Optical coherence tomography of the human skin
Open accessThe DOI matches its Crossref record; 11 of the 11 references it cites were found in a registry or an index.
Abstract
BACKGROUND Optical coherence tomography (OCT) is a new diagnostic method for tissue characterization. OBJECTIVE We investigated normal and pathologic structures in human skin in several locations to evaluate the potential application of this technique to dermatology. METHODS Based on the principle of low-coherence interferometry, cross-sectional images of the human skin can be obtained in vivo with a high spatial resolution of about 15 microns. Within a penetration depth of 0.5 to 1.5 mm, structures of the stratum corneum, the living epidermis, and the papillary dermis can be distinguished. RESULTS Different layers could be detected that were differentiated by induction of experimental blisters and by comparison with corresponding histologic sections. Furthermore, OCT images of several skin diseases and tumors were obtained. CONCLUSION OCT is a promising new imaging method for visualization of morphologic changes of superficial layers of the human skin. It may be useful for noninvasive diagnosis of bullous skin diseases, skin tumors, and in vivo investigation of pharmacologic effects.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Huang, D., Swanson, E. A., Lin, C. P., Schuman, J. S., Stinson, W. G., Chang, W., Hee, M. R., Flotte, T., Gregory, K., Puliafito, C. A., et al. (1991) · Science · doi:10.1126/science.1957169 · Cited by 13,788
Optical Coherence Tomography of the Human Retina
Hee, M. R. (1995) · Archives of Ophthalmology · doi:10.1001/archopht.1995.01100030081025 · Cited by 1,634
Micrometer-Scale Resolution Imaging of the Anterior Eye In Vivo With Optical Coherence Tomography
Izatt, J. A. (1994) · Archives of Ophthalmology · doi:10.1001/archopht.1994.01090240090031 · Cited by 927
Optical biopsy and imaging using optical coherence tomography
Fujimoto, J. G., Brezinski, M. E., Tearney, G. J., Boppart, S. A., Bouma, B., Hee, M. R., Southern, J. F., & Swanson, E. A. (1995) · Nature Medicine · doi:10.1038/nm0995-970 · Cited by 887
Tearney, G. J., Brezinski, M. E., Bouma, B. E., Hee, M. R., Southern, J. F., & Fujimoto, J. G. (1995) · Optics Letters · doi:10.1364/ol.20.002258 · Cited by 562
Measurement of optical properties of biological tissues by low-coherence reflectometry
Schmitt, J. M., Knüttel, A., & Bonner, R. F. (1993) · Applied Optics · doi:10.1364/ao.32.006032 · Cited by 355
Subsurface Imaging of Living Skin with Optical Coherence Microscopy
Schmitt, J. M., Yadlowsky, M. J., & Bonner, R. F. (1995) · Dermatology · doi:10.1159/000246523 · Cited by 251
Optical Coherence Tomography of Central Serous Chorioretinopathy
HEE, M. R., PULIAFITO, C. A., WONG, C., REICHEL, E., DUKER, J. S., SCHUMAN, J. S., SWANSON, E. A., & FUJIMOTO, J. G. (1995) · American Journal of Ophthalmology · doi:10.1016/s0002-9394(14)73760-2 · Cited by 205
Optical coherence tomography of central serous chorioretinopathy
Ju-fen, H. (2010) · Cited by 183
Low-coherence optical tomography in turbid tissue: theoretical analysis
Pan, Y., Birngruber, R., Rosperich, J., & Engelhardt, R. (1995) · Applied Optics · doi:10.1364/ao.34.006564 · Cited by 149
Self-citation
Cited by
Optical coherence tomography: a review of clinical development from bench to bedside
Zysk, A. M., Nguyen, F. T., Oldenburg, A. L., Marks, D. L., & Boppart, S. A. (2007) · Journal of Biomedical Optics · doi:10.1117/1.2793736 · Cited by 547
Optical coherence tomography based angiography [Invited]
Chen, C.-L., & Wang, R. K. (2017) · Biomedical Optics Express · doi:10.1364/boe.8.001056 · Cited by 415
Ultrahigh-resolution full-field optical coherence tomography
Dubois, A., Grieve, K., Moneron, G., Lecaque, R., Vabre, L., & Boccara, C. (2004) · Applied Optics · doi:10.1364/ao.43.002874 · Cited by 394
Lasers in dermatology: Four decades of progress
Tanzi, E. L., Lupton, J. R., & Alster, T. S. (2003) · Journal of the American Academy of Dermatology · doi:10.1067/mjd.2003.582 · Cited by 350
Finger pad friction and its role in grip and touch
