Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care
Open accessThe DOI matches its Crossref record; 44 of the 44 references it cites were found in a registry or an index; 12 of the 42 sources its claims cite were checked for support.
Abstract
Sensitive sensing Neonatal care, particularly for premature babies, is complicated by the infants' fragility and by the need for a large number of tethered sensors to be attached to their tiny bodies. Chung et al. developed a pair of sensors that only require water to adhere to the skin and allow for untethered monitoring of key vital signs (see the Perspective by Guinsburg). On-board data processing allowed for efficient wireless near-field communication using standard protocols. The absence of cables makes it easier to handle the infants and allows for skin-to-skin contact between the babies and their parents or caregivers. Science , this issue p. eaau0780 ; see also p. 924
Overview
CoChat Summary
TL;DR
The paper introduces a pair of ultrathin, wireless, battery-free epidermal electronic systems (EES) for neonatal intensive care: one chest-mounted ECG device and one foot-mounted PPG device. Pilot studies in neonates show clinical-grade agreement for heart rate, respiratory rate, and SpO₂, plus added capabilities for skin temperature and pulse arrival time as a systolic blood pressure surrogate.
Key findings
- The binodal EES continuously streams ECG, PPG, and skin temperature data, yielding HR, HRV, RR, SpO₂, and PAT; wireless data rates reach up to 800 bytes/s with dual channels, NFC operation works up to 25 cm through NICU materials, and BLE range reaches up to 20 m.
- The devices have effective moduli of 200 to 300 kPa, adhere via van der Waals forces, and show peel forces about a factor of ~10 lower than conventional NICU adhesives, with maximum von Mises stress on skin reduced by a factor of 4.3; the microfluidic channel reduces interfacial stresses by up to a factor of 2.5, and adhesion energy is G = 16 N/m.
- For MRI compatibility, mesh electrodes reduce the in-plane magnetic field gradient by a factor of ~3 and the out-of-plane gradient by a factor of 4 versus commercial electrodes; simulated copper heating is 1°C with maximum skin temperature change of 0.04°C, and experimental 3-T MRI measurements show ~0.1°C temperature difference, meeting FDA “MRI-safe” requirements.
- In neonates (n = 3, gestational age 28 to 40 weeks), mean differences versus standard-of-care monitors were –0.17 beats per minute for HR, 0.76 breaths per minute for RR, and 1.02% for SpO₂; additional studies in n = 18 neonates with gestational ages as low as 28 weeks and weights as low as 1470 g showed similar performance, and scaled production cost is estimated at less than $20 USD per unit.
Methods
- Fabrication used semiconductor processing, lamination, transfer printing, chip placement, and solder bonding; each EES includes serpentine Cu traces, chip-scale ICs, fractal mesh electrodes, an ionic-liquid microfluidic chamber, silicone encapsulation, and an NFC loop antenna for simultaneous wireless power and data.
- In-sensor analytics include a streamlined Pan-Tompkins algorithm for real-time QRS detection to yield HR and HRV, dynamic baseline control and SpO₂ calculation for PPG, and time-synchronized ECG/PPG processing for PAT.
- Validation included healthy adult volunteers (n = 3) against FDA-cleared monitoring equipment, and neonates in two tertiary-level NICUs: n = 3 with synchronous standard-of-care measurements and n = 18 using related device platforms with onboard power supplies; the protocol was IRB-approved and registered on ClinicalTrials.gov (NCT02865070).
Limitations
- The paper states that results are preliminary feasibility testing and validation; comprehensive clinical studies are ongoing to verify measurements across increased age groups and ethnic backgrounds, and remaining challenges include nursing acceptance, compatibility with legacy monitoring systems, and device sterilization for reuse.
- The paper notes that additional testing is needed for altered skin barrier function, such as burn victims or patients with epidermolysis bullosa.
