Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care
Open accessThe DOI matches its Crossref record; 44 of the 44 references it cites were found in a registry or an index; 12 of the 42 sources its claims cite were checked for support.
Abstract
Sensitive sensing Neonatal care, particularly for premature babies, is complicated by the infants' fragility and by the need for a large number of tethered sensors to be attached to their tiny bodies. Chung et al. developed a pair of sensors that only require water to adhere to the skin and allow for untethered monitoring of key vital signs (see the Perspective by Guinsburg). On-board data processing allowed for efficient wireless near-field communication using standard protocols. The absence of cables makes it easier to handle the infants and allows for skin-to-skin contact between the babies and their parents or caregivers. Science , this issue p. eaau0780 ; see also p. 924
Overview
CoChat Summary
TL;DR
The paper introduces a pair of ultrathin, wireless, battery-free epidermal electronic systems (EES) for neonatal intensive care: one chest-mounted ECG device and one foot-mounted PPG device. Pilot studies in neonates show clinical-grade agreement for heart rate, respiratory rate, and SpO₂, plus added capabilities for skin temperature and pulse arrival time as a systolic blood pressure surrogate.
Key findings
- The binodal EES continuously streams ECG, PPG, and skin temperature data, yielding HR, HRV, RR, SpO₂, and PAT; wireless data rates reach up to 800 bytes/s with dual channels, NFC operation works up to 25 cm through NICU materials, and BLE range reaches up to 20 m.
- The devices have effective moduli of 200 to 300 kPa, adhere via van der Waals forces, and show peel forces about a factor of ~10 lower than conventional NICU adhesives, with maximum von Mises stress on skin reduced by a factor of 4.3; the microfluidic channel reduces interfacial stresses by up to a factor of 2.5, and adhesion energy is G = 16 N/m.
- For MRI compatibility, mesh electrodes reduce the in-plane magnetic field gradient by a factor of ~3 and the out-of-plane gradient by a factor of 4 versus commercial electrodes; simulated copper heating is 1°C with maximum skin temperature change of 0.04°C, and experimental 3-T MRI measurements show ~0.1°C temperature difference, meeting FDA “MRI-safe” requirements.
- In neonates (n = 3, gestational age 28 to 40 weeks), mean differences versus standard-of-care monitors were –0.17 beats per minute for HR, 0.76 breaths per minute for RR, and 1.02% for SpO₂; additional studies in n = 18 neonates with gestational ages as low as 28 weeks and weights as low as 1470 g showed similar performance, and scaled production cost is estimated at less than $20 USD per unit.
Methods
- Fabrication used semiconductor processing, lamination, transfer printing, chip placement, and solder bonding; each EES includes serpentine Cu traces, chip-scale ICs, fractal mesh electrodes, an ionic-liquid microfluidic chamber, silicone encapsulation, and an NFC loop antenna for simultaneous wireless power and data.
- In-sensor analytics include a streamlined Pan-Tompkins algorithm for real-time QRS detection to yield HR and HRV, dynamic baseline control and SpO₂ calculation for PPG, and time-synchronized ECG/PPG processing for PAT.
- Validation included healthy adult volunteers (n = 3) against FDA-cleared monitoring equipment, and neonates in two tertiary-level NICUs: n = 3 with synchronous standard-of-care measurements and n = 18 using related device platforms with onboard power supplies; the protocol was IRB-approved and registered on ClinicalTrials.gov (NCT02865070).
Limitations
- The paper states that results are preliminary feasibility testing and validation; comprehensive clinical studies are ongoing to verify measurements across increased age groups and ethnic backgrounds, and remaining challenges include nursing acceptance, compatibility with legacy monitoring systems, and device sterilization for reuse.
- The paper notes that additional testing is needed for altered skin barrier function, such as burn victims or patients with epidermolysis bullosa.
- Evident from the text: the neonatal validation cohorts are small (n = 3 with synchronous standard-of-care comparisons; n = 18 using related platforms with onboard power supplies), adult validation is n = 3, and temperature and PAT data in neonates are shown without reference comparators because conventional devices acquire these only periodically.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures









Verification
Claims the paper makes
6 Valid · 2 Not valid · 18 Not checked · 82 Own results and interpretations
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“A key example is the measurement of pulse arrival time (PAT), defined by the time lapse between the maximum fiducial point in the ECG signal (R peak) and the corresponding minimal fiducial point in the PPG signal at valley as in Fig. 6H, as a direct correlate to systolic blood pressure (29 , 30).”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Real-time measurements, in-sensor analytics, and data transmission
What the sources say
- Partly supportedcites [29] Kim J. S., Chee Y. J., Park J. W., Choi J. W., Park K. S., A new approach for non-intrusive monitoring of blood pressure on a toilet seat. Physiol. Meas. 27, 203– 211 (2006). doi: 10.1088/0967-3334/27/2/010; pmid: 16400206 sourceExperimental — accuracy not yet measured
Section 4.1 reports that PAT was significantly, but moderately, correlated with systolic blood pressure. Section 2.2 defines the study’s PAT using the ECG R-peak and the PPG’s maximum first-derivative point, not the PPG valley; it says the minimum point was unclear in its signal. The claimed correlation is supported, but the attributed measurement definition is not. (LLM, conf=0.87)
- Not checked — no full text of the sourcecites [30] Geddes L. A., Voelz M. H., Babbs C. F., Bourland J. D., Tacker W. A., Pulse transit time as an indicator of arterial blood pressure. Psychophysiology 18, 71– 74 (1981). doi: 10.1111/j.1469-8986.1981.tb01545.x; pmid: 7465731 source
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The inherently thin, soft mechanical properties of the sensors (Fig. 1A) reported here allow for adhesion via van derWaals forces alone.”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Low-modulus mechanics, soft interface adhesion, and implications for neonatal skin safety
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The mechanical decoupling afforded by the microfluidic channel decreases these stresses by up to a factor of 2.5 (Fig. 2B) relative to otherwise similar designs without the microfluidics.”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Low-modulus mechanics, soft interface adhesion, and implications for neonatal skin safety
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Oscillations in the force only appear for holes larger than the characteristic size of the cohesive zone (~500 mm, as in Fig. 2I).”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Low-modulus mechanics, soft interface adhesion, and implications for neonatal skin safety
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Specifically, the interface starts to delaminate when the normal stress reaches ~20 kPa (Fig. 2G, inset).”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Low-modulus mechanics, soft interface adhesion, and implications for neonatal skin safety
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
61 more claims
Create a free account to see all 108 claims and to verify the 1 reference still open.