Adams, M. J., Johnson, S. A., Lefèvre, P., Lévesque, V., Hayward, V., André, T., & Thonnard, J.-L. (2013) · Journal of The Royal Society Interface · doi:10.1098/rsif.2012.0467 · Cited by 293
Applications of optical coherence tomography in dermatology
GAMBICHLER, T., MOUSSA, G., SAND, M., SAND, D., ALTMEYER, P., & HOFFMANN, K. (2005) · Journal of Dermatological Science · doi:10.1016/j.jdermsci.2005.07.006 · Cited by 283
Optical coherence tomography in dermatology
Sattler, E., Kästle, R., & Welzel, J. (2013) · Journal of Biomedical Optics · doi:10.1117/1.jbo.18.6.061224 · Cited by 283
Self-citationDubois, A., Levecq, O., Azimani, H., Siret, D., Barut, A., Suppa, M., del Marmol, V., Malvehy, J., Cinotti, E., Rubegni, P., et al. (2018) · Journal of Biomedical Optics · doi:10.1117/1.jbo.23.10.106007 · Cited by 269
Advances in Optical Coherence Tomography Imaging for Dermatology
Pierce, M. C., Strasswimmer, J., Hyle Park, B., Cense, B., & de Boer, J. F. (2004) · Journal of Investigative Dermatology · doi:10.1111/j.0022-202x.2004.23404.x · Cited by 257
An introduction to medical imaging with coherent terahertz frequency radiation
Fitzgerald, A. J., Berry, E., Zinovev, N. N., Walker, G. C., Smith, M. A., & Chamberlain, J. M. (2002) · Physics in Medicine and Biology · doi:10.1088/0031-9155/47/7/201 · Cited by 255
Optical coherence tomography in dermatology: technical and clinical aspects
Gambichler, T., Jaedicke, V., & Terras, S. (2011) · Archives of Dermatological Research · doi:10.1007/s00403-011-1152-x · Cited by 240
“Also, dilated blood vessels can be visualised in OCT images of psoriatic skin [8, 109–111].”
BackgroundInfluentialGambichler, T., Matip, R., Moussa, G., Altmeyer, P., & Hoffmann, K. (2006) · Journal of Dermatological Science · doi:10.1016/j.jdermsci.2006.09.008 · Cited by 239
“Depending on the scattering properties of tissue and some accepted loss in resolution, a penetration depth of about 1 mm can be achieved using conventional OCT scanners [19—21].”
BackgroundKnüttel, A., & Boehlau-Godau, M. (2000) · Journal of Biomedical Optics · doi:10.1117/1.429972 · Cited by 235
Dynamic Optical Coherence Tomography in Dermatology
Ulrich, M., Themstrup, L., de Carvalho, N., Manfredi, M., Grana, C., Ciardo, S., Kästle, R., Holmes, J., Whitehead, R., Jemec, G. B. E., et al. (2016) · Dermatology · doi:10.1159/000444706 · Cited by 230
Self-citationGorzelanny, C., Mess, C., Schneider, S. W., Huck, V., & Brandner, J. M. (2020) · Pharmaceutics · doi:10.3390/pharmaceutics12070684 · Cited by 229
“OCT has been introduced in dermatology more than 20 years ago and proven as a useful tool for non-invasive morphological skin analysis [177,178].”
MethodNeerken, S., Lucassen, G. W., Bisschop, M. A., Lenderink, E., & Nuijs, T. (. M. ). (2004) · Journal of Biomedical Optics · doi:10.1117/1.1645795 · Cited by 216
Imaging thermally damaged tissue by Polarization Sensitive Optical Coherence Tomography
De Boer, J., Srinivas, S., Malekafzali, A., Chen, Z., & Nelson, J. (1998) · Optics Express · doi:10.1364/oe.3.000212 · Cited by 211
MOGENSEN, M., & JEMEC, G. B. E. (2007) · Dermatologic Surgery · doi:10.1111/j.1524-4725.2007.33251.x · Cited by 210
MOGENSEN, M., JOERGENSEN, T. M., NÜRNBERG, B. M., MORSY, H. A., THOMSEN, J. B., THRANE, L., & JEMEC, G. B. E. (2009) · Dermatologic Surgery · doi:10.1111/j.1524-4725.2009.01164.x · Cited by 205
Optoacoustic mesoscopy for biomedicine
Omar, M., Aguirre, J., & Ntziachristos, V. (2019) · Nature Biomedical Engineering · doi:10.1038/s41551-019-0377-4 · Cited by 204
OCT imaging of skin cancer and other dermatological diseases
Mogensen, M., Thrane, L., Jørgensen, T. M., Andersen, P. E., & Jemec, G. B. E. (2009) · Journal of Biophotonics · doi:10.1002/jbio.200910020 · Cited by 201
Polarization effects in optical coherence tomography of various biological tissues
de Boer, J. F., Srinivas, S. M., Park, B. H., Pham, T. H., Chen, Z., Milner, T. E., & Nelson, J. S. (1999) · IEEE Journal of Selected Topics in Quantum Electronics · doi:10.1109/2944.796347 · Cited by 200
“OCT has been used to image the eye [4], skin [5]–[7], aero-digestive tracts [8], and cervical dysplasia and carcinoma in situ [9].”
