Improved elimination of motion artifacts from a photoplethysmographic signal using a Kalman smoother with simultaneous accelerometry
Open accessThe DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.
Abstract
A photoplethysmography (PPG) signal provides very useful information about a subject's hemodynamic status in a hospital or ubiquitous environment. However, PPG is very vulnerable to motion artifacts, which can significantly distort the information belonging to the PPG signal itself. Thus, the reduction of the effects of motion artifacts is an important issue when monitoring the cardiovascular system by PPG. There have been many adaptive techniques to reduce motion artifacts from PPG signals. In the present study, we compared a method based on the fixed-interval Kalman smoother with the usual adaptive filtering algorithms, e.g. the normalized least mean squares, recursive least squares and the conventional Kalman filter. We found that the fixed-interval Kalman smoother reduced motion artifacts from the PPG signal most effectively. Therefore, the use of the fixed-interval Kalman smoother can reduce motion artifacts in PPG, thus providing the most reliable information that can be deduced from the reconstructed PPG signals.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Optimal State Estimation: Kalman, H∞, and Nonlinear Approaches
Simon, D. (2006) · doi:10.1002/0470045345 · Cited by 4,875
Mobile monitoring with wearable photoplethysmographic biosensors
Asada, H. H., Shaltis, P., Reisner, A., Rhee, S., & Hutchinson, R. C. (2003) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/memb.2003.1213624 · Cited by 466
A state-space approach to adaptive RLS filtering
Sayed, A. H., & Kailath, T. (1994) · IEEE Signal Processing Magazine · doi:10.1109/79.295229 · Cited by 449 · cited 2×
“From the point of computation and memory demand, NLMS has an order of complexity O(M) per iteration while RLS and Kalman filter have an order of O(M2) per iteration (Lanzi et al 2006, Sayed and Kailath 1994).”
BackgroundDevelopment of real-time motion artifact reduction algorithm for a wearable photoplethysmography
Han, H., Kim, M.-J., & Kim, J. (2007) · 2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2007.4352596 · Cited by 101
Cited by
Ghamari, M. (2018) · International Journal of Biosensors & Bioelectronics · doi:10.15406/ijbsbe.2018.04.00125 · Cited by 945
Wearable photoplethysmographic sensors—past and present
Tamura, T., Maeda, Y., Sekine, M., & Yoshida, M. (2014) · Electronics · doi:10.3390/electronics3020282 · Cited by 912
“, NLMS, RLS, and conventional Kalman filter algorithms), can be used to estimate motion artifact reduction, and has been shown to provide reliable information from the reconstructed PPG signals [63].”
MethodZhang, Z., Pi, Z., & Liu, B. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2359372 · Cited by 748
Wavelet-based motion artifact removal for functional near-infrared spectroscopy
Molavi, B., & Dumont, G. A. (2012) · Physiological Measurement · doi:10.1088/0967-3334/33/2/259 · Cited by 623
Zhang, Z. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2406332 · Cited by 388
“cancelation [7], [8], spectrum subtraction [9], Kalman filtering [10], wavelet denoising [11], and empirical mode decomposition [12], to name a few.”
MethodDeep PPG: Large-Scale Heart Rate Estimation with Convolutional Neural Networks
Reiss, A., Indlekofer, I., Schmidt, P., & Van Laerhoven, K. (2019) · Sensors · doi:10.3390/s19143079 · Cited by 377
Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring
Fine, J., Branan, K. L., Rodriguez, A. J., Boonya-ananta, T., Ajmal, Ramella-Roman, J. C., McShane, M. J., & Coté, G. L. (2021) · Biosensors · doi:10.3390/bios11040126 · Cited by 348
Accurate Heart Rate Monitoring During Physical Exercises Using PPG
Temko, A. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2017.2676243 · Cited by 293
Biswas, D., Everson, L., Liu, M., Panwar, M., Verhoef, B.-E., Patki, S., Kim, C. H., Acharyya, A., Van Hoof, C., Konijnenburg, M., et al. (2019) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2019.2892297 · Cited by 288
Heart Rate Estimation From Wrist-Worn Photoplethysmography: A Review
Biswas, D., Simoes-Capela, N., Van Hoof, C., & Van Helleputte, N. (2019) · IEEE Sensors Journal · doi:10.1109/jsen.2019.2914166 · Cited by 258
Current progress of photoplethysmography and SPO2 for health monitoring
Tamura, T. (2019) · Biomedical Engineering Letters · doi:10.1007/s13534-019-00097-w · Cited by 247
A review of signals used in sleep analysis
Roebuck, A., Monasterio, V, Gederi, E., Osipov, M., Behar, J., Malhotra, A., Penzel, T., & Clifford, G. D. (2014) · Physiological Measurement · doi:10.1088/0967-3334/35/1/r1 · Cited by 223
“Rather than incorporating sophisticated algorithms into pulse oximeter systems, an alternative approach is to pass the output of pulse oximeters through a postprocessing step for artefact rejection (Lee et al 2010, Krishnan et al 2010, Sukor et al 2011).”
