Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care
Open accessThe DOI matches its Crossref record; 44 of the 44 references it cites were found in a registry or an index; 12 of the 42 sources its claims cite were checked for support.
Abstract
Sensitive sensing Neonatal care, particularly for premature babies, is complicated by the infants' fragility and by the need for a large number of tethered sensors to be attached to their tiny bodies. Chung et al. developed a pair of sensors that only require water to adhere to the skin and allow for untethered monitoring of key vital signs (see the Perspective by Guinsburg). On-board data processing allowed for efficient wireless near-field communication using standard protocols. The absence of cables makes it easier to handle the infants and allows for skin-to-skin contact between the babies and their parents or caregivers. Science , this issue p. eaau0780 ; see also p. 924
Overview
CoChat Summary
TL;DR
The paper introduces a pair of ultrathin, wireless, battery-free epidermal electronic systems (EES) for neonatal intensive care: one chest-mounted ECG device and one foot-mounted PPG device. Pilot studies in neonates show clinical-grade agreement for heart rate, respiratory rate, and SpO₂, plus added capabilities for skin temperature and pulse arrival time as a systolic blood pressure surrogate.
Key findings
- The binodal EES continuously streams ECG, PPG, and skin temperature data, yielding HR, HRV, RR, SpO₂, and PAT; wireless data rates reach up to 800 bytes/s with dual channels, NFC operation works up to 25 cm through NICU materials, and BLE range reaches up to 20 m.
- The devices have effective moduli of 200 to 300 kPa, adhere via van der Waals forces, and show peel forces about a factor of ~10 lower than conventional NICU adhesives, with maximum von Mises stress on skin reduced by a factor of 4.3; the microfluidic channel reduces interfacial stresses by up to a factor of 2.5, and adhesion energy is G = 16 N/m.
- For MRI compatibility, mesh electrodes reduce the in-plane magnetic field gradient by a factor of ~3 and the out-of-plane gradient by a factor of 4 versus commercial electrodes; simulated copper heating is 1°C with maximum skin temperature change of 0.04°C, and experimental 3-T MRI measurements show ~0.1°C temperature difference, meeting FDA “MRI-safe” requirements.
- In neonates (n = 3, gestational age 28 to 40 weeks), mean differences versus standard-of-care monitors were –0.17 beats per minute for HR, 0.76 breaths per minute for RR, and 1.02% for SpO₂; additional studies in n = 18 neonates with gestational ages as low as 28 weeks and weights as low as 1470 g showed similar performance, and scaled production cost is estimated at less than $20 USD per unit.
Methods
- Fabrication used semiconductor processing, lamination, transfer printing, chip placement, and solder bonding; each EES includes serpentine Cu traces, chip-scale ICs, fractal mesh electrodes, an ionic-liquid microfluidic chamber, silicone encapsulation, and an NFC loop antenna for simultaneous wireless power and data.
- In-sensor analytics include a streamlined Pan-Tompkins algorithm for real-time QRS detection to yield HR and HRV, dynamic baseline control and SpO₂ calculation for PPG, and time-synchronized ECG/PPG processing for PAT.
- Validation included healthy adult volunteers (n = 3) against FDA-cleared monitoring equipment, and neonates in two tertiary-level NICUs: n = 3 with synchronous standard-of-care measurements and n = 18 using related device platforms with onboard power supplies; the protocol was IRB-approved and registered on ClinicalTrials.gov (NCT02865070).
Limitations
- The paper states that results are preliminary feasibility testing and validation; comprehensive clinical studies are ongoing to verify measurements across increased age groups and ethnic backgrounds, and remaining challenges include nursing acceptance, compatibility with legacy monitoring systems, and device sterilization for reuse.
- The paper notes that additional testing is needed for altered skin barrier function, such as burn victims or patients with epidermolysis bullosa.
- Evident from the text: the neonatal validation cohorts are small (n = 3 with synchronous standard-of-care comparisons; n = 18 using related platforms with onboard power supplies), adult validation is n = 3, and temperature and PAT data in neonates are shown without reference comparators because conventional devices acquire these only periodically.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures









Verification
Claims the paper makes
6 Valid · 2 Not valid · 18 Not checked · 82 Own results and interpretations
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“A key example is the measurement of pulse arrival time (PAT), defined by the time lapse between the maximum fiducial point in the ECG signal (R peak) and the corresponding minimal fiducial point in the PPG signal at valley as in Fig. 6H, as a direct correlate to systolic blood pressure (29 , 30).”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Real-time measurements, in-sensor analytics, and data transmission
What the sources say
- Partly supportedcites [29] Kim J. S., Chee Y. J., Park J. W., Choi J. W., Park K. S., A new approach for non-intrusive monitoring of blood pressure on a toilet seat. Physiol. Meas. 27, 203– 211 (2006). doi: 10.1088/0967-3334/27/2/010; pmid: 16400206 sourceExperimental — accuracy not yet measured
Section 4.1 reports that PAT was significantly, but moderately, correlated with systolic blood pressure. Section 2.2 defines the study’s PAT using the ECG R-peak and the PPG’s maximum first-derivative point, not the PPG valley; it says the minimum point was unclear in its signal. The claimed correlation is supported, but the attributed measurement definition is not. (LLM, conf=0.87)
- Not checked — no full text of the sourcecites [30] Geddes L. A., Voelz M. H., Babbs C. F., Bourland J. D., Tacker W. A., Pulse transit time as an indicator of arterial blood pressure. Psychophysiology 18, 71– 74 (1981). doi: 10.1111/j.1469-8986.1981.tb01545.x; pmid: 7465731 source
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The inherently thin, soft mechanical properties of the sensors (Fig. 1A) reported here allow for adhesion via van derWaals forces alone.”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Low-modulus mechanics, soft interface adhesion, and implications for neonatal skin safety
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The mechanical decoupling afforded by the microfluidic channel decreases these stresses by up to a factor of 2.5 (Fig. 2B) relative to otherwise similar designs without the microfluidics.”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Low-modulus mechanics, soft interface adhesion, and implications for neonatal skin safety
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Oscillations in the force only appear for holes larger than the characteristic size of the cohesive zone (~500 mm, as in Fig. 2I).”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Low-modulus mechanics, soft interface adhesion, and implications for neonatal skin safety
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Specifically, the interface starts to delaminate when the normal stress reaches ~20 kPa (Fig. 2G, inset).”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Low-modulus mechanics, soft interface adhesion, and implications for neonatal skin safety
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
61 more claims
Create a free account to see all 108 claims and to verify the 1 reference still open.
