Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care
Open accessThe DOI matches its Crossref record; 44 of the 44 references it cites were found in a registry or an index; 12 of the 42 sources its claims cite were checked for support.
Abstract
Sensitive sensing Neonatal care, particularly for premature babies, is complicated by the infants' fragility and by the need for a large number of tethered sensors to be attached to their tiny bodies. Chung et al. developed a pair of sensors that only require water to adhere to the skin and allow for untethered monitoring of key vital signs (see the Perspective by Guinsburg). On-board data processing allowed for efficient wireless near-field communication using standard protocols. The absence of cables makes it easier to handle the infants and allows for skin-to-skin contact between the babies and their parents or caregivers. Science , this issue p. eaau0780 ; see also p. 924
Overview
CoChat Summary
TL;DR
The paper introduces a pair of ultrathin, wireless, battery-free epidermal electronic systems (EES) for neonatal intensive care: one chest-mounted ECG device and one foot-mounted PPG device. Pilot studies in neonates show clinical-grade agreement for heart rate, respiratory rate, and SpO₂, plus added capabilities for skin temperature and pulse arrival time as a systolic blood pressure surrogate.
Key findings
- The binodal EES continuously streams ECG, PPG, and skin temperature data, yielding HR, HRV, RR, SpO₂, and PAT; wireless data rates reach up to 800 bytes/s with dual channels, NFC operation works up to 25 cm through NICU materials, and BLE range reaches up to 20 m.
- The devices have effective moduli of 200 to 300 kPa, adhere via van der Waals forces, and show peel forces about a factor of ~10 lower than conventional NICU adhesives, with maximum von Mises stress on skin reduced by a factor of 4.3; the microfluidic channel reduces interfacial stresses by up to a factor of 2.5, and adhesion energy is G = 16 N/m.
- For MRI compatibility, mesh electrodes reduce the in-plane magnetic field gradient by a factor of ~3 and the out-of-plane gradient by a factor of 4 versus commercial electrodes; simulated copper heating is 1°C with maximum skin temperature change of 0.04°C, and experimental 3-T MRI measurements show ~0.1°C temperature difference, meeting FDA “MRI-safe” requirements.
- In neonates (n = 3, gestational age 28 to 40 weeks), mean differences versus standard-of-care monitors were –0.17 beats per minute for HR, 0.76 breaths per minute for RR, and 1.02% for SpO₂; additional studies in n = 18 neonates with gestational ages as low as 28 weeks and weights as low as 1470 g showed similar performance, and scaled production cost is estimated at less than $20 USD per unit.
Methods
- Fabrication used semiconductor processing, lamination, transfer printing, chip placement, and solder bonding; each EES includes serpentine Cu traces, chip-scale ICs, fractal mesh electrodes, an ionic-liquid microfluidic chamber, silicone encapsulation, and an NFC loop antenna for simultaneous wireless power and data.
- In-sensor analytics include a streamlined Pan-Tompkins algorithm for real-time QRS detection to yield HR and HRV, dynamic baseline control and SpO₂ calculation for PPG, and time-synchronized ECG/PPG processing for PAT.
- Validation included healthy adult volunteers (n = 3) against FDA-cleared monitoring equipment, and neonates in two tertiary-level NICUs: n = 3 with synchronous standard-of-care measurements and n = 18 using related device platforms with onboard power supplies; the protocol was IRB-approved and registered on ClinicalTrials.gov (NCT02865070).
Limitations
- The paper states that results are preliminary feasibility testing and validation; comprehensive clinical studies are ongoing to verify measurements across increased age groups and ethnic backgrounds, and remaining challenges include nursing acceptance, compatibility with legacy monitoring systems, and device sterilization for reuse.
- The paper notes that additional testing is needed for altered skin barrier function, such as burn victims or patients with epidermolysis bullosa.
- Evident from the text: the neonatal validation cohorts are small (n = 3 with synchronous standard-of-care comparisons; n = 18 using related platforms with onboard power supplies), adult validation is n = 3, and temperature and PAT data in neonates are shown without reference comparators because conventional devices acquire these only periodically.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures









Verification
Claims the paper makes
6 Valid · 2 Not valid · 18 Not checked · 82 Own results and interpretations
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“A key example is the measurement of pulse arrival time (PAT), defined by the time lapse between the maximum fiducial point in the ECG signal (R peak) and the corresponding minimal fiducial point in the PPG signal at valley as in Fig. 6H, as a direct correlate to systolic blood pressure (29 , 30).”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Real-time measurements, in-sensor analytics, and data transmission
What the sources say
- Partly supportedcites [29] Kim J. S., Chee Y. J., Park J. W., Choi J. W., Park K. S., A new approach for non-intrusive monitoring of blood pressure on a toilet seat. Physiol. Meas. 27, 203– 211 (2006). doi: 10.1088/0967-3334/27/2/010; pmid: 16400206 sourceExperimental — accuracy not yet measured
Section 4.1 reports that PAT was significantly, but moderately, correlated with systolic blood pressure. Section 2.2 defines the study’s PAT using the ECG R-peak and the PPG’s maximum first-derivative point, not the PPG valley; it says the minimum point was unclear in its signal. The claimed correlation is supported, but the attributed measurement definition is not. (LLM, conf=0.87)
- Not checked — no full text of the sourcecites [30] Geddes L. A., Voelz M. H., Babbs C. F., Bourland J. D., Tacker W. A., Pulse transit time as an indicator of arterial blood pressure. Psychophysiology 18, 71– 74 (1981). doi: 10.1111/j.1469-8986.1981.tb01545.x; pmid: 7465731 source
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The inherently thin, soft mechanical properties of the sensors (Fig. 1A) reported here allow for adhesion via van derWaals forces alone.”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Low-modulus mechanics, soft interface adhesion, and implications for neonatal skin safety
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The mechanical decoupling afforded by the microfluidic channel decreases these stresses by up to a factor of 2.5 (Fig. 2B) relative to otherwise similar designs without the microfluidics.”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Low-modulus mechanics, soft interface adhesion, and implications for neonatal skin safety
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Oscillations in the force only appear for holes larger than the characteristic size of the cohesive zone (~500 mm, as in Fig. 2I).”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Low-modulus mechanics, soft interface adhesion, and implications for neonatal skin safety
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Specifically, the interface starts to delaminate when the normal stress reaches ~20 kPa (Fig. 2G, inset).”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Low-modulus mechanics, soft interface adhesion, and implications for neonatal skin safety
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
61 more claims
Create a free account to see all 108 claims and to verify the 1 reference still open.
