Regenerated sciatic nerve axons stimulated through a chronically implanted macro-sieve electrode
Open accessAbstract
Sieve electrodes provide a chronic interface for stimulating peripheral nerve axons. Yet, successful utilization requires robust axonal regeneration through the implanted electrode. The present study determined the effect of large transit zones in enhancing axonal regeneration and revealed an intimate neural interface with an implanted sieve electrode. Fabrication of the polyimide sieve electrodes employed sacrificial photolithography. The manufactured macro-sieve electrode (MSE) contained nine large transit zones with areas of ~0.285 mm2 surrounded by eight Pt-Ir metallized electrode sites. Prior to implantation, saline or glial derived neurotropic factor (GDNF) was injected into nerve guidance silicone-conduits with or without a MSE. The MSE assembly or a nerve guidance conduit was implanted between transected ends of the sciatic nerve in adult male Lewis rats. At 3 months’ post-operation, fiber counts were similar through both implant types. Likewise, stimulation of nerves regenerated through a MSE or an open silicone conduit evoked comparable muscle forces. These results showed that nerve regeneration was comparable through MSE transit zones and an open conduit. GDNF had a minimal positive effect on the quality and morphology of fibers regenerating through the MSE; thus, the MSE may reduce reliance on GDNF to augment axonal regeneration. Selective stimulation of several individual muscles was achieved through monopolar stimulation of individual electrodes sites suggesting that the MSE might be an optimal platform for functional neuromuscular stimulation.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
A small amount of pores makes it possible to create relatively large transit zones, resulting in faster regeneration.
“A small amount of pores makes it possible to create relatively large transit zones, which results in faster regeneration (MacEwan et al., 2016).”
From Update on Peripheral Nerve Electrodes for Closed-Loop Neuroprosthetics (2018) · cites this paper as MacEwan et al., 2016
The Abstract describes a macro-sieve electrode with nine large transit zones, and the Introduction contrasts it with earlier electrodes containing hundreds of small holes. But the Conclusions report regeneration comparable to that through empty conduits, not faster regeneration; the cited paper does not establish that fewer pores increased the regeneration rate. No page numbers were provided in our copy. (LLM, conf=0.88)
Experimental — accuracy not yet measuredIn vivo experiments have shown fascicles regenerated through the regenerative electrode for every implant, after which stable long-term recording or stimulation up to 3 months was possible.
“However, in vivo experiments have been performed in which the fascicles regenerated through the electrode for every implant after which stable long term recording or stimulation (up to 3 months) was possible (Musick et al., 2015; MacEwan et al., 2016).”
From Update on Peripheral Nerve Electrodes for Closed-Loop Neuroprosthetics (2018) · cites this paper as MacEwan et al., 2016
The Discussion (Viability of the macro-sieve electrode design; page not provided) reports widespread nerve regeneration through the implanted electrodes and chronic stability of the assemblies. The Methods (Surgical procedures) place terminal testing at 3 months, but the Results (Electrical activation of re-innervated musculature) say only three-quarters of implanted electrodes effectively recruited muscle. The paper therefore does not establish stable recording or stimulation through every implant for the full 3 months. (LLM, conf=0.89)
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Functional Evaluation of Complete Sciatic, Peroneal, and Posterior Tibial Nerve Lesions in the Rat
Bain, J. R., Mackinnon, S. E., & Hunter, D. A. (1989) · Plastic & Reconstructive Surgery · doi:10.1097/00006534-198901000-00024 · Cited by 1,262 · see citing works
de Medinaceli, L., Freed, W. J., & Wyatt, R. J. (1982) · Experimental Neurology · doi:10.1016/0014-4886(82)90234-5 · Cited by 1,156 · see citing works
Navarro, X., Krueger, T. B., Lago, N., Micera, S., Stieglitz, T., & Dario, P. (2005) · Journal of the Peripheral Nervous System · doi:10.1111/j.1085-9489.2005.10303.x · Cited by 845 · see citing works
Development of fibrin derivatives for controlled release of heparin-binding growth factors
Sakiyama‐Elbert, S. E., & Hubbell, J. A. (2000) · Journal of Controlled Release · doi:10.1016/s0168-3659(99)00221-7 · Cited by 565 · see citing works
Sakiyama‐Elbert, S. E., & Hubbell, J. A. (2000) · Journal of Controlled Release · doi:10.1016/s0168-3659(00)00296-0 · Cited by 424 · see citing works
Functionally selective peripheral nerve stimulation with a flat interface nerve electrode
Tyler, D. J., & Durand, D. M. (2002) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2002.806840 · Cited by 398 · cited 1× · see citing works
“, 1993; Loeb and Peck, 1996), slowly penetrating interfascicular nerve electrodes (SPINEs; Tyler and Durand, 1997), and flat-interface nerve electrodes (FINEs; Tyler and Durand, 2002).”