- Probably not a substantial claim. 14
- It may state two separate claims as one. 33
- The quote only partly states this claim. 2
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Researchers have developed various versatile epidermal electrodes with high performance that can universally measure EEG, EMG, EOG, and ECG through placing electrodes on different areas of the skin.[25,414]”
From Flexible Electronics and Devices as Human–Machine Interfaces for Medical Robotics (2022) · cites this paper as [414]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Previous reports validated the performance of an FDA-cleared physiological monitoring system against gold standard monitoring platforms [18–23].”
From Pregnancy outcomes in infertility patients diagnosed with sleep disordered breathing with wireless wearable sensors (2022) · cites this paper as [18]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“In previous work, we demonstrate the ability for the chest- and finger-mounted sensors to monitor vital signs in both children and adults; the abdominal sensor allows for maternal–fetal monitoring of pregnant women in both high- and low-resource settings as well as the clinical and home settings (22, 23).”
From Comprehensive pregnancy monitoring with a network of wireless, soft, and flexible sensors in high- and low-resource health settings (2021) · cites this paper as [23]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“and have been used in neonatal and pediatric intensive care units. 4,5”
From Advanced wireless monitoring for children in the cardiac perioperative setting (2023) · cites this paper as [4]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“However, this system is limited by (1) the modest maximum operating distances (~30 cm) supported by near-field communication (NFC) protocols used for power transfer and data communication, (2) the mechanically fragile nature of the ultrathin, compliant mechanics designs, (3) the sufficient, but constrained range of measurement capabilities, and (4) the demand for highly advanced device configur”
From Skin-interfaced biosensors for advanced wireless physiological monitoring in neonatal and pediatric intensive-care units (2020) · cites this paper as [9]
Experimental — accuracy not yet measured
5 more claims about this paper
Create a free account to see every claim other papers make about this one.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Compatibility with medical imaging techniques used in the NICU
“eliminating any parasitic heating from magnetically induced eddy currents.”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Compatibility with medical imaging techniques used in the NICU
“Full three-dimensional multi-physicsmodeling shows that, at the end of a single scan for 0.5 ms, the copper layer of an ECG EES undergoes heating by only 1°C (Fig. 3H).”
Citations
Cites
A Real-Time QRS Detection Algorithm
Pan, J., & Tompkins, W. J. (1985) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1985.325532 · Cited by 7,855
Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis
Gao, W., Emaminejad, S., Nyein, H. Y. Y., Challa, S., Chen, K., Peck, A., Fahad, H. M., Ota, H., Shiraki, H., Kiriya, D., et al. (2016) · Nature · doi:10.1038/nature16521 · Cited by 5,087
Kim, D.-H., Lu, N., Ma, R., Kim, Y.-S., Kim, R.-H., Wang, S., Wu, J., Won, S. M., Tao, H., Islam, A., et al. (2011) · Science · doi:10.1126/science.1206157 · Cited by 4,836
Self-citationWearable Sensors for Remote Health Monitoring
Majumder, S., Mondal, T., & Deen, M. (2017) · Sensors · doi:10.3390/s17010130 · Cited by 1,375
Kiserud, T., Piaggio, G., Carroli, G., Widmer, M., Carvalho, J., Neerup Jensen, L., Giordano, D., Cecatti, J. G., Abdel Aleem, H., Talegawkar, S. A., et al. (2017) · PLOS Medicine · doi:10.1371/journal.pmed.1002220 · Cited by 729
Thermochemistry of ionic liquid heat-transfer fluids
Valkenburg, M. E. V., Vaughn, R. L., Williams, M., & Wilkes, J. S. (2005) · Thermochimica Acta · doi:10.1016/j.tca.2004.11.013 · Cited by 595
Halcox, J. P. J., Wareham, K., Cardew, A., Gilmore, M., Barry, J. P., Phillips, C., & Gravenor, M. B. (2017) · Circulation · doi:10.1161/circulationaha.117.030583 · Cited by 580
Self-assembled three dimensional network designs for soft electronics
Jang, K.-I., Li, K., Chung, H. U., Xu, S., Jung, H. N., Yang, Y., Kwak, J. W., Jung, H. H., Song, J., Yang, C., et al. (2017) · Nature Communications · doi:10.1038/ncomms15894 · Cited by 445
Self-citationChen, W., Kobayashi, T., Ichikawa, S., Takeuchi, Y., & Togawa, T. (2000) · Medical & Biological Engineering & Computing · doi:10.1007/bf02345755 · Cited by 425
Harris, K. D., Elias, A. L., & Chung, H.-J. (2016) · Journal of Materials Science · doi:10.1007/s10853-015-9643-3 · Cited by 403
Self-citation
Research works printed but not linked: 2
- O. Sharma, S. N. Lewis, U. Telang, L. D’Almeida, L. E. S. Lewis, Design of a Bluetooth Enabled Health Monitoring System for Infants Using Wearable Technology. J. Adv. Res. Dyn. Contrl. Syst. 15, 887–894 (2017). http://jardcs.org/papers/v9/sp/20181075.pdf
- C. H. Lund J. A. Tucker in Neonatal Skin: Structure and Function S. B. Hoath H. I. Maibach Eds. (Dekker ed. 2 2003) pp. 299–324.
