Comparison of Mono-, Bi-, and Tripolar Configurations for Stimulation and Recording With an Interfascicular Interface
Abstract
Previous studies have indicated that electrodes placed between fascicles can provide nerve recruitment with high topological selectivity if the areas of interest in the nerve are separated with passive elements. In this study, we investigated if this separation of fascicles also can provide topologically selective nerve recordings and compared the performance of mono-, bi-, and tripolar configurations for stimulation and recording with an intra-neural interface. The interface was implanted in the sciatic nerve of 10 rabbits and achieved a median selectivity of Ŝ=0.98-0.99 for all stimulation configurations, while recording selectivity configurations was in the range of Ŝ=0.70-0.80 with the monopolar configuration providing the lowest and the average reference configuration the highest recording selectivity. Interfascicular electrodes could provide an interesting addition to the bulk of peripheral nerve interfaces available for neural prosthetic devices. The separation of the nerve into chambers by the passive elements of the electrode could ensure a higher selectivity than comparable cuff electrodes and the intra-neural location could provide an option of targeting mainly central fascicles. Further studies are, however, still required to develop biocompatible electrodes and test their stability and safety in chronic experiments.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Two interfascicular electrode designs tested in rabbit experiments confirmed the capability of interfascicular electrodes in interfacing with subpopulations of nerve fibers.
“These studies also confirmed the capability of interfascicular electrodes in interfacing with subpopulations of nerve fibers.”
From Interfaces with the peripheral nervous system for the control of a neuroprosthetic limb: a review (2020) · cites this paper as [83]
Cites
Navarro, X., Krueger, T. B., Lago, N., Micera, S., Stieglitz, T., & Dario, P. (2005) · Journal of the Peripheral Nervous System · doi:10.1111/j.1085-9489.2005.10303.x · Cited by 845 · see citing works
Simpson, J. A. (1979) · Journal of Neurology Neurosurgery & Psychiatry · doi:10.1136/jnnp.42.6.580 · Cited by 676 · see citing works
Electrical stimulation of sacral spinal nerves for treatment of faecal incontinence
Matzel, K. E., Stadelmaie, U., Gall, F. P., et al. (1995) · The Lancet · doi:10.1016/s0140-6736(95)91799-3 · Cited by 591 · see citing works
Selective Electrical Interfaces with the Nervous System
Rutten, W. L. C. (2002) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.4.020702.153427 · Cited by 276 · see citing works
Cutaneous whole nerve recordings used for correction of footdrop in hemiplegic man
Haugland, M. K., & Sinkjaer, T. (1995) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.481970 · Cited by 168 · see citing works
Sensory nerve recording for closed‐loop control to restore motor functions
Popovic, D. B., Stein, R. B., Jovanovic, K. L., Dai, R., Kostov, A., & Armstrong, W. W. (1993) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.247801 · Cited by 159 · see citing works
Experimental correction of footdrop by electrical stimulation of the peroneal nerve
Waters, R. L., McNeal, D. R., & Perry, J. (1975) · Journal of Bone and Joint Surgery · doi:10.2106/00004623-197557080-00002 · Cited by 152 · see citing works
Citi, L., Carpaneto, J., Yoshida, K., Hoffmann, K., Koch, K., Dario, P., & Micera, S. (2008) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2008.04.025 · Cited by 117 · see citing works
A slowly penetrating interfascicular nerve electrode for selective activation of peripheral nerves
Tyler, D. J., & Durand, D. M. (1997) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.559349 · Cited by 108 · cited 2× · see citing works
“Existing literature on peripheral nerve electrodes focus mainly on extra-neural and intrafascicular electrodes, but studies have also indicated that electrodes placed between the fascicles can provide topologically selective recruitment of subareas of the nerve, provided that passive elements are used to compartmental…”
BackgroundYoo, P. B., & Durand, D. M. (2005) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2005.851482 · Cited by 106 · see citing works
Sacral neurostimulation for urinary retention: 10‐year experience from one UK centre
Datta, S. N., Chaliha, C., Singh, A., et al. (2007) · British Journal of Urology · doi:10.1111/j.1464-410x.2007.07282.x · Cited by 104 · see citing works
Micera, S., Rossini, P. M., Rigosa, J., Citi, L., Carpaneto, J., Raspopovic, S., Tombini, M., Cipriani, C., Assenza, G., Carrozza, M. C., et al. (2011) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-8-53 · Cited by 100 · see citing works
A technique for anodally blocking large nerve fibres through chronically implanted electrodes.
