Visual percepts evoked with an intracortical 96-channel microelectrode array inserted in human occipital cortex
Open accessThe DOI matches its Crossref record; 59 of the 60 references it cites were found in a registry or an index; 35 of the 69 sources its claims cite were checked for support.
Abstract
BACKGROUNDA long-held goal of vision therapy is to transfer information directly to the visual cortex of blind individuals, thereby restoring a rudimentary form of sight. However, no clinically available cortical visual prosthesis yet exists.METHODSWe implanted an intracortical microelectrode array consisting of 96 electrodes in the visual cortex of a 57-year-old person with complete blindness for a 6-month period. We measured thresholds and the characteristics of the visual percepts elicited by intracortical microstimulation.RESULTSImplantation and subsequent explantation of intracortical microelectrodes were carried out without complications. The mean stimulation threshold for single electrodes was 66.8 ± 36.5 μA. We consistently obtained high-quality recordings from visually deprived neurons and the stimulation parameters remained stable over time. Simultaneous stimulation via multiple electrodes was associated with a significant reduction in thresholds (P < 0.001, ANOVA) and evoked discriminable phosphene percepts, allowing the blind participant to identify some letters and recognize object boundaries.CONCLUSIONSOur results demonstrate the safety and efficacy of chronic intracortical microstimulation via a large number of electrodes in human visual cortex, showing its high potential for restoring functional vision in the blind.TRIAL REGISTRATIONClinicalTrials.gov identifier NCT02983370.FUNDINGThe Spanish Ministerio de Ciencia Innovación y Universidades, the Generalitat Valenciana (Spain), the Europan Union's Horizon 2020 programme, the Bidons Egara Research Chair of the University Miguel Hernández (Spain), and the John Moran Eye Center of the University of Utah.
Overview
CoChat Summary
TL;DR
In a 57-year-old woman with complete blindness, a 96-channel Utah Electrode Array implanted in right occipital cortex for 6 months was safely implanted and explanted, and intracortical microstimulation evoked phosphenes via 88 of 96 electrodes. Multi-electrode stimulation produced discriminable patterned percepts that allowed identification of some letters and object boundaries, though percepts were variable and required training and calibration.
Key findings
- Implantation and explantation occurred without complications; electrode impedances were mostly 37–56 kΩ (47 ± 4.8 kΩ), and single-electrode thresholds increased by 19.4% over the last 4 months, from 64.8 to 80.4 μA. Phosphenes were reliably evoked with 88 of 96 electrodes, with a mean single-electrode threshold of 66.8 ± 36.5 μA (range 7–128 μA).
- Simultaneous stimulation of larger electrode groups significantly reduced thresholds (P < 0.001, ANOVA); 4-electrode thresholds averaged 48.0 ± 22.6 μA, 28% lower than single electrodes. Two-point discrimination was limited: 23.6% of pair stimulations evoked 2 discrete phosphenes, while 76.4% evoked a single phosphene that was 1.9 ± 0.1 times larger; for electrodes 400 μm apart, temporal separation >250 ms yielded 2 distinct phosphenes 90% of the time, reaching 100% at ≥565 μm separation.
- The subject discriminated horizontal versus vertical patterns with 81.4% overall success across 22 sessions on 4 days, later reaching 100%; she reliably discriminated letters such as i, L, C, V, and O with overall success always >70%. In an O-versus-o task, performance improved to 90%–95% after 70 trials and 100% in the latest sessions.
- With a bio-inspired retinal encoder, the subject localized a large white square with 100% accuracy across 70 trials and a smaller square with 100% accuracy across 150 trials over 5 days; reaction time decreased from 8.6 ± 0.6 seconds to 5.4 ± 0.3 seconds after 2 training days.
Methods
- Single-subject feasibility study in a 57-year-old right-handed female with toxic optic neuropathy and complete blindness for 16 years. A Utah Electrode Array with 96 electrodes (1.5-mm shanks, 10 × 10 grid) was implanted in the right occipital cortex near the V1/V2 border and explanted after 6 months.
- Stimulation and recording used the CereStim 96 and NeuroPort systems, with up to 16 simultaneously stimulated electrodes; sessions occurred generally 5 days per week, 1–2 times per day, up to 4 hours per session, totaling approximately 540 hours of stimulation. Thresholds were measured with a binary search procedure (1–128 μA, 1-μA minimum step, 10% catch trials), and psychometric curves were fit with a Weibull cumulative distribution function to estimate 50% detection thresholds.
