Detailed Analysis of Insulin Absorption Variability and the Tissue Response to Continuous Subcutaneous Insulin Infusion Catheter Implantation in Swine
Open accessAbstract
BACKGROUND: Worldwide, ∼1 million people manage their type 1 diabetes with an insulin pump and a continuous subcutaneous insulin infusion (CSII) catheter. Patients routinely insert a new catheter every 2-3 days due to increasing variability of insulin absorption over time. Catheter insertion and maintenance damage capillaries, lymphatics, cells, and connective tissue leading to an acute inflammatory response. METHODS: We compared an investigational CSII catheter (IC) and a commercial CSII catheter (CC) regarding insulin absorption pharmacokinetics (PK) and tissue inflammation. The two different catheter designs were implanted into the subcutaneous tissue of six swine for 5 days. Insulin boluses were given on days 1, 3, and 5 of wear-time to assess PK. Tissue around catheters was excised and stained to visualize inflammation and morphological changes of adjacent tissue. RESULTS: values on day 5 of catheter wear-time. Both catheter types produced high intra- and intersubject day-to-day insulin absorption variability. The IC caused significantly more tissue disruption and lead to irregular changes in tissue morphology. Both catheter types were surrounded by a layer of inflammatory tissue that varied in composition, thickness, and density over time. A catheter that was manually inserted by pushing a sharp tip through the skin caused more trauma and variability than a 90° Teflon cannula with automated insertion. CONCLUSIONS: Insulin absorption variability could be attributed to the layer of inflammatory tissue, which may function as a mechanical barrier to insulin flow into adjacent vascular tissue. The impact of the acute inflammatory tissue response on insulin absorption has to be considered in future catheter designs. A catheter that was manually inserted by pushing a sharp tip through the skin caused more trauma and variability than a 90° Teflon cannula with automated insertion.
AI snapshot
Written from the paper’s full text.
TL;DR
In a 5-day swine study, insulin absorption from a commercial CSII catheter was fastest and highest on day 5, while an investigational catheter caused more tissue trauma and more variable absorption. The surrounding inflammatory tissue layer is proposed as a mechanical barrier contributing to day-to-day insulin absorption variability.
Key findings
- Commercial catheters (CCs) produced the fastest insulin absorption on day 5: tmax = 12.0 ± 4.5 min, highest Cmax = 114.7 ± 30.1 mU/L, and highest AUC60 = 54.9 ± 20.8 mU/L·h.
- Both catheter types showed high intra- and intersubject day-to-day variability; for example, IC intersubject CV for AUC was 52.5% on day 1, 82.1% on day 3, and 93.8% on day 5.
- Investigational catheters (ICs) caused significantly more tissue disruption: larger debris field (4.9 ± 1.4 mm vs 3.2 ± 1.1 mm for CCs), more irregular topography, and more reticular fiber disruption, fat necrosis, and hemorrhage at the tip base (P < 0.001 for several comparisons).
- Both catheter types were surrounded by an inflammatory tissue layer that varied in thickness, density, composition, and continuity; the paper concludes this layer may act as a mechanical barrier to insulin flow into adjacent vascular tissue.
Methods
- Observational study in six nondiabetic ambulatory female swine (68.5 ± 3.6 kg); 72 CSII catheters were implanted for 5 days: 36 commercial CSII catheters and 36 investigational CSII catheters.
- Insulin lispro (U-100 Humalog) was infused via pumps; 5-U boluses were given on days 1, 3, and 5, with PK measured as Cmax, tmax, AUC, and AUC60.
- Tissue surrounding catheters was excised after euthanasia, stained with hematoxylin and eosin, Masson’s trichrome, and reticulin, and graded by a blinded surgical pathologist for hemorrhage, reticular fiber disruption, fibrin deposition, fat necrosis, and collagen deposition.
Limitations
- Stated: small sample size; difficulty controlling variables in awake swine such as temperature, activity level, and food intake; many catheters had occlusion alarms or dye/insulin leakage; histology sometimes did not fully localize the insertion channel; no lispro-specific ELISA was commercially available, so insulin was calculated indirectly by subtracting porcine insulin from total insulin, adding to variability.
- Stated: PK data could not be analyzed for statistical significance due to the small number of studies.
- Evident from text: the swine model may not fully represent human inflammatory responses, although the authors argue swine is an adequate model for human wound healing.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
19 of 50 cited sources checked · 3 supported · 0 flagged · accuracy not yet measured
- Premisecitation
When a CSII catheter is inserted, it causes an acute inflammatory response of the adipose tissue.
“When a CSII catheter is inserted, it causes an acute inflammatory response of the adipose tissue.”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [14] source
The Introduction (page not provided) says implantation of biomaterials is followed by acute inflammation and that the response depends partly on the tissue implanted. It does not specifically address CSII catheters or establish an acute response in adipose tissue. (LLM, conf=0.86)
Experimental — accuracy not yet measured
- Premisecitation
The degree of inflammation may influence insulin absorption.
“The degree of inflammation is dependent on catheter material, shape, and wear-time and may influence insulin absorption.”
Quote in the paperSubstantial claim1 of 2 sources checked- cites [14] source
The Introduction describes inflammation following biomaterial implantation, and §2.1 describes how biomaterial surfaces affect inflammatory responses. Neither addresses whether inflammation influences insulin absorption; searches for “insulin absorption” and insulin-delivery catheters found no relevant passage. (LLM, conf=0.73)
Experimental — accuracy not yet measured - cites [17] source
- Premisecitation
Catheter insertion and maintenance damage capillaries, lymphatics, cells, and connective tissue, leading to an acute inflammatory response.
“Catheter insertion and maintenance damage capillaries, lymphatics, cells, and connective tissue leading to an acute inflammatory response.”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [14] source
The Introduction (page unavailable) says that injury to vascularized connective tissue during biomaterial implantation initiates inflammation, including an acute response. It does not specifically address catheter maintenance or say that capillaries, lymphatics, and cells are damaged. (LLM, conf=0.86)
Experimental — accuracy not yet measured
- Premisecitation
Patients routinely insert a new CSII catheter every 2-3 days because insulin absorption into the circulation becomes more variable and less reliable over time.
“Patients routinely insert a new catheter every 2–3 days due to increasing variability of insulin absorption over time.”
- Also stated in: § Introduction · plain
“Patients are instructed to insert a new CSII set every 2 to 3 days because insulin absorption into the circulation becomes more variable and less reliable over time.”
Quote in the paperSubstantial claim2 of 3 sources checked- cites [6] source
The Introduction recommends changing infusion sets every 48–72 hours, and the Discussion says prolonged use appeared to change insulin absorption and reduce glycemic control. But the study measured blood glucose, not variability in insulin absorption, and does not establish that increasing absorption variability is why patients routinely change catheters. (LLM, conf=0.84)
Experimental — accuracy not yet measured - cites [8] Shetty G, 2010, Diabetes Technol Ther source
- cites [9] source
The Discussion recommends changing infusion sets every 2–3 days, and the introductory passage under Abstract says longer use appeared to change insulin absorption and worsen glycemic control. The paper does not establish increasing variability of absorption as the reason patients routinely change catheters; it also gives skin reactions and other adverse effects as reasons for the recommended interval. (LLM, conf=0.85)
Experimental — accuracy not yet measured
- Also stated in: § Introduction · plain
- Premisecitation
Despite recent improvements in CSII catheter design and insertion method, the CSII catheter remains the weak link of an insulin infusion system for the management of type 1 diabetes.
“Despite recent improvements in CSII catheter design and method of insertion, the CSII catheter remains the weak link of an insulin infusion system for the management of type 1 diabetes.”
Quote in the paperSubstantial claim2 of 3 sources checked- cites [5] source
- cites [6] source
The Abstract (page unavailable) reports that infusion-set and injection-site problems began to occur in measurable amounts on day 3 in patients with type 1 diabetes. The Discussion (page unavailable) also describes catheter-related disruptions to insulin delivery. Our copy does not establish that catheter design or insertion methods had recently improved, or compare the catheter with other components to call it the system’s weak link. (LLM, conf=0.84)
Experimental — accuracy not yet measured - cites [7] source
The Abstract and Conclusion describe infusion sets as essential to pump therapy and implicated in many pump-related adverse events, supporting the claim that the catheter remains a vulnerable component. The Conclusion discusses a newer cannula design, but the passages on insertion in “Site Selection and Rotation” describe available techniques rather than establishing that insertion methods have recently improved. (LLM, conf=0.82)
Experimental — accuracy not yet measured
- Premisecitation
The success of insulin pump therapy may be dependent on the robustness of insertion methods to avoid catheter kinking and failed delivery of insulin into the tissue.
“The success of insulin pump therapy may further be dependent on the robustness of insertion methods to avoid issues such as catheter kinking and therefore failed delivery of insulin into the tissue.”
Quote in the paperNot a substantial claim1 of 2 sources checked - Premisecitation
The degree of inflammation is dependent on catheter material, shape, and wear-time.
“The degree of inflammation is dependent on catheter material, shape, and wear-time and may influence insulin absorption.”
Quote in the paperSubstantial claim1 of 2 sources checked- cites [14] source
Section 2.3 says material form and size can influence the foreign body reaction, and the Introduction discusses how inflammatory responses change over time after implantation. The copy does not establish that the degree of inflammation depends specifically on catheter shape or wear-time; its explicit account of the degree of response instead emphasizes implantation injury, tissue site, and provisional matrix formation. No page numbers were supplied. (LLM, conf=0.83)
Experimental — accuracy not yet measured - cites [17] source
- Premisecitation
A layer of inflammatory tissue forms around each CSII cannula that contains plasma proteins, platelets, red blood cells, white blood cells, and cellular debris.
“A layer of inflammatory tissue forms around each CSII cannula that contains plasma proteins, platelets, red blood cells, white blood cells, and cellular debris.”
