Real-time intradermal continuous glucose monitoring using a minimally invasive microneedle-based system
Open accessThe DOI matches its Crossref record; 51 of the 51 references it cites were found in a registry or an index.
Abstract
Continuous glucose monitoring (CGM) has the potential to greatly improve diabetes management. The aim of this work is to show a proof-of-concept CGM device which performs minimally invasive and minimally delayed in-situ glucose sensing in the dermal interstitial fluid, combining the advantages of microneedle-based and commercially available CGM systems. The device is based on the integration of an ultra-miniaturized electrochemical sensing probe in the lumen of a single hollow microneedle, separately realized using standard silicon microfabrication methods. By placing the sensing electrodes inside the lumen facing an opening towards the dermal space, real-time measurement purely can be performed relying on molecular diffusion over a short distance. Furthermore, the device relies only on passive capillary lumen filling without the need for complex fluid extraction mechanisms. Importantly, the transdermal portion of the device is 50 times smaller than that of commercial products. This allows access to the dermis and simultaneously reduces tissue trauma, along with being virtually painless during insertion. The three-electrode enzymatic sensor alone was previously proven to have satisfactory sensitivity (1.5 nA/mM), linearity (up to 14 mM), selectivity, and long-term stability (up to 4 days) in-vitro. In this work we combine this sensor technology with microneedles for reliable insertion in forearm skin. In-vivo human tests showed the possibility to correctly and dynamically track glycaemia over time, with approximately 10 min delay with respect to capillary blood control values, in line with the expected physiological lag time. The proposed device can thus reduce discomfort and potentially enable less invasive real-time CGM in diabetic patients.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667
Electrochemical Glucose Biosensors
Wang, J. (2008) · Chemical Reviews · doi:10.1021/cr068123a · Cited by 3,498
Electrochemical Glucose Sensors and Their Applications in Diabetes Management
Heller, A., & Feldman, B. (2008) · Chemical Reviews · doi:10.1021/cr068069y · Cited by 1,707 · cited 1×
“Moreover, several commercial CGM devices have been presented (Csöregi et al. 1995; Feldman et al. 2003; Heller and Feldman 2008; Rodbard 2016; Yoo and Lee 2010) and released to the market since the early 2000s.”
Non-invasive wearable electrochemical sensors: a review
Bandodkar, A. J., & Wang, J. (2014) · Trends in Biotechnology · doi:10.1016/j.tibtech.2014.04.005 · Cited by 1,247 · cited 1×
“In particular, to performminimally invasive glucose sensing, transdermal electrochemical sensors remain the most common and reliable solution (Bandodkar and Wang 2014; Heller and Feldman 2010; Vashist et al. 2011; Wang 2008).”
Glucose Biosensors: An Overview of Use in Clinical Practice
Yoo, E.-H., & Lee, S.-Y. (2010) · Sensors · doi:10.3390/s100504558 · Cited by 1,137 · cited 1×
“Moreover, several commercial CGM devices have been presented (Csöregi et al. 1995; Feldman et al. 2003; Heller and Feldman 2008; Rodbard 2016; Yoo and Lee 2010) and released to the market since the early 2000s.”
Larrañeta, E., Lutton, R. E. M., Woolfson, A. D., & Donnelly, R. F. (2016) · Materials Science and Engineering: R: Reports · doi:10.1016/j.mser.2016.03.001 · Cited by 964 · cited 1×
“Finally, microneedle arrays for intradermal drug delivery, such as insulin injection, have previously been reported (Ita 2015; Larrañeta et al. 2016; Van Der Maaden et al. 2012; Roxhed et al. 2008), and this feature combined with microneedle-based CGM could enable a minimally invasive closed-loop…”
Continuous Glucose Monitoring: A Review of Successes, Challenges, and Opportunities
Rodbard, D. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0417 · Cited by 737 · cited 1×
“Moreover, several commercial CGM devices have been presented (Csöregi et al. 1995; Feldman et al. 2003; Heller and Feldman 2008; Rodbard 2016; Yoo and Lee 2010) and released to the market since the early 2000s.”