- Behavioral testing included sham-controlled detection, 2-alternative forced-choice discrimination, phosphene mapping on a digitizing tablet, subjective brightness/size ratings, two-point discrimination with varied electrode spacing and stimulus delays, and preliminary tests with a bio-inspired retinal encoder. Statistical analysis used IBM SPSS 20.0 with t tests or ANOVA and Scheffe’s procedure, with P < 0.05 considered significant.
Limitations
- Stated: the study involved a single subject over a 6-month period, and the authors state that further investigations in more subjects and over a longer period are required; simultaneous stimulation currents can summate, possibly requiring temporal interleaving; percepts sometimes did not clearly correlate with electrode locations, suggesting spatial remapping or fine calibration may be needed; not all letters could be induced, and many stimulation patterns were not recognized as letters; wireless systems are not yet sufficiently robust for clinical use.
- Stated: the CereStim 96 limited simultaneous stimulation to 16 electrodes, constraining the complexity of patterns that could be tested.
- Evident from the text: outcomes depend heavily on the single participant’s subjective reports of phosphene location, brightness, size, and shape, and there was no blinding or control group, limiting generalizability and objectivity.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures










Verification
Claims the paper makes
19 Valid · 2 Not valid · 5 Not checked · 144 Own results and interpretations
Some cited sources support this claim and some don’t — read each source’s verdict below.
In the paper
“The results therefore suggest that silicon-based, penetrating microelectrode arrays could provide the stability required in a clinical device (11, 32-34) but further investigations in more subjects and over a longer period of time are needed.”
What the sources say
- Partly supportedcites [11] Normann RA, Fernandez E. Clinical applications of penetrating neural interfaces and Utah Electrode Array technologies. J Neural Eng . 2016;13(6):061003. https://doi.org/10.1088/1741-2560/13/6/061003 sourceExperimental — accuracy not yet measured
Section 2.5 describes silicon as a mechanically strong substrate and says silicon carbide may provide long-term mechanical stability in vivo. But it also says materials have not yet been found that protect arrays for the years required in clinical devices. The Conclusions report limited current lifetimes and say further work is needed, so the source supports the possibility, not evidence that silicon-based arrays provide the required clinical stability. (LLM, conf=0.84)
- Disputedcites [32] Marin C, Fernandez E. Biocompatibility of intracortical microelectrodes: current status and future prospects. Front Neuroeng . 2010;3:8. sourceExperimental — accuracy not yet measured
The Introduction says long-term viability and stability remain a problem for available neural electrodes. Challenges and Future Perspectives adds that keeping intracortical microelectrodes biologically and electrically viable for many years remains difficult. Although the paper discusses penetrating arrays, it does not present silicon-based arrays as likely to provide the stability required for a clinical device. (LLM, conf=0.79)
Checked by a reasoning model — may be wrong.
- Not checked — no full text of the sourcecites [33] Fernandez E, Botella C. Biotolerability of intracortical microelectrodes. Advanced Biosystems . 2017;2(1):1-14. https://doi.org/10.1002/adbi.201700115 source
- Could not verifycites [34] Fernandez E. Development of visual neuroprostheses: trends and challenges. Bioelectron Med . 2018;4(8):1-8. sourceExperimental — accuracy not yet measured
The ‘Electrode-to-tissue Interface issues’ section discusses long-term viability as a challenge, including corrosion, glial scarring, and the need for improved protection of implant electronics. Searches for silicon-based penetrating arrays and clinical-device stability found no passage reporting results that suggest those arrays could provide the required stability. (LLM, conf=0.72)
hedgedstates a possible cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Current-induced tissue damage during electrical stimulation was avoided by keeping stimulus charge density below the damage threshold boundary (20, 21).”
What the sources say
- Partly supportedcites [20] Cogan SF, et al. Tissue damage thresholds during therapeutic electrical stimulation. J Neural Eng . 2016;13(2):021001. https://doi.org/10.1088/1741-2560/13/2/021001 sourceExperimental — accuracy not yet measured
The Introduction (page unavailable) says the Shannon boundary often distinguishes damaging from non-damaging stimulation for large electrodes. But the paper’s section on the origin of the 30 μC cm−2 threshold says charge density alone is insufficient to predict damage: tissue damage can occur below that limit. The claim overstates charge density as a general way to avoid damage. (LLM, conf=0.87)
- Not checked — no full text of the sourcecites [21] Shannon RV. A model of safe levels for electrical stimulation. IEEE Trans Biomed Eng . 1992;39(4):424-426. https://doi.org/10.1109/10.126616 source
plainstates a causeSingle case- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“and several studies indicate that UEAs retain their functionality in humans for more than 1000 days (29, 31).”
