A Low-Cost, Accurate, and Easily-Worn E-Skin and IMU Hand Kinematic Measurement System
Open accessAbstract
Stroke and other neurological injuries impair hand function. Although rehabilitation therapists encourage reintegration of the affected hand into daily activities, there are few tools that can be worn during everyday life that provide quantitative feedback on how much or how effectively the hand is used. Wearable sensors that can accurately track hand movements and are easily applied and removed can present intuitive feedback that could motivate hand use similarly to how pedometers encourage walking. While tracking all the hand and finger joints is needed for scientific studies, under-sensorization, or using fewer sensors than required for tracking all degrees of freedom, may suffice for providing users feedback about hand use in everyday life. Further, it may enable a wearable device to be more easily donned and doffed, more power efficient, and more cost efficient. Here we develop a low-cost, multi-sensor, wireless wearable system for tracking selected hand and wrist movements during everyday life. The system includes fabricated soft, stretchable “e-skin” bend sensors and off-the-shelf inertial measurement units (IMUs) that accurately measure finger bend angles and wrist movements. The system also includes an application and removal protocol that enabled naïve unimpaired participants to apply and remove the system in ~5 min and ~4 min, respectively. The system cost was $111 per device, with prices falling to an estimate of $55 when manufactured at scale. This hand wearable demonstrates accurate kinematic tracking and user-friendly donning/doffing workflows for unimpaired participants, making it a promising platform for everyday hand tracking. Future work will extend this platform to the movement-impaired population for neurorehabilitation applications.
Overview
CoChat Summary
TL;DR
The paper presents a low-cost, wireless, under-sensorized hand/wrist kinematic system combining soft liquid-metal e-skin bend sensors and IMUs, with a user-friendly donning/doffing workflow. It reports benchtop and on-hand accuracy, usability in unimpaired participants, a ~9 h wear demonstration, and a $111 prototype cost ($55 estimated at scale).
Key findings
- Benchtop strain-sensor accuracy: mean absolute error (MAE) 1.39° and RMSE 1.75° across 13 sensors; 99.7% of predictions had absolute error ≤5°. Under worst-case misalignment, MAE was 2.31° (RMSE 2.80°), with 91.3% of Model 1 predictions ≤5°.
- On-hand strain-sensor accuracy for index MCP: overall MAE 4.91° (mean RMSE 7.30°) across first-application self-trained and second-application Model 1 conditions; 88% of pooled predictions were ≤10°. On-hand wrist flexion/extension with two IMUs: MAE 6.73° (RMSE 8.80°); 79.4% of predictions ≤10°, 92.7% ≤15°, and 98% ≤20°.
- Usability: naïve unimpaired participants applied the full system in ~5 min and removed it in ~4 min. Toupee tape (TT) was significantly faster to apply/remove than Derma-Tac (DT) and had fewer failures: TT 0/24 total failures and 1/24 partial failures vs DT 4/24 total failures and 4/24 partial failures.
- System cost: $111 per prototype device, estimated $118 for a future custom-PCB iteration, and estimated $55 per unit at manufacturing scale. A single participant wore the system continuously for ~9 h during daily activities, exceeding the conservative ~8.2 h battery-life estimate.
Methods
- The system used in-lab-fabricated soft silicone microfluidic “e-skin” strain sensors with eutectic gallium–indium channels, plus off-the-shelf BNO055 IMUs, custom data acquisition electronics, a watch-style wrist unit, and a dorsal hand-back unit; data streamed wirelessly via Bluetooth.
- Strain sensors were characterized on a custom benchtop “autobender” against a rotary encoder, then tested on-hand in N=4 unimpaired participants using an overhead camera and DeepLabCut as ground truth for index MCP and wrist angles; calibration used quadratic fits between normalized ADC values and known bend angles.
- A separate N=4 unimpaired usability study compared TT vs DT attachment over repeated application/removal cycles, including splinted dexterity and simulated ADLs; a single participant completed a ~9 h long-term wear protocol. Statistical analysis used ANOVA, paired t-tests, and linear mixed-effects models with effect sizes and confidence intervals.
Limitations
- The paper states that accuracy and usability were tested only in small unimpaired populations; future studies in larger movement-impaired populations are required, and the on-hand accuracy of the thumb MCP remains uncharacterized.
