Lancet Commission on Hypertension group position statement on the global improvement of accuracy standards for devices that measure blood pressure
Open accessThe DOI matches its Crossref record; 57 of the 58 references it cites were found in a registry or an index.
Abstract
: The Lancet Commission on Hypertension identified that a key action to address the worldwide burden of high blood pressure (BP) was to improve the quality of BP measurements by using BP devices that have been validated for accuracy. Currently, there are over 3000 commercially available BP devices, but many do not have published data on accuracy testing according to established scientific standards. This problem is enabled through weak or absent regulations that allow clearance of devices for commercial use without formal validation. In addition, new BP technologies have emerged (e.g. cuffless sensors) for which there is no scientific consensus regarding BP measurement accuracy standards. Altogether, these issues contribute to the widespread availability of clinic and home BP devices with limited or uncertain accuracy, leading to inappropriate hypertension diagnosis, management and drug treatment on a global scale. The most significant problems relating to the accuracy of BP devices can be resolved by the regulatory requirement for mandatory independent validation of BP devices according to the universally-accepted International Organisation for Standardization Standard. This is a primary recommendation for which there is an urgent international need. Other key recommendations are development of validation standards specifically for new BP technologies and online lists of accurate devices that are accessible to consumers and health professionals. Recommendations are aligned with WHO policies on medical devices and universal healthcare. Adherence to recommendations would increase the global availability of accurate BP devices and result in better diagnosis and treatment of hypertension, thus decreasing the worldwide burden from high BP.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“hundreds of companies manufacturing more than 3500 different models of automated BPMDs, many of which are without evidence of having passed validation testing [25–29].”
From Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure (2022) · cites this paper as [29]
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“The lack of standardization in BP measurement is a serious and frequent problem in contemporary clinical practice that impacts patient safety and health system performance.8”
From International Consensus on Standardized Clinic Blood Pressure Measurement – A Call to Action (2023) · cites this paper as [8]
The cited source says what the claim attributes to it.
In the citing paper
“To date, no regulatory requirements exist for independent validation of all BP devices that are put on the market (using the universally accepted International Organization for Standardization (ISO) standard), despite increasing calls to mandate validation of these devices16.”
From Blood pressure and its variability: classic and novel measurement techniques (2022) · cites this paper as [16]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Pickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., & Roccella, E. J. (2005) · Circulation · doi:10.1161/01.cir.0000154900.76284.f6 · Cited by 3,985
Ettehad, D., Emdin, C. A., Kiran, A., Anderson, S. G., Callender, T., Emberson, J., Chalmers, J., Rodgers, A., & Rahimi, K. (2016) · The Lancet · doi:10.1016/s0140-6736(15)01225-8 · Cited by 3,755 · cited 1×
“Reassuringly, if the presence of high BP can be correctly identified with appropriate use of accurate BP devices and BP measurement protocols, the risk of future cardiovascular events can be reduced significantly with BP-lowering medications [4], dietary and lifestyle interventions [5].”
BackgroundSelf-citationWilliams, B., Mancia, G., Spiering, W., Agabiti Rosei, E., Azizi, M., Burnier, M., Clement, D. L., Coca, A., de Simone, G., Dominiczak, A., et al. (2018) · Journal of Hypertension · doi:10.1097/hjh.0000000000001940 · Cited by 2,885
Self-citationGakidou, E., Afshin, A., Abajobir, A. A., Abate, K. H., Abbafati, C., Abbas, K. M., Abd-Allah, F., Abdulle, A. M., Abera, S. F., Aboyans, V., et al. (2017) · The Lancet · doi:10.1016/s0140-6736(17)32366-8 · Cited by 2,406 · cited 1×
“H igh blood pressure (BP) is the leading modifiable risk factor for cardiovascular disease, contributing to the greatest global burden of disease [1,2].”
BackgroundPickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., & Roccella, E. J. (2005) · Hypertension · doi:10.1161/01.hyp.0000150859.47929.8e · Cited by 2,051 · cited 2×
“Thus, accurate BP measurement has been cited as one of the most important tests in clinical medicine [10].”
