Neuronal loss due to prolonged controlled-current stimulation with chronically implanted microelectrodes in the cat cerebral cortex
Open accessThe DOI matches its Crossref record; 14 of the 14 references it cites were found in a registry or an index.
Abstract
Activated iridium microelectrodes were implanted for 450-1282 days in the sensorimotor cortex of seven adult domestic cats and then pulsed for 240 h (8 h per day for 30 days) at 50 Hz. Continuous stimulation at 2 nC/phase and with a geometric charge density of 100 microC cm(-2) produced no detectable change in neuronal density in the tissue surrounding the microelectrode tips. However, pulsing with a continuous 100% duty cycle at 4 nC/phase and with a geometric charge density of 200 microC cm(-2) induced loss of cortical neurons over a radius of at least 150 microm from the electrode tips. The same stimulus regimen but with a duty cycle of 50% (1 s of stimulation, and then 1 s without stimulation repeated for 8 h) produced neuronal loss within a smaller radius, approximately 60 microm from the center of the electrode tips. However, there also was significant loss of neurons surrounding the unpulsed electrodes, presumably as a result of mechanical injury due to their insertion into and long-term residence in the tissue, and this was responsible for most of the neuronal loss within 150 microm of the electrodes pulsed with the 50% duty cycle.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Neural Stimulation and Recording Electrodes
Cogan, S. F. (2008) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.10.061807.160518 · Cited by 2,243 · cited 2×
“By returning the stimulating electrodes to the bias potential, the stimulator ensures that the electrodes are returned to a specific electrochemical state prior to the next stimulus pulse and thus ensures that no net charge is injected into the tissue, provided that all charge injection is by reversible processes (24).”
BackgroundResponse of brain tissue to chronically implanted neural electrodes
Polikov, V. S., Tresco, P. A., & Reichert, W. M. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2005.08.015 · Cited by 1,896 · cited 1×
“The latter might be mitigated by controlling the sub-acute phase of the inflammatory response induced by the electrodes (27,30-32)}.”
BackgroundBrain–machine interfaces: past, present and future
Lebedev, M. A., & Nicolelis, M. A. L. (2006) · Trends in Neurosciences · doi:10.1016/j.tins.2006.07.004 · Cited by 1,875 · cited 1×
“There are many application in the basic neurosciences that require microstimulation in the cerebral cortex, and several emerging clinical applications, including sensory feedback for a brain-machine interface (14,15) and cortical-level visual prostheses (4,5,16-18).”
BackgroundLearning to control a brain–machine interface for reaching and grasping by primates
Carmena, J. M., Lebedev, M. A., Crist, R. E., O'Doherty, J. E., Santucci, D. M., Dimitrov, D. F., Patil, P. G., Henriquez, C. S., & Nicolelis, M. A. L. (2003) · PLoS Biology · doi:10.1371/journal.pbio.0000042 · Cited by 1,761 · cited 1×
“There are many application in the basic neurosciences that require microstimulation in the cerebral cortex, and several emerging clinical applications, including sensory feedback for a brain-machine interface (14,15) and cortical-level visual prostheses (4,5,16-18).”
BackgroundCharge density and charge per phase as cofactors in neural injury induced by electrical stimulation
McCreery, D. B., Agnew, W. F., Yuen, T. G. H., & Bullara, L. (1990) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.102812 · Cited by 708 · cited 1×
“Thus, electrical stimulation is an effective but also an unnatural means of inducing neuronal activity, and appropriate precautions must be observed in order to minimize neuronal injury (8-13).”
BackgroundSelf-citationSchmidt, E. M., Bak, M. J., Hambrecht, F. T., Kufta, C. V., O'Rourke, D. K., & Vallabhanath, P. (1996) · Brain · doi:10.1093/brain/119.2.507 · Cited by 672
Electrical stimulation of neural tissue to evoke behavioral responses
Tehovnik, E. J. (1996) · Journal of Neuroscience Methods · doi:10.1016/0165-0270(95)00131-x · Cited by 478 · cited 2×
“Microstimulation by penetrating (intraparenchymal) microelectrodes affords better spatial resolution of the electrical stimulus than is possible with stimulating microelectrodes on the surface of the brain (1-3) which may be essential for successful implementation of a cortical-level visual prosthesis (2,4-7).”
