Deep Eutectic Solvent-Based Microextraction of Lead(II) Traces from Water and Aqueous Extracts before FAAS Measurements
Open accessAbstract
Microextraction procedures for the separation of Pb(II) from water and food samples extracts were developed. A deep eutectic solvent composed of α-benzoin oxime and iron(III) chloride dissolved in phenol was applied as a phase separator support. In addition, this deep eutectic mixture worked as an efficient extractor of Pb(II). The developed microextraction process showed a high ability to tolerate the common coexisting ions in the real samples. The optimum conditions for quantitative recoveries of Pb(II) from aqueous extracts were at pH 2.0, conducted by adding 150 µL from the deep eutectic solvent. The quantitative recoveries were obtained with various initial sample volumes up to 30 mL. Limits of detection and limits of quantification of 0.008 and 0.025 µg L−1 were achieved with a relative standard deviation (RSD%) of 2.9, which indicates the accuracy and sensitivity of the developed procedure. Recoveries from the reference materials, including TMDA 64.2, TMDA 53.3, and NCSDC-73349, were 100%, 97%, and 102%, respectively. Real samples, such as tap, lake, and river water, as well as food samples, including salted peanuts, chickpeas, roasted yellow corn, pistachios, and almonds, were successfully applied for Pb(II) analysis by atomic absorption spectroscopy (AAS) after applying the developed deep eutectic solvent-based microextraction procedures.
Overview
CoChat Summary
TL;DR
A deep eutectic solvent made of α-benzoin oxime, iron(III) chloride, and phenol was used for microextraction of Pb(II) from water and food extracts before FAAS. The method achieved LOD 0.008 µg L−1, LOQ 0.025 µg L−1, RSD 2.9%, and quantitative recovery under optimized conditions.
Key findings
- Optimum recovery occurred at pH 2.0 with 150 µL of deep eutectic solvent; 50–100 µL gave approximately 80% recovery, while at least 150 µL gave approximately 100%.
- LOD and LOQ were 0.008 and 0.025 µg L−1, respectively, with RSD 2.9%; the preconcentration factor was 60 using sample volumes up to 30 mL and a final extracted volume of 500 µL.
- Certified reference material recoveries were 100% for TMDA 64.2, 97% for TMDA 53.3, and 102% for NCSDC-73349.
- Foreign ions gave recoveries of 94–103%; addition/recovery experiments gave 92–101% in the Results section, while the Conclusions state 92–106%.
Methods
- The DES was prepared by mixing iron(III) chloride with phenol at a 1:5 molar ratio, warming at 50 °C for 10 min, stirring for 60 min at room temperature, then adding 0.01 g α-benzoin oxime and stirring another 60 min; melting point was 28.6 °C and viscosity was 47 cps.
- Pb(II) samples were adjusted to pH 2, injected with approximately 150 µL DES, vortexed for 5 min, centrifuged for 12 min at 4000 rpm, and the remaining phase was dissolved in nitric acid to 500 µL for FAAS.
- Validation used TMDA 64.2, TMDA 53.3, and NCSDC-73349; water samples were filtered through 0.45-µm filters and 10 mL was used, while food samples were digested from 1.0 g to 10 mL and 2 mL extract was used; spiking was tested at 0.5, 1.0, and 2.0 mg L−1 with three replicates.
Limitations
- Not stated in the paper; there is no explicit limitations section.
- Evident from the text: the addition/recovery range is reported inconsistently, with Results stating 92–101% and Conclusions stating 92–106%.
- Evident from the text: the narrative does not report individual foreign-ion concentrations or full interference data, and real-sample validation is limited to the listed water and food samples and spiking levels.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures



Verification
Claims the paper makes
1 Valid · 2 Not valid · 7 Not checked · 27 Own results and interpretations
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“The sample volume is an important factor for obtaining a high preconcentration factor [51,52,53,54,55,56,57,58].”
