13,154], especially when the implant is much smaller than its separation distance from the TX coil [155].”
BackgroundMethodDesign and Safety Considerations of an Implantable Rectenna for Far-Field Wireless Power Transfer
Liu, C.-R., Guo, Y.-X., Sun, H.-C., & Xiao, S. (2014) · IEEE Transactions on Antennas and Propagation · doi:10.1109/tap.2014.2352363 · Cited by 222 · cited 6×
“The re eived power, PRX, can be calculated using Friis radio link formula as [169]:”
BackgroundMethodInfluentialLong-term implants of Parylene-C coated microelectrodes
Schmidt, E., Mcintosh, J. S., & Bak, M. (2006) · Medical and Biological Engineering and Computing · doi:10.1007/bf02441836 · Cited by 216
The Development of the Nucleus® Freedom™ Cochlear Implant System
Patrick, J., Busby, P., & Gibson, P. (2006) · Trends in Amplification · doi:10.1177/1084713806296386 · Cited by 214
Wireless Power Transfer: Metamaterials and Array of Coupled Resonators
Wang, B., Yerazunis, W. S., & Teo, K. H. (2013) · Proceedings of the IEEE · doi:10.1109/jproc.2013.2245611 · Cited by 211 · cited 1×
“The possibility of using metamaterials for WPT was reviewed in [45,46].”
BackgroundAn inductive power link for a wireless endoscope.
Lenaerts, B., & Puers, R. (2007) · Biosensors & bioelectronics · doi:10.1016/j.bios.2006.06.015 · Cited by 210 · cited 2×
“021 2007/[99] Capsule 10 mm × 13 mm 1 1 Air 0.”
BackgroundAn inductive power system with integrated bi-directional data-transmission
Catrysse, M., Hermans, B., & Puers, R. (2004) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2004.02.016 · Cited by 205
Ho, S. L., Wang, J., Fu, W. N., & Sun, M. (2011) · IEEE Transactions on Magnetics · doi:10.1109/tmag.2010.2091495 · Cited by 204 · cited 1×
“Furthermore, non-radiative techniques include inductive coupling, which includes witricity [14], magnetic resonance coupling [15] and capacitive coupling [16].”
MethodLow-Power Wireless Power Delivery
Falkenstein, E., Roberg, M., & Popović, Z. (2012) · IEEE transactions on microwave theory and techniques · doi:10.1109/tmtt.2012.2193594 · Cited by 204 · cited 3×
“ethodology is proposed in [164] to deter ine the opti al i pedance of the rectifier in order to accurately design and easure the output of the rectifier circuit for RFF.”
BackgroundMethodVilla, J., Sallán, J., Llombart, A., & Sanz, J. (2009) · Applied Energy · doi:10.1016/j.apenergy.2008.05.009 · Cited by 203
Inductor Modeling in Wireless Links for Implantable Electronics
Yang, Z., Liu, W.-T., & Basham, E. (2007) · IEEE transactions on magnetics · doi:10.1109/tmag.2007.904189 · Cited by 198 · cited 3×
“Another approach is to calculate iteratively the AC resistance of WWC using [65]:”
BackgroundThe functional performance of the Argus II retinal prosthesis
Stronks, H. C., & Dagnelie, G. (2014) · Expert Review of Medical Devices · doi:10.1586/17434440.2014.862494 · Cited by 189 · cited 2×
“A commercial retinal prosthesis system [114,115], ARGUS II, developed by Second Sight Medical Products, is presented in Figure 18a,b.”
MethodWireless Power Delivery to Flexible Subcutaneous Implants Using Capacitive Coupling
Jegadeesan, R., Agarwal, K., Guo, Y., Yen, S.-C., & Thakor, N. V. (2016) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2016.2615623 · Cited by 189 · cited 3×
“The optimized PTE, η, can be approximated using the dielectric tissue dispersion of the Cole-Cole relaxation model [55,57]: η = RL RL + RT (1− |τ|2) (1)”
MethodMED-EL Cochlear Implants: State of the Art and a Glimpse Into the Future
Hochmair, I. J., Nopp, P., Jolly, C., Schmidt, M., Schößer, H., Garnham, C., & Anderson, I. (2006) · Trends in Amplification · doi:10.1177/1084713806296720 · Cited by 187 · cited 6×
“Figure 13 shows the first FDA approved spinal cord stimulation implant (Freedom SystemTM) by StimWave.”
InfluentialAn Overview of Resonant Circuits for Wireless Power Transfer
Jiang, C.-Q., Chau, K., Liu, C.-H., & Lee, C. H. T. (2017) · Energies · doi:10.3390/en10070894 · Cited by 185 · cited 1×
“An extensive review of different TX circuit topologies for NRIC is presented in [95].”
BackgroundUltrasonic vs. Inductive Power Delivery for Miniature Biomedical Implants
Denisov, A., & Yeatman, E. (2010) · 2010 International Conference on Body Sensor Networks · doi:10.1109/bsn.2010.27 · Cited by 185
Interphase Mutual Inductance in Polyphase Inductive Power Transfer Systems
Kissin, M. L. G., Boys, J., & Covic, G. (2009) · IEEE transactions on industrial electronics (1982. Print) · doi:10.1109/tie.2009.2020076 · Cited by 175
A New Concept: Asymmetrical Pick-Ups for Inductively Coupled Power Transfer Monorail Systems
Elliott, G. A. J., Covic, G., Kacprzak, D., & Boys, J. (2006) · IEEE transactions on magnetics · doi:10.1109/tmag.2006.879619 · Cited by 173
Swallowable Wireless Capsule Endoscopy: Progress and Technical Challenges
Pan, G., & Wang, L.-T. (2011) · Gastroenterology Research and Practice · doi:10.1155/2012/841691 · Cited by 168 · cited 1×
“In [49,50], the possibility and few applications of WPT for capsule endoscopy were demonstrated.”
BackgroundCapacitive coupling for power and data telemetry to implantable biomedical microsystems
Sodagar, A. M., & Amiri, P. (2009) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner.2009.5109320 · Cited by 164 · cited 2×
“Non-radiative capacitive coupling (NRCC) of IMDs was first demonstrated for the powering of subcutaneous implants in [52] and was later extended to flexible implants.”
