Validation of photoplethysmography as a method to detect heart rate during rest and exercise
Abstract
Despite their enhanced marketplace visibility, validity of wearable photoplethysmographic heart rate monitoring is scarce. Forty-seven healthy participants performed seven, 6-min exercise bouts and completed a valid skin type scale. Participants wore an Omron HR500U (OHR) and a Mio Alpha (MA), two commercial wearable photoplethysmographic heart rate monitors. Data were compared to a Polar RS800CX (PRS). Means and error were calculated between devices using minutes 2-5. Compared to PRS, MA data was significantly different in walking, biking (2.41 ± 3.99 bpm and 3.26 ± 11.38 bpm, p < 0.05) and weight lifting (23.30 ± 31.94 bpm, p < 0.01). OHR differed from PRS in walking (4.95 ± 7.53 bpm, p < 0.05) and weight lifting (4.67 ± 8.95 bpm, p < 0.05). MA during elliptical, stair climbing and biking conditions demonstrated a strong correlation between jogging speed and error (r = 0.55, p < 0.0001), and showed differences in participants with less photosensitive skin.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
PPG sensors used to measure HR are liable to poor accuracy during activities with increased physical exertion or activities involving repetitive contractions of forearm skeletal muscles.
“Prior research has suggested that PPG sensors used to measure the HR are liable to poor accuracy during activities with increased physical exertion or activities involving repetitive contractions of forearm skeletal muscles [36-38].”
From Accuracy of Wrist-Worn Activity Monitors During Common Daily Physical Activities and Types of Structured Exercise: Evaluation Study (2018) · cites this paper as [38]
The remaining two of the ten included studies observed mixed results across different wearable devices.
“The remaining two studies showed mixed results.”
From Accuracy of heart rate measurement with wrist-worn wearable devices in various skin tones: a systematic review (2023) · cites this paper as [16]
Spierer et al. found that the Mio Alpha device had a higher mean average error in Fitzpatrick 6 patients compared to Fitzpatrick 1 patients (16 beats/min compared to 3 beats/min).
“Spierer et al. found that one of the two devices tested (Mio Alpha) had a higher mean average error in Fitzpatrick 6 patients compared to Fitzpatrick 1 patients (16 beats/min compared to 3 beats/min, respectively) [16].”
From Accuracy of heart rate measurement with wrist-worn wearable devices in various skin tones: a systematic review (2023) · cites this paper as [16]
Heart rate measurement error increases with activity intensity for the Omron HR-500U and Mio Alpha when jogging, stair climbing, and using the stationary bike.
“Investigators have found that heart rate measurement error increases with activity intensity for the Omron HR-500U and Mio Alpha when jogging, stair climbing, and using the stationary bike [21].”
From Estimating Accuracy at Exercise Intensities: A Comparative Study of Self-Monitoring Heart Rate and Physical Activity Wearable Devices (2017) · cites this paper as [21]
Cites
Heart Disease and Stroke Statistics—2014 Update
Go, A. S., Mozaffarian, D., Roger, V. L., Benjamin, E. J., Berry, J. D., Blaha, M. J., Dai, S., Ford, E. S., Fox, C. S., Franco, S., et al. (2013) · Circulation · doi:10.1161/01.cir.0000441139.02102.80 · Cited by 5,207 · see citing works
(2005) · JAMA · doi:10.1001/jama.294.19.2423-c · Cited by 2,325 · see citing works
PHYSICAL ACTIVITY AS AN INDEX OF HEART ATTACK RISK IN COLLEGE ALUMNI1
PAFFENBARGER, R. S., WING, A. L., & HYDE, R. T. (1978) · American Journal of Epidemiology · doi:10.1093/oxfordjournals.aje.a112608 · Cited by 1,980 · see citing works
