CoChat

A Chest‐Conformable, Wireless Electro‐Mechanical E‐Tattoo for Measuring Multiple Cardiac Time Intervals

Open access

The DOI matches its Crossref record; 42 of the 42 references it cites were found in a registry or an index; 23 of the 54 sources its claims cite were checked for support.

Abstract

Abstract Cardiovascular diseases are the leading cause of death globally. Noninvasive, accurate, and continuous cardiovascular monitoring can enable the preemptive detection of heart diseases and timely intervention to prevent serious cardiac complications. However, unobtrusive, ambulatory, and comprehensive cardiac monitoring is still a challenge as conventional electronics are rigid, heavy, or consume too much power for long‐term measurement. This work presents a thin (200 µm), stretchable (20%), lightweight (2.5 g), wireless, and low‐power (<3 mW) cardiac monitoring device that conforms to the human chest like a temporary tattoo sticker, correspondingly known as an e‐tattoo. This chest e‐tattoo features dual‐mode electro‐mechanical sensing—bio‐electric cardiac signals via electrocardiography and mechanical cardiac rhythm via seismocardiography. A unique peripheral synchronization strategy between the two sensors enables the measurement of systolic time intervals like the pre‐ejection period and the left ventricular ejection time with high accuracy (error = −0.44 ± 8.74 ms) while consuming very low power. The e‐tattoo is validated against clinically approved gold‐standard instruments on five human subjects. The good wearability and low power consumption of this e‐tattoo permit 24‐h continuous ambulatory monitoring.

Overview

CoChat Summary

TL;DR

This paper presents a chest-conformable, wireless electro-mechanical e-tattoo that combines ECG and SCG for measuring cardiac time intervals such as PEP and LVET. It reports low-power, stretchable, reusable operation and validates STI measurement against a clinical monitor in five healthy subjects, with 24-h ambulatory feasibility in one subject.

Key findings

  • The e-tattoo is thin (200 µm), stretchable (20%), lightweight (2.5 g including battery), wireless, and low-power (<3 mW; average current draw 940 µA), with a CR1220 coin cell supporting up to 40 h of operation.
  • Dual ECG/SCG sensing with peripheral synchronization enabled LVET measurement against NICOM with a mean difference of −0.44 ± 8.74 ms and Pearson correlation of 0.955; HR differences were 0.07 ± 1.21 bpm during static poses and 0.02 ± 1.24 bpm during cycling.
  • The e-tattoo showed no significant signal degradation after 20% off-chest stretching and reapplication, and no perceptible damage after ten cycles of electrode-layer reuse.
  • In posture transitions, statistically significant changes were observed in PEP (p < 0.05, paired t-test) and LVET (p < 0.005, paired t-test) from supine to upright/standing.

Methods

  • Device design: a flexible printed circuit with serpentine interconnects, graphite polyurethane dry ECG electrodes on Tegaderm, an ADXL355 accelerometer for SCG, a MAX30003 ECG front end, and an nRF52832 BLE CPU; ECG and SCG were sampled at 125 Hz with hardware synchronization using an external 32.768 kHz crystal and ECG-generated sync pulses.
  • Human testing: IRB-approved study at UT Austin; five healthy males aged 20–28 for static posture and incremental cycling experiments compared against NICOM, plus one 27-year-old male for >24 h ambulatory monitoring compared against a Samsung Galaxy Watch 3 for HR.
  • Signal processing: ECG band-pass filtered at 0.5–40 Hz, SCG at 1–40 Hz, R-peak extraction, beat-by-beat SCG segmentation, signal quality indexing (λ = 25, threshold = 0.5), and AO/AC peak tracking to compute HR, PEP, and LVET.