BackgroundVisual sensations produced by intracortical microstimulation of the human occipital cortex
Bak, M., Girvin, J. P., Hambrecht, F. T., Kufta, C. V., Loeb, G. E., & Schmidt, E. M. (1990) · Medical & Biological Engineering & Computing · doi:10.1007/bf02442682 · Cited by 299 · cited 1×
“who had been totally blind for 22 years [4, 5].”
BackgroundHe, W., McConnell, G. C., & Bellamkonda, R. V. (2006) · Journal of Neural Engineering · doi:10.1088/1741-2560/3/4/009 · Cited by 177 · cited 1×
“Tissue injury might also be mitigated by a judicious change in the physical configuration of the electrodes and/or by modifying the physical properties of the electrode array surface, including appropriate modifications to the shanks’ tip region and fabrication of the shanks from other materials such as nanoporous sil…”
BackgroundA model for intracortical visual prosthesis research
Troyk, P., Bak, M., Berg, J., Bradley, D., Cogan, S., Erickson, R., Kufta, C., McCreery, D., Schmidt, E., & Towle, V. (2003) · Artificial Organs · doi:10.1046/j.1525-1594.2003.07308.x · Cited by 166 · cited 1×
“There are many application in the basic neurosciences that require microstimulation in the cerebral cortex, and several emerging clinical applications, including sensory feedback for a brain-machine interface (14,15) and cortical-level visual prostheses (4,5,16-18).”
BackgroundSelf-citation
Cited by
Intracortical microstimulation of human somatosensory cortex
Flesher, S. N., Collinger, J. L., Foldes, S. T., Weiss, J. M., Downey, J. E., Tyler-Kabara, E. C., Bensmaia, S. J., Schwartz, A. B., Boninger, M. L., & Gaunt, R. A. (2016) · Science Translational Medicine · doi:10.1126/scitranslmed.aaf8083 · Cited by 704
“Intracortical microstimulation has the potential to damage the electrode (43), surrounding tissue (44), or both, depending on the injected charge and charge density.”
BackgroundTissue damage thresholds during therapeutic electrical stimulation
Cogan, S. F., Ludwig, K. A., Welle, C. G., & Takmakov, P. (2016) · Journal of Neural Engineering · doi:10.1088/1741-2560/13/2/021001 · Cited by 386
“Although stimulus frequencies for clinical DBS (130–185 Hz) are higher those used in most studies of stimulation-induced neural damage, minimal adverse tissue response has been reported (Haberler et al 2000, Burbaud et al 2002). As suggested by Kuncel and Grill (2004), the absence of clinically relevant tissue damage…”
BackgroundInfluentialRestoring sensorimotor function through intracortical interfaces: progress and looming challenges
Bensmaia, S. J., & Miller, L. E. (2014) · Nature Reviews Neuroscience · doi:10.1038/nrn3724 · Cited by 368
Akhavan, O., & Ghaderi, E. (2013) · Nanoscale · doi:10.1039/c3nr02161k · Cited by 238
Lewis, P. M., Ackland, H. M., Lowery, A. J., & Rosenfeld, J. V. (2015) · Brain Research · doi:10.1016/j.brainres.2014.11.020 · Cited by 228
Heo, C., Yoo, J., Lee, S., Jo, A., Jung, S., Yoo, H., Lee, Y. H., & Suh, M. (2011) · Biomaterials · doi:10.1016/j.biomaterials.2010.08.095 · Cited by 214
Nolta, N. F., Christensen, M. B., Crane, P. D., Skousen, J. L., & Tresco, P. A. (2015) · Biomaterials · doi:10.1016/j.biomaterials.2015.02.081 · Cited by 175
Implantable microcoils for intracortical magnetic stimulation
Lee, S. W., Fallegger, F., Casse, B. D. F., & Fried, S. I. (2016) · Science Advances · doi:10.1126/sciadv.1600889 · Cited by 151
“2E, magnetic stimulation from these coils can modulate activity well beyond the region over which encapsulation and cell migration are expected to occur (11, 22, 34, 35), thereby suggesting that these coils can remain viable during chronic implantation.”
Kolarcik, C. L., Catt, K., Rost, E., Albrecht, I. N., Bourbeau, D., Du, Z., Kozai, T. D. Y., Luo, X., Weber, D. J., & Tracy Cui, X. (2015) · Journal of Neural Engineering · doi:10.1088/1741-2560/12/1/016008 · Cited by 126
“The stimulation protocol employed was modeled after those utilized for intracortical microstimulation of visual prosthetic devices [45] and for sensorimotor cortex stimulation [43] and differs from that used in prior studies [42].”
