Motion artifacts in MRI: A complex problem with many partial solutions
Open accessAbstract
Subject motion during magnetic resonance imaging (MRI) has been problematic since its introduction as a clinical imaging modality. While sensitivity to particle motion or blood flow can be used to provide useful image contrast, bulk motion presents a considerable problem in the majority of clinical applications. It is one of the most frequent sources of artifacts. Over 30 years of research have produced numerous methods to mitigate or correct for motion artifacts, but no single method can be applied in all imaging situations. Instead, a “toolbox” of methods exists, where each tool is suitable for some tasks, but not for others. This article reviews the origins of motion artifacts and presents current mitigation and correction methods. In some imaging situations, the currently available motion correction tools are highly effective; in other cases, appropriate tools still need to be developed. It seems likely that this multifaceted approach will be what eventually solves the motion sensitivity problem in MRI, rather than a single solution that is effective in all situations. This review places a strong emphasis on explaining the physics behind the occurrence of such artifacts, with the aim of aiding artifact detection and mitigation in particular clinical situations. J. Magn. Reson. Imaging 2015;42:887–901.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Sparse MRI: The application of compressed sensing for rapid MR imaging
Lustig, M., Donoho, D., & Pauly, J. M. (2007) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21391 · Cited by 6,952
SENSE: Sensitivity encoding for fast MRI
Pruessmann, K. P., Weiger, M., Scheidegger, M. B., & Boesiger, P. (1999) · Magnetic Resonance in Medicine · doi:10.1002/(sici)1522-2594(199911)42:5<952::aid-mrm16>3.0.co;2-s · Cited by 6,251
Generalized autocalibrating partially parallel acquisitions (GRAPPA)
Griswold, M. A., Jakob, P. M., Heidemann, R. M., Nittka, M., Jellus, V., Wang, J., Kiefer, B., & Haase, A. (2002) · Magnetic Resonance in Medicine · doi:10.1002/mrm.10171 · Cited by 5,303
Simultaneous acquisition of spatial harmonics (SMASH): Fast imaging with radiofrequency coil arrays
Sodickson, D. K., & Manning, W. J. (1997) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910380414 · Cited by 2,142
RARE imaging: A fast imaging method for clinical MR
Hennig, J., Nauerth, A., & Friedburg, H. (1986) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910030602 · Cited by 2,081
Multi-planar image formation using NMR spin echoes
Mansfield, P. (1977) · Journal of Physics C: Solid State Physics · doi:10.1088/0022-3719/10/3/004 · Cited by 1,813
Handbook of MRI Pulse Sequences
(2004) · Elsevier eBooks · doi:10.1016/b978-0-12-092861-3.x5000-6 · Cited by 1,610
Advances in sensitivity encoding with arbitrary k‐space trajectories
Pruessmann, K. P., Weiger, M., Börnert, P., & Boesiger, P. (2001) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1241 · Cited by 1,211
FLASH imaging: Rapid NMR imaging using low flip-angle pulses
Haase, A., Frahm, J., Matthaei, D., Hänicke, W., & Merboldt, K.-D. (2011) · Journal of Magnetic Resonance · doi:10.1016/j.jmr.2011.09.021 · Cited by 1,153
Motion correction with PROPELLER MRI: Application to head motion and free-breathing cardiac imaging
Pipe, J. G. (1999) · Magnetic Resonance in Medicine · doi:10.1002/(sici)1522-2594(199911)42:5<963::aid-mrm17>3.0.co;2-l · Cited by 928
Time‐resolved contrast‐enhanced 3D MR angiography
Korosec, F. R., Frayne, R., Grist, T. M., & Mistretta, C. A. (1996) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910360304 · Cited by 877
FLASH imaging. Rapid NMR imaging using low flip-angle pulses
Haase, A., Frahm, J., Matthaei, D., Hanicke, W., & Merboldt, K.-D. (1986) · Journal of Magnetic Resonance (1969) · doi:10.1016/0022-2364(86)90433-6 · Cited by 826
Adaptive technique for high-definition MR imaging of moving structures.