BackgroundSystem Optimization for Improved Accuracy in Transcutaneous Signal and Power Transmission
Hochmair, E. (1984) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1984.325404 · Cited by 163
Sherrit, S., Leary, S., Dolgin, B., & Bar-Cohen, Y. (1999) · 1999 IEEE Ultrasonics Symposium. Proceedings. International Symposium (Cat. No.99CH37027) · doi:10.1109/ultsym.1999.849139 · Cited by 161
An implantable power supply with an optically rechargeable lithium battery
Goto, K., Nakagawa, T., Nakamura, O., & Kawata, S. (2001) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.930908 · Cited by 159 · cited 2×
“Optical power transfer (OPT) is one of the least studied techniques to transfer power wirelessly to a biomedical implant [19,39].”
BackgroundLiu, X., & Hui, S. (2007) · IEEE transactions on power electronics · doi:10.1109/tpel.2007.909301 · Cited by 159
Model validation of untethered, ultrasonic neural dust motes for cortical recording
Seo, D., Carmena, J. M., Rabaey, J. M., Maharbiz, M. M., & Alon, E. (2015) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2014.07.025 · Cited by 158
A mm-sized implantable power receiver with adaptive link compensation
O'Driscoll, S., Poon, A. S. Y., & Meng, T. (2009) · IEEE International Solid-State Circuits Conference · doi:10.1109/isscc.2009.4977424 · Cited by 155 · cited 5×
“Midfield WPT, characterized by a separation distance of 100 to 500 mm, lies in between the non-radiative and radiative regions [13].”
BackgroundMethodInfluentialAn Equivalent Complex Permeability Model for Litz-Wire Windings
Nan, X., & Sullivan, C. R. (2005) · IEEE transactions on industry applications · doi:10.1109/tia.2009.2013594 · Cited by 153 · cited 1×
“The inner diameter of the TX is 41 cm and is made of rectangular Litz cable [120].”
BackgroundA Modular 32-site wireless neural stimulation microsystem
Ghovanloo, M., & Najafi, K. (2004) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2004.837026 · Cited by 149
Sun, T., Xie, X., Li, G., Gu, Y., Deng, Y., & Wang, Z.-H. (2012) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2012.2206809 · Cited by 148 · cited 3×
“6% [103], 87% [104] and 86% [105] efficiencies for MIDs.”
MethodA new IEEE standard for safety levels with respect to human exposure to radio-frequency radiation
Lin, J. C. (2006) · IEEE Antennas and Propagation Magazine · doi:10.1109/map.2006.1645601 · Cited by 146 · cited 1×
“(4) The large power requirement of some multi-functional implantable devices might prohibit the use of a WPT system that could generate an SAR above 2 W/kg, the recommended limit regarding safety of human tissues [117,127].”
BackgroundA Figure-of-Merit for Designing High-Performance Inductive Power Transmission Links
Kiani, M., & Ghovanloo, M. (2013) · IEEE Transactions on Industrial Electronics · doi:10.1109/tie.2012.2227914 · Cited by 146 · cited 8×
“The PTE of four-coil-based NRMRC WPT system using LCT is given as [137,142,144]: Sensors 2020, 20, 3487 22 of 58 η4−coil = ( k212Q1Q2 )( k223Q2Q3 )( k234Q3Q4 ) ( 1 + k212Q1Q2 )( 1 + k234Q3Q4 ) + k223Q2Q3 ( 1 + k223Q2Q3+k234Q3Q4 ) (32) The PTE of four-coil-based NRMRC WPT system using RLT can be written as [67,75,82,14…”
MethodInfluentialAn Ultrasonically Powered Implantable Micro-Oxygen Generator (IMOG)
Maleki, T., Cao, N., Song, S., Kao, C. H., Ko, S., & Ziaie, B. (2011) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2011.2163634 · Cited by 144
Contactless Energy Transfer Systems Using Antiparallel Resonant Loops
Lee, W.-S., Son, W., Oh, K., & Yu, J.-W. (2013) · IEEE transactions on industrial electronics (1982. Print) · doi:10.1109/tie.2011.2177611 · Cited by 144 · cited 1×
“Recent research regarding the use of antiparallel resonant loops [133,134] demonstrate a higher PTE at a larger separation distance.”
BackgroundDas, R., & Yoo, H. (2017) · IEEE transactions on microwave theory and techniques · doi:10.1109/tmtt.2017.2647945 · Cited by 142
Lumped parameter models for single- and multiple-layer inductors
Massarini, A., Kazimierczuk, M. K., & Grandi, G. (1996) · PESC Record. 27th Annual IEEE Power Electronics Specialists Conference · doi:10.1109/pesc.1996.548595 · Cited by 137 · cited 1×
“where ς is the thickness of the insulating wire, ro is the radius of the conducting section of the wire, Dx is the average diameter of the insulating coating shell given by (Din + Dc)/2 [83].”
BackgroundLiu, M., Fu, M.-F., & Ma, C. (2016) · IEEE transactions on power electronics · doi:10.1109/tpel.2015.2472565 · Cited by 132 · cited 2×
“A detailed analysis and optimization of the parameters is provided in [96].”
BackgroundMethodRobust Wireless Power Transmission to mm-Sized Free-Floating Distributed Implants
Mirbozorgi, S. A., Yeon, P., & Ghovanloo, M. (2017) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2017.2663358 · Cited by 131 · cited 3×
“For very small-sized RX coils, a high Q-factor intermediate resonator can significantly improve the PTE, which is explained analytically in [145].”
BackgroundInfluentialNa, K., Jang, H., Ma, H., & Bien, F. (2015) · IEEE transactions on microwave theory and techniques · doi:10.1109/tmtt.2014.2365475 · Cited by 130 · cited 4×
“Figure 27a shows a four-coil WPT system for CE [148].”