Quote in the paperSubstantial claim2 of 3 sources checked- cites [14] source
The Introduction (page unavailable) describes a blood-based provisional matrix forming around implanted biomaterials, followed by inflammation at the implant site; the protein-adsorption section identifies host plasma proteins. The paper also discusses platelets and inflammatory white blood cells, but the cited passages do not establish this composition around every CSII cannula or identify red blood cells and cellular debris as components of the layer. (LLM, conf=0.86)
Experimental — accuracy not yet measured - cites [15] source
- cites [16] source
The FBR Overview (page not provided) describes protein adsorption, inflammatory-cell recruitment, and eventual fibrous encapsulation around implanted biomaterials. It does not specifically address CSII cannulas, establish that a layer forms around each one, or describe that layer as containing red blood cells and cellular debris. (LLM, conf=0.85)
Experimental — accuracy not yet measured
- Premisecitation
The mechanisms that cause variable and unreliable insulin absorption from a CSII catheter into the circulation are poorly understood.
“The mechanisms that cause variable and unreliable insulin absorption from a CSII catheter into the circulation are poorly understood.”
Quote in the paperSubstantial claim1 of 4 sources checked- cites [10] source
- cites [11] source
- cites [12] source
The Methods describe intravenous insulin infusion in dogs, and the Discussion concerns insulin moving from blood into the interstitium. Neither addresses variable absorption in the opposite direction—from a CSII catheter into circulation—or whether its mechanisms are poorly understood. Searches for CSII catheter absorption and subcutaneous infusion did not locate a relevant passage in our copy; page numbers were unavailable. (LLM, conf=0.77)
Experimental — accuracy not yet measured - cites [13] source
- Findingown data
The IC caused significantly more tissue disruption than the CC.
“The IC caused significantly more tissue disruption and lead to irregular changes in tissue morphology.”
Quote in the paperSubstantial claim - Findingown data
Both catheter types produced high intra- and intersubject day-to-day insulin absorption variability.
“Both catheter types produced high intra- and intersubject day-to-day insulin absorption variability.”
Quote in the paperSubstantial claim - Findingown data
The IC led to irregular changes in tissue morphology.
“The IC caused significantly more tissue disruption and lead to irregular changes in tissue morphology.”
Quote in the paperSubstantial claim - Findingown data
A catheter that was manually inserted by pushing a sharp tip through the skin caused more trauma and variability than a 90° Teflon cannula with automated insertion.
“A catheter that was manually inserted by pushing a sharp tip through the skin caused more trauma and variability than a 90° Teflon cannula with automated insertion.”
- Also stated in: § Discussion · plain
“A catheter that was manually inserted by pushing a sharp tip through the skin caused more trauma and variability than a 90° Teflon cannula with automated insertion.”
Quote in the paperSubstantial claim- Rests on an assumption: The differences in tissue trauma and insulin absorption variability between the IC and CC are attributable to the catheter insertion method (manual sharp tip vs 90° Teflon cannula with automated insertion) rather than to other differences in catheter material, shape, or design.
- Also stated in: § Discussion · plain
- Findingown data
Both catheter types were surrounded by a layer of inflammatory tissue that varied in composition, thickness, density, and continuity over time.
“Both catheter types were surrounded by a layer of inflammatory tissue that varied in composition, thickness, and density over time.”
- Also stated in: § Tissue histology surrounding a CSII catheter · plain
“The layer of inflammatory tissue surrounding the CC and IC cannula varied in thickness, density, composition, and continuity.”
Quote in the paperSubstantial claim - Also stated in: § Tissue histology surrounding a CSII catheter · plain
- Findingown data
The mean Cmax was higher for insulin administered through the CCs compared to the ICs.
“The mean AUC and Cmax were higher and tmax shorter for insulin administered through the CCs compared to the ICs.”
Quote in the paperSubstantial claim - Findingown data
Five of the six ICs produced a lower AUC on day 5 compared with day 3.
“Five of the six ICs produced a lower AUC on day 5 compared with day 3 (Table 3).”
Quote in the paperNot a substantial claim - Findingown data
Ten of the 36 PK studies resulted in a limited and/or delayed insulin absorption curve.
“Ten of the 36 PK studies resulted in a limited and/or delayed insulin absorption curve.”
Quote in the paperSubstantial claim - Findingown data
Two CCs produced low and delayed insulin absorption on day 1.
“Two CCs produced low and delayed insulin absorption on day 1, followed by higher insulin absorption on subsequent days.”
Quote in the paperNot a substantial claim - Findingown data
Fibroblasts produced a variable amount of fresh collagen that added thickness and density to the surrounding layer of inflammatory tissue.
“Fibroblasts produced a variable amount of fresh collagen (stained light blue in the trichrome stains compared to the darker blue of mature collagen) that added thickness and density to this surrounding layer of inflammatory tissue (Fig. 1A–C, center, Supplementary Figs. S6–S17).”
Quote in the paperNot a substantial claim - Findingown data
The tissue histopathology surrounding the CSII infused with lispro insulin was generally similar to the histology surrounding catheters infused with saline or not infused.
“The tissue histopathology surrounding the CSII infused with lispro insulin had a greater amount of fibrin in the reservoir, but was generally similar to the histology surrounding catheters infused with saline or not infused (Table 2 and Supplementary Table S2).”
Quote in the paperSubstantial claim - Findingown data
Insulin absorption from the CCs was most variable on day 1.
“Insulin absorption from the CCs and was least variable on day 5 and most variable on day 1 (Fig. 2 and Table 4).”
Quote in the paperNot a substantial claim - Findingown data
The mean AUC was higher for insulin administered through the CCs compared to the ICs.
“The mean AUC and Cmax were higher and tmax shorter for insulin administered through the CCs compared to the ICs.”
Quote in the paperSubstantial claim - Findingown data
Cells, capillaries, lymph vessels, connective tissues (collagen, elastin, reticular fibers), and ECM were damaged along the path of cannula insertion.
“Cells, capillaries, lymph vessels, connective tissues (collagen, elastin, reticular fibers), and ECM were damaged along the path of cannula insertion.”
Quote in the paperSubstantial claim - Findingown data
The tissue surrounding the CSII cannula on day 5 of implantation resembled a clean surgical wound with a foreign body, rather than normal dermis and subcutaneous tissue.
“In general, the tissue surrounding the CSII cannula on day 5 of implantation resembled a clean surgical wound with a foreign body, rather than normal dermis and subcutaneous tissue (Supplementary Fig. S3).”
Quote in the paperNot a substantial claim - Findingown data
No other CC used for the PK study developed an occlusion alarm.
“No other CC used for the PK study developed an occlusion alarm.”
Quote in the paperNot a substantial claim - Findingown data
Thrombus formed around many of the cannulas due to damaged arterioles, capillaries, and venules.
“Thrombus formed around many of the cannulas due to damaged arterioles, capillaries, and venules (Fig. 1A center).”
Quote in the paperNot a substantial claim - Findingown data
In the other five CCs, on day 5, AUC60 was 54.9 ± 20.8 mU/L·h.
“The other five CCs produced the fastest insulin absorption (tmax = 12.0 ± 4.5 min), the highest plasma insulin concentration (Cmax = 114.7 ± 30.1 mU/L), and the highest AUC60 (54.9 ± 20.8 mU/L·h) on day 5 of implantation.”
Quote in the paperNot a substantial claim - Findingown data
Dye leakage and occlusion alarms occurred in CCs that were infused with insulin lispro, saline, and not infused.
“Dye leakage and occlusion alarms occurred in CCs that were infused with insulin lispro, saline, and not infused (Supplementary Table S5).”
Quote in the paperNot a substantial claim - Findingown data
No statistical relationship was detected between the observed histology results and insulin absorption PK.
“We were unable to detect a statistical relationship between the observed histology results and insulin absorption PK due to the small sample size and high variability.”
Quote in the paperSubstantial claim - Findingreasoning
The commercial and ICs produced distinct patterns of insulin absorption.
“The commercial and ICs had distinct designs, materials, and methods of insertion that produced a significant difference in the inflammatory tissue response and distinct patterns of insulin absorption.”
Quote in the paperSubstantial claim - Findingown data
The CCs elicited significantly less reticular fiber disruption than the ICs.
“The CCs elicited significantly less reticular fiber disruption, fat necrosis, hemorrhage at the cannula tip, a regular tissue topography, and a smaller debris field compared with the ICs.”
Quote in the paperSubstantial claim - Findingown data
The tissue histopathology surrounding the CSII infused with lispro insulin had a greater amount of fibrin in the reservoir than catheters infused with saline or not infused.
“The tissue histopathology surrounding the CSII infused with lispro insulin had a greater amount of fibrin in the reservoir, but was generally similar to the histology surrounding catheters infused with saline or not infused (Table 2 and Supplementary Table S2).”
Quote in the paperNot a substantial claim - Findingown data
The six ICs produced the lowest AUC on day 5.
“The six ICs produced the highest AUC60, AUC, and Cmax on day 1, the fastest insulin absorption on day 3, and the lowest AUC60, AUC, and Cmax on day 5.”
Quote in the paperNot a substantial claim - Findingown data
The CCs elicited a regular tissue topography more often than the ICs.
“The CCs elicited significantly less reticular fiber disruption, fat necrosis, hemorrhage at the cannula tip, a regular tissue topography, and a smaller debris field compared with the ICs.”
Quote in the paperSubstantial claim - Findingown data
The six ICs produced the highest Cmax on day 1.
“The six ICs produced the highest AUC60, AUC, and Cmax on day 1, the fastest insulin absorption on day 3, and the lowest AUC60, AUC, and Cmax on day 5.”
Quote in the paperNot a substantial claim - Findingown data
The ICs of swine 1, 2, 3, and 4 used for the PK study produced one or more occlusion alarms.
“The ICs of swine 1, 2, 3, and 4 used for the PK study produced one or more occlusion alarms.”
Quote in the paperNot a substantial claim - Findingown data
In the other five CCs, on day 5, tmax was 12.0 ± 4.5 min.
“The other five CCs produced the fastest insulin absorption (tmax = 12.0 ± 4.5 min), the highest plasma insulin concentration (Cmax = 114.7 ± 30.1 mU/L), and the highest AUC60 (54.9 ± 20.8 mU/L·h) on day 5 of implantation.”
