Wearable Accelerometer and Gyroscope Sensors for Estimating the Severity of Essential Tremor
Open accessThe DOI matches its Crossref record; 73 of the 73 references it cites were found in a registry or an index; 12 of the 21 sources its claims cite were checked for support.
Abstract
BACKGROUND: Several validated clinical scales measure the severity of essential tremor (ET). Their assessments are subjective and can depend on familiarity and training with scoring systems. METHOD: We propose a multi-modal sensing using a wearable inertial measurement unit for estimating scores on the Fahn-Tolosa-Marin tremor rating scale (FTM) and determine the classification accuracy within the tremor type. 17 ET participants and 18 healthy controls were recruited for the study. Two movement disorder neurologists who were blinded to prior clinical information viewed video recordings and scored the FTM. Participants drew a guided Archimedes spiral while wearing an inertial measurement unit placed at the mid-point between the lateral epicondyle of the humerus and the anatomical snuff box. Acceleration and gyroscope recordings were analyzed. The ratio of the power spectral density between frequency bands 0.5-4 Hz and 4-12 Hz, and the sum of power spectrum density over the entire spectrum of 2-74 Hz, for both accelerometer and gyroscope data, were computed. FTM was estimated using regression model and classification using SVM was validated using the leave-one-out method. RESULTS: Regression analysis showed a moderate to good correlation when individual features were used, while correlation was high ([Formula: see text] = 0.818) when suitable features of the gyro and accelerometer were combined. The accuracy for two-class classification of the combined features using SVM was 91.42% while for four-class it was 68.57%. CONCLUSION: Potential applications of this novel wearable sensing method using a wearable Inertial Measurement Unit (IMU) include monitoring of ET and clinical trials of new treatments for the disorder.
Overview
CoChat Summary
TL;DR
This study combined wearable accelerometer and gyroscope data from a forearm IMU during spiral drawing to estimate Fahn-Tolosa-Marin (FTM) tremor severity in essential tremor (ET). A regression model estimated FTM with \(r^2 = 0.818\), and SVM classification achieved 91.42% accuracy for ET vs controls but only 68.57% across four classes.
Key findings
- Multiple regression using accelerometer ratio (AR), gyroscope ratio (GR), and summed gyroscope PSD (GS) estimated clinical FTM score with \(r^2 = 0.818\) and RMSE = 5.25.
- SVM classification for two classes (ET vs age-matched controls) reached 91.42% accuracy with several feature combinations, including (AS, GS), (AS, GS, GR), (AR, GR, Tt), and (AR, GR, AS, GS).
- Four-class classification (controls, ET-0, ET+1, ET+2) achieved best mean accuracy of 68.57% using AS, GS, and Tt.
- Combining accelerometer and gyroscope improved classification over accelerometer alone: two-class accuracy increased from 85% to 91.42%, and four-class accuracy increased from 57.14% to 68.57%.
- Inter-rater FTM scoring by two blinded neurologists showed very strong correlation (\(r^2 = 0.95\)), and all computerized sensor features had significantly higher mean values in ET than controls.
Methods
- Seventeen participants with ET (eight male; mean age 67.28 ± 13.39 years; mean tremor duration 22.94 ± 19.65 years) and 18 healthy age-matched controls (eight male; mean age 62.8 ± 11.6 years) were recruited. ET participants were classified as ET-0 (6), ET+1 (5), or ET+2 (6) based on Consensus Axis 1 plus features.
- Participants drew a guided Archimedes spiral on a digital tablet while wearing a Delsys Trigno IMU at the midpoint between the lateral epicondyle of the humerus and the anatomical snuff box; accelerometer and gyroscope data were recorded at 148.1 samples/sec. Features included PSD ratios AR and GR (4–12 Hz / 0.5–4 Hz), summed PSD AS and GS (2–74 Hz), and total task time Tt.
- Statistical analysis used Mann-Whitney U tests, linear and multiple least-squares regression, and SVM classifiers with both leave-one-out and 5-fold cross-validation, using linear and radial basis function kernels.
Limitations
- The paper states that further improvement is necessary for the four-class problem, where accuracy was 68.57%.
- Evident from the text: the sample was small (17 ET participants and 18 controls), with very small ET-plus subgroups (5 ET+1 and 6 ET+2), limiting robustness of the four-class classification.