Limitations

  • The paper states that SCG was corrupted by motion during cycling, so STIs were computed only during rest segments after exercise; long-term LVET/PEP could not be compared against a reference device because NICOM is non-portable and no wearable LVET/PEP device was available.
  • Evident from the text: the primary validation sample was small (five healthy males aged 20–28), the long-term ambulatory study used only one participant, and participants with known cardiovascular disease were excluded, limiting generalizability to patients, women, older adults, and higher-BMI or diseased populations.
  • The paper states “Research data are not shared,” which limits independent verification; the text also notes that SCG is unusable during significant motion, restricting continuous STI extraction to restful segments.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.1002/aelm.202201284

Cites

42 of 42 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Figures

Figure 1 Introduction to systolic time intervals (STIs), e-tattoo overview, and comparison to related work. a) Different heart events captured via different modalities and used to calculate pre-ejection period (PEP) and left ventricular ejection time (LVET). b) Exploded schematic illustration of the wireless, dual-mode chest e-tattoo. Three layers from top to bottom are cover layer, flexible printed circuit (FPC), and electrode layer. c) Synchronized real electrocardiogram (ECG) and seismocardiogram (SCG) signals with important features annotated including QRS complex and T-peak on ECG, aortic valve opening (AO), aortic valve closing (AC), mitral valve closing (MC), and mitral valve opening (MO) on SCG. d) Photograph of the e-tattoo conforming to the curvature of the chest. e) Comparison of e-tattoo device weight and system current draw with similar wearable cardiac devices reported in the literature. f) Custom-designed mobile application interface for recording, visualizing, and annotating data.
Figure 1 Introduction to systolic time intervals (STIs), e-tattoo overview, and comparison to related work. a) Different heart events captured via different modalities and used to calculate pre-ejection period (PEP) and left ventricular ejection time (LVET). b) Exploded schematic illustration of the wireless, dual-mode chest e-tattoo. Three layers from top to bottom are cover layer, flexible printed circuit (FPC), and electrode layer. c) Synchronized real electrocardiogram (ECG) and seismocardiogram (SCG) signals with important features annotated including QRS complex and T-peak on ECG, aortic valve opening (AO), aortic valve closing (AC), mitral valve closing (MC), and mitral valve opening (MO) on SCG. d) Photograph of the e-tattoo conforming to the curvature of the chest. e) Comparison of e-tattoo device weight and system current draw with similar wearable cardiac devices reported in the literature. f) Custom-designed mobile application interface for recording, visualizing, and annotating data.CC-BY · From the PDF
Image 2
CC-BY · From the PDF
Figure 2 Circuit design, firmware flow, and power draw of the e-tattoo. a) A block diagram of the major hardware components of the e-tattoo. b) A diagram of the firmware flow for data collection and transmission. c) Active power draw of the e-tattoo during operation. The average current draw is shown in blue.
Figure 2 Circuit design, firmware flow, and power draw of the e-tattoo. a) A block diagram of the major hardware components of the e-tattoo. b) A diagram of the firmware flow for data collection and transmission. c) Active power draw of the e-tattoo during operation. The average current draw is shown in blue.CC-BY · From the PDF
Image 4
CC-BY · From the PDF
Figure 3 Mechanical performance of the chest e-tattoo. a) Optical image of the assembled flexible printed circuit (FPC) of the e-tattoo. b) Serpentine interconnects enable stretching up to 20% without damage. c) Signal quality before and after off-chest stretching and reapplication on the chest shows no significant degradation. Photographs comparing d) the e-tattoo and e) its rigid counterpart mounted on the same human chest where the e-tattoo shows conformable contact while the rigid device shows a gap from the sternum. f) Schematic illustration of the re-usability of the e-tattoo FPC by peeling off the used electrode layer and cover layer and reassembling with new ones. Photos illustrating g) a new e-tattoo FPC layer and h) the FPC layer after undergoing ten cycles of reassembly and reuse. i) electrocardiogram signals measured by new FPC (blue) and ten-time-reused FPC (green) do not show any perceptible differences.