Microneedle technologies for (trans)dermal drug and vaccine delivery
van der Maaden, K., Jiskoot, W., & Bouwstra, J. (2012) · Journal of Controlled Release · doi:10.1016/j.jconrel.2012.01.042 · Cited by 651 · cited 2×
“Finally, microneedle arrays for intradermal drug delivery, such as insulin injection, have previously been reported (Ita 2015; Larrañeta et al. 2016; Van Der Maaden et al. 2012; Roxhed et al. 2008), and this feature combined with microneedle-based CGM could enable a minimally invasive closed-loop system.”
Transdermal Delivery of Drugs with Microneedles—Potential and Challenges
Ita, K. (2015) · Pharmaceutics · doi:10.3390/pharmaceutics7030090 · Cited by 530 · cited 1×
“Finally, microneedle arrays for intradermal drug delivery, such as insulin injection, have previously been reported (Ita 2015; Larrañeta et al. 2016; Van Der Maaden et al. 2012; Roxhed et al. 2008), and this feature combined with microneedle-based CGM could enable a minimally invasive closed-loop…”
Boyne, M. S., Silver, D. M., Kaplan, J., & Saudek, C. D. (2003) · Diabetes · doi:10.2337/diabetes.52.11.2790 · Cited by 404 · cited 1×
“…lag time between blood glucose and interstitial fluid glucose concentrations, which is of the order of 4 to 12 min (Aussedat et al. 2000; Boyne et al. 2003), and the CGMS delay itself impedes real-time monitoring, and hence the possibility to take medical decisions, such as insulin…”
A Tale of Two Compartments: Interstitial Versus Blood Glucose Monitoring
Cengiz, E., & Tamborlane, W. V. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2009.0002 · Cited by 390 · cited 1×
“Additionally, due to their size, monitoring is performed in the hypodermis, which is not an ideal location for reliable and fast glycaemia tracking, due to its heterogeneity and the scarce local distribution of capillaries compared to the dermal region (Cengiz and Tamborlane 2009).”
Microneedle array for transdermal biological fluid extraction and in situ analysis
Mukerjee, E. V., Collins, S. D., Isseroff, R. R., & Smith, R. L. (2004) · Sensors and Actuators A: Physical · doi:10.1016/j.sna.2003.11.008 · Cited by 340 · cited 1×
“…measurement delay is typically introduced by the time required for the glucose molecules to diffuse from the interstitial fluid to the electrode area, typically situated in a compartment on the backside of the microneedle chip (Jina et al. 2014; Mukerjee et al. 2004; Strambini et al. 2015).”
Microfabricated Reference Electrodes and their Biosensing Applications
Shinwari, M. W., Zhitomirsky, D., Deen, I. A., Selvaganapathy, P. R., Deen, M. J., & Landheer, D. (2010) · Sensors · doi:10.3390/s100301679 · Cited by 333 · cited 2×
“…working electrode, biased with respect to an integrated iridium oxide (IrOx) pseudoreference electrode (pseudo-RE) (Ribet et al. 2017; Waleed Shinwari et al. 2010), thus providing a current proportional to the initial glucose concentration, according to the reactions: Glucoseþ O2 →GOx…”
Minimally Invasive Extraction of Dermal Interstitial Fluid for Glucose Monitoring Using Microneedles
Wang, P. M., Cornwell, M., & Prausnitz, M. R. (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.131 · Cited by 267 · cited 2×
“In fact, interstitial fluid (ISF) is more abundant in the dermis, and dermal glucose concentration dynamics more closely follow oscillations in the blood (Groenendaal et al. 2008; Wang et al. 2005).”
Electrochemistry in Diabetes Management
Heller, A., & Feldman, B. (2010) · Accounts of Chemical Research · doi:10.1021/ar9002015 · Cited by 255 · cited 1×
“In particular, to performminimally invasive glucose sensing, transdermal electrochemical sensors remain the most common and reliable solution (Bandodkar and Wang 2014; Heller and Feldman 2010; Vashist et al. 2011; Wang 2008).”