Heart-rate Recovery Immediately After Exercise as a Predictor of Mortality
Cole, C. R., Blackstone, E. H., Pashkow, F. J., Snader, C. E., & Lauer, M. S. (1999) · New England Journal of Medicine · doi:10.1056/nejm199910283411804 · Cited by 1,935 · see citing works
Medicine & science in sports & exercise
Walle, J. M. L., Solá, I. B., Castillo, I., & Rodríguez, J. L. T. (1994) · Journal of the American Dietetic Association · doi:10.1016/0002-8223(94)92575-5 · Cited by 1,309 · see citing works
The Technology of Accelerometry-Based Activity Monitors: Current and Future
CHEN, K. Y., & BASSETT, D. R. (2005) · Medicine & Science in Sports & Exercise · doi:10.1249/01.mss.0000185571.49104.82 · Cited by 996 · see citing works
Multiple Chronic Conditions Among US Adults: A 2012 Update
Ward, B. W., Schiller, J. S., & Goodman, R. A. (2014) · Preventing Chronic Disease · doi:10.5888/pcd11.130389 · Cited by 848 · see citing works
Nishime, E. O. (2000) · JAMA · doi:10.1001/jama.284.11.1392 · Cited by 713 · see citing works
Cole, C. R., Foody, J. M., Blackstone, E. H., & Lauer, M. S. (2000) · Annals of Internal Medicine · doi:10.7326/0003-4819-132-7-200004040-00007 · Cited by 586 · see citing works
Validation of the COSMED K4 b2 Portable Metabolic System
McLaughlin, J. E., King, G. A., Howley, E. T., Bassett, D. R., & Ainsworth, B. (2001) · International Journal of Sports Medicine · doi:10.1055/s-2001-13816 · Cited by 500 · see citing works
Brage, S., Brage, N., Franks, P. W., Ekelund, U., Wong, M.-Y., Andersen, L. B., Froberg, K., & Wareham, N. J. (2004) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00703.2003 · Cited by 473 · cited 1× · see citing works
“Measuring heart rate works well in combination with accelerometry, as the measurement error of these two approaches is not positively correlated, resulting in more accurate measures of overall activity [16,17].”
BackgroundChronic Care: Making the Case for Ongoing Care
Anderson, G. F. (2010) · Cited by 457 · see citing works
Reliability of accelerometry-based activity monitors: a generalizability study.
Welk, G. J., Schaben, J. A., & Morrow, J. R. (2004) · PubMed · Cited by 394 · see citing works
Estimating energy expenditure using accelerometers
Crouter, S. E., Churilla, J. R., & Bassett, D. R. (2006) · European Journal of Applied Physiology · doi:10.1007/s00421-006-0307-5 · Cited by 366 · see citing works
Predicting Activity Energy Expenditure Using the Actical® Activity Monitor
Heil, D. P. (2006) · Research Quarterly for Exercise and Sport · doi:10.1080/02701367.2006.10599333 · Cited by 353 · see citing works
Technical Reliability Assessment of Three Accelerometer Models in a Mechanical Setup
Esliger, D., & Tremblay, M. S. (2006) · Medicine & Science in Sports & Exercise · doi:10.1249/01.mss.0000239394.55461.08 · Cited by 244 · see citing works
Accuracy of the Actiheart for the assessment of energy expenditure in adults
Crouter, S. E., Churilla, J. R., & Bassett, D. R. (2008) · European Journal of Clinical Nutrition · doi:10.1038/sj.ejcn.1602766 · Cited by 236 · see citing works
Validation and Calibration of the Actical Accelerometer in Preschool Children
Pfeiffer, K. A., McIver, K. L., Dowda, M. C., Almeida, M. J., & Pate, R. R. (2005) · Medicine & Science in Sports & Exercise · doi:10.1249/01.mss.0000183219.44127.e7 · Cited by 233 · see citing works
Heart Rate Monitors: Validity, Stability, and Functionality
Léger, L., & Thivierge, M. (1988) · The Physician and Sportsmedicine · doi:10.1080/00913847.1988.11709511 · Cited by 210 · see citing works
