Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care
Open accessThe DOI matches its Crossref record; 44 of the 44 references it cites were found in a registry or an index; 12 of the 42 sources its claims cite were checked for support.
Abstract
Sensitive sensing Neonatal care, particularly for premature babies, is complicated by the infants' fragility and by the need for a large number of tethered sensors to be attached to their tiny bodies. Chung et al. developed a pair of sensors that only require water to adhere to the skin and allow for untethered monitoring of key vital signs (see the Perspective by Guinsburg). On-board data processing allowed for efficient wireless near-field communication using standard protocols. The absence of cables makes it easier to handle the infants and allows for skin-to-skin contact between the babies and their parents or caregivers. Science , this issue p. eaau0780 ; see also p. 924
Overview
CoChat Summary
TL;DR
The paper introduces a pair of ultrathin, wireless, battery-free epidermal electronic systems (EES) for neonatal intensive care: one chest-mounted ECG device and one foot-mounted PPG device. Pilot studies in neonates show clinical-grade agreement for heart rate, respiratory rate, and SpO₂, plus added capabilities for skin temperature and pulse arrival time as a systolic blood pressure surrogate.
Key findings
- The binodal EES continuously streams ECG, PPG, and skin temperature data, yielding HR, HRV, RR, SpO₂, and PAT; wireless data rates reach up to 800 bytes/s with dual channels, NFC operation works up to 25 cm through NICU materials, and BLE range reaches up to 20 m.
- The devices have effective moduli of 200 to 300 kPa, adhere via van der Waals forces, and show peel forces about a factor of ~10 lower than conventional NICU adhesives, with maximum von Mises stress on skin reduced by a factor of 4.3; the microfluidic channel reduces interfacial stresses by up to a factor of 2.5, and adhesion energy is G = 16 N/m.
- For MRI compatibility, mesh electrodes reduce the in-plane magnetic field gradient by a factor of ~3 and the out-of-plane gradient by a factor of 4 versus commercial electrodes; simulated copper heating is 1°C with maximum skin temperature change of 0.04°C, and experimental 3-T MRI measurements show ~0.1°C temperature difference, meeting FDA “MRI-safe” requirements.
- In neonates (n = 3, gestational age 28 to 40 weeks), mean differences versus standard-of-care monitors were –0.17 beats per minute for HR, 0.76 breaths per minute for RR, and 1.02% for SpO₂; additional studies in n = 18 neonates with gestational ages as low as 28 weeks and weights as low as 1470 g showed similar performance, and scaled production cost is estimated at less than $20 USD per unit.
Methods
- Fabrication used semiconductor processing, lamination, transfer printing, chip placement, and solder bonding; each EES includes serpentine Cu traces, chip-scale ICs, fractal mesh electrodes, an ionic-liquid microfluidic chamber, silicone encapsulation, and an NFC loop antenna for simultaneous wireless power and data.
- In-sensor analytics include a streamlined Pan-Tompkins algorithm for real-time QRS detection to yield HR and HRV, dynamic baseline control and SpO₂ calculation for PPG, and time-synchronized ECG/PPG processing for PAT.
- Validation included healthy adult volunteers (n = 3) against FDA-cleared monitoring equipment, and neonates in two tertiary-level NICUs: n = 3 with synchronous standard-of-care measurements and n = 18 using related device platforms with onboard power supplies; the protocol was IRB-approved and registered on ClinicalTrials.gov (NCT02865070).
Limitations
- The paper states that results are preliminary feasibility testing and validation; comprehensive clinical studies are ongoing to verify measurements across increased age groups and ethnic backgrounds, and remaining challenges include nursing acceptance, compatibility with legacy monitoring systems, and device sterilization for reuse.
- The paper notes that additional testing is needed for altered skin barrier function, such as burn victims or patients with epidermolysis bullosa.
- Evident from the text: the neonatal validation cohorts are small (n = 3 with synchronous standard-of-care comparisons; n = 18 using related platforms with onboard power supplies), adult validation is n = 3, and temperature and PAT data in neonates are shown without reference comparators because conventional devices acquire these only periodically.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures









Verification
Claims the paper makes
6 Valid · 2 Not valid · 18 Not checked · 82 Own results and interpretations
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“A key example is the measurement of pulse arrival time (PAT), defined by the time lapse between the maximum fiducial point in the ECG signal (R peak) and the corresponding minimal fiducial point in the PPG signal at valley as in Fig. 6H, as a direct correlate to systolic blood pressure (29 , 30).”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Real-time measurements, in-sensor analytics, and data transmission
What the sources say
- Partly supportedcites [29] Kim J. S., Chee Y. J., Park J. W., Choi J. W., Park K. S., A new approach for non-intrusive monitoring of blood pressure on a toilet seat. Physiol. Meas. 27, 203– 211 (2006). doi: 10.1088/0967-3334/27/2/010; pmid: 16400206 sourceExperimental — accuracy not yet measured
Section 4.1 reports that PAT was significantly, but moderately, correlated with systolic blood pressure. Section 2.2 defines the study’s PAT using the ECG R-peak and the PPG’s maximum first-derivative point, not the PPG valley; it says the minimum point was unclear in its signal. The claimed correlation is supported, but the attributed measurement definition is not. (LLM, conf=0.87)
- Not checked — no full text of the sourcecites [30] Geddes L. A., Voelz M. H., Babbs C. F., Bourland J. D., Tacker W. A., Pulse transit time as an indicator of arterial blood pressure. Psychophysiology 18, 71– 74 (1981). doi: 10.1111/j.1469-8986.1981.tb01545.x; pmid: 7465731 source
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The inherently thin, soft mechanical properties of the sensors (Fig. 1A) reported here allow for adhesion via van derWaals forces alone.”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Low-modulus mechanics, soft interface adhesion, and implications for neonatal skin safety
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The mechanical decoupling afforded by the microfluidic channel decreases these stresses by up to a factor of 2.5 (Fig. 2B) relative to otherwise similar designs without the microfluidics.”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Low-modulus mechanics, soft interface adhesion, and implications for neonatal skin safety
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Oscillations in the force only appear for holes larger than the characteristic size of the cohesive zone (~500 mm, as in Fig. 2I).”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Low-modulus mechanics, soft interface adhesion, and implications for neonatal skin safety
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Specifically, the interface starts to delaminate when the normal stress reaches ~20 kPa (Fig. 2G, inset).”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Low-modulus mechanics, soft interface adhesion, and implications for neonatal skin safety
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
61 more claims
Create a free account to see all 108 claims and to verify the 1 reference still open.