Quote in the paperNot a substantial claim - Findingown data
Several specimens showed a downward migration of epidermis/dermis into the insertion channel.
“Several specimens showed a downward migration of epidermis/dermis into the insertion channel (Supplementary Figs. S8–S10).”
Quote in the paperNot a substantial claim - Findingown data
Insulin absorption from the CCs was least variable on day 5.
“Insulin absorption from the CCs and was least variable on day 5 and most variable on day 1 (Fig. 2 and Table 4).”
Quote in the paperNot a substantial claim - Findingown data
In the other five CCs, on day 5, Cmax was 114.7 ± 30.1 mU/L.
“The other five CCs produced the fastest insulin absorption (tmax = 12.0 ± 4.5 min), the highest plasma insulin concentration (Cmax = 114.7 ± 30.1 mU/L), and the highest AUC60 (54.9 ± 20.8 mU/L·h) on day 5 of implantation.”
Quote in the paperNot a substantial claim - Findingown data
Insulin absorption from the ICs was most variable on day 3 (range 9-132 mU/L*h, intersubject CV = 93.8%).
“Insulin absorption from the ICs was least variable (range 34–124 mU/L*h, intersubject CV = 52.5%) on day 1 and most variable on day 3 (range 9–132 mU/L*h, intersubject CV = 93.8%).”
Quote in the paperNot a substantial claim - Findingown data
The CCs had a smaller debris field than the ICs.
“The CCs elicited significantly less reticular fiber disruption, fat necrosis, hemorrhage at the cannula tip, a regular tissue topography, and a smaller debris field compared with the ICs.”
- Also stated in: § Tissue histology surrounding a CSII catheter · plain
“The ICs produced a larger debris field with irregular topography of tissue planes and irregular reservoir geometry (P < 0.001).”
Quote in the paperNot a substantial claim - Also stated in: § Tissue histology surrounding a CSII catheter · plain
- Findingown data
The six ICs produced the highest AUC on day 1.
“The six ICs produced the highest AUC60, AUC, and Cmax on day 1, the fastest insulin absorption on day 3, and the lowest AUC60, AUC, and Cmax on day 5.”
Quote in the paperNot a substantial claim - Findingown data
The individual PK curves are highly variable in contrast with the average PK curves.
“The individual PK curves are highly variable in contrast with the average PK curves.”
Quote in the paperSubstantial claim - Findingown data
Insulin absorption from the ICs was least variable on day 1 (range 34-124 mU/L*h, intersubject CV = 52.5%).
“Insulin absorption from the ICs was least variable (range 34–124 mU/L*h, intersubject CV = 52.5%) on day 1 and most variable on day 3 (range 9–132 mU/L*h, intersubject CV = 93.8%).”
Quote in the paperNot a substantial claim - Findingown data
The six ICs produced the fastest insulin absorption on day 3.
“The six ICs produced the highest AUC60, AUC, and Cmax on day 1, the fastest insulin absorption on day 3, and the lowest AUC60, AUC, and Cmax on day 5.”
Quote in the paperNot a substantial claim - Findingown data
In general, the ICs produced a similar, yet more pronounced, tissue response characterized by a central core of adipose tissue debris surrounded by a layer of fibrin thrombus infiltrated with inflammatory cells.
“In general, the ICs produced a similar, yet more pronounced, tissue response characterized by a central core of adipose tissue debris surrounded by a layer of fibrin thrombus infiltrated with inflammatory cells.”
Quote in the paperSubstantial claim - Findingown data
Neutrophils, macrophages, fibroblasts, and lymphocytes migrated into the areas of tissue damage from the local capillaries.
“Neutrophils, macrophages, fibroblasts, and lymphocytes migrated into the areas of tissue damage from the local capillaries (dense purple and pink areas within subcutaneous tissue in hematoxylin and eosin sections, Fig. 1A–D, Supplementary Figs. S4–S15).”
Quote in the paperNot a substantial claim - Findingown data
Those two CCs had higher insulin absorption on subsequent days.
“Two CCs produced low and delayed insulin absorption on day 1, followed by higher insulin absorption on subsequent days.”
Quote in the paperNot a substantial claim - Findingown data
Methylene blue dye often accumulated in the hub indicating a technical obstruction of insulin flow.
“Methylene blue dye often accumulated in the hub indicating a technical obstruction of insulin flow.”
Quote in the paperSubstantial claim - Findingown data
The six ICs produced the lowest AUC60 on day 5.
“The six ICs produced the highest AUC60, AUC, and Cmax on day 1, the fastest insulin absorption on day 3, and the lowest AUC60, AUC, and Cmax on day 5.”
Quote in the paperNot a substantial claim - Findingown data
No other CC leaked dye onto the skin surface (35/36).
“No other CC leaked dye onto the skin surface (35/36).”
Quote in the paperNot a substantial claim - Findingown data
All of the specimens had regions of adipose cell necrosis and acute inflammation.
“All of the specimens had regions of adipose cell necrosis (Fig. 1A–D) and acute inflammation.”
Quote in the paperNot a substantial claim - Findingown data
The tissue response around CCs was characterized by a central area of adipose tissue debris surrounded by a layer of thrombus infiltrated with immune cells.
“The tissue response around CCs was characterized by a central area of adipose tissue debris surrounded by a layer of thrombus infiltrated with immune cells.”
Quote in the paperNot a substantial claim - Findingown data
Insulin absorption (PK) in this swine study was variable when insulin was infused through either type of CSII catheter on days 1, 3, and 5.
“Insulin absorption (PK) in this swine study was variable when insulin was infused through either type of CSII catheter on days 1, 3, and 5.”
Quote in the paperSubstantial claim - Findingown data
All of the CCs and ICs produced a high intrasubject and intersubject day-to-day PK variability.
“All of the CCs and ICs produced a high intrasubject and intersubject day to day PK variability (Table 4).”
Quote in the paperSubstantial claim - Findingown data
Thrombus and immune cells produced a layer of fibrin and acute inflammatory tissue that partially or completely surrounded the implanted cannula.
“Thrombus and immune cells produced a layer of fibrin and acute inflammatory tissue that partially or completely surrounded the implanted cannula (Fig. 1A, B, center, Supplementary Figs. S6–S17).”
Quote in the paperSubstantial claim - Findingown data
Four CCs had a pump occlusion alarm (11%).
“One of the CCs leaked a large amount of dye into the hub (3%), five CCs leaked a small amount into the hub (14%), and four CCs had a pump occlusion alarm (11%).”
Quote in the paperNot a substantial claim - Findingown data
The ICs produced irregular reservoir geometry.
“The ICs produced a larger debris field with irregular topography of tissue planes and irregular reservoir geometry (P < 0.001).”
Quote in the paperNot a substantial claim - Findingown data
The six ICs produced the lowest Cmax on day 5.
“The six ICs produced the highest AUC60, AUC, and Cmax on day 1, the fastest insulin absorption on day 3, and the lowest AUC60, AUC, and Cmax on day 5.”
Quote in the paperNot a substantial claim - Findingown data
The CCs elicited significantly less hemorrhage at the cannula tip than the ICs.
“The CCs elicited significantly less reticular fiber disruption, fat necrosis, hemorrhage at the cannula tip, a regular tissue topography, and a smaller debris field compared with the ICs.”
Quote in the paperSubstantial claim - Findingown data
The CCs elicited significantly less fat necrosis than the ICs.
“The CCs elicited significantly less reticular fiber disruption, fat necrosis, hemorrhage at the cannula tip, a regular tissue topography, and a smaller debris field compared with the ICs.”
Quote in the paperSubstantial claim - Findingown data
The six ICs produced the highest AUC60 on day 1.
“The six ICs produced the highest AUC60, AUC, and Cmax on day 1, the fastest insulin absorption on day 3, and the lowest AUC60, AUC, and Cmax on day 5.”
Quote in the paperNot a substantial claim - Findingown data
The mean tmax was shorter for insulin administered through the CCs compared to the ICs.
“The mean AUC and Cmax were higher and tmax shorter for insulin administered through the CCs compared to the ICs.”
Quote in the paperSubstantial claim - Findingown data
The ICs produced irregular topography of tissue planes.
“The ICs produced a larger debris field with irregular topography of tissue planes and irregular reservoir geometry (P < 0.001).”
Quote in the paperSubstantial claim - Findingown data
The commercial and ICs had distinct designs, materials, and methods of insertion that produced a significant difference in the inflammatory tissue response.
“The commercial and ICs had distinct designs, materials, and methods of insertion that produced a significant difference in the inflammatory tissue response and distinct patterns of insulin absorption.”
Quote in the paperSubstantial claim - Findingnone given
One CC kinked below the skin, leading to inadequate insulin delivery.
“One CC kinked below the skin, leading to inadequate insulin delivery.”
Quote in the paperNot a substantial claim - Findingown data
We found an increase in intra- and interindividual variability of insulin absorption over catheter wear-time.
“In conclusion, we found substantial differences in the inflammatory response to two different catheter shapes and materials and an increase in intra- and interindividual variability of insulin absorption over catheter wear-time.”
Quote in the paperSubstantial claim - Findingown data
We found substantial differences in the inflammatory response to two different catheter shapes and materials.
“In conclusion, we found substantial differences in the inflammatory response to two different catheter shapes and materials and an increase in intra- and interindividual variability of insulin absorption over catheter wear-time.”
Quote in the paperSubstantial claim - Findingnone given
Issues with the catheter hub can lead to physical obstructions of insulin flow into the catheter and tissue.
“We saw that issues with the catheter hub can lead to physical obstructions of insulin flow into the catheter and tissue.”
Quote in the paperNot a substantial claim - Interpretationreasoning
The layer of inflammatory tissue may function as a mechanical barrier to insulin flow into adjacent vascular tissue.
“Insulin absorption variability could be attributed to the layer of inflammatory tissue, which may function as a mechanical barrier to insulin flow into adjacent vascular tissue.”
Quote in the paperSubstantial claim - Interpretationreasoning
Insulin absorption variability could be attributed to the layer of inflammatory tissue.
“Insulin absorption variability could be attributed to the layer of inflammatory tissue, which may function as a mechanical barrier to insulin flow into adjacent vascular tissue.”