Adam Noah, J., Spierer, D. K., Gu, J., & Bronner, S. (2013) · Journal of Medical Engineering & Technology · doi:10.3109/03091902.2013.831135 · Cited by 202 · see citing works
Self-citationThe Cosmed K4 Telemetry System as an Accurate Device for Oxygen Uptake Measurements during Exercise
Hausswirth, C., Bigard, A. X., & Chevalier, J.-M. L. (1997) · International Journal of Sports Medicine · doi:10.1055/s-2007-972662 · Cited by 195 · see citing works
Comparison of PAEE from Combined and Separate Heart Rate and Movement Models in Children
Corder, K., Brage, S., Wareham, N. J., & Ekelund, U. (2005) · Medicine & Science in Sports & Exercise · doi:10.1249/01.mss.0000176466.78408.cc · Cited by 170 · see citing works
Influence of skin type and wavelength on light wave reflectance
Fallow, B. A., Tarumi, T., & Tanaka, H. (2013) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-013-9436-7 · Cited by 162 · see citing works
Simultaneous measurement of heart rate and body motion to quantitate physical activity
Haskell, W. L., Yee, M. C., EVANS, A., & IRBY, P. J. (1993) · Medicine & Science in Sports & Exercise · doi:10.1249/00005768-199301000-00015 · Cited by 155 · see citing works
Physical Training and Glucose Tolerance in Middle-aged Men with Chemical Diabetes
Saltin, B., Lindgärde, F., Houston, M. E., Hörlin, R., Nygaard, E., & Gad, P. (1979) · Diabetes · doi:10.2337/diab.28.1.s30 · Cited by 154 · see citing works
A photoelectric plethysmograph for the measurement of cutaneous blood flow
CHALLONER, A. V. J., & Ramsay, C. A. (1974) · Physics in Medicine and Biology · doi:10.1088/0031-9155/19/3/003 · Cited by 145 · see citing works
The role of tissue pressure measurement in diagnosing chronic anterior compartment syndrome
Rorabeck, C. H., Bourne, R. B., Fowler, P. J., Finlay, J. B., & Nott, L. (1988) · The American Journal of Sports Medicine · doi:10.1177/036354658801600209 · Cited by 140 · see citing works
Goodie, J. L., Larkin, K. T., & Schauss, S. L. (2000) · Journal of Psychophysiology · doi:10.1027//0269-8803.14.3.159 · Cited by 132 · see citing works
Thompson, D., Batterham, A. M., Bock, S., Robson, C., & Stokes, K. (2006) · The Journal of Nutrition · doi:10.1093/jn/136.4.1037 · Cited by 129 · see citing works
Validation of a heart rate monitor with children in laboratory and field settings.
Treiber, F. A., Musante, L., Hartdagan, S., Davis, H., Levy, M., & Strong, W. B. (1989) · PubMed · Cited by 116 · see citing works
PAFFENBARGER, J. R. S., Brand, R., Sholtz, R. I., & Jung, D. L. (1978) · American Journal of Epidemiology · doi:10.1093/oxfordjournals.aje.a112578 · Cited by 101 · see citing works
Validation of a heart rate monitor with children in laboratory and field settings
Treiber, F. A., Musante, L., Hartdagan, S., Davis, H., LEVY, M., & Strong, W. B. (1989) · Medicine & Science in Sports & Exercise · doi:10.1249/00005768-198906000-00019 · Cited by 83 · see citing works
Validation of the Actiheart Monitor for the Measurement of Physical Activity
Barreira, T. V., Kang, M., Caputo, J. L., Farley, R. S., & Renfrow, M. S. (2009) · International journal of exercise science · doi:10.70252/ltrb1450 · Cited by 76 · see citing works
Perrin, O., Terrier, P., Ladetto, Q., et al. (2000) · Medical & Biological Engineering & Computing · doi:10.1007/bf02344771 · Cited by 76 · see citing works
Spierer, D. K., Hagins, M., Rundle, A., & Pappas, E. (2010) · European Journal of Applied Physiology · doi:10.1007/s00421-010-1672-7 · Cited by 73 · see citing works
Self-citationMoon, J., & Butte, N. F. (1996) · Journal of Applied Physiology · doi:10.1152/jappl.1996.81.4.1754 · Cited by 68 · see citing works
The precision and accuracy of a portable heart rate monitor.