- Evident from the text: assessment relied on a single drawing task and one forearm sensor location, so findings may not generalize to other tasks, sensor placements, or free-living movement.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures



Verification
Claims the paper makes
9 Valid · 0 Not valid · 6 Not checked · 19 Own results and interpretations
Every source the paper cites for this claim was checked and supports it.
In the paper
“Shapiro-Wilk test was conducted to check for the normality distribution of the accelerometer and gyrometer parameters [61] and it showed a non-Gaussian distribution.”
§ Wearable Accelerometer and Gyroscope Sensors for Estimating the Severity of Essential Tremor > Data Analysis > Statistical Analysis
What the sources say
- Supportedcites [61] matched to “Normality Tests for Statistical Analysis: A Guide for Non-Statisticians” (2012) source
Section 3, “Normality Tests,” identifies the Shapiro–Wilk test as a test for assessing whether data are normally distributed. The citation supports the choice of test; the non-Gaussian result for the accelerometer and gyrometer parameters is presented as the citing study’s own finding, not as a finding of this source. (LLM, conf=0.91)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
Every source the paper cites for this claim was checked and supports it.
In the paper
“Spiral drawing, often used by clinicians as a pen-and-paper assessment of tremor disorders [40], is well-suited to the study of electronic recording of ET [41].”
§ Wearable Accelerometer and Gyroscope Sensors for Estimating the Severity of Essential Tremor > Introduction
What the sources say
- Supportedcites [40] matched to “How to use pen and paper tasks to aid tremor diagnosis in the clinic” (2017) source
The Archimedes spiral section describes spiral drawing as a pen-and-paper way to capture tremor characteristics, and the Essential tremor section applies it to clinical assessment. In the citing sentence, this paper supports the pen-and-paper assessment clause [40]; the separate claim about electronic recording is attached to [41]. (LLM, conf=0.90)
- Supportedcites [41] matched to “Digitizing Tablet and Fahn–Tolosa–Marı´n Ratings of Archimedes Spirals have Comparable Minimum Detectable Change in Essential Tremor” (2017) source
The Abstract Background (p.1) states that 'Drawing Archimedes spirals is a popular and valid method of assessing action tremor in the upper limbs,' and the paper is itself a blinded comparison of Fahn–Tolosa–Marín spiral ratings with digitizing-tablet measures in essential tremor. The Introduction passage notes that 'tablets are capable of providing linear objective measures of tremor in writing and drawings,' with ±0.25 mm accuracy and 100 samples/s 'adequate for measuring the amplitude and frequency' of visible tremor, and the Abstract Discussion concludes the digitizing tablet 'is a robust method of quantifying tremor.' Together these treat spiral drawing as well suited to electronic/digitized recording in ET, which is what the citing sentence attributes; the pen-and-paper clinician usage is credited to a different citation ([40]) and is not part of this judgment. (LLM, conf=0.82)
boosted- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“The total PSD in ET is higher than for controls [68], suggesting that, when performing similar tasks, people with ET have increased overall motor activity.”
§ Wearable Accelerometer and Gyroscope Sensors for Estimating the Severity of Essential Tremor > Discussion
What the sources say
- Partly supportedcites [68] matched to “A Power Spectral Density-Based Method to Detect Tremor and Tremor Intermittency in Movement Disorders” (2019) sourceExperimental — accuracy not yet measured
The Results section reports that power was several times lower in a healthy control than in a single tremor patient during the resting and postural tasks. It does not identify that patient as having ET or establish that total PSD is higher across the ET group than across controls. (LLM, conf=0.80)
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
Every source the paper cites for this claim was checked and supports it.
In the paper
“The suitable location of the sensors has been reported in our previous study, selected experimentally based on the highest performance in detecting the tremor during spiral drawin [46].”
§ Wearable Accelerometer and Gyroscope Sensors for Estimating the Severity of Essential Tremor > Methodology > Equipment and Data Recording
What the sources say
- Supportedcites [46] matched to “Wearable sensors during drawing tasks to measure the severity of essential tremor” (2022) source
The Discussion section ('Placement of wearable sensors') says the authors tested three sites and found that the mid-forearm sensor performed best for estimating tremor severity during spiral drawing. The Conclusion likewise identifies that forearm position as the most suitable of the three tested. (LLM, conf=0.92)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“SVM Classifier showed a high accuracy of 91.42% for the 2-class ET and AMC in the following combination of the features (AS, GS), (AS, GS, GR), (AR, GR, Tt) and (AR, GR, AS, GS).”