BackgroundMethodInfluentialOn the Efficiency of Far-Field Wireless Power Transfer
Xia, M.-H., & Aïssa, S. (2015) · IEEE Transactions on Signal Processing · doi:10.1109/tsp.2015.2417497 · Cited by 129 · cited 3×
“The end to end power transfer efficiency of the RFF WPT system can be expressed as [167]:”
BackgroundMethodA Smart Multicoil Inductively Coupled Array for Wireless Power Transmission
Mirbozorgi, S. A., Bahrami, H., Sawan, M., & Gosselin, B. J. (2014) · IEEE Transactions on Industrial Electronics · doi:10.1109/tie.2014.2308138 · Cited by 127
Efficient weakly-radiative wireless energy transfer: An EIT-like approach
Hamam, R. E., Karalis, A., Joannopoulos, J. D., & Marin Soljacic, A. (2009) · Annals of Physics · doi:10.1016/j.aop.2009.05.005 · Cited by 126
Misalignment Sensitivity of Strongly Coupled Wireless Power Transfer Systems
Liu, D., Hu, H., & Georgakopoulos, S. V. (2017) · IEEE transactions on power electronics · doi:10.1109/tpel.2016.2605698 · Cited by 122
Torp, H. (1976) · Springer US · doi:10.1007/978-1-4613-4307-3 · Cited by 122
Wireless Power Transfer to Miniature Implants: Transmitter Optimization
Kim, S., Ho, J. S., & Poon, A. S. Y. (2012) · IEEE Transactions on Antennas and Propagation · doi:10.1109/tap.2012.2207341 · Cited by 121 · cited 6×
“The maximum PTE is independent of the antenna structure and relies solely on the maximization of the coupling parameter γ [158]: γ = |Z21|(2) R11 (35)”
BackgroundMethodInfluentialA Planar Magnetically Coupled Resonant Wireless Power Transfer System Using Printed Spiral Coils
Jolani, F., Yu, Y.-Q., & Chen, Z. (2014) · IEEE Antennas and Wireless Propagation Letters · doi:10.1109/lawp.2014.2349481 · Cited by 118 · cited 3×
“The optimization of NRMRC follows a similar methodology as the one presented in Figure 5 [67,75,82,137,140,141].”
MethodCut the Cord: Low-Power Far-Field Wireless Powering
Popović, Z. (2013) · IEEE Microwave Magazine · doi:10.1109/mmm.2012.2234638 · Cited by 113 · cited 1×
“In [44], the challenges, possible design procedures and future of far-field WPT were examined.”
BackgroundImaging and steering an optical wireless nanoantenna link
Drégely, D., Lindfors, K., Lippitz, M., Engheta, N., Totzeck, M., & Giessen, H. (2014) · Nature Communications · doi:10.1038/ncomms5354 · Cited by 112
Chow, E., Morris, M., & Irazoqui, P. (2013) · IEEE Microwave Magazine · doi:10.1109/mmm.2013.2248586 · Cited by 111 · cited 1×
“Radio frequency (RF) modeling, different frequency bands and a brief study on the RF implantable device testing inside phantoms, ex vivo and in vivo were presented in [47].”
BackgroundAccurate Modeling of Coil Inductance for Near-Field Wireless Power Transfer
Khan, S. R., Pavuluri, S. K., & Desmulliez, M. P. Y. (2018) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2018.2854190 · Cited by 107 · cited 1×
“However, recent research on 3D WPT coils demonstrated it is a promising solution for the misalignment issue of NRIC [80,135].”
BackgroundSelf-citationApplication of Wireless Power Transmission Systems in Wireless Capsule Endoscopy: An Overview
Basar, M. R., Ahmad, M. Y., Cho, J., & Ibrahim, F. (2014) · Italian National Conference on Sensors · doi:10.3390/s140610929 · Cited by 102 · cited 1×
“In [49,50], the possibility and few applications of WPT for capsule endoscopy were demonstrated.”
BackgroundOzeri, S., Shmilovitz, D., Singer, S., & Wang, C.-C. (2010) · Ultrasonics · doi:10.1016/j.ultras.2010.01.004 · Cited by 102
A multi-coil inductive powering system for an endoscopic capsule with vibratory actuation
Carta, R., Sfakiotakis, M., Pateromichelakis, N., Thoné, J., Tsakiris, D., & Puers, R. (2011) · doi:10.1016/j.sna.2011.03.036 · Cited by 97
Scalable high lead-count parylene package for retinal prostheses
Rodger, D. C., Weiland, J. D., Humayun, M. S., & Tai, Y.-C. (2006) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2005.11.010 · Cited by 96
Far-Field RF Powering of Implantable Devices: Safety Considerations
Bercich, R. A., Duffy, D. R., & Irazoqui, P. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2246787 · Cited by 95
Omnidirectional Ultrasonic Powering for Millimeter-Scale Implantable Devices
Song, S.-H., Kim, A., & Ziaie, B. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2444854 · Cited by 95
A Wireless Magnetic Resonance Energy Transfer System for Micro Implantable Medical Sensors
Li, X., Zhang, H., Peng, F., Li, Y., Yang, T., Wang, B., & Fang, D.-M. (2012) · Italian National Conference on Sensors · doi:10.3390/s120810292 · Cited by 90 · cited 2×
“In [67,140,141] and [137,142], L is calculated using Equations (6) and (8), and Mij is determined using (9) and (11).”
MethodOrientation insensitive power transfer by magnetic resonance for mobile devices
Jonah, O., Georgakopoulos, S. V., & Tentzeris, M. (2013) · 2013 IEEE Wireless Power Transfer (WPT) · doi:10.1109/wpt.2013.6556868 · Cited by 89
Liu, H., Shao, Q., & Fang, X.-L. (2017) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2016.2538320 · Cited by 88 · cited 3×
“where XL is the reactance of the load loop [94].”
BackgroundStudy of a wireless power transmission system for an active capsule endoscope
Xin, W., Yan, G., & Wang, W.-X. (2010) · The international journal of medical robotics + computer assisted surgery : MRCAS · doi:10.1002/rcs.298 · Cited by 88 · cited 3×
“3D WPT system for capsule endoscopy (CE) [122].”
BackgroundMethodFaerber, J., Cummins, G., Pavuluri, S., Record, P., Rodriguez, A. R. A., Lay, H. S., McPhillips, R., Cox, B., Connor, C., Gregson, R., et al. (2017) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2017.2759254 · Cited by 87 · cited 3×
“Capsule endoscopy (CE) promises early stage minimally invasive detection and rapid diagnosis of GI diseases [117] to limit the adverse effects of gastrointestinal (GI) disorders [118,119].”
BackgroundSelf-citationMiniaturized Dual-Band Antenna for Implantable Wireless Communications
Xu, L.-J., Guo, Y.-X., & Wu, W. (2014) · IEEE Antennas and Wireless Propagation Letters · doi:10.1109/lawp.2014.2329937 · Cited by 87
A Novel Mat-Based System for Position-Varying Wireless Power Transfer to Biomedical Implants
Xu, Q., Wang, H., Gao, Z., Mao, Z.-H., He, J.-P., & Sun, M. (2013) · IEEE transactions on magnetics · doi:10.1109/tmag.2013.2245335 · Cited by 85 · cited 2×
“Multiple-TX [150– 152] and RX [153]-based NRMRC WPT systems show great potential for future IMDs.”
