Sputtered iridium oxide films (SIROFs) for low-impedance neural stimulation and recording electrodes
Open accessAbstract
Iridium oxide films formed by electrochemical activation of iridium metal (AIROF) or by electrochemical deposition (EIROF) are being evaluated as low-impedance charge-injection coatings for neural stimulation and recording. Iridium oxide may also be deposited by reactive sputtering from iridium metal in an oxidizing plasma. The characterization of sputtered iridium oxide films (SIROFs) as coatings for nerve electrodes is reported. SIROFs were characterized by cyclic voltammetry, electrochemical impedance spectroscopy, and potential transient measurements during charge-injection. The surface morphology of the SIROF transitions from smooth to highly nodular with increasing film thickness from 80 nm to 4600 nm. Charge-injection capacities exceed 0.75 mC/cm(2) with 0.75 ms current pulses in thicker films. The SIROF was deposited on both planar and non-planar substrates and photolithographically patterned by lift-off.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Sputtered iridium oxide films are mechanically stable and can resist corrosion and degradation.
“In contrast, sputtered iridium oxide films have been shown to be mechanically stable and can resist corrosion and degradation [11].”
From Flexible Nerve Stimulation Electrode With Iridium Oxide Sputtered on Liquid Crystal Polymer (2009) · cites this paper as [11]
Sputtered iridium oxide films can adhere to various substrates.
“Sputtered iridium oxide films have superior charge injection capabilities versus noble metals and can adhere to various substrates.”
From Flexible Nerve Stimulation Electrode With Iridium Oxide Sputtered on Liquid Crystal Polymer (2009) · cites this paper as [11]
Sputtered iridium oxide films have been deposited onto platinum and silicon substrates.
“Sputtered iridium oxide films have been deposited onto platinum [11] and silicon [16] substrates.”
From Flexible Nerve Stimulation Electrode With Iridium Oxide Sputtered on Liquid Crystal Polymer (2009) · cites this paper as [11]
The reduction of oxygen is known to produce superoxides, peroxides, and peroxide ions that can initiate free radical reactions and result in tissue damage.
“The reduction of oxygen is known to produce superoxides, peroxides, and peroxide ions that can initiate free radical reactions and result in tissue damage [10] [11].”
From Flexible Nerve Stimulation Electrode With Iridium Oxide Sputtered on Liquid Crystal Polymer (2009) · cites this paper as [11]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Electrical stimulation of excitable tissue: design of efficacious and safe protocols
Merrill, D. R., Bikson, M., & Jefferys, J. G. R. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2004.10.020 · Cited by 2,078 · see citing works
Visual perception in a blind subject with a chronic microelectronic retinal prosthesis
Humayun, M. S., Weiland, J. D., Fujii, G. Y., Greenberg, R., Williamson, R., Little, J., Mech, B., Cimmarusti, V., Van Boemel, G., Dagnelie, G., et al. (2003) · Vision Research · doi:10.1016/s0042-6989(03)00457-7 · Cited by 856 · see citing works
Schmidt, E. M., Bak, M. J., Hambrecht, F. T., Kufta, C. V., O'Rourke, D. K., & Vallabhanath, P. (1996) · Brain · doi:10.1093/brain/119.2.507 · Cited by 672 · see citing works
Ludwig, K. A., Uram, J. D., Yang, J., Martin, D. C., & Kipke, D. R. (2006) · Journal of Neural Engineering · doi:10.1088/1741-2560/3/1/007 · Cited by 638 · see citing works
Electrodeposited iridium oxide for neural stimulation and recording electrodes
Meyer, R. D., Cogan, S. F., Nguyen, T. H., & Rauh, R. D. (2001) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/7333.918271 · Cited by 255 · see citing works
Self-citationOver-pulsing degrades activated iridium oxide films used for intracortical neural stimulation
Cogan, S. F., Guzelian, A. A., Agnew, W. F., Yuen, T. G. H., & McCreery, D. B. (2004) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2004.02.019 · Cited by 210 · see citing works
Self-citationSputtered Iridium Oxide Films as Charge Injection Material for Functional Electrostimulation
Slavcheva, E., Vitushinsky, R., Mokwa, W., & Schnakenberg, U. (2004) · Journal of The Electrochemical Society · doi:10.1149/1.1747881 · Cited by 154 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Neural Stimulation and Recording Electrodes
Cogan, S. F. (2008) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.10.061807.160518 · Cited by 2,243 · see citing works