§ Deep Eutectic Solvent-Based Microextraction of Lead(II) Traces from Water and Aqueous Extracts before FAAS Measurements > 2. Results and Discussion > 2.4. Analytical Merit for the Developed DES-ME of Pb(II)
What the sources say
- Not checked yetcites [51] Tuzen, 2009, A preconcentration system for determination of copper and nickel in water and food samples employing flame atomic absorption spectrometry, J. Hazard. Mater. source
- Not checked yetcites [52] Koosha, E., Shamsipur, M., Salimi, F., and Ramezani, M. (2020). A microextraction method based on precipitation for the simultaneous separation and preconcentration of cadmium and lead before their determination by FAAS: Experimental design methodology. Sep. Sci. Technol., 1–9. source
- Not checked yetcites [53] Soylak, 1997, Determination of trace amounts of cobalt in natural water samples as 4-(2-Thiazolylazo) recorcinol complex after adsorptive preconcentration, Anal. Lett. source
- Not checked yetcites [54] Movahhedi, 2020, Response surface methodology for heavy metals removal by tioglycolic-modified Zn–Fe layer double hydroxide as a magnetic recyclable adsorbent, Chem. Pap. source
- Not checked yetcites [55] Tuzen, 2004, Column solid phase extraction of copper, iron, and zinc ions at trace levels in environmental samples on amberlite XAD-7 for their flame atomic absorption spectrometric determinations, Anal. Lett. source
- Partly supportedcites [56] Sadeghi, 2017, Simultaneous determination of ultra-low traces of lead and cadmium in food and environmental samples using dispersive solid-phase extraction (DSPE) combined with ultrasound-assisted emulsification microextraction based on the solidification of floating organic drop (UAEME-SFO) followed by GFAAS, RSC Adv. sourceExperimental — accuracy not yet measured
Section 3.2.1.1 ('Effect of the breakthrough volume') states that 'the sample volume is one of the important parameters influencing the extraction recovery' and that 'in order to achieve a high extraction recovery for sample with very low analyte concentration, a large volume of samples solution is required' (volumes 50–2000 mL tested, tolerated breakthrough at 1500 mL, 100 mL optimized). This supports sample volume as an important optimization parameter, but the paper ties it to extraction recovery/breakthrough, not to the preconcentration (enrichment) factor; the enrichment factor is mentioned only generally in Section 3.2 ('this combination lead to very high enrichment factor', 'in order to obtain a high extraction recovery and enrichment factor... the DSPE and USAEME conditions were optimized'), with no stated dependence on sample volume. So the specific attribution to a high preconcentration factor is not made by this source. (LLM, conf=0.72)
- Not checked yetcites [57] Rivas, 2009, Determination of traces of lead and cadmium using dispersive liquid-liquid microextraction followed by electrothermal atomic absorption spectrometry, Microchim. Acta source
- Not checked yetcites [58] Pawliszyn, 1996, Determination of Tetraethyllead and Inorganic Lead in Water by Solid Phase Microextraction/Gas Chromatography, Anal. Chem. source
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
The cited sources that could be checked couldn’t confirm or dispute this claim.
In the paper
“The deep eutectic solvents facilitate the phase separation during the microextraction process.”
§ Deep Eutectic Solvent-Based Microextraction of Lead(II) Traces from Water and Aqueous Extracts before FAAS Measurements > 2. Results and Discussion > 2.2. Optimization of the Deep Eutectic Solvent Quantity
What the sources say
- Could not verifycites [34] Smith, 2014, Deep Eutectic Solvents (DESs) and Their Applications, Chem. Rev. sourceExperimental — accuracy not yet measured
The paper discusses DES use in extraction, but the passages found do not describe microextraction or DESs facilitating phase separation during it. Section 3.3 (p. 15) instead notes that the Type III DESs discussed there are miscible with water and cannot be used for biphasic extraction. Searches for microextraction and phase separation did not establish the cited claim. (LLM, conf=0.75)
- Could not verifycites [35] Kalhor, P., and Ghandi, K. (2019). Deep eutectic solvents for pretreatment, extraction, and catalysis of biomass and food waste. Molecules, 24. sourceExperimental — accuracy not yet measured
Section 4.1.2 discusses DES-based extraction of pyrethroids from beverages and juices, but reports short extraction time and a high enrichment factor, not facilitated phase separation. Searches for phase separation during microextraction found no passage in our copy addressing that effect. (LLM, conf=0.72)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Could not verify
None of the sources cited for this claim have been checked yet.
In the paper
“The limits of detection of 0.008 µg L−1 for the developed DES-ME were between those of the other methods, as shown in Table 5.”