What the sources say
- Supportedcites [29] Willett FR, et al. High-performance brain-totext communication via handwriting. Nature . 2021;593(7858):249-254. https://doi.org/10.1038/s41586-021-03506-2 source
The Discussion (page not provided) states that microelectrode arrays have retained functionality for more than 1,000 days after implantation, citing prior studies and this study’s Extended Data Fig. 7. This supports the claim about long-term functionality in humans. (LLM, conf=0.93)
- Partly supportedcites [31] Simeral JD, et al. Home use of a percutaneous wireless intracortical brain-computer interface by individuals with tetraplegia. IEEE Trans Biomed Eng . 2021;68(7):2313-2325. https://doi.org/10.1109/TBME.2021.3069119 sourceExperimental — accuracy not yet measured
The Results section (page unavailable) shows that implanted intracortical arrays supported computer use in two people with tetraplegia, including T5 on trial day 588. The References section cites a study reporting cursor control 1000 days after array implantation, but this paper does not report functionality for more than 1000 days. (LLM, conf=0.85)
hedged- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“There was also a good agreement between predicted phosphene locations based on our presurgical analysis (refs. 22, 23; yellow area in Figure 5A) and the reported locations of the evoked phosphenes (blue dots with electrode numbers in Figure 5A).”
What the sources say
- Supportedcites [22] Benson NC, et al. The retinotopic organization of striate cortex is well predicted by surface topology. Curr Biol . 2012;22(21):2081-2085. https://doi.org/10.1016/j.cub.2012.09.014 source
The citation is attached to the citing authors’ presurgical prediction method, not to their observed agreement with evoked phosphenes. The source’s Results describe predicting an individual’s retinotopic map from an anatomical brain image, which supports its use for that method; the phosphene comparison is the citing authors’ own finding. (LLM, conf=0.86)
- Partly supportedcites [23] Butt OH, et al. The fine-scale functional correlation of striate cortex in sighted and blind people. J Neurosci . 2013;33(41):16209-16219. https://doi.org/10.1523/JNEUROSCI.0363-13.2013 sourceExperimental — accuracy not yet measured
The Introduction and Discussion describe using cortical anatomy to predict retinotopic positions in blind and sighted people, which supports the basis for predicting visual-field locations. The paper does not report presurgical phosphene predictions, evoked phosphenes, or agreement between predicted and reported locations; that agreement is presented as the citing authors’ own finding. (LLM, conf=0.76)
plainstates an associationSingle case- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Individuals with profound blindness generally do not live in a world of total blackness but experience a variety of spontaneous phosphenes (18, 19)”
What the sources say
- Partly supportedcites [18] Bittner AK, et al. Photopsias are related in part to perceived stress and positive mood in retinitis pigmentosa. Eye (Lond) . 2012;26(1):101-108. https://doi.org/10.1038/eye.2011.247 sourceExperimental — accuracy not yet measured
The Results report several types of spontaneous photopsias in people with retinitis pigmentosa, including flashes, dots or shapes, glare, and background glow. But the study concerns people with retinitis pigmentosa who underwent vision testing; it does not establish that individuals with profound blindness generally have these experiences or do not perceive total blackness (Results and Data collection; page not available). (LLM, conf=0.87)
- Not checked — no full text of the sourcecites [19] Boller F, et al. Charles bonnet syndrome and other hallucinatory phenomena. Front Neurol Neurosci . 2018;41:117-124. source
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
90 more claims · 1 with their cited sources checked
Create a free account to see all 170 claims and every source verdict.
- Probably not a substantial claim. 34
- It may state two separate claims as one. 50
- The quote only partly states this claim. 1
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“However, since the goal is not primarily to create natural vision, it suffices that phosphene characteristics are distinct and stable over time, which is the case for current implants (Luo et al., 2016; Fernández et al., 2021).”
From Towards a Smart Bionic Eye : AI-powered artificial vision for the treatment of incurable blindness (2022) · cites this paper as Fernández et al., 2021
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“However, empirical data show that phosphenes are not symmetric, either for retinal101 or cortical42 implants.”
