Visual percepts evoked with an intracortical 96-channel microelectrode array inserted in human occipital cortex
Open accessThe DOI matches its Crossref record; 59 of the 60 references it cites were found in a registry or an index; 35 of the 69 sources its claims cite were checked for support.
Abstract
BACKGROUNDA long-held goal of vision therapy is to transfer information directly to the visual cortex of blind individuals, thereby restoring a rudimentary form of sight. However, no clinically available cortical visual prosthesis yet exists.METHODSWe implanted an intracortical microelectrode array consisting of 96 electrodes in the visual cortex of a 57-year-old person with complete blindness for a 6-month period. We measured thresholds and the characteristics of the visual percepts elicited by intracortical microstimulation.RESULTSImplantation and subsequent explantation of intracortical microelectrodes were carried out without complications. The mean stimulation threshold for single electrodes was 66.8 ± 36.5 μA. We consistently obtained high-quality recordings from visually deprived neurons and the stimulation parameters remained stable over time. Simultaneous stimulation via multiple electrodes was associated with a significant reduction in thresholds (P < 0.001, ANOVA) and evoked discriminable phosphene percepts, allowing the blind participant to identify some letters and recognize object boundaries.CONCLUSIONSOur results demonstrate the safety and efficacy of chronic intracortical microstimulation via a large number of electrodes in human visual cortex, showing its high potential for restoring functional vision in the blind.TRIAL REGISTRATIONClinicalTrials.gov identifier NCT02983370.FUNDINGThe Spanish Ministerio de Ciencia Innovación y Universidades, the Generalitat Valenciana (Spain), the Europan Union's Horizon 2020 programme, the Bidons Egara Research Chair of the University Miguel Hernández (Spain), and the John Moran Eye Center of the University of Utah.
Overview
CoChat Summary
TL;DR
In a 57-year-old woman with complete blindness, a 96-channel Utah Electrode Array implanted in right occipital cortex for 6 months was safely implanted and explanted, and intracortical microstimulation evoked phosphenes via 88 of 96 electrodes. Multi-electrode stimulation produced discriminable patterned percepts that allowed identification of some letters and object boundaries, though percepts were variable and required training and calibration.
Key findings
- Implantation and explantation occurred without complications; electrode impedances were mostly 37–56 kΩ (47 ± 4.8 kΩ), and single-electrode thresholds increased by 19.4% over the last 4 months, from 64.8 to 80.4 μA. Phosphenes were reliably evoked with 88 of 96 electrodes, with a mean single-electrode threshold of 66.8 ± 36.5 μA (range 7–128 μA).
- Simultaneous stimulation of larger electrode groups significantly reduced thresholds (P < 0.001, ANOVA); 4-electrode thresholds averaged 48.0 ± 22.6 μA, 28% lower than single electrodes. Two-point discrimination was limited: 23.6% of pair stimulations evoked 2 discrete phosphenes, while 76.4% evoked a single phosphene that was 1.9 ± 0.1 times larger; for electrodes 400 μm apart, temporal separation >250 ms yielded 2 distinct phosphenes 90% of the time, reaching 100% at ≥565 μm separation.
- The subject discriminated horizontal versus vertical patterns with 81.4% overall success across 22 sessions on 4 days, later reaching 100%; she reliably discriminated letters such as i, L, C, V, and O with overall success always >70%. In an O-versus-o task, performance improved to 90%–95% after 70 trials and 100% in the latest sessions.
- With a bio-inspired retinal encoder, the subject localized a large white square with 100% accuracy across 70 trials and a smaller square with 100% accuracy across 150 trials over 5 days; reaction time decreased from 8.6 ± 0.6 seconds to 5.4 ± 0.3 seconds after 2 training days.
Methods
- Single-subject feasibility study in a 57-year-old right-handed female with toxic optic neuropathy and complete blindness for 16 years. A Utah Electrode Array with 96 electrodes (1.5-mm shanks, 10 × 10 grid) was implanted in the right occipital cortex near the V1/V2 border and explanted after 6 months.
- Stimulation and recording used the CereStim 96 and NeuroPort systems, with up to 16 simultaneously stimulated electrodes; sessions occurred generally 5 days per week, 1–2 times per day, up to 4 hours per session, totaling approximately 540 hours of stimulation. Thresholds were measured with a binary search procedure (1–128 μA, 1-μA minimum step, 10% catch trials), and psychometric curves were fit with a Weibull cumulative distribution function to estimate 50% detection thresholds.
- Behavioral testing included sham-controlled detection, 2-alternative forced-choice discrimination, phosphene mapping on a digitizing tablet, subjective brightness/size ratings, two-point discrimination with varied electrode spacing and stimulus delays, and preliminary tests with a bio-inspired retinal encoder. Statistical analysis used IBM SPSS 20.0 with t tests or ANOVA and Scheffe’s procedure, with P < 0.05 considered significant.
Limitations
- Stated: the study involved a single subject over a 6-month period, and the authors state that further investigations in more subjects and over a longer period are required; simultaneous stimulation currents can summate, possibly requiring temporal interleaving; percepts sometimes did not clearly correlate with electrode locations, suggesting spatial remapping or fine calibration may be needed; not all letters could be induced, and many stimulation patterns were not recognized as letters; wireless systems are not yet sufficiently robust for clinical use.
- Stated: the CereStim 96 limited simultaneous stimulation to 16 electrodes, constraining the complexity of patterns that could be tested.
- Evident from the text: outcomes depend heavily on the single participant’s subjective reports of phosphene location, brightness, size, and shape, and there was no blinding or control group, limiting generalizability and objectivity.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures










Verification
Claims the paper makes
19 Valid · 2 Not valid · 5 Not checked · 144 Own results and interpretations
Some cited sources support this claim and some don’t — read each source’s verdict below.
In the paper
“The results therefore suggest that silicon-based, penetrating microelectrode arrays could provide the stability required in a clinical device (11, 32-34) but further investigations in more subjects and over a longer period of time are needed.”
What the sources say
- Partly supportedcites [11] Normann RA, Fernandez E. Clinical applications of penetrating neural interfaces and Utah Electrode Array technologies. J Neural Eng . 2016;13(6):061003. https://doi.org/10.1088/1741-2560/13/6/061003 sourceExperimental — accuracy not yet measured
Section 2.5 describes silicon as a mechanically strong substrate and says silicon carbide may provide long-term mechanical stability in vivo. But it also says materials have not yet been found that protect arrays for the years required in clinical devices. The Conclusions report limited current lifetimes and say further work is needed, so the source supports the possibility, not evidence that silicon-based arrays provide the required clinical stability. (LLM, conf=0.84)
- Disputedcites [32] Marin C, Fernandez E. Biocompatibility of intracortical microelectrodes: current status and future prospects. Front Neuroeng . 2010;3:8. sourceExperimental — accuracy not yet measured
The Introduction says long-term viability and stability remain a problem for available neural electrodes. Challenges and Future Perspectives adds that keeping intracortical microelectrodes biologically and electrically viable for many years remains difficult. Although the paper discusses penetrating arrays, it does not present silicon-based arrays as likely to provide the stability required for a clinical device. (LLM, conf=0.79)
Checked by a reasoning model — may be wrong.
