Continuous Glucose Monitoring: Real-Time Algorithms for Calibration, Filtering, and Alarms
Open accessThe DOI matches its Crossref record; 60 of the 60 references it cites were found in a registry or an index.
Abstract
Algorithms for real-time use in continuous glucose monitors are reviewed, including calibration, filtering of noisy signals, glucose predictions for hypoglycemic and hyperglycemic alarms, compensation for capillary blood glucose to sensor time lags, and fault detection for sensor degradation and dropouts. A tutorial on Kalman filtering for real-time estimation, prediction, and lag compensation is presented and demonstrated via simulation examples. A limited number of fault detection methods for signal degradation and dropout have been published, making that an important area for future work.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Optimal Control and Estimation
Stengel, R. F. (1994) · Cited by 1,617
Evaluating clinical accuracy of systems for self-monitoring of blood glucose
Clarke, W. L., Cox, D., Gonder-Frederick, L. A., Carter, W., & Pohl, S. L. (1987) · Diabetes Care · doi:10.2337/diacare.10.5.622 · Cited by 1,435 · cited 5×
“Gani and colleagues52 stated that a disadvantage to Kalman filtering is that it requires a high-fidelity, firstprinciples model for meals and physical activity, but it should be clear from Equations (20), (21), and (31) that this is not the case.”
BackgroundInfluentialKlonoff, D. C. (2005) · Diabetes Care · doi:10.2337/diacare.28.5.1231 · Cited by 724
Glucose sensors: a review of current and emerging technology
Oliver, N. S., Toumazou, C., Cass, A. E. G., & Johnston, D. G. (2009) · Diabetic Medicine · doi:10.1111/j.1464-5491.2008.02642.x · Cited by 584
Diabetes: Models, Signals, and Control
Cobelli, C., Dalla Man, C., Sparacino, G., Magni, L., De Nicolao, G., & Kovatchev, B. P. (2009) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2009.2036073 · Cited by 511
Rebrin, K., Steil, G. M., van Antwerp, W. P., & Mastrototaro, J. J. (1999) · American Journal of Physiology-Endocrinology and Metabolism · doi:10.1152/ajpendo.1999.277.3.e561 · Cited by 366
Factors affecting blood glucose monitoring: sources of errors in measurement
Ginsberg, B. H. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300438 · Cited by 348
Sparacino, G., Zanderigo, F., Corazza, S., Maran, A., Facchinetti, A., & Cobelli, C. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2006.889774 · Cited by 338
Bequette, B. W. (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.28 · Cited by 270
Self-citationPhysiological Differences Between Interstitial Glucose and Blood Glucose Measured in Human Subjects
Kulcu, E., Tamada, J. A., Reach, G., Potts, R. O., & Lesho, M. J. (2003) · Diabetes Care · doi:10.2337/diacare.26.8.2405 · Cited by 251
Accuracy of the 5-Day FreeStyle Navigator Continuous Glucose Monitoring System
Weinstein, R. L., Schwartz, S. L., Brazg, R. L., Bugler, J. R., Peyser, T. A., & McGarraugh, G. V. (2007) · Diabetes Care · doi:10.2337/dc06-1602 · Cited by 228
Delays in Minimally Invasive Continuous Glucose Monitoring Devices: A Review of Current Technology
Keenan, D. B., Mastrototaro, J. J., Voskanyan, G., & Steil, G. M. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300528 · Cited by 198 · cited 1×
“gk+1 = 2gk − gk−1 + wk yk = gk + vk , (29)”
Is blood glucose predictable from previous values? A solicitation for data.