Figure 3 Mechanical performance of the chest e-tattoo. a) Optical image of the assembled flexible printed circuit (FPC) of the e-tattoo. b) Serpentine interconnects enable stretching up to 20% without damage. c) Signal quality before and after off-chest stretching and reapplication on the chest shows no significant degradation. Photographs comparing d) the e-tattoo and e) its rigid counterpart mounted on the same human chest where the e-tattoo shows conformable contact while the rigid device shows a gap from the sternum. f) Schematic illustration of the re-usability of the e-tattoo FPC by peeling off the used electrode layer and cover layer and reassembling with new ones. Photos illustrating g) a new e-tattoo FPC layer and h) the FPC layer after undergoing ten cycles of reassembly and reuse. i) electrocardiogram signals measured by new FPC (blue) and ten-time-reused FPC (green) do not show any perceptible differences.CC-BY · From the PDF
Image 6
CC-BY · From the PDF
Figure 4 Individual validation of electrocardiogram (ECG) and seismocardiogram (SCG) sensors. a) Gel electrodes and b) dry e-tattoo electrodes applied on a human chest. c) ECG signals from both electrodes have similar SNR and well-defined QRS complex. d) SCG signal compared with M-mode echocardiogram. Aortic valve opening (AO) and aortic valve closing (AC) events in the echocardiogram correlate to defined features on the SCG signal. e) Orientation of the human subject can be identified from the SCG signals. f) Photograph of the platform used to test the performance and signal quality of the SCG sensor ex vivo. The SCG sensor is able to detect the AO and AC peak from simulated chest vibrations generated from g) synthetic and h) real human SCG signals, both implemented by the platform.
Figure 4 Individual validation of electrocardiogram (ECG) and seismocardiogram (SCG) sensors. a) Gel electrodes and b) dry e-tattoo electrodes applied on a human chest. c) ECG signals from both electrodes have similar SNR and well-defined QRS complex. d) SCG signal compared with M-mode echocardiogram. Aortic valve opening (AO) and aortic valve closing (AC) events in the echocardiogram correlate to defined features on the SCG signal. e) Orientation of the human subject can be identified from the SCG signals. f) Photograph of the platform used to test the performance and signal quality of the SCG sensor ex vivo. The SCG sensor is able to detect the AO and AC peak from simulated chest vibrations generated from g) synthetic and h) real human SCG signals, both implemented by the platform.CC-BY · From the PDF
Image 8
CC-BY · From the PDF
Figure 5 Human subject testing protocol and results. a) Illustration of Institutional Review Board-approved human subject testing protocol for inducing small and large changes in heart rate (HR) and systolic time intervals. b) Bland-Altman plot comparing HR from e-tattoo to HR from the non-invasive cardiac output monitor (NICOM) during static posture testing. c) Bar plot illustrating pre-ejection period (PEP) change for different participants in different postures. d) Bar plot illustrating left ventricular ejection time (LVET) change for different participants in different postures. e) Time series HR data from e-tattoo and NICOM during incremental cycling experiment. f) Bland-Altman plot comparing LVET from e-tattoo to LVET from NICOM during the rest intervals after cycling exercise. g) Correlation between LVET simultaneously measured by e-tattoo versus NICOM during the same intervals.
Figure 5 Human subject testing protocol and results. a) Illustration of Institutional Review Board-approved human subject testing protocol for inducing small and large changes in heart rate (HR) and systolic time intervals. b) Bland-Altman plot comparing HR from e-tattoo to HR from the non-invasive cardiac output monitor (NICOM) during static posture testing. c) Bar plot illustrating pre-ejection period (PEP) change for different participants in different postures. d) Bar plot illustrating left ventricular ejection time (LVET) change for different participants in different postures. e) Time series HR data from e-tattoo and NICOM during incremental cycling experiment. f) Bland-Altman plot comparing LVET from e-tattoo to LVET from NICOM during the rest intervals after cycling exercise. g) Correlation between LVET simultaneously measured by e-tattoo versus NICOM during the same intervals.CC-BY · From the PDF
Image 10
CC-BY · From the PDF
Figure 6 Long-term continuous and ambulatory sensing using e-tattoo. a) Time series heart rate (HR) data from e-tattoo and Samsung Galaxy smartwatch during a 24-h period. b-d) The electrocardiogram signal (green) and the seismoocardiogram signal (blue) from the rest intervals after different activities performed in the 24-h period.
Figure 6 Long-term continuous and ambulatory sensing using e-tattoo. a) Time series heart rate (HR) data from e-tattoo and Samsung Galaxy smartwatch during a 24-h period. b-d) The electrocardiogram signal (green) and the seismoocardiogram signal (blue) from the rest intervals after different activities performed in the 24-h period.CC-BY · From the PDF

Verification

Claims the paper makes

13 Valid · 6 Not valid · 10 Not checked · 59 Own results and interpretations

42 more claims · 6 with their cited sources checked

Create a free account to see all 88 claims and to verify the 1 reference still open.