- Not checked — no full text of the sourcecites [33] Fernandez E, Botella C. Biotolerability of intracortical microelectrodes. Advanced Biosystems . 2017;2(1):1-14. https://doi.org/10.1002/adbi.201700115 source
- Could not verifycites [34] Fernandez E. Development of visual neuroprostheses: trends and challenges. Bioelectron Med . 2018;4(8):1-8. sourceExperimental — accuracy not yet measured
The ‘Electrode-to-tissue Interface issues’ section discusses long-term viability as a challenge, including corrosion, glial scarring, and the need for improved protection of implant electronics. Searches for silicon-based penetrating arrays and clinical-device stability found no passage reporting results that suggest those arrays could provide the required stability. (LLM, conf=0.72)
hedgedstates a possible cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Current-induced tissue damage during electrical stimulation was avoided by keeping stimulus charge density below the damage threshold boundary (20, 21).”
What the sources say
- Partly supportedcites [20] Cogan SF, et al. Tissue damage thresholds during therapeutic electrical stimulation. J Neural Eng . 2016;13(2):021001. https://doi.org/10.1088/1741-2560/13/2/021001 sourceExperimental — accuracy not yet measured
The Introduction (page unavailable) says the Shannon boundary often distinguishes damaging from non-damaging stimulation for large electrodes. But the paper’s section on the origin of the 30 μC cm−2 threshold says charge density alone is insufficient to predict damage: tissue damage can occur below that limit. The claim overstates charge density as a general way to avoid damage. (LLM, conf=0.87)
- Not checked — no full text of the sourcecites [21] Shannon RV. A model of safe levels for electrical stimulation. IEEE Trans Biomed Eng . 1992;39(4):424-426. https://doi.org/10.1109/10.126616 source
plainstates a causeSingle case- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“and several studies indicate that UEAs retain their functionality in humans for more than 1000 days (29, 31).”
What the sources say
- Supportedcites [29] Willett FR, et al. High-performance brain-totext communication via handwriting. Nature . 2021;593(7858):249-254. https://doi.org/10.1038/s41586-021-03506-2 source
The Discussion (page not provided) states that microelectrode arrays have retained functionality for more than 1,000 days after implantation, citing prior studies and this study’s Extended Data Fig. 7. This supports the claim about long-term functionality in humans. (LLM, conf=0.93)
- Partly supportedcites [31] Simeral JD, et al. Home use of a percutaneous wireless intracortical brain-computer interface by individuals with tetraplegia. IEEE Trans Biomed Eng . 2021;68(7):2313-2325. https://doi.org/10.1109/TBME.2021.3069119 sourceExperimental — accuracy not yet measured
The Results section (page unavailable) shows that implanted intracortical arrays supported computer use in two people with tetraplegia, including T5 on trial day 588. The References section cites a study reporting cursor control 1000 days after array implantation, but this paper does not report functionality for more than 1000 days. (LLM, conf=0.85)
hedged- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“There was also a good agreement between predicted phosphene locations based on our presurgical analysis (refs. 22, 23; yellow area in Figure 5A) and the reported locations of the evoked phosphenes (blue dots with electrode numbers in Figure 5A).”
What the sources say
- Supportedcites [22] Benson NC, et al. The retinotopic organization of striate cortex is well predicted by surface topology. Curr Biol . 2012;22(21):2081-2085. https://doi.org/10.1016/j.cub.2012.09.014 source
The citation is attached to the citing authors’ presurgical prediction method, not to their observed agreement with evoked phosphenes. The source’s Results describe predicting an individual’s retinotopic map from an anatomical brain image, which supports its use for that method; the phosphene comparison is the citing authors’ own finding. (LLM, conf=0.86)
- Partly supportedcites [23] Butt OH, et al. The fine-scale functional correlation of striate cortex in sighted and blind people. J Neurosci . 2013;33(41):16209-16219. https://doi.org/10.1523/JNEUROSCI.0363-13.2013 sourceExperimental — accuracy not yet measured
The Introduction and Discussion describe using cortical anatomy to predict retinotopic positions in blind and sighted people, which supports the basis for predicting visual-field locations. The paper does not report presurgical phosphene predictions, evoked phosphenes, or agreement between predicted and reported locations; that agreement is presented as the citing authors’ own finding. (LLM, conf=0.76)
plainstates an associationSingle case- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Individuals with profound blindness generally do not live in a world of total blackness but experience a variety of spontaneous phosphenes (18, 19)”
What the sources say
- Partly supportedcites [18] Bittner AK, et al. Photopsias are related in part to perceived stress and positive mood in retinitis pigmentosa. Eye (Lond) . 2012;26(1):101-108. https://doi.org/10.1038/eye.2011.247 sourceExperimental — accuracy not yet measured
The Results report several types of spontaneous photopsias in people with retinitis pigmentosa, including flashes, dots or shapes, glare, and background glow. But the study concerns people with retinitis pigmentosa who underwent vision testing; it does not establish that individuals with profound blindness generally have these experiences or do not perceive total blackness (Results and Data collection; page not available). (LLM, conf=0.87)
- Not checked — no full text of the sourcecites [19] Boller F, et al. Charles bonnet syndrome and other hallucinatory phenomena. Front Neurol Neurosci . 2018;41:117-124. source
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
90 more claims · 1 with their cited sources checked
Create a free account to see all 170 claims and every source verdict.
- Probably not a substantial claim. 34
- It may state two separate claims as one. 50
- The quote only partly states this claim. 1
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“However, since the goal is not primarily to create natural vision, it suffices that phosphene characteristics are distinct and stable over time, which is the case for current implants (Luo et al., 2016; Fernández et al., 2021).”
From Towards a Smart Bionic Eye : AI-powered artificial vision for the treatment of incurable blindness (2022) · cites this paper as Fernández et al., 2021
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“However, empirical data show that phosphenes are not symmetric, either for retinal101 or cortical42 implants.”
From Visual Prostheses in the Era of Artificial Intelligence Technology (2025) · cites this paper as [42]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Note that the effects found by Fernández et al., 2021 (that were modeled by us) are consistent with findings by other studies, which report brighter phosphenes for higher stimulation strengths (Schmidt et al., 1996), and a lower stimulation efficiency (i.e. higher total charge thresholds) for longer pulse trains or higher pulse widths and frequencies (Niketeghad et al., 2020).”
From Towards biologically plausible phosphene simulation for the differentiable optimization of visual cortical prostheses (2024) · cites this paper as Fernández et al., 2021
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“The conscious perception of phosphenes requires considerable training and the detection process is influenced by attention (Fernández et al., 2021).”
From Towards biologically plausible phosphene simulation for the differentiable optimization of visual cortical prostheses (2024) · cites this paper as Fernández et al., 2021
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“more recent work has highlighted that the percepts resulting from multi-electrode stimulation cannot be explained by a summative model based on single-electrode phosphenes (Beauchamp et al., 2020; Barry et al., 2020; Fernández et al., 2021).”