- The long-term wear demonstration was N=1 and qualitative; real-world artifacts such as sensor slip, wire motion, direct pressure, impacts, temperature-dependent drift, and electromagnetic interference were not fully characterized.
- The authors note risk of underpowered statistics in small-sample device testing; cost figures for the custom PCB and scaled manufacturing are engineering estimates, and the comfort study’s participant numbering appears inconsistent in the text (N=4 but listed as P5–P9).
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures

![Figure 2 Benchtop testing of strain samples using an autobender. (A) Schematic of the autobender. The strain sensor was attached to both the fixed and moving plates using Velcro. As the moving plate was actuated by a servo motor, the sensor bent over a curved radius in 15 deg increments from 0 deg to 90 deg and back, completing five full cycles. The ground truth bend angle was measured using a rotary encoder mounted coaxially with the servo motor’s axis of rotation. (B) Change in resistance (ΔR/R0) versus estimated mechanical strain (ϵXX). Experimental data (points) are compared against the predicted theoretical model (dashed line) for three different cross-sectional areas (CXA). The inset equation shows the theoretical relationship between the normalized resistance and mechanical strain. In the legend, “thick” corresponds to the thickness of the sample. Agreement between the experimental measurements and the theoretical model was quantified using variance accounted for (VAF), which was 0.999, 0.994, and 0.993 for the three samples. (C) Normalized sensor values versus bend angle and the quadratic fit. The data shown are the same experimental data as the black curve in (B). (D) Absolute angular error box plots for 13 different sensor samples grouped into five sensor-length categories. A one-way ANOVA did not detect a statistically significant effect of sensor length on mean absolute error (F(4, 8) = 1.41, p = 0.31, ω² = 0.11). (E) Schematic illustrating the reapplication process. Numbers indicate the application numbers: (1) apply sensor for the first time, and train Model 1, (2–3) apply sensor for the second and third times and train “Self-Trained Model i” models, and apply self-trained models and Model 1. (F) Reapplication absolute angular error box plots for three different samples in the sensor reapplication experiments. The split blue/orange boxes correspond to application (1) in (E). Blue-only boxes correspond to errors from self-trained models in reapplication steps 2–3, and orange-only boxes correspond to errors obtained by applying the first-application calibration model (Model 1). When MAEs were calculated by averaging within each sample across reapplications, Model 1 produced significantly higher errors than the corresponding self-trained models (paired t-test, t(2) = 6.72, p = 0.02). The estimated mean paired difference was 0.64° (95% CI [0.23°, 1.06°], Cohen’s d = 3.83, Hedges’ g = 2.19). No statistically significant difference was detected between the second and third reapplications when using Model 1 (paired t-test, t(2) = −0.82, p = 0.50). The estimated mean difference was −0.43° (95% CI [−2.66°, 1.81°], Cohen’s d = −0.47, Hedges’ g = −0.27). (G) Schematic of misalignment angular error testing on the autobender. The same sample was intentionally misaligned by 10 deg and 20 deg to evaluate the resulting angular error. (H) Misalignment absolute angular error box plots for three different samples. Split blue/orange boxes correspond to the 0° aligned condition. Blue-only boxes represent errors from self-trained models, while orange-only boxes represent errors obtained by applying the 0° calibration model (“Model 1”) to the misaligned data. No statistically significant difference was detected between errors derived from self-trained models and those obtained using Model 1 (paired t-test, t(2) = 1.29, p = 0.33). The estimated mean difference was 1.67° (95% CI [−3.79°, 7.13°], Cohen’s d = 0.74, Hedges’ g = 0.42). Partially created in BioRender. Oppenheim, T. (2026). https://biorender.com/86pe6ky (accessed on 19 June 2026).](https://figures.cochat.ai/paper-figures/1e257bd2a6304ff88a4694f7dca9ccd3/a-low-cost-accurate-and-easily-worn-e-skin-and-imu-hand-kinematic-measurement-figure-2.jpg)