- Probably not a substantial claim. 14
- It may state two separate claims as one. 33
- The quote only partly states this claim. 2
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Researchers have developed various versatile epidermal electrodes with high performance that can universally measure EEG, EMG, EOG, and ECG through placing electrodes on different areas of the skin.[25,414]”
From Flexible Electronics and Devices as Human–Machine Interfaces for Medical Robotics (2022) · cites this paper as [414]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Previous reports validated the performance of an FDA-cleared physiological monitoring system against gold standard monitoring platforms [18–23].”
From Pregnancy outcomes in infertility patients diagnosed with sleep disordered breathing with wireless wearable sensors (2022) · cites this paper as [18]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“In previous work, we demonstrate the ability for the chest- and finger-mounted sensors to monitor vital signs in both children and adults; the abdominal sensor allows for maternal–fetal monitoring of pregnant women in both high- and low-resource settings as well as the clinical and home settings (22, 23).”
From Comprehensive pregnancy monitoring with a network of wireless, soft, and flexible sensors in high- and low-resource health settings (2021) · cites this paper as [23]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“and have been used in neonatal and pediatric intensive care units. 4,5”
From Advanced wireless monitoring for children in the cardiac perioperative setting (2023) · cites this paper as [4]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“However, this system is limited by (1) the modest maximum operating distances (~30 cm) supported by near-field communication (NFC) protocols used for power transfer and data communication, (2) the mechanically fragile nature of the ultrathin, compliant mechanics designs, (3) the sufficient, but constrained range of measurement capabilities, and (4) the demand for highly advanced device configur”
From Skin-interfaced biosensors for advanced wireless physiological monitoring in neonatal and pediatric intensive-care units (2020) · cites this paper as [9]
Experimental — accuracy not yet measured
5 more claims about this paper
Create a free account to see every claim other papers make about this one.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Compatibility with medical imaging techniques used in the NICU
“eliminating any parasitic heating from magnetically induced eddy currents.”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Compatibility with medical imaging techniques used in the NICU
“Full three-dimensional multi-physicsmodeling shows that, at the end of a single scan for 0.5 ms, the copper layer of an ECG EES undergoes heating by only 1°C (Fig. 3H).”
Citations
Cites
A Real-Time QRS Detection Algorithm
Pan, J., & Tompkins, W. J. (1985) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1985.325532 · Cited by 7,855
Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis
Gao, W., Emaminejad, S., Nyein, H. Y. Y., Challa, S., Chen, K., Peck, A., Fahad, H. M., Ota, H., Shiraki, H., Kiriya, D., et al. (2016) · Nature · doi:10.1038/nature16521 · Cited by 5,087
Kim, D.-H., Lu, N., Ma, R., Kim, Y.-S., Kim, R.-H., Wang, S., Wu, J., Won, S. M., Tao, H., Islam, A., et al. (2011) · Science · doi:10.1126/science.1206157 · Cited by 4,836
Self-citationWearable Sensors for Remote Health Monitoring
Majumder, S., Mondal, T., & Deen, M. (2017) · Sensors · doi:10.3390/s17010130 · Cited by 1,375
Kiserud, T., Piaggio, G., Carroli, G., Widmer, M., Carvalho, J., Neerup Jensen, L., Giordano, D., Cecatti, J. G., Abdel Aleem, H., Talegawkar, S. A., et al. (2017) · PLOS Medicine · doi:10.1371/journal.pmed.1002220 · Cited by 729
Thermochemistry of ionic liquid heat-transfer fluids
Valkenburg, M. E. V., Vaughn, R. L., Williams, M., & Wilkes, J. S. (2005) · Thermochimica Acta · doi:10.1016/j.tca.2004.11.013 · Cited by 595
Halcox, J. P. J., Wareham, K., Cardew, A., Gilmore, M., Barry, J. P., Phillips, C., & Gravenor, M. B. (2017) · Circulation · doi:10.1161/circulationaha.117.030583 · Cited by 580
Self-assembled three dimensional network designs for soft electronics
Jang, K.-I., Li, K., Chung, H. U., Xu, S., Jung, H. N., Yang, Y., Kwak, J. W., Jung, H. H., Song, J., Yang, C., et al. (2017) · Nature Communications · doi:10.1038/ncomms15894 · Cited by 445
Self-citationChen, W., Kobayashi, T., Ichikawa, S., Takeuchi, Y., & Togawa, T. (2000) · Medical & Biological Engineering & Computing · doi:10.1007/bf02345755 · Cited by 425
Harris, K. D., Elias, A. L., & Chung, H.-J. (2016) · Journal of Materials Science · doi:10.1007/s10853-015-9643-3 · Cited by 403
Self-citation
Research works printed but not linked: 2
- O. Sharma, S. N. Lewis, U. Telang, L. D’Almeida, L. E. S. Lewis, Design of a Bluetooth Enabled Health Monitoring System for Infants Using Wearable Technology. J. Adv. Res. Dyn. Contrl. Syst. 15, 887–894 (2017). http://jardcs.org/papers/v9/sp/20181075.pdf
- C. H. Lund J. A. Tucker in Neonatal Skin: Structure and Function S. B. Hoath H. I. Maibach Eds. (Dekker ed. 2 2003) pp. 299–324.