BackgroundOn the Design of Microfluidic Implant Coil for Flexible Telemetry System
Qusba, A., Ramrakhyani, A. K., So, J.-H., Hayes, G., Dickey, M., & Lazzi, G. (2014) · IEEE Sensors Journal · doi:10.1109/jsen.2013.2293096 · Cited by 83 · cited 1×
“Severe mismatch of the resonance capacitance can occur due to the flexion characteristics of the implant coils [124].”
BackgroundEnabling Wireless Powering and Telemetry for Peripheral Nerve Implants
Jegadeesan, R., Nag, S., Agarwal, K., Thakor, N., & Guo, Y.-X. (2015) · IEEE journal of biomedical and health informatics · doi:10.1109/jbhi.2015.2424985 · Cited by 81 · cited 3×
“Figure 12 shows the stimulator implanted through a small incision made in the stomach region of a rat [111].”
BackgroundMethodInfluentialDiversity Analysis of Multiple Transmitters in Wireless Power Transfer System
Lee, K., & Cho, D. (2013) · IEEE transactions on magnetics · doi:10.1109/tmag.2012.2234132 · Cited by 80
Development of Nonsuperstrate Implantable Low-Profile CPW-Fed Ceramic Antennas
Chien, T.-F., Cheng, C.-M., Yang, H.-C., Jiang, J.-W., & Luo, C.-H. (2010) · IEEE Antennas and Wireless Propagation Letters · doi:10.1109/lawp.2010.2053342 · Cited by 80
Indoor Optical Wireless Power Transfer to Small Cells at Nighttime
Fakidis, J., Videv, S., Kučera, Š., Claußen, H., & Haas, H. (2016) · Journal of Lightwave Technology · doi:10.1109/jlt.2016.2555883 · Cited by 80
Myer, P., Mannalithara, A., Singh, G., Singh, G., Pasricha, P., & Ladabaum, U. (2013) · American Journal of Gastroenterology · doi:10.1038/ajg.2013.199 · Cited by 79 · cited 2×
“Capsule endoscopy (CE) promises early stage minimally invasive detection and rapid diagnosis of GI diseases [117] to limit the adverse effects of gastrointestinal (GI) disorders [118,119].”
BackgroundMutashar, S., Hannan, M. A., Samad, S. A., & Hussain, A. (2014) · Italian National Conference on Sensors · doi:10.3390/s140711522 · Cited by 78 · cited 1×
“As shown in Figure 5, an iterative optimization flow diagram has been adopted in most of the research literature to achieve maximum PTE [36,66,70,75,76].”
MethodA Literature Survey on Wireless Power Transfer for Biomedical Devices
Shadid, R., & Noghanian, S. (2018) · International Journal of Antennas and Propagation · doi:10.1155/2018/4382841 · Cited by 77
Inductive and Ultrasonic Multi-Tier Interface for Low-Power, Deeply Implantable Medical Devices
Sanni, A., Vilches, A., & Toumazou, C. (2012) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2011.2175390 · Cited by 75
Optimal Operating Frequency in Wireless Power Transmission for Implantable Devices
Poon, A. S. Y., O'Driscoll, S., & Meng, T. (2007) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2007.4353634 · Cited by 75 · cited 4×
“The mid-field WPT scheme for biomedical devices was initially proposed for a 2 × 2 mm2square loop RX implant antenna operating at 915 MHz [13,37].”
BackgroundMethodInfluentialBackscatter Communications: Inception of the Battery-Free Era—A Comprehensive Survey
Memon, M. L., Saxena, N., Roy, A., & Shin, D. R. (2019) · Electronics · doi:10.3390/electronics8020129 · Cited by 75
Inductive Power Transfer System With Self-Calibrated Primary Resonant Frequency
Trigui, A., Hached, S., Mounaim, F., Ammari, A., & Sawan, M. (2015) · IEEE transactions on power electronics · doi:10.1109/tpel.2015.2399417 · Cited by 74 · cited 1×
“Some attempts have been reported to address the self-tuning of the NRIC WPT resonators [125,126].”
BackgroundVibration therapy: clinical applications in bone
Thompson, W. R., Yen, S. S., & Rubin, J. (2014) · Current Opinion in Endocrinology, Diabetes & Obesity · doi:10.1097/med.0000000000000111 · Cited by 74
A Hybrid Patch/Slot Implantable Antenna for Biotelemetry Devices
Liu, C.-R., Guo, Y.-X., & Xiao, S. (2012) · IEEE Antennas and Wireless Propagation Letters · doi:10.1109/lawp.2013.2237879 · Cited by 74
An Analytical Model for Inductively Coupled Implantable Biomedical Devices With Ferrite Rods
Theilmann, P., & Asbeck, P. (2009) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2008.2004776 · Cited by 73 · cited 1×
“Furthermore, alternative approaches, such as 3D orthogonal structures [99,129–131] and ferrite materials [132] have attracted the attention of researchers as a means of improving the PTE.”
BackgroundPosition and Orientation Insensitive Wireless Power Transmission for EnerCage-Homecage System
Jia, Y., Mirbozorgi, S. A., Wang, Z., Hsu, C.-C., Madsen, T. E., Rainnie, D. G., & Ghovanloo, M. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2017.2691720 · Cited by 68
On-Chip Implantable Antennas for Wireless Power and Data Transfer in a Glaucoma-Monitoring SoC
Marnat, L., Ouda, M. H., Arsalan, M., Salama, K., & Shamim, A. (2012) · IEEE Antennas and Wireless Propagation Letters · doi:10.1109/lawp.2013.2240253 · Cited by 66
Atluri, S., & Ghovanloo, M. (2005) · Conference Proceedings. 2nd International IEEE EMBS Conference on Neural Engineering, 2005. · doi:10.1109/cne.2005.1419677 · Cited by 64 · cited 4×
“For a multi-loop, multi-layer WWC, the self-inductance is the combination of the self-inductances of all the loops and layers and the mutual inductances among different loops and layers [77,78].”
BackgroundInfluentialHigh-performance wireless powering for peripheral nerve neuromodulation systems
Tanabe, Y., Ho, J. S., Liu, J., Liao, S.-Y., Zhen, Z., Hsu, S., Shuto, C., Zhu, Z.-Y., Ma, A., Vassos, C., et al. (2017) · PLOS ONE · doi:10.1371/journal.pone.0186698 · Cited by 63 · cited 6×
“Neurostimulator Implant Figure 33a shows the design of a WPT conformal transmitter consisting of reactively loaded rings laser-cut from copper film and encapsulated in soft silicon [160].”