BackgroundChow, C. K. (2013) · JAMA · doi:10.1001/jama.2013.184182 · Cited by 1,991 · cited 1×
“There are major deficiencies with respect to optimal awareness, diagnosis and treatment of high BP, and these problems persist across low-income, middle-income and highincome countries, altogether emphasizing the need for widespread population-level improvement [3].”
BackgroundSelf-citationWhelton, P. K. (2002) · JAMA · doi:10.1001/jama.288.15.1882 · Cited by 1,473 · cited 1×
“Reassuringly, if the presence of high BP can be correctly identified with appropriate use of accurate BP devices and BP measurement protocols, the risk of future cardiovascular events can be reduced significantly with BP-lowering medications [4], dietary and lifestyle interventions [5].”
BackgroundVlachopoulos, C., Aznaouridis, K., O'Rourke, M. F., Safar, M. E., Baou, K., & Stefanadis, C. (2010) · European Heart Journal · doi:10.1093/eurheartj/ehq024 · Cited by 1,363
Olsen, M. H., Angell, S. Y., Asma, S., Boutouyrie, P., Burger, D., Chirinos, J. A., Damasceno, A., Delles, C., Gimenez-Roqueplo, A.-P., Hering, D., et al. (2016) · The Lancet · doi:10.1016/s0140-6736(16)31134-5 · Cited by 939 · cited 1×
“One of the key actions of The Lancet Commission on Hypertension was to improve BP evaluation from better quality of BP measurements through endorsed protocols and validated (accurate) BP monitors [11].”
BackgroundSelf-citationToward ubiquitous blood pressure monitoring via pulse transit time: Theory and practice
Mukkamala, R., Hahn, J.-O., Inan, O. T., Mestha, L. K., Kim, C.-S., Toreyin, H., & Kyal, S. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2441951 · Cited by 935
Monitoring of the central blood pressure waveform via a conformal ultrasonic device
Wang, C., Li, X., Hu, H., Zhang, L., Huang, Z., Lin, M., Zhang, Z., Yin, Z., Huang, B., Gong, H., et al. (2018) · Nature Biomedical Engineering · doi:10.1038/s41551-018-0287-x · Cited by 895
Central Pulse Pressure and Mortality in End-Stage Renal Disease
Safar, M. E., Blacher, J., Pannier, B., Guerin, A. P., Marchais, S. J., Guyonvarc’h, P.-M., & London, G. M. (2002) · Hypertension · doi:10.1161/hy0202.098325 · Cited by 805
O??Brien, E., Petrie, J., Littler, W., de Swiet, M., Padfield, P. L., O??Malley, K., Jamieson, M., Altman, D., Bland, M., & Atkins, N. (1990) · Journal of Hypertension · doi:10.1097/00004872-199007000-00004 · Cited by 772 · cited 1×
“In 1987, the US Association for the advancement of medical instrumentation (AAMI) developed a clinical validation protocol for BP devices [28,29], which was followed by a similar protocol from the British Hypertension Society [30,31].”
MethodSelf-citationOʼBrien, E., Pickering, T., Asmar, R., Myers, M., Parati, G., Staessen, J., Mengden, T., Imai, Y., Waeber, B., Palatini, P., et al. (2002) · Blood Pressure Monitoring · doi:10.1097/00126097-200202000-00002 · Cited by 744 · cited 1×
“In 1999, the German Hypertension League developed another protocol [32] and in 2002 the European Society of Hypertension (ESH) Working Group on BP Monitoring developed their International Protocol, which they revised in 2010 [33,34].”
BackgroundSelf-citationO'Brien, E., Atkins, N., Stergiou, G., Karpettas, N., Parati, G., Asmar, R., Imai, Y., Wang, J., Mengden, T., & Shennan, A. (2010) · Blood Pressure Monitoring · doi:10.1097/mbp.0b013e3283360e98 · Cited by 610 · cited 2×
“These protocols have been revised by adopting more stringent criteria [29,31,34].”
