A brain-computer interface that evokes tactile sensations improves robotic arm control
Open accessPublished version of a preprint
The DOI matches its Crossref record; 51 of the 51 references it cites were found in a registry or an index; 5 of the 15 sources its claims cite were checked for support.
Abstract
Prosthetic arms controlled by a brain-computer interface can enable people with tetraplegia to perform functional movements. However, vision provides limited feedback because information about grasping objects is best relayed through tactile feedback. We supplemented vision with tactile percepts evoked using a bidirectional brain-computer interface that records neural activity from the motor cortex and generates tactile sensations through intracortical microstimulation of the somatosensory cortex. This enabled a person with tetraplegia to substantially improve performance with a robotic limb; trial times on a clinical upper-limb assessment were reduced by half, from a median time of 20.9 to 10.2 seconds. Faster times were primarily due to less time spent attempting to grasp objects, revealing that mimicking known biological control principles results in task performance that is closer to able-bodied human abilities.
Overview
CoChat Summary
TL;DR
A bidirectional brain-computer interface that decoded motor intent and delivered intracortical microstimulation to somatosensory cortex improved a person with tetraplegia’s control of a robotic arm. Adding tactile feedback reduced median ARAT trial time from 20.9 to 10.2 s and improved median ARAT score from 17 to 21.
Key findings
- With ICMS-evoked tactile feedback, median ARAT score was 21 versus 17 without ICMS (p = 0.029, Wilcoxon rank-sum test), and higher than 23 prior sessions without ICMS (p = 0.005).
- Successful ARAT trial times decreased by 51.2%, from a median of 20.9 s to 10.2 s (p < 0.0001); grasp time decreased by 66%, from 13.3 s to 4.6 s, accounting for 88% of the total improvement.
- A score of 3, indicating object transfer in under 5 s, was achieved 15 times with tactile feedback versus once across all 108 trials without it; 14% of trials with tactile feedback were faster than the fastest trial without ICMS.
- In the object transfer task, time in the object zone decreased by 30.3%, from 3.3 ± 1.2 s to 2.3 ± 0.4 s per transfer (p = 0.002), and hand movement in the object zone decreased from 44.2 ± 13.1 to 32.4 ± 5.9 cm per transfer (p = 0.0007); total transfers increased from 315 without ICMS to 352 with ICMS, but transfers per trial did not significantly differ (15.8 ± 3.8 to 17.8 ± 2.4, p = 0.050).
Methods
- Single participant: a 28-year-old male with tetraplegia due to C5 motor/C6 sensory ASIA B spinal cord injury sustained 10 years before consent; study conducted under FDA Investigational Device Exemption and registered at ClinicalTrials.gov (NCT01894802).
- Two 88-electrode arrays were implanted in motor cortex hand/arm region to decode movement intent, and two 32-electrode arrays were implanted in area 1 of somatosensory cortex; a new 5 degree-of-freedom velocity decoder was trained daily, and ICMS was driven in real time by robotic hand sensors responding to object contact and grasp force.
- Performance was compared with and without ICMS on a modified ARAT with nine tasks, each object attempted three times per session and scored 0–3 with maximum score 27, using four sessions with ICMS and four without plus 23 prior sessions; an object transfer task was also performed in four sessions with and four without ICMS, five times per session, with Wilcoxon rank-sum and t-tests used for comparisons.
Limitations
- The paper states that this is a single-subject study and it is not guaranteed that findings will generalize to future experiments and participants.
- The paper states that future work should examine how stimulation encoding design, such as proportional versus on-off feedback, affects performance, and notes that many objects were rigid with no penalty for grasping too firmly, so fragile objects or more precise hand posture/grasp force control could depend more on specific sensory encoding schemes.
- Evident from the text: the participant had only one grasp dimension that opened and closed all fingers as a group, and some smaller objects were difficult to pick up because the fingertips often contacted objects rather than the distal phalanx pads, making grasp less stable; the increase in transfers per trial with ICMS was not statistically significant (p = 0.050).
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
3 Valid · 0 Not valid · 7 Not checked · 32 Own results and interpretations
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“First, using the same fundamental neural decoding and control methods, two participants have achieved similar scores on functional tasks with vision alone (5, 6, 28).”
§ A brain-computer interface that evokes tactile sensations improves robotic arm control > Abstract
What the sources say
- Partly supportedcites [5] Wodlinger B, J. Neural Eng 12, 016011 (2015). 25514320 sourceExperimental — accuracy not yet measured
Section 3.4 reports similar functional-task scores under 7D and 10D control, but both scores are from the same participant. Section 2.2 identifies a single participant; this paper therefore does not establish the claim about two participants achieving similar scores with vision alone. Page numbers are unavailable in our copy. (LLM, conf=0.88)
- Partly supportedcites [6] Collinger JL, Lancet (London, England) 381, 557– 64 (2013). PMC3641862 23253623 sourceExperimental — accuracy not yet measured
The Methods (Observation-based calibration and neural decoding; Functional brain-control tasks) describe neural decoding and functional testing, and the Results (Clinical assessments) report an ARAT score of 15–17 for one participant. This paper does not establish that two participants achieved similar scores, and the retrieved passages do not explicitly say the tasks were performed with vision alone. (LLM, conf=0.84)
- Partly supportedcites [28] Downey JE, Sci. Rep 7, 16947 (2017). 29209023 PMC5717217 sourceExperimental — accuracy not yet measured
The Methods (Study participants; Neural recording and BCI decoder calibration, page unavailable) describe two participants using motor-cortex decoding to control a robotic arm; somatosensory arrays were not used in this experiment. The Results (Online population-level scaling improves BCI object grasping, page unavailable) report fairly close object-interaction success rates with scaling, but the more functional object-transport scores differ substantially: 1.09 versus 5.28 transfers per minute. Thus, the paper does not support a general claim that the participants achieved similar scores on functional tasks. (LLM, conf=0.86)
plainn = 2- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
Every source the paper cites for this claim was checked and supports it.
In the paper
“In contrast, ICMS feedback in a relevant area of cortex, but in an unintuitive way, requires considerable learning time (thousands of trials) to use effectively (33).”