![Figure 3 Strain sensor error analysis on the hand. (A) Camera and stabilizing fixture for user study to investigate the on-hand sensor accuracy of measuring index MCP flexion/extension. (B) Example calibration data from the first sensor application on the third participant (P3). (C) Example time-series comparison of the predicted bend angles from strain sensors and the corresponding camera-measured bend angle for the index MCP during the slowest bend rate (30 BPM). (D) Absolute angular error box plot computed from the time-series comparison shown in (C). (E) Schematic showing placement of the strain sensor over the index and thumb MCP joints. (F,G) Similar to (C,D), except the time-series data/absolute angular error box plot correspond to the 240 BPM MCP bend rate. (H) Summary angular error box plots from the first sensor application for each participant and each MCP bend rate. The bar plot on the far right shows the MAE across all participants for each MCP bend rate. (I) Second-application summary. Blue boxes represent errors derived from a calibration curve fit using second-application data (“self-trained model”), while orange boxes represent errors obtained by applying the first-application calibration to the second-application data (“Model 1”). When MAEs were averaged across speeds within each participant, no statistically significant difference was detected between the two calibration approaches (paired t-test, t(2) = 1.20, p = 0.35). The estimated mean difference was −0.61° (95% CI [−2.79°, 1.57°], Cohen’s d = −0.70, Hedges’ g = −0.40). A linear mixed-effects model with speed as a fixed effect and participant as a random effect on intercept, pooled across participants and applications, revealed no significant effect of speed on the MAE (slope = 0.096, CI [−0.252, 0.443], z = 0.539, N = 45, p = 0.59, R2 marginal = 0.007, R2 conditional = 0.007). Partially created in BioRender. Schlegel, H. (2026). https://biorender.com/t2ywslt (accessed on 19 June 2026).](https://figures.cochat.ai/paper-figures/42a603fd1e2f418498fd1f2086820e3c/a-low-cost-accurate-and-easily-worn-e-skin-and-imu-hand-kinematic-measurement-figure-3.jpg)
![Figure 4 IMU error analysis on the hand. (A) Schematic illustrating the plane-method calculation used to measure the palm flexion angle from two IMUs. (B) Photograph of the customized breakout board designed for the BNO055 IMU. (C) Example time-series comparison of predicted bend angles from the IMU (orange) and the corresponding camera-measured bend angle (blue) for wrist flexion/extension during the slowest bend rate (30 BPM). (D) Absolute angular error box plot computed from the time-series comparison shown in (C). (E) Schematic showing IMU placement within a watch case and on the dorsal surface of the palm, with both IMUs oriented in the same direction. (F,G) Similar to (C,D), except the time-series data/absolute angular error box plot correspond to the 240 BPM bend rate. (H) Summary angular error box plots from the first application for each participant and each wrist bend rate. The bar plot on the far right shows the mean angular error across all participants for each wrist bend rate. (I) Same as (H) but for the second application. After averaging the MAE across the first and second applications within each participant and within each movement speed, a linear mixed-effects model with speed as a fixed effect and participant as a random intercept revealed a significant effect of speed on the mean IMU MAE (slope = 0.585 deg per speed category, z = 2.018, N = 19, p = 0.04356, 95% CI [0.017, 1.152], R2 marginal = 0.114, R2 conditional = 0.508). Partially created in BioRender. Schlegel, H. (2026). https://biorender.com/cxnk361 (accessed on 19 June 2026).](https://figures.cochat.ai/paper-figures/0a9f24be603047e698e2473b37c5bcc5/a-low-cost-accurate-and-easily-worn-e-skin-and-imu-hand-kinematic-measurement-figure-4.jpg)