- Probably not a substantial claim. 14
- It may state two separate claims as one. 33
- The quote only partly states this claim. 2
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Researchers have developed various versatile epidermal electrodes with high performance that can universally measure EEG, EMG, EOG, and ECG through placing electrodes on different areas of the skin.[25,414]”
From Flexible Electronics and Devices as Human–Machine Interfaces for Medical Robotics (2022) · cites this paper as [414]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Previous reports validated the performance of an FDA-cleared physiological monitoring system against gold standard monitoring platforms [18–23].”
From Pregnancy outcomes in infertility patients diagnosed with sleep disordered breathing with wireless wearable sensors (2022) · cites this paper as [18]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“In previous work, we demonstrate the ability for the chest- and finger-mounted sensors to monitor vital signs in both children and adults; the abdominal sensor allows for maternal–fetal monitoring of pregnant women in both high- and low-resource settings as well as the clinical and home settings (22, 23).”
From Comprehensive pregnancy monitoring with a network of wireless, soft, and flexible sensors in high- and low-resource health settings (2021) · cites this paper as [23]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“and have been used in neonatal and pediatric intensive care units. 4,5”
From Advanced wireless monitoring for children in the cardiac perioperative setting (2023) · cites this paper as [4]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“However, this system is limited by (1) the modest maximum operating distances (~30 cm) supported by near-field communication (NFC) protocols used for power transfer and data communication, (2) the mechanically fragile nature of the ultrathin, compliant mechanics designs, (3) the sufficient, but constrained range of measurement capabilities, and (4) the demand for highly advanced device configur”
From Skin-interfaced biosensors for advanced wireless physiological monitoring in neonatal and pediatric intensive-care units (2020) · cites this paper as [9]
Experimental — accuracy not yet measured
5 more claims about this paper
Create a free account to see every claim other papers make about this one.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Compatibility with medical imaging techniques used in the NICU
“eliminating any parasitic heating from magnetically induced eddy currents.”
§ Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care > Compatibility with medical imaging techniques used in the NICU
“Full three-dimensional multi-physicsmodeling shows that, at the end of a single scan for 0.5 ms, the copper layer of an ECG EES undergoes heating by only 1°C (Fig. 3H).”
Citations
Cites
A Real-Time QRS Detection Algorithm
Pan, J., & Tompkins, W. J. (1985) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1985.325532 · Cited by 7,855
Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis
Gao, W., Emaminejad, S., Nyein, H. Y. Y., Challa, S., Chen, K., Peck, A., Fahad, H. M., Ota, H., Shiraki, H., Kiriya, D., et al. (2016) · Nature · doi:10.1038/nature16521 · Cited by 5,087
Kim, D.-H., Lu, N., Ma, R., Kim, Y.-S., Kim, R.-H., Wang, S., Wu, J., Won, S. M., Tao, H., Islam, A., et al. (2011) · Science · doi:10.1126/science.1206157 · Cited by 4,836
Self-citationWearable Sensors for Remote Health Monitoring
Majumder, S., Mondal, T., & Deen, M. (2017) · Sensors · doi:10.3390/s17010130 · Cited by 1,375
Kiserud, T., Piaggio, G., Carroli, G., Widmer, M., Carvalho, J., Neerup Jensen, L., Giordano, D., Cecatti, J. G., Abdel Aleem, H., Talegawkar, S. A., et al. (2017) · PLOS Medicine · doi:10.1371/journal.pmed.1002220 · Cited by 729
Thermochemistry of ionic liquid heat-transfer fluids
Valkenburg, M. E. V., Vaughn, R. L., Williams, M., & Wilkes, J. S. (2005) · Thermochimica Acta · doi:10.1016/j.tca.2004.11.013 · Cited by 595
Halcox, J. P. J., Wareham, K., Cardew, A., Gilmore, M., Barry, J. P., Phillips, C., & Gravenor, M. B. (2017) · Circulation · doi:10.1161/circulationaha.117.030583 · Cited by 580
Self-assembled three dimensional network designs for soft electronics
Jang, K.-I., Li, K., Chung, H. U., Xu, S., Jung, H. N., Yang, Y., Kwak, J. W., Jung, H. H., Song, J., Yang, C., et al. (2017) · Nature Communications · doi:10.1038/ncomms15894 · Cited by 445
Self-citationChen, W., Kobayashi, T., Ichikawa, S., Takeuchi, Y., & Togawa, T. (2000) · Medical & Biological Engineering & Computing · doi:10.1007/bf02345755 · Cited by 425
Harris, K. D., Elias, A. L., & Chung, H.-J. (2016) · Journal of Materials Science · doi:10.1007/s10853-015-9643-3 · Cited by 403
Self-citation
Research works printed but not linked: 2
- O. Sharma, S. N. Lewis, U. Telang, L. D’Almeida, L. E. S. Lewis, Design of a Bluetooth Enabled Health Monitoring System for Infants Using Wearable Technology. J. Adv. Res. Dyn. Contrl. Syst. 15, 887–894 (2017). http://jardcs.org/papers/v9/sp/20181075.pdf
- C. H. Lund J. A. Tucker in Neonatal Skin: Structure and Function S. B. Hoath H. I. Maibach Eds. (Dekker ed. 2 2003) pp. 299–324.