§ A brain-computer interface that evokes tactile sensations improves robotic arm control > Abstract
What the sources say
- Supportedcites [33] Dadarlat MC, O’Doherty JE, Sabes PN, Nat. Neurosci 18, 138– 44 (2015). 25420067 PMC4282864 source
The Results (Behavioral task and feedback signals) describe ICMS in somatosensory cortex, with electrode directions assigned independently of local neuronal responses. The Discussion (Learning to integrate a novel sensory signal) says the patterns did not mimic naturally occurring signals and had to be learned. Online Methods (Experimental Design) reports approximately 20,000 and 40,000 training trials before the two monkeys showed clear evidence of integrating ICMS with vision. (LLM, conf=0.94)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
The sources with a verdict support this claim; the rest weren’t checked yet, have no full text we could read, or their verdicts aren’t rated yet.
In the paper
“it is only in recent studies of humans with chronically implanted microelectrode arrays that the perceptual characteristics of microstimulation have been elucidated (19, 20).”
§ A brain-computer interface that evokes tactile sensations improves robotic arm control > Abstract
What the sources say
- Not checked — no full text of the sourcecites [19] Flesher SN, Sci. Transl. Med 8, 361ra141 (2016). 27738096 source
- Supportedcites [20] Salas MA, Elife 7 (2018), doi: 10.7554/eLife.32904. source
The Abstract describes recent human ICMS studies and reports sensations elicited in a person with two implanted microelectrode arrays. Materials and methods (Task) describes how the participant reported each percept’s location, intensity, duration, and qualitative character; Results and discussion reports the resulting percept measurements. These passages support the claim that such human studies elucidated the perceptual characteristics of microstimulation. (LLM, conf=0.87)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Further, 14% of the trials with tactile feedback were completed more quickly than the very fastest trial without ICMS (Fig. 2C).”
§ A brain-computer interface that evokes tactile sensations improves robotic arm control > Abstract
plainSingle case- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The time spent in the object zone (Fig. 1G) decreased by 30.3%, from 3.3 ± 1.2 s per transfer without ICMS to 2.3 ± 0.4 s per transfer with ICMS (p = 0.002, t-test, Fig. 3A).”
§ A brain-computer interface that evokes tactile sensations improves robotic arm control > Abstract
plainstates a causeSingle case ·- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
19 more claims
Create a free account to see all 42 claims and every source verdict.
- It may state two separate claims as one. 15
- Probably not a substantial claim. 3
Citations
Cites
PENFIELD, W., & BOLDREY, E. (1937) · Brain · doi:10.1093/brain/60.4.389 · Cited by 4,188
Neuronal ensemble control of prosthetic devices by a human with tetraplegia
Hochberg, L. R., Serruya, M. D., Friehs, G. M., Mukand, J. A., Saleh, M., Caplan, A. H., Branner, A., Chen, D., Penn, R. D., & Donoghue, J. P. (2006) · Nature · doi:10.1038/nature04970 · Cited by 3,295
Reach and grasp by people with tetraplegia using a neurally controlled robotic arm
Hochberg, L. R., Bacher, D., Jarosiewicz, B., Masse, N. Y., Simeral, J. D., Vogel, J., Haddadin, S., Liu, J., Cash, S. S., van der Smagt, P., et al. (2012) · Nature · doi:10.1038/nature11076 · Cited by 2,719
Generalized Inverses, Ridge Regression, Biased Linear Estimation, and Nonlinear Estimation
Marquardt, D. W. (1970) · Technometrics · doi:10.1080/00401706.1970.10488699 · Cited by 2,276
Targeting Recovery: Priorities of the Spinal Cord-Injured Population
Anderson, K. D. (2004) · Journal of Neurotrauma · doi:10.1089/neu.2004.21.1371 · Cited by 2,094
Coding and use of tactile signals from the fingertips in object manipulation tasks
Johansson, R. S., & Flanagan, J. R. (2009) · Nature Reviews Neuroscience · doi:10.1038/nrn2621 · Cited by 2,063
High-performance neuroprosthetic control by an individual with tetraplegia
Collinger, J. L., Wodlinger, B., Downey, J. E., Wang, W., Tyler-Kabara, E. C., Weber, D. J., McMorland, A. J., Velliste, M., Boninger, M. L., & Schwartz, A. B. (2013) · The Lancet · doi:10.1016/s0140-6736(12)61816-9 · Cited by 1,848
Self-citationCortical control of a prosthetic arm for self-feeding
Velliste, M., Perel, S., Spalding, M. C., Whitford, A. S., & Schwartz, A. B. (2008) · Nature · doi:10.1038/nature06996 · Cited by 1,591
Lyle, R. C. (1981) · International Journal of Rehabilitation Research · doi:10.1097/00004356-198112000-00001 · Cited by 1,326
Restoring natural sensory feedback in real-time bidirectional hand prostheses
Raspopovic, S., Capogrosso, M., Petrini, F. M., Bonizzato, M., Rigosa, J., Di Pino, G., Carpaneto, J., Controzzi, M., Boretius, T., Fernandez, E., et al. (2014) · Science Translational Medicine · doi:10.1126/scitranslmed.3006820 · Cited by 1,001
Research works printed but not linked: 2
- J. M. Loomis R. L. Klatzky N. A. Giudice in Assistive Technology for Blindness and Low Vision R. Manduchi S. Kurniawan Eds. (CRC Press 2012).
- M. S. Johannes, J. D. Bigelow, J. M. Burck, S. D. Harshbarger, M. V. Kozlowski, T. Van Doren, An overview of the developmental process for the modular prosthetic limb. Johns Hopkins APL Tech. Dig. 30, 207–216 (2011).