MethodResultInfluentialArtificial vision through neuronal stimulation
Fernandes, R. A. B., Diniz, B., Ribeiro, R., & Humayun, M. (2012) · Neuroscience Letters · doi:10.1016/j.neulet.2012.01.063 · Cited by 90
Torab, K., Davis, T. S., Warren, D. J., House, P. A., Normann, R. A., & Greger, B. (2011) · Journal of Neural Engineering · doi:10.1088/1741-2560/8/3/035001 · Cited by 90
Selectivity and Longevity of Peripheral-Nerve and Machine Interfaces: A Review.
Ghafoor, U., Kim, S., & Hong, K.-S. (2017) · Frontiers in Neurorobotics · doi:10.3389/fnbot.2017.00059 · Cited by 88
“The fundamental problem in cortical interfaces is longevity (Warren et al., 2016): The neurons (or neuronal tissues) at the implanted surface and the electrodes (implanted interface) degrade over time (McCreery et al., 2010; Prasad et al., 2012; Kane et al., 2013; Chen et al., 2014).”
Restoring tactile and proprioceptive sensation through a brain interface
Tabot, G. A., Kim, S. S., Winberry, J. E., & Bensmaia, S. J. (2015) · Neurobiology of Disease · doi:10.1016/j.nbd.2014.08.029 · Cited by 83
The effects of chronic intracortical microstimulation on neural tissue and fine motor behavior
Rajan, A. T., Boback, J. L., Dammann, J. F., Tenore, F. V., Wester, B. A., Otto, K. J., Gaunt, R. A., & Bensmaia, S. J. (2015) · Journal of Neural Engineering · doi:10.1088/1741-2560/12/6/066018 · Cited by 76
Long-term recording reliability of liquid crystal polymer µ ECoG arrays
Woods, V., Trumpis, M., Bent, B., Palopoli-Trojani, K., Chiang, C.-H., Wang, C., Yu, C., Insanally, M. N., Froemke, R. C., & Viventi, J. (2018) · Journal of Neural Engineering · doi:10.1088/1741-2552/aae39d · Cited by 63
The effect of chronic intracortical microstimulation on the electrode–tissue interface
Chen, K. H., Dammann, J. F., Boback, J. L., Tenore, F. V., Otto, K. J., Gaunt, R. A., & Bensmaia, S. J. (2014) · Journal of Neural Engineering · doi:10.1088/1741-2560/11/2/026004 · Cited by 59
“Charge amplitudes—20 nC/phase (100 μA), 4 nC/phase (20 μA), and 2 nC/phase (10 μA)—were selected to span a range of amplitudes that have been shown to elicit a wide range of sensations [10, 13] and to test some previously established parameters [23].”
BackgroundRosenfeld, J. V., Wong, Y. T., Yan, E., Szlawski, J., Mohan, A., Clark, J. C., Rosa, M., & Lowery, A. (2020) · Journal of Neural Engineering · doi:10.1088/1741-2552/ab9e1c · Cited by 35
Approaches to a cortical vision prosthesis: implications of electrode size and placement
Christie, B. P., Ashmont, K. R., House, P. A., & Greger, B. (2016) · Journal of Neural Engineering · doi:10.1088/1741-2560/13/2/025003 · Cited by 26
“prosthesis is time-variable, the majority of the neuronal loss could be attributed to the mechanical injury resulting from electrode insertion [48].”
BackgroundMethodNeurophysiological considerations for visual implants
Meikle, S. J., & Wong, Y. T. (2022) · Brain Structure and Function · doi:10.1007/s00429-021-02417-2 · Cited by 15
Local field potential phase modulates neural responses to intracortical electrical stimulation
Allison-Walker, T. J., Ann Hagan, M., Chiang Price, N. S., & Tat Wong, Y. (2020) · 2020 42nd Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC) · doi:10.1109/embc44109.2020.9176186 · Cited by 9
Metadata
- Authors
- McCreery, D., Pikov, V., & Troyk, P. R.
- Venue
- Journal of Neural Engineering
- Publisher
- IOP Publishing
- Published
- 2010-05-11
- Volume
- 7
- Issue
- 3
- Pages
- 036005
- ISSN
- 1741-2552 (ISSN-L)
- 1741-2560
- OpenAlex
- W2091951519
- Cited by
- 137
- References
- 14
- FWCI
- 3.28
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.