MethodWearable-Based Affect Recognition—A Review
Schmidt, P., Reiss, A., Dürichen, R., & Laerhoven, K. V. (2019) · Sensors · doi:10.3390/s19194079 · Cited by 218
“In order to remove motion artefacts, adaptive (filtering) approaches can be applied [142,143].”
MethodGarbarino, M., Lai, M., Tognetti, S., Picard, R., & Bender, D. (2014) · Proceedings of the 4th International Conference on Wireless Mobile Communication and Healthcare - "Transforming healthcare through innovations in mobile and wireless technologies" · doi:10.4108/icst.mobihealth.2014.257418 · Cited by 215
“Standard techniques for artifact removal in wrist-worn PPG systems make use of accelerometers [1], [9].”
MethodKim, K. B., & Baek, H. J. (2023) · Electronics · doi:10.3390/electronics12132923 · Cited by 212
Self-citationA Smart Health Monitoring Chair for Nonintrusive Measurement of Biological Signals
Baek, H. J., Chung, G. S., Kim, K. K., & Park, K. S. (2012) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2011.2175742 · Cited by 199
“Remarkably, the estimated BP tended to conform to the actual measured BP.”
BackgroundSelf-citationSalehizadeh, S., Dao, D., Bolkhovsky, J., Cho, C., Mendelson, Y., & Chon, K. (2015) · Sensors · doi:10.3390/s16010010 · Cited by 173
“Acceleration data can be also used to reconstruct the observation model for Kalman filtering [25] to remove MA.”
MethodKhan, E., Al Hossain, F., Uddin, S. Z., Alam, S. K., & Hasan, M. K. (2016) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2466075 · Cited by 153
Photoplethysmogram signal quality estimation using repeated Gaussian filters and cross-correlation
Karlen, W., Kobayashi, K., Ansermino, J. M., & Dumont, G. A. (2012) · Physiological Measurement · doi:10.1088/0967-3334/33/10/1617 · Cited by 123
Chong, J. W., Dao, D. K., Salehizadeh, S. M. A., McManus, D. D., Darling, C. E., Chon, K. H., & Mendelson, Y. (2014) · Annals of Biomedical Engineering · doi:10.1007/s10439-014-1080-y · Cited by 79
A Review of Methods for Non-Invasive Heart Rate Measurement on Wrist
De Pinho Ferreira, N., Gehin, C., & Massot, B. (2021) · IRBM · doi:10.1016/j.irbm.2020.04.001 · Cited by 49
Reducing motion artifacts in photoplethysmograms by using relative sensor motion: Phantom study
Wijshoff, R. W. C. G. R., Mischi, M., Veen, J., & Van Der Lee, A. M. (2012) · Journal of Biomedical Optics · doi:10.1117/1.jbo.17.11.117007 · Cited by 40
Xu, K., Jiang, X., Lin, S., Dai, C., & Chen, W. (2020) · IEEE Transactions on Industrial Informatics · doi:10.1109/tii.2020.2988097 · Cited by 37
“It should be noticed that the previous research on Kalman filter [24] involves only a random walking model for signal modeling, which has”
Method
Metadata
- Authors
- Lee, B., Han, J., Baek, H. J., Shin, J. H., Park, K. S., & Yi, W. J.
- Venue
- Physiological Measurement
- Publisher
- IOP Publishing
- Published
- 2010-10-27
- Volume
- 31
- Issue
- 12
- Pages
- 1585-1603
- ISSN
- 0967-3334 (ISSN-L)
- 1361-6579
- OpenAlex
- W2107994016
- Cited by
- 154
- References
- 4
- FWCI
- 1.27
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.