Cited by
An Integrated Brain-Machine Interface Platform With Thousands of Channels
Musk, E., & Neuralink (2019) · Journal of Medical Internet Research · doi:10.2196/16194 · Cited by 1,206
Neuromorphic sensorimotor loop embodied by monolithically integrated, low-voltage, soft e-skin
Wang, W., Jiang, Y., Zhong, D., Zhang, Z., Choudhury, S., Lai, J.-C., Gong, H., Niu, S., Yan, X., Zheng, Y., et al. (2023) · Science · doi:10.1126/science.ade0086 · Cited by 788
Flexible brain–computer interfaces
Tang, X., Shen, H., Zhao, S., Li, N., & Liu, J. (2023) · Nature Electronics · doi:10.1038/s41928-022-00913-9 · Cited by 373
“Flexible BCIs create new opportunities for the integration of natural intelligence and artificial intelligence, with potential applications such as self-programmable and long-term stable real-time brain decoding with bio-inspired machine learning algorithms.”
BackgroundMethodRestoration of sensory information via bionic hands
Bensmaia, S. J., Tyler, D. J., & Micera, S. (2020) · Nature Biomedical Engineering · doi:10.1038/s41551-020-00630-8 · Cited by 257
Recent advances in regenerative biomaterials
Cao, D., & Ding, J. (2022) · Regenerative Biomaterials · doi:10.1093/rb/rbac098 · Cited by 259
The neural mechanisms of manual dexterity
Sobinov, A. R., & Bensmaia, S. J. (2021) · Nature Reviews Neuroscience · doi:10.1038/s41583-021-00528-7 · Cited by 248
An Accurate and Rapidly Calibrating Speech Neuroprosthesis
Card, N. S., Wairagkar, M., Iacobacci, C., Hou, X., Singer-Clark, T., Willett, F. R., Kunz, E. M., Fan, C., Vahdati Nia, M., Deo, D. R., et al. (2024) · New England Journal of Medicine · doi:10.1056/nejmoa2314132 · Cited by 268
Non-Invasive Brain-Computer Interfaces: State of the Art and Trends
Edelman, B. J., Zhang, S., Schalk, G., Brunner, P., Müller-Putz, G., Guan, C., & He, B. (2025) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2024.3449790 · Cited by 170
“…how sensory feedback, whether provided to the peripheral (i.e., haptic feedback) [23], [24] or central (i.e., intracortical micro-stimulation [25], transcranial electrical stimulation [26], or focused ultrasound stimulation [27]) nervous system, impacts the performance of these systems for…”
Brain–computer interfaces: the innovative key to unlocking neurological conditions
Zhang, H., Jiao, L., Yang, S., Li, H., Jiang, X., Feng, J., Zou, S., Xu, Q., Gu, J., Wang, X., et al. (2024) · International Journal of Surgery · doi:10.1097/js9.0000000000002022 · Cited by 169
“Robotics in BCI applications provide physical feedback through devices like robotic arms controlled by imagined movements, enhancing tactile feedback and facilitating system control learning [51 – 53] .”
An Integrated Brain-Machine Interface Platform With Thousands of ChannelsPreprint
Musk, E., & Neuralink (2019) · bioRxiv · doi:10.1101/703801 · Cited by 1,036
Brain–computer interface: trend, challenges, and threats
Maiseli, B., Abdalla, A. T., Massawe, L. V., Mbise, M., Mkocha, K., Nassor, N. A., Ismail, M., Michael, J., & Kimambo, S. (2023) · Brain Informatics · doi:10.1186/s40708-023-00199-3 · Cited by 168
“Feature translation generates commands that can con-trol external devices (BCI applications): cursor [57–60] for letter and text selection on the computer screen [44, 45, 61], wheelchair [62, 63], and robotic arm [64, 65].”
BackgroundSoft Fiber Electronics Based on Semiconducting Polymer
Sun, F., Jiang, H., Wang, H., Zhong, Y., Xu, Y., Xing, Y., Yu, M., Feng, L.-W., Tang, Z., Liu, J., et al. (2023) · Chemical Reviews · doi:10.1021/acs.chemrev.2c00720 · Cited by 164
Chen, S., Fan, S., Qiao, Z., Wu, Z., Lin, B., Li, Z., Riegler, M. A., Wong, M. Y. H., Opheim, A., Korostynska, O., et al. (2025) · Advanced Materials · doi:10.1002/adma.202500412 · Cited by 164
Bioelectronics for electrical stimulation: materials, devices and biomedical applications
Huang, Y., Yao, K., Zhang, Q., Huang, X., Chen, Z., Zhou, Y., & Yu, X. (2024) · Chemical Society Reviews · doi:10.1039/d4cs00413b · Cited by 180
Restoring the Sense of Touch Using a Sensorimotor Demultiplexing Neural Interface
Ganzer, P. D., Colachis, S. C., Schwemmer, M. A., Friedenberg, D. A., Dunlap, C. F., Swiftney, C. E., Jacobowitz, A. F., Weber, D. J., Bockbrader, M. A., & Sharma, G. (2020) · Cell · doi:10.1016/j.cell.2020.03.054 · Cited by 156
A Skin-Inspired Artificial Mechanoreceptor for Tactile Enhancement and Integration
Li, F., Wang, R., Song, C., Zhao, M., Ren, H., Wang, S., Liang, K., Li, D., Ma, X., Zhu, B., et al. (2021) · ACS Nano · doi:10.1021/acsnano.1c05836 · Cited by 146
Xu, Z., Zhou, F., Yan, H., Gao, G., Li, H., Li, R., & Chen, T. (2021) · Nano Energy · doi:10.1016/j.nanoen.2021.106614 · Cited by 150
Recent advances in multi-mode haptic feedback technologies towards wearable interfaces
Huang, Y., Yao, K., Li, J., Li, D., Jia, H., Liu, Y., Yiu, C. K., Park, W., & Yu, X. (2021) · Materials Today Physics · doi:10.1016/j.mtphys.2021.100602 · Cited by 157
Bio-inspired electronics: Soft, biohybrid, and “living” neural interfaces
Boufidis, D., Garg, R., Angelopoulos, E., Cullen, D. K., & Vitale, F. (2025) · Nature Communications · doi:10.1038/s41467-025-57016-0 · Cited by 199
“…20 , responsive neurostimulation 21,22 , speech decoding 23 – 25 , and motor recovery following spinal cord injury 26 , closed-loop sensory-motor interfaces for prosthetic control 27,28 , as well as deep brain stimulation (DBS) for Parkinson ’ s Disease 29,30 and neuropsychiatric conditions 14,31 .”