BackgroundControlled release of nerve growth factor enhances sciatic nerve regeneration
Lee, A. C., Yu, V. M., Lowe, J. B., Brenner, M. J., Hunter, D. A., Mackinnon, S. E., & Sakiyama‐Elbert, S. E. (2003) · Experimental Neurology · doi:10.1016/s0014-4886(03)00258-9 · Cited by 369 · see citing works
Monkey median nerve repaired by nerve graft or collagen nerve guide tube
Archibald, S., Shefner, J. M., Krarup, C., & Madison, R. (1995) · Journal of Neuroscience · doi:10.1523/jneurosci.15-05-04109.1995 · Cited by 359 · see citing works
Selective Stimulation of Cat Sciatic Nerve Using an Array of Varying-Length Microelectrodes
Branner, A., Stein, R. B., & Normann, R. A. (2001) · Journal of Neurophysiology · doi:10.1152/jn.2001.85.4.1585 · Cited by 358 · see citing works
Selective control of muscle activation with a multipolar nerve cuff electrode
Veraart, C., Grill, W. M., & Mortimer, J. T. (1993) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.237694 · Cited by 320 · see citing works
Efficient testing of motor function in spinal cord injured rats
Metz, G. A. S., Merkler, D., Dietz, V., Schwab, M. E., & Fouad, K. (2000) · Brain Research · doi:10.1016/s0006-8993(00)02778-5 · Cited by 312 · see citing works
Long-term stimulation and recording with a penetrating microelectrode array in cat sciatic nerve
Branner, A., Stein, R. B., Fernandez, E., Aoyagi, Y., & Normann, R. A. (2004) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2003.820321 · Cited by 301 · see citing works
Changes in nerve fiber numbers distal to a nerve repair in the rat sciatic nerve model
Mackinnon, S. E., Dellon, A. L., & O'Brien, J. P. (1991) · Muscle & Nerve · doi:10.1002/mus.880141113 · Cited by 270 · see citing works
Outcome measures of peripheral nerve regeneration
Wood, M. D., Kemp, S. W. P., Weber, C., Borschel, G. H., & Gordon, T. (2011) · Annals of Anatomy - Anatomischer Anzeiger · doi:10.1016/j.aanat.2011.04.008 · Cited by 264 · see citing works
Cuff electrodes for chronic stimulation and recording of peripheral nerve activity
Loeb, G. E., & Peck, R. A. (1996) · Journal of Neuroscience Methods · doi:10.1016/0165-0270(95)00123-9 · Cited by 215 · see citing works
Selective stimulation of peripheral nerve fibers using dual intrafascicular electrodes
Yoshida, K., & Horch, K. (1993) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.243412 · Cited by 208 · see citing works
Binary imaging analysis for comprehensive quantitative histomorphometry of peripheral nerve
Hunter, D. A., Moradzadeh, A., Whitlock, E. L., Brenner, M. J., Myckatyn, T. M., Wei, C., Tung, T. H. H., & Mackinnon, S. E. (2007) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2007.06.018 · Cited by 190 · see citing works
Development of growth factor fusion proteins for cell‐triggered drug delivery
Sakiyama‐Elbert, S. E., Panitch, A., & Hubbell, J. A. (2001) · The FASEB Journal · doi:10.1096/fj.00-0564fje · Cited by 181 · see citing works
Histologic Assessment of Nerve Regeneration in the Rat
Mackinnon, S. E., Hudson, A. R., & Hunter, D. A. (1985) · Plastic & Reconstructive Surgery · doi:10.1097/00006534-198503000-00014 · Cited by 179 · see citing works
Technology Insight: future neuroprosthetic therapies for disorders of the nervous system
Normann, R. A. (2007) · Nature Clinical Practice Neurology · doi:10.1038/ncpneuro0556 · Cited by 167 · see citing works
Regeneration microelectrode array for peripheral nerve recording and stimulation
Kovacs, G. T. A., Storment, C. W., & Rosen, J. M. (1992) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.256422 · Cited by 164 · cited 1× · see citing works
“Prior sieve electrodes had transit zones with hundreds of small holes (10–60 μm diameter) that facilitated recording from small groups of peripheral axons (Kovacs et al., 1992).”