Cited by
Items 701–800 of 800
Elias, V., Rabih, A., Bin, S., et al. (2021) · IEEE International Conference on Design & Test of Integrated Micro & Nano-Systems (DTS) · doi:10.1109/dts52014.2021.9498235 · Cited by 1
Optical Properties of Biopolymers
Eldhose, M., George, C., John, S., et al. (2023) · Springer eBooks · doi:10.1007/978-981-16-6603-2_9-1 · Cited by 1
Optical Properties of Biopolymers
Eldhose, M., George, C., John, S., et al. (2023) · Springer eBooks · doi:10.1007/978-981-19-0710-4_9 · Cited by 1
Wearables – A Revolution in Neonatal Monitoring?
Sivanandan, S. (2023) · The Indian Journal of Pediatrics · doi:10.1007/s12098-023-04791-7 · Cited by 1
“ANNE monitoring system is another wireless vital sign monitoring system that uses small, light sensors attached to the neonate’s skin [3].”
BackgroundLauw, C. J., Rahman, J. S., Brankovic, A., et al. (2023) · Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/embc40787.2023.10340576 · Cited by 1
“However, since the 1970s, there have been minimal improvements made to neonatal monitoring systems [8], coupled with limited adaptations to adult system monitors being used for neonates [9].”
BackgroundIncorporation of soft materials for flexible electronics
Lee, S. M., Lee, J., & Kang, S. (2024) · Elsevier eBooks · doi:10.1016/b978-0-323-99147-6.00007-7 · Cited by 1
Self-citationAdvances in Maternity Technology
Bates, M. (2023) · IEEE Pulse · doi:10.1109/mpuls.2024.3353653 · Cited by 1
“The data gathered by the sensors can be wirelessly transmitted via radio frequencies to nurses’ station displays and tablets [6].”
BackgroundNan, Y. J., Ning, D. J., Cao, Y., Yi, H., Zhan, L., Gao, Y., Ge, K., Ji, H., Li, M., & Feng, H. (2024) · Sensors · doi:10.3390/s25010037 · Cited by 1
Biodegradable and sustainable sensors: materials and applications
Silva‐Neto, H. A., Rocha, D. S., Souza, L. K. A., & Coltro, W. K. T. (2025) · Elsevier eBooks · doi:10.1016/b978-0-443-22002-9.00006-3 · Cited by 2
Williams, E. E., Ascherl, R., Gaertner, V. D., Sibrecht, G., Kurul, Ş., Herrmann, M.-L., Szakmár, E., Raffaeli, G., Bresesti, I., & Jost, K. (2025) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-025-01310-1 · Cited by 3
“Such modules are water resistant and compatible with imaging techniques, additionally they have a high mechanical compliance and a non-invasive skin adhesive interface [18], thus making them better suited for the fragile skin of extreme preterm infants.”
Biopolymers in portable electronics applications and case study
Ghosh, K., Singh, C. J., Srivastava, M., Behera, S., Kaushik, S. S., Samanta, A., & Venkatesan, H. (2025) · Elsevier eBooks · doi:10.1016/b978-0-443-31522-0.00002-6 · Cited by 1
Developments in perioperative management: The yin to the yang of congenital heart surgery
Ceneri, N. M., Desai, M. H., & Yerebakan, C. (2021) · Journal of Thoracic and Cardiovascular Surgery · doi:10.1016/j.jtcvs.2021.05.027 · Cited by 1
Networks and near-field communication: up-close but far away
Kim, J. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-818914-6.00019-3 · Cited by 2
Self-citationCamera-Based Remote Photoplethysmography for Physiological Monitoring in Neonatal Intensive Care
Sweely, B., Liu, Z., Wyatt, T. H., Newnam, K. M., Qi, H., & Zhao, X. (2021) · IEEE International Conference on Signal and Image Processing Applications (ICSIPA) · doi:10.1109/icsipa52582.2021.9576779 · Cited by 1
Self-citationWalter, J. R., Kim, D. H., Myers, D., Hill, M. A., Snoll, B., Lee, J. Y., Kulikova, E., Fagan, K., Cauinian, R., Nguyen, L., et al. (2022) · Journal of the American Geriatrics Society · doi:10.1111/jgs.17673 · Cited by 1
“The two units are time-synchronized generating pulse transit time and continuous blood pressure.8 Sensors record data and automatically download it to the cloud and tablet via encrypted Bluetooth.7 Patients with active skin conditions, major psychiatric disorders, or inability to consent were excluded.”