Quote in the paperSubstantial claim - Study design and conductown data
Data for swine 1 were excluded before PK analysis due to faulty insertion (kinking) causing inadequate penetration into the subcutaneous tissue, leading to limited insulin absorption.
“Data for swine 1 were excluded before PK analysis, due to faulty insertion (kinking) causing inadequate penetration into the subcutaneous tissue, leading to limited insulin absorption.”
Quote in the paperNot a substantial claim - Study design and conductown data
The PK data could not be analyzed for statistical significance due to the small number of studies.
“However, these data could not be analyzed for statistical significance due to the small number of studies.”
Quote in the paperNot a substantial claim - Interpretationnone given
This is the first study comparing insulin PK and tissue histopathology surrounding two different types of CSII catheters in the same swine while infusing insulin, saline, or no infusion.
“To our knowledge, this is the first study comparing insulin PK and tissue histopathology surrounding two different types of CSII catheters in the same swine while infusing insulin, saline, or no infusion.”
Quote in the paperSubstantial claim - Interpretationown data
The decreased uptake of insulin when administered through the IC may primarily be explained by issues in the connector between catheter and tubing.
“The decreased uptake of insulin when administered through the IC may primarily be explained by issues in the connector between catheter and tubing.”
Quote in the paperSubstantial claim - Interpretationreasoning
The shape of the IC possibly contributed substantially to the extent of trauma due to its sharp tip, which may have continuously disrupted tissue in the vicinity.
“The shape of the IC possibly contributed substantially to the extent of trauma due to its sharp tip, which may have continuously disrupted tissue in the vicinity.”
Quote in the paperNot a substantial claim - Premisecitation
The swine has been described as the most adequate model for human wound healing.
“Although the inflammatory response may be different from humans, the swine has been described as the most adequate model for human wound healing.26–29”
Quote in the paperNot a substantial claim0 of 4 sources checked - Interpretationreasoning
The indirect calculation of insulin concentrations adds to the calculated intra- and intersubject variability.
“Unfortunately, this indirect calculation adds to the calculated intra- and intersubject variability.”
Quote in the paperNot a substantial claim - Interpretationreasoning
This high variability may be explained by the inconsistency of the mechanical inflammatory layer surrounding the cannula of a CSII catheter.
“This high variability may be explained by the inconsistency of the mechanical inflammatory layer surrounding the cannula of a CSII catheter.”
Quote in the paperSubstantial claim - Interpretationcitation
The intrapatient CV of 37% for Cmax and 39% for tmax are comparable with recently published human data.
“The intrapatient CV of 37% for Cmax and 39% for tmax are comparable with recently published human data.32”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [32] source
- May combine two claims. 25
- The quote only partly states this claim. 2
- Classified as probably not a substantial claim. 20
May combine two claims.
PremisecitationPatients commonly experience an insulin bolus that fails to produce the expected glucose lowering effect, leading to persistent hyperglycemia.
“Patients commonly experience an insulin bolus that fails to produce the expected glucose lowering effect, leading to persistent hyperglycemia.6,19”
Quote in the paperSubstantial claim2 of 2 sources checked- cites [6] source
The Discussion (page unavailable) says disrupted pump delivery or prolonged catheter use can impair insulin absorption, and reports rising mean blood glucose with longer catheter use. It does not establish that a bolus commonly fails to lower glucose as expected or that this leads to persistent hyperglycemia. (LLM, conf=0.82)
Experimental — accuracy not yet measured - cites [19] Patel PJ, 2013, Diabetes Technol Ther source
May combine two claims.
InterpretationcitationThis observation is consistent with published animal studies, human clinical trials, and routine clinical care.
“This observation is consistent with published animal studies, human clinical trials, and routine clinical care.5,6,22,23”
Quote in the paperNot a substantial claim1 of 4 sources checkedMay combine two claims.
InterpretationcitationInsulin may flow into the connective tissue septa that surround the adipose cells, remain within the surrounding layer of inflammatory tissue, and/or flow upward along the cannula shaft onto the skin surface.
“Insulin may flow into the connective tissue septa that surround the adipose cells, remain within the surrounding layer of inflammatory tissue, and/or flow upward along the cannula shaft onto the skin surface.30,31”
Quote in the paperSubstantial claim1 of 2 sources checked- cites [30] source
The Results (page not available) say injected insulin distributes between fat lobules, and the Introduction describes connective tissue separating those lobules. The Discussion mentions back-flow through the injection channel. The paper does not describe insulin remaining in a surrounding layer of inflammatory tissue or flowing along a cannula shaft onto the skin surface. (LLM, conf=0.84)
Experimental — accuracy not yet measured - cites [31] source
May combine two claims.
InterpretationcitationThe acute inflammatory response and the infusion of insulin may directly increase local blood flow, which may contribute to increased insulin uptake on day 5 of catheter wear-time.
“Furthermore, the acute inflammatory response and the infusion of insulin may directly increase local blood flow,33–38 which may contribute to increased insulin uptake on day 5 of catheter wear-time.”
Quote in the paperNot a substantial claim2 of 6 sources checked- cites [33] source
- cites [34] source
- cites [35] source
- cites [36] source
The Abstract (page unavailable) reports that inserting a dermal microdialysis probe significantly increased skin blood flow, supporting the proposed link between insertion trauma and local perfusion. It does not address insulin infusion, insulin uptake, or day 5 of catheter wear-time. (LLM, conf=0.84)
Experimental — accuracy not yet measured - cites [37] source
The Conclusions propose that increased blood flow around an aging infusion site could explain earlier peak insulin action. They do not identify an acute inflammatory response or insulin infusion as the direct causes of that blood flow. The Methods and Results compare day 1 with day 4, not day 5; however, the day-5 uptake finding appears to be the citing authors’ application of the proposed mechanism, rather than a finding attributed to this source. (LLM, conf=0.84)
Experimental — accuracy not yet measured - cites [38] source
May combine two claims.
InterpretationcitationSlow increases and abrupt decreases in insulin pump back pressure suggest that a bolus dose of insulin may accumulate within the layer of inflammatory tissue, distend the tissue, and then abruptly travel into adjacent tissue through a pathway with lower resistance.
“Slow increases and abrupt decreases in insulin pump back pressure suggest that a bolus dose of insulin may accumulate within the layer of inflammatory tissue, distend the tissue, and then abruptly travel into adjacent tissue through a pathway with lower resistance.30”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [30] source
Split into 2 atoms — 0 confirmed, 1 partially supported, 0 disputed, 0 misattributed, 0 unverifiable, 1 not checked. Expand the claim to see per-atom verdicts.
Experimental — accuracy not yet measured
May combine two claims.
Findingown dataThe CC had the highest Cmax and fastest tmax values on day 5 of catheter wear-time.
“Insulin absorption was better when infused through a CC with highest Cmax and fastest tmax values on day 5 of catheter wear-time.”
Quote in the paperSubstantial claimThe quote only partly states this claim.
Findingown dataInsulin absorption was better when infused through a CC than through an IC.
“Insulin absorption was better when infused through a CC with highest Cmax and fastest tmax values on day 5 of catheter wear-time.”
Quote in the paperSubstantial claimMay combine two claims.
Findingnone givenInsulin absorption from the CCs was least variable on day 5 and most variable on days 1 and 3.
“Insulin absorption from the CCs was least variable on day 5 and most variable on days 1 and 3.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataMany of the specimens had a continuous or near-continuous layer of inflammatory tissue surrounding the CSII cannula, while other specimens had a discontinuous layer with normal adipose tissue adjacent to the CSII cannula.
“Many of the specimens had a continuous or near-continuous layer of inflammatory tissue surrounding the CSII cannula, while other specimens had a discontinuous layer with normal adipose tissue adjacent to the CSII cannula.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataThe IC of swine 1, 2, and 6 leaked dye into the CSII hub and produced limited insulin absorption on days 3 and/or 5.
“The IC of swine 1, 2, and 6 leaked dye into the CSII hub and produced limited insulin absorption on days 3 and/or 5 (Fig. 2D).”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataFive CCs leaked a small amount of dye into the hub (14%).
“One of the CCs leaked a large amount of dye into the hub (3%), five CCs leaked a small amount into the hub (14%), and four CCs had a pump occlusion alarm (11%).”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Findingnone givenMany CCs and ICs produced one or more pump occlusion alarms, leaked dye/insulin into the hub or onto the skin surface.
“Many CCs and ICs produced one or more pump occlusion alarms, leaked dye/insulin into the hub or onto the skin surface.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataThe CC of swine 1 used for the PK study did not insert fully into the subcutaneous tissue, kinked, leaked a large amount of dye onto the skin surface, and produced very limited insulin absorption into the circulation.
“The CC of swine 1 used for the PK study did not insert fully into the subcutaneous tissue, kinked, leaked a large amount of dye onto the skin surface, and produced very limited insulin absorption into the circulation (Fig. 2B).”
Quote in the paperNot a substantial claimMay combine two claims.
FindingreasoningCompared with the tissue response to IC implantation, the tissue layer surrounding the Teflon cannula had more regular connective tissue topography, more regular reservoir geometry, a smaller debris field, less reticular fiber disruption, less fat necrosis, and less hemorrhage within the reservoir.
“Compared with the tissue response to IC implantation, the tissue layer surrounding the Teflon cannula had more regular connective tissue topography, more regular reservoir geometry, a smaller debris field, less reticular fiber disruption, less fat necrosis, and less hemorrhage within the reservoir.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataThe CC of swine 2 used for the PK study produced a pump occlusion alarm on day 3 only.
“The CC of swine 2 used for the PK study produced a pump occlusion alarm on day 3 only.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataNo difference in tissue response was observed between infusion types, except for increased fibrin when insulin was infused.
“We did not observe a difference in tissue response between infusion types, except for increased fibrin when insulin was infused.”
Quote in the paperSubstantial claimThe quote only partly states this claim.
Findingown dataThere was no statistically significant difference in fibrin or collagen deposition at the cannula tip between CCs and ICs.