Cited by
Items 301–400 of 800
Wearing comfortable and high electrical output TENGs woven with PTFE core–shell nanofiber yarns
Qiu, S., Teng, X., Zhang, Y., Wang, X., Chen, K., Zhao, J. Q., & Huang, Q. (2025) · Chemical Engineering Journal · doi:10.1016/j.cej.2025.159501 · Cited by 46
Self-citationAl-Halhouli, A. T., Al‐Ghussain, L., Bouri, S. E., Liu, H., & Zheng, D. (2019) · Polymers · doi:10.3390/polym11091518 · Cited by 46
“In addition, the accuracy of the IJP PDMS-based sensor was comparable with or even better than the accuracy of the stretchable RR sensors in the literature [32,33].”
BackgroundMethodFerrari, L. M., Keller, K., Burtscher, B., & Greco, F. (2020) · Multifunctional Materials · doi:10.1088/2399-7532/aba6e3 · Cited by 44
Array Integration and Far‐Field Detection of Biocompatible Wireless LC Pressure Sensors
Wen, H., Chen, C., Li, S., Shi, Y., Wang, H., Guo, W., & Liu, X. (2021) · Small Methods · doi:10.1002/smtd.202001055 · Cited by 46
Zhou, J., Shao, Q., Tang, C., Qiao, F., Lu, T., Li, X., Liu, X.-J., & Zhao, H. (2022) · Advanced Materials Technologies · doi:10.1002/admt.202200595 · Cited by 44
Silk nanofibrous iontronic sensors for accurate blood pressure monitoring
Wang, S., Xiao, J., Liu, H., & Zhang, L. (2022) · Chemical Engineering Journal · doi:10.1016/j.cej.2022.139815 · Cited by 48
Advanced thermal sensing techniques for characterizing the physical properties of skin
Madhvapathy, S. R., Arafa, H., Patel, M. J., Winograd, J., Kong, J., Zhu, J., Xu, S., & Rogers, J. A. (2022) · Applied Physics Reviews · doi:10.1063/5.0095157 · Cited by 48
Self-citationBacterial cellulose/MWCNT coatings for highly sensitive and flexible paper-based humidity sensors
Ma, H., Liu, Z., Lou, J., Ding, Q., Jiang, Y., Li, X., & Han, W. (2022) · Cellulose · doi:10.1007/s10570-022-04960-5 · Cited by 44
Self-citationThermosensitive hydrogel-based, high performance flexible sensors for multi-functional e-skins
Lu, D., Zhu, M., Li, X., Zhu, Z., Lin, X., Guo, C. F., & Xiang, X. ‐. (2023) · Journal of Materials Chemistry A · doi:10.1039/d3ta02948d · Cited by 45
Mariello, M., & Proctor, C. M. (2024) · Advanced Materials Technologies · doi:10.1002/admt.202400797 · Cited by 56
Flexible Electronics and Healthcare Applications
Wang, Z., Sun, L., Ni, Y., Liu, L., & Xu, W. (2021) · Frontiers in Nanotechnology · doi:10.3389/fnano.2021.625989 · Cited by 44
“The discomfort brought by electronic skin to the skin can be eliminated by optimizing the mechanical properties, so optimized electronic skin can be attached to a newborn’s skin for intensive care (Figure 11) (Chung et al., 2019).”
BackgroundMechanically Durable Memristor Arrays Based on a Discrete Structure Design
Wang, T., Cui, Z., Liu, Y., Lu, D., Wang, M., Wan, C., Leow, W. R., Wang, C., Pan, L., Cao, X., et al. (2022) · Advanced Materials · doi:10.1002/adma.202106212 · Cited by 42
Self-citationFeng, Q., Zhang, C., Yin, R., Yin, A., Chen, Y., Wang, H., Yang, Z., Li, K., & Zhao, W. (2022) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.1c25077 · Cited by 41
Ye, Z., Zhao, G., Yang, M., Xu, Y., Ren, Y., Chen, Z., Andrabi, S. M., Xie, J., Gao, W. F., Yan, Z., et al. (2025) · Science Advances · doi:10.1126/sciadv.adt4923 · Cited by 52
Self-citationImperceptible liquid metal based tattoo for Human-Machine interface on hairy skin
Lin, W., Ai, L., Wang, Y., Yang, X., Liao, J., Pan, Q., Hong, Y., Liu, S., Long, Z., Khoo, B. L., et al. (2024) · Chemical Engineering Journal · doi:10.1016/j.cej.2024.151595 · Cited by 40
Nguyen, T. T. P., & Khine, M. (2020) · Polymers · doi:10.3390/polym12071454 · Cited by 45
Wireless monitoring and real-time adaptive predictive indicator of deterioration
Duncan, H. P., Fule, B. K., Rice, I., Sitch, A. J., & Lowe, D. (2020) · Scientific Reports · doi:10.1038/s41598-020-67835-4 · Cited by 42
“A number of sensors and monitoring systems have been developed and shown to be feasible but few have been sufficiently developed and tested in large clinical trials[7-15].”