BackgroundMethodSelf-citationO??Brien, E., Petrie, J., Little, W., de Swiet, M., Padfield, P. L., Altma, D. G., Bland, M., Coats, A., & Atkins, N. (1993) · Journal of Hypertension · doi:10.1097/00004872-199306000-00013 · Cited by 519
Self-citationIncreased Absenteeism from Work after Detection and Labeling of Hypertensive Patients
Haynes, R. B., Sackett, D. L., Taylor, D. W., Gibson, E. S., & Johnson, A. L. (1978) · New England Journal of Medicine · doi:10.1056/nejm197810052991403 · Cited by 510 · cited 1×
“An overestimation of BP based on inaccurate measurement could lead to the initiation, and potential lifelong continuation, of unnecessary medications with possible side effects as well as unwanted social effects including anxiety, workplace absenteeism [7] and increased costs from insurance and medications [8].”
BackgroundStergiou, G. S., Alpert, B., Mieke, S., Asmar, R., Atkins, N., Eckert, S., Frick, G., Friedman, B., Graßl, T., Ichikawa, T., et al. (2018) · Hypertension · doi:10.1161/hypertensionaha.117.10237 · Cited by 469
Self-citationKELLY, R. P., GIBS, H. H., O'ROURKE, M. F., DALEY, J. E., MANG, K., MORGEN, J. J., & AVOLID, A. P. (1990) · European Heart Journal · doi:10.1093/oxfordjournals.eurheartj.a059669 · Cited by 312
Blood pressure and the global burden of disease 2000. Part II: Estimates of attributable burden
Lawes, C. M., Hoorn, S. V., Law, M. R., Elliott, P., MacMahon, S., & Rodgers, A. (2006) · Journal of Hypertension · doi:10.1097/01.hjh.0000209973.67746.f0 · Cited by 296
Accuracy of Cuff-Measured Blood Pressure
Picone, D. S., Schultz, M. G., Otahal, P., Aakhus, S., Al-Jumaily, A. M., Black, J. A., Bos, W. J., Chambers, J. B., Chen, C.-H., Cheng, H.-M., et al. (2017) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2017.05.064 · Cited by 277
Self-citationA Short History of Blood Pressure Measurement
Booth, J. (1977) · Proceedings of the Royal Society of Medicine · doi:10.1177/003591577707001112 · Cited by 268 · cited 1×
“These methods rely on analysis of haemodynamic signals in the brachial artery, which were originally believed to be a good estimate of the central aortic pressure load [48] as the most clinically relevant BP [49–53].”
MethodMeasuring Blood Pressure Accurately: New and Persistent Challenges
Jones, D. W., Appel, L. J., Sheps, S. G., Roccella, E. J., & Lenfant, C. (2003) · JAMA · doi:10.1001/jama.289.8.1027 · Cited by 264 · cited 1×
“These are nontrivial problems, in that relatively small systematic inaccuracies can lead to misclassification of many millions of people at the population level [9].”
BackgroundRecommendations for Human Blood Pressure Determinations by Sphygmomanometers
BORDLEY, J., CONNOR, C. A. R., HAMILTON, W. F., KERR, W. J., & WIGGERS, C. J. (1951) · Circulation · doi:10.1161/01.cir.4.4.503 · Cited by 262 · cited 1×
“Many important articles on the deficiencies of BP measurement have been published over many decades [10,46], but the information is rarely disseminated beyond the researchers and health professionals who are already familiar with the problems.”