- Evident from the text: the neonatal validation cohorts are small (n = 3 with synchronous standard-of-care comparisons; n = 18 using related platforms with onboard power supplies), adult validation is n = 3, and temperature and PAT data in neonates are shown without reference comparators because conventional devices acquire these only periodically.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures









Verification
Claims the paper makes
6 Valid · 2 Not valid · 18 Not checked · 82 Own results and interpretations
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“A key example is the measurement of pulse arrival time (PAT), defined by the time lapse between the maximum fiducial point in the ECG signal (R peak) and the corresponding minimal fiducial point in the PPG signal at valley as in Fig. 6H, as a direct correlate to systolic blood pressure (29 , 30).”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Real-time measurements, in-sensor analytics, and data transmission
What the sources say
- Partly supportedcites [29] Kim J. S., Chee Y. J., Park J. W., Choi J. W., Park K. S., A new approach for non-intrusive monitoring of blood pressure on a toilet seat. Physiol. Meas. 27, 203– 211 (2006). doi: 10.1088/0967-3334/27/2/010; pmid: 16400206 sourceExperimental — accuracy not yet measured
Section 4.1 reports that PAT was significantly, but moderately, correlated with systolic blood pressure. Section 2.2 defines the study’s PAT using the ECG R-peak and the PPG’s maximum first-derivative point, not the PPG valley; it says the minimum point was unclear in its signal. The claimed correlation is supported, but the attributed measurement definition is not. (LLM, conf=0.87)
- Not checked — no full text of the sourcecites [30] Geddes L. A., Voelz M. H., Babbs C. F., Bourland J. D., Tacker W. A., Pulse transit time as an indicator of arterial blood pressure. Psychophysiology 18, 71– 74 (1981). doi: 10.1111/j.1469-8986.1981.tb01545.x; pmid: 7465731 source
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The inherently thin, soft mechanical properties of the sensors (Fig. 1A) reported here allow for adhesion via van derWaals forces alone.”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Low-modulus mechanics, soft interface adhesion, and implications for neonatal skin safety
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The mechanical decoupling afforded by the microfluidic channel decreases these stresses by up to a factor of 2.5 (Fig. 2B) relative to otherwise similar designs without the microfluidics.”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Low-modulus mechanics, soft interface adhesion, and implications for neonatal skin safety
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Oscillations in the force only appear for holes larger than the characteristic size of the cohesive zone (~500 mm, as in Fig. 2I).”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Low-modulus mechanics, soft interface adhesion, and implications for neonatal skin safety
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Specifically, the interface starts to delaminate when the normal stress reaches ~20 kPa (Fig. 2G, inset).”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Low-modulus mechanics, soft interface adhesion, and implications for neonatal skin safety
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
61 more claims
Create a free account to see all 108 claims and to verify the 1 reference still open.
- Probably not a substantial claim. 14
- It may state two separate claims as one. 33
- The quote only partly states this claim. 2
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Researchers have developed various versatile epidermal electrodes with high performance that can universally measure EEG, EMG, EOG, and ECG through placing electrodes on different areas of the skin.[25,414]”
From Flexible Electronics and Devices as Human–Machine Interfaces for Medical Robotics (2022) · cites this paper as [414]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Previous reports validated the performance of an FDA-cleared physiological monitoring system against gold standard monitoring platforms [18–23].”
From Pregnancy outcomes in infertility patients diagnosed with sleep disordered breathing with wireless wearable sensors (2022) · cites this paper as [18]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“In previous work, we demonstrate the ability for the chest- and finger-mounted sensors to monitor vital signs in both children and adults; the abdominal sensor allows for maternal–fetal monitoring of pregnant women in both high- and low-resource settings as well as the clinical and home settings (22, 23).”
From Comprehensive pregnancy monitoring with a network of wireless, soft, and flexible sensors in high- and low-resource health settings (2021) · cites this paper as [23]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“and have been used in neonatal and pediatric intensive care units. 4,5”
From Advanced wireless monitoring for children in the cardiac perioperative setting (2023) · cites this paper as [4]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“However, this system is limited by (1) the modest maximum operating distances (~30 cm) supported by near-field communication (NFC) protocols used for power transfer and data communication, (2) the mechanically fragile nature of the ultrathin, compliant mechanics designs, (3) the sufficient, but constrained range of measurement capabilities, and (4) the demand for highly advanced device configur”
From Skin-interfaced biosensors for advanced wireless physiological monitoring in neonatal and pediatric intensive-care units (2020) · cites this paper as [9]
Experimental — accuracy not yet measured
5 more claims about this paper
Create a free account to see every claim other papers make about this one.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Compatibility with medical imaging techniques used in the NICU
“eliminating any parasitic heating from magnetically induced eddy currents.”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Compatibility with medical imaging techniques used in the NICU
“Full three-dimensional multi-physicsmodeling shows that, at the end of a single scan for 0.5 ms, the copper layer of an ECG EES undergoes heating by only 1°C (Fig. 3H).”