Seaward, B. L., SLEAMAKER, R. H., McAuliffe, T. L., & Jf, C. (1990) · PubMed · Cited by 65 · see citing works
Tiwari, T., Casciello, A., Gansky, S. A., Henshaw, M., Ramos‐Gomez, F., Rasmussen, M., García, R. I., Albino, J. E. N., & Batliner, T. S. (2014) · Preventing Chronic Disease · doi:10.5888/pcd11.140140 · Cited by 50 · see citing works
Validity of the Smarthealth Watch to Measure Heart Rate During Rest and Exercise
Lee, C. M., & Gorelick, M. (2011) · Measurement in Physical Education and Exercise Science · doi:10.1080/1091367x.2011.539089 · Cited by 42 · see citing works
Validation du Sport Tester PE 3000 en fonction de l'enregistrement Holter
Vogelaere, P. P., Meyer, F. D., Duquet, W., & Vandevelde, P. (1987) · Science & Sports · doi:10.1016/s0765-1597(87)80024-2 · Cited by 21 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Henriksen, A., Haugen Mikalsen, M., Woldaregay, A. Z., Muzny, M., Hartvigsen, G., Hopstock, L. A., & Grimsgaard, S. (2018) · Journal of Medical Internet Research · doi:10.2196/jmir.9157 · Cited by 542 · see citing works
Accuracy of Heart Rate Watches: Implications for Weight Management
Wallen, M. P., Gomersall, S. R., Keating, S. E., Wisløff, U., & Coombes, J. S. (2016) · PLOS ONE · doi:10.1371/journal.pone.0154420 · Cited by 336 · see citing works
Nelson, B. W., & Allen, N. B. (2019) · JMIR mHealth and uHealth · doi:10.2196/10828 · Cited by 310 · see citing works
Stahl, S. E., An, H.-S., Dinkel, D. M., Noble, J. M., & Lee, J.-M. (2016) · BMJ Open Sport & Exercise Medicine · doi:10.1136/bmjsem-2015-000106 · Cited by 276 · see citing works
Guidelines for wrist-worn consumer wearable assessment of heart rate in biobehavioral research
Nelson, B. W., Low, C. A., Jacobson, N., Areán, P., Torous, J., & Allen, N. B. (2020) · npj Digital Medicine · doi:10.1038/s41746-020-0297-4 · Cited by 241 · see citing works
Dooley, E. E., Golaszewski, N. M., & Bartholomew, J. B. (2017) · JMIR mHealth and uHealth · doi:10.2196/mhealth.7043 · Cited by 242 · see citing works
Walch, O., Huang, Y., Forger, D., & Goldstein, C. (2019) · Sleep · doi:10.1093/sleep/zsz180 · Cited by 240 · see citing works
Reddy, R. K., Pooni, R., Zaharieva, D. P., Senf, B., El Youssef, J., Dassau, E., Doyle III, F. J., Clements, M. A., Rickels, M. R., Patton, S. R., et al. (2018) · JMIR mHealth and uHealth · doi:10.2196/10338 · Cited by 157 · see citing works
Assessment of the Fitbit Charge 2 for monitoring heart rate
Benedetto, S., Caldato, C., Bazzan, E., Greenwood, D. C., Pensabene, V., & Actis, P. (2018) · PLOS ONE · doi:10.1371/journal.pone.0192691 · Cited by 149 · see citing works
Fonseca, P., Weysen, T., Goelema, M. S., Møst, E. I. S., Radha, M., Lunsingh Scheurleer, C., van den Heuvel, L., & Aarts, R. M. (2017) · SLEEP · doi:10.1093/sleep/zsx097 · Cited by 148 · see citing works
Chow, H.-W., & Yang, C.-C. (2020) · JMIR mHealth and uHealth · doi:10.2196/14707 · Cited by 140 · see citing works
“To compare the accuracy of test devices, various statistical methods were chosen based on recommendations from relevant studies [2,10,26,38,51].”
