The DOI matches its Crossref record; 18 of the 18 references it cites were found in a registry or an index.
Auto-Titrating Versus Standard Continuous Positive Airway Pressure for the Treatment of Obstructive Sleep Apnea: Results of a Meta-analysis
Abstract
STUDY OBJECTIVE: To compare the effectiveness of auto-titrating continuous positive airway pressure (APAP) versus conventional continuous positive airway pressure (CPAP) in reducing the apnea-hypopnea index (AHI), reducing the mean airway pressure, improving subjective sleepiness, and improving treatment adherence in patients with obstructive sleep apnea (OSA). DESIGN: Meta-analysis and metaregression of published randomized trials comparing APAP to CPAP. SETTING: N/A. PARTICIPANTS: N/A. INTERVENTIONS: N/A. RESULTS: We identified 9 randomized trials studying a total of 282 patients. Compared to CPAP, there was no significant advantage of APAP in reducing AHI or sleepiness (pooled APAP-CPAP posttreatment AHI and Epworth Sleepiness Scale score = -0.20 events per hour, 95% confidence interval:[-0.74,0.35], and -0.56 [-1.4,0.3] respectively). The use of APAP reduced the mean applied pressure across the night by 2.2 cm water [1.9,2.5] compared to CPAP. Adherence with therapy was not substantially improved with APAP; pooled estimate of improvement was 0.20 hours per night ([-0.16,0.57], P = .28) using a random-effects model. CONCLUSIONS: Compared to standard CPAP, APAP is associated with a reduction in mean pressure. However, APAP and standard CPAP were similar in adherence and their ability to eliminate respiratory events and to improve subjective sleepiness. Given that APAP is more costly than standard CPAP, APAP should not be considered first-line chronic therapy in all patients with OSA. However, APAP may be useful in other situations (eg, home titrations, detection of mouth leak) or in certain subgroups of patients with OSA. Identifying circumstances in which APAP is a definite improvement over CPAP in terms of costs or effects should be the focus of future studies.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Meta-analysis in clinical trials
DerSimonian, R., & Laird, N. (1986) · Controlled Clinical Trials · doi:10.1016/0197-2456(86)90046-2 · Cited by 39,552
A New Method for Measuring Daytime Sleepiness: The Epworth Sleepiness Scale
Johns, M. W. (1991) · Sleep · doi:10.1093/sleep/14.6.540 · Cited by 17,364
Objective Measurement of Patterns of Nasal CPAP Use by Patients with Obstructive Sleep Apnea
Kribbs, N. B., Pack, A. I., Kline, L. R., Smith, P. L., Schwartz, A. R., Schubert, N. M., Redline, S., Henry, J. N., Getsy, J. E., & Dinges, D. F. (1993) · American Review of Respiratory Disease · doi:10.1164/ajrccm/147.4.887 · Cited by 1,339
The Association between Sleep Apnea and the Risk of Traffic Accidents
Terán-Santos, J., Jimenez-Gomez, A., & Cordero-Guevara, J. (1999) · New England Journal of Medicine · doi:10.1056/nejm199903183401104 · Cited by 1,255
An American Academy of Sleep Medicine Review
Sateia, M. J., Doghramji, K., Hauri, P., & Morin, C. M. (2000) · Cited by 370
Littner, M., Hirshkowitz, M., Davila, D., Anderson, W. M., Kushida, C. A., Woodson, B. T., Johnson, S. F., & Wise, M. S. (2002) · Sleep · doi:10.1093/sleep/25.2.143 · Cited by 183
Comparison between Automatic and Fixed Positive Airway Pressure Therapy in the Home
Massie, C. A., McArdle, N., Hart, R. W., Schmidt-Nowara, W. W., Lankford, A., Hudgel, D. W., Gordon, N., & Douglas, N. J. (2003) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.200201-022oc · Cited by 177
Efficacy of auto-CPAP in the treatment of obstructive sleep apnea/hypopnea syndrome.