From Towards a Smart Bionic Eye : AI-powered artificial vision for the treatment of incurable blindness (2022) · cites this paper as Fernández et al., 2021
Experimental — accuracy not yet measured
11 more claims about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
SciPy 1.0: fundamental algorithms for scientific computing in Python
Virtanen, P., Gommers, R., Oliphant, T. E., Haberland, M., Reddy, T., Cournapeau, D., Burovski, E., Peterson, P., Weckesser, W., Bright, J., et al. (2020) · Nature Methods · doi:10.1038/s41592-019-0686-2 · Cited by 38,430
SciPy 1.0: fundamental algorithms for scientific computing in Python
Virtanen, P., Gommers, R., Oliphant, T. E., Haberland, M., Reddy, T., Cournapeau, D., Burovski, E., Peterson, P., Weckesser, W., Bright, J., et al. (2019) · Apollo - University of Cambridge Repository · doi:10.17863/cam.64134 · Cited by 10,460
The sensations produced by electrical stimulation of the visual cortex
Brindley, G. S., & Lewin, W. S. (1968) · The Journal of Physiology · doi:10.1113/jphysiol.1968.sp008519 · Cited by 1,170
High-performance brain-to-text communication via handwriting
Willett, F. R., Avansino, D. T., Hochberg, L. R., Henderson, J. M., & Shenoy, K. V. (2021) · Nature · doi:10.1038/s41586-021-03506-2 · Cited by 1,024 · cited 1×
“A recent systematic review of human neuroprosthetic research using the UEA reported that the mean number of days of UEA implantation across all participants at the time of the study was at least 578 days (30), and several studies indicate that UEAs retain their functionality in humans for more than 1000 days (29, 31).”
Restoring cortical control of functional movement in a human with quadriplegia
Bouton, C. E., Shaikhouni, A., Annetta, N. V., Bockbrader, M. A., Friedenberg, D. A., Nielson, D. M., Sharma, G., Sederberg, P. B., Glenn, B. C., Mysiw, W. J., et al. (2016) · Nature · doi:10.1038/nature17435 · Cited by 924
Intracortical microstimulation of human somatosensory cortex
Flesher, S. N., Collinger, J. L., Foldes, S. T., Weiss, J. M., Downey, J. E., Tyler-Kabara, E. C., Bensmaia, S. J., Schwartz, A. B., Boninger, M. L., & Gaunt, R. A. (2016) · Science Translational Medicine · doi:10.1126/scitranslmed.aaf8083 · Cited by 704 · cited 1×
“Although the duration of this study was limited to 6 months, UEAs have been successfully implanted in human subjects in numerous studies (25–29) and these devices have continued to work for many years.”
Schmidt, E. M., Bak, M. J., Hambrecht, F. T., Kufta, C. V., O'Rourke, D. K., & Vallabhanath, P. (1996) · Brain · doi:10.1093/brain/119.2.507 · Cited by 672 · cited 6×
“We found similar results in previous animal experiments (35), and the visual prosthesis work with intracortical microelectrodes of Schmidt and colleagues also indicates that there is some variability phosphene location (10).”
InfluentialDecoding motor imagery from the posterior parietal cortex of a tetraplegic human
Aflalo, T., Kellis, S., Klaes, C., Lee, B., Shi, Y., Pejsa, K., Shanfield, K., Hayes-Jackson, S., Aisen, M., Heck, C., et al. (2015) · Science · doi:10.1126/science.aaa5417 · Cited by 649
A model of safe levels for electrical stimulation
Shannon, R. V. (1992) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.126616 · Cited by 526 · cited 3×
“Current-induced tissue damage during electrical stimulation was avoided by keeping stimulus charge density below the damage threshold boundary (20, 21).”
Dobelle, W. H., & Mladejovsky, M. G. (1974) · The Journal of Physiology · doi:10.1113/jphysiol.1974.sp010766 · Cited by 468
Research works printed but not linked: 1
- Brindley, 1968, Short and long-term stability of cortical electrical phosphenes, J Physiol, 196, 479, 10.1113/jphysiol.1968.sp008519 · DOI · Paper page
Cited by
Celebrating the one millionth cochlear implant
Zeng, F.-G. (2022) · JASA Express Letters · doi:10.1121/10.0012825 · Cited by 195
Game changers in science and technology - now and beyond
Betz, U. A. K., Arora, L., Assal, R. A., de Azevedo, H. F. Z., Baldwin, J., Becker, M. S., Bostock, S., Cheng, V. W. T., Egle, T., Ferrari, N., et al. (2023) · Technological Forecasting and Social Change · doi:10.1016/j.techfore.2023.122588 · Cited by 182
Modulating Brain Activity with Invasive Brain–Computer Interface: A Narrative Review
Zhao, Z., Nie, C., Jiang, C.-T., Cao, S.-H., Tian, K.-X., Yu, S., & Gu, J. (2023) · Brain Sciences · doi:10.3390/brainsci13010134 · Cited by 99
“The research team developed an invasive BCI with ICMS in the occipital cortex and successfully helped the patient to identify some letters and object boundaries (see Figure 3b) [85].”
MethodLycke, R., Kim, R., Zolotavin, P., Montes, J., Sun, Y., Kőszeghy, Á., Altun, E., Noble, B., Yin, R., He, F., et al. (2023) · Cell Reports · doi:10.1016/j.celrep.2023.112554 · Cited by 81
A review on microelectrode array fabrication techniques and their applications
Tanwar, A., Gandhi, H. A., Kushwaha, D., & Bhattacharya, J. (2022) · Materials Today Chemistry · doi:10.1016/j.mtchem.2022.101153 · Cited by 92
Cadoni, S., Demené, C., Alcala, I., Provansal, M., Nguyen, D., Nelidova, D., Labernède, G., Lubetzki, J., Goulet, R., Burban, E., et al. (2023) · Nature Nanotechnology · doi:10.1038/s41565-023-01359-6 · Cited by 96
Recent applications of EEG-based brain-computer-interface in the medical field
Liu, X., Wang, W.-L., Liu, M., Chen, M., Pereira, T., Doda, D. Y., Ke, Y., Wang, S., Dong, W., Tong, X., et al. (2025) · Military Medical Research · doi:10.1186/s40779-025-00598-z · Cited by 171
“However, preliminary research suggests that BCIs may offer a promising avenue for treating blindness [309].”
Self-citationTactile edges and motion via patterned microstimulation of the human somatosensory cortex
Valle, G., Alamri, A. H., Downey, J. E., Lienkämper, R., Jordan, P. M., Sobinov, A. R., Endsley, L. J., Prasad, D., Boninger, M. L., Collinger, J. L., et al. (2025) · Science · doi:10.1126/science.adq5978 · Cited by 71
Advances in visual prostheses: engineering and biological challenges
Borda, E., & Ghezzi, D. (2022) · Progress in Biomedical Engineering · doi:10.1088/2516-1091/ac812c · Cited by 44
“The CORTIVIS (Miguel Hernández University of Elche) uses a Utah array with 100 electrodes [107].”
