Improved Accuracy of Continuous Glucose Monitoring Systems in Pediatric Patients with Diabetes Mellitus: Results from Two Studies
Abstract
Objective: This study was designed to evaluate accuracy, performance, and safety of the Dexcom (San Diego, CA) G4 ® Platinum continuous glucose monitoring (CGM) system (G4P) compared with the Dexcom G4 Platinum with Software 505 algorithm (SW505) when used as adjunctive management to blood glucose (BG) monitoring over a 7-day period in youth, 2–17 years of age, with diabetes. Research Design and Methods: Youth wore either one or two sensors placed on the abdomen or upper buttocks for 7 days, calibrating the device twice daily with a uniform BG meter. Participants had one in-clinic session on Day 1, 4, or 7, during which fingerstick BG measurements (self-monitoring of blood glucose [SMBG]) were obtained every 30 ± 5 min for comparison with CGM, and in youth 6–17 years of age, reference YSI glucose measurements were obtained from arterialized venous blood collected every 15 ± 5 min for comparison with CGM. The sensor was removed by the participant/family after 7 days. Results: In comparison of 2,922 temporally paired points of CGM with the reference YSI measurement for G4P and 2,262 paired points for SW505, the mean absolute relative difference (MARD) was 17% for G4P versus 10% for SW505 ( P < 0.0001). In comparison of 16,318 temporally paired points of CGM with SMBG for G4P and 4,264 paired points for SW505, MARD was 15% for G4P versus 13% for SW505 ( P < 0.0001). Similarly, error grid analyses indicated superior performance with SW505 compared with G4P in comparison of CGM with YSI and CGM with SMBG results, with greater percentages of SW505 results falling within error grid Zone A or the combined Zones A plus B. There were no serious adverse events or device-related serious adverse events for either the G4P or the SW505, and there was no sensor breakoff. Conclusions: The updated algorithm offers substantial improvements in accuracy and performance in pediatric patients with diabetes. Use of CGM with improved performance has potential to increase glucose time in range and improve glycemic outcomes for youth.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“CGM utilizes subcutaneous sensors for continuous monitoring of glucose levels with a high degree of accuracy. 17 , 18”
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Continuous glucose monitoring and intensive treatment of type 1 diabetes
(2008) · New England Journal of Medicine · doi:10.1056/nejmoa0805017 · Cited by 1,559
Miller, K. M., Foster, N. C., Beck, R. W., Bergenstal, R. M., DuBose, S. N., DiMeglio, L. A., Maahs, D. M., & Tamborlane, W. V. (2015) · Diabetes Care · doi:10.2337/dc15-0078 · Cited by 1,254
Effectiveness of Sensor-Augmented Insulin-Pump Therapy in Type 1 Diabetes
Bergenstal, R. M., Tamborlane, W. V., Ahmann, A., Buse, J. B., Dailey, G., Davis, S. N., Joyce, C., Peoples, T., Perkins, B. A., Welsh, J. B., et al. (2010) · New England Journal of Medicine · doi:10.1056/nejmoa1002853 · Cited by 894
Deiss, D., Bolinder, J., Riveline, J.-P., Battelino, T., Bosi, E., Tubiana-Rufi, N., Kerr, D., & Phillip, M. (2006) · Diabetes Care · doi:10.2337/dc06-1134 · Cited by 539
the SWITCH Study Group, Battelino, T., Conget, I., Olsen, B., Schütz-Fuhrmann, I., Hommel, E., Hoogma, R., Schierloh, U., Sulli, N., & Bolinder, J. (2012) · Diabetologia · doi:10.1007/s00125-012-2708-9 · Cited by 494
The effect of continuous glucose monitoring in well-controlled type 1 diabetes
Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group (2009) · Diabetes Care · doi:10.2337/dc09-0108 · Cited by 399
Glycemic control in youth with diabetes: the SEARCH for Diabetes in Youth Study
Petitti, D. B., Klingensmith, G. J., Bell, R. A., Andrews, J. S., Dabelea, D., Imperatore, G., Marcovina, S., Pihoker, C., Standiford, D., Waitzfelder, B., et al. (2009) · The Journal of Pediatrics · doi:10.1016/j.jpeds.2009.05.025 · Cited by 397
Effectiveness of Continuous Glucose Monitoring in a Clinical Care Environment
Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group (2010) · Diabetes Care · doi:10.2337/dc09-1502 · Cited by 313
Mauras, N., Beck, R., Xing, D., Ruedy, K., Buckingham, B., Tansey, M., White, N. H., Weinzimer, S. A., Tamborlane, W., Kollman, C., et al. (2012) · Diabetes Care · doi:10.2337/dc11-1746 · Cited by 204
Clinical accuracy of a continuous glucose monitoring system with an advanced algorithm
Bailey, T. S., Chang, A., & Christiansen, M. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814559746 · Cited by 168 · cited 2×
“In pediatric patients, the SW505 has an MARD of 10% compared with reference YSI values, closely matching the favorable MARD of 9% in adults with diabetes.(28) Similarly, clinical accuracy based on error grid analyses yielded 98–100% of values falling within Zones A and B in the pediatric patients compared with 99.”