Meurice, J. C., Marc, I., & Sériès, F. (1996) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/ajrccm.153.2.8564134 · Cited by 177
Autoadjusting CPAP Therapy Based on Impedance Efficacy, Compliance and Acceptance
RANDERATH, W. J., SCHRAEDER, O., GALETKE, W., FELDMEYER, F., & RÜHLE, K.-H. (2001) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/ajrccm.163.3.2006168 · Cited by 143
Cited by
Meschia, J. F., Bushnell, C., Boden-Albala, B., Braun, L. T., Bravata, D. M., Chaturvedi, S., Creager, M. A., Eckel, R. H., Elkind, M. S. V., Fornage, M., et al. (2014) · Stroke · doi:10.1161/str.0000000000000046 · Cited by 2,650
Adult obstructive sleep apnoea
Jordan, A. S., McSharry, D. G., & Malhotra, A. (2014) · The Lancet · doi:10.1016/s0140-6736(13)60734-5 · Cited by 1,735
Self-citationAdherence to Continuous Positive Airway Pressure Therapy: The Challenge to Effective Treatment
Weaver, T. E., & Grunstein, R. R. (2008) · Proceedings of the American Thoracic Society · doi:10.1513/pats.200708-119mg · Cited by 1,680
Sawyer, A. M., Gooneratne, N. S., Marcus, C. L., Ofer, D., Richards, K. C., & Weaver, T. E. (2011) · Sleep Medicine Reviews · doi:10.1016/j.smrv.2011.01.003 · Cited by 900
Mulgrew, A. T., Fox, N., Ayas, N. T., & Ryan, C. F. (2007) · Annals of Internal Medicine · doi:10.7326/0003-4819-146-3-200702060-00004 · Cited by 368
Self-citationPhilippe, C., Stoïca-Herman, M., Drouot, X., Raffestin, B., Escourrou, P., Hittinger, L., Michel, P.-L., Rouault, S., & d’Ortho, M.-P. (2006) · Heart · doi:10.1136/hrt.2005.060038 · Cited by 346
Morgenthaler, T. I., Aurora, R. N., Brown, T., Zak, R., Alessi, C., Boehlecke, B., Chesson, A. L., Friedman, L., Kapur, V., Maganti, R., et al. (2008) · Sleep · doi:10.1093/sleep/31.1.141 · Cited by 334
Robinson, G. V., Smith, D. M., Langford, B. A., Davies, R. J. O., & Stradling, J. R. (2006) · European Respiratory Journal · doi:10.1183/09031936.06.00062805 · Cited by 314
Positive Airway Pressure Treatment for Obstructive Sleep Apnea
Kakkar, R. K., & Berry, R. B. (2007) · Chest · doi:10.1378/chest.06-2432 · Cited by 258
Predictors of Continuous Positive Airway Pressure Adherence in Patients with Obstructive Sleep Apnea
Mehrtash, M., Bakker, J. P., & Ayas, N. (2019) · Lung · doi:10.1007/s00408-018-00193-1 · Cited by 231
Self-citationIp, S., D'Ambrosio, C., Patel, K., Obadan, N., Kitsios, G. D., Chung, M., & Balk, E. M. (2012) · Systematic Reviews · doi:10.1186/2046-4053-1-20 · Cited by 148
Patient Engagement Using New Technology to Improve Adherence to Positive Airway Pressure Therapy
Malhotra, A., Crocker, M. E., Willes, L., Kelly, C., Lynch, S., & Benjafield, A. V. (2018) · Chest · doi:10.1016/j.chest.2017.11.005 · Cited by 138
Self-citationWest, S. D. (2006) · Thorax · doi:10.1136/thx.2005.046300 · Cited by 93
Compliance in sleep apnoea therapy: influence of home care support and pressure mode
Damjanovic, D., Fluck, A., Bremer, H., Müller-Quernheim, J., Idzko, M., & Sorichter, S. (2009) · European Respiratory Journal · doi:10.1183/09031936.00023408 · Cited by 81
Guidelines for sleep studies in adults-a position statement of the Australasian Sleep Association
Douglas, J. A., Chai-Coetzer, C. L., McEvoy, D., Naughton, M. T., Neill, A. M., Rochford, P., Wheatley, J., & Worsnop, C. (2017) · Sleep Medicine · doi:10.1016/j.sleep.2017.03.019 · Cited by 78