InfluentialGorny, A. W., Liew, S. J., Tan, C. S., & Müller-Riemenschneider, F. (2017) · JMIR mHealth and uHealth · doi:10.2196/mhealth.8233 · Cited by 94 · see citing works
Koerber, D., Khan, S., Shamsheri, T., Kirubarajan, A., & Mehta, S. (2023) · Journal of Racial and Ethnic Health Disparities · doi:10.1007/s40615-022-01446-9 · Cited by 91 · see citing works
Zhang, Y., Weaver, R. G., Armstrong, B., Burkart, S., Zhang, S., & Beets, M. W. (2020) · Journal of Sports Sciences · doi:10.1080/02640414.2020.1767348 · Cited by 88 · see citing works
Fan, Y.-Y., Li, Y.-G., Li, J., Cheng, W.-K., Shan, Z.-L., Wang, Y.-T., & Guo, Y.-T. (2019) · JMIR mHealth and uHealth · doi:10.2196/11437 · Cited by 81 · see citing works
Comparison of Polar M600 Optical Heart Rate and ECG Heart Rate during Exercise
HORTON, J. F., STERGIOU, P., FUNG, T. S., & KATZ, L. (2017) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000001388 · Cited by 79 · see citing works
Sartor, F., Gelissen, J., van Dinther, R., Roovers, D., Papini, G. B., & Coppola, G. (2018) · BMC Sports Science, Medicine and Rehabilitation · doi:10.1186/s13102-018-0098-0 · Cited by 65 · see citing works
Validation of a Photoplethysmographic Heart Rate Monitor: Polar OH1
Hermand, E., Cassirame, J., Ennequin, G., & Hue, O. (2019) · International Journal of Sports Medicine · doi:10.1055/a-0875-4033 · Cited by 59 · see citing works
Sañudo, B., De Hoyo, M., Muñoz-López, A., Perry, J., & Abt, G. (2019) · Journal of Medical Systems · doi:10.1007/s10916-019-1325-2 · Cited by 48 · see citing works
Béres, S., Holczer, L., & Hejjel, L. (2019) · Measurement Science Review · doi:10.2478/msr-2019-0030 · Cited by 33 · see citing works
Skin tone, confidence, and data quality of heart rate sensing in WearOS smartwatches
Ray, I., Liaqat, D., Gabel, M., & de Lara, E. (2021) · 2021 IEEE International Conference on Pervasive Computing and Communications Workshops and other Affiliated Events (PerCom Workshops) · doi:10.1109/percomworkshops51409.2021.9431120 · Cited by 23 · see citing works
Herkert, C., Kraal, J. J., van Loon, E. M. A., van Hooff, M., & Kemps, H. M. C. (2019) · JMIR mHealth and uHealth · doi:10.2196/15045 · Cited by 19 · see citing works
Validity of heart rate measurements in wrist-based monitors across skin tones during exercise
Hung, S. H., Serwa, K., Rosenthal, G., & Eng, J. J. (2025) · PLOS ONE · doi:10.1371/journal.pone.0318724 · Cited by 21 · see citing works
Støve, M. P., & Larsen, B. T. (2020) · European Journal of Physiotherapy · doi:10.1080/21679169.2019.1628300 · Cited by 5 · see citing works
- Authors
- Spierer, D. K., Rosen, Z., Litman, L. L., & Fujii, K.
- Affiliations
- Long Island University (US)
- New York Medical College (US)
- Lander University (US)
- Omron (Japan) (JP)
- Corresponding author
- Spierer, David K.
- Venue
- Journal of Medical Engineering & Technology
- Publisher
- Informa UK Limited
- Published
- 2015-06-26
- Volume
- 39
- Issue
- 5
- Pages
- 264-271
- ISSN
- 0309-1902 (ISSN-L)
- 1464-522X
- OpenAlex
- W2119121237
- Cited by
- 139
- References
- 40
- Citations by year
- 2026: 4 · 2025: 10 · 2024: 9 · 2023: 9 · 2022: 11 · 2021: 19 · 2020: 14 · 2019: 21 · 2018: 21 · 2017: 15 · 2016: 5 · 2015: 1
- FWCI
- 6.96
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Journal of Medical Engineering & Technology (publishedVersion)
- PubMed (publishedVersion)
- Touro Scholar (Touro College) (submittedVersion)
- Funders
- Omron Healthcare
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Health Sciences › Medicine › Complementary and alternative medicine › Cardiovascular and exercise physiology
- Keywords
- photoplethysmography, wearable heart rate monitor, heart rate response to exercise, device comparison, walking, cycling, weightlifting
- MeSH terms
- Adult, Algorithms, Diagnosis, Computer-Assisted, Equipment Design, Female, Heart Rate, Humans, Male, Motor Activity, Rest, Sensitivity and Specificity, Reproducibility of Results, Exercise, Photoplethysmography, Monitoring, Ambulatory, Equipment Failure Analysis, Actigraphy
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review