BackgroundInfluentialA 10.5 cm Ultrasound Link for Deep Implanted Medical Devices
Mazzilli, F., Lafon, C., & Dehollain, C. (2014) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2013.2295403 · Cited by 63
Multicoil Telemetry System for Compensation of Coil Misalignment Effects in Implantable Systems
Ramrakhyani, A. K., & Lazzi, G. (2012) · IEEE Antennas and Wireless Propagation Letters · doi:10.1109/lawp.2013.2238993 · Cited by 63 · cited 3×
“High (possible to implement in commercial IMDs) 2012/[146] Ocular Diameter: 15 mm 10 3.”
BackgroundMethodInfluentialResonance-based wireless power delivery for implantable devices
Kumar, A., Mirabbasi, S., & Mu, C. (2009) · 2009 IEEE Biomedical Circuits and Systems Conference · doi:10.1109/biocas.2009.5372092 · Cited by 63 · cited 1×
“The PTE of four-coil-based NRMRC WPT system using LCT is given as [137,142,144]:”
MethodManoufali, M., Bialkowski, K., Mohammed, B., Mills, P. C., & Abbosh, A. (2018) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2017.2778729 · Cited by 62 · cited 6×
“eas re e t of the proposed NRIC WPT link for a piglet [106].”
BackgroundInfluentialSun, G.-L., Muneer, B., Li, Y., & Zhu, Q. (2018) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2017.2787649 · Cited by 58 · cited 1×
“The demand for miniaturized MIDs has led to an increase in research on the development of chip-scale-based [40,128] NRIC where the WPT coils are embedded with the integrated circuits (ICs).”
BackgroundMiniaturized 3-D Cross-Type Receiver for Wirelessly Powered Capsule Endoscopy
Khan, S. R., Pavuluri, S. K., Cummins, G., & Desmulliez, M. P. Y. (2019) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2019.2893204 · Cited by 58 · cited 1×
“Therefore, recent research has focused on the 3D orthogonal WPT receiver architecture to mitigate the misalignment effect [123].”
BackgroundSelf-citationInductive Power Link for a Wireless Cortical Implant With Two-Body Packaging
Silay, K. M., Dehollain, C., & Declercq, M. J. (2011) · IEEE Sensors Journal · doi:10.1109/jsen.2011.2170676 · Cited by 56 · cited 6×
“Figure 10a shows the transceiver part of the silicone elastomer package (Sylgard 184 PDMS, Dow corning) coated with a 2 μm thick layer of Parylene-C to improve its biocompatibility [108–110].”
BackgroundMethodInfluentialYi, Y., Buttner, U., Fan, Y., & Foulds, I. (2015) · International journal of circuit theory and applications · doi:10.1002/cta.2024 · Cited by 55 · cited 3×
“Compared with the NRIC method, the NRMRC technique uses three [136] or four coil-based [137] architectures, as hown in Figure 23a,b, respectively, depending on the practical requirements, such as separation distance, PTE and load power rating [63].”
MethodOptical power transfer and communication methods for wireless implantable sensing platforms
Mujeeb-U-Rahman, M., Adalian, D., Chang, C.-F., & Scherer, A. (2015) · Journal of Biomedical Optics · doi:10.1117/1.jbo.20.9.095012 · Cited by 55 · cited 4×
“The predominant non-EM technique is acoustic-based power transfer [18] and is a far-field (more than 500 mm distance) transmission technique similar to optical coupling [19].”
BackgroundMethodInfluentialEfficiency optimization of wireless power transmission systems for active capsule endoscopes
Jia, Z., Yan, G.-Z., Jiangpingping, Wang, Z., & Liu, H. (2011) · Physiological Measurement · doi:10.1088/0967-3334/32/10/005 · Cited by 54 · cited 3×
“Experimental setup of optimized 3D receiver [121].”
BackgroundAn Implantable Pressure Sensing System With Electromechanical Interrogation Scheme
Kim, A., Powell, C., & Ziaie, B. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2318023 · Cited by 53
Chip-Scale Coils for Millimeter-Sized Bio-Implants
Feng, P., Yeon, P., Cheng, Y., Ghovanloo, M., & Constandinou, T. G. (2018) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2018.2853670 · Cited by 53 · cited 5×
“Chip-scale coils have been proposed for implantable neural microsystems [40].”
BackgroundInfluentialAnalysis and Optimization of Four-Coil Planar Magnetically Coupled Printed Spiral Resonators
Khan, S., & Choi, G.-S. (2016) · Italian National Conference on Sensors · doi:10.3390/s16081219 · Cited by 52 · cited 5×
“In [67,140,141] and [137,142], L is calculated using Equations (6) and (8), and Mij is determined using (9) and (11).”
BackgroundMethodInfluentialSelf-citationKim, D.-H., Kim, J., & Park, Y.-J. (2016) · IEEE transactions on microwave theory and techniques · doi:10.1109/tmtt.2016.2582874 · Cited by 52 · cited 6×
“where w is the wire diameter of the loop, or [64,77,78]: = 2 ln 8 − 2 (8)”
BackgroundMethodInfluentialLocalization and Tracking of Implantable Biomedical Sensors
Umay, I., Fi̇dan, B., & Barshan, B. (2017) · Sensors · doi:10.3390/s17030583 · Cited by 52
Wideband Near-Field Data Transmission Using Pulse Harmonic Modulation
Inanlou, F., & Ghovanloo, M. (2010) · IEEE Transactions on Circuits and Systems I Regular Papers · doi:10.1109/tcsi.2010.2055351 · Cited by 51
Recent progress in mid-range wireless power transfer
Lee, C., Zhong, W.-X., & Hui, S. (2012) · European Conference on Cognitive Ergonomics · doi:10.1109/ecce.2012.6342288 · Cited by 49 · cited 1×
“Furthermore, the authors in [42,43] described the theory behind mid-field WPT and its recent improvements.”
BackgroundNarayanamoorthi, R., Vimala, A., Chokkalingam, B., Padmanaban, S., & Leonowicz, Z. (2017) · Energies · doi:10.3390/en10091409 · Cited by 48 · cited 2×
“In [59], a nominal condition-based class-E power amplifier (PA) is proposed for NRCC presenting a load resistance RL, as shown in Figure 3.”
BackgroundSwitching Technique for Inductive Power Transfer at High- $Q$ Regimes
Ju, X., Dong, L., Huang, X., & Liao, X. (2014) · IEEE Transactions on Industrial Electronics · doi:10.1109/tie.2014.2361806 · Cited by 47 · cited 2×
“t i t , t reso t circ i is use as an energy storage ele e t [89].”