BackgroundResultOptimal sampling intervals to assess long-term glycemic control using continuous glucose monitoring
Xing, D., Kollman, C., Beck, R. W., Tamborlane, W. V., Laffel, L., Buckingham, B. A., Wilson, D. M., Weinzimer, S., Fiallo-Scharer, R., & Ruedy for the Juvenile Diabetes Research Foundation Continuo, K. J. (2011) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2010.0156 · Cited by 131
Sensor-Augmented Pump Therapy for A1C Reduction (STAR 3) Study
Bergenstal, R. M., Tamborlane, W. V., Ahmann, A., Buse, J. B., Dailey, G., Davis, S. N., Joyce, C., Perkins, B. A., Welsh, J. B., Willi, S. M., et al. (2011) · Diabetes Care · doi:10.2337/dc11-1248 · Cited by 118
Relationship of A1C to Glucose Concentrations in Children With Type 1 Diabetes
Diabetes Research in Children Network (DirecNet) Study Group (2008) · Diabetes Care · doi:10.2337/dc07-1835 · Cited by 115
Research works printed but not linked: 25
- 10.1210/jc.2012-1561 · DOI · Paper page
- 10.2337/dc12-1959 · DOI · Paper page
- 10.1016/j.jpeds.2007.03.047 · DOI
- 10.2337/dc07-1995 · DOI
- 10.1111/j.1399-5448.2008.00476.x · DOI
- 10.1089/dia.2010.0021 · DOI
- 10.1111/j.1399-5448.2011.00837.x · DOI · Paper page
- 10.2337/dc09-0889 · DOI · Paper page
- 10.1111/j.1464-5491.2011.03368.x · DOI
- 10.1177/193229681100500407 · DOI
- 10.2337/dc10-1054 · DOI
- 10.1177/1932296814559302 · DOI
- 10.1177/1932296814532203 · DOI
- 10.1089/dia.2013.0077 · DOI
- 10.1159/000411605 · doi:10.1159/000411605
- 10.2337/diacare.10.5.622 · DOI · Paper page
- 10.2337/diacare.23.8.1143 · DOI · Paper page
- 10.1056/NEJMoa1314474 · DOI · Paper page
- 10.2337/dc13-2076 · DOI · Paper page
- 10.2337/diacare.28.8.1929 · DOI
- 10.1111/j.1399-543X.2006.00142.x · DOI
- 10.1089/dia.2011.0201 · DOI
- 10.2337/dc12-2021 · DOI
- 10.1111/j.1399-5448.2011.00793.x · DOI
- 10.1177/193229681300700211 · DOI
Cited by
Heinemann, L., Freckmann, G., Ehrmann, D., Faber-Heinemann, G., Guerra, S., Waldenmaier, D., & Hermanns, N. (2018) · The Lancet · doi:10.1016/s0140-6736(18)30297-6 · Cited by 522
Chiang, J. L., Maahs, D. M., Garvey, K. C., Hood, K. K., Laffel, L. M., Weinzimer, S. A., Wolfsdorf, J. I., & Schatz, D. (2018) · Diabetes Care · doi:10.2337/dci18-0023 · Cited by 509
Self-citationContinuous glucose monitoring: a review of recent studies demonstrating improved glycemic outcomes
Rodbard, D. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0035 · Cited by 458
Measures of Accuracy for Continuous Glucose Monitoring and Blood Glucose Monitoring Devices
Freckmann, G., Pleus, S., Grady, M., Setford, S., & Levy, B. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818812062 · Cited by 202
ISPAD Clinical Practice Consensus Guidelines 2018: diabetes technologies
Sherr, J. L., Tauschmann, M., Battelino, T., de Bock, M., Forlenza, G., Roman, R., Hood, K. K., & Maahs, D. M. (2018) · Pediatric Diabetes · doi:10.1111/pedi.12731 · Cited by 197
Metadata
- Authors
- Laffel, L.
- Affiliations
- Joslin Diabetes Center
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2016-02
- Volume
- 18
- Issue
- 2_suppl
- ISSN
- 1520-9156
- 1557-8593
- Cited by
- 70
- References
- 15
- Language
- en
- Licence
- creativecommons.org/licenses/by/4.0/ (vor)
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.