Dunn, T. C., Xu, Y., Hayter, G., & Ajjan, R. A. (2018) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2017.12.015 · Cited by 239 · cited 1×
“However, continuous glucose monitoring (CGM) has been shown to significantly improve the quality of diabetes treatment, providing continuous real-time measurements and a complete temporal overview (Dunn et al. 2018; Nathan et al. 1993).”
Microneedles for Transdermal Biosensing: Current Picture and Future Direction
Ventrelli, L., Marsilio Strambini, L., & Barillaro, G. (2015) · Advanced Healthcare Materials · doi:10.1002/adhm.201500450 · Cited by 235 · cited 1×
“Therefore, microneedle-based CGMS are gaining attention because of their reduced invasiveness (Invernale et al. 2014; Miller et al. 2012, 2016b), resulting in a virtually painless insertion procedure and a reduced tissue trauma and inflammation (Ventrelli et al. 2015).”
Improved piercing of microneedle arrays in dermatomed human skin by an impact insertion method
Verbaan, F. J., Bal, S. M., van den Berg, D. J., Dijksman, J. A., van Hecke, M., Verpoorten, H., van den Berg, A., Luttge, R., & Bouwstra, J. A. (2008) · Journal of Controlled Release · doi:10.1016/j.jconrel.2008.02.009 · Cited by 233 · cited 1×
“In fact, the insertion speed plays a fundamental role in piercing the skin and achieving deep and reliable insertion (Verbaan et al. 2008).”
Technology behind commercial devices for blood glucose monitoring in diabetes management: A review
Vashist, S. K., Zheng, D., Al-Rubeaan, K., Luong, J. H. T., & Sheu, F.-S. (2011) · Analytica Chimica Acta · doi:10.1016/j.aca.2011.07.024 · Cited by 209 · cited 1×
“In particular, to performminimally invasive glucose sensing, transdermal electrochemical sensors remain the most common and reliable solution (Bandodkar and Wang 2014; Heller and Feldman 2010; Vashist et al. 2011; Wang 2008).”
Roxhed, N., Gasser, T. C., Griss, P., Holzapfel, G. A., & Stemme, G. (2007) · Journal of Microelectromechanical Systems · doi:10.1109/jmems.2007.907461 · Cited by 199 · cited 2×
“Themicroneedle geometry was previously developed to have a sharp tip geometry for easy penetration and a side opening to avoid tissue clogging the lumen during insertion (Roxhed et al. 2007).”
Self-citationAussedat, B., Dupire-Angel, M., Gifford, R., Klein, J. C., Wilson, G. S., & Reach, G. (2000) · American Journal of Physiology-Endocrinology and Metabolism · doi:10.1152/ajpendo.2000.278.4.e716 · Cited by 186 · cited 1×
“…the sum of the physiological lag time between blood glucose and interstitial fluid glucose concentrations, which is of the order of 4 to 12 min (Aussedat et al. 2000; Boyne et al. 2003), and the CGMS delay itself impedes real-time monitoring, and hence the possibility to take medical…”
Feldman, B., Brazg, R., Schwartz, S., & Weinstein, R. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503322526978 · Cited by 184 · cited 1×
“Moreover, several commercial CGM devices have been presented (Csöregi et al. 1995; Feldman et al. 2003; Heller and Feldman 2008; Rodbard 2016; Yoo and Lee 2010) and released to the market since the early 2000s.”
Diabetes Mellitus and its Chronic Complications
Bailes, B. K. (2002) · AORN Journal · doi:10.1016/s0001-2092(06)61065-x · Cited by 178 · cited 1×
“In order to avoid diabetes-induced complications such as retinal damage, cardiovasculardiseasesandneuropathies,carefulglycaemiamonitoring andmanagement isessential (Bailes2002).”