Bremer, T. M., & Gough, D. A. (1999) · Diabetes · doi:10.2337/diabetes.48.3.445 · Cited by 160
Estimation of Future Glucose Concentrations with Subject-Specific Recursive Linear Models
Eren-Oruklu, M., Cinar, A., Quinn, L., & Smith, D. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0065 · Cited by 149
Real-Time Hypoglycemia Prediction Suite Using Continuous Glucose Monitoring
Dassau, E., Cameron, F., Lee, H., Bequette, B. W., Zisser, H., Jovanovič, L., Chase, H. P., Wilson, D. M., Buckingham, B. A., & Doyle, F. J. (2010) · Diabetes Care · doi:10.2337/dc09-1487 · Cited by 143
Self-citationIntroduction to Applied Statistical Signal Analysis
Sengupta, S. K., & Shiavi, R. (1994) · Technometrics · doi:10.2307/1269207 · Cited by 139 · cited 1×
“Another linear projection method (used in the voting method presented by Cameron and colleagues40 and Dassau and colleagues41 and the pump shutoff studies of Buckingham and associates42) is to consider the end points of a data window ratek = yk − yk− N NΔt (15)”
MethodPredictive Monitoring for Improved Management of Glucose Levels
Reifman, J., Rajaraman, S., Gribok, A. V., & Ward, W. K. (2007) · Journal of Diabetes Science and Technology · doi:10.1177/193229680700100405 · Cited by 135
Preventing Hypoglycemia Using Predictive Alarm Algorithms and Insulin Pump Suspension
Buckingham, B., Cobry, E., Clinton, P., Gage, V., Caswell, K., Kunselman, E., Cameron, F., & Chase, H. P. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0032 · Cited by 132
Predicting Subcutaneous Glucose Concentration in Humans: Data-Driven Glucose Modeling
Gani, A., Gribok, A. V., Rajaraman, S., Ward, W. K., & Reifman, J. (2009) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2008.2005937 · Cited by 129
Universal Glucose Models for Predicting Subcutaneous Glucose Concentration in Humans
Gani, A., Gribok, A. V., Lu, Y., Ward, W. K., Vigersky, R. A., & Reifman, J. (2010) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2009.2034141 · Cited by 124
Development of a Neural Network for Prediction of Glucose Concentration in Type 1 Diabetes Patients
Pappada, S. M., Cameron, B. D., & Rosman, P. M. (2008) · Journal of Diabetes Science and Technology · doi:10.1177/193229680800200507 · Cited by 124
Schmidtke, D. W., Freeland, A. C., Heller, A., & Bonnecaze, R. T. (1998) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.95.1.294 · Cited by 118
JDRF CGM Study Group (2008) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2007.0302 · Cited by 116
Evaluation of Factors Affecting CGMS Calibration
Diabetes Research In Children Network (Direcnet) Study Group (2006) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2006.8.318 · Cited by 116 · cited 1×
“For these simulation studies, the steady-state Kalman gain is obtained by iterating Equations (24) and (25) until they converge to constant values.”
MethodDetermination of Plasma Glucose During Rapid Glucose Excursions with a Subcutaneous Glucose Sensor
Steil, G. M., Rebrin, K., Mastrototaro, J., Bernaba, B., & Saad, M. F. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503763816436 · Cited by 104
Poitout, V., Moatti-Sirat, D., Reach, G., Zhang, Y., Wilson, G. S., Lemonnier, F., & Klein, J. C. (1993) · Diabetologia · doi:10.1007/bf00404077 · Cited by 103
The Extended Kalman Filter for Continuous Glucose Monitoring
Knobbe, E. J., & Buckingham, B. (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.15 · Cited by 95 · cited 2×
“The two- and three-state models of Equations (21) and (31) are nothing more than basic principles of differential calculus applied to glucose; that is, the two-state model related “distance” (glucose) and “velocity” (rate of change) of glucose, whereas the threestate model relates distance, velocity and “acceleration”…”
BackgroundA Review of Continuous Glucose Monitoring Technology
Klonoff, D. C. (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.770 · Cited by 95
Hypoglycemia Prediction and Detection Using Optimal Estimation
Palerm, C. C., Willis, J. P., Desemone, J., & Bequette, B. W. (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.3 · Cited by 94
Self-citationPalerm, C. C., & Bequette, B. W. (2007) · Journal of Diabetes Science and Technology · doi:10.1177/193229680700100505 · Cited by 89
Self-citationCalibration of a subcutaneous amperometric glucose sensor implanted for 7 days in diabetic patients
Choleau, C., Klein, J. C., Reach, G., Aussedat, B., Demaria-Pesce, V., Wilson, G. S., Gifford, R., & Ward, W. K. (2002) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(01)00304-9 · Cited by 76 · cited 4×
“For the two-state model in Equation (21), this is equivalent to assuming that the rate of change of glucose is constant in the future.”
BackgroundMethodInfluentialContinuous Glucose Monitoring: An Overview of Its Development
Skyler, J. S. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2009.0045 · Cited by 72
Hypoglycemia Prediction with Subject-Specific Recursive Time-Series Models
Eren-Oruklu, M., Çınar, A., & Quinn, L. T. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400104 · Cited by 70
King, C., Anderson, S. M., Breton, M., Clarke, W. L., & Kovatchev, B. P. (2007) · Journal of Diabetes Science and Technology · doi:10.1177/193229680700100302 · Cited by 66 · cited 7×
“Gani and colleagues52 stated that a disadvantage to Kalman filtering is that it requires a high-fidelity, firstprinciples model for meals and physical activity, but it should be clear from Equations (20), (21), and (31) that this is not the case.”