Conformal in-ear bioelectronics for visual and auditory brain-computer interfaces
Wang, Z., Shi, N., Zhang, Y., Zheng, N., Li, H., Jiao, Y., Cheng, J., Wang, Y., Zhang, X., Chen, Y., et al. (2023) · Nature Communications · doi:10.1038/s41467-023-39814-6 · Cited by 155
“2e, Supplementary Movies 4 and 5.”
InfluentialMiniature battery-free bioelectronics
Nair, V., Dalrymple, A. N., Yu, Z., Balakrishnan, G., Bettinger, C. J., Weber, D. J., Yang, K., & Robinson, J. T. (2023) · Science · doi:10.1126/science.abn4732 · Cited by 128
“Al-ternatively, neural recordings from electrodes implanted into the motor cortex can be de-coded to control a robotic limb, and sensory feedback can be provided by electrodes implanted in the somatosensory cortex ( 7 ).”
BackgroundNeurosurgery and artificial intelligence
Mofatteh, M. (2021) · AIMS neuroscience · doi:10.3934/neuroscience.2021025 · Cited by 132
“A recent bidirectional BCI, which provides patients with tetraplegia with substantially improved control over the robotic arm by recording neural activity from the cortex and generating tactile sensation through intracortical stimulation of the somatosensory cortex [102], can have a substantial impact on the advanceme…”
BackgroundMXene-Enabled Self-Adaptive Hydrogel Interface for Active Electroencephalogram Interactions
Luo, J., Sun, C., Chang, B., Jing, Y., Li, K., Li, Y., Zhang, Q., Wang, H., & Hou, C. (2022) · ACS Nano · doi:10.1021/acsnano.2c08961 · Cited by 115
Current Solutions and Future Trends for Robotic Prosthetic Hands
Mendez, V., Iberite, F., Shokur, S., & Micera, S. (2021) · Annual Review of Control Robotics and Autonomous Systems · doi:10.1146/annurev-control-071020-104336 · Cited by 126
Soft Electronics for Health Monitoring Assisted by Machine Learning
Qiao, Y., Luo, J., Cui, T., Liu, H., Tang, H., Zeng, Y.-F., Liu, C., Li, Y., Jian, J., Wu, J., et al. (2023) · Nano-Micro Letters · doi:10.1007/s40820-023-01029-1 · Cited by 129
The state of clinical trials of implantable brain–computer interfaces
Patrick-Krueger, K. M., Burkhart, I., & Contreras-Vidal, J. L. (2024) · Nature Reviews Bioengineering · doi:10.1038/s44222-024-00239-5 · Cited by 125
Against cortical reorganisation
Makin, T. R., & Krakauer, J. W. (2023) · eLife · doi:10.7554/elife.84716 · Cited by 144
Nanomaterials for Flexible Neuromorphics
Ding, G., Li, H., Zhao, J., Zhou, K., Zhai, Y., Lv, Z., Zhang, M., Yan, Y., Han, S., & Zhou, Y. (2024) · Chemical Reviews · doi:10.1021/acs.chemrev.4c00369 · Cited by 113
Fifer, M. S., McMullen, D. P., Osborn, L. E., Thomas, T. M., Christie, B., Nickl, R. W., Candrea, D. N., Pohlmeyer, E. A., Thompson, M. C., Anaya, M. A., et al. (2022) · Neurology · doi:10.1212/wnl.0000000000013173 · Cited by 103
Brain–Computer Interface-Controlled Exoskeletons in Clinical Neurorehabilitation: Ready or Not?
Colucci, A., Vermehren, M., Cavallo, A., Angerhöfer, C., Peekhaus, N., Zollo, L., Kim, W., Paik, N., & Soekadar, S. R. (2022) · Neurorehabilitation and neural repair · doi:10.1177/15459683221138751 · Cited by 124
“53 The main risks opposing successful adoption are (1) unmet user expectations, for example, because the system’s robustness or assistive function are insufficient, (2) lower effect size in motor rehabilitation than anticipated, (3) lack of availability across the whole rehabilitation process (particularly, when trans…”
BackgroundNeuroprosthetics: from sensorimotor to cognitive disorders
Gupta, A., Vardalakis, N., & Wagner, F. B. (2023) · Communications Biology · doi:10.1038/s42003-022-04390-w · Cited by 120
Greenspon, C. M., Valle, G., Shelchkova, N. D., Hobbs, T. G., Verbaarschot, C., Callier, T., Berger-Wolf, E. I., Okorokova, E. V., Hutchison, B. C., Dogruoz, E., et al. (2024) · Nature Biomedical Engineering · doi:10.1038/s41551-024-01299-z · Cited by 112
Self-citationPerception of microstimulation frequency in human somatosensory cortex
Hughes, C. L., Flesher, S. N., Weiss, J. M., Boninger, M. L., Collinger, J. L., & Gaunt, R. A. (2021) · eLife · doi:10.7554/elife.65128 · Cited by 99
“BCI users can control an end effector using neural activity recorded from motor cortex and receive sensory feedback through intracortical microstimulation (ICMS) in somato-sensory cortex ( Flesher et al., 2021 ).”
Self-citationA memristor-based adaptive neuromorphic decoder for brain–computer interfaces
Liu, Z., Mei, J., Tang, J., Xu, M., Gao, B., Wang, K., Ding, S., Liu, Q., Qin, Q., Chen, W., et al. (2025) · Nature Electronics · doi:10.1038/s41928-025-01340-2 · Cited by 138
Explant Analysis of Utah Electrode Arrays Implanted in Human Cortex for Brain-Computer-Interfaces
Woeppel, K., Hughes, C., Herrera, A. J., Eles, J. R., Tyler-Kabara, E. C., Gaunt, R. A., Collinger, J. L., & Cui, X. T. (2021) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2021.759711 · Cited by 90
Self-citationNeural interfaces: Bridging the brain to the world beyond healthcare
Xu, S., Liu, Y., Lee, H., & Li, W. (2024) · Exploration · doi:10.1002/exp.20230146 · Cited by 116
Cognitive neuroscience and robotics: Advancements and future research directions
Liu, S., Wang, L., & Gao, R. X. (2023) · Robotics and Computer-Integrated Manufacturing · doi:10.1016/j.rcim.2023.102610 · Cited by 96
Interim Safety Profile From the Feasibility Study of the BrainGate Neural Interface System
Rubin, D. B., Ajiboye, A. B., Barefoot, L., Bowker, M., Cash, S. S., Chen, D., Donoghue, J. P., Eskandar, E. N., Friehs, G., Grant, C., et al. (2023) · Neurology · doi:10.1212/wnl.0000000000201707 · Cited by 81
Neural Decoding for Intracortical Brain–Computer Interfaces
Dong, Y., Wang, S., Huang, Q., Berg, R. W., Li, G., & He, J. (2023) · Cyborg and Bionic Systems · doi:10.34133/cbsystems.0044 · Cited by 90
“Evoked tactile signals effectively reduced the time to attempt to grasp objects, reveal ing that mimicking known biological control principles leads to task performance that more closely resembles ablebodied human abilities [8].”