- Probably not a substantial claim. 14
- It may state two separate claims as one. 33
- The quote only partly states this claim. 2
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Researchers have developed various versatile epidermal electrodes with high performance that can universally measure EEG, EMG, EOG, and ECG through placing electrodes on different areas of the skin.[25,414]”
From Flexible Electronics and Devices as Human–Machine Interfaces for Medical Robotics (2022) · cites this paper as [414]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Previous reports validated the performance of an FDA-cleared physiological monitoring system against gold standard monitoring platforms [18–23].”
From Pregnancy outcomes in infertility patients diagnosed with sleep disordered breathing with wireless wearable sensors (2022) · cites this paper as [18]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“In previous work, we demonstrate the ability for the chest- and finger-mounted sensors to monitor vital signs in both children and adults; the abdominal sensor allows for maternal–fetal monitoring of pregnant women in both high- and low-resource settings as well as the clinical and home settings (22, 23).”
From Comprehensive pregnancy monitoring with a network of wireless, soft, and flexible sensors in high- and low-resource health settings (2021) · cites this paper as [23]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“and have been used in neonatal and pediatric intensive care units. 4,5”
From Advanced wireless monitoring for children in the cardiac perioperative setting (2023) · cites this paper as [4]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“However, this system is limited by (1) the modest maximum operating distances (~30 cm) supported by near-field communication (NFC) protocols used for power transfer and data communication, (2) the mechanically fragile nature of the ultrathin, compliant mechanics designs, (3) the sufficient, but constrained range of measurement capabilities, and (4) the demand for highly advanced device configur”
From Skin-interfaced biosensors for advanced wireless physiological monitoring in neonatal and pediatric intensive-care units (2020) · cites this paper as [9]
Experimental — accuracy not yet measured
5 more claims about this paper
Create a free account to see every claim other papers make about this one.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Compatibility with medical imaging techniques used in the NICU
“eliminating any parasitic heating from magnetically induced eddy currents.”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Compatibility with medical imaging techniques used in the NICU
“Full three-dimensional multi-physicsmodeling shows that, at the end of a single scan for 0.5 ms, the copper layer of an ECG EES undergoes heating by only 1°C (Fig. 3H).”
Citations
Cites
A Real-Time QRS Detection Algorithm
Pan, J., & Tompkins, W. J. (1985) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1985.325532 · Cited by 7,855
Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis
Gao, W., Emaminejad, S., Nyein, H. Y. Y., Challa, S., Chen, K., Peck, A., Fahad, H. M., Ota, H., Shiraki, H., Kiriya, D., et al. (2016) · Nature · doi:10.1038/nature16521 · Cited by 5,087
Kim, D.-H., Lu, N., Ma, R., Kim, Y.-S., Kim, R.-H., Wang, S., Wu, J., Won, S. M., Tao, H., Islam, A., et al. (2011) · Science · doi:10.1126/science.1206157 · Cited by 4,836
Self-citationWearable Sensors for Remote Health Monitoring
Majumder, S., Mondal, T., & Deen, M. (2017) · Sensors · doi:10.3390/s17010130 · Cited by 1,375
Kiserud, T., Piaggio, G., Carroli, G., Widmer, M., Carvalho, J., Neerup Jensen, L., Giordano, D., Cecatti, J. G., Abdel Aleem, H., Talegawkar, S. A., et al. (2017) · PLOS Medicine · doi:10.1371/journal.pmed.1002220 · Cited by 729
Thermochemistry of ionic liquid heat-transfer fluids
Valkenburg, M. E. V., Vaughn, R. L., Williams, M., & Wilkes, J. S. (2005) · Thermochimica Acta · doi:10.1016/j.tca.2004.11.013 · Cited by 595
Halcox, J. P. J., Wareham, K., Cardew, A., Gilmore, M., Barry, J. P., Phillips, C., & Gravenor, M. B. (2017) · Circulation · doi:10.1161/circulationaha.117.030583 · Cited by 580
Self-assembled three dimensional network designs for soft electronics
Jang, K.-I., Li, K., Chung, H. U., Xu, S., Jung, H. N., Yang, Y., Kwak, J. W., Jung, H. H., Song, J., Yang, C., et al. (2017) · Nature Communications · doi:10.1038/ncomms15894 · Cited by 445
Self-citationChen, W., Kobayashi, T., Ichikawa, S., Takeuchi, Y., & Togawa, T. (2000) · Medical & Biological Engineering & Computing · doi:10.1007/bf02345755 · Cited by 425
Harris, K. D., Elias, A. L., & Chung, H.-J. (2016) · Journal of Materials Science · doi:10.1007/s10853-015-9643-3 · Cited by 403
Self-citation
Research works printed but not linked: 2
- O. Sharma, S. N. Lewis, U. Telang, L. D’Almeida, L. E. S. Lewis, Design of a Bluetooth Enabled Health Monitoring System for Infants Using Wearable Technology. J. Adv. Res. Dyn. Contrl. Syst. 15, 887–894 (2017). http://jardcs.org/papers/v9/sp/20181075.pdf
- C. H. Lund J. A. Tucker in Neonatal Skin: Structure and Function S. B. Hoath H. I. Maibach Eds. (Dekker ed. 2 2003) pp. 299–324.