BackgroundWood, M. D., Moore, A. M., Hunter, D. A., TUFFAHA, S. H., Borschel, G. H., Mackinnon, S. E., & Sakiyama‐Elbert, S. E. (2008) · Acta Biomaterialia · doi:10.1016/j.actbio.2008.11.008 · Cited by 156 · see citing works
A micromachined silicon sieve electrode for nerve regeneration applications
Akin, T., Najafi, K., Smoke, R. H., & Bradley, R. M. (1994) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.284958 · Cited by 144 · see citing works
Johnson, P. J., Parker, S. R., & Sakiyama‐Elbert, S. E. (2009) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.32343 · Cited by 128 · see citing works
Stieglitz, T., Beutel, H., & Meyer, J.-U. (1997) · Sensors and Actuators A: Physical · doi:10.1016/s0924-4247(97)01494-5 · Cited by 125 · see citing works
Numbers of regenerating axons in parent and tributary peripheral nerves in the rat
Jenq, C., & Coggeshall, R. E. (1985) · Brain Research · doi:10.1016/0006-8993(85)91381-2 · Cited by 123 · see citing works
Self‐evaluation of walking‐track measurement using a sciatic function index
Brown, C. J., Mackinnon, S. E., Evans, P. J., Bain, J. R., Makino, A. P., Hunter, O. A., & Hare, G. M. T. (1989) · Microsurgery · doi:10.1002/micr.1920100317 · Cited by 109 · see citing works
A slowly penetrating interfascicular nerve electrode for selective activation of peripheral nerves
Tyler, D. J., & Durand, D. M. (1997) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.559349 · Cited by 108 · cited 1× · see citing works
“, 1993; Loeb and Peck, 1996), slowly penetrating interfascicular nerve electrodes (SPINEs; Tyler and Durand, 1997), and flat-interface nerve electrodes (FINEs; Tyler and Durand, 2002).”