“There was no statistically significant difference in fibrin or collagen deposition at the cannula tip.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingnone givenThe ICs produced the highest AUC, AUC60, and Cmax on day 1, the fastest tmax on day 3, and the lowest AUC, AUC60, and Cmax on day 5.
“In contrast, the ICs produced the highest AUC, AUC60, and Cmax on day 1, the fastest tmax on day 3, and the lowest AUC, AUC60, and Cmax on day 5.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataThe CCs produced the fastest insulin absorption, the highest plasma insulin concentration, and the greatest AUC60 on day 5 of implantation.
“Of interest, the CCs produced the fastest insulin absorption, the highest plasma insulin concentration, and the greatest AUC60 on day 5 of implantation.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataThe three ICs that produced low insulin absorption on day 5 had a pump occlusion alarm and/or leaked dye into the hub.
“The three ICs that produced low insulin absorption on day 5 had a pump occlusion alarm and/or leaked dye into the hub.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataOne of the CCs leaked a large amount of dye into the hub (3%).
“One of the CCs leaked a large amount of dye into the hub (3%), five CCs leaked a small amount into the hub (14%), and four CCs had a pump occlusion alarm (11%).”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataResults from this study indicate faster and better absorption of insulin on day 5, with a fairly high interindividual CV of 41% for the AUC of the commercial and 94% for the investigational catheter.
“Results from this study indicate faster and better absorption of insulin on day 5, however, with a fairly high interindividual CV of 41% for the AUC of the commercial and 94% for the investigational catheter.”
Quote in the paperSubstantial claimMay combine two claims.
Findingnone givenMaximum insulin plasma concentration levels did not change or increased with wear-time.
“Maximum insulin plasma concentration levels did not change or increased with wear-time.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductown dataAUC for the CC in swine 6 could not be calculated because blood samples 20 min after bolus administration could not be collected for a period of 1.5 h in this animal.
“We furthermore could not calculate the AUC for the CC in swine 6, as we were unable to collect blood samples 20 min after bolus administration for a period of 1.5 h in this animal.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim.
Study design and conductown dataThe order of catheter type for the 5-U insulin bolus was randomized.
“We then infused a 5-U bolus of insulin into a CC or an IC in a randomized order.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe pathologist remained blinded throughout the study.
“The pathologist remained blinded throughout the study.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenBlood glucose concentration was measured in duplicate using the YSI analyzer.
“The concentration of blood glucose (BG) was measured in duplicate using the YSI analyzer.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductnone givenThe tissue histology grading system was newly established by the lead pathologist, who predefined the criteria before analyzing all samples.
“This grading system was newly established by the lead pathologist of this study who predefined these criteria before analyzing all samples.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim.
Study design and conductown dataSubcutaneous tissue glucose sensors were calibrated using CVC blood samples and a reference glucose analyzer.
“The subcutaneous tissue glucose concentration was monitored using commercial glucose sensors (SEVEN PLUS; Dexcom, Inc.) calibrated using CVC blood samples and a reference glucose analyzer (YSI-2300STAT Plus™; Yellow Springs Instruments).”
Quote in the paperSubstantial claimClassified as probably not a substantial claim.
Study design and conductnone givenCSII catheters were carefully inspected to confirm correct insertion through the skin and into the subcutaneous tissue.
“CSII catheters were carefully inspected to confirm correct insertion through the skin and into the subcutaneous tissue.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe study protocol was approved by the Institutional Animal Care & Use Committee of Thomas Jefferson University.
“The protocol was approved by the Institutional Animal Care & Use Committee of Thomas Jefferson University.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim. May combine two claims.
Study design and conductnone givenA surgical pathologist analyzed the tissue specimens using an objective grading system with each swine acting as its own control.
“A surgical pathologist analyzed the tissue specimens using an objective grading system with each swine acting as its own control.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
InterpretationreasoningLysed adipose cells would have released their contents (cytoplasm, nucleus, organelles, and triglycerides) into the wound.
“Although not visualized, lysed adipose cells would have released their contents (cytoplasm, nucleus, organelles, and triglycerides) into the wound.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenHistology did not always provide a complete picture of the tissue surrounding the CSII cannula, as localization of the full insertion channel was sometimes difficult.
“In addition, histology did not always provide a complete picture of the tissue surrounding the CSII cannula, as localization of the full insertion channel was sometimes difficult.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim. May combine two claims.
Study design and conductnone givenAt the time of plasma insulin measurement, no lispro-specific ELISA was commercially available and insulin concentrations were calculated indirectly by subtracting porcine insulin from total insulin content.
“At the time of plasma insulin measurement, no lispro-specific ELISA was commercially available and insulin concentrations were calculated indirectly by subtracting porcine insulin from total insulin content.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationSome patients successfully wear CSII catheters up to 7 days.
“Studies have shown that some patients indeed successfully wear CSII catheters up to 7 days.19,24”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [19] Patel PJ, 2013, Diabetes Technol Ther
Not checked in the free run.
- cites [24] source
May combine two claims.
Study design and conductnone givenThe study was limited by the small sample size and the difficulty in controlling variables that may affect insulin absorption into the subcutaneous vascular tissue in awake swine, such as temperature, activity level, and food intake.
“The study was limited by the small sample size and the difficulty in controlling variables that may affect insulin absorption into the subcutaneous vascular tissue in awake swine, such as temperature, activity level, and food intake.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationLike humans, swine has a thick epidermis (70-150 μm), sparse body hair, and a similar composition of the skin lipid layer.
“Like humans it has a thick epidermis (70–150 μm), sparse body hair, and a similar composition of the skin lipid layer.26,29”
Quote in the paperSubstantial claim0 of 2 sources checkedProbably not a substantial claim: may not pass S3 (load-bearing).
InterpretationreasoningInsulin absorption can further be negatively influenced by insulin traveling upward onto the skin due to an obstruction of insulin flow into the adjacent tissue.
“Insulin absorption can further be negatively influenced by insulin traveling upward onto the skin due to an obstruction of insulin flow into the adjacent tissue.”
Quote in the paperNot a substantial claimMay combine two claims.
Study design and conductnone givenThe clam-shell mechanism of the IC was sometimes difficult to handle and the insertion angle of the cannula was not consistent among samples.
“The clam-shell mechanism of the IC was sometimes difficult to handle and the insertion angle of the cannula was not consistent among samples.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
InterpretationreasoningWhen unnoticed by the patient, blood glucose control may become challenging.
“When unnoticed by the patient, blood glucose control may become challenging.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationThe subcutaneous adipose tissue of the swine has a similarly high elastic and collagen fiber content and the animals show a healing mechanism comparable to humans.
“The subcutaneous adipose tissue of the swine has a similarly high elastic and collagen fiber content and the animals show a healing mechanism comparable to humans.26,27,29”
Quote in the paperSubstantial claim0 of 3 sources checked
The paper relies on this without support.
The tissue response and insulin absorption pharmacokinetics observed in nondiabetic swine are representative of those in humans with type 1 diabetes using CSII.
“Detailed Analysis of Insulin Absorption Variability and the Tissue Response to Continuous Subcutaneous Insulin Infusion Catheter Implantation in Swine”
“The study was performed in six nondiabetic ambulatory female swine (68.5 ± 3.6 kg).”
The paper relies on this without support.
The differences in tissue trauma and insulin absorption variability between the IC and CC are attributable to the catheter insertion method (manual sharp tip vs 90° Teflon cannula with automated insertion) rather than to other differences in catheter material, shape, or design.
“The CC had a stable plastic platform with an adhesive base for attachment to the skin. The 6 mm Teflon cannula was inserted at a 90° angle through the skin and subcutaneous tissue over a sharp 27-gauge steel needle. Insertion force was applied to the catheter using a spring-loaded introducer.”
“The IC had a stable plastic platform with an adhesive base for attachment to the skin. The platform was an open clam-shell design with a side hinge. Closing the platform manually guided the curved cannula through the skin and into the subcutaneous tissue at an insertion angle of ∼60°.”
The paper relies on this without support.
Patients are currently educated to replace their CCs every 2 to 3 days to minimize the risk of infection and because insulin absorption from the subcutaneous tissue into the circulation becomes more variable and less reliable over time.
“Patients are currently educated to replace their CCs every 2 to 3 days to minimize the risk of infection and because insulin absorption from the subcutaneous tissue into the circulation becomes more variable and less reliable over time.”
The paper relies on this without support.
Current recommendations do not recommend the patient to wear a CSII set until day 5.
“The day 5 observations are of interest, as current recommendations do not recommend the patient to wear a CSII set until day 5, when insulin absorption may recover.”
The paper relies on this without support.
Both manual and supported insertions are prone to insertion failure and the optimal insertion angle is not always achieved.
“Both manual and supported insertions are prone to insertion failure and the optimal insertion angle is not always achieved.”
The paper relies on this without support.
When insulin is infused, an insulin pump produces a hydrostatic pressure differential between the inside of the CSII cannula and the adjacent subcutaneous tissue, causing insulin to flow into the tissue along the path of least resistance.
“When insulin is infused, an insulin pump produces a hydrostatic pressure differential between the inside of the CSII cannula and the adjacent subcutaneous tissue, causing insulin to flow into the tissue along the path of least resistance.”
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Overall Cmax comparison is reversed
MajorThe mean of the three reported daily Cmax values is 93.1 mU/L for CCs and about 113.4 mU/L for ICs. Thus, the reported Cmax values do not support the statement that mean Cmax was higher for CCs.
“The mean AUC and Cmax were higher and tmax shorter for insulin administered through the CCs compared to the ICs.”
“| CC (n = 5a swine) | 5 | 71.1 ± 29.4 | 54.9 ± 20.8 | 114.7 ± 30.1 | 12.0 ± 4.5 | | IC (n = 6 swine) | 1 | 68.5 ± 36.0 | 44.3 ± 23.2 | 120.0 ± 59.3 | 43.3 ± 28.8 | | IC (n = 6 swine) | 3 | 63.1 ± 51.8 | 42.7 ± 40.0 | 115.0 ± 86.2 | 20.0 ± 6.3 | | IC (n = 6 swine) | 5 | 40.1 ± 37.6 | 29.0 ± 31.3 | 105.1 ± 87.7 | 33.3 ± 18.6 |”
IC variability is assigned to the wrong day
MajorThe text calls day 3 the most variable for IC AUC and gives a CV of 93.8%, but Table 4 assigns 82.1% to day 3 and 93.8% to day 5.