Ehman, R. L., & Felmlee, J. P. (1989) · Radiology · doi:10.1148/radiology.173.1.2781017 · Cited by 801
Artifacts and pitfalls in diffusion MRI
Le Bihan, D., Poupon, C., Amadon, A., & Lethimonnier, F. (2006) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.20683 · Cited by 778
Principles and applications of balanced SSFP techniques
Scheffler, K., & Lehnhardt, S. (2003) · European Radiology · doi:10.1007/s00330-003-1957-x · Cited by 714
Breuer, F. A., Blaimer, M., Heidemann, R. M., Mueller, M. F., Griswold, M. A., & Jakob, P. M. (2005) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20401 · Cited by 633
Physiological noise in oxygenation‐sensitive magnetic resonance imaging
Krüger, G., & Glover, G. H. (2001) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1240 · Cited by 629
“Keyhole” method for accelerating imaging of contrast agent uptake
Van Vaals, J. J., Brummer, M. E., Thomas Dixon, W., Tuithof, H. H., Engels, H., Nelson, R. C., Gerety, B. M., Chezmar, J. L., & Den Boer, J. A. (1993) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.1880030419 · Cited by 602
Prospective acquisition correction for head motion with image-based tracking for real-time fMRI
Thesen, S., Heid, O., Mueller, E., & Schad, L. R. (2000) · Magnetic Resonance in Medicine · doi:10.1002/1522-2594(200009)44:3<457::aid-mrm17>3.0.co;2-r · Cited by 572
Projection Reconstruction Techniques for Reduction of Motion Effects in MRI
Glover, G. H., & Pauly, J. M. (1992) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910280209 · Cited by 563
Wang, Y., Riederer, S. J., & Ehman, R. L. (1995) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910330517 · Cited by 491
Self‐navigated spiral fMRI: Interleaved versus single‐shot
Glover, G. H., & Lai, S. (1998) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910390305 · Cited by 476
Tisdall, M. D., Hess, A. T., Reuter, M., Meintjes, E. M., Fischl, B., & van der Kouwe, A. J. W. (2012) · Magnetic Resonance in Medicine · doi:10.1002/mrm.23228 · Cited by 443
Ultra-fast MRI of the human brain with simultaneous multi-slice imaging
Feinberg, D. A., & Setsompop, K. (2013) · Journal of Magnetic Resonance · doi:10.1016/j.jmr.2013.02.002 · Cited by 433
Multishot diffusion‐weighted FSE using PROPELLER MRI
Pipe, J. G., Farthing, V. G., & Forbes, K. P. (2002) · Magnetic Resonance in Medicine · doi:10.1002/mrm.10014 · Cited by 428
Controlled aliasing in volumetric parallel imaging (2D CAIPIRINHA)
Breuer, F. A., Blaimer, M., Mueller, M. F., Seiberlich, N., Heidemann, R. M., Griswold, M. A., & Jakob, P. M. (2006) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20787 · Cited by 413
Prospective motion correction in brain imaging: A review
Maclaren, J., Herbst, M., Speck, O., & Zaitsev, M. (2013) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24314 · Cited by 401
Self-citationZaitsev, M., Dold, C., Sakas, G., Hennig, J., & Speck, O. (2006) · NeuroImage · doi:10.1016/j.neuroimage.2006.01.039 · Cited by 391
Self-citationPROMO: Real‐time prospective motion correction in MRI using image‐based tracking
White, N., Roddey, C., Shankaranarayanan, A., Han, E., Rettmann, D., Santos, J., Kuperman, J., & Dale, A. (2010) · Magnetic Resonance in Medicine · doi:10.1002/mrm.22176 · Cited by 376
Rapid gridding reconstruction with a minimal oversampling ratio
Beatty, P. J., Nishimura, D. G., & Pauly, J. M. (2005) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2005.848376 · Cited by 333
Parallel magnetic resonance imaging
Larkman, D. J., & Nunes, R. G. (2007) · Physics in Medicine and Biology · doi:10.1088/0031-9155/52/7/r01 · Cited by 327
Bailes, D. R., Gilderdale, D. J., Bydder, G. M., Collins, A. G., & Firmin, D. N. (1985) · Journal of Computer Assisted Tomography · doi:10.1097/00004728-198507010-00039 · Cited by 313