Citations
Cites
A Real-Time QRS Detection Algorithm
Pan, J., & Tompkins, W. J. (1985) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1985.325532 · Cited by 7,855
Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis
Gao, W., Emaminejad, S., Nyein, H. Y. Y., Challa, S., Chen, K., Peck, A., Fahad, H. M., Ota, H., Shiraki, H., Kiriya, D., et al. (2016) · Nature · doi:10.1038/nature16521 · Cited by 5,087
Kim, D.-H., Lu, N., Ma, R., Kim, Y.-S., Kim, R.-H., Wang, S., Wu, J., Won, S. M., Tao, H., Islam, A., et al. (2011) · Science · doi:10.1126/science.1206157 · Cited by 4,836
Self-citationWearable Sensors for Remote Health Monitoring
Majumder, S., Mondal, T., & Deen, M. (2017) · Sensors · doi:10.3390/s17010130 · Cited by 1,375
Kiserud, T., Piaggio, G., Carroli, G., Widmer, M., Carvalho, J., Neerup Jensen, L., Giordano, D., Cecatti, J. G., Abdel Aleem, H., Talegawkar, S. A., et al. (2017) · PLOS Medicine · doi:10.1371/journal.pmed.1002220 · Cited by 729
Thermochemistry of ionic liquid heat-transfer fluids
Valkenburg, M. E. V., Vaughn, R. L., Williams, M., & Wilkes, J. S. (2005) · Thermochimica Acta · doi:10.1016/j.tca.2004.11.013 · Cited by 595
Halcox, J. P. J., Wareham, K., Cardew, A., Gilmore, M., Barry, J. P., Phillips, C., & Gravenor, M. B. (2017) · Circulation · doi:10.1161/circulationaha.117.030583 · Cited by 580
Self-assembled three dimensional network designs for soft electronics
Jang, K.-I., Li, K., Chung, H. U., Xu, S., Jung, H. N., Yang, Y., Kwak, J. W., Jung, H. H., Song, J., Yang, C., et al. (2017) · Nature Communications · doi:10.1038/ncomms15894 · Cited by 445
Self-citationChen, W., Kobayashi, T., Ichikawa, S., Takeuchi, Y., & Togawa, T. (2000) · Medical & Biological Engineering & Computing · doi:10.1007/bf02345755 · Cited by 425
Harris, K. D., Elias, A. L., & Chung, H.-J. (2016) · Journal of Materials Science · doi:10.1007/s10853-015-9643-3 · Cited by 403
Self-citationEpidemiologic Trends in Neonatal Intensive Care, 2007-2012
Harrison, W., & Goodman, D. (2015) · JAMA Pediatrics · doi:10.1001/jamapediatrics.2015.1305 · Cited by 373
Skin photoplethysmography — a review
Kamal, A. A. R., Harness, J. B., Irving, G., & Mearns, A. J. (1989) · Computer Methods and Programs in Biomedicine · doi:10.1016/0169-2607(89)90159-4 · Cited by 328
Stretchable wireless system for sweat pH monitoring
Dang, W., Manjakkal, L., Navaraj, W. T., Lorenzelli, L., Vinciguerra, V., & Dahiya, R. (2018) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2018.02.025 · Cited by 326
Development of the Epidermis in the Newborn
Evans, N. J., & Rutter, N. (2009) · Biology of the Neonate · doi:10.1159/000242513 · Cited by 305
Pulse Transit Time as an Indicator of Arterial Blood Pressure
Geddes, L. A., Voelz, M. H., Babbs, C. F., Bourland, J. D., & Tacker, W. A. (1981) · Psychophysiology · doi:10.1111/j.1469-8986.1981.tb01545.x · Cited by 293
Gao, L., Zhang, Y., Malyarchuk, V., Jia, L., Jang, K.-I., Chad Webb, R., Fu, H., Shi, Y., Zhou, G., Shi, L., et al. (2014) · Nature Communications · doi:10.1038/ncomms5938 · Cited by 272
Self-citationNovel Wireless Devices for Cardiac Monitoring
Walsh, J. A., Topol, E. J., & Steinhubl, S. R. (2014) · Circulation · doi:10.1161/circulationaha.114.009024 · Cited by 175
Zhang, Y., Webb, R. C., Luo, H., Xue, Y., Kurniawan, J. F., Cho, N. H., Krishnan, S. R., Li, Y., Huang, Y. Y., & Rogers, J. A. (2015) · Advanced Healthcare Materials · doi:10.1002/adhm.201500110 · Cited by 131
Self-citationThe 2003 National Pediatric Pressure Ulcer and Skin Breakdown Prevalence Survey: A Multisite Study
McLane, K. M., Bookout, K., McCord, S., McCain, J., & Jefferson, L. S. (2004) · Journal of Wound, Ostomy and Continence Nursing · doi:10.1097/00152192-200407000-00004 · Cited by 129
Smart Jacket Design for Neonatal Monitoring with Wearable Sensors
Bouwstra, S., Chen, W., Feijs, L., & Oetomo, S. B. (2009) · 2009 Sixth International Workshop on Wearable and Implantable Body Sensor Networks · doi:10.1109/bsn.2009.40 · Cited by 129
Lund, C. (2014) · Newborn and Infant Nursing Reviews · doi:10.1053/j.nainr.2014.10.001 · Cited by 128
The Survey on Near Field Communication
Coskun, V., Ozdenizci, B., & Ok, K. (2015) · Sensors · doi:10.3390/s150613348 · Cited by 122 · cited 1×
“56 MHz supports only low-speed, low-fidelity applications such as contactless payments and wireless identification (15); thus, substantial modification in both the transponder and host reader systems at ISO15693was required to support data transfer rates sufficient for NICU monitoring (hundreds of Hz).”