BackgroundMethodInfluentialSelf-citationSmith, H. E. (2019) · Advances in Neonatal Care · doi:10.1097/anc.0000000000000637 · Cited by 1
Flexible Piezoelectric Sensors
(2022) · Wiley eBooks · doi:10.1002/9783527833115.ch10 · Cited by 1
Materials for Developing Future Flexible Electronic Device
Ahsan, M., Tasnim, M. N., Haider, J., Shahjalal, M., & Hashmi, S. (2023) · Elsevier eBooks · doi:10.1016/b978-0-12-819728-8.00095-4 · Cited by 1
Lee, S. H., Lee, K.-R., Kim, T., Im, S., Lee, Y. J., Jeong, S., Shin, H., Kim, M., Lee, J., Kim, D., et al. (2023) · Research Square · doi:10.21203/rs.3.rs-2844027/v1 · Cited by 1
Self-citationA Wearable and Unobstructive Qi-Compatible Wireless ChargerPreprint
Lu, N., Kim, S., Wells, J., Bhattacharya, S., Nathan, H., He, J., Tubilla, I., Huh, H., Kakani, P., Farshkaran, A., et al. (2023) · Research Square · doi:10.21203/rs.3.rs-3509616/v1 · Cited by 2
Imperceptible Liquid Metal Based Tattoo for Human-Machine Interface on Hairy SkinPreprint
Lin, W., Ai, L., Wang, Y., et al. (2024) · SSRN Electronic Journal · doi:10.2139/ssrn.4748583 · Cited by 1
Structural Design of Flexible and Stretchable Electronics
Xie, Z., Zhao, Z., & Avila, R. O. (2024) · Wiley eBooks · doi:10.1002/9783527842308.ch6 · Cited by 1
Self-citationAn artificial nervous system for communication between wearable and implantable therapeuticsPreprint
Ghanim, R. W., Lee, Y. J., Byun, G., Jackson, J., Ding, J. Z., Feller, E. K., Kim, E., Aygun, D., Kaushik, A., Cig, A., et al. (2025) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2025.06.04.657863 · Cited by 1
“Other wireless communication and power transfer methods, such as near-field communication (NFC), require bulky external antennae that limit patient adherence and require centimeter-scale alignment with targeted devices [19, 20].”
Self-citationA Wireless Neonatal Intensive Care Unit: Fiction or Closer Reality?
Sénéchal, E., Chung, H. U., Rogers, J. A., & SantʼAnna, G. (2025) · Neonatology · doi:10.1159/000547056 · Cited by 2
“Recent work at Northwestern University shows that a pair of devices of this type, each of which gently and non-invasively adheres to the fragile skin of a premature neonate, is capable of capturing complete, clinical grade vital signs information without any wires or cables [24, 25].”
Self-citationImproving care of critically ill newborns
Guinsburg, R. (2019) · Science · doi:10.1126/science.aaw2085
Position: Smart KT Tape -- A Bendable Wearable System for Muscle Fatigue Sensing
Chen, J., Liu, C. Y.-S., Chen, H. J., et al. (2019) · doi:10.1145/3325424.3329671
“Emerged recently are total system solutions where the sensing, processing, and communication capabilities are integrated on a System on Chip (SoC) [4] or an FPCB [5].”
BackgroundSelf-citationDing, H., Lv, G., Shi, Z. H., et al. (2020) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2020.07.05.188391
Self-citationControlled Wave Structure of the Microfiber for Stretchable Optical Sensors
Zhu, H., & Xu, F. (2020) · Asia Communications and Photonics Conference · doi:10.1364/acpc.2020.t3a.2
“The devices which possess the flexibility and even stretchability show great application potentials in smart wearable devices[1], medical equipment[2], motion monitoring[3], and robotics.”