MR image artifacts from periodic motion
Wood, M. L., & Henkelman, R. M. (1985) · Medical Physics · doi:10.1118/1.595782 · Cited by 302
Bailes, D. R., Gilderdale, D. J., Bydder, G. M., et al. (1985) · PubMed · Cited by 300
Magnetic resonance imaging with respiratory gating: techniques and advantages
Ehman, R., McNamara, M., Pallack, M., Hricak, H., & Higgins, C. (1984) · American Journal of Roentgenology · doi:10.2214/ajr.143.6.1175 · Cited by 299
Barger, A. V., Block, W. F., Toropov, Y., Grist, T. M., & Mistretta, C. A. (2002) · Magnetic Resonance in Medicine · doi:10.1002/mrm.10212 · Cited by 291
Self‐navigated interleaved spiral (SNAILS): Application to high‐resolution diffusion tensor imaging
Liu, C., Bammer, R., Kim, D., & Moseley, M. E. (2004) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20288 · Cited by 247
Diffusion‐weighted interleaved echo‐planar imaging with a pair of orthogonal navigator echoes
Butts, K., de Crespigny, A., Pauly, J. M., & Moseley, M. (1996) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910350518 · Cited by 244
Respiration‐induced B0fluctuations and their spatial distribution in the human brain at 7 Tesla
Van de Moortele, P., Pfeuffer, J., Glover, G. H., Ugurbil, K., & Hu, X. (2002) · Magnetic Resonance in Medicine · doi:10.1002/mrm.10145 · Cited by 240
Real‐time rigid body motion correction and shimming using cloverleaf navigators
van der Kouwe, A. J. W., Benner, T., & Dale, A. M. (2006) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21038 · Cited by 230
Measurement and Correction of Microscopic Head Motion during Magnetic Resonance Imaging of the Brain
Maclaren, J., Armstrong, B. S. R., Barrows, R. T., Danishad, K. A., Ernst, T., Foster, C. L., Gumus, K., Herbst, M., Kadashevich, I. Y., Kusik, T. P., et al. (2012) · PLoS ONE · doi:10.1371/journal.pone.0048088 · Cited by 215
Self-citationSpherical navigator echoes for full 3D rigid body motion measurement in MRI
Welch, E. B., Manduca, A., Grimm, R. C., Ward, H. A., & Jack Jr., C. R. (2002) · Magnetic Resonance in Medicine · doi:10.1002/mrm.10012 · Cited by 212
Generalized Reconstruction by Inversion of Coupled Systems (GRICS) applied to free‐breathing MRI
Odille, F., Vuissoz, P., Marie, P., & Felblinger, J. (2008) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21623 · Cited by 192
Motion corrected compressed sensing for free‐breathing dynamic cardiac MRI
Usman, M., Atkinson, D., Odille, F., Kolbitsch, C., Vaillant, G., Schaeffter, T., Batchelor, P. G., & Prieto, C. (2013) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24463 · Cited by 185
Bellon, E., Haacke, E., Coleman, P., Sacco, D., Steiger, D., & Gangarosa, R. (1986) · American Journal of Roentgenology · doi:10.2214/ajr.147.6.1271 · Cited by 174
Magnetic field changes in the human brain due to swallowing or speaking
Birn, R. M., Bandettini, P. A., Cox, R. W., Jesmanowicz, A., & Shaker, R. (1998) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910400108 · Cited by 172
Is the brain cortex a fractal?
Kiselev, V. G., Hahn, K. R., & Auer, D. P. (2003) · NeuroImage · doi:10.1016/s1053-8119(03)00380-x · Cited by 167
Prospective real‐time correction for arbitrary head motion using active markers
Ooi, M. B., Krueger, S., Thomas, W. J., Swaminathan, S. V., & Brown, T. R. (2009) · Magnetic Resonance in Medicine · doi:10.1002/mrm.22082 · Cited by 165
Lim, R. P., Shapiro, M., Wang, E. Y., Law, M., Babb, J. S., Rueff, L. E., Jacob, J. S., Kim, S., Carson, R. H., Mulholland, T. P., et al. (2008) · American Journal of Neuroradiology · doi:10.3174/ajnr.a1252 · Cited by 164
Forbes, K. P. N., Pipe, J. G., Bird, C. R., & Heiserman, J. E. (2001) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.1176 · Cited by 159
Real‐time motion and B 0 corrected single voxel spectroscopy using volumetric navigators