Background陳, 文. (2001) · Medical Entomology and Zoology · Cited by 116
Pressure Ulcers in the Hospitalized Neonate: Rates and Risk Factors
Visscher, M., & Taylor, T. (2014) · Scientific Reports · doi:10.1038/srep07429 · Cited by 104
Ahlstrom, C., Johansson, A., Uhlin, F., Länne, T., & Ask, P. (2005) · Journal of Artificial Organs · doi:10.1007/s10047-005-0301-4 · Cited by 104 · cited 1×
“The Moens-Korteweg equation provides a linear relationship between PAT and BP (38, 39).”
MethodMa, Y., Pharr, M., Wang, L., Kim, J., Liu, Y., Xue, Y., Ning, R., Wang, X., Chung, H. U., Feng, X., et al. (2017) · Small · doi:10.1002/smll.201602954 · Cited by 100 · cited 2×
“between the interconnected components and the skin (14).”
BackgroundSelf-citationThe scars of newborn intensive care
Cartlidge, P. H. T., Fox, P. E., & Rutter, N. (1990) · Early Human Development · doi:10.1016/0378-3782(90)90105-r · Cited by 95
Medical Device-Related Hospital-Acquired Pressure Ulcers in Children: An Integrative Review
Murray, J. S., Noonan, C., Quigley, S., & Curley, M. A. Q. (2013) · Journal of Pediatric Nursing · doi:10.1016/j.pedn.2013.05.004 · Cited by 91
Blood pressure disorders in the neonate: Hypotension and hypertension
Fanaroff, J. M., & Fanaroff, A. A. (2006) · Seminars in Fetal and Neonatal Medicine · doi:10.1016/j.siny.2006.01.002 · Cited by 84
Gent, A. N., & Hamed, G. R. (1975) · The Journal of Adhesion · doi:10.1080/00218467508075041 · Cited by 81
Radiation Exposure in 212 Very Low and Extremely Low Birth Weight Infants
Puch-Kapst, K., Juran, R., Stoever, B., & Wauer, R. R. (2009) · Pediatrics · doi:10.1542/peds.2008-1028 · Cited by 71 · cited 1×
“The EES eliminates radio-opaque wires, thereby improving evaluation by x-ray imaging, a modality required for 90% of low–birth weight neonates (26).”
BackgroundStimuli-responsive liquid crystal elastomers for dynamic cell culture
Agrawal, A., Adetiba, O., Kim, H., Chen, H., Jacot, J. G., & Verduzco, R. (2015) · Journal of materials research/Pratt's guide to venture capital sources · doi:10.1557/jmr.2014.392 · Cited by 66
A new approach for non-intrusive monitoring of blood pressure on a toilet seat
Kim, J. S., Chee, Y. J., Park, J. W., Choi, J. W., & Park, K. S. (2006) · Physiological Measurement · doi:10.1088/0967-3334/27/2/010 · Cited by 64
Self-citationBaserga, M. C., Puri, A., & Sola, A. (2002) · Journal of Perinatology · doi:10.1038/sj.jp.7210713 · Cited by 64 · cited 1×
“Although arterial lines offer a continuous measurement of blood pressure in neonates, these invasive interventions can cause thrombosis, hematomas, infection, and even death (34).”