MethodBio-Inspired Vision and Neuromorphic Image Processing Using Printable Metal Oxide Photonic Synapses
Sun, L., Qu, S., Du, Y., Lu, Y., Li, Y., Wang, Z., & Xu, W. (2022) · ACS Photonics · doi:10.1021/acsphotonics.2c01583 · Cited by 47
van der Grinten, M., de Ruyter van Steveninck, J., Lozano, A., Pijnacker, L., Rückauer, B., Roelfsema, P., van Gerven, M., van Wezel, R., Güçlü, U., & Güçlütürk, Y. (2022) · doi:10.1101/2022.12.23.521749 · Cited by 39
Self-citationvan der Grinten, M., de Ruyter van Steveninck, J., Lozano, A., Pijnacker, L., Rueckauer, B., Roelfsema, P., van Gerven, M., van Wezel, R., Güçlü, U., & Güçlütürk, Y. (2024) · eLife · doi:10.7554/elife.85812 · Cited by 39
“Our simulator uses a thresholding model based on psychometric data from Fernández et al., 2021 .”
BackgroundMethodInfluentialSelf-citationTowards a Smart Bionic Eye : AI-powered artificial vision for the treatment of incurable blindness
Beyeler, M., & Sanchez-Garcia, M. (2022) · Journal of Neural Engineering · doi:10.1088/1741-2552/aca69d · Cited by 35
“…current prosthesis users report seeing highly distorted phosphenes, which vary in shape across subjects as well as electrodes and often fail to assemble into more complex percepts (Wilke et al 2011, Beyeler et al 2019, Beauchamp et al 2020, Erickson-Davis and Korzybska 2021, Fernández et al 2021).”
Chronic stability of a neuroprosthesis comprising multiple adjacent Utah arrays in monkeys
Chen, X., Wang, F., Kooijmans, R. N., Klink, P. C., Boehler, C., Asplund, M., & Roelfsema, P. R. (2023) · Journal of Neural Engineering · doi:10.1088/1741-2552/ace07e · Cited by 49
“Recently, stimulation of the visual cortex via a 96-channel Utah array allowed a blind volunteer to perceive simple shapes [19] within a small region of the visual field (less than 4 degrees of visual angle (dva)).”
BackgroundInfluentialSelf-citationNeural electrodes for brain‐computer interface system: From rigid to soft
Yang, D., Tian, G., Chen, J., Liu, Y., Fatima, E., Qiu, J., Malek, N. A. N. N., & Qi, D. (2025) · BMEMat · doi:10.1002/bmm2.12130 · Cited by 41
Percepts evoked by multi-electrode stimulation of human visual cortex
Bosking, W. H., Oswalt, D. N., Foster, B. L., Sun, P., Beauchamp, M. S., & Yoshor, D. (2022) · Brain stimulation · doi:10.1016/j.brs.2022.08.007 · Cited by 34
“Similar levels of pattern discrimination performance have been obtained in recent 637 experiments using penetrating electrodes [21,22].”
BackgroundMethodResultInfluentialHughes, C. L., & Kozai, T. D. Y. (2023) · Brain stimulation · doi:10.1016/j.brs.2023.05.013 · Cited by 31
“For vision, ICMS has consisted of short stimulus trains at high frequencies (50 pulses at 300 Hz) [18,42].”
BackgroundMethodEisenkolb, V. M., Held, L. M., Utzschmid, A., Lin, X., Krieg, S. M., Meyer, B., Gempt, J., & Jacob, S. N. (2023) · Cell Reports · doi:10.1016/j.celrep.2023.112467 · Cited by 32
Flexible Polymer Electrodes for Stable Prosthetic Visual Perception in Mice
Orlemann, C., Boehler, C., Kooijmans, R. N., Li, B., Asplund, M., & Roelfsema, P. R. (2024) · Advanced Healthcare Materials · doi:10.1002/adhm.202304169 · Cited by 27
Self-citationBjånes, D. A., Kellis, S., Nickl, R., Baker, B., Aflalo, T., Bashford, L., Chivukula, S., Fifer, M. S., Osborn, L. E., Christie, B., et al. (2025) · Acta Biomaterialia · doi:10.1016/j.actbio.2025.02.030 · Cited by 43
Liu, X., Chen, P., Ding, X., Liu, A., Li, P., Sun, C., & Guan, H. (2022) · Annals of Translational Medicine · doi:10.21037/atm-22-2858 · Cited by 38
“gov identifier NCT02983370) implanted a 96-electrode intracortical microelectrode array in the visual cortex of a 57-year-old man who was completely blind for 6 months (73).”
BackgroundIn vivo microelectrode arrays for neuroscience
Williams, N. P., Vöröslakos, M., Shi, D., Pwint, M. Y., Lanzio, V., Mao, H., Zolotavin, P., Yoon, E., Stieglitz, T., Xie, C., et al. (2025) · Nature Reviews Methods Primers · doi:10.1038/s43586-025-00399-7 · Cited by 39
“Similar principles apply to eliciting percepts in other sensory domains 176 .”
Implantable neural electrodes: from preparation optimization to application
Zhang, B., Xie, R., Jiang, J., Hao, S., Fang, B., Zhang, J., Bai, H., Peng, B., Li, L., Liu, Z., et al. (2023) · Journal of Materials Chemistry C · doi:10.1039/d2tc05162a · Cited by 29
Electrode sharpness and insertion speed reduce tissue damage near high-density penetrating arrays
McNamara, I., Wellman, S. M., Li, L., Eles, J. R., Savya, S. P., Sohal, H. S., Angle, M. R., & Kozai, T. D. Y. (2024) · Journal of Neural Engineering · doi:10.1088/1741-2552/ad36e1 · Cited by 25
“These technologies have enabled a deeper understanding of brain function through fundamental neuroscience research [3–6] and showcasing promising potential for the restoration of functional motor control in individuals with tetraplegia as well as sensory restoration [7–18].”
BackgroundVision Restoration by Optogenetic Therapy and Developments Toward Sonogenetic Therapy
Provansal, M., Marazova, K., Sahel, J., & Picaud, S. A. (2022) · Translational Vision Science & Technology · doi:10.1167/tvst.11.1.18 · Cited by 26
Wu, G. K., Ardeshirpour, Y., Mastracchio, C., Kent, J., Caiola, M. J., & Ye, M. (2023) · iScience · doi:10.1016/j.isci.2023.108140 · Cited by 24
“4,9,10,31 A recent study also implied that different patterns of stimulation pulses may be needed for different therapeutic applications.”
BackgroundPatterned electrical brain stimulation by a wireless network of implantable microdevices
Lee, A., Lee, J.-H., Leung, V., Larson, L. E., & Nurmikko, A. V. (2024) · Nature Communications · doi:10.1038/s41467-024-54542-1 · Cited by 31
“Since the period of active current injection in most neuromodulation schemes is relatively short, on the order of 100 µ s 26,27 , the wireless RF energy transfer can likewise be kept short to reduce the average RF power low.”
Thin flexible arrays for long-term multi-electrode recordings in macaque primary visual cortex
Merken, L., Schelles, M., Ceyssens, F., Kraft, M. D., & Janssen, P. (2022) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac98e2 · Cited by 21
“Electrical microstimulation of primary visual cortex by means of multiple Utah arrays shows promising results for restoring vision [30, 69, 70], but the human primary visual cortex is largely confined to the mesial surface of the brain around the calcarine sulcus, which cannot be reached by Utah arrays [71].”