MethodCritical Q analysis of a current-fed resonant converter for ICPT applications
Boys, J., Hu, A., & Covic, G. (2000) · doi:10.1049/el:20001030 · Cited by 47
Longevity of Dual Chamber Pacemakers: Device and Patient Related Determinants
Kindermann, M., Schwaab, B., Berg, M., & Fröhlig, G. (2001) · Pacing and clinical electrophysiology : PACE · doi:10.1046/j.1460-9592.2001.00810.x · Cited by 44 · cited 2×
“Packaged bat eries are limited by thei operational lifetime [31], necessit ting repeated surgical interv ntions to replace them.”
BackgroundCharacterization of In-Body Radio Channels for Wireless Implants
Li, J., Nie, Z., Liu, Y., Wang, L., & Hao, Y. (2016) · IEEE Sensors Journal · doi:10.1109/jsen.2016.2635700 · Cited by 44
A video wireless capsule endoscopy system powered wirelessly: design, analysis and experiment
Pan, G., Xin, W., Yan, G., & Chen, J.-L. (2011) · doi:10.1088/0957-0233/22/6/065802 · Cited by 42
Multi-disciplinary challenges in tissue modeling for wireless electromagnetic powering: a review
Bocan, K. N., Mickle, M. H., & Sejdic, E. (2017) · IEEE Sensors Journal · doi:10.1109/jsen.2017.2748338 · Cited by 41 · cited 4×
“A lower operating frequency range of 100 kHz to 50 MHz reduces tissue losses provided that the WPT systems are of sufficient large size for optimum operation [60].”
BackgroundInfluentialA Closed-Loop Remote Powering Link for Wireless Cortical Implants
Silay, K. M., Dehollain, C., & Declercq, M. (2013) · IEEE Sensors Journal · doi:10.1109/jsen.2013.2259621 · Cited by 40
Mohamed, M., Fahmy, G., Abdel-Rahman, A., Allam, A., Barakat, A., Abo-Zahhad, M., Jia, H., & Pokharel, R. (2018) · IEEE Access · doi:10.1109/access.2018.2866457 · Cited by 36 · cited 2×
“6% [103], 87% [104] and 86% [105] efficiencies for MIDs.”
Ultrasonically powered piezoelectric generators for bio-implantable sensors: Plate versus diaphragm
Christensen, D. B., & Roundy, S. (2015) · Journal of Intelligent Material Systems and Structures · doi:10.1177/1045389x15585897 · Cited by 36
Interaction of electromagnetic waves with humans in wearable and biomedical implant antennas
Agarwal, K., & Guo, Y.-X. (2015) · Asia-Pacific Software Engineering Conference · doi:10.1109/apemc.2015.7175377 · Cited by 35
Eyeglasses-powered, contact lens-like platform with high power transfer efficiency
Kim, Y., Maeng, J., & Irazoqui, P. P. (2015) · Biomedical Microdevices · doi:10.1007/s10544-015-9979-0 · Cited by 35 · cited 5×
“Another ocular diagnostics device mbedded into a contact lens [113] and fabricated n a Parylene substrate is shown in Figures 16 and 17.”
BackgroundInfluentialResonantly coupled antennas for passive sensors
Rindorf, L., Lading, L., & Breinbjerg, O. (2008) · Italian National Conference on Sensors · doi:10.1109/icsens.2008.4716759 · Cited by 34
Highly efficient wireless energy harvesting system using metamaterial based compact CP antenna
Agarwal, K., Mishra, T., Karim, M., Nasimuddin, Chuen, M., Guo, Y.-X., & Panda, S. K. (2013) · 2013 IEEE MTT-S International Microwave Symposium Digest (MTT) · doi:10.1109/mwsym.2013.6697693 · Cited by 33 · cited 1×
“In [165], a single-stage Dickson charge pump is used as a rectifier.”
MethodPersonalizing therapies for gastric cancer: molecular mechanisms and novel targeted therapies.
Luís, M., Tavares, A., Carvalho, L. S., Lara-Santos, L., Araújo, A., & de Mello, R. D. (2013) · World Journal of Gastroenterology · doi:10.3748/wjg.v19.i38.6383 · Cited by 32 · cited 2×
“Capsule endoscopy (CE) promises early stage minimally invasive detection and rapid diagnosis of GI diseases [117] to limit the adverse effects of gastrointestinal (GI) disorders [118,119].”
BackgroundUltrasound energy harvesting system for deep implanted-medical-devices (IMDs)
Mazzilli, F., Thoppay, P., Praplan, V., & Dehollain, C. (2012) · 2012 IEEE International Symposium on Circuits and Systems · doi:10.1109/iscas.2012.6271911 · Cited by 32
Wu, W., & Fang, Q. (2011) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2011.6090998 · Cited by 32 · cited 3×
“Its performance is commonly limited by the skin effect and proximity effect resistances [36,69,72], large separation distance between TX and RX, translational and angular misalignment of the coils and tissue losses.”
BackgroundModeling of a capacitive link for data telemetry to biomedical implants
Takhti, M., Asgarian, F., & Sodagar, A. M. (2011) · Biomedical Circuits and Systems Conference · doi:10.1109/biocas.2011.6107757 · Cited by 30 · cited 3×
“In [53], an NRCC WPT link was tested on a non-human primate cadaver between 100 to 150 MHz for a 40 × 40 mm2 external TX patch size.”
MethodInfluentialHayati, M., Lotfi, A., Kazimierczuk, M., & Sekiya, H. (2014) · IEEE Transactions on Circuits and Systems Part 1: Regular Papers · doi:10.1109/tcsi.2013.2283692 · Cited by 28 · cited 3×
“Besides the nominal-based class-E PA [90–92], a higher effici cy sub-nomin l lass-E PA is presented in its gen r l cont xt [93] n for IDs [94].”
BackgroundDESIGN OF SPIRAL CIRCULAR COILS IN WET AND DRY TISSUE FOR BIO-IMPLANTED MICRO-SYSTEM APPLICATIONS
Abbas, S. M., Hannan, M. A., Samad, S. A., & Hussain, A. (2013) · Progress In Electromagnetics Research M · doi:10.2528/pierm13052707 · Cited by 26 · cited 1×
“As shown in Figure 5, an iterative optimization flow diagram has been adopted in most of the research literature to achieve maximum PTE [36,66,70,75,76].”
MethodA Magnetic-Balanced Inductive Link for the Simultaneous Uplink Data and Power Telemetry
Gong, C., Liu, D.-K., Miao, Z., & Li, M. (2017) · Italian National Conference on Sensors · doi:10.3390/s17081768 · Cited by 23 · cited 6×
“Prototype and measurement setup of the proposed system [112].”
