Achieving Routine Submillisievert CT Scanning: Report from the Summit on Management of Radiation Dose in CT
Open accessAbstract
This Special Report presents the consensus of the Summit on Management of Radiation Dose in Computed Tomography (CT) (held in February 2011), which brought together participants from academia, clinical practice, industry, and regulatory and funding agencies to identify the steps required to reduce the effective dose from routine CT examinations to less than 1 mSv. The most promising technologies and methods discussed at the summit include innovations and developments in x-ray sources; detectors; and image reconstruction, noise reduction, and postprocessing algorithms. Access to raw projection data and standard data sets for algorithm validation and optimization is a clear need, as is the need for new, clinically relevant metrics of image quality and diagnostic performance. Current commercially available techniques such as automatic exposure control, optimization of tube potential, beam-shaping filters, and dynamic z-axis collimators are important, and education to successfully implement these methods routinely is critically needed. Other methods that are just becoming widely available, such as iterative reconstruction, noise reduction, and postprocessing algorithms, will also have an important role. Together, these existing techniques can reduce dose by a factor of two to four. Technical advances that show considerable promise for additional dose reduction but are several years or more from commercial availability include compressed sensing, volume of interest and interior tomography techniques, and photon-counting detectors. This report offers a strategic roadmap for the CT user and research and manufacturer communities toward routinely achieving effective doses of less than 1 mSv, which is well below the average annual dose from naturally occurring sources of radiation.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Donoho, D. L. (2006) · IEEE Transactions on Information Theory · doi:10.1109/tit.2006.871582 · Cited by 23,994
Donoho, D. L. (2004) · Cited by 16,703
Candes, E. J., Romberg, J., & Tao, T. (2006) · IEEE Transactions on Information Theory · doi:10.1109/tit.2005.862083 · Cited by 15,639
Near-Optimal Signal Recovery From Random Projections: Universal Encoding Strategies?
Candes, E. J., & Tao, T. (2006) · IEEE Transactions on Information Theory · doi:10.1109/tit.2006.885507 · Cited by 6,839
Preface, Executive Summary and Glossary
(2007) · Annals of the ICRP · doi:10.1016/j.icrp.2007.10.003 · Cited by 2,106
Sparsity and incoherence in compressive sampling
Candès, E., & Romberg, J. (2007) · Inverse Problems · doi:10.1088/0266-5611/23/3/008 · Cited by 2,075
Sidky, E. Y., & Pan, X. (2008) · Physics in Medicine and Biology · doi:10.1088/0031-9155/53/17/021 · Cited by 1,995
Computed Tomography: Principles, Design, Artifacts, and Recent Advances
Hsieh, J. (2015) · doi:10.1117/3.2197756 · Cited by 1,400
Chen, G.-H., Tang, J., & Leng, S. (2008) · Medical Physics · doi:10.1118/1.2836423 · Cited by 1,141
Self-citationAccurate image reconstruction from few-views and limited-angle data in divergent-beam CT
Sidky, E. Y., Kao, C.-M., & Pan, X. (2006) · Journal of X-Ray Science and Technology: Clinical Applications of Diagnosis and Therapeutics · doi:10.3233/xst-2006-00155 · Cited by 1,091
A three‐dimensional statistical approach to improved image quality for multislice helical CT
Thibault, J., Sauer, K. D., Bouman, C. A., & Hsieh, J. (2007) · Medical Physics · doi:10.1118/1.2789499 · Cited by 937