Cited by
Items 201–300 of 800
Stretchable and biodegradable self-healing conductors for multifunctional electronics
Jang, T., Han, W. B., Han, S., Dutta, A., Lim, J. H., Kim, T.-K., Lim, B. H., Ko, G., Shin, J., Rajaram, K., et al. (2024) · Science Advances · doi:10.1126/sciadv.adp9818 · Cited by 89
Self-citationConformable bioelectronic interfaces: Mapping the road ahead
Schiavone, G., & Lacour, S. P. (2019) · Science Translational Medicine · doi:10.1126/scitranslmed.aaw5858 · Cited by 77
Ultrasensitive and ultrastretchable electrically self-healing conductors
Li, Y., Fang, T., Zhang, J., Zhu, H., Sun, Y., Wang, S., Lu, Y., & Kong, D. (2023) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.2300953120 · Cited by 86
Li, Y., Wang, S., Zhang, J., Ma, X., Cao, S., Sun, Y., Feng, S., Fang, T., & Kong, D. (2022) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.1c25206 · Cited by 72
Ultrastretchable MXene Microsupercapacitors
Wang, M., Feng, S., Bai, C., Ji, K., Zhang, J., Wang, S., Lu, Y., & Kong, D. (2023) · Small · doi:10.1002/smll.202300386 · Cited by 73
Hong, Y., Jin, L., Wang, B., Liao, J., He, B., Yang, T., Long, Z., Li, P., Zhang, Z., Liu, S., et al. (2021) · Energy & Environmental Science · doi:10.1039/d1ee02353e · Cited by 71
Self-citationKwak, S. S., Yoo, S., Avila, R. O., Chung, H. U., Jeong, H., Liu, C. Y., Vogl, J. L., Kim, J., Yoon, H., Park, Y., et al. (2021) · Advanced Materials · doi:10.1002/adma.202103974 · Cited by 74
Self-citationTunable Hydrogel Electronics for Diagnosis of Peripheral Neuropathy
Yang, G., Hu, Y., Guo, W., Wei, L., Liu, W., Guo, G., Geng, C., Liu, Y., & Wu, H. (2023) · Advanced Materials · doi:10.1002/adma.202308831 · Cited by 79
Highly Permeable and Ultrastretchable Liquid Metal Micromesh for Skin-Attachable Electronics
Ma, X., Zhang, M., Zhang, J., Wang, S., Cao, S., Li, Y., Hu, G., & Kong, D. (2022) · ACS Materials Letters · doi:10.1021/acsmaterialslett.1c00763 · Cited by 70
All‐day wearable health monitoring system
Han, S. A., Naqi, M., Kim, S., & Kim, J. H. (2022) · EcoMat · doi:10.1002/eom2.12198 · Cited by 77
Self-citationQiu, J., Yu, T., Zhang, W., Zhao, Z., Zhang, Y., Guo, Y., Zhao, Y., Du, X., Liu, X., Yang, L., et al. (2020) · ACS Materials Letters · doi:10.1021/acsmaterialslett.0c00203 · Cited by 69
Self-citationHan, W. B., Yang, S. M., Rajaram, K., & Hwang, S. (2021) · Advanced Sustainable Systems · doi:10.1002/adsu.202100075 · Cited by 76
Yang, S. M., Shim, J. H., Cho, H., Jang, T., Ko, G., Shim, J., Kim, T. H., Zhu, J., Park, S. U., Kim, Y. S., et al. (2022) · Advanced Materials · doi:10.1002/adma.202108203 · Cited by 75
Self-citationSkin-interfaced colorimetric microfluidic devices for on-demand sweat analysis
Liu, W., Cheng, H., & Wang, X. (2023) · npj Flexible Electronics · doi:10.1038/s41528-023-00275-y · Cited by 80
Wang, Y., Wang, Z., Zhang, Y., Yang, J., Zhang, Z., Zhou, P., Xu, Y., Sun, Q., Zheng, M., Yan, W., et al. (2025) · Science Advances · doi:10.1126/sciadv.adt2286 · Cited by 88
Self-citationJi, B., Xie, Z., Hong, W., Jiang, C., Guo, Z., Wang, L., Wang, X., Yang, B. T., & Liu, J. (2019) · Journal of Materiomics · doi:10.1016/j.jmat.2019.11.006 · Cited by 71
Self-citationAl‐Qatatsheh, A., Morsi, Y., Zavabeti, A., Zolfagharian, A., Salim, N. V., Kouzani, A. Z., Mosadegh, B., & Gharaie, S. (2020) · Sensors · doi:10.3390/s20164484 · Cited by 71
Flexible, wearable biosensors for digital health
Zhu, P., Peng, H., & Rwei, A. Y. (2022) · Medicine in Novel Technology and Devices · doi:10.1016/j.medntd.2022.100118 · Cited by 81
Lin, Y. Q., Fang, T., Bai, C., Sun, Y., Yang, C., Hu, G., Guo, H., Qiu, W., Huang, W., Wang, L., et al. (2023) · Nano Letters · doi:10.1021/acs.nanolett.3c03670 · Cited by 69
Materials, Devices, and Applications for Wearable and Implantable Electronics
Han, W. B., Ko, G., Jang, T., & Hwang, S. (2021) · ACS Applied Electronic Materials · doi:10.1021/acsaelm.0c00724 · Cited by 67
Yang, A., Song, J., Liu, H., Zhao, Z., Li, L., & Yan, F. (2023) · Advanced Functional Materials · doi:10.1002/adfm.202215037 · Cited by 74
A Controlled Biodegradable Triboelectric Nanogenerator Based on PEGDA/Laponite Hydrogels
Li, Z. Y., Li, C., Sun, W., Bai, Y., Li, Z., & Deng, Y. (2023) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.2c22359 · Cited by 70