§ Wearable Accelerometer and Gyroscope Sensors for Estimating the Severity of Essential Tremor > Results > Wearable Sensors Et Classification
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
12 more claims
Create a free account to see all 34 claims and every source verdict.
- Borderline: one of the tests for a substantial claim is unclear. 1
- It may state two separate claims as one. 10
- Probably not a substantial claim. 7
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ Wearable Accelerometer and Gyroscope Sensors for Estimating the Severity of Essential Tremor > Results > Wearable Sensors Et Classification
“the best SVM classification mean accuracy was 68.57% using combination of three features: AS, GS, Tt.”
§ Wearable Accelerometer and Gyroscope Sensors for Estimating the Severity of Essential Tremor > Conclusion
“the four class ET-0, ET+1, ET+2, Controls achieved a classification accuracy of 68.57% using the features of AG, GS, Tt.”
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ Wearable Accelerometer and Gyroscope Sensors for Estimating the Severity of Essential Tremor > Results > Wearable Sensors Et Classification
“SVM Classifier showed a high accuracy of 91.42% for the 2-class ET and AMC in the following combination of the features (AS, GS), (AS, GS, GR), (AR, GR, Tt) and (AR, GR, AS, GS).”
§ Wearable Accelerometer and Gyroscope Sensors for Estimating the Severity of Essential Tremor > Results > Wearable Sensors Et Classification
“85.71% in (AR, GR, AS, GS), (AR, GR, AS)”
Citations
Cites
Sensitivity and False Alarm Rate of a Fall Sensor in Long-Term Fall Detection in the Elderly
Kangas, M., Korpelainen, R., Vikman, I., Nyberg, L., & Jämsä, T. (2015) · Gerontology · doi:10.1159/000362720 · Cited by 18,909
Normality Tests for Statistical Analysis: A Guide for Non-Statisticians
Ghasemi, A., & Zahediasl, S. (2012) · International Journal of Endocrinology and Metabolism · doi:10.5812/ijem.3505 · Cited by 4,368 · cited 1×
“A. STATISTICAL ANALYSIS Shapiro-Wilk test was conducted to check for the normality distribution of the accelerometer and gyrometer parameters [61] and it showed a non-Gaussian distribution.”
Activity Recognition from User-Annotated Acceleration Data
Bao, L., & Intille, S. S. (2004) · Lecture Notes in Computer Science · doi:10.1007/978-3-540-24646-6_1 · Cited by 3,147
Activity recognition using cell phone accelerometers
Kwapisz, J. R., Weiss, G. M., & Moore, S. A. (2011) · ACM SIGKDD Explorations Newsletter · doi:10.1145/1964897.1964918 · Cited by 2,912
Bhatia, K. P., Bain, P., Bajaj, N., Elble, R. J., Hallett, M., Louis, E. D., Raethjen, J., Stamelou, M., Testa, C. M., Deuschl, G., et al. (2018) · Movement Disorders · doi:10.1002/mds.27121 · Cited by 1,575
Performance evaluation of classification algorithms by k-fold and leave-one-out cross validation
Wong, T.-T. (2015) · Pattern Recognition · doi:10.1016/j.patcog.2015.03.009 · Cited by 1,562 · cited 1×
“Studies have supported the Leave-One-Out Cross-Validation (LOOCV) for a more reliable performance for small datasets [62].”