BackgroundInfluentialWireless power delivery for retinal prostheses
Ng, D., Williams, C. E., Allen, P., Bai, S., Boyd, C. S., Meffin, H., Halpern, M., & Skafidas, E. (2011) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2011.6092061 · Cited by 23 · cited 3×
“The maximum achievable efficiency of the proposed system [147] is 8.”
BackgroundInfluentialCotté, B., Lafon, C., Dehollain, C., & Chapelon, J. (2012) · IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control · doi:10.1109/tuffc.2012.2373 · Cited by 22
Xiu, Z., Ho, S. L., & Fu, W. N. (2013) · IEEE Transactions on Magnetics · doi:10.1109/tmag.2013.2244585 · Cited by 22
Factors Affecting the Synthesis of Autonomous Sensors with RFID Interface
Węglarski, M., & Jankowski-Mihułowicz, P. (2019) · Italian National Conference on Sensors · doi:10.3390/s19204392 · Cited by 20
Khan, S., & Choi, G.-S. (2017) · Microelectronics Journal · doi:10.1016/j.mejo.2017.06.002 · Cited by 20
Self-citationHigh efficiency low power rectifier design using zero bias schottky diodes
Mabrouki, A., Latrach, M., & Lorrain, V. (2014) · IEEE Faible Tension Faible Consommation · doi:10.1109/ftfc.2014.6828604 · Cited by 20 · cited 2×
“However, the efficiency of this type of rectifier varies between 30% to 80% and significantly lower at higher operating frequencies [102] due to an increase in leakage current.”
BackgroundHigh-Efficiency Mid-Range Inductive Power Transfer Employing Alternative-Winding Coils
Yi, Z.-X., Li, M., Muneer, B., & Zhu, Q. (2019) · IEEE transactions on power electronics · doi:10.1109/tpel.2018.2872047 · Cited by 19 · cited 1×
“Recent research regarding the use of antiparallel resonant loops [133,134] demonstrate a higher PTE at a larger separation distance.”
BackgroundEffects of coil misalignment in a four coil implantable wireless power transfer system
Varghese, B. J., Bobba, P. B., & Kavitha, M. (2016) · IEEE Power India International Conference · doi:10.1109/poweri.2016.8077155 · Cited by 19 · cited 1×
“Compared with NRIC and NRMRC, this transfer technique demonstrates robustness against misalignment of the TX and RX coils [162,163].”
MethodAn optical wireless power transfer system for rapid charging
Raavi, S., Arigong, B., Zhou, R., Jung, S., Jin, M. H. ‐., Zhang, H., & Kim, H. (2013) · IEEE Texas Symposium on Wireless and Microwave Circuits and Systems (WMCS) · doi:10.1109/wmcas.2013.6563551 · Cited by 18
Study of midfield wireless power transfer for implantable medical devices
Keerthi., K. S., Ilango., K., & Manjula., G. N. (2018) · 2018 2nd International Conference on Biomedical Engineering (IBIOMED) · doi:10.1109/ibiomed.2018.8534820 · Cited by 17 · cited 2×
“The mid-field WPT for implantable systems was analyzed in [48] and its powering method for cardiac implants was demonstrated.”
MethodSpatial carrier position modulation based multichannel capacitive link for bioelectronic implants
Hassan, A., Sawma, C., Hasanuzzaman, M., Gosselin, B., & Sawan, M. (2015) · Biomedical Circuits and Systems Conference · doi:10.1109/biocas.2015.7348292 · Cited by 17
Optimization of planar strongly coupled wireless power transfer system for biomedical applications
Khan, S. R., & Choi, G.-S. (2016) · doi:10.1002/mop.29934 · Cited by 16 · cited 4×
“The optimization of NRMRC follows a similar methodology as the one presented in Figure 5 [67,75,82,137,140,141].”
MethodInfluentialSelf-citationMoorey, C. L., Holderbaum, W., & Potter, B. (2015) · doi:10.3390/en81011342 · Cited by 15 · cited 2×
“where QL (=2πfLRX/RL) [15,87,88] is the loaded Q-factor calculated by using the inductance of the RX coil, LRX, and the load resistance, RL, and the imensi nless figure-of-merit, FOM, defined as [87]: = , (26)”
Method3-Coil orientation insensitive wireless power transfer for capsule endoscope
Ghotbi, I., Najjarzadegan, M., Ashtiani, S., Shoaei, O., & Shahabadi, M. (2015) · Iranian Conference on Electrical Engineering · doi:10.1109/iraniancee.2015.7146407 · Cited by 15 · cited 2×
“A similar rectifier topology with Schottky diodes is presented in [97,98] for three dimensional (3D) NRIC WPT for MIDs.”
MethodRudge, A. (1982) · Electronics and Power · doi:10.1049/ep.1982.0113 · Cited by 15
Neural Computation, Neural Devices, and Neural Prosthesis
Yang, Z. (2014) · Springer: New York · doi:10.1007/978-1-4614-8151-5 · Cited by 15 · cited 4×
“However, the level of miniaturization required for NRIC and NRMRC implantable systems affects the PTE performance at these low frequencies [13,154], especially when the implant is much smaller than its separation distance from the TX coil [155].”
BackgroundMethodInfluentialThangasamy, V., Kamsani, N., Thiruchelvam, V., Hamidon, M. N., Hashim, S., Bukhori, M. F., & Yusoff, Z. (2015) · Student Conference on Research and Development · doi:10.1109/scored.2015.7449370 · Cited by 14
Wireless Power Transfer at Sub-GHz Frequency for Capsule Endoscope
Ding, K., Yu, Y., Lin, H., & Xie, J. (2016) · doi:10.2528/pierc16042302 · Cited by 13 · cited 3×
“NRMRC PT sub-GHz experi ental setup [149].”
InfluentialPowering low-power implants using PZT transducer discs operated in the radial mode
Sanni, A., & Vilches, A. (2013) · doi:10.1088/0964-1726/22/11/115005 · Cited by 12
New analytical model for the characterisation of printed spiral coils for wireless power transfer
Khan, S., Pavuluri, S., & Desmulliez, M. (2018) · doi:10.1049/cp.2018.1122 · Cited by 12 · cited 7×
“A nested multidimensional optimization algorithm has recently been proposed in [79].”