“Insulin absorption from the ICs was least variable (range 34–124 mU/L*h, intersubject CV = 52.5%) on day 1 and most variable on day 3 (range 9–132 mU/L*h, intersubject CV = 93.8%).”
“| Between animals | IC | 3 | 82.1 | 93.7 | 75.0 | 31.6 | | Between animals | IC | 5 | 93.8 | 107.9 | 83.4 | 55.9 |”
Incorrect percentage for seven specimens in Table 2
MinorThe 'None' infusion group has 24 specimens, so seven specimens are 29.2%, not 29.7%.
Reported 29.7 · recomputed 29.166667
“| Fibrin deposition in reservoir | None | 0 | 7 (29.7) | 14 (58.3) | 3 (12.5) | * |”
Table 2 percentage denominator is misstated
MinorTable 2 reports percentages within each 24-catheter infusion group: for example, 9 insulin-infused specimens are listed as 37.5%. Its note instead says the percentages are of the 36 CCs and 36 ICs total.
“Note: Values in brackets are a percent of 36 CCs and 36 ICs total; *P < 0.05 significant difference between infusion types.”
“| Reticular fiber disruption | Insulin | 1 (4.2) | 9 (37.5) | 6 (25.0) | 8 (33.3) | |”
CC variability was not generally greatest on day 1
MinorTable 4 shows higher between-animal CVs on day 3 than on day 1 for CC AUC60, Cmax, and tmax. Only the AUC CV is greatest on day 1.
“Insulin absorption from the CCs and was least variable on day 5 and most variable on day 1 (Fig. 2 and Table 4).”
“| Between animals | CC | 1 | 45.6 | 25.3 | 29.5 | 49.5 | | Between animals | CC | 3 | 36.1 | 54.5 | 55.6 | 61.2 | | Between animals | CC | 5 | 41.3 | 38.0 | 26.2 | 37.3 |”
Cites
Foreign body reaction to biomaterials
Anderson, J. M., Rodriguez, A., & Chang, D. T. (2008) · Seminars in Immunology · doi:10.1016/j.smim.2007.11.004 · Cited by 5,138 · cited 2× · see citing works
“The degree of inflammation is dependent on catheter material, shape, and weartime and may influence insulin absorption.(14,17) The success of insulin pump therapy may further be dependent on the robustness of insertion methods to avoid issues such as catheter kinking and therefore failed delivery of insulin into the t…”
BackgroundPKSolver: An add-in program for pharmacokinetic and pharmacodynamic data analysis in Microsoft Excel
Zhang, Y., Huo, M., Zhou, J., & Xie, S. (2010) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2010.01.007 · Cited by 2,073 · cited 2× · see citing works
“We applied the linear trapezoidal method using the PKSolver add-in for Microsoft Excel.20 For a given time interval (t2-t1), the integral was calculated by ½$(C1 + C2)$(t2-t1).”
MethodThe pig as a model for human wound healing
Sullivan, T. P., Eaglstein, W. H., Davis, S. C., & Mertz, P. (2001) · Wound Repair and Regeneration · doi:10.1046/j.1524-475x.2001.00066.x · Cited by 1,091 · cited 2× · see citing works
“The subcutaneous adipose tissue of the swine has a similarly high elastic and collagen fiber content and the animals show a healing mechanism comparable to humans.(26,27,29) When insulin is infused, an insulin pump produces a hydrostatic pressure differential between the inside of the CSII cannula and the adjacent sub…”
BackgroundMeyer, W., Schwarz, R., & Neurand, K. (2015) · Current Problems in Dermatology · doi:10.1159/000401274 · Cited by 454 · cited 2× · see citing works
“The subcutaneous adipose tissue of the swine has a similarly high elastic and collagen fiber content and the animals show a healing mechanism comparable to humans.(26,27,29) When insulin is infused, an insulin pump produces a hydrostatic pressure differential between the inside of the CSII cannula and the adjacent sub…”
BackgroundBinder, C., Lauritzen, T., Faber, O., & Pramming, S. (1984) · Diabetes Care · doi:10.2337/diacare.7.2.188 · Cited by 341 · see citing works
Insulin transport across capillaries is rate limiting for insulin action in dogs.
Yang, Y. J., Hope, I. D., Ader, M., & Bergman, R. N. (1989) · Journal of Clinical Investigation · doi:10.1172/jci114339 · Cited by 303 · see citing works
Tissue response to implanted polymers: The significance of sample shape
Matlaga, B. F., Yasenchak, L. P., & Salthouse, T. N. (1976) · Journal of Biomedical Materials Research · doi:10.1002/jbm.820100308 · Cited by 295 · cited 2× · see citing works
“Round, soft shapes are known to be better tolerated that sharp edges.(17,25) In general, the ICs produced a similar, yet more pronounced, tissue response characterized by a central core of adipose tissue debris surrounded by a layer of fibrin thrombus infiltrated with inflammatory cells.”
BackgroundEngineering biomaterials to integrate and heal: The biocompatibility paradigm shifts
Bryers, J. D., Giachelli, C. M., & Ratner, B. D. (2012) · Biotechnology and Bioengineering · doi:10.1002/bit.24559 · Cited by 271 · see citing works
Variability of Insulin Absorption and Insulin Action
Heinemann, L. (2002) · Diabetes Technology & Therapeutics · doi:10.1089/152091502320798312 · Cited by 267 · see citing works
Porcine Models of Cutaneous Wound Healing
Seaton, M., Hocking, A., & Gibran, N. S. (2015) · ILAR Journal · doi:10.1093/ilar/ilv016 · Cited by 249 · see citing works
Continuous Subcutaneous Insulin Infusion
Lenhard, M. J., & Reeves, G. D. (2001) · Archives of Internal Medicine · doi:10.1001/archinte.161.19.2293 · Cited by 204 · see citing works
Some aspects of macrophage behavior at the implant interface
Salthouse, T. N. (1984) · Journal of Biomedical Materials Research · doi:10.1002/jbm.820180407 · Cited by 176 · cited 1× · see citing works
“Round, soft shapes are known to be better tolerated that sharp edges.(17,25) In general, the ICs produced a similar, yet more pronounced, tissue response characterized by a central core of adipose tissue debris surrounded by a layer of fibrin thrombus infiltrated with inflammatory cells.”
BackgroundInsulin Infusion Set: The Achilles Heel of Continuous Subcutaneous Insulin Infusion
Heinemann, L., & Krinelke, L. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600429 · Cited by 148 · cited 1× · see citing works
“This observation is consistent with published animal studies, human clinical trials, and routine clinical care.(5,6,22,23) Of interest, the CCs produced the fastest insulin absorption, the highest plasma insulin concentration, and the greatest AUC60 on day 5 of implantation.”
ResultNonmetabolic Complications of Continuous Subcutaneous Insulin Infusion: A Patient Survey
Pickup, J. C., Yemane, N., Brackenridge, A., & Pender, S. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0192 · Cited by 123 · cited 1× · see citing works
“The success of insulin pump therapy may further be dependent on the robustness of insertion methods to avoid issues such as catheter kinking and therefore failed delivery of insulin into the tissue.(18,19) We hypothesize that the flow of insulin from a CSII catheter into adjacent vascular tissue is affected by the cha…”
BackgroundProtein interactions with subcutaneously implanted biosensors
Gifford, R., Kehoe, J. J., Barnes, S. L., Kornilayev, B. A., Alterman, M. A., & Wilson, G. S. (2006) · Biomaterials · doi:10.1016/j.biomaterials.2005.11.033 · Cited by 117 · see citing works
Dermatological Complications of Continuous Subcutaneous Insulin Infusion in Children and Adolescents
Conwell, L. S., Pope, E., Artiles, A. M., Mohanta, A., Daneman, A., & Daneman, D. (2008) · The Journal of Pediatrics · doi:10.1016/j.jpeds.2007.10.006 · Cited by 111 · see citing works
Schmid, V., Hohberg, C., Borchert, M., Forst, T., & Pfützner, A. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400429 · Cited by 97 · see citing works
(2001) · PubMed · Cited by 93 · see citing works
Randomized Trial of Infusion Set Function: Steel Versus Teflon
Patel, P. J., Benasi, K., Ferrari, G., Evans, M. G., Shanmugham, S., Wilson, D. M., & Buckingham, B. A. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0119 · Cited by 90 · cited 3× · see citing works
“The success of insulin pump therapy may further be dependent on the robustness of insertion methods to avoid issues such as catheter kinking and therefore failed delivery of insulin into the tissue.(18,19) We hypothesize that the flow of insulin from a CSII catheter into adjacent vascular tissue is affected by the cha…”
BackgroundKobayashi, T., Sawano, S., Itoh, T., Kosaka, K., Hirayama, H., & Kasuya, Y. (1983) · Diabetes · doi:10.2337/diab.32.4.331 · Cited by 86 · see citing works
Swan, K. L., Dziura, J. D., Steil, G. M., Voskanyan, G. R., Sikes, K. A., Steffen, A. T., Martin, M. L., Tamborlane, W. V., & Weinzimer, S. A. (2009) · Diabetes Care · doi:10.2337/dc08-0595 · Cited by 82 · see citing works
Golden, S. H., & Sapir, T. (2012) · Journal of Managed Care Pharmacy · doi:10.18553/jmcp.2012.18.s6-a.1 · Cited by 73 · see citing works
Serup, L. G. (1998) · Acta Dermato-Venereologica · doi:10.1080/00015559850135733 · Cited by 73 · see citing works
Ruan, Y., Wilinska, M. E., Thabit, H., & Hovorka, R. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2016.2590498 · Cited by 71 · cited 1× · see citing works
“The intrapatient CV of 37% for Cmax and 39% for tmax are comparable with recently published human data.(32) This high variability may be explained by the inconsistency of the mechanical inflammatory layer surrounding the cannula of a CSII catheter.”