Design, Development, and Evaluation of a Novel Microneedle Array-based Continuous Glucose Monitor
Jina, A., Tierney, M. J., Tamada, J. A., McGill, S., Desai, S., Chua, B., Chang, A., & Christiansen, M. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814526191 · Cited by 176 · cited 1×
“…measurement delay is typically introduced by the time required for the glucose molecules to diffuse from the interstitial fluid to the electrode area, typically situated in a compartment on the backside of the microneedle chip (Jina et al. 2014; Mukerjee et al. 2004; Strambini et al. 2015).”
Crichton, M. L., Donose, B. C., Chen, X., Raphael, A. P., Huang, H., & Kendall, M. A. F. (2011) · Biomaterials · doi:10.1016/j.biomaterials.2011.03.012 · Cited by 171 · cited 1×
“This is due to the ability of the skin to elastically deform under the applied pressure (Crichton et al. 2011).”
Characterization of Microchannels Created by Metal Microneedles: Formation and Closure
Kalluri, H., Kolli, C. S., & Banga, A. K. (2011) · The AAPS Journal · doi:10.1208/s12248-011-9288-3 · Cited by 165 · cited 1×
“There, the skin is easily accessible, the epidermis relatively thin and easy to penetrate, with an average thickness of 74 μm (Hwang et al. 2016), and the 1-mm-thick dermal region can be reliably reached with 700 μm long microneedles (Coulman et al. 2011; Kalluri et al. 2011).”
Glucose Variability: A Review of Clinical Applications and Research Developments
Rodbard, D. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0092 · Cited by 154
Strambini, L. M., Longo, A., Scarano, S., Prescimone, T., Palchetti, I., Minunni, M., Giannessi, D., & Barillaro, G. (2014) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2014.11.010 · Cited by 154 · cited 1×
“…measurement delay is typically introduced by the time required for the glucose molecules to diffuse from the interstitial fluid to the electrode area, typically situated in a compartment on the backside of the microneedle chip (Jina et al. 2014; Mukerjee et al. 2004; Strambini et al. 2015).”
Painless Drug Delivery Through Microneedle-Based Transdermal Patches Featuring Active Infusion
Roxhed, N., Samel, B., Nordquist, L., Griss, P., & Stemme, G. (2008) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2007.906492 · Cited by 151 · cited 2×
“The same microneedles were previously successfully used for insulin delivery (Roxhed et al. 2008), as well as intradermal injection of rabies vaccine in humans in a clinical study (Vescovo et al. 2017).”
Self-citationMicroneedle-based sensors for medical diagnosis
Miller, P. R., Narayan, R. J., & Polsky, R. (2016) · Journal of Materials Chemistry B · doi:10.1039/c5tb02421h · Cited by 143 · cited 2×
“…need for complex interstitial fluid extraction mechanisms based on vacuum suction, microfluidic valves or liquid pre-filling (Gattiker et al. 2005; Miller et al. 2016a; Ricci et al. 2008; Vesper et al. 2006; Wang et al. 2005; Yu et al. 2012; Zimmermann et al. 2004), the risk of clogging of the…”
Coulman, S. A., Birchall, J. C., Alex, A., Pearton, M., Höfer, B., O’Mahony, C., Drexler, W., & Považay, B. (2010) · Pharmaceutical Research · doi:10.1007/s11095-010-0167-x · Cited by 139 · cited 1×
“There, the skin is easily accessible, the epidermis relatively thin and easy to penetrate, with an average thickness of 74 μm (Hwang et al. 2016), and the 1-mm-thick dermal region can be reliably reached with 700 μm long microneedles (Coulman et al. 2011; Kalluri et al. 2011).”