MethodWearable Sensors for Human Activity Monitoring: A Review
Mukhopadhyay, S. C. (2015) · IEEE Sensors Journal · doi:10.1109/jsen.2014.2370945 · Cited by 1,401
Evaluation of a threshold-based tri-axial accelerometer fall detection algorithm
Bourke, A. K., O’Brien, J. V., & Lyons, G. M. (2007) · Gait & Posture · doi:10.1016/j.gaitpost.2006.09.012 · Cited by 821
Najafi, B., Aminian, K., Paraschiv-Ionescu, A., Loew, F., Bula, C. J., & Robert, P. (2003) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2003.812189 · Cited by 762
Mayagoitia, R. E., Nene, A. V., & Veltink, P. H. (2002) · Journal of Biomechanics · doi:10.1016/s0021-9290(01)00231-7 · Cited by 655
The Mobile Sensing Platform: An Embedded Activity Recognition System
Choudhury, T., Borriello, G., Consolvo, S., Haehnel, D., Harrison, B., Hemingway, B., Hightower, J., Klasnja, P. "., Koscher, K., LaMarca, A., et al. (2008) · IEEE Pervasive Computing · doi:10.1109/mprv.2008.39 · Cited by 598
Estimation of energy expenditure using CSA accelerometers at hip and wrist sites
SWARTZ, A. M., STRATH, S. J., BASSETT, D. R., O???BRIEN, W. L., KING, G. A., & AINSWORTH, B. E. (2000) · Medicine & Science in Sports & Exercise · doi:10.1097/00005768-200009001-00003 · Cited by 513
A practical gait analysis system using gyroscopes
Tong, K., & Granat, M. H. (1999) · Medical Engineering & Physics · doi:10.1016/s1350-4533(99)00030-2 · Cited by 460 · cited 1×
“Gyroscopes in combination with accelerometers have the advantage that both linear and angular motion is recorded [36], [37], reducing calibration biases.”
BackgroundA study of hereditary essential tremor
Bain, P. G., Findley, L. J., Thompson, P. D., Gresty, M. A., Rothwell, J. C., Harding, A. E., & Marsden, C. D. (1994) · Brain · doi:10.1093/brain/117.4.805 · Cited by 392
On-body localization of wearable devices: An investigation of position-aware activity recognition
Sztyler, T., & Stuckenschmidt, H. (2016) · 2016 IEEE International Conference on Pervasive Computing and Communications (PerCom) · doi:10.1109/percom.2016.7456521 · Cited by 369 · cited 1×
“Wearable accelerometers have been used to study activities of daily living [13], [14], [15], [16], inferring metabolic energy expenditure [17], [18], [19], measuring gait parameters [20] and predicting falls [21].”
MethodSensor Positioning for Activity Recognition Using Wearable Accelerometers
Atallah, L., Lo, B., King, R., & Yang, G.-Z. (2011) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2011.2160540 · Cited by 346
A novel method for using accelerometer data to predict energy expenditure
Crouter, S. E., Clowers, K. G., & Bassett, D. R. (2006) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00818.2005 · Cited by 345
Stacy, M. A., Elble, R. J., Ondo, W. G., Wu, S., Hulihan, J., & TRS Study Group (2007) · Movement Disorders · doi:10.1002/mds.21412 · Cited by 296 · cited 1×
“These tools must perform well on tests of intra-and inter-rater reliability [7], and be sensitive to clinically meaningful changes in tremor severity.”
MethodStanding balance evaluation using a triaxial accelerometer
Mayagoitia, R. E., Lötters, J. C., Veltink, P. H., & Hermens, H. (2002) · Gait & Posture · doi:10.1016/s0966-6362(01)00199-0 · Cited by 249
Shoe-integrated sensor system for wireless gait analysis and real-time feedback
Morris, S. J., & Paradiso, J. A. (2002) · Proceedings of the Second Joint 24th Annual Conference and the Annual Fall Meeting of the Biomedical Engineering Society] [Engineering in Medicine and Biology · doi:10.1109/iembs.2002.1053379 · Cited by 177
Position-aware activity recognition with wearable devices
Sztyler, T., Stuckenschmidt, H., & Petrich, W. H. (2017) · Pervasive and Mobile Computing · doi:10.1016/j.pmcj.2017.01.008 · Cited by 148
Kim, H. B., Lee, W. W., Kim, A., Lee, H. J., Park, H. Y., Jeon, H. S., Kim, S. K., Jeon, B., & Park, K. S. (2018) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2018.02.007 · Cited by 124
Surface EMG and acceleration signals in Parkinson’s disease: feature extraction and cluster analysis
Rissanen, S. M., Kankaanpaa, M. J., Meigal, A. Y., Tarvainen, M. P., Nuutinen, J., Tarkka, I. M., Airaksinen, O., & Karjalainen, P. A. (2008) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-008-0369-0 · Cited by 105
Essential tremor—a neurodegenerative disorder associated with cognitive defects?