Self-citationFolding and Bending Planar Coils for Highly Precise Soft Angle Sensing
Wang, H., Totaro, M., Veerapandian, S., Ilyas, M., Kong, M., Jeong, U., & Beccai, L. (2020) · Advanced Materials Technologies · doi:10.1002/admt.202000659 · Cited by 64
Near‐Sensor Reservoir Computing for Gait Recognition via a Multi‐Gate Electrolyte‐Gated Transistor
Liu, X., Sun, C., Guo, Z., Xia, X., Jiang, Q., Ye, X., Shang, J., Zhang, Y., Zhu, X., & Li, R. (2023) · Advanced Science · doi:10.1002/advs.202300471 · Cited by 65
Self-citationE-Skin: The Dawn of a New Era of On-Body Monitoring Systems
Bunea, A.-C., Dediu, V., Laszlo, E. A., Pistriţu, F., Carp, M., Iliescu, F. S., Ionescu, O. N., & Iliescu, C. (2021) · Micromachines · doi:10.3390/mi12091091 · Cited by 66
“The performance of this class of integrated electronic e-skin sensors, mounted onto epidermis based only on van der Waals interactions, improved through increased contact area and fewer motion artefacts [27,28].”
BackgroundUltrasensitive and Ultraprecise Pressure Sensors for Soft Systems
Shi, L., Li, Z., Chen, M., Zhu, T., & Wu, L. (2023) · Advanced Materials · doi:10.1002/adma.202210091 · Cited by 63
Cheng, A., Li, X., Li, D., Chen, Z., Cui, T., Tao, L., Jian, J., Xiao, H., Shao, W., Tang, Z.-Y., et al. (2025) · Nature Communications · doi:10.1038/s41467-024-55649-1 · Cited by 89
3D active-matrix multimodal sensor arrays for independent detection of pressure and temperature
Jo, Y., Lee, Y., Kwon, J., Kim, S., Ryu, G., Yun, S., Baek, S., Ko, H., & Jung, S. (2025) · Science Advances · doi:10.1126/sciadv.ads4516 · Cited by 78
Self-citationFlexible bioelectronics for physiological signals sensing and disease treatment
Yao, G., Yin, C., Wang, Q., Zhang, T., Chen, S., Lu, C., Zhao, K., Xu, W., Pan, T., Gao, M., et al. (2019) · Journal of Materiomics · doi:10.1016/j.jmat.2019.12.005 · Cited by 71
Emerging Frontier of Peripheral Nerve and Organ Interfaces
Shahriari, D., Rosenfeld, D., & Anikeeva, P. (2020) · Neuron · doi:10.1016/j.neuron.2020.09.025 · Cited by 67
Electronically integrated microcatheters based on self-assembling polymer films
Rivkin, B. B., Becker, C., Singh, B., Aziz, A., Akbar, F., Egunov, A. I., Karnaushenko, D. D., Naumann, R., Schaefer, R., Medina‐Sánchez, M., et al. (2021) · Science Advances · doi:10.1126/sciadv.abl5408 · Cited by 61
Skin-interfaced wireless biosensors for perinatal and paediatric health
Kim, J., Yoo, S., Liu, C. Y., Kwak, S. S., Walter, J. R., Xu, S., & Rogers, J. A. (2023) · Nature Reviews Bioengineering · doi:10.1038/s44222-023-00090-0 · Cited by 74
Self-citationLi, Y., Chen, K., Pang, Y., Zhang, J., Wu, M., Xu, Y., Cao, S., Zhang, X., Wang, S., Sun, Y., et al. (2023) · ACS Nano · doi:10.1021/acsnano.3c04758 · Cited by 64
Self-citationNan, K., Wong, K., Li, D., Ying, B., McRae, J. C., Feig, V. R., Wang, S., Du, N., Liang, Y., Mao, Q., et al. (2024) · Nature Communications · doi:10.1038/s41467-024-51102-5 · Cited by 83
Flexible pressure sensors with microstructures
Tang, R., Lu, F., Liu, L., Yan, Y., Du, Q., Zhang, B., Zhou, T., & Fu, H. (2021) · Nano Select · doi:10.1002/nano.202100003 · Cited by 63
Ionic communication for implantable bioelectronics
Zhao, Z., Spyropoulos, G. D., Cea, C., Gelinas, J. N., & Khodagholy, D. (2022) · Science Advances · doi:10.1126/sciadv.abm7851 · Cited by 61
Stretchable Electronics with Strain‐Resistive Performance
Hou, S., Chen, C., Bai, L., Yu, J., Cheng, Y., & Huang, W. (2023) · Small · doi:10.1002/smll.202306749 · Cited by 65
Biocompatible Light Guide‐Assisted Wearable Devices for Enhanced UV Light Delivery in Deep Skin
Zhang, H., Zhao, H., Zhao, X., Xu, C., Franklin, D., Vázquez‐Guardado, A., Bai, W., Zhao, J. W., Li, K., Monti, G., et al. (2021) · Advanced Functional Materials · doi:10.1002/adfm.202100576 · Cited by 61
Self-citationLiu, C., Kim, J., Kwak, S. S., Hourlier‐Fargette, A., Avila, R., Vogl, J., Tzavelis, A., Chung, H. U., Lee, J. Y., Kim, D. H., et al. (2021) · Advanced Healthcare Materials · doi:10.1002/adhm.202100383 · Cited by 57
Self-citationNear-Field Communication in Biomedical Applications
Kang, S.-G., Song, M.-S., Kim, J., Lee, J. W., & Kim, J. (2021) · Sensors · doi:10.3390/s21030703 · Cited by 65
“Biosignals are among the most important indicators needed for disease prevention, especially in infants and seniors [36,37].”