Evaluation of a minimally invasive glucose biosensor for continuous tissue monitoring
Sharma, S., Huang, Z., Rogers, M., Boutelle, M., & Cass, A. E. G. (2016) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-016-9961-6 · Cited by 132 · cited 1×
“…et al. 2005; Yu et al. 2012; Zimmermann et al. 2004), the risk of clogging of the needle opening during insertion or unreliable skin penetration (Chua et al. 2013), and the need for complex electrode functionalization procedures on non-flat surfaces (Sharma et al. 2016; Windmiller et al. 2011).”
Use of Microneedle Array Devices for Continuous Glucose Monitoring: A Review
El-Laboudi, A., Oliver, N. S., Cass, A., & Johnston, D. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2012.0188 · Cited by 131 · cited 1×
“The size of a transdermally implanted device affects the tissue trauma during insertion and the degree of inflammation over time, both of which cause an increased foreign body reaction that can be detrimental for long-termmeasurements (El-Laboudi et al. 2013;Wang et al. 2015).”
Csoeregi, E., Schmidtke, D. W., & Heller, A. (1995) · Analytical Chemistry · doi:10.1021/ac00103a015 · Cited by 129 · cited 1×
“Moreover, several commercial CGM devices have been presented (Csöregi et al. 1995; Feldman et al. 2003; Heller and Feldman 2008; Rodbard 2016; Yoo and Lee 2010) and released to the market since the early 2000s.”
Bicomponent Microneedle Array Biosensor for Minimally‐Invasive Glutamate Monitoring
Windmiller, J. R., Valdés‐Ramírez, G., Zhou, N., Zhou, M., Miller, P. R., Jin, C., Brozik, S. M., Polsky, R., Katz, E., Narayan, R., et al. (2011) · Electroanalysis · doi:10.1002/elan.201100361 · Cited by 129 · cited 1×
“…et al. 2005; Yu et al. 2012; Zimmermann et al. 2004), the risk of clogging of the needle opening during insertion or unreliable skin penetration (Chua et al. 2013), and the need for complex electrode functionalization procedures on non-flat surfaces (Sharma et al. 2016; Windmiller et al. 2011).”
Chua, B., Desai, S. P., Tierney, M. J., Tamada, J. A., & Jina, A. N. (2013) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2013.09.026 · Cited by 121 · cited 1×
“…et al. 2005; Yu et al. 2012; Zimmermann et al. 2004), the risk of clogging of the needle opening during insertion or unreliable skin penetration (Chua et al. 2013), and the need for complex electrode functionalization procedures on non-flat surfaces (Sharma et al. 2016; Windmiller et al. 2011).”
Foreign Body Reaction to Implantable Biosensors: Effects of Tissue Trauma and Implant Size
Wang, Y., Vaddiraju, S., Gu, B., Papadimitrakopoulos, F., & Burgess, D. J. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815601869 · Cited by 107 · cited 1×
“The size of a transdermally implanted device affects the tissue trauma during insertion and the degree of inflammation over time, both of which cause an increased foreign body reaction that can be detrimental for long-termmeasurements (El-Laboudi et al. 2013;Wang et al. 2015).”
Real-Time Improvement of Continuous Glucose Monitoring Accuracy
Facchinetti, A., Sparacino, G., Guerra, S., Luijf, Y. M., DeVries, J. H., Mader, J. K., Ellmerer, M., Benesch, C., Heinemann, L., Bruttomesso, D., et al. (2013) · Diabetes Care · doi:10.2337/dc12-0736 · Cited by 106 · cited 1×
“Such delay can be considered acceptable and potentially further mitigated by smart modeling and data trend analysis (Facchinetti et al. 2013).”
Microneedle Electrodes Toward an Amperometric Glucose‐Sensing Smart Patch
Invernale, M. A., Tang, B. C., York, R. L., Le, L., Hou, D. Y., & Anderson, D. G. (2014) · Advanced Healthcare Materials · doi:10.1002/adhm.201300142 · Cited by 99 · cited 1×
“Therefore, microneedle-based CGMS are gaining attention because of their reduced invasiveness (Invernale et al. 2014; Miller et al. 2012, 2016b), resulting in a virtually painless insertion procedure and a reduced tissue trauma and inflammation (Ventrelli et al. 2015).”