BackgroundFine, I., & Boynton, G. M. (2024) · Scientific Reports · doi:10.1038/s41598-024-65337-1 · Cited by 19
“(1234567890)”
The risks and challenges of neurotechnologies for human rights
M, S. N., & C, G. (2023) · UNESCO eBooks · doi:10.54678/pogs7778 · Cited by 44
Hughes, C. L., Stieger, K. C., Chen, K., Vazquez, A. L., & Kozai, T. D. Y. (2025) · iScience · doi:10.1016/j.isci.2025.112707 · Cited by 24
Nanotechnology for vision restoration
Lanzani, G., Chiaravalli, G., Colombo, E., Manfredi, G., Marco, S. D., Vurro, V., & Benfenati, F. (2024) · Nature Reviews Bioengineering · doi:10.1038/s44222-024-00210-4 · Cited by 19
“Nevertheless, chronic intracortical microstimulation using deep microelectrode arrays in the primary visual cortex has shown potential for restoring functional vision in macaques and humans with blindness 82–84 .”
Effects of central nervous system electrical stimulation on non-neuronal cells
Williams, N. P., Kushwah, N. S., Dhawan, V., Zheng, X. S., & Cui, X. T. (2022) · Frontiers in Neuroscience · doi:10.3389/fnins.2022.967491 · Cited by 20
Attitudes of potential recipients toward emerging visual prosthesis technologies
Karadima, V., Pezaris, E., & Pezaris, J. S. (2023) · Scientific Reports · doi:10.1038/s41598-023-36913-8 · Cited by 22
“Nevertheless, significant work continues in both pre-clinical 16 and clinical levels 66,67 .”
Toward a personalized closed-loop stimulation of the visual cortex: Advances and challenges
Grani, F., Soto-Sánchez, C., Fimia, A., & Fernández, E. (2022) · Frontiers in Cellular Neuroscience · doi:10.3389/fncel.2022.1034270 · Cited by 16
“…be good targets (Nguyen et al., 2016; Gaillet et al., 2020; Borda et al., 2022; Rassia et al., 2022), several groups are trying to develop visual prostheses designed to directly stimulate the visual cortex (Lee et al., 2016; Beauchamp et al., 2020; Fernandez et al., 2021; Bosking et al., 2022).”
BackgroundSelf-citationSequence of visual cortex stimulation affects phosphene brightness in blind subjects
Salas, M. A., Bell, J. S., Niketeghad, S., Oswalt, D. N., Bosking, W. H., Patel, U. K., Dorn, J. D., Yoshor, D., Greenberg, R. J., Bari, A., et al. (2022) · Brain stimulation · doi:10.1016/j.brs.2022.03.008 · Cited by 19
“A number of these groups are currently testing the safety of their systems as a long-term VCP and evaluating their functionality inside and outside clinical settings [1,6].”
BackgroundFeyerick, M., & Dehaene, W. (2023) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2023.3287294 · Cited by 22
“4 µ A have been reported, with typical T φ 1 in the 100 µ s to 800 µ s range [34], [35], [36].”
BackgroundMethodAligning Visual Prosthetic Development With Implantee Needs
Nadolskis, L., Turkstra, L. M., Larnyo, E., & Beyeler, M. (2024) · Translational Vision Science & Technology · doi:10.1167/tvst.13.11.28 · Cited by 19
Smith, T. J., Wu, Y., Cheon, C., Khan, A. A., Srinivasan, H., Capadona, J. R., Cogan, S. F., Pancrazio, J. J., Engineer, C. T., & Hernandez-Reynoso, A. G. (2023) · Frontiers in Neuroscience · doi:10.3389/fnins.2023.1202258 · Cited by 14
Axonal stimulation affects the linear summation of single-point perception in three Argus II users
Hou, Y., Nanduri, D., Granley, J., Weiland, J. D., & Beyeler, M. (2024) · Journal of Neural Engineering · doi:10.1088/1741-2552/ad31c4 · Cited by 14
“Although a recent study with intra-cortical electrodes (Chen et al 2020) showed that macaques could successfully identify the intended shape of a patterned electrical stimulus, human participants implanted with the same technology could not always do that (Fernández et al 2021).”
Barbruni, G. L., Rodino, F., Ros, P. M., Demarchi, D., Ghezzi, D., & Carrara, S. (2024) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2024.3357626 · Cited by 17
“This power is enough for [34].”
MethodInfluentialA novel interface for cortical columnar neuromodulation with multipoint infrared neural stimulation
Tian, F., Zhang, Y., Schriver, K. E., Hu, J. M., & Roe, A. W. (2024) · Nature Communications · doi:10.1038/s41467-024-50375-0 · Cited by 21
“These non-topographic projections also project in a bilaminar pattern avoiding layer 4, supporting an interaction that generates higher order response 40 .”
InfluentialGrani, F., Soto‐Sánchez, C., Farfán, F. D., Alfaro, A., Grima, M. D., Doblado, A. R., & Fernández, E. (2022) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac801d · Cited by 15
“The methods are described in greater detail in a previous paper [30] and are only summarized here.”
InfluentialSelf-citationSuematsu, N., Vazquez, A. L., & Kozai, T. D. Y. (2024) · Journal of Neural Engineering · doi:10.1088/1741-2552/ad3853 · Cited by 18
Continuity within the somatosensory cortical map facilitates learning
Lassagne, H., Goueytes, D., Shulz, D. E., Estebanez, L., & Ego‐Stengel, V. (2022) · Cell Reports · doi:10.1016/j.celrep.2022.110617 · Cited by 15
“, 2015), as well as to visual inputs (Dobelle, 2000; Fernández et al., 2021).”
BackgroundThe potential power of Neuralink – how brain-machine interfaces can revolutionize medicine
Kumar, R., Waisberg, E., Ong, J., & Lee, A. G. (2025) · Expert Review of Medical Devices · doi:10.1080/17434440.2025.2498457 · Cited by 22
“Other companies, such as Precision Neuroscience with its epicortical neural interface (Layer 7 Cortical Interface), have developed alternative approaches, and Blackrock Microsystems have made significant strides in brain–computer interface (BCI) technology as well [7,8].”
InfluentialThorn, J. T., Chenais, N. A. L., Hinrichs, S., Chatelain, M., & Ghezzi, D. (2022) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac5a5c · Cited by 11
Electrode Arrays for Detecting and Modulating Deep Brain Neural Information in Primates: A Review
Zhang, S., Song, Y., Lv, S., Jing, L., Wang, M., Liu, Y., Xu, W., Jiao, P., Zhang, S., Wang, M., et al. (2025) · Cyborg and Bionic Systems · doi:10.34133/cbsystems.0249 · Cited by 16
Optogenetic activation of visual thalamus generates artificial visual percepts
Wang, J., Azimi, H., Zhao, Y., Kaeser, M., Sánchez, P. V., Vázquez‐Guardado, A., Rogers, J. A., Harvey, M., & Rainer, G. (2023) · eLife · doi:10.7554/elife.90431 · Cited by 12
Comparison of electrical microstimulation artifact removal methods for high-channel-count prostheses
Wang, F., Chen, X., & Roelfsema, P. R. (2024) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2024.110169 · Cited by 11
“…the video feed into instructions for electrical stimulation, and a stimulator that delivers electrical pulses to electrodes in brain structures that process visual information (Barry et al., 2023; Dobelle et al., 1974, 1976; Fern ´ andez et al., 2021; Humayun et al., 1999; Salas et al., 2022) (Fig.”