Health Risks from Exposure to Low Levels of Ionizing Radiation: BEIR VII Phase 2
(2006) · National Academies Press eBooks · doi:10.17226/11340 · Cited by 904
1990 Recommendations of the International Commission on Radiological Protection.
(1991) · PubMed · Cited by 904
Schlomka, J. P., Roessl, E., Dorscheid, R., Dill, S., Martens, G., Istel, T., Bäumer, C., Herrmann, C., Steadman, R., Zeitler, G., et al. (2008) · Physics in Medicine and Biology · doi:10.1088/0031-9155/53/15/002 · Cited by 801
CT Dose Reduction and Dose Management Tools: Overview of Available Options
McCollough, C. H., Bruesewitz, M. R., & Kofler, J. M. (2006) · RadioGraphics · doi:10.1148/rg.262055138 · Cited by 768
Self-citationNCRP report no. 93: Ionizing radiation exposure of the population of the United States
Eckelman, W. C. (1988) · International Journal of Radiation Applications and Instrumentation. Part B. Nuclear Medicine and Biology · doi:10.1016/0883-2897(88)90020-7 · Cited by 755
Techniques and Applications of Automatic Tube Current Modulation for CT
Kalra, M. K., Maher, M. M., Toth, T. L., Schmidt, B., Westerman, B. L., Morgan, H. T., & Saini, S. (2004) · Radiology · doi:10.1148/radiol.2333031150 · Cited by 725
Jakobs, T. F., Becker, C. R., Ohnesorge, B., Flohr, T., Suess, C., Schoepf, U. J., & Reiser, M. F. (2002) · European Radiology · doi:10.1007/s00330-001-1278-x · Cited by 615
Silva, A. C., Lawder, H. J., Hara, A., Kujak, J., & Pavlicek, W. (2010) · American Journal of Roentgenology · doi:10.2214/ajr.09.2953 · Cited by 560
K-edge imaging in x-ray computed tomography using multi-bin photon counting detectors
Roessl, E., & Proksa, R. (2007) · Physics in Medicine and Biology · doi:10.1088/0031-9155/52/15/020 · Cited by 558
Penalized weighted least-squares image reconstruction for dual energy X-ray transmission tomography
Sukovic, P., & Clinthorne, N. H. (2000) · IEEE Transactions on Medical Imaging · doi:10.1109/42.896783 · Cited by 552
Wang, J., Li, T., Lu, H., & Liang, Z. (2006) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2006.882141 · Cited by 533
Model observers for assessment of image quality.
Barrett, H. H., Yao, J., Rolland, J. P., & Myers, K. J. (1993) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.90.21.9758 · Cited by 531
Compressed sensing based interior tomography
Yu, H., & Wang, G. (2009) · Physics in Medicine and Biology · doi:10.1088/0031-9155/54/9/014 · Cited by 519
Self-citationRealistic reference phantoms: An ICRP/ICRU joint effort
(2009) · Annals of the ICRP · doi:10.1016/j.icrp.2009.09.001 · Cited by 496
Human observer detection experiments with mammograms and power‐law noise
Burgess, A. E., Jacobson, F. L., & Judy, P. F. (2001) · Medical Physics · doi:10.1118/1.1355308 · Cited by 453
Singh, S., Kalra, M. K., Hsieh, J., Licato, P. E., Do, S., Pien, H. H., & Blake, M. A. (2010) · Radiology · doi:10.1148/radiol.10092212 · Cited by 435
Projection space denoising with bilateral filtering and CT noise modeling for dose reduction in CT
Manduca, A., Yu, L., Trzasko, J. D., Khaylova, N., Kofler, J. M., McCollough, C. M., & Fletcher, J. G. (2009) · Medical Physics · doi:10.1118/1.3232004 · Cited by 360
Self-citationPrimak, A. N., Ramirez Giraldo, J. C., Liu, X., Yu, L., & McCollough, C. H. (2009) · Medical Physics · doi:10.1118/1.3083567 · Cited by 318
Self-citationKachelrieß, M., Watzke, O., & Kalender, W. A. (2001) · Medical Physics · doi:10.1118/1.1358303 · Cited by 311
Self-citationNonlinear sinogram smoothing for low-dose X-ray CT
Li, T., Li, X., Wang, J., Wen, J., Lu, H., Hsieh, J., & Liang, Z. (2004) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2004.834824 · Cited by 309
Siegel, M. J., Schmidt, B., Bradley, D., Suess, C., & Hildebolt, C. (2004) · Radiology · doi:10.1148/radiol.2332032107 · Cited by 294
McCollough, C. H., & Schueler, B. A. (2000) · Medical Physics · doi:10.1118/1.598948 · Cited by 291
Self-citationFast Model-Based X-Ray CT Reconstruction Using Spatially Nonhomogeneous ICD Optimization
Yu, Z., Thibault, J.-B., Bouman, C. A., Sauer, K. D., & Hsieh, J. (2011) · IEEE Transactions on Image Processing · doi:10.1109/tip.2010.2058811 · Cited by 290
Tang, J., Nett, B. E., & Chen, G.-H. (2009) · Physics in Medicine and Biology · doi:10.1088/0031-9155/54/19/008 · Cited by 282
Self-citationSolitary pulmonary nodules: evaluation of blood flow patterns with dynamic CT.