See all claims — free account
  • It may state two separate claims as one. 27
  • Probably not a substantial claim. 21

Create a free account to see these claims.

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

Each contradiction between two passages listed here was confirmed by an independent check.

1 of 1

Citations

Cites

Show all 42 references

Research works printed but not linked: 10

  1. Bruss Z. S., 2022, Physiology, Stroke Volume, StatPearls
  2. Preejith S., 2016, 2016 38th Annual Int. Conf. of the IEEE Engineering in Medicine and Biology Society (EMBC), 623
  3. Li J., 2022, 2022 44th Annual Int. Conf. of the IEEE Engineering in Medicine and Biology Society (EMBC), 4312
  4. Zhong Y., 2016, 2016 38th Annual Int. Conf. of the IEEE Engineering in Medicine and Biology Society (EMBC), 603
  5. Lee S., 2019, 2019 41st Annual Int. Conf. of the IEEE Engineering in Medicine and Biology Society (EMBC), 1290
  6. Di Rienzo M., 2014, 2014 36th Annual Int. Conf. of the IEEE Engineering in Medicine and Biology Society, 6089
  7. Tadi M. J., 2015, Sixth Int. Conf. on Graphic and Image Processing (ICGIP 2014), 571
  8. Hansen I. H., 2015, 2015 37th Annual Int. Conf. of the IEEE Engineering in Medicine and Biology Society (EMBC), 330
  9. Gazi A. H., 2021, 43rd Annual Int. Conf. of the IEEE Engineering in Medicine and Biology Society (EMBC), 1444
  10. Yang Y.‐F., 2019, Proc. of Int. Conf. on Biomedical and Health Informatics, 363

Cited by

See all 42

Metadata

Authors
Bhattacharya, S., Nikbakht, M., Alden, A., Tan, P., Wang, J., Alhalimi, T. A., Kim, S., Wang, P., Tanaka, H., Tandon, A., Coyle, E. F., Inan, O. T., & Lu, N.
Affiliations
  • The University of Texas at Austin (US)
  • Georgia Institute of Technology (US)
  • Stretch Med Inc. Austin TX 78750 USA
  • Cleveland Clinic (US)
  • Texas Materials Institute (US)
Corresponding author
Lu, Nanshu
Venue
Advanced Electronic Materials
Publisher
Wiley
Published
2023-04-18
Volume
9
Issue
9
ISSN
2199-160X (ISSN-L)
OpenAlex
W4366245263
Cited by
42
References
42
Citations by year
2026: 8 · 2025: 17 · 2024: 10 · 2023: 7
FWCI
3.31
Type
journal article
Language
en
Open access
Locations
  • Advanced Electronic Materials (publishedVersion, cc-by)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
Funders
  • National Science Foundation (STTR‐1938254, ASCENT‐2133106)
  • Office of Naval Research Global (N00014‐20‐1‐2112)
Topics
  • Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
  • Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
  • Physical Sciences › Engineering › Biomedical Engineering › Wireless Body Area Networks
Keywords
stretchable electronics, electrocardiogram, seismocardiography, cardiac time intervals, systolic time intervals, pre-ejection period, left ventricular ejection time, low-power sensors, sensor synchronization, ambulatory monitoring, wearable health monitoring

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

The claim verdicts this report counts are experimental: their accuracy is not yet measured.

A language model drew the claims from the paper on 2026-10-12, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).

A second model labelled each claim on 2026-10-12 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).

Cited sources are checked with a language model’s help (gpt-6-sol).

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.