§ Deep Eutectic Solvent-Based Microextraction of Lead(II) Traces from Water and Aqueous Extracts before FAAS Measurements > 2. Results and Discussion > 2.5. Real Food and Water Samples Analyses after Applying the Developed Deep Eutectic-Based Microextraction of Pb(II)
What the sources say
- Not checked yetcites [21] AlOthman, 2015, Solid phase extraction of metal ions in environmental samples on 1-(2-pyridylazo)-2-naphthol impregnated activated carbon cloth, Ecotoxicol. Environ. Saf. source
- Not checked yetcites [23] Habila, 2014, Fe3O4nanoparticles and ultrasound assisted dispersive liquid-liquid microextraction of lead(ii) for its microsampling flame atomic absorption spectrometric determination in food and environmental samples, RSC Adv. source
- Not checked yetcites [24] Zhou, 2011, Determination of lead in environmental waters with dispersive liquid-liquid microextraction prior to atomic fluorescence spectrometry, J. Hazard. Mater. source
- Could not verifycites [30] Mandlate, 2017, Determination of cadmium and lead at sub-ppt level in soft drinks: An efficient combination between dispersive liquid-liquid microextraction and graphite furnace atomic absorption spectrometry, Food Chem. sourceExperimental — accuracy not yet measured
Section 3.7 reports LODs of 0.006 and 0.072 ng L−1 for the paper’s DLLME method and compares them with literature values. It does not address the developed DES-ME method or establish where its reported 0.008 µg L−1 LOD falls among other methods (searched: DES-ME, 0.008, and LOD comparisons). (LLM, conf=0.77)
- Not checked yetcites [33] Shah, 2012, Single step in-syringe system for ionic liquid based liquid microextraction combined with flame atomic absorption spectrometry for lead determination, J. Anal. At. Spectrom. source
- Could not verifycites [34] Smith, 2014, Deep Eutectic Solvents (DESs) and Their Applications, Chem. Rev. sourceExperimental — accuracy not yet measured
The review discusses DES applications, including metal extraction (section 3.3, p. 15), but the passages do not address the developed DES-ME method or compare analytical limits of detection. Searches for 0.008 µg L−1 and for microextraction limits of detection found no passage supporting the claimed comparison. (LLM, conf=0.76)
- Not checked yetcites [42] Alavi, 2017, Deep eutectic liquid organic salt as a new solvent for carrier-mediated hollow fiber liquid phase microextraction of lead from whole blood followed by electrothermal atomic absorption spectrometry, New J. Chem. source
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The developed microextraction process showed a high ability to tolerate the common coexisting ions in the real samples.”
- Also stated in: § 2.3. Tolerance of Foreign Ions in Relation to Recovery · hedged
“The recoveries were in the range between 94% and 103% (Table 1), indicating that the developed DES-ME process for Pb(II) is matrix-independent.”
- Also stated in: § 4. Conclusions · boosted
“The developed DES-ME of Pb(II) exhibited high tolerance for various ions, including F−, PO43−, Ni2+, NO3−, Co2+, Zn2+, CO32−, K+/Cl−, Mg2+, Ca2+, SO42−, and Fe2+.”
§ Deep Eutectic Solvent-Based Microextraction of Lead(II) Traces from Water and Aqueous Extracts before FAAS Measurements > Abstract
boosted- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
- Also stated in: § 2.3. Tolerance of Foreign Ions in Relation to Recovery · hedged
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Limits of detection and limits of quantification of 0.008 and 0.025 µg L−1 were achieved with a relative standard deviation (RSD%) of 2.9, which indicates the accuracy and sensitivity of the developed procedure.”
- Also stated in: § 2.4. Analytical Merit for the Developed DES-ME of Pb(II) · plain
“The developed DES-ME of Pb(II) exhibited limits of detection (LODs) and limits of quantification (LOQs) of 0.008 and 0.025 µg·L−1, respectively, with an RSD% of 2.9, which indicates accuracy and sensitivity.”
§ Deep Eutectic Solvent-Based Microextraction of Lead(II) Traces from Water and Aqueous Extracts before FAAS Measurements > Abstract
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Also stated in: § 2.4. Analytical Merit for the Developed DES-ME of Pb(II) · plain
21 more claims
Create a free account to see all 37 claims and to verify the 27 references still open.
- Probably not a substantial claim. 7
- It may state two separate claims as one. 4
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Electromembrane extraction integrates DES-based systems with electrically driven analyte transport, providing improved selectivity and more effective preconcentration of ionic species compared with many traditional liquid-phase extraction techniques [52,78].”