Spatiotemporal hemodynamic monitoring via configurable skin-like microfiber Bragg grating group
Zhu, H., Luo, J., Dai, Q., Zhu, S., Yang, H., Zhou, K., Zhan, L., Xu, B., Chen, Y., Lu, Y., et al. (2023) · Opto-Electronic Advances · doi:10.29026/oea.2023.230018 · Cited by 40
Wei, C., Lin, W., Wang, L., Cao, Z., Huang, Z., Liao, Q., Guo, Z., Su, Y., Zheng, Y., Liao, X., et al. (2023) · Nano-Micro Letters · doi:10.1007/s40820-023-01176-5 · Cited by 42
“As if creating a virtual world, humans still can use their fingers and eyes to interact with the virtual world through IE devices, which makes the life of the avatar more real and the interaction more direct [8–11].”
BackgroundWang, Q., Li, Y., Lin, Y. Q., Sun, Y., Bai, C., Guo, H., Fang, T., Hu, G., Lu, Y., & Kong, D. (2024) · Nano-Micro Letters · doi:10.1007/s40820-023-01314-z · Cited by 47
“An attractive platform, called epidermal electronics, is established through the direct lamination of stretchable devices on the skin [4, 5], achieving robust body integration for emerging areas in health monitoring [6–8], wearable therapy [9], and human–machine interfaces [10, 11].”
BackgroundAdvancements and artificial intelligence approaches in antennas for environmental sensing
Lalbakhsh, A., Simorangkir, R. B. V. B., Bayat-Makou, N., Kishk, A. A., & Esselle, K. P. (2022) · Elsevier eBooks · doi:10.1016/b978-0-323-90508-4.00004-6 · Cited by 56
Liu, C. Y., Kim, J., Yang, D. S., Cho, D., Yoo, S., Madhvapathy, S. R., Jeong, H., Yang, T., Luan, H., Avila, R. O., et al. (2023) · Advanced Functional Materials · doi:10.1002/adfm.202302256 · Cited by 40
Self-citationZeng, T., Wu, Y., & Lei, M. (2024) · Advanced Composites and Hybrid Materials · doi:10.1007/s42114-024-00949-9 · Cited by 50
“[88] The system consists of a pair of ultrathin, low-modulus measurement modules, which were capable of non-invasively and accurately monitoring the vital signs of newborns, including heart rate (HR), respiratory rate (RR), temperature, blood oxygen (SpO 2 ), and blood pressure (BP).”
E-Skin and Its Advanced Applications in Ubiquitous Health Monitoring
Sun, X., Guo, X., Gao, J., Wu, J., Huang, F., Zhang, J., Huang, F., Lu, X. J., Shi, Y., & Pan, L. (2024) · Biomedicines · doi:10.3390/biomedicines12102307 · Cited by 48
“…reduces unnecessary interventions, providing parents and healthcare professionals with scienti fi c evidence and health assessments and Rogers’ team has developed wireless, battery-free vital sign monitoring systems based on ultrathin “skin-like” measurement modules [207]. predictions [205,206].”
Background3D printed microstructures for flexible electronic devices
Liu, Y., Xu, Y., Avila, R. O., Liu, C., Xie, Z., Wang, L., & Yu, X. (2019) · Nanotechnology · doi:10.1088/1361-6528/ab2d5d · Cited by 41
Self-citationFully roll-to-roll gravure printed electronics: challenges and the way to integrating logic gates
Tiara, A. M., Moon, H., Cho, G., & Lee, J. (2022) · Japanese Journal of Applied Physics · doi:10.35848/1347-4065/ac575e · Cited by 45
Self-citationBrenneman, J., Tansel, D. Z., Fedder, G. K., & Panat, R. P. (2021) · Extreme Mechanics Letters · doi:10.1016/j.eml.2021.101199 · Cited by 39
Highly stretchable, deformation-stable wireless powering antenna for wearable electronics
He, Z., Wang, Y., Xiao, H., Wu, Y., Xia, X., Li, S., Liu, J., Huang, K., Wang, F., Shang, J., et al. (2023) · Nano Energy · doi:10.1016/j.nanoen.2023.108461 · Cited by 40
A Bioinspired Wireless Epidermal Photoreceptor for Artificial Skin Vision
Zhang, Y., & Tao, T. H. (2020) · Advanced Functional Materials · doi:10.1002/adfm.202000381 · Cited by 41
Self-citationZhu, J., Zhang, S., Yi, N., Song, C., Qiu, D., Hu, Z., Li, B., Xing, C., Yang, H., Wang, Q., et al. (2021) · Nano-Micro Letters · doi:10.1007/s40820-021-00631-5 · Cited by 42
“For instance, near-field communication (NFC) has been widely used in wireless epidermal electronic systems [26] to monitor various physiologically relevant signals.”