Bermejo-Pareja, F. (2011) · Nature Reviews Neurology · doi:10.1038/nrneurol.2011.44 · Cited by 98 · cited 1×
“Essential tremor affects about 2% of the population, begin 5 to 10 times more prevalent compared with Parkinson’s disease [1], [2].”
BackgroundUsing Portable Transducers to Measure Tremor Severity
Elble, R. J., & McNames, J. (2016) · Tremor and Other Hyperkinetic Movements · doi:10.5334/tohm.320 · Cited by 95
Estimating fugl-meyer clinical scores in stroke survivors using wearable sensors
Del Din, S., Patel, S., Cobelli, C., & Bonato, P. (2011) · 2011 Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2011.6091444 · Cited by 90
Distributed segmentation and classification of human actions using a wearable motion sensor network
Yang, A. Y., Iyengar, S., Kuryloski, P., & Jafari, R. (2008) · 2008 IEEE Computer Society Conference on Computer Vision and Pattern Recognition Workshops · doi:10.1109/cvprw.2008.4563176 · Cited by 88
Norman, K. E., Edwards, R., & Beuter, A. C. (1999) · Journal of Neuroscience Methods · doi:10.1016/s0165-0270(99)00091-6 · Cited by 88
Primary writing tremor. A form of focal dystonia?
Elble, R. J., Moody, C., & Higgins, C. (1990) · Movement Disorders · doi:10.1002/mds.870050205 · Cited by 88
Transducer‐based evaluation of tremor
Haubenberger, D., Abbruzzese, G., Bain, P. G., Bajaj, N., Benito‐León, J., Bhatia, K. P., Deuschl, G., Forjaz, M. J., Hallett, M., Louis, E. D., et al. (2016) · Movement Disorders · doi:10.1002/mds.26671 · Cited by 87
Ondo, W., Hashem, V., LeWitt, P. A., Pahwa, R., Shih, L., Tarsy, D., Zesiewicz, T., & Elble, R. (2018) · Movement Disorders Clinical Practice · doi:10.1002/mdc3.12560 · Cited by 87
How to use pen and paper tasks to aid tremor diagnosis in the clinic
Alty, J., Cosgrove, J., Thorpe, D., & Kempster, P. (2017) · Practical Neurology · doi:10.1136/practneurol-2017-001719 · Cited by 78
Self-citationHeida, T., Wentink, E., & Marani, E. (2013) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-10-70 · Cited by 66 · cited 1×
“The principle has been used to examine various clinical aspects of ET [63].”
BackgroundGyroscope vs. accelerometer measurements of motion from wrist PPG during physical exercise
Casson, A. J., Vazquez Galvez, A., & Jarchi, D. (2016) · ICT Express · doi:10.1016/j.icte.2016.11.003 · Cited by 57
Lemoyne, R. C., Tomycz, N. D., Mastroianni, T., McCandless, C., Cozza, M., & Peduto, D. (2015) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2015.7319948 · Cited by 56
Essential tremor is a neurodegenerative disease
Bonuccelli, U. (2012) · Journal of Neural Transmission · doi:10.1007/s00702-012-0878-8 · Cited by 51
Development of a System for Measurement and Analysis of Tremor Using a Three-axis Accelerometer
Iizuka, T., Takeuchi, A., Shirataka, M., Ikeda, N., & Mamorita, N. (2009) · Methods of Information in Medicine · doi:10.3414/me9243 · Cited by 49
A Fusion Health Monitoring Using ECG and Accelerometer sensors for Elderly Persons at Home
Chung, W.-Y., Bhardwaj, S., Punvar, A., Lee, D.-S., & Myllylae, R. (2007) · 2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2007.4353164 · Cited by 47
Quantification of Finger-Tapping Angle Based on Wearable Sensors
Djuric-Jovicic, M. D., Jovicic, N. S., Roby-Brami, A., Popović, M., Kostić, V., & Djordjevic, A. R. (2017) · Sensors · doi:10.3390/s17020203 · Cited by 46
Continuous Monitoring of Essential Tremor Using a Portable System Based on Smartwatch
Zheng, X., Vieira Campos, A., Ordieres-Meré, J., Balseiro, J., Labrador Marcos, S., & Aladro, Y. (2017) · Frontiers in Neurology · doi:10.3389/fneur.2017.00096 · Cited by 44 · cited 4×
“ET is associated with heightened activity in the frequency range of 4-12 Hz; in healthy people, the 0-4 Hz band is dominant [49], [59].”