![Figure 6 User comfort study protocol and comparative performance of two sensor-attachment methods (DT, blue vs. TT, orange). (A) Study timeline. Participants completed an allergy check, experimenter-guided APP0/REM0, then three timed cycles of application and removal without any experimenter feedback (APP1–3, REM1–3), indicated by the green clock icon. Fine motor tasks (pinch/grasp, writing, and typing) and daily-living tasks (eating, phone use, hand sanitizer application, and jacket on/off) followed APP2; APP2/REM2 were performed with finger splints to limit finger range of motion. A ~60 min free-time block preceded REM3. Cycle-specific questions followed each APP/REM cycle (indicated by the clipboard icon); final open-ended questions were asked prior to concluding the session. (B) Application time by method and cycle with participant traces in gray and mixed-effects learning lines in black. DT and TT both get significantly faster with each application (linear mixed-effects model, random intercept by participant, DT: slope = −39.5 s per application, 95% CI [−74.2, −4.78], t = −2.23, p = 0.026, R2 marginal = 0.092, R2 conditional = 0.796, n = 12, TT: slope = −60.8 s per application, 95% CI [−99.5, −22.0], t = −3.07, p = 0.0021, R2 marginal = 0.143, R2 conditional = 0.833, n = 12). (C) Pooled over APP1-3, TT applications were significantly faster than DT applications (difference between mean application times with DT and TT was 130.6 s, 95% CI [−180.06, −81.11], paired, two-sided t-test, t = −5.810, p = 0.0001, n = 12 paired observations from 4 participants × 3 APP, Cohen’s d = −1.677, Hedges’ g = −1.560). (D) Removal time by method and cycle with participant traces in gray and mixed-effects learning lines in black. DT did not significantly change in removal time but TT significantly improved in speed (linear mixed-effects model, random intercept by participant, DT: 95% CI on slope [−60.48, 7.98], t = −1.503, p = 0.133, R2 marginal = 0.036, R2 conditional = 0.824, n = 12, TT slope: −44.4 s per removal, 95% CI [−87.23, −1.52], t = −2.030, p = 0.0424, R2 marginal = 0.070, R2 conditional = 0.812, n = 12). (E) Pooled over REM1–3, TT removals were significantly faster than DT removals (difference between mean removal times between DT and TT of 157.9 s, 95% CI [−226.60, −89.23], paired, two-sided t-test, t = −5.060, p = 0.0004; n = 12 paired observations = 4 participants × 3 REM, Cohen’s d = −1.461, Hedges’ g = −1.359). (F) Survey ratings (1–10; paired, two-sided t-tests; n = 12 pairs/metric) favor TT on ease (+0.75, 95% CI [0.33, 1.17], t = 3.954, p = 0.0023, Cohen’s d = 1.141, Hedges’ g = 1.062), stability (+1.33, 95% CI [0.79, 1.88], t = 5.360, p = 0.00023, Cohen’s d = 1.547, Hedges’ g = 1.439), and adhesion (+1.33, 95% CI [0.60, 2.07], t = 4.000, p = 0.0021, Cohen’s d = 1.155, Hedges’ g = 1.074). (G) Wear failure summary (counts across WEAR1–3): total failure (fully off) and partial failure (lifting) for DT and TT; y-axis shows counts (0–24). The 24 count arises from 4 participants × 3 APP × 2 sensors/application. For full figure, * indicates p < 0.05, ** indicates p < 0.01, *** indicates p < 0.001. Partially created in BioRender. Schlegel, H. (2026). https://biorender.com/cndws0y and https://biorender.com/xn0edbu (accessed on 19 June 2026).](https://figures.cochat.ai/paper-figures/2ce33cdfeaf14df69da11e928860735b/a-low-cost-accurate-and-easily-worn-e-skin-and-imu-hand-kinematic-measurement-figure-6.jpg)

Verification
Claims the paper makes
0 Valid · 2 Not valid · 28 Not checked · 144 Own results and interpretations
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The R2 (variance accounted for) between the data and model was 0.999, 0.994, and 0.993 for the three samples.”
§ A Low-Cost, Accurate, and Easily-Worn E-Skin and IMU Hand Kinematic Measurement System > 4. Results > 4.1. Benchtop Testing of Strain Sensors
plainstates an associationn = 3- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper states — it cites no source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“a demonstration of multi-hour out-of-lab hand tracking”
§ A Low-Cost, Accurate, and Easily-Worn E-Skin and IMU Hand Kinematic Measurement System > Abstract > Highlights
boosted- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“No statistically significant association between the bend speed and MAE was detected (slope = 0.096, 95% CI [−0.252, 0.443], z = 0.539, p = 0.59, R2 marginal = 0.007, R2 conditional = 0.007).”
§ A Low-Cost, Accurate, and Easily-Worn E-Skin and IMU Hand Kinematic Measurement System > 4. Results > 4.2. On-Hand Testing of Strain Sensors
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
In the paper
“Consistent with previous work [15], a linear mixed-effects model revealed a significant effect of increasing angular error with increasing movement speed (Figure 4H,I).”