BackgroundSharman, J. E., Avolio, A. P., Baulmann, J., Benetos, A., Blacher, J., Blizzard, C. L., Boutouyrie, P., Chen, C.-H., Chowienczyk, P., Cockcroft, J. R., et al. (2017) · European Heart Journal · doi:10.1093/eurheartj/ehw632 · Cited by 231
Self-citationSmartphone-based blood pressure monitoring via the oscillometric finger-pressing method
Chandrasekhar, A., Kim, C.-S., Naji, M., Natarajan, K., Hahn, J.-O., & Mukkamala, R. (2018) · Science Translational Medicine · doi:10.1126/scitranslmed.aap8674 · Cited by 212
A content analysis of smartphone–based applications for hypertension management
Kumar, N., Khunger, M., Gupta, A., & Garg, N. (2015) · Journal of the American Society of Hypertension · doi:10.1016/j.jash.2014.12.001 · Cited by 208
FDA Regulation of Mobile Medical Apps
Shuren, J., Patel, B., & Gottlieb, S. (2018) · JAMA · doi:10.1001/jama.2018.8832 · Cited by 155 · cited 2×
“Recommendation 2: Development of specific standards for the validation of new BP measuring technologies that cannot be tested using the ISO Standard (ISO 810602:2018) The worldwide inundation of new technology that claims to measure BP brings challenges to regulatory authorities that are also encouraging ways to find…”
BackgroundBilo, G., Zorzi, C., Ochoa Munera, J. E., Torlasco, C., Giuli, V., & Parati, G. (2015) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000124 · Cited by 143
Self-citationNair, D., Tan, S.-Y., Gan, H. ‐., Lim, S.-F., Tan, J., Zhu, M., Gao, H., Chua, N.-H., Peh, W. L., & Mak, K.-H. (2007) · Journal of Human Hypertension · doi:10.1038/sj.jhh.1002306 · Cited by 112
Accurate blood pressure measurement: why does it matter?
Campbell, N. R. C., & McKay, D. W. (1999) · Europe PMC (PubMed Central) · Cited by 104
Self-citationValidation protocols for blood pressure measuring devices in the 21st century
Stergiou, G. S., Alpert, B. S., Mieke, S., Wang, J., & O'Brien, E. (2018) · The Journal of Clinical Hypertension · doi:10.1111/jch.13294 · Cited by 92
Self-citationNew Approaches to Evaluating and Monitoring Blood Pressure
Goldberg, E. M., & Levy, P. D. (2016) · Current Hypertension Reports · doi:10.1007/s11906-016-0650-9 · Cited by 87
Danchin, N. (2004) · American Journal of Hypertension · doi:10.1016/j.amjhyper.2003.09.010 · Cited by 80
An Assessment of the Accuracy of Home Blood Pressure Monitors When Used in Device Owners
Ringrose, J. S., Polley, G., McLean, D., Thompson, A., Morales, F., & Padwal, R. (2017) · American Journal of Hypertension · doi:10.1093/ajh/hpx041 · Cited by 64
Self-citationReliability of home blood pressure monitoring: in the context of validation and accuracy
Jung, M.-H., Kim, G.-H., Kim, J.-H., Moon, K.-W., Yoo, K.-D., Rho, T.-H., & Kim, C.-M. (2015) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000121 · Cited by 55
Stergiou, G. S., Asmar, R., Myers, M., Palatini, P., Parati, G., Shennan, A., Wang, J., & O’Brien, E. (2018) · Journal of Hypertension · doi:10.1097/hjh.0000000000001635 · Cited by 55 · cited 2×
“Indeed, endorsement of a standardized validation protocol by a hypertension society has been shown to influence manufacturers to pursue independent validation of BP devices [39].”