BackgroundPediatric Cerebral Hemodynamic Monitoring Goes Wireless
Hurley, D. (2021) · Neurology Today · doi:10.1097/01.nt.0000733412.90842.87
Weese‐Mayer, D. E., & Gonik, R. (2021) · Clinical Autonomic Research · doi:10.1007/s10286-021-00807-4
Self-citationBorrero, I., & Maria, L. (2021)
Ishac, D., Matta, S., Bin, S., Aziz, H., Karam, E. G., Abche, A., & Nassar, G. (2021) · Cellular and Molecular Bioengineering · doi:10.1007/s12195-021-00712-w · Cited by 1
Song, Y., Gao, W. F., & Zhang, H. (2021) · Wiley eBooks · doi:10.1002/9783527833474.ch1
Li, H., Zhang, S., Zuo, S., Zhang, Z., Wang, B., Cao, L., Chen, X., Wang, R., Gong, Z., & Prades, J. D. (2020) · doi:10.2196/preprints.22527
Minimally-invasive and non-invasive flexible devices for robust characterizations of deep tissues
Yuan, Y., Tang, W., Wu, H., Chen, S., Ren, X., Hu, B., & Gu, N. (2022) · Scientia Sinica Chimica · doi:10.1360/ssc-2022-0100
Seeds on a parachute: the technology of greening
Zdeshchyts, A. V., & Zdeshchyts, V. M. (2022) · IOP Conference Series Earth and Environmental Science · doi:10.1088/1755-1315/1049/1/012024 · Cited by 1
Sensors and Platforms for Flexible Bioelectronics
Sonkusale, S., Baghini, M. S., & Aeron, S. (2022) · Springer eBooks · doi:10.1007/978-3-030-98538-7_2
Fei, Q. (2021) · MOspace Institutional Repository (University of Missouri) · doi:10.32469/10355/90962
Silk Nanofibrous Iontronic Sensors for Accurate Blood Pressure MonitoringPreprint
Wang, S., Xiao, J., Liu, H., et al. (2022) · SSRN Electronic Journal · doi:10.2139/ssrn.4094622
Highly adhesive stretchable polymer and highly dynamic stable human electrophysiological monitoring
Hu, M., & Yang, H. (2022) · Scientia Sinica Chimica · doi:10.1360/ssc-2022-0155
Next Generation of High‐Performance Printed Flexible Electronics
Dahiya, A. S., Kumaresan, Y., & Dahiya, R. (2022) · Wiley eBooks · doi:10.1002/9781119869610.ch9
The Flexible and Wearable Pressure Sensing Microsystems for Medical Diagnostics
Li, H., Lan, R., Chen, J., & Li, L. (2022) · Springer eBooks · doi:10.1007/978-3-031-16084-4_10 · Cited by 2
Structural Engineering of Flexible Electronics
Huang, Y., Su, Y., & Jiang, S. (2022) · Springer eBooks · doi:10.1007/978-981-19-6623-1_1
Self-citationLiu, Z., Li, J., & Li, G. (2022) · International Conference on Microwave and Millimeter Wave Technology (ICMMT) · doi:10.1109/icmmt55580.2022.10022567
“save conventional healthy monitoring costs and significantly improve quality of life [1].”
BackgroundSelf-citationSheng, X., Wenxin, 赵. Z., Lizhu, 李. L., Yunxiang, 黄. H., & Ding, H. (2023) · Chinese Journal of Lasers · doi:10.3788/cjl221562 · Cited by 1
Devices based on the ECG signal
Bolić, M. (2023) · Elsevier eBooks · doi:10.1016/b978-0-12-820947-9.00001-5
Devices based on the time difference between signals
Bolić, M. (2023) · Elsevier eBooks · doi:10.1016/b978-0-12-820947-9.00007-6
Photonic Skin Based on Microfiber Bragg Grating for Pulse Wave Detection
Zhu, H., Luo, J., Zhu, S., Yang, H., & Xu, F. (2023) · Asia Communications and Photonics Conference · doi:10.1109/acp/poem59049.2023.10369435
“They have been utilized for monitoring various parameters such as blood pressure[4], blood glucose[5], intraocular pressure[6, 7], and so on.”
MethodFlexible Electronics for Wearable Sensing Systems
Fu, Q. (2024) · Wiley eBooks · doi:10.1002/9783527841080.ch6 · Cited by 3
Direct mask printing of highly stretchable liquid metal electrodes for strain sensors
Sun, W., Luo, Z., & Li, X. (2024) · doi:10.1117/12.3025078
“Flexible electronics have exhibited great potential for applications in the fields of electronic skin [1], [2], soft robots [3], human computer interaction [4], wearable devices [5], [6], and flexible sensors[7]–[9].”
BackgroundZhang, Y., Wang, S., Chen, Z., Liu, B., Liu, J., Yang, H. W., Hu, Y., Fu, Y., & Wu, Q. (2024) · SSRN Electronic Journal · doi:10.2139/ssrn.4874890
Self-citationA Mechanics and Electromagnetic Scaling Law for Highly Stretchable Radio Frequency ElectronicsPreprint
Huang, Y. Y., Zhao, Z., Avila, R. O., et al. (2024) · SSRN Electronic Journal · doi:10.2139/ssrn.4881995
Self-citationEnhanced Prediction of Swimmer Fitness Using Modified Resilient PSO Algorithm
Poornima, K. G., & Prasad, K. K. (2024) · Transactions of Indian National Academy of Engineering · doi:10.1007/s41403-024-00495-2 · Cited by 1
Zhu, M., & Zhang, Y. (2024) · 中国科学通报:英文版 · doi:10.1016/j.scib.2024.08.020 · Cited by 1
Self-citationDigital environmental health: a digital platform for preliminary prevention and intervention
Seo, S. (2024) · Women s Health Nursing · doi:10.4069/whn.2024.08.31
“The digital transformation began with binary computing systems and has evolved through various stages, including circuit chips, personal computers, the Internet, the World Wide Web, social media platforms, and smartphones [2].”
MethodHigh Sensitivity Resistive Vibration Sensing: Optimization for Wearable Heart Rate Monitoring
Abeeleh, A. A., Albagdady, A., Al‐Mahmodi, M., Al‐Shishani, G., & Al-Halhouli, A. T. (2024) · International Conference on Research and Education in Mechatronics (REM) · doi:10.1109/rem63063.2024.10735693
“Chung et al. [7], presented stretchable and wearable sensor capable of measuring several vital signs such as HR, HRV, oxygen saturation (SPO2) and respiratory rate.”