BackgroundMethodInfluentialSelf-citationArtificial intelligence-powered smart vision glasses for the visually impaired
Udayakumar, D., Gopalakrishnan, S., Raghuram, A., Kartha, A., Krishnan, A. K., Ramamirtham, R., Muthangi, R., & Raju, R. (2025) · Indian Journal of Ophthalmology · doi:10.4103/ijo.ijo_1621_24 · Cited by 13
“[28] 29–31] Argus II provides a rudimentary vision for ULV/blindness, but costs are prohibitively high and unavailable in India.”
Losing, preserving, and restoring vision from neurodegeneration in the eye
Kerschensteiner, D. (2023) · Current Biology · doi:10.1016/j.cub.2023.08.044 · Cited by 12
“Using modern brain-computer interfaces (e.g., Utah arrays), efforts are underway to dynamically steer stimuli to shape phosphenes into useful representations, with early successes enabling blind patients to recognize letters 310,311 .”
BackgroundSequential epiretinal stimulation improves discrimination in simple shape discrimination tasks only
Christie, B., Sadeghi, R., Kartha, A., Caspi, A., Tenore, F. V., Klatzky, R. L., Dagnelie, G., & Billings, S. (2022) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac7326 · Cited by 10
Clinical Progress and Optimization of Information Processing in Artificial Visual Prostheses
Wang, J., Zhao, R., Li, P., Fang, Z., Li, Q., Han, Y., Zhou, R., & Zhang, Y. (2022) · Sensors · doi:10.3390/s22176544 · Cited by 12
“Schematic diagram of poor phosphenes induced by CORTIVIS electrode implantati n. (A) Simultaneous stimulatio of four electr des arrang d in a square may prod ce the perception; (B) Immediately after implantation, the induced phosphenes may cause poor perception of objects, such as the letter “E” in the figure.”
MethodInfluentialThe role of the visual field size in artificial vision
Ghezzi, D. (2023) · Journal of Neural Engineering · doi:10.1088/1741-2552/acc7cd · Cited by 14
“Technologically, increasing the VF size implies developing scalable electrode arrays with a substantially higher number of channels than what is available today [15, 57].”
Grani, F., Soto‐Sánchez, C., Doblado, A. R., López-Peco, R., González-López, P., & Fernández, E. (2025) · Science Advances · doi:10.1126/sciadv.adv8846 · Cited by 12
“Previous works in different animal models ( 5 – 14 ) and humans ( 15 – 25 ) have shown that microelectrodes implanted in the visual cortex allow the recognition of simple shapes, motion, and letters.”
InfluentialSelf-citationChronically implantable μLED arrays for optogenetic cortical surface stimulation in mice
Greer, R. J., Verdier, A., Butt, E., Cheng, Y., Callas, E., McAlinden, N., Aniorte, A., Kakaouia, E. M., Pereyra, M., Dawson, M. D., et al. (2026) · Nature Communications · doi:10.1038/s41467-025-68191-5 · Cited by 12
“Transforma-tive experiments in humans have restored lost sensation through intracortical electrical stimulation 2,46 .”
Raising the stakes for cortical visual prostheses
Beauchamp, M. S., Bosking, W. H., Oswalt, D. N., & Yoshor, D. (2021) · Journal of Clinical Investigation · doi:10.1172/jci154983 · Cited by 10
Rocca, A., Lehner, C., Wekesa, E., Luna, E., Fernández-Cornejo, V., Abarca-Olivas, J., Soto‐Sánchez, C., Fernández-Jover, E., & González-López, P. (2023) · Journal of neurosurgery · doi:10.3171/2023.8.jns23772 · Cited by 11
Self-citationde Ruyter van Steveninck, J., Nipshagen, M., van Gerven, M., Güçlü, U., Güçlüturk, Y., & van Wezel, R. (2024) · Journal of Neural Engineering · doi:10.1088/1741-2552/ad357d · Cited by 8
“Even given the rapid recent developments (for instance, see [7, 8]) it is clear that the quality of the prosthetic vision will be relatively elementary compared to natural vision.”
BackgroundAn actor-model framework for visual sensory encoding
Leong, F., Rahmani, B., Psaltis, D., Moser, C., & Ghezzi, D. (2024) · Nature Communications · doi:10.1038/s41467-024-45105-5 · Cited by 8
“Arti fi cial visual encoding converts high resolution images captured by an external camera into a spatiotemporal pattern of arti-fi cial stimuli delivered by a retinal 9 – 12 , optic nerve 13,14 , or cortical prosthesis 15,16 .”
Changes Over Time in the Electrode/Brain Interface Impedance: An Ex-Vivo Study
Iannucci, L., Barbruni, G. L., Ghezzi, D., Parvis, M., Grassini, S., & Carrara, S. (2023) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2023.3284691 · Cited by 10
A 128-channel neural stimulation and recording ASIC for scalable cortical visual prosthesis
Raducanu, B. C., Aymerich, J., Hsu, W.-Y., Hendrickx, P., & López, C. M. (2023) · European Solid-State Circuits Conference (ESSCIRC) · doi:10.1109/esscirc59616.2023.10268749 · Cited by 8
“Damages to the eye and top layers of the retina may be alleviated by using retinal prostheses [1]–[4], which stimulate the retina to create a perception of vision [5], [6].”
BackgroundInfluentialSharif, N. A. (2022) · OBM Neurobiology · doi:10.21926/obm.neurobiol.2201116 · Cited by 10
Measuring visual information gathering in individuals with ultra low vision using virtual reality
Kartha, A., Sadeghi, R., Bradley, C., Tran, C. T. T., Gee, W., & Dagnelie, G. (2023) · Scientific Reports · doi:10.1038/s41598-023-30249-z · Cited by 9
Neural mechanisms underlying intracortical microstimulation for sensory restoration
Hughes, C. L., Chen, X., Grill, W. M., & Kozai, T. D. Y. (2026) · Nature Biomedical Engineering · doi:10.1038/s41551-025-01583-6 · Cited by 14
“Previous clinical research predominantly focused on single-electrode ICMS 12,13,16,21,22,25 , which drives more intense and focal activation than physiological sensory input 119 .”
InfluentialSelf-citationCortical layering disrupts multi-electrode current steering
Meikle, S. J., Hagan, M. A., Price, N. S. C., & Wong, Y. T. (2023) · Journal of Neural Engineering · doi:10.1088/1741-2552/acdaf7 · Cited by 8
Halfmann, C., Rüland, T., Müller, F. U., Jehasse, K., & Kampa, B. M. (2023) · Frontiers in Cellular Neuroscience · doi:10.3389/fncel.2023.1258773 · Cited by 6
“, 2021) also provide a promising approach for evoking visual percepts by electrical stimulation of the visual cortex (Beauchamp et al., 2020; Roelfsema, 2020; Fernández et al., 2021).”