BackgroundSelf-citationNon-invasive Estimation of Aortic Blood Pressures: A Close Look at Current Devices and Methods
Millasseau, S., & Agnoletti, D. (2014) · Current Pharmaceutical Design · doi:10.2174/1381612820666141023163748 · Cited by 50
Butlin, M., Shirbani, F., Barin, E. S., Tan, I., Spronck, B., & Avolio, A. P. (2018) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2018.2823333 · Cited by 47
Home sphygmomanometers: validation versus accuracy
Akpolat, T., Dilek, M., Aydogdu, T., Adibelli, Z., Erdem, D. G., & Erdem, E. (2009) · Blood Pressure Monitoring · doi:10.1097/mbp.0b013e3283262f31 · Cited by 46
Campbell, N. R. C., Gelfer, M., Stergiou, G. S., Alpert, B. S., Myers, M. G., Rakotz, M. K., Padwal, R., Schutte, A. E., O'Brien, E., Lackland, D. T., et al. (2016) · The Journal of Clinical Hypertension · doi:10.1111/jch.12782 · Cited by 45
Self-citationMeasurement of Blood Pressure Using an Arterial Pulsimeter Equipped with a Hall Device
Lee, S., Nam, D., Hong, Y.-S., Lee, W.-B., Son, I.-H., Kim, K.-H., & Choi, J. (2011) · Sensors · doi:10.3390/s110201784 · Cited by 44
The pursuit of accurate blood pressure measurement: A 35‐year travail
O'Brien, E., & Stergiou, G. S. (2017) · The Journal of Clinical Hypertension · doi:10.1111/jch.13005 · Cited by 39 · cited 1×
“Efforts to validate BP devices formally started in the 1980s [27].”
BackgroundSelf-citationRaichle, C. J., Eckstein, J., Lapaire, O., Leonardi, L., Brasier, N., Vischer, A. S., & Burkard, T. (2018) · Hypertension · doi:10.1161/hypertensionaha.117.10647 · Cited by 39
Guidelines for blood pressure measurement: development over 30 years
Stergiou, G. S., Parati, G., McManus, R. J., Head, G. A., Myers, M. G., & Whelton, P. K. (2018) · Journal of Clinical Hypertension · doi:10.1111/jch.13295 · Cited by 35
Self-citationSharman, J. E., & Marwick, T. H. (2018) · Journal of Human Hypertension · doi:10.1038/s41371-018-0122-6 · Cited by 34 · cited 1×
“The effects of unknown BP errors can influence the clinical environment, epidemiology and research, and even contribute to discrepancies among hypertension guidelines [20].”
BackgroundSelf-citationTholl, U., Lüders, S., Bramlage, P., Dechend, R., Eckert, S., Mengden, T., Nürnberger, J., Sanner, B., & Anlauf, M. (2016) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000186 · Cited by 33 · cited 1×
“In 1999, the German Hypertension League developed another protocol [32] and in 2002 the European Society of Hypertension (ESH) Working Group on BP Monitoring developed their International Protocol, which they revised in 2010 [33,34].”
BackgroundKim, H., Seo, J.-B., Chung, W., Kim, S., Kim, M.-A., & Zo, J. (2014) · American Journal of Hypertension · doi:10.1093/ajh/hpu146 · Cited by 31
Can ‘FDA-cleared’ blood pressure devices be trusted? A call to action
Alpert, B. S. (2017) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000279 · Cited by 31
Self-citationSelf‐measurement of blood pressure at home: Is it reliable?
Dilek, M., Adıbelli, Z., Aydogdu, T., Köksal, A. R., Çakar, B., & Akpolat, T. (2008) · Blood Pressure · doi:10.1080/08037050701758018 · Cited by 24
Achieving reliable blood pressure measurements in clinical practice: It's time to meet the challenge
O'Brien, E., Dolan, E., & Stergiou, G. S. (2018) · The Journal of Clinical Hypertension · doi:10.1111/jch.13323 · Cited by 23 · cited 2×
“This can be achieved through online lists of BP devices that detail the results of validation studies carried out according to international standards [43].”
BackgroundSelf-citationAccurate blood pressure measuring devices: Influencing users in the 21st century
O'Brien, E. T., Alpert, B. S., & Stergiou, G. S. (2018) · Journal of Clinical Hypertension · doi:10.1111/jch.13278 · Cited by 20
Self-citationThe quest for accuracy of blood pressure measuring devices
O'Brien, E., Stergiou, G. S., & Turner, M. J. (2018) · The Journal of Clinical Hypertension · doi:10.1111/jch.13279 · Cited by 18 · cited 2×
“Regulations can also be difficult to enforce due to ambiguities and high levels of complexity [15].”