Cited by
Items 501–600 of 800
The use of wireless sensors in the neonatal intensive care unit: a study protocol
Sénéchal, E., Radeschi, D. J., Tao, L., Lv, S., Jeanne, E., Kearney, R. E., Shalish, W., & SantʼAnna, G. (2023) · PeerJ · doi:10.7717/peerj.15578 · Cited by 17
Ionic hydrogels-based triboelectric nanogenerators for self-powered human–machine interfaces
Liang, S., Li, C., Niu, M., Zhu, P., Pan, Z., & Mao, Y. (2023) · Journal of Physics Materials · doi:10.1088/2515-7639/ad05e7 · Cited by 18
“These self-powered HMIs can be integrated with medical sensors and devices, eliminating the need for external power sources and enabling real-time health monitoring and data transmission [173–175].”
BackgroundSpray-on electronic tattoos with MXene and liquid metal nanocomposites
Wang, L., Lin, Y. Q., Yang, C., Wang, Q., Fang, T., Bai, C., Wang, J., & Kong, D. (2024) · Chemical Engineering Journal · doi:10.1016/j.cej.2024.157504 · Cited by 16
Wei, B., Chen, G., & Wang, Q.-W. (2022) · IEEE Sensors Journal · doi:10.1109/jsen.2022.3189053 · Cited by 14
A wireless fully-passive acquisition of biopotentials
Liu, S., Meng, X., Zhang, J., & Chae, J. (2019) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2019.111336 · Cited by 16
Athanasiadis, M., Afanasenkau, D., Derks, W. J. A., Tondera, C., Murganti, F., Busskamp, V., Bergmann, O., & Minev, I. R. (2020) · npj Flexible Electronics · doi:10.1038/s41528-020-0075-z · Cited by 15
Engineering Steep Subthreshold Swings in High‐Performance Organic Field‐Effect Transistor Sensors
Wang, Y., Huang, W., Li, J., Liu, S., Fu, J., Wang, L., Wang, H., Li, W., Xie, L., Ling, H., et al. (2024) · Small · doi:10.1002/smll.202406522 · Cited by 15
A Mechanically Durable Hydrogel Synapse
Tian, H., Zhou, R., Ke, L., Qian, K., Liu, Y., & Hao, J. (2025) · Advanced Functional Materials · doi:10.1002/adfm.202505232 · Cited by 13
Thiruchelvam, P., & Dasmahapatra, A. K. (2025) · Small · doi:10.1002/smll.202502286 · Cited by 17
Liao, Z., Golparvar, A. J., Bathaei, M. J., Cardoso, F. A., & Boutry, C. M. (2025) · Advanced Science · doi:10.1002/advs.202509934 · Cited by 20
Wu, S., Han, Y., Kang, M., Zhou, Y., & Zhang, Y. (2025) · Nano Energy · doi:10.1016/j.nanoen.2025.111615 · Cited by 14
Self-citationMechanics of transfer printing for elastomeric stamps with collapse cavities
Zhang, Y., Lu, T., Zhao, J., Ai, J., Chen, Y., Shi, M., Ma, Y., & Feng, X. (2023) · Extreme Mechanics Letters · doi:10.1016/j.eml.2023.101956 · Cited by 14
Self-citationUltra-Low Power Wearable Infant Sleep Position Sensor
Yun, I., Jeung, J., Kim, M., Kim, Y.-S., & Chung, Y. (2019) · Sensors · doi:10.3390/s20010061 · Cited by 13
Guo, X. M., Avila, R. O., Huang, Y. Y., & Xie, Z. (2021) · International journal of mechanical system dynamics · doi:10.1002/msd2.12017 · Cited by 13
Self-citationMaranha, M., Silva, A. F., Lopes, P. A., de Almeida, A. T., & Tavakoli, M. (2022) · Advanced Sensor Research · doi:10.1002/adsr.202200025 · Cited by 12
Supercooled Liquid Ga Stretchable Electronics
Wang, B., Mašlík, J., Hellman, O., Gumiero, A., & Hjort, K. (2023) · Advanced Functional Materials · doi:10.1002/adfm.202300036 · Cited by 20
Biopolymers-based skin-interfaced triboelectric sensors
Xu, S., Barbosa, P. H. D. S., & Wu, W. (2023) · Nano Research · doi:10.1007/s12274-023-5784-x · Cited by 16
Self-citationNonprinted IC Textiles for Wearable Electronics
Wei, X., Zhang, W., & Fan, X. H. (2022) · Accounts of Materials Research · doi:10.1021/accountsmr.2c00150 · Cited by 13
Recent Advances in Materials, Designs and Applications of Skin Electronics
Yao, K., Yang, Y., Wu, P., Zhao, G., Wang, L., & Yu, X. (2022) · IEEE Open Journal of Nanotechnology · doi:10.1109/ojnano.2022.3218960 · Cited by 13
“[45] (c) Stretchable silicon nanomembrane using fractal patterns in Peano-curve.”
