Long Micro-Channel Electrode Arrays: A Novel Type of Regenerative Peripheral Nerve Interface
Open accessThe DOI matches its Crossref record; 15 of the 15 references it cites were found in a registry or an index.
Abstract
We have demonstrated that micro-channel electrode arrays with 100 microm x 100 microm cross-section channels support axon regeneration well, and that micro-channels of similar calibre and up to 5 mm long can support axon regeneration and vascularisation. They may be microfabricated using silicon, silicone, or polyimide and thin metal films to form 3-D bundles of long micro-channels. Arrays of "mini-nerves," i.e., miniature nerve fascicles with their own blood vessels, successfully grew through implants 0.5-5 mm long. Furthermore, guiding the regenerating nerve fibres into the small insulating channels allows for a significant increase of the extracellular (recordable) amplitude of action potentials, which promises considerable improvement for in vivo electrophysiology.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“While axon regeneration can occur in microchannel devices with channels as small as small as 55 μm in diameter (similar to sieve electrode dimensions), optimal axon regeneration is seen when coupled with growth of one or more blood vessel, requiring larger microchannels of 100 μm diameter (Lacour et al., 2009).”
From Foreign Body Reaction to Implanted Biomaterials and Its Impact in Nerve Neuroprosthetics (2021) · cites this paper as Lacour et al., 2009
In the citing paper
“The MCEA device allowed for the regeneration of miniature nerve fascicles up to 5 mm long with their own vascular structures to grow through the implanted device.”
From Regenerative Electrode Interfaces for Neural Prostheses. (2016) · cites this paper as [30]
In the citing paper
“These fascicles within the MCEA maintained a similar structure and function to nerves regenerated through an ordinary silicone tube.”
From Regenerative Electrode Interfaces for Neural Prostheses. (2016) · cites this paper as [30]
Citations
Cites
Navarro, X., Krueger, T. B., Lago, N., Micera, S., Stieglitz, T., & Dario, P. (2005) · Journal of the Peripheral Nervous System · doi:10.1111/j.1085-9489.2005.10303.x · Cited by 845
Self-citationNerve regeneration through artificial tubular implants
Fields, R. D., Le Beau, J. M., Longo, F. M., & Ellisman, M. H. (1989) · Progress in Neurobiology · doi:10.1016/0301-0082(89)90036-1 · Cited by 308 · cited 1×
“This is a typical observation of a nerve regenerating through a tube [ 12 ].”
BackgroundA multielectrode array for intrafascicular recording and stimulation in sciatic nerve of cats
Branner, A., & Normann, R. A. (2000) · Brain Research Bulletin · doi:10.1016/s0361-9230(99)00231-2 · Cited by 234 · cited 1×
“Current interfaces typically have tens (sieve electrodes) [2]–[4] to a hundred or so (Utah pin arrays) electrodes [5].”
BackgroundCuff electrodes for chronic stimulation and recording of peripheral nerve activity
Loeb, G. E., & Peck, R. A. (1996) · Journal of Neuroscience Methods · doi:10.1016/0165-0270(95)00123-9 · Cited by 215 · cited 1×
“We have overcome this by guiding the axons into small micro-channels made of insulating material: the micro-channel restricts the extracellular space, augmenting the resulting extracellular impedance thus increasing the recordable amplitude of the action potential [6], [7].”
BackgroundLago, N., Ceballos, D., J Rodrı́guez, F., Stieglitz, T., & Navarro, X. (2005) · Biomaterials · doi:10.1016/j.biomaterials.2004.06.025 · Cited by 178
Self-citationStieglitz, T., Beutel, H., & Meyer, J.-U. (1997) · Sensors and Actuators A: Physical · doi:10.1016/s0924-4247(97)01494-5 · Cited by 125
Numbers of regenerating axons in parent and tributary peripheral nerves in the rat
Jenq, C., & Coggeshall, R. E. (1985) · Brain Research · doi:10.1016/0006-8993(85)91381-2 · Cited by 123 · cited 2×
“uninjured sciatic nerves contained an average of 7,395 26 myelinated fibres ; nerves regenerated through 3 mm gap silicone tube had an average of 7494 MF [13].”
BackgroundAssessment of the biocompatibility of photosensitive polyimide for implantable medical device use
Sun, Y., Lacour, S. P., Brooks, R. A., Rushton, N., Fawcett, J., & Cameron, R. E. (2009) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.32125 · Cited by 122
Self-citationMicrochannels as Axonal Amplifiers
FitzGerald, J. J., Lacour, S. P., McMahon, S. B., & Fawcett, J. W. (2008) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2007.909533 · Cited by 76 · cited 2×
“For example, single-unit extracellular voltage swings of up to 500 can be recorded by the midpoint electrode in a 90- -diameter, 1-cm-long channel [7].”