From Visual Prostheses in the Era of Artificial Intelligence Technology (2025) · cites this paper as [42]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Note that the effects found by Fernández et al., 2021 (that were modeled by us) are consistent with findings by other studies, which report brighter phosphenes for higher stimulation strengths (Schmidt et al., 1996), and a lower stimulation efficiency (i.e. higher total charge thresholds) for longer pulse trains or higher pulse widths and frequencies (Niketeghad et al., 2020).”
From Towards biologically plausible phosphene simulation for the differentiable optimization of visual cortical prostheses (2024) · cites this paper as Fernández et al., 2021
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“The conscious perception of phosphenes requires considerable training and the detection process is influenced by attention (Fernández et al., 2021).”
From Towards biologically plausible phosphene simulation for the differentiable optimization of visual cortical prostheses (2024) · cites this paper as Fernández et al., 2021
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“more recent work has highlighted that the percepts resulting from multi-electrode stimulation cannot be explained by a summative model based on single-electrode phosphenes (Beauchamp et al., 2020; Barry et al., 2020; Fernández et al., 2021).”
From Towards a Smart Bionic Eye : AI-powered artificial vision for the treatment of incurable blindness (2022) · cites this paper as Fernández et al., 2021
Experimental — accuracy not yet measured
11 more claims about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
SciPy 1.0: fundamental algorithms for scientific computing in Python
Virtanen, P., Gommers, R., Oliphant, T. E., Haberland, M., Reddy, T., Cournapeau, D., Burovski, E., Peterson, P., Weckesser, W., Bright, J., et al. (2020) · Nature Methods · doi:10.1038/s41592-019-0686-2 · Cited by 38,430
SciPy 1.0: fundamental algorithms for scientific computing in Python
Virtanen, P., Gommers, R., Oliphant, T. E., Haberland, M., Reddy, T., Cournapeau, D., Burovski, E., Peterson, P., Weckesser, W., Bright, J., et al. (2019) · Apollo - University of Cambridge Repository · doi:10.17863/cam.64134 · Cited by 10,460
The sensations produced by electrical stimulation of the visual cortex
Brindley, G. S., & Lewin, W. S. (1968) · The Journal of Physiology · doi:10.1113/jphysiol.1968.sp008519 · Cited by 1,170
High-performance brain-to-text communication via handwriting
Willett, F. R., Avansino, D. T., Hochberg, L. R., Henderson, J. M., & Shenoy, K. V. (2021) · Nature · doi:10.1038/s41586-021-03506-2 · Cited by 1,024 · cited 1×
“A recent systematic review of human neuroprosthetic research using the UEA reported that the mean number of days of UEA implantation across all participants at the time of the study was at least 578 days (30), and several studies indicate that UEAs retain their functionality in humans for more than 1000 days (29, 31).”
Restoring cortical control of functional movement in a human with quadriplegia
Bouton, C. E., Shaikhouni, A., Annetta, N. V., Bockbrader, M. A., Friedenberg, D. A., Nielson, D. M., Sharma, G., Sederberg, P. B., Glenn, B. C., Mysiw, W. J., et al. (2016) · Nature · doi:10.1038/nature17435 · Cited by 924
Intracortical microstimulation of human somatosensory cortex
Flesher, S. N., Collinger, J. L., Foldes, S. T., Weiss, J. M., Downey, J. E., Tyler-Kabara, E. C., Bensmaia, S. J., Schwartz, A. B., Boninger, M. L., & Gaunt, R. A. (2016) · Science Translational Medicine · doi:10.1126/scitranslmed.aaf8083 · Cited by 704 · cited 1×
“Although the duration of this study was limited to 6 months, UEAs have been successfully implanted in human subjects in numerous studies (25–29) and these devices have continued to work for many years.”
Schmidt, E. M., Bak, M. J., Hambrecht, F. T., Kufta, C. V., O'Rourke, D. K., & Vallabhanath, P. (1996) · Brain · doi:10.1093/brain/119.2.507 · Cited by 672 · cited 6×
“We found similar results in previous animal experiments (35), and the visual prosthesis work with intracortical microelectrodes of Schmidt and colleagues also indicates that there is some variability phosphene location (10).”
