CoChat

Reduced platelet activation and thrombosis in extracorporeal circuits coated with nitric oxide release polymers

Annich, Gail M., Meinhardt, Jürgen P., Mowery, Kelly A., Ashton, Brian A., Merz, Scott I., Hirschl, Ronald B., Meyerhoff, Mark E., Bartlett, Robert H. · 2000 · Critical Care Medicine
Cited by 152

The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.

Abstract

OBJECTIVE: To determine whether the use of nitric oxide (NO)-releasing polymers coated onto the inner surface of extracorporeal circuits can reduce platelet consumption and activation in the absence of systemic heparinization using a rabbit model of venovenous extracorporeal circulation. DESIGN: Prospective, controlled trial. SETTING: Research laboratory at an academic medical institution. SUBJECTS: New Zealand White Rabbits. INTERVENTIONS: Anesthetized, tracheotomized, and ventilated New Zealand White rabbits were injected with freshly prepared, 111In(oxine)3 labeled single donor platelets through the external jugular vein. After baseline measurements, these animals were placed on venovenous extracorporeal circulation through a 1-m control circuit or NO test circuit for 4 hrs at a blood flow rate of 109-118 mL/min via roller pump. Four groups were studied: systemically heparinized control circuits, systemically heparinized NO test circuits, nonheparinized control circuits, and nonheparinized NO test circuits. Platelet counts, fibrinogen levels, and plasma free indium levels were measured hourly. Circuits were rinsed and retained for gamma counting after the 4-hr run or when the circuit clotted. Four animals, one from each group, did not receive radiolabeled platelets so that the circuits could be preserved for scanning electron microscopic examination after the 4-hr study. MEASUREMENTS AND MAIN RESULTS: Platelet consumption was significantly reduced in both the heparinized and nonheparinized NO test groups when compared with the controls (p < .0001 and p < .0004, respectively). Platelet adhesion to the extracorporeal circuits was significantly reduced in the nonheparinized test circuits when compared with the controls (p < .05). Scanning electron microscopic examination of the circuits revealed that in the absence of heparin and in the presence of a NO-releasing surface, platelets retained their spherical nonactivated shape. CONCLUSIONS: The incorporation of NO into the surface of extracorporeal circuits reduces platelet consumption and eliminates the need for systemic heparinization in a rabbit model of extracorporeal circulation.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/00003246-200004000-00001

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

  • Nitric Oxide-Releasing Polymers Containing the [N(O)NO]-Group

    Smith, D. J., Chakravarthy, D., Pulfer, S., Simmons, M. L., Hrabie, J. A., Citro, M. L., Saavedra, J. E., Davies, K. M., Hutsell, T. C., Mooradian, D. L., et al. (1996) · Journal of Medicinal Chemistry · doi:10.1021/jm950652b · Cited by 208 · cited 1×

    “(15) have shown that specific diazeniumdiolate NO donors can be incorporated into polymers to dramatically improve their biocompatibility.”
    Background

Show all 1 references

Cited by

Metadata

Authors
Annich, G. M., Meinhardt, J. P., Mowery, K. A., Ashton, B. A., Merz, S. I., Hirschl, R. B., Meyerhoff, M. E., & Bartlett, R. H.
Venue
Critical Care Medicine
Published
2000-04
Volume
28
Issue
4
Pages
915-920
ISSN
0090-3493 (ISSN-L)
OpenAlex
W2094682388
Cited by
152
References
1
FWCI
8.08
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.