BackgroundThe science and engineering behind sensitized brain-controlled bionic hands
Pandarinath, C., & Bensmaïa, S. J. (2021) · Physiological Reviews · doi:10.1152/physrev.00034.2020 · Cited by 86
Willsey, M. S., Shah, N. P., Avansino, D. T., Hahn, N. V., Jamiolkowski, R. M., Kamdar, F. B., Hochberg, L. R., Willett, F. R., & Henderson, J. M. (2025) · Nature Medicine · doi:10.1038/s41591-024-03341-8 · Cited by 116
Modulating Brain Activity with Invasive Brain–Computer Interface: A Narrative Review
Zhao, Z.-P., Nie, C., Jiang, C.-T., Cao, S.-H., Tian, K., Yu, S., & Gu, J.-W. (2023) · Brain Science · doi:10.3390/brainsci13010134 · Cited by 77
“…evoked using a bidirectional BCI that records neural activity from the motor cortex and generates tactile sensations through intracortical microstimulation of the somatosensory cortex, en ‐ abling subjects with tetraplegia to substantially improve performance with a robotic limb [73]. on a 4.2.”
BackgroundMethodInfluentialDecoding the brain: From neural representations to mechanistic models
Mathis, M. W., Rotondo, A. P., Chang, E. F., Tolias, A. S., & Mathis, A. (2024) · Cell · doi:10.1016/j.cell.2024.08.051 · Cited by 128
EEG-based brain-computer interface enables real-time robotic hand control at individual finger level
Ding, Y., Udompanyawit, C., Zhang, Y., & He, B. (2025) · Nature Communications · doi:10.1038/s41467-025-61064-x · Cited by 138
Going beyond primary motor cortex to improve brain–computer interfaces
Gallego, J. Á., Makin, T. R., & McDougle, S. D. (2022) · Trends in Neurosciences · doi:10.1016/j.tins.2021.12.006 · Cited by 78
“’ Yet despite the impressive advances in BCI technology over the past several years [2,6,7], even seemingly simple tasks such as controlling a computer cursor with a BCI turn out to be challenging, often more so than performing analogous tasks with a limb [8].”
BackgroundAn Ultra‐Thin MXene Film for Multimodal Sensing of Neuroelectrical Signals with Artifacts Removal
Song, D., Li, X., Jang, M., Lee, Y., Zhai, Y., Hu, W., Yan, H., Zhang, S., Chen, L., Lu, C., et al. (2023) · Advanced Materials · doi:10.1002/adma.202304956 · Cited by 67
Valle, G., Katic, N., Eggemann, D., Gorskii, O., Pavlova, N., Petrini, F. M., Čvančara, P., Stieglitz, T., Евгеньевич, М. П., Bumbaširević, M., et al. (2024) · Nature Communications · doi:10.1038/s41467-024-45190-6 · Cited by 83
“The presented neuromodulation framework based on biomimetic encoding could also be very relevant for other neuroprostheses in the CNS (e.g., Deep Brain Stimulation 83 , epidural stimulation 13 , ICMS 5,18 ) and for bioelectronic medicine applications (e.g., vagus stimulation 84 , stimulation of the autonomic nervous s…”
BackgroundInfluentialKhan, A. A., Laghari, A. A., Shaikh, A. A., Dootio, M. A., Estrela, V. V., & Lopes, R. T. (2021) · Neuroscience Informatics · doi:10.1016/j.neuri.2021.100030 · Cited by 66
Gao, Z., Pang, Z., Chen, Y., Lei, G., Zhu, S., Li, G., Shen, Y., & Xu, W. (2022) · Neuroscience Bulletin · doi:10.1007/s12264-022-00959-x · Cited by 83
“Recent studies have reported a novel bidirectional BCI that records neural activity from the motor cortex and generates sensations by intracortical microsimulation of the sensory cortex [192, 199].”
BackgroundShelchkova, N. D., Downey, J. E., Greenspon, C. M., Okorokova, E. V., Sobinov, A. R., Verbaarschot, C., He, Q., Collinger, J. L., Tortolani, A. F., Moore, D. D., et al. (2023) · Nature Communications · doi:10.1016/j.brs.2023.01.203 · Cited by 77
“These sensations can be used to pro-videtactilefeedbackaboutobjectinteractionsandhavebeenshownto improve the functionality of a brain-controlled robotic hand 15 .”
MethodSelf-citationGallego-Carracedo, C., Perich, M. G., Chowdhury, R. H., Miller, L. E., & Gallego, J. Á. (2022) · eLife · doi:10.7554/elife.73155 · Cited by 75
Rochford, A. E., Carnicer‐Lombarte, A., Kawan, M., Jin, A. T., Hilton, S., Curto, V. F., Rutz, A. L., Moreau, T., Kotter, M., Malliaras, G. G., et al. (2023) · Science Advances · doi:10.1126/sciadv.add8162 · Cited by 69
“Some recent systems tested in human patients make use of brain recording interfaces to acquire and interpret motor signals from the patient cortex and drive a robotic prosthesis (22, 23).”