Hess, A. T., Tisdall, M. D., Andronesi, O. C., Meintjes, E. M., & van der Kouwe, A. J. W. (2011) · Magnetic Resonance in Medicine · doi:10.1002/mrm.22805 · Cited by 131
Prospective motion correction of high-resolution magnetic resonance imaging data in children
Brown, T. T., Kuperman, J. M., Erhart, M. J., White, N. S., Roddey, J. C., Shankaranarayanan, A., Han, E. T., Rettmann, D. W., & Dale, A. M. (2010) · NeuroImage · doi:10.1016/j.neuroimage.2010.06.017 · Cited by 130
Automatic compensation of motion artifacts in MRI
Atkinson, D., Hill, D., Stoyle, P. N. R., Summers, P. E., Clare, S., Bowtell, R. W., & Keevil, S. (1999) · Magnetic Resonance in Medicine · doi:10.1002/(sici)1522-2594(199901)41:1<163::aid-mrm23>3.0.co;2-9 · Cited by 122
Prospective Real-Time Slice-by-Slice Motion Correction for fMRI in Freely Moving Subjects
Speck, O., Hennig, J., & Zaitsev, M. (2006) · Magnetic Resonance Materials in Physics Biology and Medicine · doi:10.1007/s10334-006-0027-1 · Cited by 113
Self-citationBrau, A. C. S., & Brittain, J. H. (2006) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20785 · Cited by 109
Paschal, C. B., & Morris, H. D. (2004) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.10451 · Cited by 107
Augmented generalized SENSE reconstruction to correct for rigid body motion
Bammer, R., Aksoy, M., & Liu, C. (2006) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21106 · Cited by 105
Forbes, K. P. N., Pipe, J. G., Karis, J. P., Farthing, V. G., & Heiserman, J. E. (2003) · PubMed Central · Cited by 97
Turboprop: Improved PROPELLER imaging
Pipe, J. G., & Zwart, N. R. (2006) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20768 · Cited by 94
Zhang, T., Chowdhury, S., Lustig, M. S., Barth, R. A., Alley, M. T., Grafendorfer, T., Calderon, P. D., Robb, F. J. L., Pauly, J. M., & Vasanawala, S. S. (2013) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.24333 · Cited by 91
Blind retrospective motion correction of MR images
Loktyushin, A., Nickisch, H., Pohmann, R., et al. (2013) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24615 · Cited by 90
Delattre, B. M. A., Heidemann, R. M., Crowe, L. A., Vallée, J., & Hyacinthe, J. (2010) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2010.03.036 · Cited by 86
Adaptive technique for three-dimensional MR imaging of moving structures.
Korin, H. W., Felmlee, J. P., Ehman, R. L., & Riederer, S. J. (1990) · Radiology · doi:10.1148/radiology.177.1.2399320 · Cited by 78
Prospective correction of affine motion for arbitrary MR sequences on a clinical scanner
Nehrke, K., & Börnert, P. (2005) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20686 · Cited by 78
Hybrid ultrasound MRI for improved cardiac imaging and real‐time respiration control
Feinberg, D. A., Giese, D., Bongers, D. A., Ramanna, S., Zaitsev, M., Markl, M., & Günther, M. (2009) · Magnetic Resonance in Medicine · doi:10.1002/mrm.22250 · Cited by 74
Self-citationNonrigid retrospective respiratory motion correction in whole‐heart coronary MRA
Schmidt, J. F. M., Buehrer, M., Boesiger, P., & Kozerke, S. (2011) · Magnetic Resonance in Medicine · doi:10.1002/mrm.22939 · Cited by 69
Hybrid prospective and retrospective head motion correction to mitigate cross‐calibration errors
Aksoy, M., Forman, C., Straka, M., Çukur, T., Joachim Hornegger, A., & Bammer, R. (2011) · Magnetic Resonance in Medicine · doi:10.1002/mrm.23101 · Cited by 69
Quantification of fMRI artifact reduction by a novel plaster cast head holder
Edward, V., Windischberger, C., Cunnington, R., Erdler, M., Lanzenberger, R. R., Mayer, D., Endl, W., & Beisteiner, R. (2000) · Human Brain Mapping · Cited by 68
Prospective motion correction with continuous gradient updates in diffusion weighted imaging