Zhang, X., Hu, M., Zhang, Y., Zhai, G., & Zhang, X.-P. S. (2023) · Advanced Intelligent Systems · doi:10.1002/aisy.202200345 · Cited by 39
Self-citationArmadillo-inspired ultra-sensitive flexible sensor for wearable electronics
Lü, Y., Zhao, D., He, J., & Zou, L. (2023) · Chemical Engineering Journal · doi:10.1016/j.cej.2023.146171 · Cited by 36
A Noninvasive Miniaturized Transcutaneous Oxygen Monitor
Costanzo, I., Sen, D., Adegite, J. O., Rao, P. M., & Güler, Ü. (2021) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2021.3094931 · Cited by 37
Huttunen, O., Behfar, M. H., Hiitola‐Keinänen, J., & Hiltunen, J. A. (2022) · Advanced Materials Technologies · doi:10.1002/admt.202101496 · Cited by 42
Yang, G., Lan, Z., Gong, H., Wen, J., Pang, B., Qiu, Y., Zhang, Y., Guo, W., Bu, T., Xie, B., et al. (2024) · Advanced Functional Materials · doi:10.1002/adfm.202417841 · Cited by 47
Self-citationMotion-unrestricted dynamic electrocardiogram system utilizing imperceptible electronics
Li, D., Cui, T., Liu, J., Shao, W., Liu, X., Chen, Z., Xu, Z., Li, X., Xu, S., Xie, Z., et al. (2025) · Nature Communications · doi:10.1038/s41467-025-58390-5 · Cited by 56
Self-citationMultifunctional Biomedical Devices with Closed‐Loop Systems for Precision Therapy
Leng, Y., & Sun, R. (2025) · Advanced Healthcare Materials · doi:10.1002/adhm.202500860 · Cited by 51
Bai, C., Ji, K., Feng, S., Zhang, J., & Kong, D. (2022) · Energy storage materials · doi:10.1016/j.ensm.2022.02.032 · Cited by 36
Yu, B., Yue, D., Hou, K., Ju, L., Chen, H., Ding, C., Liu, Z., Dai, Y., Bisoyi, H. K., Guan, Y., et al. (2022) · Light Science & Applications · doi:10.1038/s41377-022-01005-1 · Cited by 44
Self-citationPlant‐Wearable Sensors for Intelligent Forestry Monitoring
Li, Y., Xu, H., Han, C., Bai, Y., Wang, Y., Yu, H., Song, W., & Sun, Z. (2022) · Advanced Sustainable Systems · doi:10.1002/adsu.202200333 · Cited by 41
Yuan, J., Feng, Z., Li, D., Luo, Y., & Zhou, Y. (2024) · Rare Metals · doi:10.1007/s12598-023-02543-2 · Cited by 39
“Flexible electronic devices that are bendable, twistable, and stretchable can conformally cover complex curved surfaces, which have become an increasing advancement of electronic devices [1–6].”
Lighting the Path to Precision Healthcare: Advances and Applications of Wearable Photonic Sensors
Song, R., Cho, S., Khan, S. A., Park, I., & Gao, W. F. (2025) · Advanced Materials · doi:10.1002/adma.202419161 · Cited by 54
Yi, Z., Huang, J., Liu, Z., Liu, J., & Yang, B. T. (2019) · IEEE Electron Device Letters · doi:10.1109/led.2019.2954878 · Cited by 35
“However, NFC sensors are generally used to complete a static measurement [20].”
BackgroundSelf-citationStretchable Polymerized High Internal Phase Emulsion Separators for High Performance Soft Batteries
Danninger, D., Hartmann, F., Paschinger, W., Pruckner, R., Schwödiauer, R., Demchyshyn, S., Bismarck, A., Bauer, S., & Kaltenbrunner, M. (2020) · Advanced Energy Materials · doi:10.1002/aenm.202000467 · Cited by 34
Ink‐Based Additive Nanomanufacturing of Functional Materials for Human‐Integrated Smart Wearables
Xu, S., & Wu, W. (2020) · Advanced Intelligent Systems · doi:10.1002/aisy.202000117 · Cited by 36
Self-citationLee, W. H., Lee, Y., Na, J. Y., Kim, S. H., Lee, H. J., Lim, Y., Cho, S. H., Cho, S. H., & Park, H. (2020) · PLoS ONE · doi:10.1371/journal.pone.0243939 · Cited by 35
Self-citationHuang, Q., Wang, J., Li, C., Zhu, J., Wang, W., Huang, Y., Zhang, Y., Jiao, H., Zhang, S., Meng, H., et al. (2022) · npj Flexible Electronics · doi:10.1038/s41528-022-00190-8 · Cited by 37
Self-citationLuo, Y., Chen, X., Li, X., Tian, H., Li, S., Wang, L., He, J., Yang, Z., & Shao, J. (2022) · Advanced Materials · doi:10.1002/adma.202207141 · Cited by 35
Flexible organic photodetectors and their use in wearable systems
Guo, H., Saifi, S., Fukuda, K., Cheng, H., Lou, Z., & Xu, X. (2021) · Digital Signal Processing · doi:10.1016/j.dsp.2021.103145 · Cited by 37
A general self-powered wireless sensing solution based on triboelectric-discharge effect
Wang, H., Xia, X., Fu, J., Li, J., Chen, C., Dai, Y., Fan, Z., Hu, G., & Zi, Y. (2022) · Nano Energy · doi:10.1016/j.nanoen.2022.107982 · Cited by 37
The newborn delivery room of tomorrow: emerging and future technologies
Batey, N., Henry, C., Garg, S., Wagner, M., Malhotra, A., Valstar, M., Smith, T., Sharkey, D., the European Society for Paediatric Research (ESPR) Neonatal Resuscitation Section Writing Group, Niemuth, M., et al. (2024) · Pediatric Research · doi:10.1038/s41390-022-01988-y · Cited by 31
“There have been some exciting developments with wireless vital signs monitoring in the neonatal unit which offer real potential.(108) These must be suitably adapted to cope with the different demands of the DR with a transitioning infant undergoing DCC.”
BackgroundEmerging intelligent wearable devices for cardiovascular health monitoring
Wang, Y., Zou, Y., & Li, Z. (2024) · Nano Today · doi:10.1016/j.nantod.2024.102544 · Cited by 40
Biomimetic electroactive materials and devices for regenerative engineering
Liu, Y., Bai, Y., Heng, B. C., Wang, Y., Zhang, X., Nguyen, T. D., & Deng, X. (2025) · Nature Reviews Electrical Engineering · doi:10.1038/s44287-025-00145-x · Cited by 44
Soft–Hard Composites for Bioelectric Interfaces
Lin, Y., Fang, Y., Yue, J., & Tian, B. (2020) · Trends in Chemistry · doi:10.1016/j.trechm.2020.03.005 · Cited by 36
“A wireless epidermal electronic system developed for a neonatal intensive care unit was used to monitor and analyze data in-sensor from neonates in real time without a battery [47].”