§ A Low-Cost, Accurate, and Easily-Worn E-Skin and IMU Hand Kinematic Measurement System > 4. Results > 4.4. On-Hand Testing of IMUs
What the sources say
- Not checked yetcites [15] Shenoy, 2022, Design and Validation of an IMU Based Full Hand Kinematic Measurement System, IEEE Access source
plainstates a causen = 19 ·- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“While our strain sensors did not demonstrate considerable changes in signals with the index finger abduction/adduction movement (Figure S4)”
§ A Low-Cost, Accurate, and Easily-Worn E-Skin and IMU Hand Kinematic Measurement System > 5. Discussion > 5.9. Future Directions
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
63 more claims
Create a free account to see all 174 claims and to verify the 50 references still open.
- It may state two separate claims as one. 71
- The quote only partly states this claim. 2
- Probably not a substantial claim. 51
Statistics recomputed
Every reported test result, recomputed from its test statistic and degrees of freedom. A result shows as not adding up only once an independent check has confirmed it.
- ConsistentF(4, 8) = 1.41reported p = .31 · recomputed p = .314
“F(4, 8) = 1.41, p = .31”
- Consistentt(2) = 6.72reported p = .02 · recomputed p = .021
“t(2) = 6.72, p = .02”
- Consistentt(2) = -0.82reported p = .5 · recomputed p = .498
“t(2) = −0.82, p = .50”
- Consistentt(2) = 1.29reported p = .33 · recomputed p = .326
“t(2) = 1.29, p = .33”
- Consistentt(2) = 1.2reported p = .35 · recomputed p = .353
“t(2) = 1.20, p = .35”
1 more recomputed test
Create a free account to see every test the Stats check recomputed.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
“a shift in 0.16 Ohms, which is 2% of our strain sensor resistance, measured as 0.7 Ohms.”
§ A Low-Cost, Accurate, and Easily-Worn E-Skin and IMU Hand Kinematic Measurement System > 5. Discussion > 5.8. Robustness to Real-World Disturbances
Reported 2 · recomputed 22.857143
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ A Low-Cost, Accurate, and Easily-Worn E-Skin and IMU Hand Kinematic Measurement System > 4. Results > 4.1. Benchtop Testing of Strain Sensors
“A one-way ANOVA performed on per-sample mean absolute errors showed no statistically significant effect of the sensor length (L0) on the absolute angular error”
§ A Low-Cost, Accurate, and Easily-Worn E-Skin and IMU Hand Kinematic Measurement System > 2. Materials and Methods > 2.6. Sensor Length Testing on the Autobender
“Thirteen sensors with identical liquid-metal channel lengths (L0) but varying sensor lengths (L1) to account for different hand sizes were tested”
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ A Low-Cost, Accurate, and Easily-Worn E-Skin and IMU Hand Kinematic Measurement System > 4. Results > 4.6. Testing System Application in a User Study
“separate participants (N = 4, P5–P9), were brought to the lab”
§ A Low-Cost, Accurate, and Easily-Worn E-Skin and IMU Hand Kinematic Measurement System > 3. User Comfort Study Protocol > 3.1. Participants
“We recruited n = 4 adults with typical healthy upper-limb movement”
Citations
Cites
DeepLabCut: markerless pose estimation of user-defined body parts with deep learning
Mathis, A., Mamidanna, P., Cury, K. M., Abe, T., Murthy, V. N., Mathis, M. W., & Bethge, M. (2018) · Nature Neuroscience · doi:10.1038/s41593-018-0209-y · Cited by 5,187
Materials and Mechanics for Stretchable Electronics
Rogers, J. A., Someya, T., & Huang, Y. (2010) · Science · doi:10.1126/science.1182383 · Cited by 5,042