Zhang, M., & Kono, M. (1997) · Radiology · doi:10.1148/radiology.205.2.9356631 · Cited by 271
(2007) · Annals of the ICRP · doi:10.1016/j.icrp.2007.11.001 · Cited by 256
Automatic selection of tube potential for radiation dose reduction in CT: A general strategy
Yu, L., Li, H., Fletcher, J. G., & McCollough, C. H. (2010) · Medical Physics · doi:10.1118/1.3264614 · Cited by 230
Self-citationFunama, Y., Awai, K., Nakayama, Y., Kakei, K., Nagasue, N., Shimamura, M., Sato, N., Sultana, S., Morishita, S., & Yamashita, Y. (2005) · Radiology · doi:10.1148/radiol.2373041643 · Cited by 226
Yu, L., Bruesewitz, M. R., Thomas, K. B., Fletcher, J. G., Kofler, J. M., & McCollough, C. H. (2011) · RadioGraphics · doi:10.1148/rg.313105079 · Cited by 217
Self-citationApplication‐ and patient size‐dependent optimization of x‐ray spectra for CT
Kalender, W. A., Deak, P., Kellermeier, M., van Straten, M., & Vollmar, S. V. (2009) · Medical Physics · doi:10.1118/1.3075901 · Cited by 206
Self-citationFeuerlein, S., Roessl, E., Proksa, R., Martens, G., Klass, O., Jeltsch, M., Rasche, V., Brambs, H.-J., Hoffmann, M. H. K., & Schlomka, J.-P. (2008) · Radiology · doi:10.1148/radiol.2492080560 · Cited by 195
(1991) · Annals of the ICRP · doi:10.1016/0146-6453(91)90065-o · Cited by 189
Computer-Simulated Radiation Dose Reduction for Abdominal Multidetector CT of Pediatric Patients
Frush, D. P., Slack, C. C., Hollingsworth, C. L., Bisset, G. S., Donnelly, L. F., Hsieh, J., Lavin-Wensell, T., & Mayo, J. R. (2002) · American Journal of Roentgenology · doi:10.2214/ajr.179.5.1791107 · Cited by 174
Wavelet Based Noise Reduction in CT-Images Using Correlation Analysis
Borsdorf, A., Raupach, R., Flohr, T., & Hornegger, J. (2008) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2008.923983 · Cited by 164
Channelized hotelling and human observer correlation for lesion detection in hepatic SPECT imaging.