MethodWearable Label‐Free Optical Biodetectors: Progress and Perspectives
Dhanabalan, S. S., Sriram, S., Walia, S., Avaninathan, S. R., Carrasco, M. F., & Bhaskaran, M. C. (2020) · Advanced Photonics Research · doi:10.1002/adpr.202000076 · Cited by 35
Fu, F., Wang, J., Tan, Y., & Yu, J. (2021) · Macromolecular Rapid Communications · doi:10.1002/marc.202100776 · Cited by 33
Han, H., Oh, Y. S., Cho, S., Park, H., Lee, S., Ko, K., Park, J., Choi, J., Ha, J., Han, C., et al. (2022) · Small · doi:10.1002/smll.202205048 · Cited by 34
Self-citationWearable Humidity Sensor for Continuous Sweat Rate Monitoring
Aggarwal, A., Dautta, M., Ayala‐Cardona, L. F., Wudaru, A., & Javey, A. (2023) · Advanced Materials Technologies · doi:10.1002/admt.202300385 · Cited by 33
Mechanics designs-performance relationships in epidermal triboelectric nanogenerators
Yao, K., Liu, Y., Li, D., He, J., Li, J., Lam, R. H. W., Xie, Z., Wang, L., & Yu, X. (2020) · Nano Energy · doi:10.1016/j.nanoen.2020.105017 · Cited by 32
Self-citationNFC-Based Wearable Optoelectronics Working with Smartphone Application for Untact Healthcare
Kang, M. H., Lee, G. J., Yun, J. H., & Song, Y. M. (2021) · Sensors · doi:10.3390/s21030878 · Cited by 36
Low‐Threshold, External‐Cavity‐Free Flexible Perovskite Lasers
Cao, X., Xing, S., Lai, R., Lian, Y., Wang, Y., Xu, J., Zou, C., Zhao, B., & Di, D. (2023) · Advanced Functional Materials · doi:10.1002/adfm.202211841 · Cited by 35
Applications of flexible electronics related to cardiocerebral vascular system
Lin, R., Lei, M., Ding, S., Cheng, Q., Ma, Z., Wang, L., Tang, Z., Zhou, B., & Zhou, Y. (2023) · Materials Today Bio · doi:10.1016/j.mtbio.2023.100787 · Cited by 35
Wang, Q., Sun, Y., Qin, C., Lin, Y. Q., Fang, T., Yang, C., Zhang, J., Lu, Y., & Kong, D. (2025) · Advanced Materials · doi:10.1002/adma.202417799 · Cited by 38
Skin‐Inspired Electronics and Its Applications in Advanced Intelligent Systems
Li, J., Ma, Z., Wang, H., Gao, X., Zhou, Z., Tao, R., Pan, L., & Shi, Y. (2019) · Advanced Intelligent Systems · doi:10.1002/aisy.201900063 · Cited by 34
Nano energy for miniaturized systems
Zhu, M., Zhu, F., & Schmidt, O. G. (2020) · Nano Materials Science · doi:10.1016/j.nanoms.2020.10.001 · Cited by 31
Tensor-Based ECG Anomaly Detection toward Cardiac Monitoring in the Internet of Health Things
Zhou, H., & Kan, C. (2021) · Sensors · doi:10.3390/s21124173 · Cited by 31
“, textiles [4] and skin-like sensors [5]) have been increasingly developed to record the ECG in out-of-hospital settings.”
BackgroundPark, Y., Jang, J., Kim, H., Hwang, J. C., Kwon, Y. W., & Park, J. (2021) · Advanced Electronic Materials · doi:10.1002/aelm.202101034 · Cited by 31
Self-citationHui, Z., Wang, P., Yang, J., Zhou, J., Huang, W., & Sun, G. (2022) · Advanced Materials Interfaces · doi:10.1002/admi.202200261 · Cited by 32
Chu, X., Wang, R., Zhao, H. C., Kuang, M., Yan, J., Wang, B., Ma, H., Cui, M., & Zhang, X. (2022) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.2c00828 · Cited by 30
Sun, J., Zhang, L., Gong, S., Chen, J., & Guo, H. (2023) · Nano Energy · doi:10.1016/j.nanoen.2023.109010 · Cited by 35
Emerging Blood Gas Monitors: How They Can Help With COVID-19
Güler, Ü., Costanzo, I., & Sen, D. (2020) · IEEE Solid-State Circuits Magazine · doi:10.1109/mssc.2020.3021839 · Cited by 30
“Microfluidic channels further reduce the forces needed to remove the patch [20].”
BackgroundInfluentialAI‐Based Metamaterial Design for Wearables
Yigci, D., Ahmadpour, A., & Taşoğlu, S. (2023) · Advanced Sensor Research · doi:10.1002/adsr.202300109 · Cited by 35
Liquid Metal-Based Self-Healing Conductors for Flexible and Stretchable Electronics
Lin, Y. Q., Wang, H., Qiu, W., Ye, C., & Kong, D. (2024) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.4c10541 · Cited by 33
Hydrogel Cryo‐Microtomy Continuously Making Soft Electronic Devices
Fu, X., Li, J., Tang, C., Xie, S., Sun, X., Wang, B., & Peng, H. (2020) · Advanced Functional Materials · doi:10.1002/adfm.202008355 · Cited by 30
Soft Wireless Bioelectronics and Differential Electrodermal Activity for Home Sleep Monitoring
Kim, H., Kwon, S., Kwon, Y., & Yeo, W. (2021) · Sensors · doi:10.3390/s21020354 · Cited by 35
Wang, Y., Li, Z., & Hu, Q. (2021) · Nano Today · doi:10.1016/j.nantod.2021.101127 · Cited by 37
Recent Advances in 2D Wearable Flexible Sensors
Aftab, S., Al‐Kahtani, A. A., Iqbal, M. Z., Hussain, S., & Koyyada, G. (2023) · Advanced Materials Technologies · doi:10.1002/admt.202202168 · Cited by 32
Designs and Applications for the Multimodal Flexible Hybrid Epidermal Electronic Systems
Li, D., Cui, T., Xu, Z., Xu, S., Dong, Z., Tao, L., Liu, H., Yang, Y., & Ren, T. (2024) · Research · doi:10.34133/research.0424 · Cited by 39
“Second, since conventional IC chips can be regarded as rigid islands themselves, they can be directly inter-connected by stretchable bridges to form the island-bridge structure [37,38].”