BackgroundInfluentialSelf-citationSweat-Resistant Silk Fibroin-Based Double Network Hydrogel Adhesives
Wang, J., Zhang, N., Tan, Y., Fu, F., Liu, G., Fang, Y., Zhang, X., Liu, M., Cheng, Y., & Yu, J. (2022) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.2c02534 · Cited by 59
Intelligent fibers and textiles for wearable biosensors
Fu, W., Yu, B., Ji, D., Zhou, Z., Li, X., Wang, R., Lu, W., Sun, Y., & Dai, Y. (2024) · Responsive materials · doi:10.1002/rpm.20240018 · Cited by 78
Jang, J., Ji, S., Grandhi, G. K., Cho, H. B., Im, W. B., & Park, J. (2021) · Advanced Materials · doi:10.1002/adma.202008539 · Cited by 56
Self-citationXu, S., Rwei, A. Y., Vwalika, B., Chisembele, M. P., Stringer, J. S. A., Ginsburg, A. S., & Rogers, J. A. (2021) · The Lancet Digital Health · doi:10.1016/s2589-7500(21)00001-7 · Cited by 57
Self-citationRecent progress of self-powered respiration monitoring systems
Dai, J., Li, L., Shi, B., & Li, Z. (2021) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2021.113609 · Cited by 56
An artificial triboelectricity-brain-behavior closed loop for intelligent olfactory substitution
Zhong, T., Zhang, M., Fu, Y., Han, Y., Guan, H., He, H., Zhao, T., Xing, L., Xue, X., Zhang, Y., et al. (2019) · Nano Energy · doi:10.1016/j.nanoen.2019.103884 · Cited by 58
Self-citationMa, H., Li, X., Lou, J., Gu, Y., Zhang, Y., Jiang, Y., Cheng, H., & Han, W. (2022) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.1c20163 · Cited by 56
Self-citationGhanim, R. W., Kaushik, A., Park, J., & Abramson, A. (2023) · Device · doi:10.1016/j.device.2023.100092 · Cited by 60
Self-citationLi, W., Pu, J., Zhao, X., Jia, J. F., Ke, K., Bao, R., Liu, Z., Yang, M., & Yang, W. M. (2020) · Journal of Materials Chemistry C · doi:10.1039/d0tc03961f · Cited by 55
Self-citationRFID-Based Dual-Chip Epidermal Sensing Platform for Human Skin Monitoring
Occhiuzzi, C., Parrella, S., Camera, F., Nappi, S., & Marrocco, G. (2020) · IEEE Sensors Journal · doi:10.1109/jsen.2020.3031664 · Cited by 54
“ONGOING progress in low power wearable [1] and epidermal electronics [2], [3] are aimed at probing personal health throughout the human skin analysis [4]–[7].”
BackgroundChallenges in Materials and Devices of Electronic Skin
Zhang, J., Li, J., Cheng, W. M., Zhang, J., Zhou, Z., Sun, X., Li, L., Liang, J., Shi, Y., & Pan, L. (2022) · ACS Materials Letters · doi:10.1021/acsmaterialslett.1c00799 · Cited by 57
Responsive materials and mechanisms as thermal safety systems for skin-interfaced electronic devices
Yoo, S., Yang, T., Park, M., Jeong, H., Lee, Y. J., Cho, D., Kim, J., Kwak, S. S., Shin, J., Park, Y., et al. (2023) · Nature Communications · doi:10.1038/s41467-023-36690-y · Cited by 63
Self-citationKim, Y., Lim, J., Lim, J. H., Hwang, E., Lee, H., Kim, M., Ha, I., Cho, H., Kwon, J., Oh, J., et al. (2023) · Advanced Materials · doi:10.1002/adma.202212294 · Cited by 58
Annapooranan, R., Jeyakumar, S. S., Chambers, R. J., Long, R., & Cai, S. (2023) · Advanced Functional Materials · doi:10.1002/adfm.202309123 · Cited by 55
Zhuang, Q., Yao, K., Song, X., Zhang, Q., Zhang, C., Wang, H., Yang, R., Zhao, G., Li, S., Shu, H., et al. (2025) · Science Advances · doi:10.1126/sciadv.adu3146 · Cited by 66
Heat-shedding with photonic structures: radiative cooling and its potential
Heo, S., Lee, G. J., & Song, Y. M. (2022) · Journal of Materials Chemistry C · doi:10.1039/d2tc00318j · Cited by 58
Three-Dimensionally Architected Tactile Electronic Skins
Hu, X., Liu, Z. Y., & Zhang, Y. (2025) · ACS Nano · doi:10.1021/acsnano.5c02232 · Cited by 57
Self-citationSkin-like drift-free biosensors with stretchable diode-connected organic field-effect transistors
Zhao, C., Park, J., Maulà, D., Yuan, Y., Zhong, D., Wang, W., Liu, Q., Xu, C., Zheng, Y., Mow, R. K., et al. (2025) · Nature Electronics · doi:10.1038/s41928-025-01465-4 · Cited by 80
Self-citationHow is flexible electronics advancing neuroscience research?