Background(1998) · PubMed · Cited by 107 · see citing works
McDonnall, D., Clark, G. A., & Normann, R. A. (2004) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2004.828425 · Cited by 98 · see citing works
Fabrication and characteristics of an implantable, polymer-based, intrafascicular electrode
Lawrence, S. M., Dhillon, G. S., & Horch, K. W. (2003) · Journal of Neuroscience Methods · doi:10.1016/s0165-0270(03)00231-0 · Cited by 93 · see citing works
Lago, N., Udina, E., Ramachandran, A., & Navarro, X. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2007.891168 · Cited by 68 · see citing works
Rat sciatic nerve regeneration through a micromachined silicon chip
Zhao, Q., Drott, J., Laurell, T., Wallman, L., Lindström, K., Bjursten, L. M., Lundborg, G., Montelius, L., & Danielsen, N. (1997) · Biomaterials · doi:10.1016/s0142-9612(96)00112-3 · Cited by 67 · see citing works
Analysis of high-perimeter planar electrodes for efficient neural stimulation
Wei, X. F. (2009) · Frontiers in Neuroengineering · doi:10.3389/neuro.16.015.2009 · Cited by 61 · see citing works
Lago, N., & Navarro, X. (2006) · Journal of Neurotrauma · doi:10.1089/neu.2006.23.227 · Cited by 57 · see citing works
Selective motor unit recruitment via intrafascicular multielectrode stimulation
McDonnall, D., Clark, G. A., & Normann, R. A. (2004) · Canadian Journal of Physiology and Pharmacology · doi:10.1139/y04-047 · Cited by 53 · see citing works
The Effect of Reinnervation on Force Production and Power Output in Skeletal Muscle
Yoshimura, K., Asato, H., Cederna, P. S., Urbanchek, M. G., & Kuzon, W. M. (1999) · Journal of Surgical Research · doi:10.1006/jsre.1998.5498 · Cited by 52 · see citing works
Wallman, L. (2001) · Biomaterials · doi:10.1016/s0142-9612(00)00342-2 · Cited by 49 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Foreign Body Reaction to Implanted Biomaterials and Its Impact in Nerve Neuroprosthetics
Carnicer-Lombarte, A., Chen, S.-T., Malliaras, G. G., & Barone, D. G. (2021) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2021.622524 · Cited by 454 · see citing works
“More recently, additional features such as the incorporation of porous “transit” zones in the center of the sieve and the use of glial derived neurotrophic factor (GDNF) to improve nerve regeneration and survival have been implemented (MacEwan et al., 2016).”
BackgroundSong, J. W., Ryu, H., Bai, W., Xie, Z., Vázquez‐Guardado, A., Nandoliya, K. R., Avila, R. O., Lee, G., Song, Z. F., Kim, J., et al. (2023) · Science Advances · doi:10.1126/sciadv.ade4687 · Cited by 202 · see citing works
Wang, J., Xiong, H., Zhu, T., Liu, Y., Pan, H., Fan, C., Zhao, X., & Lu, W. W. (2020) · ACS Nano · doi:10.1021/acsnano.0c03570 · Cited by 200 · see citing works
Qian, Y., Song, J., Zhao, X., Chen, W., Ouyang, Y., Yuan, W., & Fan, C. (2018) · Advanced Science · doi:10.1002/advs.201700499 · Cited by 198 · see citing works
Interfaces with the peripheral nervous system for the control of a neuroprosthetic limb: a review
Yildiz, K. A., Shin, A. Y., & Kaufman, K. R. (2020) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-020-00667-5 · Cited by 113 · see citing works
“The electrode enabled axonal regeneration, and the stimulation of different electrode sites recruited specific muscles with a continuing improvement [192].”
BackgroundUpdate on Peripheral Nerve Electrodes for Closed-Loop Neuroprosthetics
Rijnbeek, E. H., Eleveld, N., & Olthuis, W. (2018) · Frontiers in Neuroscience · doi:10.3389/fnins.2018.00350 · Cited by 91 · see citing works
“A small amount of pores makes it possible to create relatively large transit zones, which results in faster regeneration (MacEwan et al., 2016).”
BackgroundJoshi, A., Choudhury, S., Baghel, V. S., Ghosh, S., Gupta, S., Debrupa Lahiri, A., Ananthasuresh, G. K., & Chatterjee, K. (2023) · Advanced Healthcare Materials · doi:10.1002/adhm.202300701 · Cited by 86 · see citing works
Compliant peripheral nerve interfaces.
Paggi, V., Akouissi, O., Micera, S., & Lacour, S. P. (2021) · Journal of Neural Engineering · doi:10.1088/1741-2552/abcdbe · Cited by 60 · see citing works
“Multiple configurations have been exposed in literature, using perforated arrays [15, 119, 120] and slotted structures [121].”