InfluentialDecoding motor imagery from the posterior parietal cortex of a tetraplegic human
Aflalo, T., Kellis, S., Klaes, C., Lee, B., Shi, Y., Pejsa, K., Shanfield, K., Hayes-Jackson, S., Aisen, M., Heck, C., et al. (2015) · Science · doi:10.1126/science.aaa5417 · Cited by 649
A model of safe levels for electrical stimulation
Shannon, R. V. (1992) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.126616 · Cited by 526 · cited 3×
“Current-induced tissue damage during electrical stimulation was avoided by keeping stimulus charge density below the damage threshold boundary (20, 21).”
Dobelle, W. H., & Mladejovsky, M. G. (1974) · The Journal of Physiology · doi:10.1113/jphysiol.1974.sp010766 · Cited by 468
Research works printed but not linked: 1
- Brindley, 1968, Short and long-term stability of cortical electrical phosphenes, J Physiol, 196, 479, 10.1113/jphysiol.1968.sp008519 · DOI · Paper page
Cited by
Items 201–232 of 232
Kozai, T. D. Y., Hooks, B. M., Vazquez, A. L., Gharbawie, O. A., Huang, C., Graczyk, E. L., Smith, M. A., Taylor, D. M., Hernandez-Reynoso, A. G., London, A. J., et al. (2026) · Journal of Neural Engineering · doi:10.1088/1741-2552/ae54cf · Cited by 4
“Over the past decade, human studies progressed from proof-of-concept demonstrations of tactile percepts (figure 2(A)) to practical applications, including improved robotic prosthesis control [41] and emerging approaches for vision restoration [42, 59, 131].”
Self-citationZhou, J., Hu, D., Wu, M., Hong, Z., li, R., Chen, Q., Mi, X.-S., & Zhang, J. (2026) · Advances in Ophthalmology Practice and Research · doi:10.1016/j.aopr.2026.03.005
Fabrication of full soft diamond implants for functional rehabilitation
Wilfinger, C. (2023) · theses.fr (ABES) · doi:10.70675/66cda50dz7fa2z4482z8f5bz1e96fb082bf1 · Cited by 3
Optimizing cortical feedback strategies for a closed-loop brain machine interface
Lassagne, H. (2023) · theses.fr (ABES) · doi:10.70675/779d5bbaz5470z4faczb771z288a715634b1
A multi-region flexible neural interface for behavioral state decoding in freely moving mice
Tian, Y., Li, G., Su, H., Jiang, L., Luo, Y., Yang, Y., Huang, L., Li, J., Jin, S., Chen, P., et al. (2026) · Microsystems & Nanoengineering · doi:10.1038/s41378-026-01258-5 · Cited by 1
Qian, X., Liu, C., Zhang, Y., Subramaniam, S., Fang, Q., Deen, J., & Jiang, W. (2026) · Progress in Biomedical Engineering · doi:10.1088/2516-1091/ae67c5 · Cited by 1
“With the Utah-style MEAs, researchers were able to decode neural signals from patients with neurodegenerative diseases into actions such as moving a cursor [47], texting [19], and handwriting [20], even stimulating visually deprived neurons in a sustained amount of time [48].”
Wireless image transfer across opaque cornea: A novel approach to visual rehabilitation
Natarajan, R. (2026) · Indian Journal of Ophthalmology · doi:10.4103/ijo.ijo_368_26
Phillips, A., Lotlikar, A., Kling, A., Hays, M. R., Vasireddy, P., Hofflich, B., Sharon, A., Sommeling, M., Brown, J. B., Baum, C., et al. (2026) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.64898/2026.05.04.721188 · Cited by 2
“Several different methods have been proposed and tested to leverage these percepts in useful ways via static or dynamic multielectrode stimulation strategies [34,35], but these methods are limited to simple patterns and do not extend to complex or complete visual scenes.”