Chen, Y., Rommelfanger, N. J., Mahdi, A. I., Wu, X., Keene, S. T., Obaid, A., Salleo, A., Wang, H., & Hong, G. (2020) · Biomaterials · doi:10.1016/j.biomaterials.2020.120559 · Cited by 61
Yao, H., Sun, T., Chiam, J. S., Tan, M., Ho, K. Y., Liu, Z., & Tee, B. C. K. (2021) · Advanced Functional Materials · doi:10.1002/adfm.202008650 · Cited by 56
Self-citationFeng, B., Zou, G., Wang, W., Dong, M., Xiao, Y., Ren, H., Zhao, X., Zhao, G., Wu, A., Zhu, H., et al. (2020) · Nano Energy · doi:10.1016/j.nanoen.2020.104847 · Cited by 52
Li, J., Chu, H., Chen, Z., Yiu, C. K., Qu, Q., Li, Z.-Y., & Yu, X. (2024) · ACS Nano · doi:10.1021/acsnano.4c04291 · Cited by 50
Wearable Near-Field Communication Sensors for Healthcare: Materials, Fabrication and Application
Sun, X., Zhao, C., Li, H., Yu, H., Zhang, J., Qiu, H., Liang, J., Wu, J., Su, M., Shi, Y., et al. (2022) · Micromachines · doi:10.3390/mi13050784 · Cited by 62
“…are great prospects for mercial applications in smartphone sensing, food safety and health monitoring [23] 3 of 22 NFC sensors are a promising tool for real ‐ time health monitoring, featuring non ‐ contact data transmission under passive tags, wireless communication and a rapid response [20–22].”
BackgroundMethodBroadband Photodetectors and Imagers in Stretchable Electronics Packaging
Araki, T., Li, K., Suzuki, D., Abe, T., Kawabata, R., Uemura, T., Izumi, S., Tsuruta, S., Terasaki, N., Kawano, Y., et al. (2023) · Advanced Materials · doi:10.1002/adma.202304048 · Cited by 61
Skin-interfacing wearable biosensors for smart health monitoring of infants and neonates
Zhou, L., Guess, M., Kim, K. R., & Yeo, W. (2024) · Communications Materials · doi:10.1038/s43246-024-00511-6 · Cited by 64
On-Skin Epidermal Electronics for Next-Generation Health Management
Xu, J., Chen, X., Li, S., Luo, Y., Deng, S., Yang, B. L., Lv, J., Tian, H., Li, X., & Shao, J. (2025) · Nano-Micro Letters · doi:10.1007/s40820-025-01871-5 · Cited by 75
“The integrated and wireless EES allow for the gentle, continuous, and noninvasive monitoring of ECG signals for highly vulnerable babies in intensive care units, exhibiting excellent performance on par with the most sophisticated clinical-standard monitoring devices [164].”
Wang, H., Xiang, Z., Zhao, P., Wan, J., Miao, L., Guo, H., Xu, C., Zhao, W. G., Han, M., & Zhang, H. (2022) · ACS Nano · doi:10.1021/acsnano.2c05299 · Cited by 54
Zhang, G., Li, P., Wang, X., et al. (2022) · Advanced Functional Materials · doi:10.1002/adfm.202208900 · Cited by 50
A dual function flexible sensor for independent temperature and pressure sensing
Hu, R., Li, J., Wu, F., Lu, Z., & Deng, C. (2024) · Chemical Engineering Journal · doi:10.1016/j.cej.2024.152135 · Cited by 52
Connected Health Technology for Cardiovascular Disease Prevention and Management
Wongvibulsin, S., Martin, S. S., Steinhubl, S. R., & Muse, E. D. (2019) · Current Treatment Options in Cardiovascular Medicine · doi:10.1007/s11936-019-0729-0 · Cited by 61
A Self-Powered Biosensor for Monitoring Maximal Lactate Steady State in Sport Training
Mao, Y., Yue, W., Zhao, T., Shen, M., Liu, B., & Chen, S. (2020) · Biosensors · doi:10.3390/bios10070075 · Cited by 55
Highly Selective Biomimetic Flexible Tactile Sensor for Neuroprosthetics
Li, Y., Cao, Z., Li, T., Sun, F., Bai, Y., Lu, Q., Wang, S., Yang, X., Hao, M., Lan, N., et al. (2020) · Research · doi:10.34133/2020/8910692 · Cited by 53
“In the past several years, there have been many studies on flexible force sensors with great performance [13, 14], some of which have already surpassed the sensitivities of human beings [15, 16].”