BackgroundCutrone, A., & Micera, S. (2019) · Advanced Healthcare Materials · doi:10.1002/adhm.201801345 · Cited by 57 · see citing works
A guide towards long-term functional electrodes interfacing neuronal tissue
Renz, A. F., Reichmuth, A. M., Stauffer, F., Thompson‐Steckel, G., & Vörös, J. (2018) · Journal of Neural Engineering · doi:10.1088/1741-2552/aae0c2 · Cited by 50 · see citing works
Rezza, A. S., Külahçı, Y., Gorantla, V. S., Zor, F., & Drzeniek, N. M. (2022) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2022.863969 · Cited by 47 · see citing works
Wang, J., Yuan, Y., Zhang, S., Lu, S., Han, G., Bian, M., Huang, L., Meng, D., Su, D., Xiao, L., et al. (2024) · Advanced Science · doi:10.1002/advs.202302988 · Cited by 43 · see citing works
Phantom limb pain: peripheral neuromodulatory and neuroprosthetic approaches to treatment
Petersen, B. A., Nanivadekar, A. C., Chandrasekaran, S., & Fisher, L. E. (2019) · Muscle & Nerve · doi:10.1002/mus.26294 · Cited by 39 · see citing works
Fascicular nerve stimulation and recording using a novel double-aisle regenerative electrode
Delgado-Martínez, I., Righi, M., Santos, D., Cutrone, A., Bossi, S., D’Amico, S., Del Valle, J., Micera, S., & Navarro, X. (2017) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa6bac · Cited by 35 · see citing works
Transient Implantable Electronics for Postsurgery Preventive Medicine
Mirzajani, H., Zolfaghari, P., Nakhjavani, S. A., Koca, B. Y., Khodapanahandeh, M., & Ürey, H. (2024) · Advanced Functional Materials · doi:10.1002/adfm.202413324 · Cited by 26 · see citing works
Wei, S., Hu, Q., Ma, J., DAI, X. H., Sun, Y., Han, G., Meng, H., Xu, W., Zhang, L., Ma, X., et al. (2022) · Bioactive Materials · doi:10.1016/j.bioactmat.2022.03.014 · Cited by 22 · see citing works
PCL NGCs integrated with urolithin-A-loaded hydrogels for nerve regeneration
Jin, X., Fang, J., Wang, J., Xu, B., Wang, X., Liu, S., Chen, F., & Liu, J. (2022) · Journal of Materials Chemistry B · doi:10.1039/d2tb01624a · Cited by 19 · see citing works
Kim, H., Dingle, A. M., Ness, J. P., Baek, D.-H., Bong, J., Lee, I. K., Shulzhenko, N. O., Zeng, W., Israel, J. S., Pisaniello, J. A., et al. (2020) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2020.108602 · Cited by 19 · see citing works
Vėbraitė, I., & Hanein, Y. (2021) · Frontiers in Medical Technology · doi:10.3389/fmedt.2021.675744 · Cited by 16 · see citing works
Hasiba-Pappas, S. K., Kamolz, L., Luze, H., Nischwitz, S. P., Holzer-Geissler, J. C. J., Tuca, A., Rienmüller, T. M., Polz, M., Ziesel, D., & Winter, R. (2023) · Journal of Personalized Medicine · doi:10.3390/jpm13030414 · Cited by 15 · see citing works
Flexible and Soft Materials and Devices for Neural Interface
Castagnola, E., Zheng, X. S., & Cui, X. T. (2023) · Handbook of Neuroengineering · doi:10.1007/978-981-16-5540-1_5 · Cited by 14 · see citing works
Coker, R., Zellmer, E. R., & Moran, D. W. (2018) · Journal of Neural Engineering · doi:10.1088/1741-2552/aaefab · Cited by 15 · see citing works
Karczewski, A. M., Dingle, A. M., & Poore, S. O. (2021) · Frontiers in Neurorobotics · doi:10.3389/fnbot.2021.711028 · Cited by 13 · see citing works
A Microclip Peripheral Nerve Interface (μcPNI) for Bioelectronic Interfacing with Small Nerves
Rowan, C. C., Graudejus, O., & Otchy, T. M. (2021) · Advanced Science · doi:10.1002/advs.202102945 · Cited by 11 · see citing works