Brindley, G. S., & Craggs, M. D. (1980) · Journal of Neurology Neurosurgery & Psychiatry · doi:10.1136/jnnp.43.12.1083 · Cited by 97 · see citing works
A modeling study of nerve fascicle stimulation
Veltink, P. H., van Veen, B. K., Struijk, J. J., Holsheimer, J., & Boom, H. B. K. (1989) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.32100 · Cited by 95 · see citing works
Self-citationStruijk, J. J. (1997) · Biophysical Journal · doi:10.1016/s0006-3495(97)78890-8 · Cited by 93 · see citing works
Self-citationProgrammed and Magnet-Induced Vagus Nerve Stimulation for Refractory Epilepsy
Boon, P., Vonck, K. E. J., Walleghem, P. V., D’Havé, M., Goossens, L., Vandekerckhove, T., Caemaert, J., & Reuck, J. D. (2001) · Journal of Clinical Neurophysiology · doi:10.1097/00004691-200109000-00003 · Cited by 87 · see citing works
Multiple-electrode nerve cuffs for low-velocity and velocity-selective neural recording
Taylor, J. T., Donaldson, N., & Winter, J. (2004) · Medical & Biological Engineering & Computing · doi:10.1007/bf02347545 · Cited by 79 · see citing works
A flexible method for fabrication of nerve cuff electrodes
Haugland, M. K. (2002) · Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.1996.656992 · Cited by 79 · see citing works
Neuroprostheses to treat neurogenic bladder dysfunction: current status and future perspectives
Rijkhoff, N. J. M. (2004) · Child's Nervous System · doi:10.1007/s00381-003-0859-1 · Cited by 66 · see citing works
Recruitment characteristics of nerve fascicles stimulated by a multigroove electrode
Koole, P., Holsheimer, J., Struijk, J. J., & Verloop, A. J. (1997) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.559348 · Cited by 56 · see citing works
Self-citationSeparation of action potentials in multiunit intrafascicular recordings
Goodall, E. V., & Horch, K. W. (1992) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.125014 · Cited by 53 · see citing works
Selective Stimulation of the Canine Hypoglossal Nerve Using a Multi-contact Cuff Electrode
Yoo, P. B., Şahin, M., & Durand, D. M. (2004) · Annals of Biomedical Engineering · doi:10.1023/b:abme.0000019170.74375.fb · Cited by 48 · see citing works
Technology and Long-term Application of an Epineural Electrode
THOMA, H., CIRSCH, W., HOLLE, J., & Mayr, W. (1989) · ASAIO Transactions · doi:10.1097/00002480-198907000-00103 · Cited by 43 · see citing works
Yoshida, K., Kurstjens, G. A. M., & Hennings, K. (2009) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2009.08.005 · Cited by 43 · see citing works
Transverse tripolar stimulation of peripheral nerve: a modelling study of spatial selectivity
Deurloo, K. E. I., Holsheimer, J., & Boom, H. B. K. (1998) · Medical & Biological Engineering & Computing · doi:10.1007/bf02522860 · Cited by 40 · see citing works
Early seizure detection in rats based on vagus nerve activity
Harreby, K. R., Sevcencu, C., & Struijk, J. J. (2010) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-010-0683-1 · Cited by 36 · see citing works
Self-citationNeurophysiological Effects of Left Vagal Stimulation in Man
Upton, A. R. M., Tougas, G. H., Talalla, A., et al. (1991) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1991.tb04050.x · Cited by 28 · see citing works
Fascicle selective recording with a nerve cuff electrode
Struijk, J. J., Haugland, M. K., & Thomsen, M. (2002) · Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.1996.656993 · Cited by 26 · see citing works
Self-citationNielsen, T. N., Kurstjens, G. A. M., & Struijk, J. J. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2010.2090879 · Cited by 25 · see citing works
Self-citationMeier, J. H., Rutten, W. L. C., & Boom, H. B. K. (1998) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.709558 · Cited by 22 · see citing works
Schuettler, M., Seetohul, V., Rijkhoff, N. J. M., Moeller, F. V., Donaldson, N., & Taylor, J. (2011) · 2011 Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2011.6090847 · Cited by 19 · see citing works
Nerve Stimulation With A Multi-Contact Cuff Electrode: Validation Of Model Predictions