Ribet, F., Stemme, G., & Roxhed, N. (2016) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2016.10.007 · Cited by 69 · cited 6×
“…in the lumen during measurement, which is especially important for in-vivo experiments; the latter allows for distinct functionalization of the three electrodes when dispensing the membrane solutions, potentially improving the sensor performance, as discussed in previous work (Ribet et al. 2017).”
Self-citationDunn, T. C., Eastman, R. C., & Tamada, J. A. (2004) · Diabetes Care · doi:10.2337/diacare.27.9.2161 · Cited by 55 · cited 1×
“Moreover, this translates into a maximum sensing variation rate of 9.5 mg/ dL/min, which is well above the typical maximum physiological glucose variation rate (3 mg/dL/min) (Dunn et al. 2004), even allowing tracking of extreme glycemic variations.”
InfluentialGattiker, G. E., Kaler, K. V. I. S., & Mintchev, M. P. (2005) · Microsystem Technologies · doi:10.1007/s00542-005-0015-9 · Cited by 52 · cited 1×
“…issues include the need for complex interstitial fluid extraction mechanisms based on vacuum suction, microfluidic valves or liquid pre-filling (Gattiker et al. 2005; Miller et al. 2016a; Ricci et al. 2008; Vesper et al. 2006; Wang et al. 2005; Yu et al. 2012; Zimmermann et al. 2004), the risk…”
Towards an Integrated Microneedle Total Analysis Chip for Protein Detection
Miller, P. R., Moorman, M. W., Manginell, R., Ashlee, C., Brener, I., Wheeler, D. R., Narayan, R. J., & Polsky, R. (2016) · Electroanalysis · doi:10.1002/elan.201600063 · Cited by 48 · cited 2×
“…need for complex interstitial fluid extraction mechanisms based on vacuum suction, microfluidic valves or liquid pre-filling (Gattiker et al. 2005; Miller et al. 2016a; Ricci et al. 2008; Vesper et al. 2006; Wang et al. 2005; Yu et al. 2012; Zimmermann et al. 2004), the risk of clogging of the…”
Safety, tolerability and efficacy of intradermal rabies immunization with DebioJect™
Vescovo, P., Rettby, N., Ramaniraka, N., Liberman, J., Hart, K., Cachemaille, A., Piveteau, L.-D., Zanoni, R., Bart, P., & Pantaleo, G. (2017) · Vaccine · doi:10.1016/j.vaccine.2016.09.069 · Cited by 42 · cited 1×
“The same microneedles were previously successfully used for insulin delivery (Roxhed et al. 2008), as well as intradermal injection of rabies vaccine in humans in a clinical study (Vescovo et al. 2017).”
Hollow Microneedle-based Sensor for Multiplexed Transdermal Electrochemical Sensing
Miller, P. R., Skoog, S. A., Edwards, T. L., Wheeler, D. R., Xiao, X., Brozik, S. M., Polsky, R., & Narayan, R. J. (2012) · Journal of Visualized Experiments · doi:10.3791/4067 · Cited by 40
Modeling Glucose and Water Dynamics in Human Skin
Groenendaal, W., Schmidt, K. A., von Basum, G., van Riel, N. A. W., & Hilbers, P. A. J. (2008) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2007.0290 · Cited by 34 · cited 1×
“In fact, interstitial fluid (ISF) is more abundant in the dermis, and dermal glucose concentration dynamics more closely follow oscillations in the blood (Groenendaal et al. 2008; Wang et al. 2005).”
Current Trends in Continuous Glucose Monitoring
Lodwig, V., Kulzer, B., Schnell, O., & Heinemann, L. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814525826 · Cited by 33 · cited 1×
“In addition, CGMcan also reduce collateral effects of finger pricks, such as pain, infections, and sensory loss (Le Floch et al. 2008; Lodwig et al. 2014), as well as the patients’ discomfort and thus potentially improve their quality of life.”