BackgroundKosnoff, J., Yu, K., Liu, C., & He, B. (2024) · Nature Communications · doi:10.1038/s41467-024-48576-8 · Cited by 70
“The N200 theta, alpha, beta, and low gamma powers were extracted by taking the [4, 8), [8, 12), [12, 30), or [30, 40) Hz frequency power response, respectively, in the100to250mspost-stimuluswindow.”
Fifer, M. S., McMullen, D. P., Osborn, L. E., Thomas, T. M., Christie, B. P., Nickl, R. W., Candrea, D. N., Pohlmeyer, E. A., Thompson, M. C., Anaya, M. A., et al. (2022) · Utrecht University Repository (Utrecht University) · Cited by 60
Shared Three-Dimensional Robotic Arm Control Based on Asynchronous BCI and Computer Vision
Zhou, Y., Yu, T., Gao, W., Huang, W., Lu, Z., Huang, Q., & Li, Y. (2023) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2023.3299350 · Cited by 62
“Since then, numerous types of studies have successfully converted animal or human motor cortical activities into intentional movements of a mechanized arm [12], [13], [14], [15].”
BackgroundMa, H., Liang, J., Qiu, J. R., Li, J., Ma, L., Sheng, H., Shao, M., Wang, Q., Li, F., Fu, Y., et al. (2023) · Advanced Materials · doi:10.1002/adma.202301218 · Cited by 61
Tactile edges and motion via patterned microstimulation of the human somatosensory cortex
Valle, G., Alamri, A. H., Downey, J. E., Lienkämper, R., Jordan, P. M., Sobinov, A. R., Endsley, L. J., Prasad, D., Boninger, M. L., Collinger, J. L., et al. (2025) · Science · doi:10.1126/science.adq5978 · Cited by 66
Self-citationThe frequency of cortical microstimulation shapes artificial touch
Callier, T., Brantly, N. W., Caravelli, A., & Bensmaïa, S. J. (2019) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1916453117 · Cited by 62
Shared Control of Bimanual Robotic Limbs With a Brain-Machine Interface for Self-Feeding
Handelman, D. A., Osborn, L. E., Thomas, T. M., Badger, A. R., Thompson, M. C., Nickl, R. W., Anaya, M. A., Wormley, J. M., Cantarero, G. L., McMullen, D. P., et al. (2022) · Frontiers in Neurorobotics · doi:10.3389/fnbot.2022.918001 · Cited by 62
“, 2022), controlling external robotic limbs with closedloop sensory feedback (Flesher et al., 2021), or combined with other assistive technologies (McMullen et al.”
BackgroundA Review: Research Progress of Neural Probes for Brain Research and Brain–Computer Interface
Luo, J., Xue, N., & Chen, J. (2022) · Biosensors · doi:10.3390/bios12121167 · Cited by 66
Zhang, J., Li, J., Huang, Z., Huang, D., Yu, H., & Li, Z. (2023) · Health Data Science · doi:10.34133/hds.0096 · Cited by 67
Sombeck, J. T., Heye, J. O., Kumaravelu, K., Goetz, S. M., Peterchev, A. V., Grill, W. M., Bensmaïa, S. J., & Miller, L. E. (2022) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac63e8 · Cited by 52
Muret, D., Root, V., Kieliba, P., Clode, D., & Makin, T. R. (2022) · Cell Reports · doi:10.1016/j.celrep.2022.110523 · Cited by 54
“Even if not functionally valuable, this information could be exploited for brain-machine interfaces (Flesher et al., 2021), where specific parts of the homunculus might not be as directly accessible (e.g., the medial foot region).”
Song, D., Zeng, M., Min, P., Jia, X., Gao, F., Yu, Z., & Li, X. (2022) · Journal of Material Science and Technology · doi:10.1016/j.jmst.2022.09.050 · Cited by 54
Advanced Electrode Technologies for Noninvasive Brain–Computer Interfaces
Lin, S., Jiang, J., Huang, K., Li, L., He, X., Du, P. Y., Wu, Y., Liu, J., Li, X., Huang, Z., et al. (2023) · ACS Nano · doi:10.1021/acsnano.3c06781 · Cited by 58
A bioinspired MXene-based flexible sensory neuron for tactile near-sensor computing
Huang, J., Feng, J., Chen, Z., Dai, Z., Yang, S., Chen, Z., Zhang, H., Zhou, Z., Zeng, Z., Li, X., et al. (2024) · Nano Energy · doi:10.1016/j.nanoen.2024.109684 · Cited by 56
Automated Pulmonary Nodule Classification and Detection Using Deep Learning Architectures
Ahmed, I., Chehri, A., Jeon, G., & Piccialli, F. (2022) · IEEE Transactions on Computational Biology and Bioinformatics · doi:10.1109/tcbb.2022.3192139 · Cited by 56
“For instance, by examining the transmission of infections and social practices in relation to environmental influences or to performing a brain-machine interface such as controlling sensations of robotic arm [8], and stroke neurorehabilitation [9].”