Wolf, S. L., Winstein, C. J., Miller, J. P., Taub, E., Uswatte, G., Morris, D., Giuliani, C., Light, K. E., Nichols-Larsen, D., & EXCITE Investigators, F. T. (2006) · JAMA · doi:10.1001/jama.296.17.2095 · Cited by 1,889
Olejnik, S., & Algina, J. (2003) · Psychological Methods · doi:10.1037/1082-989x.8.4.434 · Cited by 1,622
Embedded 3D Printing of Strain Sensors within Highly Stretchable Elastomers
Muth, J. T., Vogt, D. M., Truby, R. L., Mengüç, Y., Kolesky, D. B., Wood, R. J., & Lewis, J. A. (2014) · Advanced Materials · doi:10.1002/adma.201400334 · Cited by 1,611
Dickey, M. D., Chiechi, R. C., Larsen, R. J., Weiss, E. A., Weitz, D. A., & Whitesides, G. M. (2008) · Advanced Functional Materials · doi:10.1002/adfm.200701216 · Cited by 1,468
End-to-end design of wearable sensors
Ates, H. C., Nguyen, P. Q., Gonzalez-Macia, L., Morales-Narváez, E., Güder, F., Collins, J. J., & Dincer, C. (2022) · Nature Reviews Materials · doi:10.1038/s41578-022-00460-x · Cited by 1,306
Postural Hand Synergies for Tool Use
Santello, M., Flanders, M., & Soechting, J. F. (1998) · The Journal of Neuroscience · doi:10.1523/jneurosci.18-23-10105.1998 · Cited by 1,215
Deep Learning in Human Activity Recognition with Wearable Sensors: A Review on Advances
Zhang, S., Li, Y., Zhang, S., Shahabi, F., Xia, S., Deng, Y., & Alshurafa, N. (2022) · Sensors · doi:10.3390/s22041476 · Cited by 625
Wearable Microfluidic Diaphragm Pressure Sensor for Health and Tactile Touch Monitoring
Gao, Y., Ota, H., Schaler, E. W., Chen, K., Zhao, A., Gao, W., Fahad, H. M., Leng, Y., Zheng, A., Xiong, F., et al. (2017) · Advanced Materials · doi:10.1002/adma.201701985 · Cited by 614 · cited 1×
“Other studies have developed customized application strategies for wearables that include the development of customized hand-molds [30] or printing directly on the skin [46], but these are not scalable nor cost-effective.”
Constraint-induced movement therapy after stroke
Kwakkel, G., Veerbeek, J. M., van Wegen, E. E. H., & Wolf, S. L. (2015) · The Lancet Neurology · doi:10.1016/s1474-4422(14)70160-7 · Cited by 591
MediaPipe Hands: On-device Real-time Hand Tracking
Zhang, F., Bazarevsky, V., Vakunov, A., Tkachenka, A., Sung, G., Chang, C.-L., & Grundmann, M. (2020) · arXiv (Cornell University) · doi:10.48550/arxiv.2006.10214 · Cited by 557
Predictors of adherence to home-based physical therapies: A systematic review
Essery, R., Geraghty, A. W. A., Kirby, S., & Yardley, L. (2017) · Disability and Rehabilitation · doi:10.3109/09638288.2016.1153160 · Cited by 391
Intensive upper limb neurorehabilitation in chronic stroke: outcomes from the Queen Square programme
Ward, N. S., Brander, F., & Kelly, K. (2019) · Journal of Neurology, Neurosurgery & Psychiatry · doi:10.1136/jnnp-2018-319954 · Cited by 341
McCabe, J., Monkiewicz, M., Holcomb, J., Pundik, S., & Daly, J. J. (2015) · Archives of Physical Medicine and Rehabilitation · doi:10.1016/j.apmr.2014.10.022 · Cited by 221
Analysis of the synergies underlying complex hand manipulation
Todorov, E., & Ghahramani, Z. (2005) · The 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2004.1404285 · Cited by 178
A substrate-less nanomesh receptor with meta-learning for rapid hand task recognition
Kim, K. K., Kim, M., Pyun, K., Kim, J., Min, J., Koh, S., Root, S. E., Kim, J., Nguyen, B.-N. T., Nishio, Y., et al. (2022) · Nature Electronics · doi:10.1038/s41928-022-00888-7 · Cited by 175
Assessment of hand kinematics using inertial and magnetic sensors
Kortier, H. G., Sluiter, V. I., Roetenberg, D., & Veltink, P. H. (2014) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-11-70 · Cited by 173