МОДЕЛЮВАННЯ ПРОЦЕСІВ ФОРМУВАННЯ ФОСФЕННИХ ОБРАЗІВ У СИСТЕМАХ ВІЗУАЛЬНИХ НЕЙРОІНТЕРФЕЙСІВ
Yeroshenko, O. (2026) · Системи управління навігації та зв’язку Збірник наукових праць · doi:10.26906/sunz.2026.2.081 · Cited by 1
Advances in neuroprostheses: interfaces, materials, and applications
He, E., Chen, K., Liu, S., Chen, H., Xiao, Y., Chen, R., Tu, P., Pan, G., & Lin, P. (2026) · Nano Convergence · doi:10.1186/s40580-026-00552-2 · Cited by 1
Aniorte, A., Gea, M., Garrigós, J., Martínez, J. J., & Ferrández, J. M. (2026) · Lecture notes in computer science · doi:10.1007/978-3-032-27314-7_38
Visual Cortex Stimulation Modulates Muscle Networks Underlying Postural Control in Blindness
Albarracín, A. L., Cano, L. A., Farfán, F. D., et al. (2026) · Lecture notes in computer science · doi:10.1007/978-3-032-27314-7_40
Retinitis pigmentosa: Diagnosis and management
Rosin, B., & Sahel, J. (2026) · Handbook of clinical neurology · doi:10.1016/b978-0-443-22212-2.00005-7
Vision restoration: From prostheses to genetic-based brain–machine interfaces
Alcala, I., DESAILLY, E., Arcizet, F., Marazova, K., Gauvain, G., & Picaud, S. A. (2026) · Handbook of clinical neurology · doi:10.1016/b978-0-443-22212-2.00017-3
Kim, R., Lycke, R., Zolotavin, P., Montes, J., Xie, C., & Luan, L. (2026) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.64898/2026.06.05.730421
Stimulus Optimization Using the Artificial Bee Colony Algorithm for Visual Neuroprostheses
Pratiwi, A. M., Jara-Balsera, J., Guzmán-Miranda, H., Suaning, G. J., & Barriga‐Rivera, A. (2026) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2026.3706302
“This would be a potential solution to the variations in the perceptual threshold over time [48], obviating the needs for lengthy stimulation tuning sessions.”
Artem, I., Ma, H. F., Li, F. H., Chen, Z., Xu, M., YU, C., Li, R. K., Wu, J., & He, F. (2026) · Cell Reports Physical Science · doi:10.64898/2026.06.17.733032
Fukuda, S., & Hayashida, Y. (2025) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner61569.2025.11589044
“However, the energy required to induce these phosphenes varies significantly depending on the parameters of the stimulus pulse [1][2].”
A Programmable System for Concurrent Neural Stimulation and Recording with Artifact Mitigation
Schmitz, C. M., & Viventi, J. (2025) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner61569.2025.11588950
“Stimulation currents above 200 µA can be delivered without saturation, which is sufficient for modulating neural activity with intracortical [41]-[44] and µECoG electrodes [16], [45]- [47].”
Singh, S. L. R. R., & Chakrabarty, S. (2026) · Studies in Rhythm Engineering · doi:10.1007/978-3-032-19974-4_10
Trends in neuroprosthetic technologies to restore sensations
Park, B., Butterworth, W., Lee, N., & Kim, S. (2026) · Biomedical Engineering Letters · doi:10.1007/s13534-026-00603-x
“A study using a 96-channel intracortical array reported that phosphene-based stimulation enabled the identification of some letters and object boundaries [95].”
InfluentialScialom, E., Lonnqvist, B., Bornet, A., & Herzog, M. H. (2026) · Journal of Neural Engineering · doi:10.1088/1741-2552/ae9227
“This observation is particularly relevant given that the number of electrodes that can be simultaneously stimulated in current cortical visual prostheses remains severely limited in humans (Fernández et al 2021, Bosking et al 2022).”
Valle, G., Alamri, A. H., Downey, J. E., Lienkämper, R., Jordan, P. M., Miller, L. E., Gaunt, R. A., Greenspon, C. M., & Bensmaïa, S. J. (2026) · Springer briefs in electrical and computer engineering · doi:10.1007/978-3-032-25828-1_4
Fares, H., Ronchini, M., Zamani, M., Farkhani, H., Chiappalone, M., Neftci, E., & Moradi, F. (2026) · npj Biomedical Innovations. · doi:10.1038/s44385-026-00074-w · Cited by 2
Kim, R., Lycke, R., Zolotavin, P., Montes, J., Xie, C., & Luan, L. (2026) · Science Advances · doi:10.1126/sciadv.aef0343 · Cited by 1
Huggins, J. E., Mattia, D., Aarnoutse, E. J., Allison, B. Z., Berezutskaya, J., Bianchi, L., Chavarriaga, R., Corsi, M., Dean, H., Fernandes, F., et al. (2026) · Research in Biomedical Engineering and Technology · doi:10.1080/29960355.2026.2731643
Decoding natural scenes from patterned optogenetic responses in mouse visual cortexPreprint
Jia, S., Xin, T., Li, C., Ji, H., Wu, R., Wang, S., Guo, Q., Yu, Z., Liu, J. K., & Li, Y.-T. (2026) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.64898/2026.09.21.753135
“On one hand, substantial progress has been made in stimulation technologies, including electrical interfaces with increasingly high spatial resolution (7) and, more recently, optogenetic stimulation approaches that offer cell-type specificity and finer spatial control (8; 9).”