Dingle, A. M., Ness, J. P., Novello, J., Israel, J. S., Sanchez, R., Millevolte, A. X. T., Brodnick, S. K., Krugner-Higby, L. A., Nemke, B. W., Lü, Y., et al. (2019) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2019.108504 · Cited by 12 · see citing works
Zellmer, E. R., MacEwan, M. R., & Moran, D. W. (2017) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa9e96 · Cited by 12 · see citing works
Coker, R., Zellmer, E. R., & Moran, D. W. (2018) · Journal of Neural Engineering · doi:10.1088/1741-2552/aaefcf · Cited by 11 · see citing works
Dingle, A. M., Ness, J. P., Novello, J., Millevolte, A. X. T., Zeng, W., Sanchez, R., Nemke, B. W., Lü, Y., Suminski, A. J., Markel, M. D., et al. (2020) · Military Medicine · doi:10.1093/milmed/usz246 · Cited by 10 · see citing works
Veith, A. P., Li, X., Modi, H., Abbaspour, A. R., Luan, L., Xie, C., & Baker, A. B. (2021) · Biomaterials · doi:10.1016/j.biomaterials.2021.120924 · Cited by 9 · see citing works
Atkinson, E. W., Kuliasha, C. A., Kasper, M., Furniturewalla, A., Lim, A. S., Jiracek-Sapieha, L., Brake, A., Gormaley, A., Rivera-Llabres, V. G., Singh, I., et al. (2022) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac8bfe · Cited by 9 · see citing works
Advances in wearable and implantable devices for wireless electrical stimulation therapy
Zhang, Y., Wang, Y., Zhao, K., Yang, M., Ye, Z., & Zhang, X. (2025) · Discover Electronics · doi:10.1007/s44291-025-00046-1 · Cited by 8 · see citing works
Cut wires: The Electrophysiology of Regenerated Tissue
Lowe, A. L., & Thakor, N. V. (2021) · Bioelectronic Medicine · doi:10.1186/s42234-021-00062-y · Cited by 6 · see citing works
Millevolte, A. X. T., Dingle, A. M., Ness, J. P., Novello, J., Zeng, W., Lü, Y., Minor, R. L., Nemke, B. W., Markel, M. D., Suminski, A. J., et al. (2021) · Frontiers in Neuroscience · doi:10.3389/fnins.2021.613844 · Cited by 6 · see citing works
Interfacing peripheral nerve with macro-sieve electrodes following spinal cord injury
Ray, W. Z., Birenbaum, N. K., & MacEwan, M. R. (2017) · Neural Regeneration Research · doi:10.4103/1673-5374.208565 · Cited by 5 · see citing works
Clinical Basis for Creating an Osseointegrated Neural Interface
Karczewski, A. M., Zeng, W., Stratchko, L. M., Bachus, K. N., Poore, S. O., & Dingle, A. M. (2022) · Frontiers in Neuroscience · doi:10.3389/fnins.2022.828593 · Cited by 5 · see citing works
Evaluation of Commercial Connectors for Active Neural Implants
Lancashire, H. T., Habibollahi, M., Jiang, D., & Demosthenous, A. (2021) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner49283.2021.9441072 · Cited by 5 · see citing works
Regenerative Electrodes for Peripheral Nerve Interfacing
Zellmer, E. R., & Moran, D. W. (2021) · Handbook of Neuroengineering · doi:10.1007/978-981-15-2848-4_18-1 · Cited by 4 · see citing works
STEER: 3D Printed Guide for Nerve Regrowth Control and Neural Interface in Non-Human Primate Model
Blasiak, A., Ng, K. A., Wong, M. D. S., Tsai, C.-W., Rusly, A., Gammad, G. G. L., Voges, K., Libedinsky, C., Yen, S., Thakor, N. V., et al. (2021) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2021.3113653 · Cited by 4 · see citing works
Flexible and Soft Materials and Devices for Neural Interface
Castagnola, E., Zheng, X. S., & Cui, X. T. (2021) · Handbook of Neuroengineering · doi:10.1007/978-981-15-2848-4_5-1 · Cited by 3 · see citing works