MethodLong-distance multifunctional wireless sensing platform for identifying and ranging
Si, J., Han, L., Wang, R., Wu, C., Guo, M., & Li, J. (2023) · Nano Energy · doi:10.1016/j.nanoen.2023.108267 · Cited by 14
Multifunctional Nanocomposite Yield-Stress Fluids for Printable and Stretchable Electronics
Lu, Q., Sun, Y., Wu, M., Wang, Q., Feng, S., Fang, T., Hu, G., Huang, W., Li, Z., Kong, D., et al. (2024) · ACS Nano · doi:10.1021/acsnano.4c01668 · Cited by 12
Horii, T., Yamashita, K., Ito, M., Okada, K., & Fujie, T. (2024) · NPG Asia Materials · doi:10.1038/s41427-024-00553-9 · Cited by 17
Novel Conformal Skin Patch with Embedded Thin-Film Electrodes for Early Detection of Extravasation
Lim, R., Damalerio, R., Bong, C. L., Tan, S. K., & Cheng, M. (2021) · Sensors · doi:10.3390/s21103429 · Cited by 11
Ivić, J., Petritz, A., Irimia, C. V., Kahraman, B., Kanbur, Y., Bednorz, M., Yumusak, C., Aslam, M. A., Matković, A., Saller, K. M., et al. (2022) · Advanced Sustainable Systems · doi:10.1002/adsu.202200234 · Cited by 12
Bioelectronic devices for light-based diagnostics and therapies
Kim, H., Sritandi, W., Xiong, Z., & Ho, J. S. (2023) · Biophysics Reviews · doi:10.1063/5.0102811 · Cited by 12
Continuous sensing and quantification of body motion in infants: A systematic review
Peng, Z. Q., Kommers, D. R., Liang, R.-H., Long, X., Cottaar, W., Niemarkt, H. J., Andriessen, P., & van Pul, C. (2023) · Heliyon · doi:10.1016/j.heliyon.2023.e18234 · Cited by 12
Sensors for Neonatal Monitoring
Güler, Ü., Sen, D., Costanzo, I., Tufan, T. B., & Rhein, L. M. (2022) · Elsevier eBooks · doi:10.1016/b978-0-12-822548-6.00104-7 · Cited by 16
A Fully-Analogue Light-to-Frequency Converter Circuit for Optical Sensing Applications
Stanchieri, G. D. P., De Marcellis, A., Faccio, M., Palange, E., & Guler, U. (2022) · IEEE Sensors Journal · doi:10.1109/jsen.2022.3187943 · Cited by 11
Li, M., Chen, D., Deng, X., Xu, B., Li, M., Liang, H., Wang, M., Song, G., Zhang, T., & Liu, Y. (2023) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.2c23002 · Cited by 17
Ghomrawi, H., Many, B. T., Holl, J. L., Ahmed, A. G., Jackson, M. E., Sibley, J., Khan, R., Kaufmann, E. E., Appeadu‐Mensah, W., & Abdullah, F. (2023) · International Journal of Medical Informatics · doi:10.1016/j.ijmedinf.2023.105046 · Cited by 12
Anti-Motion Interference Electrocardiograph Monitoring System: A Review
Shen, J., Li, X., Wang, Y., Li, Y., Bian, J., Zhu, X., He, X., & Li, J. (2024) · IEEE Sensors Journal · doi:10.1109/jsen.2024.3383872 · Cited by 13
“Chung et al. [157] also developed an FPCB for the wireless and battery-free human body biological signals monitoring system, which exploits a bimodal pair of ultrathin and low-modulus measurement modules that refer to an epidermal electronic system (EES).”
BackgroundZhou, K., Ding, R., Ye, W., Huang, J. T., Zhou, Z., Huang, L., Ma, X., Fan, Z., & Lin, Y. (2024) · Nano Letters · doi:10.1021/acs.nanolett.4c04328 · Cited by 11
An Infra-Red-Based Prototype for a Miniaturized Transcutaneous Carbon Dioxide Monitor
Tufan, T. B., Sen, D., & Güler, Ü. (2021) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc46164.2021.9630469 · Cited by 12
“New developments in next-generation biomedical device technologies bring healthcare to homes with wireless, remote and continuous monitoring of vital signs such as cardiac and respiratory parameters [1], [2].”