BackgroundHerbert, R., Lim, H., Park, S., Kim, J., & Yeo, W. (2021) · Advanced Healthcare Materials · doi:10.1002/adhm.202100158 · Cited by 55
Self-citationLuo, Y., Chen, X., Tian, H., Li, X., Lu, Y., Liu, Y., & Shao, J. (2022) · Research · doi:10.34133/2022/9852138 · Cited by 54
Liu, F., Feng, Y., Qi, Y., Liu, G., Zhou, H., Lin, Y., Fan, B., Zhang, Z., Dong, S., & Zhang, C. (2023) · InfoMat · doi:10.1002/inf2.12428 · Cited by 58
Wireless Battery-free and Fully Implantable Organ Interfaces
Bhatia, A., Hanna, J., Stuart, T., Kasper, K. A., Clausen, D. M., & Gutruf, P. (2024) · Chemical Reviews · doi:10.1021/acs.chemrev.3c00425 · Cited by 53
Wireless Technologies for Wearable Electronics: A Review
Kim, C. Y., Lee, J., Jeong, E. Y., Jang, Y., Kim, H., Choi, B., Han, D., Oh, Y., & Jeong, J. (2025) · Advanced Electronic Materials · doi:10.1002/aelm.202400884 · Cited by 69
Self-citationFunctional materials for powering and implementing next-generation miniature sensors
Wu, B., Ouro-Koura, H., Lu, S., Li, H., Wang, X. S., Xiao, J., & Deng, Z. D. (2023) · Materials Today · doi:10.1016/j.mattod.2023.09.001 · Cited by 56
Dewolf, A. H., Labini, F. S., Cappellini, G., Lacquaniti, F., & Ivanenko, Y. P. (2020) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2020.00473 · Cited by 53
Li, X., Zhan, C., Huang, Q., He, M., Yang, C., Yang, C., Huang, X., Chen, M., Xie, X., & Chen, H. (2022) · ACS Applied Nano Materials · doi:10.1021/acsanm.1c04220 · Cited by 54
Spider-inspired tunable mechanosensor for biomedical applications
Kim, T., Hong, I., Roh, Y., Kim, D.-J., Kim, S., Im, S., Kim, C., Jang, K., Kim, S., Kim, M., et al. (2023) · npj Flexible Electronics · doi:10.1038/s41528-023-00247-2 · Cited by 54
Self-citationIntelligent block copolymer self-assembly towards IoT hardware components
Yang, G. G., Choi, H. J., Li, S., Kim, J. H., Kwon, K., Jin, H. M., Kim, B. H., & Kim, S. O. (2024) · Nature Reviews Electrical Engineering · doi:10.1038/s44287-024-00017-w · Cited by 55
Self-citationChen, S., Ouyang, Q., Meng, X., Yang, Y., Li, C., Miao, X.-B., Chen, Z., Zhao, G., Lei, Y., Ghanem, B., et al. (2025) · Science Advances · doi:10.1126/sciadv.adv2406 · Cited by 58
Kim, K., Yoo, J., Shim, J., Ryu, Y., Choi, S.-Y., Lee, J., Lee, H. M., Koo, J., & Kang, S. (2021) · Advanced Materials Technologies · doi:10.1002/admt.202001297 · Cited by 57
Self-citationLantern‐Inspired On‐Skin Helical Interconnects for Epidermal Electronic Sensors
Li, D., Cui, T., Jian, J., Yan, J., Xu, J., Li, X., Li, Z., Yan, A., Chen, Z., Shao, W., et al. (2023) · Advanced Functional Materials · doi:10.1002/adfm.202213335 · Cited by 52
Flexible Supercapacitor Integrated Systems
He, A., He, J., Cao, L., Chen, J., Cheng, B., Ma, R., Ding, Y., Yang, G., & Fang, Y. (2024) · Advanced Materials Technologies · doi:10.1002/admt.202301931 · Cited by 50
Capacitive in-sensor tactile computing
Chen, Y., Cao, J., Qiu, J., Yang, D., Liu, M., Zhang, M., Li, C., Yu, J., Yu, J., Zhang, X., et al. (2025) · Nature Communications · doi:10.1038/s41467-025-60703-7 · Cited by 76
Implantable bioelectronics and wearable sensors for kidney health and disease
Madhvapathy, S. R., Cho, S., Gessaroli, E., Forte, E., Xiong, Y., Gallon, L. G., & Rogers, J. A. (2025) · Nature Reviews Nephrology · doi:10.1038/s41581-025-00961-2 · Cited by 72
Self-citationA stretchable laminated GNRs/BNNSs nanocomposite with high electrical and thermal conductivity
Tan, C., Zhu, H., Ma, T., Guo, W., Liu, X., Huang, X., Zhao, H., Long, Y., Jiang, P., & Sun, B. (2019) · Nanoscale · doi:10.1039/c9nr06060j · Cited by 46
Recent progress of skin-integrated electronics for intelligent sensing
Li, D., Yao, K., Gao, Z., Liu, Y., & Yu, X. (2021) · Light Advanced Manufacturing · doi:10.37188/lam.2021.004 · Cited by 51
3D designed battery-free wireless origami pressure sensor
Kim, T., Kalhori, A. H., Kim, T., Bao, C., & Kim, W. S. (2022) · Microsystems & Nanoengineering · doi:10.1038/s41378-022-00465-0 · Cited by 47
Novel fabrication techniques for ultra-thin silicon based flexible electronics
Lee, J. Y., Ju, J. E., Lee, C., Won, S. M., & Yu, K. J. (2024) · International Journal of Extreme Manufacturing · doi:10.1088/2631-7990/ad492e · Cited by 57
Self-citationDeformable and Stretchable Electrodes for Soft Electronic Devices
Kim, Y. H., Kweon, O. Y., Won, Y., & Oh, J. H. (2019) · Macromolecular Research · doi:10.1007/s13233-019-7175-4 · Cited by 45
Self-powered flexible and transparent smart patch for temperature sensing
Chen, X., Ren, Z., Guo, H., Cheng, X., & Zhang, H. (2020) · Applied Physics Letters · doi:10.1063/1.5134526 · Cited by 44
Applications of Cellulose Nanomaterials in Stimuli-Responsive Optics
Peng, Z., Lin, Q., Tai, Y.-A. A., & Wang, Y. (2020) · Journal of Agricultural and Food Chemistry · doi:10.1021/acs.jafc.0c04742 · Cited by 46
Marks, H. L., Bucknor, A. E. M., Roussakis, E., Nowell, N. H., Kamali, P., Cascales, J. P., Kazei, D. D., Lin, S. J., & Evans, C. L. (2020) · Science Advances · doi:10.1126/sciadv.abd1061 · Cited by 43
“Recently, skin interfaced wireless sensors for wearable (%stO 2 ) monitoring in neonates have been pioneered by Chung et al. ( 19 , 20 ) and show great promise for overcoming the typical setbacks of continuous (%stO 2 ) monitoring, once made cost effective and widely available.”