Reconstructing Hands in 3D with Transformers
Pavlakos, G., Shan, D., Radosavovic, I., Kanazawa, A., Fouhey, D., & Malik, J. (2024) · 2024 IEEE/CVF Conference on Computer Vision and Pattern Recognition (CVPR) · doi:10.1109/cvpr52733.2024.00938 · Cited by 163
AI-Driven Wearable Bioelectronics in Digital Healthcare
Huang, G., Chen, X., & Liao, C. (2025) · Biosensors · doi:10.3390/bios15070410 · Cited by 138
Tashakori, A., Jiang, Z., Servati, A., Soltanian, S., Narayana, H., Le, K., Nakayama, C., Yang, C.-L., Wang, Z. J., Eng, J. J., et al. (2024) · Nature Machine Intelligence · doi:10.1038/s42256-023-00780-9 · Cited by 137
Li, X., Rytkin, E., Zhao, Q., Bhat, P., Pfenniger, A., Yin, L., Huang, X., Yang, L., Yang, B., Burrell, A. R., et al. (2025) · Science Advances · doi:10.1126/sciadv.adw3044 · Cited by 98
Design of an Inertial-Sensor-Based Data Glove for Hand Function Evaluation
Lin, B.-S., Lee, I.-J., Yang, S.-Y., Lo, Y.-C., Lee, J., & Chen, J.-L. (2018) · Sensors · doi:10.3390/s18051545 · Cited by 96
A Soft Sensor-Based Three-Dimensional (3-D) Finger Motion Measurement System
Park, W., Ro, K., Kim, S., & Bae, J. (2017) · Sensors · doi:10.3390/s17020420 · Cited by 77
3D printed water-soluble scaffolds for rapid production of PDMS micro-fluidic flow chambers
Dahlberg, T., Stangner, T., Zhang, H., Wiklund, K., Lundberg, P., Edman, L., & Andersson, M. J. (2018) · Scientific Reports · doi:10.1038/s41598-018-21638-w · Cited by 77
Unexpected complexity of everyday manual behaviors
Yan, Y., Goodman, J. M., Moore, D. D., Solla, S. A., & Bensmaia, S. J. (2020) · Nature Communications · doi:10.1038/s41467-020-17404-0 · Cited by 73
Kim, J., Kim, B., Jang, M., Kang, K., Kim, D., Ju, B.-K., & Kim, J. (2020) · Sensors · doi:10.3390/s20071921 · Cited by 68
Mugler, E. M., Tomic, G., Singh, A., Hameed, S., Lindberg, E. W., Gaide, J., Alqadi, M. M., Robinson, E., Dalzotto, K., Limoli, C., et al. (2019) · Neurorehabilitation and neural repair · doi:10.1177/1545968319834903 · Cited by 66
Luo, Y., Fan, H., Lai, X., Zeng, Z., Lan, X., Lin, P., Tang, L., Wang, W., Chen, Y., & Tang, Y. (2024) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2023.115905 · Cited by 63
Park, M., Park, T., Park, S., Yoon, S. J., Koo, S. H., & Park, Y.-L. (2024) · Nature Communications · doi:10.1038/s41467-024-50101-w · Cited by 58
Ganguly, K., Khanna, P., Morecraft, R. J., & Lin, D. J. (2022) · Neuron · doi:10.1016/j.neuron.2022.06.024 · Cited by 49
Self-citationSun, Y.-C., Boero, G., & Brugger, J. (2021) · ACS Applied Electronic Materials · doi:10.1021/acsaelm.1c00884 · Cited by 39
Laganà, F. (2026) · Electronics · doi:10.3390/electronics15040833 · Cited by 38
Integrating wearable health devices with AI and edge computing for personalized rehabilitation
Xi, L., Li, C., Anari, M. S., & Rezaee, K. (2025) · Journal of Cloud Computing · doi:10.1186/s13677-025-00795-0 · Cited by 36
Data-driven glove calibration for hand motion capture
Wang, Y., & Neff, M. (2013) · Proceedings of the 12th ACM SIGGRAPH/Eurographics Symposium on Computer Animation · doi:10.1145/2485895.2485901 · Cited by 35
Hybrid AI–Taguchi–ANOVA Approach for Thermographic Monitoring of Electronic Devices
Laganà, F., Pratticò, D., Quattrone, M. F., Pullano, S. A., & Calcagno, S. (2026) · Eng · doi:10.3390/eng7010028 · Cited by 33
Marin‐Pardo, O., Phanord, C. S., Donnelly, M. R., Laine, C. M., & Liew, S. (2021) · Sensors · doi:10.3390/s21051806 · Cited by 26
Linear and Nonlinear Kinematic Synergies in the Grasping Hand
Patel, V., & Burns, M. (2015) · Journal of Bioengineering & Biomedical Science · doi:10.4172/2155-9538.1000163 · Cited by 19