BackgroundGe, C., Duan, Z., Li, R., Chen, H., Li, T., Hu, P., Wang, Z., Zhao, J., & Zhang, J. (2022) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2022.113676 · Cited by 11
Yang, J., Li, Z., Zhang, X., Chen, B., Feng, J., & Wang, T. (2022) · Cell Reports Physical Science · doi:10.1016/j.xcrp.2022.101101 · Cited by 11
Wireless monitoring and artificial intelligence: A bright future in cardiothoracic surgery
Kalfa, D. M., Agrawal, S. K., Goldshtrom, N., LaPar, D. J., & BACHA, E. A. (2019) · Journal of Thoracic and Cardiovascular Surgery · doi:10.1016/j.jtcvs.2019.08.141 · Cited by 12
Bai, W., Irie, M., Liu, Z., Luan, H., Franklin, D., Nandoliya, K. R., Guo, H., Zang, H., Weng, Y., Lu, D., et al. (2021) · BME Frontiers · doi:10.34133/2021/8653218 · Cited by 13
“…of tissue microenvironments can serve as simple, yet important, diagnostic metrics relevant to a wide range of diseases and disorders [1 – 6], including those associated with chronic in fl ammation, traumatic injury, immunological irregularities, infections, and transplant rejection…”
BackgroundSelf-citationPhysical Intelligence in Small‐Scale Robots and Machines
Chen, H., & Sitti, M. (2025) · Advanced Materials · doi:10.1002/adma.202510332 · Cited by 27
Yang, B., Xia, Y., Zhi, X., Liu, K., Li, M., & Wang, X. (2024) · Device · doi:10.1016/j.device.2023.100246 · Cited by 10
Advancing rehabilitation: Knittable fiber-shaped sensors for monitoring rotator cuff injury recovery
Zhang, H., Li, G., Wei, H., Wu, X., Yan, S., Xia, Y., Yang, S., Schmuki, P., Cao, R., & Zhu, M. (2024) · Chemical Engineering Journal · doi:10.1016/j.cej.2024.150470 · Cited by 12
3D Woven Liquid Metals for Radio‐Frequency Stretchable Circuits
Rahman, M. S., Tiwari, A. P., Agnew, S. A., & Scheideler, W. J. (2024) · Advanced Materials Technologies · doi:10.1002/admt.202400339 · Cited by 10
Guo, X., Liu, Y., Zhou, Z., Peng, H., Gao, Y., Shao, M., & Yu, Y. (2025) · Advanced Science · doi:10.1002/advs.202414285 · Cited by 10
Self-citationYang, F., Tong, Y., Zhang, S., Zheng, J., Zhu, J., Liu, H., Hu, Y., Zhang, C., Chen, Z., Chen, G., et al. (2025) · IEEE Transactions on Antennas and Propagation · doi:10.1109/tap.2025.3564785 · Cited by 13
Skin-inspired electrochemical tactility and luminescence
Kim, T. Y., Choi, H., Choi, H.-S., Kim, J. S., Kim, D. H., & Jeong, U. (2022) · Electrochimica Acta · doi:10.1016/j.electacta.2022.140259 · Cited by 9
Self-citationHigh‐Conductivity Crack‐Free 3D Electrical Interconnects Directly Printed on Soft PDMS Substrates
Brenneman, J., Tansel, D. Z., Fedder, G. K., & Panat, R. P. (2022) · Advanced Materials Technologies · doi:10.1002/admt.202200396 · Cited by 10
Design and Analysis of Flexible Capacitive Power Transfer With Stable Output Capability
Li, F., Zhou, H. L., Hu, W., et al. (2022) · IEEE Transactions on Circuits and Systems I Regular Papers · doi:10.1109/tcsi.2022.3194647 · Cited by 10
Yan, C. Y., Li, X., Yang, Z., Wang, X., Ran, H., Zhang, R., Tian, H., Wang, C., Chen, X., & Shao, J. (2023) · Materials Horizons · doi:10.1039/d3mh00325f · Cited by 10
Bioinspired adhesive patch with octopus vulgaris micro-sucker and hexagonal tree-frog pad structures
Ahmad, N. N., Ghazali, N. N. N., Rani, A. T. A., Kee, C. C., & Wong, Y. H. (2023) · Materials Science in Semiconductor Processing · doi:10.1016/j.mssp.2023.107731 · Cited by 11
A New Multilevel Pulsed Modulation Technique for Low Power High Data Rate Optical Biotelemetry
Stanchieri, G. D. P., Battisti, G., Marcellis, A. D., Faccio, M., Palange, E., & Constandinou, T. G. (2021) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas49922.2021.9644996 · Cited by 11
Thermal analysis of the flexible electronics affixed on large curvature myocardium surface
Yin, Y., Li, Y., & Li, M. (2019) · International Journal of Heat and Mass Transfer · doi:10.1016/j.ijheatmasstransfer.2019.118983 · Cited by 10
Self-powered wearable biosensors
Gao, W. F. (2021) · Energy Harvesting and Storage: Materials, Devices, and Applications · doi:10.1117/12.2588899 · Cited by 9
Hybrid materials approaches for bioelectronics
Bai, H., Hu, Z., & Rogers, J. A. (2023) · MRS Bulletin · doi:10.1557/s43577-023-00620-6 · Cited by 13
Self-citationNext generation material interfaces for neural engineering
Wunderlich, H., & Kozielski, K. (2021) · Current Opinion in Biotechnology · doi:10.1016/j.copbio.2021.09.005 · Cited by 9
“Chung et al. designed skin-adhesive PDMS soft electrodes with copper conductive layers that where mechanically separated from the source of strain (i.e. the skin) with a nontoxic ionic liquid encased in the PDMS [13] (Figure 1b,c).”
BackgroundAdditively Fabricated Electronic Skin with High Performance in Dynamic Sensing as Human Skin
Zhan, L., Cao, Y., Gao, Y., Meng, S., Ji, H., Li, M., Wei, J., & Feng, H. (2023) · ACS Applied Electronic Materials · doi:10.1021/acsaelm.2c01654 · Cited by 9
Xu, J. B., Feng, H., Cao, J., Zhang, B., Bao, W., Wang, L., Yin, Y., Li, H., & Zhu, G. (2023) · Nano Energy · doi:10.1016/j.nanoen.2023.108522 · Cited by 8
Wireless interfaces for brain neurotechnologies
Kim, H., & Ho, J. S. (2022) · Philosophical Transactions of the Royal Society A Mathematical Physical and Engineering Sciences · doi:10.1098/rsta.2021.0020 · Cited by 18
“Structuring rigid materials using motifs such as out-of-plane buckling structures [193], serpentines [194,195], helical coils [196], microcracks [197,198], kirigami patterns [199] and open mesh [200] can confer stretchability.”