BackgroundSelf-citationPolyimide micro-channel arrays for peripheral nerve regenerative implants
Lacour, S. P., Atta, R., FitzGerald, J. J., Blamire, M., Tarte, E., & Fawcett, J. (2008) · Sensors and Actuators A: Physical · doi:10.1016/j.sna.2008.05.031 · Cited by 56
Self-citationRamachandran, A., Schuettler, M., Lago, N., Doerge, T., Koch, K. P., Navarro, X., Hoffmann, K.-P., & Stieglitz, T. (2006) · Journal of Neural Engineering · doi:10.1088/1741-2560/3/2/005 · Cited by 46
Self-citationLoeb, G. E., Marks, W. B., & Beatty, P. G. (1977) · Medical & Biological Engineering & Computing · doi:10.1007/bf02442964 · Cited by 46
Microchannel Electrodes for Recording and Stimulation:In Vitro Evaluation
FitzGerald, J. J., Lacour, S. P., McMahon, S. B., & Fawcett, J. W. (2009) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2009.2013960 · Cited by 45
Self-citationBenmerah, S., Lacour, S. P., & Tarte, E. J. (2009) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2009.5333723 · Cited by 14
Self-citationRecording with microchannel electrodes in a noisy environment
FitzGerald, J. J., Lacour, S. P., & Fawcett, J. W. (2008) · 2008 30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2008.4649084 · Cited by 10
Self-citation
Cited by
Hassler, C., Boretius, T., & Stieglitz, T. (2011) · Journal of Polymer Science Part B: Polymer Physics · doi:10.1002/polb.22169 · Cited by 501
Foreign Body Reaction to Implanted Biomaterials and Its Impact in Nerve Neuroprosthetics
Carnicer-Lombarte, A., Chen, S.-T., Malliaras, G. G., & Barone, D. G. (2021) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2021.622524 · Cited by 454
“This is, however, followed by a sharp decline in axon numbers at the 12 month time point (Lago et al., 2005).”
BackgroundFlexible and stretchable micro-electrodes for in vitro and in vivo neural interfaces
Lacour, S. P., Benmerah, S., Tarte, E., FitzGerald, J., Serra, J., McMahon, S., Fawcett, J., Graudejus, O., Yu, Z., & Morrison, B. (2010) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-010-0644-8 · Cited by 263
“Robust vascularisation is observed through most micro-channels [ 21 ].”
BackgroundMethodInfluentialSelf-citationPast, Present, and Future of Nerve Conduits in the Treatment of Peripheral Nerve Injury
Muheremu, A., & Ao, Q. (2015) · BioMed Research International · doi:10.1155/2015/237507 · Cited by 199
“Bioabsorbable and compound conduits (consisting of neurotrophic factors [63], nerve supporters [64, 65], SCs [66], and neural stem cells [67–69]) have promoted chemotactic regeneration of peripheral nerves and enhanced the effectiveness of nerve repair.”
BackgroundSafety of long-term electrical peripheral nerve stimulation: review of the state of the art
Günter, C., Delbeke, J., & Ortiz-Catalan, M. (2019) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-018-0474-8 · Cited by 198
Metadata
- Authors
- Lacour, S. P., Fitzgerald, J. J., Lago, N., Tarte, E., McMahon, S., & Fawcett, J.
- Affiliations
- University of Cambridge (GB)
- King's College London (GB)
- University of Birmingham (GB)
- Venue
- IEEE Transactions on Neural Systems and Rehabilitation Engineering
- Published
- 2009-10
- Volume
- 17
- Issue
- 5
- Pages
- 454-460
- ISSN
- 1534-4320 (ISSN-L)
- 1558-0210
- OpenAlex
- W2139282950
- Cited by
- 75
- References
- 15
- Citations by year
- 2026: 2 · 2025: 1 · 2024: 7 · 2023: 3 · 2022: 1 · 2021: 3 · 2020: 2 · 2019: 3 · 2018: 7 · 2017: 6 · 2016: 7 · 2015: 7 · 2014: 4 · 2013: 5 · 2012: 4
- FWCI
- 2.35
- Type
- journal article
- Language
- en
- Open access
- green
- infoscience.epfl.ch
- Locations
- IEEE Transactions on Neural Systems and Rehabilitation Engineering (publishedVersion)
- PubMed (publishedVersion)
- Infoscience (Ecole Polytechnique Fédérale de Lausanne) (submittedVersion)
- Infoscience (Ecole Polytechnique Fédérale de Lausanne) (submittedVersion, cc-by-nc-nd)
- Research Portal (King's College London) (submittedVersion)
- University of Birmingham Research Portal (University of Birmingham) (submittedVersion)
- Funders
- Engineering and Physical Sciences Research Council (EP/C52330X/1, EP/H00727X/1)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Nerve injury and regeneration
- Health Sciences › Medicine › Surgery › Tissue Engineering and Regenerative Medicine
- Keywords
- peripheral nerve interfaces, axon regeneration, nerve guidance channels, silicon microchannels, action potential amplitude, in vivo electrophysiology, peripheral nerve stimulation
- MeSH terms
- Action Potentials, Animals, Electrodes, Implanted, Equipment Design, Male, Microelectrodes, Nerve Regeneration, Sensitivity and Specificity, Reproducibility of Results, Equipment Failure Analysis, Sciatic Neuropathy, Guided Tissue Regeneration, Rats
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.