BackgroundNeuron devices: emerging prospects in neural interfaces and recognition
Wang, Y., Liu, S., Wang, H., Zhao, Y., & Zhang, X. (2022) · Microsystems & Nanoengineering · doi:10.1038/s41378-022-00453-4 · Cited by 56
Invasive Brain Computer Interface for Motor Restoration in Spinal Cord Injury: A Systematic Review
Levett, J. J., Elkaim, L. M., Niazi, F., Weber, M. H., Iorio‐Morin, C., Bonizzato, M., & Weil, A. G. (2023) · Neuromodulation Technology at the Neural Interface · doi:10.1016/j.neurom.2023.10.006 · Cited by 70
Stoney vs. Histed: Quantifying the spatial effects of intracortical microstimulation
Kumaravelu, K., Sombeck, J. T., Miller, L. E., Bensmaïa, S. J., & Grill, W. M. (2021) · Brain stimulation · doi:10.1016/j.brs.2021.11.015 · Cited by 56
A skin-integrated multimodal haptic interface for immersive tactile feedback
Huang, Y., Zhou, J., Ke, P., Guo, X. M., Yiu, C. K., Yao, K., Cai, S., Li, D., Zhou, Y., Li, J., et al. (2023) · Nature Electronics · doi:10.1038/s41928-023-01074-z · Cited by 72
Ten-Hour Stable Noninvasive Brain-Computer Interface Realized by Semidry Hydrogel-Based Electrodes
Liu, J., Lin, S., Li, W., Zhao, Y., Liu, D., He, Z., Wang, D., Lei, M., Hong, B., & Wu, H. (2022) · Research · doi:10.34133/2022/9830457 · Cited by 50
In vivo spatiotemporal dynamics of astrocyte reactivity following neural electrode implantation
Savya, S. P., Li, F., Lam, S., Wellman, S. M., Stieger, K. C., Chen, K., Eles, J. R., & Kozai, T. D. Y. (2022) · Biomaterials · doi:10.1016/j.biomaterials.2022.121784 · Cited by 48
Bergeron, D., Iorio‐Morin, C., Bonizzato, M., Lajoie, G., Gaucher, N. O., Racine, É., & Weil, A. G. (2023) · Journal of Child Neurology · doi:10.1177/08830738231167736 · Cited by 58
“The brain-computer interface loop is closed when the user receives feedback of this action directly through their natural senses (visual feedback of the performed action) or artificially, via neural feedback (intracortical microstimulation for tactile information of a prosthetic limb, for instance).(13) Using this fee…”
BackgroundElectrocorticogram (ECoG): Engineering Approaches and Clinical Challenges for Translational Medicine
Moon, H., Kwon, J., Eun, J., Chung, C. K., Kim, J. S., Chou, N., & Kim, S. (2024) · Advanced Materials Technologies · doi:10.1002/admt.202301692 · Cited by 58
Neuromorphic devices for electronic skin applications
Patil, C. S., Ghode, S. B., Kim, J., Kamble, G. U., Kundale, S. S., Mannan, A., Ko, Y., Noman, M. A., Saqib, Q. M., Patil, S. R., et al. (2025) · Materials Horizons · doi:10.1039/d4mh01848f · Cited by 58
Advances in materials for haptic skin electronics
Gao, Y., Yao, K., Jia, S., Huang, Y., Zhao, G., Zhang, B., Liu, Y., & Yu, X. (2024) · Matter · doi:10.1016/j.matt.2024.06.010 · Cited by 58
Saway, B. F., Palmer, C., Hughes, C. L., Triano, M. J., Suresh, R. E., Gilmore, J., George, M. S., Kautz, S. A., & Rowland, N. C. (2024) · Neurotherapeutics · doi:10.1016/j.neurot.2024.e00337 · Cited by 58
“Simulating sensory feedback has been shown to improve robotic arm control, which can lead to better rehabilitation for people with movement limitations [66].”
BackgroundSelf-citationA tactile oral pad based on carbon nanotubes for multimodal haptic interaction
Hou, B., Yang, D., Ren, X., Yi, L., & Liu, X. (2024) · Nature Electronics · doi:10.1038/s41928-024-01234-9 · Cited by 50
“Touchscreen devices such as smartphones and tablets have helped make tasks like typing and gaming more accessible 1–3 .”
Enhancing Neuroplasticity for Post-Stroke Motor Recovery: Mechanisms, Models, and Neurotechnology
Yan, W., Lin, Y., Chen, Y.-F., Wang, Y., Wang, J., & Zhang, M. (2025) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2025.3551753 · Cited by 56
“Research has examined the clinical efficacy of various feedback modalities, including robotic [92], functional electrical stimulation (FES) and NMES [70].”
Herring, E. Z., Graczyk, E. L., Memberg, W. D., Adams, R., Baca-Vaca, G. F., Hutchison, B. C., Krall, J. T., Alexander, B. J., Conlan, E. C., Alfaro, K. E., et al. (2023) · Neurosurgery · doi:10.1227/neu.0000000000002769 · Cited by 46
“Previous studies have used focal brain stimulation to produce arti fi cial sensory percepts for improving control of brain-controlled robotic limbs and BMI user experience, 18,19,45 but these and other 17 studies did not combine sensory feedback with brain-controlled FES.”
BackgroundNatraj, N., Seko, S., Abiri, R., Miao, R., Yan, H., Graham, Y., Tu-Chan, A., Chang, E. F., & Ganguly, K. (2025) · Cell · doi:10.1016/j.cell.2025.02.001 · Cited by 48
Wang, T., Chen, Y., & Cui, H. (2022) · Neuroscience Bulletin · doi:10.1007/s12264-022-00832-x · Cited by 43
Neurotechnologies to restore hand functions
Losanno, E., Mender, M. J., Chestek, C. A., Shokur, S., & Micera, S. (2023) · Nature Reviews Bioengineering · doi:10.1038/s44222-023-00054-4 · Cited by 47
Wei, C., Lin, W., Wang, L., Cao, Z., Huang, Z., Liao, Q., Guo, Z., Su, Y., Zheng, Y., Liao, X., et al. (2023) · Nano-Micro Letters · doi:10.1007/s40820-023-01176-5 · Cited by 41
“Interactive electronic devices (IE devices) in the metaverse enable users to enhance the quality of interactions and create real-time closed-loop feedback through an intuitive interface [1–7].”
BackgroundBidirectional Sim-to-Real Transfer for GelSight Tactile Sensors With CycleGAN
Chen, W., Xu, Y., Chen, Z., Zeng, P., Dang, R., Chen, R., & Xu, J. L. (2022) · IEEE Robotics and Automation Letters · doi:10.1109/lra.2022.3167064 · Cited by 43
“T ACTILE sensing is essential in robot’s interaction with objects and the environment [1], as it directly provides contact states and offers complementary information aside from visual sensors.”
BackgroundExploring Cognition with Brain–Machine Interfaces
Andersen, R. A., Aflalo, T., Bashford, L., Bjånes, D. A., & Kellis, S. (2022) · Annual Review of Psychology · doi:10.1146/annurev-psych-030221-030214 · Cited by 46
“Furthermore, the addition of sensory feedback by ICMS to human S1 improves the time to grasp objects with a robotic limb controlled by motor cortex activity in a human, compared with the previous standard of visual feedback only (Flesher et al. 2021).”