BackgroundPossible application of agricultural robotics in rabbit farming under smart animal husbandry
Jiang, W., Hao, H., Wang, H., & Wang, L. (2025) · Journal of Cleaner Production · doi:10.1016/j.jclepro.2025.145301 · Cited by 10
Wang, Z., Chang, X., Xu, Y., Gao, Y., Peng, Y., Wang, Y., Feng, Z. H., & Wang, H. (2025) · Materials Today Physics · doi:10.1016/j.mtphys.2025.101755 · Cited by 9
Self-citationMicrofluidic-enabled three-dimensional stretchable thermoelectrics
Huang, Z., Chen, T., Jiang, Y., Zhou, R., Wang, Y., Ji, J., Xie, H., Pan, T., Fan, D., Liang, L., et al. (2025) · npj Flexible Electronics · doi:10.1038/s41528-025-00429-0 · Cited by 10
Nano-bio interfaces for electrical and biochemical signal transduction
Yang, X., Tsai, C., Yang, Y., Zhang, W., You, H., Forró, C., Pașca, S. P., & Cui, B. (2025) · Nature Reviews Bioengineering · doi:10.1038/s44222-025-00374-7 · Cited by 20
Microstructure engineering for tactile-enabled embodied intelligence
Pang, H., Hua, Q., & Shen, G. (2026) · Materials Horizons · doi:10.1039/d5mh01906k · Cited by 10
AI-enabled flexible electronic systems via near-sensor and in-sensor computing
Xu, Z., Xie, E., Hou, C., Bai, Y., He, W., Liu, X., Zhang, F., & Huang, Y. (2026) · npj Flexible Electronics · doi:10.1038/s41528-026-00544-6 · Cited by 16
Self-citationli, S., Kong, C., Wu, E., Yang, C., Gao, Z., Zhou, X., Guo, H., Jiang, S., & Liang, J. (2026) · Soft Science · doi:10.20517/ss.2025.120 · Cited by 13
Liu, L., Zhang, H., Xu, R., et al. (2022) · IEEE Electron Device Letters · doi:10.1109/led.2022.3225302 · Cited by 8
“In addition, it can be expected as wireless electronic skins to realize biophysical tracking of fitness and wellness [14], as well as clinical management [15], [16].”
BackgroundWalter, J. R., Lee, J. Y., Yu, L., Kim, B., Martell, K., Opdycke, A., Scheffel, J., Felsl, I., Patel, S., Rangel, S. M., et al. (2024) · Scientific Reports · doi:10.1038/s41598-024-57830-4 · Cited by 13
Self-citationMonoolein-Based Wireless Capacitive Sensor for Probing Skin Hydration
Chaturvedi, V., Falk, M., Björklund, S., Gonzalez-Martinez, J. F., & Shleev, S. (2024) · Sensors · doi:10.3390/s24144449 · Cited by 9
Conformal Iontronic Sensing Clear Aligner
Sun, C., Li, S., Zhou, Z., Chen, X., Liu, Z., Liao, S., Xing, J., Sun, H., Cheng, Y., Yi, Z., et al. (2024) · Advanced Functional Materials · doi:10.1002/adfm.202408376 · Cited by 10
Self-citationActive-type piezoelectric smart textiles with antifouling performance for pathogenic control
Lee, S. E., Lee, H., Kim, J. H., Park, J. C., Kyung, S., Choi, H., Baek, S. H., Park, J. H., Park, S.-H., Kim, J., et al. (2024) · npj Flexible Electronics · doi:10.1038/s41528-024-00350-y · Cited by 11
Self-citationA Multilevel Synchronized Optical Pulsed Modulation for High Efficiency Biotelemetry
Stanchieri, G. D. P., Marcellis, A. D., Battisti, G., Faccio, M., Palange, E., & Constandinou, T. G. (2022) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2022.3209542 · Cited by 9
“[2], [3], [4], [5], [6], [7], [8], [9], [10], [11], [12], [13], [14], [15].”
BackgroundEfficient Single-Symbol Backscatter With Uncontrolled Ambient OFDM WiFi
Gong, W., Huang, Y., Wang, Q., et al. (2023) · IEEE/ACM Transactions on Networking · doi:10.1109/tnet.2023.3332220 · Cited by 8
“Placing miniature battery-free sensors on the surface of the body [33] or implanting them in the body [34] is a promising approach to health monitoring.”