ResultRemote Glucose Monitoring in Camp Setting Reduces the Risk of Prolonged Nocturnal Hypoglycemia
DeSalvo, D. J., Keith-Hynes, P., Peyser, T. A., Place, J., Caswell, K., Wilson, D. M., Harris, B., Clinton, P. K., Kovatchev, B. P., & Buckingham, B. A. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0139 · Cited by 68 · see citing works
Model Study of the Pressure Build-Up during Subcutaneous Injection
Thomsen, M., Hernandez-Garcia, A., Mathiesen, J., Poulsen, M., Sørensen, D. N., Tarnow, L., & Feidenhans'l, R. (2014) · PLoS ONE · doi:10.1371/journal.pone.0104054 · Cited by 65 · cited 2× · see citing works
“Insulin may flow into the connective tissue septa that surround the adipose cells, remain within the surrounding layer of inflammatory tissue, and/or flow upward along the cannula shaft onto the skin surface.(30,31) We hypothesize that a layer of inflammatory tissue that is thick, dense, and/or continuous may prevent…”
BackgroundInsulin Infusion Set Use: European Perspectives and Recommendations
Deiss, D., Adolfsson, P., Alkemade-van Zomeren, M., Bolli, G. B., Charpentier, G., Cobelli, C., Danne, T., Girelli, A., Mueller, H., Verderese, C. A., et al. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.07281.sf · Cited by 62 · see citing works
Plum, A., Agersø, H., & Andersen, L. (2000) · Drug Metabolism and Disposition · doi:10.1016/s0090-9556(24)15122-7 · Cited by 56 · cited 1× · see citing works
“This observation is consistent with published animal studies, human clinical trials, and routine clinical care.(5,6,22,23) Of interest, the CCs produced the fastest insulin absorption, the highest plasma insulin concentration, and the greatest AUC60 on day 5 of implantation.”
ResultPfützner, A., Sachsenheimer, D., Grenningloh, M., Heschel, M., Walther-Johannesen, L., Gharabli, R., & Klonoff, D. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815604438 · Cited by 53 · cited 1× · see citing works
“Patients are instructed to insert a new CSII set every 2 to 3 days because insulin absorption into the circulation becomes more variable and less reliable over time.(6,8,9) Catheters are routinely replaced when a bolus dose of insulin fails to produce the expected glucose lowering effect, defined as unexplained hyperg…”
BackgroundGarg, S. K. (2010) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2009.0127 · Cited by 38 · see citing works
Clausen, T. S., Kaastrup, P., & Stallknecht, B. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2009.0058 · Cited by 36 · see citing works
Karlin, A. W., Ly, T. T., Pyle, L., Forlenza, G. P., Messer, L., Wadwa, R. P., DeSalvo, D. J., Payne, S. L., Hanes, S., Clinton, P., et al. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0432 · Cited by 32 · cited 1× · see citing works
“Studies have shown that some patients indeed successfully wear CSII catheters up to 7 days.(19,24) In contrast, the ICs produced the highest AUC, AUC60, and Cmax on day 1, the fastest tmax on day 3, and the lowest AUC, AUC60, and Cmax on day 5.”
BackgroundInsulin Infusion Sets: A Critical Reappraisal
Heinemann, L. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0013 · Cited by 29 · cited 1× · see citing works
“Worldwide, *1 million people manage their diabetes with an insulin pump, rapid acting insulin, and a continuous subcutaneous insulin infusion (CSII) catheter.(1) Insulin is continuously delivered through the CSII catheter at a variable basal rate between meals and during sleep.”
BackgroundWang, X., Jolliffe, A., Carr, B. D., Zhang, Q., Bilger, M., Cui, Y., Wu, J., Wang, X., Mahoney, M., Rojas-Peña, Á., et al. (2018) · Biomaterials Science · doi:10.1039/c8bm00849c · Cited by 27 · see citing works
Højbjerre, L., Skov-Jensen, C., Kaastrup, P., Pedersen, P. E., & Stallknecht, B. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0061 · Cited by 24 · see citing works
Insulin Pump Use in Adults with Type 1 Diabetes—Practical Issues
Shetty, G. K., & Wolpert, H. A. (2010) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2010.0002 · Cited by 21 · cited 1× · see citing works
“Patients are instructed to insert a new CSII set every 2 to 3 days because insulin absorption into the circulation becomes more variable and less reliable over time.(6,8,9) Catheters are routinely replaced when a bolus dose of insulin fails to produce the expected glucose lowering effect, defined as unexplained hyperg…”
BackgroundIncorporating the Treat-to-Target Concept in Rheumatoid Arthritis
Ruderman, E., Nola, K., Ferrell, S., Sapir, T., & Cameron, D. R. (2012) · Journal of Managed Care Pharmacy · doi:10.18553/jmcp.2012.18.s9-a.1 · Cited by 17 · see citing works
The Pump Life: Patient Responses and Clinical and Technological Problems
Pickup, J. C., Sherwin, R. S., Tamborlane, W. V., Rizza, R. A., Service, F. J., & The Kroc Collaborative Study Group (1985) · Diabetes · doi:10.2337/diab.34.3.s37 · Cited by 15 · cited 1× · see citing works
“Insulin may flow into the connective tissue septa that surround the adipose cells, remain within the surrounding layer of inflammatory tissue, and/or flow upward along the cannula shaft onto the skin surface.(30,31) We hypothesize that a layer of inflammatory tissue that is thick, dense, and/or continuous may prevent…”
BackgroundInsulin Action and Distribution of Tissue Blood Flow
Bergman, R. N. (2003) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2003-031431 · Cited by 15 · see citing works
Schaeffer, N. E. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0098 · Cited by 13 · see citing works
Vlasakakis, G., Johnson, S. L., Lin, J., Yao, X., Gruenloh, C. J., Chism, J. P., & Nunez, D. J. (2015) · Advances in Therapy · doi:10.1007/s12325-015-0222-4 · Cited by 2 · cited 1× · see citing works
“This observation is consistent with published animal studies, human clinical trials, and routine clinical care.(5,6,22,23) Of interest, the CCs produced the fastest insulin absorption, the highest plasma insulin concentration, and the greatest AUC60 on day 5 of implantation.”
Result
Flagged first, then by citations
Research works printed but not linked: 2
- Patel PJ, 2013, Diabetes Technol Ther, 16, 1
- Wang JF, 2001, Comp Med, 51, 341
Cited by
Sorted by citations
Douglass, M. E., Hopkins, S. P., Chug, M. K., Kim, G., Garren, M. R. S., Ashcraft, M. M., Nguyen, D. T., Tayag, N., Handa, H., & Brisbois, E. J. (2021) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.1c18190 · Cited by 29 · see citing works
Wang, X., Jolliffe, A., Carr, B. D., Zhang, Q., Bilger, M., Cui, Y., Wu, J., Wang, X., Mahoney, M., Rojas-Peña, Á., et al. (2018) · Biomaterials Science · doi:10.1039/c8bm00849c · Cited by 27 · see citing works
Advancing continuous subcutaneous insulin infusion in vivo: New insights into tissue challenges
Kesserwan, S., Mulka, A., Sharafieh, R., Qiao, Y., Wu, R., Kreutzer, D. L., & Klueh, U. (2020) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.37097 · Cited by 25 · see citing works
Mulka, A., Lewis, B. E., Mao, L., Sharafieh, R., Kesserwan, S., Wu, R., Kreutzer, D. L., & Klueh, U. (2021) · ACS Pharmacology & Translational Science · doi:10.1021/acsptsci.1c00047 · Cited by 22 · see citing works
Evaluation of Insulin Pump Infusion Sites in Type 1 Diabetes: The DERMIS Study
Kalus, A., Shinohara, M. M., Wang, R. K., Baran, J. D., Dong, X., Khakpour, D., Lu, J., & Hirsch, I. B. (2023) · Diabetes Care · doi:10.2337/dc23-0426 · Cited by 23 · see citing works
EISLER, G. J., Kastner, J. R., Torjman, M. C., Khalf, A., Diaz, D., Dinesen, A. R., Loeum, C., Thakur, M. L., Strasma, P. J., & Joseph, J. I. (2019) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2019-000881 · Cited by 18 · see citing works
Self-citationInsulin Derived Fibrils Induce Cytotoxicity in vitro and Trigger Inflammation in Murine Models
Lewis, B. E., Mulka, A., Mao, L., Sharafieh, R., Qiao, Y., Kesserwan, S., Wu, R., Kreutzer, D., & Klueh, U. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211033868 · Cited by 17 · see citing works
Lipohypertrophy and Insulin: An Old Dog That Needs New Tricks
Huang, J. K., Yeung, A. M., Kerr, D. E., Gentile, S., Heinemann, L., Al‐Sofiani, M. E., Joseph, J. I., Seley, J. J., & Klonoff, D. C. (2023) · Endocrine Practice · doi:10.1016/j.eprac.2023.04.006 · Cited by 17 · see citing works
Self-citationKaushal, A., Zhang, Y., Ballantyne, L. L., & Fitzpatrick, L. E. (2022) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2022.959512 · Cited by 15 · see citing works
“A major limitation of CSII is the short wear times and the poor reliability of IIS, which can lead to the development of unexplained hyperglycemia and diabetic ketoacidosis (Hauzenberger et al., 2017; Hauzenberger et al., 2018).”
BackgroundThe role of Toll-like receptor signaling in the macrophage response to implanted materials
McKiel, L. A., Woodhouse, K. A., & Fitzpatrick, L. E. (2019) · MRS Communications · doi:10.1557/mrc.2019.154 · Cited by 14 · see citing works
Feasibility study of a prototype extended‐wear insulin infusion set in adults with type 1 diabetes
Kastner, J. R., Venkatesh, N., Brown, K., Muchmore, D. B., Ekinci, E. I., Fourlanos, S., Joseph, J. I., Shafeeq, M., Shi, L., Strange, P., et al. (2022) · Diabetes Obesity and Metabolism · doi:10.1111/dom.14685 · Cited by 14 · see citing works
Self-citationA survey of neurological decompression illness in commercial breath-hold divers (Ama) of Japan.