MethodDe Berker, D. A. R., André, J., & Baran, R. (2007) · International Journal of Cosmetic Science · doi:10.1111/j.1467-2494.2007.00372.x · Cited by 195
“It has some potential for examining periungual tissues, but has been little explored as a technique [252].”
BackgroundGambichler, T., Regeniter, P., Bechara, F. G., Orlikov, A., Vasa, R., Moussa, G., Stücker, M., Altmeyer, P., & Hoffmann, K. (2007) · Journal of the American Academy of Dermatology · doi:10.1016/j.jaad.2007.05.029 · Cited by 175
Polarization-sensitive optical coherence tomography of invasive basal cell carcinoma
Strasswimmer, J., Pierce, M. C., Park, B. H., Neel, V., & de Boer, J. F. (2004) · Journal of Biomedical Optics · doi:10.1117/1.1644118 · Cited by 167
Advances in optical coherence tomography in dermatology—a review
Olsen, J., Holmes, J., & Jemec, G. B. E. (2018) · Journal of Biomedical Optics · doi:10.1117/1.jbo.23.4.040901 · Cited by 165
A systematic review of objective burn scar measurements
Lee, K. C., Dretzke, J., Grover, L., Logan, A., & Moiemen, N. (2016) · Burns & Trauma · doi:10.1186/s41038-016-0036-x · Cited by 165
Skin cancer detection using non-invasive techniques
Narayanamurthy, V., Padmapriya, P., Noorasafrin, A., Pooja, B., Hema, K., Firus Khan, A. Y., Nithyakalyani, K., & Samsuri, F. (2018) · RSC Advances · doi:10.1039/c8ra04164d · Cited by 165
Optical coherence tomography in contact dermatitis and psoriasis
Welzel, J., Bruhns, M., & Wolff, H. H. (2003) · Archives of Dermatological Research · doi:10.1007/s00403-003-0390-y · Cited by 164
Self-citationJonathan, E., Enfield, J., & Leahy, M. J. (2011) · Journal of Biophotonics · doi:10.1002/jbio.201000103 · Cited by 164
Applications and future directions for optical coherence tomography in dermatology*
Wan, B., Ganier, C., Du‐Harpur, X., Harun, N., Watt, F. M., Patalay, R., & Lynch, M. D. (2021) · British Journal of Dermatology · doi:10.1111/bjd.19553 · Cited by 148
Using optical coherence tomography to evaluate skin sun damage and precancer
Korde, V. R., Bonnema, G. T., Xu, W., Krishnamurthy, C., Ranger‐Moore, J., Saboda, K., Slayton, L. D., Salasche, S. J., Warneke, J. A., Alberts, D. S., et al. (2007) · Lasers in Surgery and Medicine · doi:10.1002/lsm.20573 · Cited by 145
“These studies have shown that OCT is able to visualize the stratum corneum of glabrous skin in palmoplantar locations, the epidermis and the upper dermis of skin [4], and skin appendages and blood vessels [5].”
BackgroundOptical coherence tomography for the characterization of basal cell carcinoma in vivo: A pilot study
Olmedo, J. M., Warschaw, K. E., Schmitt, J. M., & Swanson, D. L. (2006) · Journal of the American Academy of Dermatology · doi:10.1016/j.jaad.2006.03.013 · Cited by 120
Barton, J. K., Gossage, K. W., Xu, W., Ranger-Moore, J. R., Saboda, K., Brooks, C. A., Duckett, L. D., Salasche, S. J., Warneke, J. A., & Alberts, D. S. (2003) · Technology in Cancer Research & Treatment · doi:10.1177/153303460300200605 · Cited by 85
Line-Field Confocal Optical Coherence Tomography (LC-OCT) for Skin Imaging in Dermatology
Latriglia, F., Ogien, J., Tavernier, C., Fischman, S., Suppa, M., Perrot, J.-L., & Dubois, A. (2023) · Life · doi:10.3390/life13122268 · Cited by 74
“The first use of OCT in dermatology was demonstrated in 1997 [16].”
Background
Metadata
- Authors
- Welzel, J., Lankenau, E., Birngruber, R., & Engelhardt, R.
- Venue
- Journal of the American Academy of Dermatology
- Publisher
- Elsevier BV
- Published
- 1997-12
- Volume
- 37
- Issue
- 6
- Pages
- 958-963
- ISSN
- 0190-9622 (ISSN-L)
- OpenAlex
- W2089018977
- Cited by
- 442
- References
- 11
- FWCI
- 6.94
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.