BackgroundSelf-citationWearable Sensors: An ACS Sensors Special Issue
Zhang, Y. (2025) · ACS Sensors · doi:10.1021/acssensors.5c01530 · Cited by 7
Self-citationColburn, D. A. M., Chern, T. L., Guo, V. E., Salamat, K. A., Pugliese, D. N., Bradley, C., Shimbo, D., & Sia, S. K. (2025) · npj Biosensing · doi:10.1038/s44328-024-00021-y · Cited by 10
Nano/micro-engineered metallic flexible electrodes: advancing the era of epidermal electronics
Zhu, H., Liu, H., Cho, J. H., & Liu, G. (2025) · International Journal of Extreme Manufacturing · doi:10.1088/2631-7990/ae02ca · Cited by 14
Laser-based engineering strategies for biomedical and healthcare devices
Song, S., Kim, K. Y., Lee, H., Hong, S., Yoon, S. J., Hwang, S., In, J. B., Jeon, H., & Ko, S. H. (2025) · Materials Science and Engineering R Reports · doi:10.1016/j.mser.2025.101122 · Cited by 11
Implementing clinical needs into the development of wearable health-monitoring technology
Brasier, N., Domenghino, A., Meier, C. A., Jacob, C., Xu, S., & Rogers, J. A. (2025) · Nature Nanotechnology · doi:10.1038/s41565-025-02020-0 · Cited by 9
Self-citationLee, J., Lee, M., Kim, J., Kim, H.-J., Yu, J., Kim, N., & 이재홍 (2025) · npj Flexible Electronics · doi:10.1038/s41528-025-00486-5 · Cited by 16
Self-citationCharacterization of Bilayer Tissue Moduli and Thickness Via Eccentric Rotating Mass Dynamics
Xie, Z., Bai, D., Ma, J., Jing, C., & Lu, M. (2026) · Applied Mechanics Reviews · doi:10.1115/1.4071065 · Cited by 9
Self-citationLu, X., Zhong, H., Chen, S., Yang, C., Chen, Y., Pan, Y., & Li, G. (2026) · Advanced Materials · doi:10.1002/adma.73436 · Cited by 11
Marangalou, A. H., Gonzalez, M., Reppucci, N., & Güler, Ü. (2024) · Electronics · doi:10.3390/electronics13193947 · Cited by 8
Devnath, A., Bae, J., Alimkhanuly, B., Lee, G., Lee, S., Kadyrov, A., Patil, S. V., & Lee, S. (2024) · ACS Nano · doi:10.1021/acsnano.4c08650 · Cited by 10
Self-citationLansac, Y., Choi, C., & Jang, Y. H. (2024) · Bulletin of the Korean Chemical Society · doi:10.1002/bkcs.12908 · Cited by 7
Qin, C., Wang, Q., Fang, T., Wang, L., Yang, C., Lin, Y. Q., Bai, C., He, W., Ding, L., Zhang, J., et al. (2025) · Chemical Engineering Journal · doi:10.1016/j.cej.2025.159937 · Cited by 8
Review on Wearable Antennas: Part 2: Recent advances and applications.
Ye, Z., Zhao, G., Yang, M., & Chen, P. (2025) · IEEE Antennas and Propagation Magazine · doi:10.1109/map.2025.3530415 · Cited by 8
Tunable tensile response of honeycomb plates with nanoscale thickness: Testing and modeling
Jiao, P., Nicaise, S. M., Azadi, M., et al. (2019) · Extreme Mechanics Letters · doi:10.1016/j.eml.2019.100599 · Cited by 7
Stretchable, Multi-Layered Stack Antenna for Smart/Wearable Electronic Applications
Hong, K., Cho, J., & Shin, G. (2022) · Materials · doi:10.3390/ma15093275 · Cited by 9
Shape-adaptive and recyclable radio-frequency devices based on polymer with variable stiffness
Wang, H., Yu, B., Huang, Z., Wang, F., Ju, L., Xiao, Z.-D., Zhan, J., Chen, H., Liu, Z., Lu, W., et al. (2024) · Cell Reports Physical Science · doi:10.1016/j.xcrp.2024.101882 · Cited by 7
Self-citationWireless monitoring in the ICU on the horizon
Shah, P. S. (2020) · Nature Medicine · doi:10.1038/s41591-020-0798-3 · Cited by 7
Guan, M., Zhang, Q., Qiu, R., Hu, Z., Zhang, X., Peng, B., & Liu, M. (2023) · Advanced Electronic Materials · doi:10.1002/aelm.202300323 · Cited by 6
Chen, M., Tang, P., Li, R., Li, X., Yi, X., Luan, H., Wang, X., Zhou, C., Zhou, K., & Hu, J. (2024) · Chemical Engineering Journal · doi:10.1016/j.cej.2024.153861 · Cited by 6
Xiao, N., He, L., & Wang, K. (2025) · Coatings · doi:10.3390/coatings15020149 · Cited by 7
Lin, B., Cheng, B., & Zhang, Y. (2025) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2025.137749 · Cited by 6
Self-citationLee, S. E., Lee, H., Kim, J. H., Park, S., Oh, Y. T., Choi, H., Kim, J., Jo, H., Park, J. H., Jin, H. J., et al. (2025) · Advanced Healthcare Materials · doi:10.1002/adhm.202500597 · Cited by 8
Self-citationLiu, G., Xu, T., Zhou, L., Wang, Z., Zhang, Y., Guo, S., Li, X., Chen, P., Li, J., Yu, L., et al. (2025) · ACS Nano · doi:10.1021/acsnano.5c10181 · Cited by 7
Self-citationWang, Y., Zou, Y., Xue, J., Fan, Z., Tan, P., Liu, M., Cheng, P., Ouyang, H., Fan, Y., Liu, A., et al. (2025) · Advanced Functional Materials · doi:10.1002/adfm.202514598 · Cited by 6