MethodInfluentialSelf-citationNyberg, D., Kildal, P., & Carlsson, J. (2007) · doi:10.1049/ic.2007.1256 · Cited by 11
An ultrasonically powered implantable micro-light source for localized photodynamic therapy
Zhou, J.-W., Kim, A., Song, S. H., & Ziaie, B. (2015) · International Conference on Solid-State Sensors, Actuators and Microsystems · doi:10.1109/transducers.2015.7181063 · Cited by 10
3-Coil resonance-based wireless power transfer system for implantable electronic
Yi, Y., Buttner, U., Fan, Y., & Foulds, I. (2013) · 2013 IEEE Wireless Power Transfer (WPT) · doi:10.1109/wpt.2013.6556925 · Cited by 10 · cited 1×
“The PTE of the three-coil-based system using lumped element analysis is derived as [136,143]:”
MethodA 36 W Wireless Power Transfer System with 82% Efficiency for LED Lighting Applications
Chen, W.-T., Chinga, R. A., Yoshida, S., Lin, J.-S., & Hsu, C.-K. (2013) · doi:10.5104/jiepeng.6.32 · Cited by 10
An ultrasonically controlled power management system for implantable biomedical devices
Zhou, J.-W., Kim, A., & Ziaie, B. (2015) · Biomedical Circuits and Systems Conference · doi:10.1109/biocas.2015.7348297 · Cited by 9
Analysis and design of Class-E power amplifiers at any duty ratio in frequency domain
Yang, T.-L., Liang, J., Zhao, C.-Y., & Chen, D.-Y. (2011) · doi:10.1007/s10470-010-9568-3 · Cited by 9
Wang, J.-H., Ho, S., Fu, W., & Sun, M. (2010) · Digests of the 2010 14th Biennial IEEE Conference on Electromagnetic Field Computation · doi:10.1109/cefc.2010.5481419 · Cited by 9 · cited 1×
“Witricity, a term coined for wireless electricity, is the technology to transfer power using two resonating objects synchronized at the same operating frequency [17].”
Background
Flagged first, then by citations
10 of the 219 references couldn’t be matched to a known paper.
Research works printed but not linked: 9
- Basaeri, 2016, A review of acoustic power transfer for bio-medical implants, Smart Mater. Struct., 25, 123001, 10.1088/0964-1726/25/12/123001 · DOI · Paper page
- Schuder, 1961, High-level electromagnetic energy transfer through a closed chest wall, Inst. Radio Engrs. Int Conv. Rec., 9, 119
- Sadiku, N. (1994). Elements of Electromagnetics, Sounders College Press.
- Pozar, D. (2005). Microwave Engineering Fourth Edition, Wiley.
- Volakis, J., Chen, C.-C., and Fujimoto, K. (2010). Small Antennas: Miniaturization Techniques & Applications, McGraw-Hill Professional Publishing.
- Gustafsson, M., Tayli, D., and Cismasu, M. (2015). Physical Bounds of Antennas, Springer. · DOI
- Yaghjian, A.D., and Best, S.R. (2003, January 22–27). Impedance, bandwidth, and Q of antennas. Proceedings of the IEEE Antennas and Propagation Society International Symposium. Digest. Held in Conjunction with: USNC/CNC/URSI North American Radio Sci. Meeting (Cat. No.03CH37450), Columbus, OH, USA.
- Hoskins, P., Martin, K., and Thrush, A. (2010). Diagnostic Ultrasound: Physics and Equipment, Cambridge University Press. · DOI
- Sze, S.M., and Ng, K.K. (2006). Physics of Semiconductor Devices, John Wiley & Sons, Inc. · DOI · Paper page
Cited by
Sorted by citations
Biodegradable Materials for Sustainable Health Monitoring Devices
Hosseini, E. S., Dervin, S., Ganguly, P., & Dahiya, R. (2020) · ACS Applied Bio Materials · doi:10.1021/acsabm.0c01139 · Cited by 341
Wireless power transfer based on novel physical concepts
Song, M., Jayathurathnage, P. K. S., Zanganeh, E., Krasikova, M., Smirnov, P. R., Belov, P. A., Kapitanova, P. V., Simovski, C. R., Tretyakov, S., & Krasnok, A. (2021) · Nature Electronics · doi:10.1038/s41928-021-00658-x · Cited by 201
Wireless charging systems for electric vehicles
Amjad, M. J., Farooq‐i‐Azam, M., Ni, Q., Dong, M., & Ansari, E. A. (2022) · Renewable and Sustainable Energy Reviews · doi:10.1016/j.rser.2022.112730 · Cited by 198
Miniaturised wireless power transfer systems for neurostimulation: a review
Barbruni, G. L., Ros, P. M., Demarchi, D., Carrara, S., & Ghezzi, D. (2020) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2020.3038599 · Cited by 161
“Differently from other works [8], [9], [10] this review compares the advantages as well as the critical limitations of each powering method in designing a wireless link”
Implantable bioelectronics toward long-term stability and sustainability
Li, Y., Li, N., Oliveira, N. D., & Wang, S. (2021) · Matter · doi:10.1016/j.matt.2021.02.001 · Cited by 161
Metadata
- Authors
- Khan, S. R., Pavuluri, S. K., Cummins, G., & Desmulliez, M. P. Y.
- Affiliations
- Heriot-Watt University (GB)
- University of Birmingham (GB)
- Corresponding author
- Khan, Sadeque Reza
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2020-06-19
- Volume
- 20
- Issue
- 12
- Pages
- 3487
- ISSN
- 1424-8220 (ISSN-L)
- PMC
- PMC7349694
- OpenAlex
- W3036329699
- Cited by
- 355
- References
- 219
- Citations by year
- 2026: 50 · 2025: 70 · 2024: 69 · 2023: 70 · 2022: 50 · 2021: 40 · 2020: 6
- FWCI
- 2.06
- Type
- journal article
- Language
- en
- Open access
- gold
- dx.doi.org
- Licence
- Locations
- Sensors (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- University of Birmingham Research Portal (University of Birmingham) (submittedVersion, cc-by)
- Funders
- Engineering and Physical Sciences Research Council (EP/K034537/2)
- Topics
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Wireless Power Transfer Systems
- Physical Sciences › Engineering › Biomedical Engineering › Wireless Body Area Networks
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Induction Heating and Inverter Technology
- Keywords
- wireless power transfer, biomedical implants, inductive coupling, magnetic resonance coupling, capacitive coupling, power transfer efficiency, tissue safety, separation distance, operating frequency
- MeSH terms
- Electronics, Electric Power Supplies, Prostheses and Implants, Wireless Technology
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review