BackgroundA triboelectric tactile sensor with flower-shaped holes for texture recognition
Xing, P., An, S., Wu, Y., Gui, L., Liu, S., Wang, J., Cheng, Y., Zhang, Y., & Pu, X. (2023) · Nano Energy · doi:10.1016/j.nanoen.2023.108758 · Cited by 41
Sparling, T. L., Iyer, L., Pasquina, P. F., & Petrus, E. (2024) · Journal of Neuroscience · doi:10.1523/jneurosci.1051-23.2023 · Cited by 44
Seven technologies to watch in 2024
Eisenstein, M. S. (2024) · Nature · doi:10.1038/d41586-024-00173-x · Cited by 45
Effects of stimulus pulse rate on somatosensory adaptation in the human cortex
Hughes, C. L., Flesher, S. N., & Gaunt, R. A. (2022) · Brain stimulation · doi:10.1016/j.brs.2022.05.021 · Cited by 36
Self-citationWei, Y., Marshall, A. G., McGlone, F. P., Makdani, A. D., Zhu, Y., Yan, L., Ren, L., & Wei, G. (2024) · Nature Communications · doi:10.1038/s41467-024-50616-2 · Cited by 47
“Simple pain-re fl ex or slipping reactive controls alongside object recognition were also implemented 6,8,31,33 – 35 .”
Chandrasekaran, S., Fifer, M. S., Bickel, S., Osborn, L. E., Herrero, J. L., Christie, B. P., Xu, J., Murphy, R. K. J., Singh, S., Glasser, M. F., et al. (2021) · Bioelectronic Medicine · doi:10.1186/s42234-021-00076-6 · Cited by 44
“This approach has been shown to improve prosthetic arm control performance, particularly in grasp confidence and transfer time (Flesher et al. 2021).”
BackgroundSelf-citationAcquiring new memories in neocortex of hippocampal-lesioned mice
Luo, W., Yun, D., Hu, Y., Tian, M., Yang, J., Xu, Y., Tang, Y. C., Zhan, Y., Xie, H. B., & Guan, J. (2022) · Nature Communications · doi:10.1038/s41467-022-29208-5 · Cited by 42
An, H., Nason, S. R., Lim, J., Kwon, K., Willsey, M. S., Patil, P. G., Kim, H.-S., Sylvester, D., Chestek, C. A., & Blaauw, D. T. (2022) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2022.3175926 · Cited by 34
Neural Plasticity in Sensorimotor Brain–Machine Interfaces
Dadarlat, M. C., Canfield, R. A., & Orsborn, A. L. (2023) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-110220-110833 · Cited by 36
“While feedback like vision allows almost all motor BMIs to be closed-loop systems, movement precision and real-world interactions will be limited in the absence of proprioceptive and tactile feedback (4, 103).”
BackgroundLi, F., Gallego, J., Tirko, N. N., Greaser, J., Bashe, D., Patel, R., Shaker, E., Valkenburg, G. E. V., Alsubhi, A. S., Wellman, S. M., et al. (2024) · Nature Communications · doi:10.1038/s41467-024-49709-9 · Cited by 40
“Penetrating microelectrode arrays that interface with the nervous system are front-end components of brain – computer interfaces (BCI), which have demonstrated remarkable potential for restoring motor andsensoryfunction 17 – 19 .”
Demonstration of a portable intracortical brain-computer interface
Weiss, J. M., Gaunt, R. A., Franklin, R. K., Boninger, M. L., & Collinger, J. L. (2019) · Brain-Computer Interfaces · doi:10.1080/2326263x.2019.1709260 · Cited by 37
Self-citationA modular strategy for next-generation upper-limb sensory-motor neuroprostheses
Shokur, S., Mazzoni, A., Schiavone, G., Weber, D. J., & Micera, S. (2021) · Med · doi:10.1016/j.medj.2021.05.002 · Cited by 37
Malleability of the cortical hand map following a finger nerve block
Wesselink, D. B., Sanders, Z., Edmondson, L. R., Dempsey‐Jones, H., Kieliba, P., Kikkert, S., Themistocleous, A. C., Emir, U. E., Diedrichsen, J., Saal, H. P., et al. (2022) · Science Advances · doi:10.1126/sciadv.abk2393 · Cited by 35
Metadata
- Authors
- Flesher, S. N., Downey, J. E., Weiss, J. M., Hughes, C. L., Herrera, A. J., Tyler-Kabara, E. C., Boninger, M. L., Collinger, J. L., & Gaunt, R. A.
- Affiliations
- University of Pittsburgh (US)
- Department of Neurosurgery, University of Texas at Austin, Austin, TX, USA.
- United States Department of Veterans Affairs (US)
- Corresponding author
- Flesher, Sharlene N.
- ORCID
- Venue
- Science
- Published
- 2021-05-21
- Volume
- 372
- Issue
- 6544
- Pages
- 831-836
- ISSN
- 0036-8075 (ISSN-L)
- 1095-9203
- PMC
- PMC8715714
- OpenAlex
- W3161143221
- Cited by
- 617
- References
- 40
- Citations by year
- 2026: 102 · 2025: 154 · 2024: 131 · 2023: 102 · 2022: 75 · 2021: 41
- FWCI
- 45.79
- Type
- journal article
- Language
- en
- Open access
- green
- doi.org
- Locations
- Science (publishedVersion)
- bioRxiv (Cold Spring Harbor Laboratory) (acceptedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- Defense Advanced Research Projects Agency (N66001-16-C-4051, N66001-10-C-4056)
- National Science Foundation (DGE-1247842)
- Topics
- Life Sciences › Neuroscience › Cognitive Neuroscience › EEG and Brain-Computer Interfaces
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Health Sciences › Medicine › Rehabilitation › Stroke Rehabilitation and Recovery
- Keywords
- brain-computer interface, tactile feedback, robotic arm control, tetraplegia, intracortical microstimulation, somatosensory cortex, motor cortex, upper limb prosthesis, grasping, bidirectional neural interface
- MeSH terms
- Adult, Arm, Artificial Limbs, Humans, Male, Motor Cortex, Movement, Quadriplegia, Robotics, Somatosensory Cortex, Touch, Hand Strength, Brain-Computer Interfaces
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.