High-Frequency Alternating Current Block Using Macro-Sieve Electrodes: A Pilot Study
Ray, S., Javeed, S., Khalifeh, J. M., Chandra, N. S., Birenbaum, N. K., Felder, J. M., Moran, D. W., Ray, W. Z., & MacEwan, M. R. (2021) · Cureus · doi:10.7759/cureus.13728 · Cited by 3 · see citing works
High-Porosity Sieve-Type Neural Electrodes for Motor Function Recovery and Nerve Signal Acquisition
Choi, W., Park, H., Oh, S., Seok, S., Yoon, D. S., & Kim, J. (2024) · Micromachines · doi:10.3390/mi15070862 · Cited by 2 · see citing works
Sensory Percepts Elicited by Chronic Macro-Sieve Electrode Stimulation of the Rat Sciatic Nerve
Chandra, N. S., McCarron, W. M., Yan, Y., Ruiz, L. C., Sallinger, E. G., Birenbaum, N. K., Burton, H., Green, L., Moran, D. W., Ray, W. Z., et al. (2021) · Frontiers in Neuroscience · doi:10.3389/fnins.2021.758427 · Cited by 2 · see citing works
Chandra, N. S., Birenbaum, N. K., Green, L., Ray, W. Z., & MacEwan, M. R. (2022) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2022.109528 · Cited by 1 · see citing works
Regenerative Electrodes for Peripheral Nerve Interfacing
Zellmer, E. R., & Moran, D. W. (2023) · Handbook of Neuroengineering · doi:10.1007/978-981-16-5540-1_18 · Cited by 1 · see citing works
Li, M., Ly, C. M., Wen, J., Rustogi, P., & Meng, E. F. (2026) · Bioelectronic Medicine · doi:10.1186/s42234-025-00199-0 · Cited by 1 · see citing works
Chronic Nerve Interfacing Utilizing Graft-Embedded Regenerative Macro-Sieve Electrodes
Nishtala, A. (2018) · Open Scholarship Institutional Repository (Washington University in St. Louis) · doi:10.7936/k71n80j9
Directionally Sensitive Low-noise Neural Recording Using Tetrapolar Nerve Cuff Electrode
Sabetian, P. (2020) · TSpace (University of Toronto)
Rosario, M., Zhang, J., Kaleem, M. I., Chandra, N. S., Yan, Y., Moran, D. W., Wood, M. D., Ray, W. Z., & MacEwan, M. R. (2024) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2024.110295
- Authors
- MacEwan, M. R., Zellmer, E. R., Wheeler, J. J., Burton, H., & Moran, D. W.
- Affiliations
- Washington University in St. Louis (US)
- Corresponding author
- Moran, Daniel W.
- Venue
- Frontiers in Neuroscience
- Publisher
- Frontiers Media SA
- Published
- 2016-12-08
- Volume
- 10
- Pages
- 557-557
- ISSN
- 1662-453X (ISSN-L)
- PMC
- PMC5143347
- OpenAlex
- W2560020984
- Cited by
- 48
- References
- 38
- Citations by year
- 2026: 1 · 2025: 1 · 2024: 4 · 2023: 5 · 2022: 6 · 2021: 14 · 2020: 5 · 2019: 2 · 2018: 7 · 2017: 3
- FWCI
- 2.25
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- cc-by
- Locations
- Frontiers in Neuroscience (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Digital Commons@Becker (Washington University School of Medicine) (submittedVersion, cc-by)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Funders
- Defense Advanced Research Projects Agency (HR0011-15-2-0007)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Nerve injury and regeneration
- Health Sciences › Medicine › Surgery › Tissue Engineering and Regenerative Medicine
- Keywords
- peripheral nerve stimulation, axon regeneration, sciatic nerve, nerve conduits, glial cell line-derived neurotrophic factor, polyimide electrodes, neural interfaces, functional electrical stimulation, monopolar stimulation, muscle force
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review