BackgroundMethodInfluentialCalibration of a subcutaneous amperometric glucose sensor
Choleau, C., Klein, J. C., Reach, G., Aussedat, B., Demaria-Pesce, V. H., Wilson, G. S., Gifford, R., & Ward, W. K. (2002) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(01)00306-2 · Cited by 66 · cited 14×
“Uncertainty in the future grows as [Equation (24) without the measurement feedback term] Pk = ΦPk−1Φ T + ΓwQΓwT .”
BackgroundMethodInfluentialContinuous Glucose Monitoring with Glucose Sensors: Calibration and Assessment Criteria
Lodwig, V., & Heinemann, L. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503322250596 · Cited by 63 · cited 1×
“ ŷk = −a1 ŷk−1 −− aN ŷk− N + b0 yk + b1yk−1 ++ bM yk− M, (13)”
An Online Self-Tunable Method to Denoise CGM Sensor Data
Facchinetti, A., Sparacino, G., & Cobelli, C. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2009.2033264 · Cited by 62
Artificial pancreas: an emerging approach to treat Type 1 diabetes
Kumareswaran, K., Evans, M. L., & Hovorka, R. (2009) · Expert Review of Medical Devices · doi:10.1586/erd.09.23 · Cited by 61 · cited 1×
“(3) The slope and intercept are estimated from”
BackgroundMcGarraugh, G., & Bergenstal, R. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0047 · Cited by 61 · cited 11×
“Uncertainty in the future grows as [Equation (24) without the measurement feedback term] Pk = ΦPk−1Φ T + ΓwQΓwT .”
InfluentialStatistical Hypoglycemia Prediction
Cameron, F., Niemeyer, G., Gundy-Burlet, K., & Buckingham, B. (2008) · Journal of Diabetes Science and Technology · doi:10.1177/193229680800200412 · Cited by 58
Continuous Glucose Sensing: Future Technology Developments
Brauker, J. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0137 · Cited by 58
A Dual-Rate Kalman Filter for Continuous Glucose Monitoring
Kuure-Kinsey, M., Palerm, C. C., & Bequette, B. W. (2006) · 2006 International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2006.260057 · Cited by 56 · cited 1×
“yk = a1yk−1 + a2 yk−1 + wk + c1wk−1 , (34)”
Self-citationPrevention of Hypoglycemia Using Risk Assessment With a Continuous Glucose Monitoring System
Choleau, C., Dokládal, P., Klein, J. C., Ward, W. K., Wilson, G. S., & Reach, G. (2002) · Diabetes · doi:10.2337/diabetes.51.11.3263 · Cited by 48
Zanderigo, F., Sparacino, G., Kovatchev, B. P., & Cobelli, C. (2007) · Journal of Diabetes Science and Technology · doi:10.1177/193229680700100508 · Cited by 46
Predictive monitoring for abnormal situation management
Juricek, B. C., Seborg, D. E., & Larimore, W. E. (2001) · Journal of Process Control · doi:10.1016/s0959-1524(00)00043-3 · Cited by 46
A user-friendly method for calibrating a subcutaneous glucose sensor-based hypoglycaemic alarm
Aussedat, B., Thomé-Duret, V., Reach, G., Lemmonier, F., Klein, J.-C., Hu, Y., & Wilson, G. S. (1997) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(97)00083-3 · Cited by 43 · cited 2×
“The state estimates are obtained solving Equations (22) and (23) as before.”