From Eco-Friendly Deep Eutectic Solvent-Based Extraction Technologies: A Comprehensive Review of Principles, Applications, and Comparative Insights (2026) · cites this paper as [52]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ Deep Eutectic Solvent-Based Microextraction of Lead(II) Traces from Water and Aqueous Extracts before FAAS Measurements > 2. Results and Discussion > 2.2. Optimization of the Deep Eutectic Solvent Quantity
“including phenol and iron(II) chloride, which are important for the deep eutectic solvent structure”
§ Deep Eutectic Solvent-Based Microextraction of Lead(II) Traces from Water and Aqueous Extracts before FAAS Measurements > 3. Materials and Methods > 3.2. DES-ME of Pb(II)
“The deep eutectic extraction solution was obtained by mixing iron(III) chloride with phenol at a molar ratio of 1:5”
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ Deep Eutectic Solvent-Based Microextraction of Lead(II) Traces from Water and Aqueous Extracts before FAAS Measurements > 2. Results and Discussion > 2.5. Real Food and Water Samples Analyses after Applying the Developed Deep Eutectic-Based Microextraction of Pb(II)
“Table 4 presents the recovered Pb(II) concentrations with various spiking levels, which are in agreement with the added one, with recovery between 92% and 101%.”
§ Deep Eutectic Solvent-Based Microextraction of Lead(II) Traces from Water and Aqueous Extracts before FAAS Measurements > 4. Conclusions
“The addition/recovery analysis showed close agreement between the spiked and detected concentrations, with recoveries between 92% and 106%.”
Citations
Cites
Deep Eutectic Solvents (DESs) and Their Applications
Smith, E. L., Abbott, A. P., & Ryder, K. S. (2014) · Chemical Reviews · doi:10.1021/cr300162p · Cited by 7,378
Wani, A. L., Ara, A., & Usmani, J. A. (2015) · Interdisciplinary Toxicology · doi:10.1515/intox-2015-0009 · Cited by 1,964 · cited 1×
“The possibility of the circulating lead binding with erythrocytes is high, leading to its rapid distribution throughout the body, including in organs such as the liver and brain, and in the bones [6].”
BackgroundVardhan, K. H., Kumar, P. S., & Panda, R. C. (2019) · Journal of Molecular Liquids · doi:10.1016/j.molliq.2019.111197 · Cited by 1,618
Environmental Contamination by Heavy Metals
Masindi, V., & Muedi, K. L. (2018) · Heavy Metals · doi:10.5772/intechopen.76082 · Cited by 927
Balkhair, K. S., & Ashraf, M. A. (2016) · Saudi Journal of Biological Sciences · doi:10.1016/j.sjbs.2015.09.023 · Cited by 532
Hydrophobic deep eutectic solvents in microextraction techniques–A review
Makoś, P., Słupek, E., & Gębicki, J. (2020) · Microchemical Journal · doi:10.1016/j.microc.2019.104384 · Cited by 469
Liquid-phase microextraction – The different principles and configurations
Yamini, Y., Rezazadeh, M., & Seidi, S. (2019) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2018.06.010 · Cited by 309 · cited 2×
“However, the liquid-phase microextraction process is superior to solid-phase microextraction due to its simple and quick phase separation as well as its ability to apply a direct injection of the separated phase for instrumental analysis [27,29].”
BackgroundDeep Eutectic Solvents for Pretreatment, Extraction, and Catalysis of Biomass and Food Waste
Kalhor, P., & Ghandi, K. (2019) · Molecules · doi:10.3390/molecules24224012 · Cited by 305 · cited 1×
“Deep eutectic solvents are promising for the attraction of heavy metals from aqueous solutions because they are entirely composed of nitrogen and oxygen in their structure, in addition to the main skeleton of organic derived materials including carbon and hydrogen [34,35].”
BackgroundChandra, R., Bharagava, R. N., Yadav, S., & Mohan, D. (2009) · Journal of Hazardous Materials · doi:10.1016/j.jhazmat.2008.06.040 · Cited by 265 · cited 1×
“In addition, accumulation and transference through the food chain leads to a more significant spread into soil, plants, water, and food, making its way into the human body [4,5].”