BackgroundBioderived and degradable polymers for transient electronics
Uva, A., Lin, A., Babi, J., & Tran, H. (2021) · Journal of Chemical Technology & Biotechnology · doi:10.1002/jctb.6790 · Cited by 42
Stuart, T., Kasper, K. A., Iwerunmor, I. C., McGuire, D. T., Peralta, R., Hanna, J., Johnson, M., Farley, M., LaMantia, T., Udorvich, P., et al. (2021) · Science Advances · doi:10.1126/sciadv.abj3269 · Cited by 46
Wang, P. P., Lan, Y., Huan, C., Luo, J., Cai, W., Fan, J., He, X., Huang, Z., Zhu, H. L., Ke, Q., et al. (2023) · Materials Science and Engineering R Reports · doi:10.1016/j.mser.2023.100759 · Cited by 50
Yao, G., Gan, X., & Lin, Y. (2024) · 中国科学通报:英文版 · doi:10.1016/j.scib.2024.05.012 · Cited by 46
Metadata
- Authors
- Chung, H. U., Kim, B. H., Lee, J. Y., Lee, J., Xie, Z., Ibler, E. M., Lee, K., Banks, A., Jeong, J. Y., Kim, J., Ogle, C., Grande, D., Yu, Y., Jang, H., Assem, P., Ryu, D., Kwak, J. W., Namkoong, M., Park, J. B., Lee, Y., Kim, D. H., Ryu, A., Jeong, J., You, K., Ji, B., Liu, Z., Huo, Q., Feng, X., Deng, Y., Xu, Y., Jang, K.-I., Kim, J., Zhang, Y., Ghaffari, R., Rand, C. M., Schau, M., Hamvas, A., Weese-Mayer, D. E., Huang, Y., Lee, S. M., Lee, C. H., Shanbhag, N. R., Paller, A. S., Xu, S., & Rogers, J. A.
- Affiliations
- Northwestern University (US)
- University of Illinois Urbana-Champaign (US)
- Department of Materials Science and Engineering, Northwestern University, Evanston, IL 60208, USA.
- Agency for Science, Technology and Research (SG)
- Tsinghua University (CN)
- Daegu Gyeongbuk Institute of Science and Technology (KR)
- Lurie Children's Hospital (US)
- Center for Autonomic Medicine, Department of Pediatrics, Ann & Robert H. Lurie Children’s Hospital of Chicago, Chicago, IL 60611, USA.
- Kookmin University (KR)
- Purdue University West Lafayette (US)
- Corresponding author
- Chung, Ha Uk
- Kim, Bong Hoon
- Lee, Jong Yoon
- Lee, Jungyup
- Xie, Zhaoqian
- Paller, Amy S.
- Xu, Shuai
- Rogers, John A.
- ORCID
- Chung, Ha Uk
- Kim, Bong Hoon
- Lee, Jong Yoon
- Lee, Jungyup
- Xie, Zhaoqian
- Lee, KunHyuck
- Banks, Anthony
- Jeong, Ji Yoon
- Kim, Jongwon
- Grande, Dominic
- Yu, Yongjoon
- Jang, Hokyung
- Assem, Pourya
- Kwak, Jean Won
- Namkoong, Myeong
- Park, Jun Bin
- Lee, Yechan
- Kim, Do Hoon
- Ryu, Arin
- You, Kevin
- Ji, Bowen
- Liu, Zhuangjian
- Huo, Qingze
- Feng, Xue
- Deng, Yujun
- Xu, Yeshou
- Jang, Kyung-In
- Kim, Jeonghyun
- Zhang, Yihui
- Ghaffari, Roozbeh
- Rand, Casey M.
- Hamvas, Aaron
- Huang, Yonggang
- Lee, Seung Min
- Lee, Chi Hwan
- Shanbhag, Naresh R.
- Paller, Amy S.
- Xu, Shuai
- Rogers, John A.
- Venue
- Science
- Published
- 2019-03
- Volume
- 363
- Issue
- 6430
- ISSN
- 0036-8075 (ISSN-L)
- 1095-9203
- PMC
- PMC6510306
- OpenAlex
- W2916533817
- Semantic Scholar
- b0511afac0cc252946aa8f08a078b12e48030581
- Cited by
- 807
- References
- 43
- Citations by year
- 2026: 72 · 2025: 112 · 2024: 103 · 2023: 121 · 2022: 124 · 2021: 131 · 2020: 97 · 2019: 47
- FWCI
- 35.10
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by
- Locations
- Science (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Funders
- Bill and Melinda Gates Foundation (OPP1182909)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Physical Sciences › Engineering › Biomedical Engineering › Wireless Body Area Networks
- Keywords
- epidermal electronics, neonatal intensive care, battery-free operation, neonatal monitoring
- MeSH terms
- Diagnostic Imaging, Electronics, Equipment Design, Humans, Infant, Newborn, Monitoring, Physiologic, Skin, Intensive Care, Neonatal, Vital Signs, Lab-On-A-Chip Devices, Wireless Technology
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.