BackgroundMethodInfluentialBoroojerdi, B., Ghaffari, R., Mahadevan, N., Markowitz, M., Melton, K., Morey, B., Otoul, C., Patel, S., Phillips, J., Sen-Gupta, E., et al. (2019) · Parkinsonism & Related Disorders · doi:10.1016/j.parkreldis.2018.11.024 · Cited by 43
Using Portable Transducers to Measure Tremor Severity
Elble, R., & McNames, J. (2016) · Tremor and Other Hyperkinetic Movements · doi:10.7916/d8dr2vcc · Cited by 43
Real-time elderly activity monitoring system based on a tri-axial accelerometer
Kang, D. W., Choi, J. S., Lee, J. W., Chung, S. C., Park, S. J., & Tack, G. R. (2010) · Disability and Rehabilitation: Assistive Technology · doi:10.3109/17483101003718112 · Cited by 43
An Expert System for Quantification of Bradykinesia Based on Wearable Inertial Sensors
Bobic, V. N., Djuric-Jovicic, M. D., Dragašević, N., Popović, M. B., Kostić, V., & Kvaščev, G. (2019) · Sensors · doi:10.3390/s19112644 · Cited by 41
Louis, E. D. (2020) · Neuroepidemiology · doi:10.1159/000502862 · Cited by 41 · cited 1×
“Better diagnostic criteria have been created to delineate ET and to help to separate it from other tremor disorders with which it can be confused [6].”
BackgroundCombined accelerometer and EMG analysis to differentiate essential tremor from Parkinson's disease
Ghassemi, N. H., Marxreiter, F., Pasluosta, C. F., Kugler, P., Schlachetzki, J., Schramm, A., Eskofier, B. M., & Klucken, J. (2016) · 2016 38th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2016.7590791 · Cited by 39
Estimation of Activity Energy Expenditure: Accelerometer Approach
Choi, J. H., Lee, J., Hwang, H. T., Kim, J. P., Park, J. C., & Shin, K. (2005) · 2005 IEEE Engineering in Medicine and Biology 27th Annual Conference · doi:10.1109/iembs.2005.1615295 · Cited by 36
Šprdlík, O., Hurák, Z., Hoskovcová, M., & Růžička, E. (2009) · IEEE International Conference on Information Technology and Applications in Biomedicine · doi:10.1109/itab.2009.5394320 · Cited by 35
Wearable sensors during drawing tasks to measure the severity of essential tremor
Ali, S. M., Arjunan, S. P., Peters, J., Perju-Dumbrava, L., Ding, C., Eller, M., Raghav, S., Kempster, P., Motin, M. A., Radcliffe, P. J., et al. (2022) · Scientific Reports · doi:10.1038/s41598-022-08922-6 · Cited by 35
Self-citationAssessment of Head Tremor with Accelerometers Versus Gyroscopic Transducers
Elble, R. J., Hellriegel, H., Raethjen, J., & Deuschl, G. (2017) · Movement Disorders Clinical Practice · doi:10.1002/mdc3.12379 · Cited by 35 · cited 1×
“Elble et al. [28] investigated the use of accelerometers and gyroscopes for measuring head tremors and found gyroscopes to be more effective.”
BackgroundHuang, E. J., & Onnela, J.-P. “. (2020) · Sensors · doi:10.3390/s20133706 · Cited by 28
Luft, F., Sharifi, S., Mugge, W., Schouten, A., Bour, L., van Rootselaar, A.-F., Veltink, P., & Heida, T. (2019) · Sensors · doi:10.3390/s19194301 · Cited by 27 · cited 1×
“The total PSD in ET is higher than for controls [68], suggesting that, when performing similar tasks, people with ET have increased overall motor activity.”