Design and Validation of an IMU Based Full Hand Kinematic Measurement System
Shenoy, P., Gupta, A., & S.K.M., V. (2022) · IEEE Access · doi:10.1109/access.2022.3203186 · Cited by 19
Hand tracking using wearable wrist imaging
Lu, G., Kim, S., Chen, X., Zeng, Y., Li, D., Wang, S., Liu, B., Li, S., Li, R., Deng, B., et al. (2026) · Nature Electronics · doi:10.1038/s41928-026-01594-4 · Cited by 16
Harrison, A., Jester, A., Mouli, S., Fratini, A., & Jabran, A. (2024) · Sensors · doi:10.3390/s25010002 · Cited by 15
Design and functional evaluation of an epidermal strain sensing system for hand tracking
Michaud, H. O., Dejace, L., de Mulatier, S., & Lacour, S. P. (2016) · 2016 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) · doi:10.1109/iros.2016.7759492 · Cited by 13
A Comparative Evaluation of Bend Sensors for Wearable Applications
Dunne, L. E., Smyth, B., & Caulfield, B. (2007) · Proceedings · doi:10.1109/iswc.2007.4373797 · Cited by 11
Bonifati, P., Baracca, M., Menolotto, M., Averta, G., & Bianchi, M. (2023) · Sensors · doi:10.3390/s23073716 · Cited by 10
Khanna, P., Oppenheim, T., Tu-Chan, A., Abrams, G., & Ganguly, K. (2023) · Neurorehabilitation and Neural Repair · doi:10.1177/15459683231179173 · Cited by 7
Self-citationChatzıoglou, G. N., Pinar, Y. A., & Gövsa, F. (2024) · Anatomy & Cell Biology · doi:10.5115/acb.23.310 · Cited by 3
ATHENA: Automatically tracking hands expertly with no annotations
Mulla, D. M., Costantino, M., Freud, E., & Michaels, J. A. (2025) · Journal of Neurophysiology · doi:10.1152/jn.00407.2025 · Cited by 2
TrackStudio: An integrated toolkit for markerless tracking
Dimitrov, H., Pavalkyte, V., Dominijanni, G., & Makin, T. R. (2026) · PLOS One · doi:10.1371/journal.pone.0358647
Research works printed but not linked: 1
- Kuipers, J.B. (2007). Quaternions and Rotation Sequences: A Primer with Applications to Orbits, Aerospace and Virtual Reality, Princeton University Press.
Cited by
Still being checked.
Metadata
- Authors
- Oppenheim, T., Schlegel, H., Xie, P. Y., Khokhar, Z., & Khanna, P.
- Affiliations
- California Maritime Academy (US)
- University of California, Berkeley (US)
- Corresponding author
- Khanna, Preeya
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2026-07-28
- Volume
- 26
- Issue
- 15
- Pages
- 4795
- ISSN
- 1424-8220 (ISSN-L)
- PMC
- PMC13469233
- OpenAlex
- W7171515028
- Semantic Scholar
- 62d8d0670a235ae1bdc6c4f823a9943d2302c728
- Cited by
- 0
- References
- 48
- FWCI
- 0.00
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Sensors (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion)
- Funders
- Eunice Kennedy Shriver National Institute of Child Health & Human Development of the National Institutes of Health (P2CHD101899)
- Weill Neurohub Next Great Ideas Program (1DP2NS142717-01)
- Topics
- Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Context-Aware Activity Recognition Systems
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Health Sciences › Medicine › Rehabilitation › Stroke Rehabilitation and Recovery
- Keywords
- wearable sensors, hand kinematics, electronic skin, inertial measurement unit, stroke rehabilitation, neurorehabilitation, low-cost wearable device, donning and doffing, user feedback
- MeSH terms
- Wearable Electronic Devices, Motion Capture, Biomechanical Phenomena, Hand, Humans, Movement
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
A language model drew the claims from the paper on 2026-11-18, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-11-18 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.