Gifford, H. C., King, M. A., de Vries, D. J., & Soares, E. J. (2000) · PubMed · Cited by 162
Anatomical background and generalized detectability in tomosynthesis and cone‐beam CT
Gang, G. J., Tward, D. J., Lee, J., et al. (2010) · Medical Physics · doi:10.1118/1.3352586 · Cited by 134
Image covariance and lesion detectability in direct fan-beam x-ray computed tomography
Wunderlich, A., & Noo, F. (2008) · Physics in Medicine and Biology · doi:10.1088/0031-9155/53/10/002 · Cited by 113
Region of interest reconstruction from truncated data in circular cone-beam CT
Yu, L., Zou, Y. S., Sidky, E. Y., Pelizzari, C. A., Munro, P., & Pan, X. (2006) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2006.872329 · Cited by 101
Self-citationKarmazyn, B. W., Frush, D. P., Applegate, K. E., Maxfield, C. M., Cohen, M. D., & Jones, R. P. (2008) · American Journal of Roentgenology · doi:10.2214/ajr.08.1391 · Cited by 89
Lubner, M. G., Pickhardt, P. J., Tang, J., & Chen, G. (2011) · Radiology · doi:10.1148/radiol.11101380 · Cited by 80
Self-citationPotential of high‐Z contrast agents in clinical contrast‐enhanced computed tomography
Nowak, T., Hupfer, M., Brauweiler, R., Eisa, F., & Kalender, W. A. (2011) · Medical Physics · doi:10.1118/1.3658738 · Cited by 67
Self-citationVolume‐of‐interest (VOI) imaging in C‐arm flat‐detector CT for high image quality at reduced dose
Kolditz, D., Kyriakou, Y., & Kalender, W. A. (2010) · Medical Physics · doi:10.1118/1.3427641 · Cited by 62
Self-citationChen, L., Shaw, C. C., Altunbas, M. C., et al. (2008) · Medical Physics · doi:10.1118/1.2948397 · Cited by 46
The National Council on Radiation Protection and Measurements
Rogers, L. F. (2000) · American Journal of Roentgenology · doi:10.2214/ajr.175.6.1751493 · Cited by 38
The Medical Image Perception Society Update on Key Issues for Image Perception Research
Krupinski, E. A., & Berbaum, K. S. (2009) · Radiology · doi:10.1148/radiol.2531090237 · Cited by 29
Using Human and Model Performance to Compare MRI Reconstructions
Tisdall, M. D., & Atkins, M. S. (2006) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2006.881374 · Cited by 22
Cited by
Dual- and Multi-Energy CT: Principles, Technical Approaches, and Clinical Applications
McCollough, C. H., Leng, S., Yu, L., & Fletcher, J. G. (2015) · Radiology · doi:10.1148/radiol.2015142631 · Cited by 1,598
Self-citationWang, G. (2016) · IEEE Access · doi:10.1109/access.2016.2624938 · Cited by 455
Self-citationGreffier, J., Hamard, A., Pereira, F., Barrau, C., Pasquier, H., Beregi, J. P., & Frandon, J. (2020) · European Radiology · doi:10.1007/s00330-020-06724-w · Cited by 328
McCollough, C. H., Bartley, A. C., Carter, R. E., Chen, B., Drees, T. A., Edwards, P., Holmes, D. R., Huang, A. E., Khan, F., Leng, S., et al. (2017) · Medical Physics · doi:10.1002/mp.12345 · Cited by 308
Self-citationGutjahr, R., Halaweish, A. F., Yu, Z., Leng, S., Yu, L., Li, Z., Jorgensen, S. M., Ritman, E. L., Kappler, S., & McCollough, C. H. (2016) · Investigative Radiology · doi:10.1097/rli.0000000000000251 · Cited by 282
Self-citation
Metadata
- Authors
- McCollough, C. H., Chen, G. H., Kalender, W., Leng, S., Samei, E., Taguchi, K., Wang, G., Yu, L., & Pettigrew, R. I.
- Venue
- Radiology
- Published
- 2012-08
- Volume
- 264
- Issue
- 2
- Pages
- 567-580
- ISSN
- 0033-8419 (ISSN-L)
- 1527-1315
- OpenAlex
- W2142008745
- Cited by
- 276
- References
- 57
- FWCI
- 27.59
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.