Tamaki, H., Kohshi, K., Ishitake, T., & Wong, R. M. S. (2010) · PubMed · Cited by 14 · see citing works
Swinney, M. R., Cox, A. L., Hawkins, E. D., Xue, J., Garhyan, P., Stanley, J. R. L., Sule, S. V., Adragni, K. P., & Michael, M. D. (2020) · Journal of Pharmaceutical Sciences · doi:10.1016/j.xphs.2020.12.008 · Cited by 13 · see citing works
Simic, A., Schøndorff, P. K., Stumpe, T., Heschel, M., Regittnig, W., Pöttler, T., Ninaus, D., Augustin, T., Groselj‐Strele, A., Pieber, T. R., et al. (2021) · Diabetes Obesity and Metabolism · doi:10.1111/dom.14337 · Cited by 13 · see citing works
“Our findings are consistent with those of previous studies that investigated the effects of prolonged infusion-set wear on insulin sensitivity in humans and swine.(13,14) The study in humans examined the pharmacodynamics of two rapid-acting insulins over 4 days of standard infusion-set wear and observed a significant…”
ResultSelf-citationMcKiel, L. A., Ballantyne, L. L., Negri, G. L., Woodhouse, K. A., & Fitzpatrick, L. E. (2023) · Frontiers in Immunology · doi:10.3389/fimmu.2023.1232586 · Cited by 11 · see citing works
Kastner, J. R., EISLER, G. J., Torjman, M. C., Khalf, A., Diaz, D., Dinesen, A. R., Loeum, C., Strasma, P. J., & Joseph, J. I. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221093362 · Cited by 11 · see citing works
Self-citationDorati, R., Cosco, D., Battaglia, L., Cirri, M., Conti, B., Borghi, T., Gabriele, M., Bruno, G. M., Rocco, P., Selmin, F., et al. (2025) · Drug Discovery Today · doi:10.1016/j.drudis.2025.104441 · Cited by 8 · see citing works
Degen, I., Brown, K. R., Reeve, H. W. J., & Abdallah, Z. S. (2024) · JMIRx Med · doi:10.2196/44384 · Cited by 8 · see citing works
Leveraging Mechanical Forces to Target Insulin Injection–Induced Lipohypertrophy and Fibrosis
desJardins-Park, H. E., & Wan, D. C. (2021) · Diabetes Spectrum · doi:10.2337/ds20-0048 · Cited by 7 · see citing works
Diabetes Technology Meeting 2020
Shang, T., Zhang, J. Y., Bequette, B. W., Raymond, J. K., Coté, G. L., Sherr, J. L., Castle, J. R., Pickup, J. C., Pavlovic, Y., Espinoza, J., et al. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211016480 · Cited by 7 · see citing works
Closed loop insulin delivery–Opportunities and limitations
Weiss, R. (2023) · Journal of Diabetes · doi:10.1111/1753-0407.13490 · Cited by 4 · see citing works
“20 Moreover, repeated insertion of catheters in the same vicinity, as commonly observed in many patients using insulin pumps, results in local lipohypertrophy that hampers insulin absorption even more.”
BackgroundStrollo, F., Guarino, G., & Gentile, S. (2025) · Diabetology · doi:10.3390/diabetology6030022 · Cited by 4 · see citing works
“These consist primarily of local complications, including subcutaneous nodules defined as lipodystrophies (LDs) caused by injection errors and responsible for altered activity of injected insulin [9–12].”
Jacquemier, P., Rétory, Y., Virbel-Fleischman, C., Schmidt, A., Ostertag, A., Cohen-Solal, M. E., Alzaïd, F., Potier, L., Julla, J., Gautier, J., et al. (2023) · Scientific Reports · doi:10.1038/s41598-023-46993-1 · Cited by 4 · see citing works
“Indeed, literature provides multiple examples of bolus delivery studies, either in vivo on swine using imaging 13–15 , in vivo in humans using pressure measurement 16,17 , or ex vivo using imaging 18–20 .”
A Green Light-Activated Insulin Depot with Ultrafast In Vivo Efficiency in the Subcutaneous Space
Ansong, M., Kover, K. L., Shah, P. V., et al. (2024) · ACS Biomaterials Science & Engineering · doi:10.1021/acsbiomaterials.4c00362 · Cited by 4 · see citing works
Insulin Distribution in Human Adipose Tissue via a Novel Insulin Infusion Catheter
Altendorfer‐Kroath, T., Schwingenschuh, S., Schøndorff, P. K., Heschel, M., Sinner, F. M., & Birngruber, T. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0195 · Cited by 4 · see citing works
Lal, R., & Leelarathna, L. (2020) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2020.2502 · Cited by 4 · see citing works
“These include the presence of lipohypertrophy or scarring, local trauma and inflammation around cannula site, and type of cannula (21).”
Garhyan, P., Pratt, E. J., Klein, O., Famulla, S., Zijlstra, E., Lalonde, A., Swinney, M. R., KAZDA, C. M., & Dassau, E. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221145200 · Cited by 2 · see citing works
Progressive Acceleration of Insulin Exposure Over 7 Days of Infusion Set Wear
Kastner, J. R., Bailey, T. S., Strange, P., Shi, L., Oberg, K. A., Strasma, P. J., Joseph, J. I., & Muchmore, D. B. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2022.0323 · Cited by 2 · see citing works
“Insulin absorption (PK) is known to show intrapatient variability on each of 3 days of IIS wear during insulin pump therapy.(1,4,6) In this study, we extended PK observations to 7 days of IIS wear performing 4 euglycemic clamp experiments on days 0, 3, 5, and 7 of IIS wear after SC insulin bolus administration.”
BackgroundResultSelf-citationZhang, Y., Kuo, L., Woodhouse, K. A., & Fitzpatrick, L. E. (2024) · ACS Pharmacology & Translational Science · doi:10.1021/acsptsci.4c00363 · Cited by 1 · see citing works
Performing Dual Glucose Clamp Experiments in Sedated Farm Swine: A Practical Method
Torjman, M. C., Hamilton, W., Dillon, K., et al. (2025) · Methods and Protocols · doi:10.3390/mps8050118 · Cited by 1 · see citing works
“The most complete studies investigating continuous subcutaneous insulin delivery and glucose control require examination and correlation of histological findings, an important advantage of the swine model [9,10].”
Self-citationAouba, A., Dreves, B., Morera, J., Rod, A., Guetlin, M., & Joubert, M. (2026) · Diabetes Technology & Therapeutics · doi:10.1177/15209156261423554 · Cited by 1 · see citing works
“10 Nevertheless, clinical data on the potential impact of catheter type, cannula length, or wear duration on glycemic control remain scarce, particularly in the context of AID systems.”
Pickup, J. C. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.2503 · Cited by 1 · see citing works
Insulin Delivery Hardware: Pumps and Pens
Lal, R., & Leelarathna, L. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.2503
“Absorption after delivery can be variable even within the same individual due to lipohypertrophy; scarring; local trauma; and inflammation from insulin, excipients, preservatives, or the cannula (3).”
Januszewski, A. S., SNAITH, J. R., Grzelka‐Woźniak, A., Simonsen, J. R., Sachithanandan, N., Ward, G. M., O'Neal, D. N., Gordin, D., Thorn, L. M., Groop, P., et al. (2025) · Diabetes Obesity and Metabolism · doi:10.1111/dom.16487
“1,2 This reduction has been attributed to improved insulin absorption, 3,4 lower basal insulin requirements 2 and the avoidance of insulin ‘ stacking ’ , which reduces the risk of hypoglycaemia.”
Yeh, T., Huang, J.-S., & Wu, Y. (2025) · doi:10.2196/preprints.81401
Klueh, U., Oendraite, I., Cunegundes, P. S., et al. (2026) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.64898/2026.06.05.730153
Groisman, L., Thorson, T. J., Saudi, B., Myrick, R. J., Gonthier, A. R., Nelson, G., Loeum, C., Torjman, M. C., Joseph, J. I., Mohraz, A., et al. (2026) · Biomaterials Advances · doi:10.1016/j.bioadv.2026.215155
Self-citation
- Authors
- Hauzenberger, J. R., Hipszer, B. R., Loeum, C., McCue, P. A., DeStefano, M., Torjman, M. C., Kaner, M. T., Dinesen, A. R., Chervoneva, I., Pieber, T. R., & Joseph, J. I.
- Affiliations
- Thomas Jefferson University (US)
- Chester County Historical Society (US)
- Medical University of Graz (AT)
- Corresponding author
- Joseph, Jeffrey I
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2017-11
- Volume
- 19
- Issue
- 11
- Pages
- 641-650
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- PMC
- PMC5689134
- OpenAlex
- W2762466554
- Semantic Scholar
- b04efe0b92d5a1cd334def512745195a82f869ab
- Cited by
- 39
- References
- 41
- Citations by year
- 2026: 3 · 2025: 5 · 2024: 3 · 2023: 6 · 2022: 5 · 2021: 7 · 2020: 4 · 2019: 4 · 2018: 1
- FWCI
- 1.68
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Diabetes Technology & Therapeutics (publishedVersion)
- PubMed (publishedVersion)
- Europe PMC (PubMed Central) (submittedVersion, cc-by)
- PubMed Central (submittedVersion, cc-by)
- Funders
- Thomas Jefferson University
- Bosch-Forschungsstiftung
- Austrian Science Fund (W1241)
- Karl-Franzens-Universität Graz
- Medizinische Universität Graz (W1241)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Engineering › Biomedical Engineering › Intravenous Infusion Technology and Safety
- Health Sciences › Medicine › Dermatology › Skin Diseases and Diabetes
- Keywords
- insulin pump therapy, type 1 diabetes, insulin pharmacokinetics, catheter design, porcine model
- MeSH terms
- Animals, Catheters, Indwelling, Female, Hypoglycemic Agents, Inflammation, Insulin, Insulin Infusion Systems, Swine, Subcutaneous Tissue
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-13.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-13.
Cited sources are checked by gpt-6-sol.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review