BackgroundMethodInfluentialSelf-citationAn E-Textile Respiration Sensing System for NICU Monitoring: Design and Validation
Cay, G., Ravichandran, V., Saikia, M. J., Hoffman, L., Laptook, A. R., Padbury, J. F., Salisbury, A. L., Gitelson-Kahn, A., Venkatasubramanian, K., Shahriari, Y., et al. (2021) · Journal of Signal Processing Systems · doi:10.1007/s11265-021-01669-9 · Cited by 34
Ultraflexible tattoo electrodes for epidermal and in vivo electrophysiological recording
Wei, B., Wang, Z., Guo, H., Xie, F., Cheng, S., Lou, Z., Zhou, C., Ji, H., Zhang, M., Wang, X., et al. (2023) · Cell Reports Physical Science · doi:10.1016/j.xcrp.2023.101335 · Cited by 36
Xing, L., Wang, X., Li, M., Jia, Y., Yang, G., Liu, C., Shen, C., & Liu, X. (2024) · Advanced Nanocomposites · doi:10.1016/j.adna.2024.05.001 · Cited by 32
Shin, J., Wang, H., Kwon, K., Ostojich, D., Christiansen, Z., Berkovich, J., Park, Y., Li, Z., Lee, G., Nasif, R., et al. (2022) · Advanced Healthcare Materials · doi:10.1002/adhm.202202021 · Cited by 32
Self-citationRecyclable liquid metal – Graphene supercapacitor
Sanati, A. L., Lopes, P. A., Chambel, A., Silva, A. F., Oliveira, D. M., Majidi, C., de Almeida, A. T., & Tavakoli, M. (2023) · Chemical Engineering Journal · doi:10.1016/j.cej.2023.147894 · Cited by 28
Surface‐Mediated Interconnections of Nanoparticles in Cellulosic Fibrous Materials toward 3D Sensors
Yan, S., Shan, S., Wen, J., Li, J., Kang, N., Wu, Z., Lombardi, J. P., Cheng, H., Wang, J., Luo, J., et al. (2020) · Advanced Materials · doi:10.1002/adma.202002171 · Cited by 27
Mobile health in preventive cardiology: current status and future perspective
Kozik, M., Isakadze, N., & Martin, S. S. (2021) · Current Opinion in Cardiology · doi:10.1097/hco.0000000000000891 · Cited by 29
Hwang, W., Kim, J., Park, S., Kang, T., Kim, S.-H., Lee, K., Lee, M., Kwak, R., Choi, I., & Yi, H. (2022) · Advanced Materials Technologies · doi:10.1002/admt.202200477 · Cited by 25
Self-citationPiezoelectric approaches for wearable continuous blood pressure monitoring: a review
Yi, Z., Zhang, W., & Yang, B. T. (2022) · Journal of Micromechanics and Microengineering · doi:10.1088/1361-6439/ac87ba · Cited by 30
“Wearable electrocardiogram (EGC) monitoring device [16, 17] have been presented for neonatal and paediatric intensive care units.”
Smart electronics based on 2D materials for wireless healthcare monitoring
Hu, L., Kim, B. J., Ji, S., Hong, J., Katiyar, A. K., & Ahn, J. (2022) · Applied Physics Reviews · doi:10.1063/5.0104873 · Cited by 31
Self-citationMagnetic Manipulation of Locomotive Liquid Electrodes for Wireless Active Cardiac Monitoring
Kim, S., Park, Y., Kim, J., Kim, E., Lee, D. H., Lee, J., Cheon, J., & Park, J. (2023) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.3c03724 · Cited by 27
Self-citationBan, S., Lee, H., Chen, J., Kim, H.-S., Hu, Y., Cho, S. J., & Yeo, W. (2024) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2024.116302 · Cited by 30
Wang, Z., Xu, X., Xu, Y., Lin, W., & Peng, Z. (2022) · npj Flexible Electronics · doi:10.1038/s41528-022-00175-7 · Cited by 30
Self-citationStretchable microbatteries and microsupercapacitors for next-generation wearable electronics
Bai, C., Li, S., Ji, K., Wang, M., & Kong, D. (2023) · Energy Materials · doi:10.20517/energymater.2023.31 · Cited by 26
Synthesis of shape-programmable elastomer for a bioresorbable, wireless nerve stimulator
Lim, J. H., Han, W. B., Jang, T., Ko, G., Shin, J., Han, S., Kang, H., Eom, C., Choi, S. J., Rajaram, K., et al. (2024) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2024.116222 · Cited by 29
HfO2 Memristor-Based Flexible Radio Frequency Switches
Chen, S.-C., Yang, Y.-T., Tseng, Y.-C., Chiou, K.-D., Huang, P., Chih, J.-H., Liu, H.-Y., Chou, T.-T., Jhang, Y.-Y., Chen, C., et al. (2024) · ACS Nano · doi:10.1021/acsnano.4c11846 · Cited by 32
Photonic Nanomaterials for Wearable Health Solutions
Park, T., Leem, J. W., Kim, Y. L., & Lee, C. H. (2025) · Advanced Materials · doi:10.1002/adma.202418705 · Cited by 32
Self-citationRobust Skin-Conformal Nano-Electrodes for Sustainable Health and Performance Monitoring
Kim, J., Park, S., Jeon, J., Kang, D.-H., Bryan, G. M., Broderick, T. J., Stone, M. O., & Tsukruk, V. V. (2025) · ACS Nano · doi:10.1021/acsnano.5c08540 · Cited by 33
“We increased the treadmill speed to 3 min intervals: 1.0 m/s (slow), 1.4 m/s (typical), and 1.8 m/s (brisk), which covers most of the walking cycles (Figure 8c, see the detailed protocol in Supporting Information Note 4).”