Herbst, M., Maclaren, J., Weigel, M., Korvink, J. G., Hennig, J., & Zaitsev, M. (2011) · Magnetic Resonance in Medicine · doi:10.1002/mrm.23230 · Cited by 68
Self-citationDetection and elimination of motion artifacts by regeneration of k ‐space
Bydder, M., Larkman, D. J., & Hajnal, J. V. (2002) · Magnetic Resonance in Medicine · doi:10.1002/mrm.10093 · Cited by 61
Prospective motion correction for single‐voxel 1H MR spectroscopy
Keating, B., Deng, W., Roddey, J. C., White, N. S., Dale, A. M., Stenger, V. A., & Ernst, T. (2010) · Magnetic Resonance in Medicine · doi:10.1002/mrm.22448 · Cited by 50
Norris, D. G., & Driesel, W. (2001) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1099 · Cited by 47
Bydder, M., Atkinson, D., Larkman, D. J., Hill, D., & Hajnal, J. V. (2003) · Magnetic Resonance in Medicine · doi:10.1002/mrm.10388 · Cited by 46
Direct monitoring of coronary artery motion with cardiac fat navigator echoes
Nguyen, T. D., Nuval, A., Mulukutla, S. R., & Wang, Y. (2003) · Magnetic Resonance in Medicine · doi:10.1002/mrm.10550 · Cited by 44
Boegle, R., Maclaren, J., & Zaitsev, M. (2010) · Magnetic Resonance Materials in Physics Biology and Medicine · doi:10.1007/s10334-010-0225-8 · Cited by 42
Self-citationSchulz, J., Siegert, T., Bazin, P., Maclaren, J., Herbst, M., Zaitsev, M., & Turner, R. (2013) · NeuroImage · doi:10.1016/j.neuroimage.2013.08.006 · Cited by 35
Self-citationGümüş, K., Keating, B., Poser, B. A., Armstrong, B., Chang, L., Maclaren, J., Prieto, T. E., Speck, O., Zaitsev, M., & Ernst, T. (2013) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24857 · Cited by 26
Self-citationMenon, V., Lim, K. O., Anderson, J. H., Johnson, J., & Adolf Pfefferbaum, A. (1997) · Behavior Research Methods, Instruments, & Computers · doi:10.3758/bf03210613 · Cited by 23
Reduction of motion artifacts in carotid MRI using free‐induction decay navigators
Dyverfeldt, P., Deshpande, V. S., Kober, T., et al. (2013) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.24389 · Cited by 14
Cited by
Image processing and analysis methods for the Adolescent Brain Cognitive Development Study
Hagler, D. J., Hatton, S., Cornejo, M. D., Makowski, C., Fair, D. A., Dick, A. S., Sutherland, M. T., Casey, B. J., Barch, D. M., Harms, M. P., et al. (2019) · NeuroImage · doi:10.1016/j.neuroimage.2019.116091 · Cited by 1,088
Pros and cons of ultra-high-field MRI/MRS for human application
Ladd, M. E., Bachert, P., Meyerspeer, M., Moser, E., Nagel, A. M., Norris, D. G., Schmitter, S., Speck, O., Straub, S., & Zaiss, M. (2018) · Progress in Nuclear Magnetic Resonance Spectroscopy · doi:10.1016/j.pnmrs.2018.06.001 · Cited by 544
Quantitative assessment of structural image quality
Rosen, A. F. G., Roalf, D. R., Ruparel, K., Blake, J., Seelaus, K., Villa, L. P., Ciric, R., Cook, P. A., Davatzikos, C., Elliott, M. A., et al. (2018) · NeuroImage · doi:10.1016/j.neuroimage.2017.12.059 · Cited by 424
Tanenbaum, L. N., Tsiouris, A. J., Johnson, A. N., Naidich, T. P., DeLano, M. C., Melhem, E. R., Quarterman, P., Parameswaran, S. X., Shankaranarayanan, A., Goyen, M., et al. (2017) · American Journal of Neuroradiology · doi:10.3174/ajnr.a5227 · Cited by 296
Regional, circuit and network heterogeneity of brain abnormalities in psychiatric disorders
Segal, A., Parkes, L., Aquino, K., Kia, S. M., Wolfers, T., Franke, B., Hoogman, M., Beckmann, C. F., Westlye, L. T., Andreassen, O. A., et al. (2023) · Nature Neuroscience · doi:10.1038/s41593-023-01404-6 · Cited by 271
Metadata
- Authors
- Zaitsev, M., Maclaren, J., & Herbst, M.
- Venue
- Journal of Magnetic Resonance Imaging
- Publisher
- Wiley
- Published
- 2015-01-28
- Volume
- 42
- Issue
- 4
- Pages
- 887-901
- ISSN
- 1053-1807 (ISSN-L)
- 1522-2586
- OpenAlex
- W1812490466
- Cited by
- 774
- References
- 79
- FWCI
- 21.92
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.