Learning in a closed-loop brain-machine interface with distributed optogenetic cortical feedback
Goueytes, D., Lassagne, H., Shulz, D. E., Ego‐Stengel, V., & Estebanez, L. (2022) · Journal of Neural Engineering · doi:10.1088/1741-2552/acab87 · Cited by 9
“…distributed neuronal activations possible, by using sophisticated multichannel electrical microstimula-tions (Dadarlat et al 2015, Weiss et al 2019, Fernández et al 2021, Flesher et al 2021) or by harnessing spatiotemporally patterned optogenetic stimulation of the cortex (Abbasi et al 2018,…”
Barbruni, G. L., Ghezzi, D., & Carrara, S. (2023) · IEEE Solid-State Circuits Magazine · doi:10.1109/mssc.2023.3246004 · Cited by 7
“200 A n Current amplitudes in these ranges are generally suitable for intracortical stimulation in many neural applications [6], [7].”
BackgroundPreszler, C., Stieger, K. C., Chen, K., Zhang, G., & Kozai, T. D. Y. (2026) · Journal of Neural Engineering · doi:10.1088/1741-2552/ae4652 · Cited by 9
Optimal placement of high-channel visual prostheses in human retinotopic visual cortex
van Hoof, R., Lozano, A., Wang, F., Klink, P. C., Roelfsema, P. R., & Goebel, R. (2025) · Journal of Neural Engineering · doi:10.1088/1741-2552/adaeef · Cited by 6
“Visual prostheses that interface directly with the brain may one day be a realistic treat-ment option for blindness that results from damage to the early visual pathways (Normann et al 2009, Roelfsema 2020, Rosenfeld et al 2020, Fernández et al 2021).”
Self-citationSaalmann, Y. B., Mofakham, S., Mikell, C. B., & Djurić, P. M. (2022) · Current Research in Neurobiology · doi:10.1016/j.crneur.2022.100071 · Cited by 7
“Stimulation across pairs of electrodes induced phosphenes in humans (Fernández et al., 2021).”
BackgroundBayesian optimization of cortical neuroprosthetic vision using perceptual feedback
Küçükoğlu, B., Soo, L., Leeftink, D., Grani, F., Soto‐Sánchez, C., Güçlü, U., van Gerven, M. A. J., & Fernández, E. (2025) · Journal of Neural Engineering · doi:10.1088/1741-2552/adeae9 · Cited by 5
“While higher currents tend to increase the perceived brightness ratings for phosphenes, highest brightness ratings are not reached by the largest currents tested (Fernandez et al 2021).”
InfluentialSelf-citationTowards precise synthetic neural codes: high-dimensional stimulation with flexible electrodes
Kim, R., Liu, Y., Zhang, J., Xie, C., & Luan, L. (2025) · npj Flexible Electronics · doi:10.1038/s41528-025-00447-y · Cited by 7
“Such intracortical microstimulation (ICMS) has demonstrated remarkable selectivity, evoking sensations in individual fi ngertips and generating precise letters and shapes in the visual fi eld 42,59 .”
Bjånes, D. A., Bashford, L., Pejsa, K., Lee, B., Liu, C. Y., & Andersen, R. A. (2025) · Journal of Neural Engineering · doi:10.1088/1741-2552/ae1bd8 · Cited by 10
“Electrical intra-cortical micro-stimulation (ICMS) patterns have evoked both somatosensory percepts in laboratory animals [28–35], human models [10, 11, 36] and visual percepts in animal [37–39] and human models [40].”
Chung, S.-H., Kim, C., Kim, M., Choi, D., Son, Y., Roh, H., Lee, B. C., Lee, H., Kim, Y., Lee, S.-K., et al. (2025) · Microsystems & Nanoengineering · doi:10.1038/s41378-025-01093-0 · Cited by 6
“Clinical applications include intracortical neural interfaces for functional restoration 1,2 , cochlear implants for hearing rehabilitation 3,4 , and retinal prostheses that restore visual perception in patients with photoreceptor degeneration 5,6 .”
Farfán, F. D., Soo, L., Grani, F., Murcia, M. D. G., & Fernández, E. (2025) · Cerebral Cortex · doi:10.1093/cercor/bhaf075 · Cited by 6
Self-citationEnd-to-end learning of safe stimulation parameters for cortical neuroprosthetic vision
Küçükoğlu, B., Rueckauer, B., van Steveninck, J. D. R., van der Grinten, M., Güçlütürk, Y., Roelfsema, P. R., Güçlü, U., & van Gerven, M. A. J. (2025) · Journal of Neural Engineering · doi:10.1088/1741-2552/ade918 · Cited by 7
“The resulting current going beyond the order of milliamperes can cause seizures upon simultaneous stimulation of multiple electrodes (Fernandez et al 2021), whereas current on the order of microamperes suffices to evoke phosphenes with penetrating electrodes (Bak et al 1990, Beauchamp et al 2020).”
InfluentialSelf-citationDecoding electroencephalographic responses to visual stimuli compatible with electrical stimulation
Romeni, S., Toni, L., Artoni, F., & Micera, S. (2024) · APL Bioengineering · doi:10.1063/5.0195680 · Cited by 5
Tackling visual impairment: emerging avenues in ophthalmology
Lin, F., Su, Y., Zhao, C., Akter, F., Yao, S., Huang, S., Shao, X., & Yao, Y. (2025) · Frontiers in Medicine · doi:10.3389/fmed.2025.1567159 · Cited by 10
“If successful, it could provide a new treatment option for patients with conditions like retinitis pigmentosa or optic nerve atrophy, where traditional treatments are ineffective (208).”
Evaluation of Pneumatic Insertion Stability of Utah Slanted Electrode Arrays in Rat Sciatic Nerve
Thomas, W. M., Leber, M., Crew, J., & Warren, D. J. (2022) · 2022 44th Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC) · doi:10.1109/embc48229.2022.9871237 · Cited by 5
“In the central nervous system, UEAs have been used for motor decode tasks [2], [3], brain mapping [4], and direct stimulation and recording of primary sensory areas of cortex directed towards sensory restoration [5], [6].”
Oz, R., Edelman-Klapper, H., Margalit, S. N., & Slovin, H. (2022) · Cerebral Cortex · doi:10.1093/cercor/bhac409 · Cited by 6
Schrimpf, M., McGrath, P., Margalit, E., & DiCarlo, J. J. (2024) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2024.01.09.572970 · Cited by 15
Intracortical microstimulation in humans: a decade of safety and efficacyPreprint
Greenspon, C. M., Hobbs, T. G., Verbaarschot, C., Alamri, A. H., Shelchkova, N. D., Lienkämper, R., Ye, J., Simpson, T. W., Weiss, J. M., Weir, D. M., et al. (2025) · medRxiv · doi:10.1101/2025.08.11.25332271 · Cited by 5
“Beyond somatosensation, ICMS may also restore other lost senses, such as vision for patients with retinal or subcortical visual impairments 10,11 .”
Hybrid neural autoencoders for stimulus encoding in visual and other sensory neuroprostheses
Granley, J., Relic, L., & Beyeler, M. (2022) · Advances in Neural Information Processing Systems 35 · doi:10.52202/068431-1647 · Cited by 5
“In such devices, diminished sensory modalities (e.g., hearing [3], vision [4, 5], cutaneous touch [6]) are re-enacted through streams of artificial input to the nervous system.”