BackgroundAdvances in Home Care and Real-Time Vital Signs Monitoring
Lucca, G., Dalmazo, B. L., Bertaco, D., et al. (2024) · Lecture notes in computer science · doi:10.1007/978-3-031-77738-7_28
Applications of Nanotechnology in Medicine with a Focus on Diagnostic Imaging
Farghal, M. (2024) · Premier journal of science. · doi:10.70389/pjs.100037 · Cited by 2
Neonatal General Movements Artificial Intelligence Assessment and Its Clinical Practice
Huang, X., Yi, M., & Han, T. (2025) · IntechOpen eBooks · doi:10.5772/intechopen.1009195 · Cited by 1
Chen, Z., Wang, S., Zhang, Y., Wu, Q., Zheng, F., Liu, J., Yang, H. L., Hu, Y., Fu, Y., & Liu, B. (2025) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2025.3600735 · Cited by 2
“It has a wide range of applications in various fields such as human–computer interaction (HCI), health monitoring, motion analysis, robotics, and virtual reality (VR) / augmented reality (AR) technology [5], [6], [7], [8], [9], [10].”
Self-citationYiu, C. K., Wu, P., Zhou, J., & Wooyoung, P. (2025) · Wiley eBooks · doi:10.1002/9783527845507.ch14
Wireless Colorimetric Multi-Biomarker Sensing to Enable Critical Neonatal MonitoringPreprint
Castelblanco, A., Ruggeri, E., Matzeu, G., et al. (2025) · medRxiv · doi:10.1101/2025.09.21.25336258
“Not only is the use of wired sensors in neonatal intensive care known to obstruct daily care tasks and reduce parental skin bonding [8], but the subsequent need for invasive procedures also results in limitations for health care and trauma for the newborn and its family.”
Computer-Based Wearable Devices for Remote Patient Monitoring
Anura, A., & Maji, D. (2025) · Springer eBooks · doi:10.1007/978-981-96-9346-7_7
Integrating ML Approaches for Photodetectors
Bhattacharya, S., Ghosh, S., Deb, D., Hiremath, P., & Nath, D. (2025) · Progress in optical science and photonics · doi:10.1007/978-981-95-1687-2_9
Stretchable radio-frequency electronics with programmable mechanical and electromagnetic responses
Carusetta, N., Lee, C. G., & Avila, R. O. (2025) · International Journal of Extreme Manufacturing · doi:10.1088/2631-7990/ae2e5c
“Reproduced from [2] .”
Self-citationSkin-inspired soft bioelectronic epidermal sensing, implantable applications and system integrations
Wei, C., & Zheng, K. (2025) · Materials Today Physics · doi:10.1016/j.mtphys.2025.101988 · Cited by 1
Wang, X., Guo, W., Jiao, X., Jin, Z., Chuai, C., Pan, J., Sun, H., Sun, Q., Wang, S., & Xu, P. (2025) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2025.117398
Applications for Electronic Devices by High-Resolution Printing of Functional Nanoparticle Layers
Kusaka, Y., WATANABE, Y., & Takei, A. (2025) · Journal of The Surface Finishing Society of Japan · doi:10.4139/sfj.76.313
Fu, K., Zhao, J., Deng, J., Zhao, C., Wang, X., & Qiu, L. (2026) · IEEE Sensors Journal · doi:10.1109/jsen.2026.3652555
“To address these limitations, skin-conformal “electronic tattoo” sensors have emerged as a promising platform for next-generation HCI [24], [25], [26], [27], [28].”
Design of 1V Low-Power CMOS TIA with High Performance for Integrated Biomedical Applications
Koozehkanani, R. D. (2026) · Journal of Circuits Systems and Computers · doi:10.1142/s0218126626501616
Ju, L., Yu, B., Wang, R., Chen, H., Hou, K., Zhang, C., Geng, M., Dai, Y., Zhang, Y., Guan, Y., et al. (2026) · Advanced Science · doi:10.1002/advs.202522350
Self-citationAdvanced silicon nanomembrane based bioelectronics for flexible and stretchable implantable systems
Lee, J., Lee, Y., Woo, H., Hong, M., Lee, D. J., & Sang, M. (2026) · Biomedical Engineering Letters · doi:10.1007/s13534-026-00568-x · Cited by 1
Self-citationBio‐Inspired Mechanical Amplification Block on Implantable Tactile Sensors
Choi, S., Ye, Y., Jeong, C., Mun, Y., Oh, S., & Kim, T. (2026) · Advanced Electronic Materials · doi:10.1002/aelm.202500874
“Although prior works have advanced sensor sensitivity and signal amplification [ 15 ], they do not resolve the embedded sensitivity loss driven by mechanical shielding [ 16, 17 ].”
Goedicke‐Fritz, S., Schmiech, D., Thull, R., Kaiser, E., Körbel, C., Laschke, M. W., Marnach, A., Müller, S., Tutdibi, E., Nourkami‐Tutdibi, N., et al. (2026) · Sensors · doi:10.3390/s26072139
“Moreover, electrodes, plasters, and cables interfere with freedom of movement, particularly during “Kangaroo Care” and therapeutic positioning [14,20,21].”
InfluentialReconfigurable Spoof Plasmonic Skyrmion Electronics for Deformation‐Invariant Multimode Sensing
Wang, Z., Kong, X., Huang, X., Tian, Y., Yao, W., Xiao, X., Li, Z., & Luo, Y. (2026) · Advanced Science · doi:10.1002/advs.75071 · Cited by 2
Payne, V. L., & Partridge, B. (2026) · Health informatics · doi:10.1007/978-3-032-16657-9_12
Self-powered intelligence for personalized healthcare
Cai, Y., Kong, Z., Jin, P., Wang, X., Hou, R., Ding, N., Fu, H., Yeatman, E. M., Lee, C., Deng, F., et al. (2026) · National Science Review · doi:10.1093/nsr/nwag317 · Cited by 3
“As a result, most RFEHs are designed for epidermal deployment, such as behind the ear or on the chest, where the surface curvature is moderate and mechanical deformation is limited [77 ].”