BackgroundSparacino, G., Facchinetti, A., Maran, A., & Cobelli, C. (2008) · Current Diabetes Reviews · doi:10.2174/157339908785294361 · Cited by 41
Voskanyan, G., Keenan, D. B., Mastrototaro, J. J., & Steil, G. M. (2007) · Journal of Diabetes Science and Technology · doi:10.1177/193229680700100507 · Cited by 37
I. Glucose control strategies for treating type 1 diabetes mellitus
Doyle, F., Jovanovic, L. G., & Seborg, D. E. (2007) · Journal of Process Control · doi:10.1016/j.jprocont.2007.01.013 · Cited by 37
Noujaim, S. E., Horwitz, D., Sharma, M., & Marhoul, J. (2007) · Journal of Diabetes Science and Technology · doi:10.1177/193229680700100509 · Cited by 29 · cited 3×
“Rebrin and colleagues65 presented an intuitive estimation approach by rearranging Equation (37) to solve for blood glucose (g) from the subcutaneous sensor signal (y):”
BackgroundMethodInfluentialUsing Support Vector Machines to Detect Therapeutically Incorrect Measurements by the MiniMed CGMS®
Bondía, J., Tarín, C., García-Gabín, W., Esteve, E., Fernández‐Real, J. M., Ricart, W., & Vehí, J. (2008) · Journal of Diabetes Science and Technology · doi:10.1177/193229680800200413 · Cited by 28
Hypoglycemia Warning Signal and Glucose Sensors: Requirements and Concepts
Heise, T., Koschinsky, T., Heinemann, L., & Lodwig, V. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503322250587 · Cited by 25
An Overview of Continuous Glucose Monitoring Systems
Cox, M. A. (2009) · Journal of Pediatric Health Care · doi:10.1016/j.pedhc.2009.06.002 · Cited by 25 · cited 1×
“ ŷk = a0 yk + a1yk−1 ++ aM yk− M , (11)”
Ward, W. K., Casey, H. M., Quinn, M. J., Federiuk, I. F., & Wood, M. D. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503322640980 · Cited by 23
Freeland, A. C., & Bonnecaze, R. T. (1999) · Annals of Biomedical Engineering · doi:10.1114/1.196 · Cited by 23
Computationally Efficient Neural Predictive Control Based on a Feedforward Architecture
Kuure-Kinsey, M., Cutright, R., & Bequette, B. W. (2006) · Industrial & Engineering Chemistry Research · doi:10.1021/ie060246y · Cited by 23
Self-citationThe Role of the Independent Variable to Glucose Sensor Calibration
Panteleon, A. E., Rebrin, K., & Steil, G. M. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503765691901 · Cited by 22 · cited 1×
“The underlying model is xk +1 = Φxk + Γ wk yk = Cxk + vk , (19)”
BackgroundClinical Overview of Continuous Glucose Monitoring
Buckingham, B. A. (2008) · Journal of Diabetes Science and Technology · doi:10.1177/193229680800200223 · Cited by 19 · cited 1×
“min m i yk− i − ŷk− i ( ) 2 i=0 N −1 ∑ , (36)”
Continuous real‐time glucose monitoring systems: time for a closer look
Kerr, D. E., & Fayers, K. (2008) · Practical Diabetes · doi:10.1002/pdi.1197 · Cited by 10
Performance Standards for Continuous Glucose Monitors
Hayter, P. G., Sharma, M., Dunka, L., Stout, P., Price, D. A., Horwitz, D. L., Marhoul, J. C., & Vaez-zadeh, S. (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.721 · Cited by 4
Cited by
Lee, I., Probst, D., Klonoff, D., & Sode, K. (2021) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2021.113054 · Cited by 288
Wearable Continuous Glucose Monitoring Sensors: A Revolution in Diabetes Treatment
Cappon, G., Acciaroli, G., Vettoretti, M., Facchinetti, A., & Sparacino, G. (2017) · Electronics · doi:10.3390/electronics6030065 · Cited by 237
“Currently available CGM systems already incorporate smart features, such as alerts and alarms generation in case of hypoglycemic or hyperglycemic events, incorporation of glucose trend information, detection of sensor faults and artifacts [55,56].”
BackgroundXie, X., Doloff, J. C., Yesilyurt, V., Sadraei, A., McGarrigle, J. J., Omami, M., Veiseh, O., Farah, S., Isa, D., Ghani, S., et al. (2018) · Nature Biomedical Engineering · doi:10.1038/s41551-018-0273-3 · Cited by 235
Challenges and recent progress in the development of a closed-loop artificial pancreas
Bequette, B. W. (2012) · Annual Reviews in Control · doi:10.1016/j.arcontrol.2012.09.007 · Cited by 183
“An overview of CGM calibration and sensor signal filtering algorithms is provided by Bequette (2010).”
MethodSelf-citationWu, Z., Qiao, Z., Chen, S., Fan, S., Liu, Y., Qi, J., & Lim, C. T. (2024) · Communications Materials · doi:10.1038/s43246-024-00468-6 · Cited by 183
Metadata
- Authors
- Bequette, B. W.
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2010-03
- Volume
- 4
- Issue
- 2
- Pages
- 404-418
- ISSN
- 1932-2968 (ISSN-L)
- OpenAlex
- W2096170052
- Cited by
- 192
- References
- 60
- FWCI
- 13.21
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.