BackgroundSmyser, C. D., Kidokoro, H., & Inder, T. E. (2012) · Journal of Paediatrics and Child Health · doi:10.1111/j.1440-1754.2012.02535.x · Cited by 61
A problem of generalized magneto-thermoelastic thin slim strip subjected to a moving heat source
He, T., & Cao, L. (2009) · Mathematical and Computer Modelling · doi:10.1016/j.mcm.2008.12.004 · Cited by 53
A Comparison of Blood Pressure Measurements in Newborns
O'Shea, J., & Dempsey, E. (2009) · American Journal of Perinatology · doi:10.1055/s-0028-1091391 · Cited by 49 · cited 1×
“Cuff-based blood pressure measurements with sphygmomanometers fail to provide continuous measurement, overestimate blood pressure in premature neonates (32), and pose a direct risk for pressure-related injuries (33).”
BackgroundRutter, N. (1988) · British Medical Bulletin · doi:10.1093/oxfordjournals.bmb.a072303 · Cited by 47
Variability in incubator humidity practices in the management of preterm infants
Sinclair, L., Crisp, J., & Sinn, J. (2009) · Journal of Paediatrics and Child Health · doi:10.1111/j.1440-1754.2009.01555.x · Cited by 40
Wippermann, C. F., Schranz, D., & Huth, R. G. (1995) · Journal of Clinical Monitoring · doi:10.1007/bf01616991 · Cited by 36
Galland, B. C., Tan, E., & Taylor, B. J. (2007) · Sleep · doi:10.1093/sleep/30.7.891 · Cited by 34
The use of pulse transit time in pediatric sleep studies: A systematic review
Smith, L. A., Dawes, P. J., & Galland, B. C. (2018) · Sleep Medicine Reviews · doi:10.1016/j.smrv.2016.11.006 · Cited by 22
Melbourne, L., Chang, T., Murnick, J., Zaniletti, I., Glass, P., & Massaro, A. N. (2016) · Journal of Perinatology · doi:10.1038/jp.2016.116 · Cited by 13 · cited 1×
“Magnetic resonance imaging (MRI) is essential in the NICU because of its ability to deliver precise assessment of whitematter, graymatter, and posterior fossa abnormalities with functional capabilities that exceed those of ultrasound (23, 24).”
BackgroundPresence and pattern of scarring in children born very preterm
Tottman, A. C., Alsweiler, J. M., Bloomfield, F. H., & Harding, J. E. (2018) · Archives of Disease in Childhood - Fetal and Neonatal Edition · doi:10.1136/archdischild-2016-311999 · Cited by 12
Research works printed but not linked: 2
- O. Sharma, S. N. Lewis, U. Telang, L. D’Almeida, L. E. S. Lewis, Design of a Bluetooth Enabled Health Monitoring System for Infants Using Wearable Technology. J. Adv. Res. Dyn. Contrl. Syst. 15, 887–894 (2017). http://jardcs.org/papers/v9/sp/20181075.pdf
- C. H. Lund J. A. Tucker in Neonatal Skin: Structure and Function S. B. Hoath H. I. Maibach Eds. (Dekker ed. 2 2003) pp. 299–324.