BackgroundSelf-citationAccuracy of Sphygmanometers at Pharmacies
Erdem, D., Erdem, E., Dilek, M., Aydogdu, T., Selim, N., Demiray, F., Sarı, A., & Akpolat, T. (2009) · Kidney & Blood Pressure Research · doi:10.1159/000228936 · Cited by 16
Hodgkinson, J., Koshiaris, C., Martin, U., Mant, J., Heneghan, C. J., Hobbs, R. F., & McManus, R. J. (2016) · British Journal of General Practice · doi:10.3399/bjgp16x684769 · Cited by 11
Self-citationGiles, T. D., Kostis, J. B., & Fernandez, C. (2018) · Journal of Clinical Hypertension · doi:10.1111/jch.13286 · Cited by 6
Reporting of blood pressure monitor validation studies
McManus, R. J., Lacy, P. S., Clark, C., Chapman, N., & Lewis, P. (2018) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000334 · Cited by 5
Self-citation
Research works printed but not linked: 1
- AMA. AHA developing list of validated BP monitors. https://wire.ama-assn.org/delivering-care/ama-aha-developing-list-validated-bp-monitors [Accessed 15 August 2018].
Cited by
Mancia, G., Kreutz, R., Brunström, M., Burnier, M., Grassi, G., Januszewicz, A., Muiesan, M. L., Tsioufis, K., Agabiti-Rosei, E., Algharably, E. A. E., et al. (2023) · Journal of Hypertension · doi:10.1097/hjh.0000000000003480 · Cited by 3,473
Self-citation2024 ESC Guidelines for the management of elevated blood pressure and hypertension
McEvoy, J. W., McCarthy, C. P., Bruno, R. M., Brouwers, S., Canavan, M. D., Ceconi, C., Christodorescu, R. M., Daskalopoulou, S. S., Ferro, C. J., Gerdts, E., et al. (2024) · European Heart Journal · doi:10.1093/eurheartj/ehae178 · Cited by 2,298
Self-citationStergiou, G. S., Palatini, P., Parati, G., O’Brien, E., Januszewicz, A., Lurbe, E., Persu, A., Mancia, G., Kreutz, R., & on behalf of the European Society of Hypertension Council and the European Society of Hypertension Working Group on Blood Pressure Monitoring and Cardiovascular Variability (2021) · Journal of Hypertension · doi:10.1097/hjh.0000000000002843 · Cited by 859
Self-citationHypertension in low-and middle-income countries
Schutte, A. E., Srinivasapura Venkateshmurthy, N., Mohan, S., & Prabhakaran, D. (2021) · Circulation Research · doi:10.1161/circresaha.120.318729 · Cited by 460
Self-citationBlood pressure and its variability: classic and novel measurement techniques
Schutte, A. E., Kollias, A., & Stergiou, G. S. (2022) · Nature Reviews Cardiology · doi:10.1038/s41569-022-00690-0 · Cited by 335
Self-citation
Metadata
- Authors
- Sharman, J. E., O’Brien, E., Alpert, B., Schutte, A. E., Delles, C., Hecht Olsen, M., Asmar, R., Atkins, N., Barbosa, E., Calhoun, D., Campbell, N. R. C., Chalmers, J., Benjamin, I., Jennings, G., Laurent, S., Boutouyrie, P., Lopez-Jaramillo, P., McManus, R. J., Mihailidou, A. S., Ordunez, P., Padwal, R., Palatini, P., Parati, G., Poulter, N., Rakotz, M. K., Rosendorff, C., Saladini, F., Scuteri, A., Sebba Barroso, W., Cho, M.-C., Sung, K.-C., Townsend, R. R., Wang, J.-G., Willum Hansen, T., Wozniak, G., & Stergiou, G.