BackgroundHow to Tackle Tremor – Systematic Review of the Literature and Diagnostic Work-Up
Buijink, A. W. G., Contarino, M. F., Koelman, J. H. T. M., Speelman, J. D., & van Rootselaar, A. F. (2012) · Frontiers in Neurology · doi:10.3389/fneur.2012.00146 · Cited by 24
Šprdlík, O., Hurák, Z., Hoskovcová, M., Ulmanová, O., & Růžička, E. (2010) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2010.09.004 · Cited by 21
Balance in essential tremor during tandem gait: Is the first mis-step an important finding?
Cinar, N., Sahin, S., Okluoglu Onay, T., & Karsidag, S. (2013) · Journal of Clinical Neuroscience · doi:10.1016/j.jocn.2013.01.013 · Cited by 20 · cited 1×
“The Fahn-Tolosa-Marin tremor rating scale (FTM) has a good track record with clinical observations of ET [8], [9].”
BackgroundLimited correlations between clinician-based and patient-based measures of essential tremor severity
van der Stouwe, A. M. M., Broersma, M., Buijink, A. W. G., van Rootselaar, A. F., & Maurits, N. M. (2015) · Parkinsonism & Related Disorders · doi:10.1016/j.parkreldis.2015.03.004 · Cited by 17
Nanda, S. K., Lin, W., Lee, M., & Chen, R. (2015) · IEEE International Conference on Networking, Sensing and Control (ICNSC) · doi:10.1109/icnsc.2015.7116070 · Cited by 17
Elble, R. J., & Ellenbogen, A. (2017) · Tremor and Other Hyperkinetic Movements · doi:10.5334/tohm.344 · Cited by 16
Molparia, B., Schrader, B. N., Cohen, E., Wagner, J. L., Gupta, S. R., Gould, S., Hwynn, N., Spencer, E. G., & Torkamani, A. (2018) · PeerJ · doi:10.7717/peerj.5308 · Cited by 15
Elble, R. J., & Ellenbogen, A. L. (2017) · PubMed · doi:10.7916/d825282b · Cited by 15
Machowska-Majchrzak, A., Pierzchała, K., Pietraszek, S., & Łabuz-Roszak, B. (2011) · Neurologia i Neurochirurgia Polska · doi:10.1016/s0028-3843(14)60022-0 · Cited by 13
Intermittent bilateral coherence in physiological and essential hand tremor
Chakraborty, S., Kopecká, J., Šprdlík, O., Hoskovcová, M., Ulmanová, O., Růžička, E., & Zapotocky, M. (2017) · Clinical Neurophysiology · doi:10.1016/j.clinph.2016.12.027 · Cited by 10
Development of a human movement monitoring system based on wearable devices
Zheng, X., & Ordieres-Meré, J. B. (2014) · Hispana · Cited by 10
Knowledge about Essential Tremor: A Study of Essential Tremor Families
Cristal, A. D., Chen, K. P., Hernandez, N. C., Factor-Litvak, P. R., Clark, L. N., Ottman, R., & Louis, E. D. (2018) · Frontiers in Neurology · doi:10.3389/fneur.2018.00027 · Cited by 8
Computerized Screening of Essential Tremor and Level of Severity Using Consumer Tablet
Motin, M. A., Peters, J., Perju‐Dumbrava, L., Ding, C., Eller, M., Raghav, S., Ali, S. M., Kempster, P. A., Radcliffe, P., & Kumar, D. (2021) · IEEE Access · doi:10.1109/access.2021.3052186 · Cited by 7
Self-citationPersonal digital health in Parkinson's disease: Case histories and commentary
Boving, A. T., Shuster, C. L., Walls, T. A., & Brothers, T. N. (2021) · Digital Health · doi:10.1177/20552076211061925 · Cited by 5
Clinical Tremor Severity Estimation Using an Instrumented Eating Utensil
Miocinovic, S., Shoeb, A. H., Wang, S., Byrd, E. A., Swann, N. C., Pathak, A., & Ostrem, J. L. (2017) · Journal of Parkinson s Disease · doi:10.3233/jpd-160929 · Cited by 5
Ai, L., Wang, J., Huang, L., & Wang, X. (2007) · Lecture notes in computer science · doi:10.1007/978-3-540-72393-6_142 · Cited by 5
(1967) · doi:10.1016/c2013-0-13314-4 · Cited by 3
Pourfar, M. H., & Louis, E. D. (2006) · Current Therapy in Neurologic Disease · doi:10.1016/b978-0-323-03432-6.50066-0 · Cited by 2