Self-citationGarcia, G., Cheeran, C. F., Kodakkadan, F. A., et al. (2026) · IEEE International Conference on Advanced Robotics and its Social Impacts (ARSO) · doi:10.1109/arso68304.2026.11536134
“…crying and activate a rocking mechanism until crying stops [8], some of them include physiological feedback such as heart rate and temperature [9], and even environmental feedback such as humidity and gas sensors [10], but these approaches also rely on generic engineered motions rather than…”
Dong, P., Song, Y., Li, Y., Djurić, P. M., & Yao, S. (2026) · Soft Science · doi:10.20517/ss.2026.38
Photonic integrated chip for wearable physiological and biochemical sensing
Li, X., Li, H., Ding, M., Han, M., Liu, Y., Yang, L., Zhang, C. H. I., Zhang, L., Hernández-Ramírez, F., Cao, L., et al. (2026) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2026.140406 · Cited by 1
Qiang, X., Mao, M., Wang, Y., et al. (2026) · Advanced Science · doi:10.1002/advs.75745
“Various security technologies have thus been developed, including radio-frequency identification (RFID) tags [ 7 ], fluorescent and phosphorescent materials [ 8–10 ], invisible inks [ 11 ], holograms [ 12 ] , watermarks [ 13 ], and cryptographic methods [ 14–16 ], providing multi-layered solutions…”
Analytical modeling for suction cup designs for skin-interfaced wearable devices
Li, S., Rogers, J. A., & Huang, Y. Y. (2026) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.2614670123
Self-citationPressure‐Induced Drift Artifacts in Stretchable Liquid Metal ThinFilm Electrocardiogram Electrodes
Li, D., Xu, Z., Chen, Z., Xu, S.-Y., Cui, J., Wo, S., Wang, Z., Liu, Y., Yin, J.-J., Liu, H., et al. (2026) · Advanced Science · doi:10.1002/advs.76002 · Cited by 1
Self-citationMishra, M., & Ganesan, A. (2026) · Discover Electronics · doi:10.1007/s44291-026-00249-0 · Cited by 1
“…these priorities point toward a necessary shift in research emphasis– from demonstrating that a sensor can detect a target analyte under controlled conditions to demonstrating that it can do so reliably, durably, and safely in the populations and environments where it is clinically needed [1, 82].”
Kim, W., Kim, Y., Yoo, H., & Lee, E. K. (2026) · Discover Nano · doi:10.1186/s11671-026-04790-5 · Cited by 2
Choi, S., Gu, G., Kim, Y., Lee, S. H., Sim, S.-I., Jang, Y. M., & Kim, H. (2026) · medRxiv · doi:10.64898/2026.07.06.26357087
Self-citationStretchable multimodal deformation sensor with self-mode recognition by a single Hall sensor
Zhang, C., Pan, C., Xia, N., Yang, H., Su, L., Wang, X., Nie, Z., Wu, K., Zhou, L., Han, M., et al. (2026) · Nature Communications · doi:10.1038/s41467-026-75632-2 · Cited by 1
Shi, W., Mi, L., Chen, J., Liu, B., Wang, M., Dai, Z., Liu, B., Gao, R., Lin, Y., Guo, W., et al. (2026) · Advanced Science · doi:10.1002/advs.76775
Li, Y., He, Z., Chen, X., Pan, L., Yu, Y., & Liu, C. (2026) · Advanced Materials Technologies · doi:10.1002/admt.71214 · Cited by 1
Self-citationTriboelectric Tactile Sensors for Embodied Intelligence: Design, Performances, and Applications
Yu, K., Yi, J., Wang, J., Luo, Q., Liu, T., Liu, Z., Gao, F., Cui, H., Bai, Y., Liu, Y., et al. (2026) · Advanced Materials · doi:10.1002/adma.74599 · Cited by 1
Self-citationNi, J., Lu, J., Wu, N., Tang, J., Chen, T., Zhang, X., Wang, W., & Zhao, Z. (2026) · Rare Metals · doi:10.1002/rar2.70499 · Cited by 1
Conformal mechanics of serpentine meshes for millimeter-scale neural probes
Batay, S., Zhou, Y., Pathirana, P. V., Seymour, J. P., & Avila, R. O. (2026) · Mechanics of Materials · doi:10.1016/j.mechmat.2026.105860
Bridging IC chips and deformable circuits: A packaging framework for high-density soft electronics
Lee, Y., Lee, Y., Park, S., Ham, H., Jang, H., Jung, Y., Lee, S., & Seo, J. (2026) · Materials Science and Engineering R Reports · doi:10.1016/j.mser.2026.101285
Self-citationWei, Y., Wang, T., Yan, J., Bao, Y., Ma, Z., Ma, L., Guo, X., & Zhao, J. (2026) · Advanced Materials Technologies · doi:10.1002/admt.71303