Neural–Computer Interfaces: Theory, Practice, Perspectives
Dubynin, I. A., Zemlyanskov, M. S., Shalayeva, I., Gorskii, O., Grinevich, V. P., & Евгеньевич, М. П. (2025) · Applied Sciences · doi:10.3390/app15168900 · Cited by 7
A mesoscale optogenetics system for precise and robust stimulation of the primate cortex
Li, H., Lu, D., Liao, H., Zhang, J., Tang, J., He, F., Li, L., Zhang, Y., Leng, X., Li, C., et al. (2026) · Neuron · doi:10.1016/j.neuron.2025.12.001 · Cited by 7
Alfaro, A., Soo, L., Wacławczyk, D., Ruiz, R. M., Grani, F., & Fernandez, E. (2026) · Brain Communications · doi:10.1093/braincomms/fcaf504 · Cited by 5
“Since we were aware of the possible increase of spontaneous positive visual phenomena in the first weeks after the implantation of penetrating microelectrodes in the human visual cortex, 8 we explained this to the subject and attempted to reassure him.”
InfluentialSelf-citationNeurosurgery and the Brain-Computer Interface
Rosenfeld, J. V. (2024) · Advances in experimental medicine and biology · doi:10.1007/978-3-031-64892-2_32 · Cited by 8
Integrating Eye-Tracking With Cortical Visual Prostheses in Patients Without Eyes: A Case Study
Wacławczyk, D., Soo, L., Ruiz, R. M., Caspi, A., & Fernández, E. (2025) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2025.3615286 · Cited by 5
“Theexperimental setup was designed to replicate previously published approaches, including lighting conditions, subject preparation, and software used for data acquisition [28], [30], [57], [59].”
Self-citationPaez-Escamilla, M. A., Alabek, M. L., Beale, O., Prensky, C. J., Lejoyeux, R., Friberg, T. R., Sahel, J., & Rosin, B. (2023) · Diagnostics · doi:10.3390/diagnostics13233521 · Cited by 5
Haq, W., Basavaraju, S., Speck, A., & Zrenner, E. (2022) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac8ad0 · Cited by 5
“In addition, addressing the vision processing centers directly, the concept of cortical implants are investigated (Lowery et al 2015) and their promising potential has been investigated through several ongoing clinical trials (Pouratian et al 2019, Fernandez et al 2021).”
Multichannel stimulation module as a tool for animal studies on cortical neural prostheses
Hayashida, Y., Kameda, S., Umehira, Y., Ishikawa, S., & Yagi, T. (2022) · Frontiers in Medical Technology · doi:10.3389/fmedt.2022.927581 · Cited by 5
“After decades from the detailed clinical experiments on the intracortical visual prosthesis (21), a recent clinical trial verified the feasibility of this type of visual prosthesis (22, 23).”
BackgroundUse-Dependent Inactivation of the Mouse Visual Cortex in Response to Microstimulation Train
Fukuda, S., Yoshida, K., Nomoto, T., Yagi, T., & Hayashida, Y. (2022) · E-Health and Bioengineering Conference (EHB) · doi:10.1109/ehb55594.2022.9991319 · Cited by 4
Human-in-the-Loop Optimization for Deep Stimulus Encoding in Visual Prostheses
Granley, J., Fauvel, T., Chalk, M., & Beyeler, M. (2023) · PubMed · doi:10.48550/arxiv.2306.13104 · Cited by 22
“For instance, visual prostheses have the potential to restore vision to people living with incurable blindness by bypassing damaged parts of the visual system and directly stimulating the remaining cells in order to evoke visual percepts (phosphenes) [3–6].”
BackgroundPeiroten, L., Zrenner, E., & Haq, W. (2023) · Bioengineering · doi:10.3390/bioengineering10101208 · Cited by 6
Neural activity shaping in visual prostheses with deep learning
Castro, D., Grayden, D. B., Meffin, H., & Spencer, M. (2024) · Journal of Neural Engineering · doi:10.1088/1741-2552/ad6186 · Cited by 6
“In this study, we consider a square MEA with an 8-by-8 matrix of electrodes which is within the range of existing retinal implants (Luo and Da Cruz 2016, Rosenfeld et al 2020, Fernández et al 2021, Petoe et al 2021).”
TACTILE EDGES AND MOTION VIA PATTERNED MICROSTIMULATION OF THE HUMAN CORTEXPreprint
Valle, G., Alamari, A. H., Lienkämper, R., Downey, J. E., Sobinov, A. R., Endsley, L. J., Prasad, D., Boninger, M. L., Collinger, J. L., Warnke, P. C., et al. (2024) · medRxiv · doi:10.1101/2024.05.06.24306627 · Cited by 4
Neural Activity Shaping in Visual Prostheses with Deep Learning
Castro, D., Grayden, D. B., Meffin, H., & Spencer, M. (2023) · doi:10.1101/2023.12.20.572123 · Cited by 4
Metadata
- Authors
- Fernández, E., Alfaro, A., Soto-Sánchez, C., Gonzalez-Lopez, P., Lozano, A. M., Peña, S., Grima, M. D., Rodil, A., Gómez, B., Chen, X., Roelfsema, P. R., Rolston, J. D., Davis, T. S., & Normann, R. A.
- Affiliations
- Universitat de Miguel Hernández d'Elx (ES)
- Netherlands Institute for Neuroscience (NL)
- University of Utah (US)
- ORCID
- Venue
- Journal of Clinical Investigation
- Published
- 2021-12-01
- Volume
- 131
- Issue
- 23
- ISSN
- 0021-9738 (ISSN-L)
- 1558-8238
- PMC
- PMC8631600
- OpenAlex
- W3207382873
- Semantic Scholar
- 3fc2ad1299c6210f699eec7354757ffedb0bdc7a
- Cited by
- 246
- References
- 56
- Citations by year
- 2026: 63 · 2025: 69 · 2024: 31 · 2023: 42 · 2022: 40 · 2021: 1
- FWCI
- 10.33
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Journal of Clinical Investigation (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Pure Amsterdam UMC (publishedVersion, cc-by)
- KNAW Research Portal (The Royal Netherlands Academy of Arts and Sciences) (publishedVersion, other-oa)
- KNAW research portal (Royal Academy of Art and Sciences (KNAW)) (publishedVersion, other-oa)
- VU Research Portal (publishedVersion, other-oa)
- Data Archiving and Networked Services (DANS) (submittedVersion, other-oa)
- Funders
- Spanish Ministerio de Ciencia Innovación y Universidades (RTI2018-098969-B-100)
- Generalitat Valenciana (PROMETEO/2019/119)
- Bidons Egara Research Chair of the University Miguel Hernández (2019)
- John Moran Eye Center of the University of Utah (2019)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Life Sciences › Neuroscience › Cognitive Neuroscience › Tactile and Sensory Interactions
- Life Sciences › Neuroscience › Cognitive Neuroscience › EEG and Brain-Computer Interfaces
- Keywords
- cortical visual prosthesis, intracortical microelectrode arrays, microstimulation, visual percepts, phosphenes, blindness, visual cortex, stimulation threshold, vision restoration
- MeSH terms
- Blindness, Electric Stimulation, Electrodes, Implanted, Female, Humans, Microelectrodes, Middle Aged, Occipital Lobe, Optic Nerve Diseases, Phosphenes, Retina, Vision, Ocular, Visual Cortex, Visual Perception, Treatment Outcome, Visual Prosthesis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.