Cited by
Technology Roadmap for Flexible Sensors
Luo, Y., Abidian, M. R., Ahn, J.-H., Akinwande, D., Andrews, A. M., Antonietti, M., Bao, Z., Berggren, M., Berkey, C. A., Bettinger, C. J., et al. (2023) · ACS Nano · doi:10.1021/acsnano.2c12606 · Cited by 1,383
Self-citationSkin-integrated wireless haptic interfaces for virtual and augmented reality
Yu, X., Xie, Z., Yu, Y., Lee, J., Vazquez-Guardado, A., Luan, H., Ruban, J., Ning, X., Akhtar, A., Li, D., et al. (2019) · Nature · doi:10.1038/s41586-019-1687-0 · Cited by 1,025
Self-citationA New Frontier of Printed Electronics: Flexible Hybrid Electronics
Khan, Y., Thielens, A., Muin, S., Ting, J., Baumbauer, C., & Arias, A. C. (2020) · Advanced Materials · doi:10.1002/adma.201905279 · Cited by 891
Sign-to-speech translation using machine-learning-assisted stretchable sensor arrays
Zhou, Z., Chen, K., Li, X., Zhang, S., Wu, Y., Zhou, Y., Meng, K., Sun, C., He, Q., Fan, W., et al. (2020) · Nature Electronics · doi:10.1038/s41928-020-0428-6 · Cited by 841
Skin-Interfaced Wearable Sweat Sensors for Precision Medicine
Min, J., Tu, J., Xu, C., Lukas, H., Shin, S., Yang, Y., Solomon, S. A., Mukasa, D., & Gao, W. (2023) · Chemical Reviews · doi:10.1021/acs.chemrev.2c00823 · Cited by 712
Metadata
- Authors
- Chung, H. U., Kim, B. H., Lee, J. Y., Lee, J., Xie, Z., Ibler, E. M., Lee, K., Banks, A., Jeong, J. Y., Kim, J., Ogle, C., Grande, D., Yu, Y., Jang, H., Assem, P., Ryu, D., Kwak, J. W., Namkoong, M., Park, J. B., Lee, Y., Kim, D. H., Ryu, A., Jeong, J., You, K., Ji, B., Liu, Z., Huo, Q., Feng, X., Deng, Y., Xu, Y., Jang, K.-I., Kim, J., Zhang, Y., Ghaffari, R., Rand, C. M., Schau, M., Hamvas, A., Weese-Mayer, D. E., Huang, Y., Lee, S. M., Lee, C. H., Shanbhag, N. R., Paller, A. S., Xu, S., & Rogers, J. A.
- Affiliations
- Northwestern University (US)
- University of Illinois Urbana-Champaign (US)
- Department of Materials Science and Engineering, Northwestern University, Evanston, IL 60208, USA.
- Agency for Science, Technology and Research (SG)
- Tsinghua University (CN)
- Daegu Gyeongbuk Institute of Science and Technology (KR)
- Lurie Children's Hospital (US)
- Center for Autonomic Medicine, Department of Pediatrics, Ann & Robert H. Lurie Children’s Hospital of Chicago, Chicago, IL 60611, USA.
- Kookmin University (KR)
- Purdue University West Lafayette (US)
- Corresponding author
- Chung, Ha Uk
- Kim, Bong Hoon
- Lee, Jong Yoon
- Lee, Jungyup
- Xie, Zhaoqian
- Paller, Amy S.
- Xu, Shuai
- Rogers, John A.
- ORCID
- Chung, Ha Uk
- Kim, Bong Hoon
- Lee, Jong Yoon
- Lee, Jungyup
- Xie, Zhaoqian
- Lee, KunHyuck
- Banks, Anthony
- Jeong, Ji Yoon
- Kim, Jongwon
- Grande, Dominic
- Yu, Yongjoon
- Jang, Hokyung
- Assem, Pourya
- Kwak, Jean Won
- Namkoong, Myeong
- Park, Jun Bin
- Lee, Yechan
- Kim, Do Hoon
- Ryu, Arin
- You, Kevin
- Ji, Bowen
- Liu, Zhuangjian
- Huo, Qingze
- Feng, Xue
- Deng, Yujun
- Xu, Yeshou
- Jang, Kyung-In
- Kim, Jeonghyun
- Zhang, Yihui
- Ghaffari, Roozbeh
- Rand, Casey M.
- Hamvas, Aaron
- Huang, Yonggang
- Lee, Seung Min
- Lee, Chi Hwan
- Shanbhag, Naresh R.
- Paller, Amy S.
- Xu, Shuai
- Rogers, John A.
- Venue
- Science
- Published
- 2019-03
- Volume
- 363
- Issue
- 6430
- ISSN
- 0036-8075 (ISSN-L)
- 1095-9203
- PMC
- PMC6510306
- OpenAlex
- W2916533817
- Semantic Scholar
- b0511afac0cc252946aa8f08a078b12e48030581
- Cited by
- 807
- References
- 43
- Citations by year
- 2026: 72 · 2025: 112 · 2024: 103 · 2023: 121 · 2022: 124 · 2021: 131 · 2020: 97 · 2019: 47
- FWCI
- 35.10
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by
- Locations
- Science (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Funders
- Bill and Melinda Gates Foundation (OPP1182909)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Physical Sciences › Engineering › Biomedical Engineering › Wireless Body Area Networks
- Keywords
- epidermal electronics, neonatal intensive care, battery-free operation, neonatal monitoring
- MeSH terms
- Diagnostic Imaging, Electronics, Equipment Design, Humans, Infant, Newborn, Monitoring, Physiologic, Skin, Intensive Care, Neonatal, Vital Signs, Lab-On-A-Chip Devices, Wireless Technology
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.