Lu, Y., Liang, S., Niu, J., Jie, C., Zhang, S., Zhang, H., Jia, D., & Huang, W. (2026) · Advanced Science · doi:10.1002/advs.77522
Bioinspired Low‐Power Mechanical Sensing Technologies
Zhang, X., Zhang, X., Huang, X., Li, B., Zhang, C., Meng, X., Chen, G., Chen, Y., Zhang, J., Niu, S., et al. (2026) · Advanced Materials · doi:10.1002/adma.75090
Xu, Y., Lv, N., Zhang, S., et al. (2026) · Surfaces and Interfaces · doi:10.1016/j.surfin.2026.110744
Self-citationEngineering magnetic soft materials for biomedical applications
Liu, X., Xia, N., Chan, K. F., Sung, J. J. �., Chiu, P. W. Y., Wang, L., & Zhang, L. (2026) · Nature Reviews Materials · doi:10.1038/s41578-026-00963-x
Majumdar, S., Zaheen, A., Chakraborty, P., Rahman, M., Abir, M. R. A. R., Tabassum, S., Choi, C., & Karim, M. R. (2026) · Materials Today Chemistry · doi:10.1016/j.mtchem.2026.104077
Ultralow-Drift Perovskite Detector for Multiple Vital Sign Fusion Monitoring
Chen, Y., Xu, H., Dai, Z., Huang, L., Zhao, Y., Xie, S., Xu, H., Li, Y., Qu, D., Hu, M., et al. (2026) · Nano Letters · doi:10.1021/acs.nanolett.6c03360
Self-citation
Metadata
- Authors
- Chung, H. U., Kim, B. H., Lee, J. Y., Lee, J., Xie, Z., Ibler, E. M., Lee, K., Banks, A., Jeong, J. Y., Kim, J., Ogle, C., Grande, D., Yu, Y., Jang, H., Assem, P., Ryu, D., Kwak, J. W., Namkoong, M., Park, J. B., Lee, Y., Kim, D. H., Ryu, A., Jeong, J., You, K., Ji, B., Liu, Z., Huo, Q., Feng, X., Deng, Y., Xu, Y., Jang, K.-I., Kim, J., Zhang, Y., Ghaffari, R., Rand, C. M., Schau, M., Hamvas, A., Weese-Mayer, D. E., Huang, Y., Lee, S. M., Lee, C. H., Shanbhag, N. R., Paller, A. S., Xu, S., & Rogers, J. A.
- Affiliations
- Northwestern University (US)
- University of Illinois Urbana-Champaign (US)
- Department of Materials Science and Engineering, Northwestern University, Evanston, IL 60208, USA.
- Agency for Science, Technology and Research (SG)
- Tsinghua University (CN)
- Daegu Gyeongbuk Institute of Science and Technology (KR)
- Lurie Children's Hospital (US)
- Center for Autonomic Medicine, Department of Pediatrics, Ann & Robert H. Lurie Children’s Hospital of Chicago, Chicago, IL 60611, USA.
- Kookmin University (KR)
- Purdue University West Lafayette (US)
- Corresponding author
- Chung, Ha Uk
- Kim, Bong Hoon
- Lee, Jong Yoon
- Lee, Jungyup
- Xie, Zhaoqian
- Paller, Amy S.
- Xu, Shuai
- Rogers, John A.
- ORCID
- Chung, Ha Uk
- Kim, Bong Hoon
- Lee, Jong Yoon
- Lee, Jungyup
- Xie, Zhaoqian
- Lee, KunHyuck
- Banks, Anthony
- Jeong, Ji Yoon
- Kim, Jongwon
- Grande, Dominic
- Yu, Yongjoon
- Jang, Hokyung
- Assem, Pourya
- Kwak, Jean Won
- Namkoong, Myeong
- Park, Jun Bin
- Lee, Yechan
- Kim, Do Hoon
- Ryu, Arin
- You, Kevin
- Ji, Bowen
- Liu, Zhuangjian
- Huo, Qingze
- Feng, Xue
- Deng, Yujun
- Xu, Yeshou
- Jang, Kyung-In
- Kim, Jeonghyun
- Zhang, Yihui
- Ghaffari, Roozbeh
- Rand, Casey M.
- Hamvas, Aaron
- Huang, Yonggang
- Lee, Seung Min
- Lee, Chi Hwan
- Shanbhag, Naresh R.
- Paller, Amy S.
- Xu, Shuai
- Rogers, John A.
- Venue
- Science
- Published
- 2019-03
- Volume
- 363
- Issue
- 6430
- ISSN
- 0036-8075 (ISSN-L)
- 1095-9203
- PMC
- PMC6510306
- OpenAlex
- W2916533817
- Semantic Scholar
- b0511afac0cc252946aa8f08a078b12e48030581
- Cited by
- 807
- References
- 43
- Citations by year
- 2026: 72 · 2025: 112 · 2024: 103 · 2023: 121 · 2022: 124 · 2021: 131 · 2020: 97 · 2019: 47
- FWCI
- 35.10
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by
- Locations
- Science (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Funders
- Bill and Melinda Gates Foundation (OPP1182909)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Physical Sciences › Engineering › Biomedical Engineering › Wireless Body Area Networks
- Keywords
- epidermal electronics, neonatal intensive care, battery-free operation, neonatal monitoring
- MeSH terms
- Diagnostic Imaging, Electronics, Equipment Design, Humans, Infant, Newborn, Monitoring, Physiologic, Skin, Intensive Care, Neonatal, Vital Signs, Lab-On-A-Chip Devices, Wireless Technology
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.