Ma, Y., Li, Y., Zhang, S., Liang, L., & Li, G. (2026) · Neural Networks · doi:10.1016/j.neunet.2026.109693
Restorative and regenerative therapies for vision loss
Boychev, N., Soucy, J. R., Dang, S. H., Tran, C. T., Malkin, A. G., Yeung, V., Ross, A. E., Hutcheon, A. E. K., & Kanu, L. N. (2026) · Eye · doi:10.1038/s41433-026-04913-3
Stimulating current direction produces depth-dependent spatial selectivity in cortical activation
Shumkova, D., Jiang, X., Price, N. S. C., & Wong, Y. T. (2026) · Brain stimulation · doi:10.1016/j.brs.2026.103232
Deep Learning–Based Control of Electrically Evoked Activity in Human Visual Cortex
Moure, P., Granley, J., Grani, F., Soo, L., Lozano, A., López-Peco, R., Villamarin-Ortiz, A., Soto-Sanchez, C., Liu, S.-C., Beyeler, M., et al. (2025) · bioRxiv · doi:10.1101/2025.09.24.678361
“One prominent example of neuroprostheses is the visual prosthesis, which electrically stimulates neurons in the retina [11–14], optic nerve [15], lateral geniculate nucleus [16], or early visual cortex [17–20] to restore visual perception.”
InfluentialSelf-citation
Metadata
- Authors
- Fernández, E., Alfaro, A., Soto-Sánchez, C., Gonzalez-Lopez, P., Lozano, A. M., Peña, S., Grima, M. D., Rodil, A., Gómez, B., Chen, X., Roelfsema, P. R., Rolston, J. D., Davis, T. S., & Normann, R. A.
- Affiliations
- Universitat de Miguel Hernández d'Elx (ES)
- Netherlands Institute for Neuroscience (NL)
- University of Utah (US)
- ORCID
- Venue
- Journal of Clinical Investigation
- Published
- 2021-12-01
- Volume
- 131
- Issue
- 23
- ISSN
- 0021-9738 (ISSN-L)
- 1558-8238
- PMC
- PMC8631600
- OpenAlex
- W3207382873
- Semantic Scholar
- 3fc2ad1299c6210f699eec7354757ffedb0bdc7a
- Cited by
- 246
- References
- 56
- Citations by year
- 2026: 63 · 2025: 69 · 2024: 31 · 2023: 42 · 2022: 40 · 2021: 1
- FWCI
- 10.33
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Journal of Clinical Investigation (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Pure Amsterdam UMC (publishedVersion, cc-by)
- KNAW Research Portal (The Royal Netherlands Academy of Arts and Sciences) (publishedVersion, other-oa)
- KNAW research portal (Royal Academy of Art and Sciences (KNAW)) (publishedVersion, other-oa)
- VU Research Portal (publishedVersion, other-oa)
- Data Archiving and Networked Services (DANS) (submittedVersion, other-oa)
- Funders
- Spanish Ministerio de Ciencia Innovación y Universidades (RTI2018-098969-B-100)
- Generalitat Valenciana (PROMETEO/2019/119)
- Bidons Egara Research Chair of the University Miguel Hernández (2019)
- John Moran Eye Center of the University of Utah (2019)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Life Sciences › Neuroscience › Cognitive Neuroscience › Tactile and Sensory Interactions
- Life Sciences › Neuroscience › Cognitive Neuroscience › EEG and Brain-Computer Interfaces
- Keywords
- cortical visual prosthesis, intracortical microelectrode arrays, microstimulation, visual percepts, phosphenes, blindness, visual cortex, stimulation threshold, vision restoration
- MeSH terms
- Blindness, Electric Stimulation, Electrodes, Implanted, Female, Humans, Microelectrodes, Middle Aged, Occipital Lobe, Optic Nerve Diseases, Phosphenes, Retina, Vision, Ocular, Visual Cortex, Visual Perception, Treatment Outcome, Visual Prosthesis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.