Deurloo, K. E. I., Holsheimer, J., & Bergveld, P. (2000) · Archives of Physiology and Biochemistry · doi:10.1076/apab.108.4.349.4301 · Cited by 17 · see citing works
An implanted system for multi-site nerve cuff-based ENG recording using velocity selectivity
Clarke, C. T., Xu, X., Rieger, R., Taylor, J. T., & Donaldson, N. (2008) · Analog Integrated Circuits and Signal Processing · doi:10.1007/s10470-008-9233-2 · Cited by 15 · see citing works
Fascicle-Selectivity of an Intraneural Stimulation Electrode in the Rabbit Sciatic Nerve
Nielsen, T. N., Sevcencu, C., & Struijk, J. J. (2012) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2011.2169671 · Cited by 11 · see citing works
Self-citationA summary of the theory of velocity selective neural recording
Taylor, J. T., Schüettler, M., Clarke, C., et al. (2011) · doi:10.1109/iembs.2011.6091151 · Cited by 10 · see citing works
Intraoperative Recording of Sacral Root Nerve Signals in Humans
Kurstjens, M., Rijkhoff, N., Borau, A., et al. (2005) · Artificial Organs · doi:10.1111/j.1525-1594.2005.29044.x · Cited by 9 · see citing works
New electrode layout for internal selectivity of nerves
Rossel, O., Soulier, F., & Cathébras, G. (2009) · doi:10.1109/iembs.2009.5334437 · Cited by 6 · see citing works
Yoo, P. B., & Durand, D. M. (2005) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2004.1404200 · Cited by 5 · see citing works
Fascicle-selective multi-contact cuff electrode
Rossel, O., Soulier, F., Coulombe, J., Bernard, S., & Cathébras, G. (2011) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2011.6090820 · Cited by 3 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Safety of long-term electrical peripheral nerve stimulation: review of the state of the art
Günter, C., Delbeke, J., & Ortiz-Catalan, M. (2019) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-018-0474-8 · Cited by 198 · see citing works
A review for the peripheral nerve interface designer
Larson, C. E., & Meng, E. (2020) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2019.108523 · Cited by 133 · see citing works
Interfaces with the peripheral nervous system for the control of a neuroprosthetic limb: a review
Yildiz, K. A., Shin, A. Y., & Kaufman, K. R. (2020) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-020-00667-5 · Cited by 113 · see citing works
Active Neural Interface Circuits and Systems for Selective Control of Peripheral Nerves: A Review
Habibollahi, M., Jiang, D., Lancashire, H. T., & Demosthenous, A. (2024) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2024.3430038 · Cited by 11 · see citing works
“Electrode configuration may be used to focus extracellular fields and improve stimulation selectivity and efficacy [121], [122].”
- Authors
- Nielsen, T. N., Sevcencu, C., & Struijk, J. J.
- Affiliations
- Aalborg University (DK)
- Venue
- IEEE Transactions on Neural Systems and Rehabilitation Engineering
- Published
- 2014-01
- Volume
- 22
- Issue
- 1
- Pages
- 88-95
- ISSN
- 1534-4320 (ISSN-L)
- 1558-0210
- OpenAlex
- W1980086968
- Cited by
- 15
- References
- 39
- Citations by year
- 2026: 1 · 2025: 1 · 2024: 1 · 2022: 1 · 2021: 1 · 2020: 1 · 2019: 2 · 2018: 2 · 2016: 1 · 2015: 3 · 2014: 1
- FWCI
- 0.81
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- IEEE Transactions on Neural Systems and Rehabilitation Engineering (publishedVersion)
- PubMed (publishedVersion)
- VBN Forskningsportal (Aalborg Universitet) (submittedVersion)
- Funders
- Danish National Advanced Technology Foundation
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Engineering › Biomedical Engineering › Muscle activation and electromyography studies
- Life Sciences › Neuroscience › Neurology › Vagus Nerve Stimulation Research
- Keywords
- sciatic nerve, neurostimulation, neural recording, bipolar configuration, peripheral nerve interfaces, neuroprosthetics, cuff electrodes, rabbit model
- MeSH terms
- Action Potentials, Animals, Electric Stimulation, Electrodes, Implanted, Equipment Design, Peripheral Nerves, Rabbits, Sensitivity and Specificity, Reproducibility of Results, Equipment Failure Analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review