BackgroundLiquid–phase microextraction: A review of reviews
Rutkowska, M., Płotka-Wasylka, J., Sajid, M., & Andruch, V. (2019) · Microchemical Journal · doi:10.1016/j.microc.2019.103989 · Cited by 244 · cited 1×
“However, the liquid-phase microextraction process is superior to solid-phase microextraction due to its simple and quick phase separation as well as its ability to apply a direct injection of the separated phase for instrumental analysis [27,29].”
Background
Research works printed but not linked: 1
- (2019, December 19). The Health Effects of Environmental Lead Exposure: Closing Pandora’s Box-Behavioral Measures of Neurotoxicity-NCBI Bookshelf, Available online: https://www.ncbi.nlm.nih.gov/books/NBK234984/.
Cited by
Deep eutectic solvents in liquid-phase microextraction: Contribution to green chemistry
Santos, L. B., de Assis, R. D. S., Barreto, J. A., Bezerra, M. A., Novaes, C. G., & Lemos, V. A. (2021) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2021.116478 · Cited by 161
Applications of deep eutectic solvents for hard-to-separate liquid systems
Cheng, H., & Qi, Z. (2021) · Separation and Purification Technology · doi:10.1016/j.seppur.2021.119027 · Cited by 110
Khalid, A., Tahir, S., Khalid, A. R., Hanif, M. A., Abbas, Q., & Zahid, M. (2024) · Green Chemistry · doi:10.1039/d3gc04112c · Cited by 73
Deep Eutectic Solvents Application in Food Analysis
Ortega-Zamora, C., González‐Sálamo, J., & Hernández‐Borges, J. (2021) · Molecules · doi:10.3390/molecules26226846 · Cited by 55
Application of deep eutectic solvents in atomic absorption spectrometry
Andruch, V., Halko, R., Tuček, J., & Płotka‐Wasylka, J. (2021) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2021.116510 · Cited by 42
Metadata
- Authors
- Habila, M. A., AlMasoud, N., Alomar, T. S., AlOthman, Z. A., Yilmaz, E., & Soylak, M.
- Affiliations
- King Saud University (SA)
- Princess Nourah bint Abdulrahman University (SA)
- Department of Analytical Chemistry, Faculty of Pharmacy, Erciyes University, 38039 Kayseri, Turkey
- Erciyes University (TR)
- Corresponding author
- Habila, Mohamed A.
- Alomar, Taghrid S.
- Venue
- Molecules
- Publisher
- MDPI AG
- Published
- 2020-10-19
- Volume
- 25
- Issue
- 20
- Pages
- 4794
- ISSN
- 1420-3049 (ISSN-L)
- PMC
- PMC7587555
- OpenAlex
- W3093088910
- Cited by
- 38
- References
- 57
- Citations by year
- 2026: 8 · 2025: 4 · 2024: 5 · 2023: 6 · 2022: 5 · 2021: 9 · 2020: 1
- FWCI
- 2.72
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Molecules (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Funders
- the deanship of scientific research at princess nourah bint abdulrahman university (the research groups program grant number RGP-1440-0023.)
- Topics
- Physical Sciences › Chemistry › Analytical Chemistry › Analytical chemistry methods development
- Physical Sciences › Chemical Engineering › Filtration and Separation › Chemical and Physical Properties in Aqueous Solutions
- Physical Sciences › Chemistry › Electrochemistry › Electrochemical Analysis and Applications
- Keywords
- deep eutectic solvents, lead(II) determination, preconcentration, ferric chloride, phenols, water analysis, food samples, trace metal extraction
- MeSH terms
- Animals, Ferric Compounds, Food Contamination, Humans, Lead, Milk, Spectrophotometry, Atomic, Water, Water Pollutants, Chemical, Limit of Detection, Liquid Phase Microextraction
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-09-30, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-09-30 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (qwen3.8-omni-flash and gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Deep Eutectic Solvent-Based Microextraction of Lead(II) Traces from Water and Aqueous Extracts before FAAS Measurements. https://app.cochat.ai/papers/10.3390/molecules25204794
- The paper
- Habila, M. A., AlMasoud, N., Alomar, T. S., AlOthman, Z. A., Yilmaz, E., & Soylak, M. (2020). Deep Eutectic Solvent-Based Microextraction of Lead(II) Traces from Water and Aqueous Extracts before FAAS Measurements. Molecules. https://doi.org/10.3390/molecules25204794
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.