Kim, J. W., Kwon, Y., Eom, G., Kwon, D., Koh, S., & Park, B. K. (2010) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-3-642-14515-5_326 · Cited by 2
Elble, R. J., & Ellenbogen, A. (2017) · Tremor and Other Hyperkinetic Movements · doi:10.7916/d89s20h7
Research works printed but not linked: 2
- Donald, 2015, Biomedical Engineering Handbook—Biomedical Engineering Fundamentals
- 10.3389/fneur.2012.00146 · DOI · Paper page
Cited by
Bancel, T., Béranger, B., Daniel, M., Didier, M., Santin, M. D., Rachmilevitch, I., Shapira, Y., Tanter, M., Bardinet, É., Fernández-Vidal, S., et al. (2024) · Brain stimulation · doi:10.1016/j.brs.2024.05.003 · Cited by 68
A deep learning approach to remotely assessing essential tremor with handwritten images
Peng, Y., Han, S., Wu, D., Ma, C., Zeng, Z., He, P., Yuan, T., Shi, Y., Li, L., Yang, W., et al. (2025) · Scientific Reports · doi:10.1038/s41598-025-94729-0 · Cited by 10
Keba, M., Bachmann, M., Lass, J., & Rätsep, T. (2025) · Journal of Clinical Medicine · doi:10.3390/jcm14062073 · Cited by 11
A Skin-Tone Insensitive, Fully Integrated Smartwatch for Continuous Blood Pressure Measurement
Huang, C.-Y., Sung, Y., Ni, S.-H., & Wang, T.-W. (2024) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2024.3502829 · Cited by 10
Towards improved human arm movement analysis: Advanced feature engineering and model optimization
Benachour, Y., Rehman, M. U., & Flitti, F. (2025) · Engineering Applications of Artificial Intelligence · doi:10.1016/j.engappai.2025.111194 · Cited by 8
Metadata
- Authors
- Ali, S. M., Arjunan, S. P., Peter, J., Perju-Dumbrava, L., Ding, C., Eller, M., Raghav, S., Kempster, P., Motin, M. A., Radcliffe, P. J., & Kumar, D. K.
- Affiliations
- RMIT University (AU)
- SRM Institute of Science and Technology (IN)
- Monash Medical Centre (AU)
- Neurosciences Department, Monash Health, Clayton, VIC, Australia
- Monash Health (AU)
- Rajshahi University of Engineering and Technology (BD)
- Corresponding author
- Kumar, Dinesh Kant
- ORCID
- Venue
- IEEE Journal of Translational Engineering in Health and Medicine
- Published
- 2024
- Volume
- 12
- Pages
- 194-203
- ISSN
- 2168-2372 (ISSN-L)
- PMC
- PMC10776092
- OpenAlex
- W4388125269
- Semantic Scholar
- b6dd4de4b81897f2495e35d298326f2b391fe66d
- Cited by
- 22
- References
- 71
- Citations by year
- 2026: 7 · 2025: 11 · 2024: 4
- FWCI
- 1.64
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- IEEE Journal of Translational Engineering in Health and Medicine (publishedVersion, cc-by-nc-nd)
- PubMed Central (publishedVersion, cc-by-nc-nd)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Figshare (submittedVersion, cc-by-nc-nd)
- Funders
- Scheme for Promotion of Academic and Research Collaboration (S134)
- Stiftelsen Promobilia (Wearable sensors)
- Topics
- Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Context-Aware Activity Recognition Systems
- Health Sciences › Medicine › Clinical Neurology › Deep Brain Stimulation and Movement Disorders
- Health Sciences › Health Professions › Health Information Management › Artificial Intelligence in Healthcare
- Keywords
- essential tremor, wearable inertial sensors, accelerometer, gyroscope, Fahn-Tolosa-Marin Tremor Rating Scale, tremor classification, power spectral density features, multimodal sensing, tremor monitoring
- MeSH terms
- Acceleration, Wearable Electronic Devices, Humans, Tremor, Essential Tremor, Accelerometry
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol and qwen3.8-omni-flash).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.