An Interstitial Fluid Transdermal Extraction System for Continuous Glucose Monitoring
Yu, H., Li, D., Roberts, R. C., Xu, K., & Tien, N. C. (2012) · Journal of Microelectromechanical Systems · doi:10.1109/jmems.2012.2192910 · Cited by 30 · cited 1×
“…microfluidic valves or liquid pre-filling (Gattiker et al. 2005; Miller et al. 2016a; Ricci et al. 2008; Vesper et al. 2006; Wang et al. 2005; Yu et al. 2012; Zimmermann et al. 2004), the risk of clogging of the needle opening during insertion or unreliable skin penetration (Chua et al.…”
Vesper, H. W., Wang, P. M., Archibold, E. F., Prausnitz, M. R., & Myers, G. L. (2006) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2006.8.76 · Cited by 27 · cited 1×
“…mechanisms based on vacuum suction, microfluidic valves or liquid pre-filling (Gattiker et al. 2005; Miller et al. 2016a; Ricci et al. 2008; Vesper et al. 2006; Wang et al. 2005; Yu et al. 2012; Zimmermann et al. 2004), the risk of clogging of the needle opening during insertion or…”
Ex Vivo Continuous Glucose Monitoring With Microdialysis Technique: The Example of GlucoDay
Ricci, F., Moscone, D., & Palleschi, G. (2008) · IEEE Sensors Journal · doi:10.1109/jsen.2007.912535 · Cited by 24 · cited 1×
“…fluid extraction mechanisms based on vacuum suction, microfluidic valves or liquid pre-filling (Gattiker et al. 2005; Miller et al. 2016a; Ricci et al. 2008; Vesper et al. 2006; Wang et al. 2005; Yu et al. 2012; Zimmermann et al. 2004), the risk of clogging of the needle opening during…”
Self-Monitoring of Blood Glucose, Cutaneous Finger Injury, and Sensory Loss in Diabetic Patients
Floch, J. P. L., Bauduceau, B., Lévy, M., Mosnier-Pudar, H., Sachon, C., & Kakou, B. (2008) · Diabetes Care · doi:10.2337/dc08-1174 · Cited by 23
Cited by
Continuous glucose monitoring and metrics for clinical trials: an international consensus statement
Battelino, T., Alexander, C. M., Amiel, S. A., Arreaza-Rubin, G., Beck, R. W., Bergenstal, R. M., Buckingham, B. A., Carroll, J., Ceriello, A., Chow, E., et al. (2023) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(22)00319-9 · Cited by 831
Villena Gonzales, W., Mobashsher, A., & Abbosh, A. (2019) · Sensors · doi:10.3390/s19040800 · Cited by 772
Tehrani, F., Teymourian, H., Wuerstle, B., Kavner, J., Patel, R., Furmidge, A., Aghavali, R., Hosseini-Toudeshki, H., Brown, C., Zhang, F., et al. (2022) · Nature Biomedical Engineering · doi:10.1038/s41551-022-00887-1 · Cited by 625
Non-Invasive Blood Glucose Monitoring Technology: A Review
Tang, L., Chang, S. J., Chen, C.-J., & Liu, J.-T. (2020) · Sensors · doi:10.3390/s20236925 · Cited by 481
Broza, Y. Y., Zhou, X., Yuan, M., Qu, D., Zheng, Y., Vishinkin, R., Khatib, M., Wu, W., & Haick, H. (2019) · Chemical Reviews · doi:10.1021/acs.chemrev.9b00437 · Cited by 461
Metadata
- Authors
- Ribet, F., Stemme, G., & Roxhed, N.
- Venue
- Biomedical Microdevices
- Published
- 2018-12
- Volume
- 20
- Issue
- 4
- Pages
- 101-101
- ISSN
- 1387-2176 (ISSN-L)
- 1572-8781
- OpenAlex
- W2904140750
- Cited by
- 181
- References
- 51
- FWCI
- 4.42
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.