InfluentialSelf-citationAl-Halhouli, A. T., Al‐Ghussain, L., Bouri, S. E., Liu, H., & Zheng, D. (2020) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-020-00481-3 · Cited by 28
“[38] Stretchable and wearable RR sensor 0.”
Flexible hybrid electronics: Enabling integration techniques and applications
Wu, H., Huang, Y., & Yin, Z. (2022) · Science China Technological Sciences · doi:10.1007/s11431-022-2074-8 · Cited by 32
Self-citationHow Far for the Electronic Skin: From Multifunctional Material to Advanced Applications
Chen, Y., Sun, Y., Wei, Y., & Qiu, J. (2023) · Advanced Materials Technologies · doi:10.1002/admt.202201352 · Cited by 26
Micropatterned Elastomeric Composites for Encapsulation of Transient Electronics
Han, W. B., Ko, G., Yang, S. M., Kang, H., Lee, J. H., Shin, J., Jang, T., Han, S., Kim, D., Lim, J. H., et al. (2023) · ACS Nano · doi:10.1021/acsnano.3c03063 · Cited by 24
Self-citation
Metadata
- Authors
- Chung, H. U., Kim, B. H., Lee, J. Y., Lee, J., Xie, Z., Ibler, E. M., Lee, K., Banks, A., Jeong, J. Y., Kim, J., Ogle, C., Grande, D., Yu, Y., Jang, H., Assem, P., Ryu, D., Kwak, J. W., Namkoong, M., Park, J. B., Lee, Y., Kim, D. H., Ryu, A., Jeong, J., You, K., Ji, B., Liu, Z., Huo, Q., Feng, X., Deng, Y., Xu, Y., Jang, K.-I., Kim, J., Zhang, Y., Ghaffari, R., Rand, C. M., Schau, M., Hamvas, A., Weese-Mayer, D. E., Huang, Y., Lee, S. M., Lee, C. H., Shanbhag, N. R., Paller, A. S., Xu, S., & Rogers, J. A.
- Affiliations
- Northwestern University (US)
- University of Illinois Urbana-Champaign (US)
- Department of Materials Science and Engineering, Northwestern University, Evanston, IL 60208, USA.
- Agency for Science, Technology and Research (SG)
- Tsinghua University (CN)
- Daegu Gyeongbuk Institute of Science and Technology (KR)
- Lurie Children's Hospital (US)
- Center for Autonomic Medicine, Department of Pediatrics, Ann & Robert H. Lurie Children’s Hospital of Chicago, Chicago, IL 60611, USA.
- Kookmin University (KR)
- Purdue University West Lafayette (US)
- Corresponding author
- Chung, Ha Uk
- Kim, Bong Hoon
- Lee, Jong Yoon
- Lee, Jungyup
- Xie, Zhaoqian
- Paller, Amy S.
- Xu, Shuai
- Rogers, John A.
- ORCID
- Chung, Ha Uk
- Kim, Bong Hoon
- Lee, Jong Yoon
- Lee, Jungyup
- Xie, Zhaoqian
- Lee, KunHyuck
- Banks, Anthony
- Jeong, Ji Yoon
- Kim, Jongwon
- Grande, Dominic
- Yu, Yongjoon
- Jang, Hokyung
- Assem, Pourya
- Kwak, Jean Won
- Namkoong, Myeong
- Park, Jun Bin
- Lee, Yechan
- Kim, Do Hoon
- Ryu, Arin
- You, Kevin
- Ji, Bowen
- Liu, Zhuangjian
- Huo, Qingze
- Feng, Xue
- Deng, Yujun
- Xu, Yeshou
- Jang, Kyung-In
- Kim, Jeonghyun
- Zhang, Yihui
- Ghaffari, Roozbeh
- Rand, Casey M.
- Hamvas, Aaron
- Huang, Yonggang
- Lee, Seung Min
- Lee, Chi Hwan
- Shanbhag, Naresh R.
- Paller, Amy S.
- Xu, Shuai
- Rogers, John A.
- Venue
- Science
- Published
- 2019-03
- Volume
- 363
- Issue
- 6430
- ISSN
- 0036-8075 (ISSN-L)
- 1095-9203
- PMC
- PMC6510306
- OpenAlex
- W2916533817
- Semantic Scholar
- b0511afac0cc252946aa8f08a078b12e48030581
- Cited by
- 807
- References
- 43
- Citations by year
- 2026: 72 · 2025: 112 · 2024: 103 · 2023: 121 · 2022: 124 · 2021: 131 · 2020: 97 · 2019: 47
- FWCI
- 35.10
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by
- Locations
- Science (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Funders
- Bill and Melinda Gates Foundation (OPP1182909)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Physical Sciences › Engineering › Biomedical Engineering › Wireless Body Area Networks
- Keywords
- epidermal electronics, neonatal intensive care, battery-free operation, neonatal monitoring
- MeSH terms
- Diagnostic Imaging, Electronics, Equipment Design, Humans, Infant, Newborn, Monitoring, Physiologic, Skin, Intensive Care, Neonatal, Vital Signs, Lab-On-A-Chip Devices, Wireless Technology
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.