- Affiliations
- University College Dublin (IE)
- University of Tennessee Health Science Center (US)
- North-West University (ZA)
- University of Glasgow (GB)
- University of Alabama at Birmingham (US)
- University of Calgary (CA)
- UNSW Sydney (AU)
- American Heart Association (US)
- The University of Sydney (AU)
- Inserm (FR)
- University of Oxford (GB)
- Royal North Shore Hospital (AU)
- University of Alberta (CA)
- University of Padua (IT)
- IRCCS Istituto Auxologico Italiano (IT)
- Imperial College London (GB)
- American Medical Association (US)
- Holbæk Sygehus (DK)
- University of Southern Denmark (DK)
- Corresponding author
- Alpert, Bruce
- ORCID
- Sharman, James E.
- O’Brien, Eoin
- Alpert, Bruce
- Schutte, Aletta E.
- Delles, Christian
- Asmar, Roland
- Atkins, Neil
- Barbosa, Eduardo
- Calhoun, David
- Campbell, Norm R.C.
- Chalmers, John
- Benjamin, Ivor
- Jennings, Garry
- Laurent, Stéphane
- Boutouyrie, Pierre
- McManus, Richard J.
- Mihailidou, Anastasia S.
- Padwal, Raj
- Palatini, Paolo
- Parati, Gianfranco
- Poulter, Neil
- Rakotz, Michael K.
- Rosendorff, Clive
- Saladini, Francesca
- Scuteri, Angelo
- Cho, Myeong-Chan
- Sung, Ki-Chul
- Townsend, Raymond R.
- Wang, Ji-Guang
- Wozniak, Gregory
- Stergiou, George
- Venue
- Journal of Hypertension
- Published
- 2020-01
- Volume
- 38
- Issue
- 1
- Pages
- 21-29
- ISSN
- 0263-6352 (ISSN-L)
- PMC
- PMC6919228
- OpenAlex
- W2973007128
- Semantic Scholar
- 2413d689ae41fec25f61910c94f550039881bba8
- Cited by
- 139
- References
- 58
- Citations by year
- 2026: 13 · 2025: 17 · 2024: 13 · 2023: 16 · 2022: 37 · 2021: 20 · 2020: 22
- FWCI
- 14.45
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by-nc-nd
- Locations
- Journal of Hypertension (publishedVersion, cc-by-nc-nd)
- PubMed Central (publishedVersion, cc-by-nc-nd)
- PubMed (publishedVersion)
- UTAS Research Repository (acceptedVersion)
- UNSWorks (UNSW Sydney) (submittedVersion, cc-by-nc-nd)
- UNSWorks (University of New South Wales, Sydney, Australia) (submittedVersion, cc-by-nc-nd)
- University of Southern Denmark Research Portal (University of Southern Denmark) (submittedVersion, cc-by-nc-nd)
- LA Referencia (Red Federada de Repositorios Institucionales de Publicaciones Científicas) (submittedVersion, cc-by-nc-nd)
- Boloka Institutional Repository (North-west University) (submittedVersion)
- UNICA IRIS Institutional Research Information System (University of Cagliari) (submittedVersion, other-oa)
- UnissResearch (Università degli Studi di Sassari) (submittedVersion)
- Spiral (Imperial College London) (submittedVersion, other-oa)
- Funders
- American Heart Association
- Hjerteforeningen
- Steno Diabetes Center Copenhagen (SDCC 3.A Complications)
- National Institute for Health and Care Research (NIHR200201, NIHR-RP-02-12-015)
- National Heart Foundation of Australia
- High Blood Pressure Research Council of Australia
- Société Française d’Hypertension Artérielle
- Hypertension Canada
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Clinical practice guidelines implementation
- Keywords
- blood pressure measurement accuracy, hypertension diagnosis, device validation, medical device regulation, home blood pressure monitoring, antihypertensive therapy
- MeSH terms
- Blood Pressure Determination, Humans, Hypertension, Practice Guidelines as Topic, Sphygmomanometers
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.