Positive Pressure Therapy: A Perspective on Evidence-based Outcomes and Methods of Application
Sanders, M. H., Montserrat, J. M., Farré, R., & Givelber, R. J. (2008) · Proceedings of the American Thoracic Society · doi:10.1513/pats.200709-150mg · Cited by 75
Vennelle, M., White, S., Riha, R. L., Mackay, T. W., Engleman, H. M., & Douglas, N. J. (2010) · Sleep · doi:10.1093/sleep/33.2.267 · Cited by 74
Auto-Adjusting Versus Fixed Positive Pressure Therapy in Mild to Moderate Obstructive Sleep Apnoea
Nolan, G. M., Doherty, L. S., & Mc Nicholas, W. T. (2007) · Sleep · doi:10.1093/sleep/30.2.189 · Cited by 66
Sleep apnea: a review of diagnostic sensors, algorithms, and therapies
Shokoueinejad, M., Fernandez, C., Carroll, E., Wang, F., Levin, J., Rusk, S., Glattard, N., Mulchrone, A., Zhang, X., Xie, A., et al. (2017) · Physiological Measurement · doi:10.1088/1361-6579/aa6ec6 · Cited by 64
Auto-adjusting Positive Airway Pressure Treatment for Sleep Apnea Diagnosed by Home Sleep Testing
Berry, R. B., & Sriram, P. (2014) · Journal of Clinical Sleep Medicine · doi:10.5664/jcsm.4272 · Cited by 64
Evaluation of autoCPAP devices in home treatment of sleep apnea/hypopnea syndrome
Meurice, J. C., Cornette, A., Philip-Joet, F., Pepin, J. L., Escourrou, P., Ingrand, P., & Veale, D. (2007) · Sleep Medicine · doi:10.1016/j.sleep.2007.03.019 · Cited by 52
Galetke, W., Anduleit, N., Richter, K., Stieglitz, S., & Randerath, W. J. (2008) · Respiration · doi:10.1159/000097767 · Cited by 49
Pépin, J.-L., Muir, J.-F., Gentina, T., Dauvilliers, Y., Tamisier, R., Sapene, M., Escourrou, P., Fleury, B., Philip-Joet, F., Philip, P., et al. (2009) · Chest · doi:10.1378/chest.08-2646 · Cited by 46
Flexible Positive Airway Pressure Improves Treatment Adherence Compared with Auto-adjusting PAP
Chihara, Y., Tsuboi, T., Hitomi, T., Azuma, M., Murase, K., Toyama, Y., Harada, Y., Aihara, K., Tanizawa, K., Handa, T., et al. (2013) · Sleep · doi:10.5665/sleep.2378 · Cited by 42
TO, K. W., CHAN, W. C., CHOO, K. L., LAM, W. K., WONG, K. K., & HUI, D. S. (2008) · Respirology · doi:10.1111/j.1440-1843.2007.01138.x · Cited by 42
Marshall, N. S., Neill, A. M., & Campbell, A. J. (2008) · Sleep and Breathing · doi:10.1007/s11325-008-0189-3 · Cited by 38
Donovan, L. M., Shah, A., Chai-Coetzer, C. L., Barbé, F., Ayas, N. T., & Kapur, V. K. (2020) · Chest · doi:10.1016/j.chest.2019.10.002 · Cited by 36
Self-citationAuto bi-level pressure relief–PAP is as effective as CPAP in OSA patients—a pilot study
Blau, A., Minx, M., Peter, J. G., Glos, M., Penzel, T., Baumann, G., & Fietze, I. (2012) · Sleep and Breathing · doi:10.1007/s11325-011-0574-1 · Cited by 32
Nilius, G., Cottin, U., Domanski, U., van’t Hoog, T., Franke, K.-J., Burian, S., & Ruehle, K.-H. (2012) · Somnologie - Schlafforschung und Schlafmedizin · doi:10.1007/s11818-012-0580-y · Cited by 10
Metadata
- Authors
- Ayas, N. T., Patel, S. R., Malhotra, A., Schulzer, M., Malhotra, M., Jung, D., Fleetham, J., & White, D. P.
- Venue
- Sleep
- Publisher
- Oxford University Press (OUP)
- Published
- 2004-03
- Volume
- 27
- Issue
- 2
- Pages
- 249-253
- ISSN
- 0161-8105 (ISSN-L)
- 1550-9109
- OpenAlex
- W2035872
- Cited by
- 255
- References
- 18
- FWCI
- 10.45
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.