Self-citationAn Integrated Brain-Machine Interface Platform With Thousands of Channels
Musk, E., & Neuralink (2019) · Journal of Medical Internet Research · doi:10.2196/16194 · Cited by 1,206 · see citing works
A Review of Organic and Inorganic Biomaterials for Neural Interfaces
Fattahi, P., Yang, G., Kim, G., & Abidian, M. R. (2014) · Advanced Materials · doi:10.1002/adma.201304496 · Cited by 580 · see citing works
In Vitro and In Vivo Evaluation of PEDOT Microelectrodes for Neural Stimulation and Recording
Venkatraman, S., Hendricks, J., King, Z. A., Sereno, A. J., Richardson-Burns, S., Martin, D., & Carmena, J. M. (2011) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2011.2109399 · Cited by 319 · see citing works
Nanostructured platinum grass enables superior impedance reduction for neural microelectrodes
Boehler, C., Stieglitz, T., & Asplund, M. (2015) · Biomaterials · doi:10.1016/j.biomaterials.2015.07.036 · Cited by 179 · see citing works
The future of neuroscience: flexible and wireless implantable neural electronics
McGlynn, E., Nabaei, V., Ren, E., Galeote‐Checa, G., Das, R., Curia, G., & Heidari, H. (2021) · Advanced Science · doi:10.1002/advs.202002693 · Cited by 115 · see citing works
Characteristics of prosthetic vision in rats with subretinal flat and pillar electrode arrays
Ho, E., Lei, X., Flores, T., Lorach, H., Huang, T., Galambos, L., Kamins, T., Harris, J., Mathieson, K., & Palanker, D. (2019) · Journal of Neural Engineering · doi:10.1088/1741-2552/ab34b3 · Cited by 65 · see citing works
Neural interfaces engineered via micro- and nanostructured coatings
Chen, N., Tian, L., Patil, A. C., Peng, S., Yang, I. H., Thakor, N. V., & Ramakrishna, S. (2017) · Nano Today · doi:10.1016/j.nantod.2017.04.007 · Cited by 74 · see citing works
The effect of chronic intracortical microstimulation on the electrode–tissue interface
Chen, K. H., Dammann, J. F., Boback, J. L., Tenore, F. V., Otto, K. J., Gaunt, R. A., & Bensmaia, S. J. (2014) · Journal of Neural Engineering · doi:10.1088/1741-2560/11/2/026004 · Cited by 59 · see citing works
Optimization of pillar electrodes in subretinal prosthesis for enhanced proximity to target neurons
Flores, T., Lei, X., Huang, T., Lorach, H., Dalal, R., Galambos, L., Kamins, T., Mathieson, K., & Palanker, D. (2018) · Journal of Neural Engineering · doi:10.1088/1741-2552/aaac39 · Cited by 57 · see citing works
Flexible Nerve Stimulation Electrode With Iridium Oxide Sputtered on Liquid Crystal Polymer
Wang, K., Liu, C.-C., & Durand, D. M. (2009) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2008.926691 · Cited by 57 · see citing works
Yin, P., Liu, Y., Xiao, L., & Zhang, C. (2021) · Polymers · doi:10.3390/polym13162834 · Cited by 42 · see citing works
Electrical Microstimulation of Visual Cerebral Cortex Elevates Psychophysical Detection Thresholds
Cone, J. J., Ni, A. M., Ghose, K., & Maunsell, J. H. R. (2018) · eneuro · doi:10.1523/eneuro.0311-18.2018 · Cited by 13 · see citing works
Čvančara, P., Valle, G., Müller, M., Guiho, T., Hiairrassary, A., Petrini, F., Raspopovic, S., Strauss, I., Granata, G., Fernandez, E., et al. (2019) · doi:10.1101/653964 · Cited by 12 · see citing works
- Authors
- Cogan, S. F., Plante, T. D., & Ehrlich, J.
- Affiliations
- EIC Laboratories (US)
- ORCID
- Cogan, S.F.
- Venue
- The 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society
- Publisher
- IEEE
- Published
- 2005-04-06
- Volume
- 4
- Pages
- 4153-4156
- ISSN
- 2375-7477 (ISSN-L)
- PMC
- PMC2709748
- OpenAlex
- W111354489
- Cited by
- 138
- References
- 7
- Citations by year
- 2026: 3 · 2025: 7 · 2024: 7 · 2023: 4 · 2022: 5 · 2021: 10 · 2020: 10 · 2019: 9 · 2018: 9 · 2017: 8 · 2016: 10 · 2015: 5 · 2014: 6 · 2013: 8 · 2012: 3
- FWCI
- 17.43
- Type
- conference paper
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- National Heart, Lung, and Blood Institute (R44HL071395)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Materials Science › Polymers and Plastics › Conducting polymers and applications
- Physical Sciences › Chemistry › Electrochemistry › Electrochemical Analysis and Applications
- Keywords
- iridium oxide films, reactive sputtering, neural stimulation electrodes, neural recording electrodes, electrode impedance, electrochemical activation, electrodeposition, film thickness effects, surface morphology, lift-off patterning, photolithography
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review