Self-citationHuang, X., He, Y., Shi, Y., Hao, Y., Ge, Y., Zhang, H., Li, P., Xu, F., & Yao, Y. (2025) · npj Flexible Electronics · doi:10.1038/s41528-025-00476-7 · Cited by 11
A wireless, battery-free device for electrical neuromodulation of bladder contractions
Li, J., Deng, G., Li, X., Yin, L., Yuan, C., Shao, W., Xia, X., Yan, J., & Yao, J. (2024) · Materials Today Bio · doi:10.1016/j.mtbio.2024.101233 · Cited by 10
Sénéchal, E., Radeschi, D. J., Lv, S., Jeanne, E., Ruiz, A. S., Tao, L., Dulmage, B., Shalish, W., Kearney, R. E., & SantʼAnna, G. (2025) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2025.1555882 · Cited by 6
“However many of these studies have used short monitoring durations in a small sample size under well controlled experimental conditions (Chung et al., 2019; Tanigasalam et al., 2019; Chen et al., 2020; 2024; Grooby et al., 2021; Henry et al., 2021; Kwak et al., 2021; Pike et al., 2021; Sriraam et…”
Postbuckling Analysis of Ultra-Low Rigidity Serpentine Structures
Zhao, J. Z. (2022) · Journal of Applied Mechanics · doi:10.1115/1.4053397 · Cited by 6
Wearable electronic devices in the intensive care units
Chen, J., Jiang, W., & Song, Y. (2024) · Clinical eHealth · doi:10.1016/j.ceh.2024.02.001 · Cited by 7
Electromechanical Decoupling by Porous Aerogel Conducting Polymer
You, I., & Jeong, U. (2019) · Matter · doi:10.1016/j.matt.2019.06.008 · Cited by 6
Metadata
- Authors
- Chung, H. U., Kim, B. H., Lee, J. Y., Lee, J., Xie, Z., Ibler, E. M., Lee, K., Banks, A., Jeong, J. Y., Kim, J., Ogle, C., Grande, D., Yu, Y., Jang, H., Assem, P., Ryu, D., Kwak, J. W., Namkoong, M., Park, J. B., Lee, Y., Kim, D. H., Ryu, A., Jeong, J., You, K., Ji, B., Liu, Z., Huo, Q., Feng, X., Deng, Y., Xu, Y., Jang, K.-I., Kim, J., Zhang, Y., Ghaffari, R., Rand, C. M., Schau, M., Hamvas, A., Weese-Mayer, D. E., Huang, Y., Lee, S. M., Lee, C. H., Shanbhag, N. R., Paller, A. S., Xu, S., & Rogers, J. A.
- Affiliations
- Northwestern University (US)
- University of Illinois Urbana-Champaign (US)
- Department of Materials Science and Engineering, Northwestern University, Evanston, IL 60208, USA.
- Agency for Science, Technology and Research (SG)
- Tsinghua University (CN)
- Daegu Gyeongbuk Institute of Science and Technology (KR)
- Lurie Children's Hospital (US)
- Center for Autonomic Medicine, Department of Pediatrics, Ann & Robert H. Lurie Children’s Hospital of Chicago, Chicago, IL 60611, USA.
- Kookmin University (KR)
- Purdue University West Lafayette (US)
- Corresponding author
- Chung, Ha Uk
- Kim, Bong Hoon
- Lee, Jong Yoon
- Lee, Jungyup
- Xie, Zhaoqian
- Paller, Amy S.
- Xu, Shuai
- Rogers, John A.
- ORCID
- Chung, Ha Uk
- Kim, Bong Hoon
- Lee, Jong Yoon
- Lee, Jungyup
- Xie, Zhaoqian
- Lee, KunHyuck
- Banks, Anthony
- Jeong, Ji Yoon
- Kim, Jongwon
- Grande, Dominic
- Yu, Yongjoon
- Jang, Hokyung
- Assem, Pourya
- Kwak, Jean Won
- Namkoong, Myeong
- Park, Jun Bin
- Lee, Yechan
- Kim, Do Hoon
- Ryu, Arin
- You, Kevin
- Ji, Bowen
- Liu, Zhuangjian
- Huo, Qingze
- Feng, Xue
- Deng, Yujun
- Xu, Yeshou
- Jang, Kyung-In
- Kim, Jeonghyun
- Zhang, Yihui
- Ghaffari, Roozbeh
- Rand, Casey M.
- Hamvas, Aaron
- Huang, Yonggang
- Lee, Seung Min
- Lee, Chi Hwan
- Shanbhag, Naresh R.
- Paller, Amy S.
- Xu, Shuai
- Rogers, John A.
- Venue
- Science
- Published
- 2019-03
- Volume
- 363
- Issue
- 6430
- ISSN
- 0036-8075 (ISSN-L)
- 1095-9203
- PMC
- PMC6510306
- OpenAlex
- W2916533817
- Semantic Scholar
- b0511afac0cc252946aa8f08a078b12e48030581
- Cited by
- 807
- References
- 43
- Citations by year
- 2026: 72 · 2025: 112 · 2024: 103 · 2023: 121 · 2022: 124 · 2021: 131 · 2020: 97 · 2019: 47
- FWCI
- 35.10
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by
- Locations
- Science (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Funders
- Bill and Melinda Gates Foundation (OPP1182909)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Physical Sciences › Engineering › Biomedical Engineering › Wireless Body Area Networks
- Keywords
- epidermal electronics, neonatal intensive care, battery-free operation, neonatal monitoring
- MeSH terms
- Diagnostic Imaging, Electronics, Equipment Design, Humans, Infant, Newborn, Monitoring, Physiologic, Skin, Intensive Care, Neonatal, Vital Signs, Lab-On-A-Chip Devices, Wireless Technology
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.